TEST REPORT IEC/EN 60335 & 60601 Household And Similar .

2y ago
337 Views
39 Downloads
2.83 MB
44 Pages
Last View : 22d ago
Last Download : 3m ago
Upload by : Camryn Boren
Transcription

Test Report issued under the responsibility of:NEVON SOLUTIONS PVT LTDTEST REPORTIEC/EN60335 & 60601Household and Similar Electrical AppliancesSafety and Effectiveness of Medical ElectricalEquipmentReport Reference No. .:45122 164Date of issue.:20.08.2020Total number of pages.:44Testing Authority.:Nevon Solutions Product Safety LabAddress.:C4, Laghu Udyog, IB Patel Rd, Goregaon East, Mumbai 063Test specification:Standard .:IEC 60335-1IEC 60601-1Test procedure.:As per Respective IEC Standard Page 3Receipt of Samples .:04-07-2020 (4 July 2020)Date of Test :17-07-2020 (17 July 2020)Test Result/Conclusion :PASS - See Results & Conclusion Page 5Test Sample Description:Product .:Ultraviolet Sterilizers Box/ChamberBrand/Make Name :NevonModel No :C72Manufacturer Name .:NevonSolutions Pvt LtdRatings : AC 230V 50/60 HzPossible test case verdicts:- test case does not apply to the test object . : N/A- test object does meet the requirement. : P(Pass)- test object does not meet the requirement. : F(Fail)

ContentsPage NumberSection 1 Test Results & Conclusion with Justification . . 3Section 2 Test Environment & Conditions . . .4Section 3 EUT Information (Equipment Under Test) . . 5Section 4 Test Configuration & Test Report . . .6Section 6 Appendix 1 . . 25Photographs of Test ConfigurationsSection 7 Appendix 2 . . 35Photographs of EUT

SECTION 1 SUMMARY OF TESTINGSummary of testing: PASSTests performed (name of test and test clause):Testing location:All itemsNevonSolutions Pvt Ltd, C4, LaghuUdyog, IB Patel Rd, Goregaon East,Mumbai 063Summary of compliance:We tested UV Sterilizer devices model No C72 to determine weather it is in compliance with relevant IECstandards IEC 60335-1: 2010 and IEC 60601-1. We found that the product met the below testingrequirements when tested as received.The UV-C Sterilizers were tested against the current standards:IEC 60335-1: 2010IEC 60601-1Model references and rated inputsC72 – 88W 230VCopy of Marking Plate

SECTION 2 TEST ENVIRONMENT & CONDITIONSClauseRequirementTest SiteTest Date4Suitable Devices for Measuring Voltage, CurrentPower & FrequencyNevon Product TestingLab , Mumbai17-07-20207Spirit & WaterNevon Product TestingLab , Mumbai17-07-2020Nevon Product TestingLab , Mumbai17-07-20208Test Probe Setup For Protection Against LiveParts10Voltmeter Ammeter For Voltage, CurrentMeasurement & Wattage CalculationNevon Product TestingLab , Mumbai17-07-202023Internal Wiring Flexing Test ApparatusNevon Product TestingLab , Mumbai17-07-202027Earthing Continuity Testing MeterNevon Product TestingLab , Mumbai17-07-20208Voltmeter & Ammeter Setup For Input & OutputVoltage Current & Leakage Current DetectionNevon Product TestingLab , Mumbai17-07-2020254nm UV-C Radiation Measurement InstrumentNevon Product TestingLab , Mumbai17-07-2020UVC Radiation Leakage Test InstrumentNevon Product TestingLab , Mumbai19-08-2020

SECTION 3 EUT INFORMATION (Equipment Under Test)Equipment Under Test Nevon UVLizer C72 – Ultraviolet Sterilizer(EUT):Product Type: Electronic DeviceModel No/Variant: C72Dimensions: 585x302x286mmWeight: 15kgOperating Voltage: 230V ACTotal Wattage: 88WProduct Material: Mild Steel Powder CoatedCountry Of Manufacture: IndiaProduct Color: White

SECTION 3 TEST CONFIGURATION & TEST REPORTClauseTestResult/RemarkVerdict4.11POWER INPUTInput Voltage TestingDevice Working at 230V ACPInput Current Testing0.769 AmperesPCurrent (A)Calculated WattageN/A176 WattP7MARKINGS & INSTRUCTIONS7.1Rated Voltage of Voltage Range V230VPFrequency Hz50HzPRated current (A)N/AProduct NameUVLizerPModel NumberC72PManufacturer/Company NameNevon Solutions Pvt LtdPCE RoHS MarkingsCE RoHSPElectronic Waste MarkingProvidedPCountry or ManufactureIndiaP7.1.3Durability Of MarkingsResistant to Water & EthylAlcoholP7.6Correct Symbols Used7.9Marking or placing of switches which may causea hazardNo Such SwitchesN/A7.10Indications of switches on stationary appliancesand controls on all appliances by use of figures,letters or other visual meansProvided on DisplayPPThe figure 0 indicates only OFF position, unlessno confusion with the OFF positionN/A7.11Indication for direction of adjustment of controlsN/A7.12Instructions for safe use providedProvided on User ManualP7.12.1Sufficient details for installation suppliedProvided on User ManualPSufficient details for Product Working SuppliedProvided on User ManualP

UV Safety Warning Instructions ProvidedProvided on ProductP8PROTECTION AGAINST ACCESS TO LIVE PARTS8.1Adequate protection against accidental contactwith live partsP8.1.1Requirement applies for all positions, detachableNo Detachable Partsparts removedLive parts protected at least by basic insulation before installation or assembly:N/A- built-in appliancesP- fixed appliancesP- appliances delivered in separate unitsN/AClass II appliances and constructions constructedso that there is adequate protection againstaccidental contact with basic insulation and metalparts separated from live parts by basic insulationonlyOnly possible to touch parts separated from liveparts by double or reinforced insulationP8.1.58.210POWER INPUT AND CURRENT10.1Power input at normal operating temperature,rated voltage and normal operation not deviatingfrom rated power input by more than shown intable 1Test for an appliance with one or more ratedvoltage ranges (IEC/EN 60335-1/A2)10.2Current at normal operating temperature, ratedvoltage and normal operation not deviating fromrated current by more than shown in table 2PPOnly one voltageN/A(See appended table)PTest for an appliance with one or more ratedvoltage ranges (IEC/EN 60335-1/A2)11N/AHEATING11.1No excessive temperatures in normal useP13LEAKAGE CURRENT AND ELECTRIC STRENGTH AT OPERATING TEMPERATURE13.1Leakage current not excessive and electricstrength adequateSee belowPHeating appliances operated at 1.15 times ratedpowerinput. :N/AMotor-operated appliances and combinedappliances supplied at 1,06 times ratedvoltage. :N/A

13.2Protective impedance and radio interferencefilters disconnected before carrying out thetestsN/ALeakage current measured by means of thecircuit described in figure 4 of IEC 60990N/ALeakage current measurements13.3(See appended table)The appliance is disconnected from the supplyElectric strength tests according to table 4P(See appended table)No breakdown during the tests1616.1N/APLEAKAGE CURRENT AND ELECTRIC STRENGTHLeakage current not excessive and electricstrength adequatePProtective impedance disconnected from liveparts before carrying out the tests16.2N/ASingle-phase appliances: test voltage 1,06times rated voltage. :N/A244 VThree-phase appliances: test voltage 1,06PN/A. :16.3Leakage current measurements(See appended table)PElectric strength tests according to table 7(See appended table)PNo breakdown during the tests17OVERLOAD PROTECTION OF TRANSFORMERS AND ASSOCIATED CIRCUITSNo excessive temperatures in transformer orassociated circuits in event of short-circuitslikely to occur in normal use1919.1P(See appended table)PAppliance supplied with 1,06 or 0,94 times ratedvoltage and the most unfavourable short-circuitor overload likely to occur in normal use applied. :N/ATemperature rise of insulation of the conductorsof safety extra-low voltage circuits not exceedingthe relevant value specified in table 3 by morethan 15 KN/ATemperature of the winding not exceeding thevalue specified in table 8,N/Ahowever limits do not apply to fail-safetransformers complying with sub-clause 15.5of IEC 61558-1N/AABNORMAL OPERATIONThe risk of fire or mechanical damage underabnormal or careless operation obviatedP

Electronic circuits so designed and appliedthat a fault will not render the applianceunsafeAppliances incorporating contactors orrelays subjected to the test of 19.14, beingcarried out before the tests of 19.11(IEC/EN 60335-1/A2)PNo contactors, no relaysN/A19.2Test of appliance with heating elements withrestricted heat dissipation; test voltage (V):power input of 0,85 times rated power input. :N/A19.3Test of 19.2 repeated; test voltage (V): powerinput of 1,24 times rated power input. :N/A19.4Test conditions as in cl. 11, any control limitingthe temperature during tests of cl. 11 shortcircuitedNo thermostat or similarN/A19.5Test of 19.4 repeated on Class 0I and Iappliances with tubular sheathed or embeddedheating elements. No short-circuiting, but oneend of the element connected to the elementssheathNo appropriate heatingelementsN/A19.619.11The test repeated with reversed polarity and theother end of the heating element connected tothe sheathN/AThe test is not carried out on appliancesintended to be permanently connected to fixedwiring and on appliances where an all-poledisconnection occurs during the test of 19.4N/AAppliances with PTC heating elements tested atrated voltage, establishing steady conditionsNo appropriate heatingelementsN/AThe working voltage of the PTC heating elementis increased by 5 % and the appliance isoperated until steady conditions are reestablished. The voltage is then increased insimilar steps until 1,5 times working voltage oruntil the PTC heating element rupturesN/AElectronic circuits, compliance checked byevaluation of the fault conditions specified in19.11.2 for all circuits or parts of circuits, unlessthey comply with the conditions specified in19.11.1PAppliances incorporating a protective electroniccircuit subjected to the tests of 19.11.3 and19.11.4 (IEC/EN 60335-1/A1)N/AAppliances having a switch with an off positionobtained by electronic disconnection, or a switchplacing the appliance in a stand-by mode,subjected to the tests of 19.11.4 (IEC/EN 603351/A1)N/AAppliances incorporating an electronic circuitthat relies upon a programmable component tofunction correctly, subjected to the test of19.11.4.8 (IEC/EN 60335-1/A2)N/A

19.11.119.11.219.11.3Before applying the fault conditions a) to f) in19.11.2, it is checked if circuits or parts of circuitmeet both of the following conditions:- the electronic circuit is a low-power circuit, thatis, the maximum power at low-power points doesnot exceed 15 W according to the tests specifiedN/A- the protection against electric shock, firehazard, mechanical hazard or dangerousmalfunction in other parts of the appliance doesnot rely on the correct functioning of theelectronic circuitN/AFault conditions applied one at a time, theappliance operated under conditions specified incl. 11, but supplied at rated voltage, the durationof the tests as specified:a) short circuit of functional insulation ifclearances or creepage distances are less thanthe values specified in 29N/Ab) open circuit at the terminals of any componentN/Ac) short circuit of capacitors, unless they complywith IEC 60384-14N/Ad) short circuit of any two terminals of anelectronic component, other than integratedcircuits. This fault condition is not appliedbetween the two circuits of an optocouplerN/Ae) failure of triacs in the diode modeN/Af) failure of an integrated circuit (IEC/EN 603351/A1)N/Ag) failure of an electronic power switching device(IEC/EN 60335-1/A2)N/AIf the appliance incorporates a protectiveelectronic circuit which operates to ensurecompliance with clause 19, the relevant test isrepeated with a single fault simulated, asindicated in a) to f) of 19.11.2N/ADuring and after each test the following ischecked:- the temperature rise of the windings do notexceed the values specified in table 8N/A- the appliance complies with the conditionsspecified in 19.13N/A- any current flowing through protectiveimpedance not exceeding the limits specified in8.1.4N/A

If a conductor of a printed board becomes opencircuited, the appliance is considered to havewithstood the particular test, provided all three ofthe following conditions are met:19.11.419.11.4.1- the material of the printed circuit boardwithstands the burning test of annex EN/A- any loosened conductor does not reduce theclearances or creepage distances between liveparts and accessible metal parts below thevalues specified in cl. 29N/A- the appliance withstands the tests of 19.11.2with open-circuited conductor bridgedN/AAppliances having a switch with an off positionobtained by electronic disconnection, or (IEC/EN60335-1/A1)N/Aa switch that can be placed in the stand-bymode, (IEC/EN 60335-1/A1)N/Asubjected to the tests of 19.11.4.1 to 19.11.4.7(IEC/EN 60335-1/A1)N/AAppliances incorporating a protective electroniccircuit subjected to the tests of 19.11.4.1 to19.11.4.7, except that (IEC/EN 60335-1/A1)N/AAppliances operated for 30 s or 5 min during thetest of 19.7 are not subjected to the tests forelectromagnetic phenomena. (IEC/EN 603351/A1)N/AAppliances having a device with an off positionobtained by electronic disconnection, orN/Aa device that can be placed in the stand-bymode,N/Asubjected to the tests of 19.11.4.1 to 19.11.4.7(IEC/EN 60335-1/A2)N/AThe appliance is subjected to electrostaticdischarges in accordance with IEC 61000-4-2,test level 4 (IEC/EN 60335-1/A1)N/A

19.11.4.2The appliance is subjected to radiated fields inaccordance with IEC 61000-4-3, test level 3(IEC/EN 60335-1/A1)21MECHANICAL STRENGTH21.1Appliance has adequate mechanical strengthand is constructed as to withstand roughhandlingPChecked by applying blows to the appliance inaccordance with test Ehb of IEC 60068-2-75,spring hammer test, impact energy 0,5 J(IEC/EN 60335-1/A1)N/AIf necessary, supplementary or reinforcedinsulation subjected to the electric strengthtest of 16.3N/AIf necessary, repetition of groups of three blowson a new sampleN/AThe requirement is not applicable to the glassenvelopes of heat lamps incorporated in ceilingmounted heat lamp appliances.(IEC/EN 60335-2-30/A2)N/A21.2N/AFor appliances with heating elements indirect contact with accessible panels madeof glass, ceramic or similar material : appliedto the panel a force of 2 J (IEC/EN 60335-230/A2)See aboveN/AAccessible parts of solid insulation havingstrength to prevent penetration by sharpimplements(IEC/EN 60335-1/A1)No accessibleinsulation (heating foil)N/AThe insulation is tested as specified, unless(IEC/EN 60335-1/A1)N/Athe thickness of supplementary insulation is atleast 1 mm and reinforced insulation is at least2 mm (IEC/EN 60335-1/A1)N/A21.101For visibly glowing radiant heaters: mass 5 kgplaced for 1 min on central part of the fireguard(IEC/EN 60335-2-30)No appropriate heaterN/A21.102Test for heaters having a part fixed to the walland other part hinged to it. (IEC/EN 60335-2-30)No appropriate heaterN/AAfter the test, compliance with 8.1 and Clause29.1 shall not be impaired.2222.1N/ACONSTRUCTIONAppliance marked with the first numeral ofthe IP system, relevant requirements of IEC60529 are fulfilledN/A

22.2Stationary appliance: means to provide allpole disconnection from the supply provided,the following means being available:- a supply cord fitted with a plugP- a switch complying with 24.3N/A- a statement in the instruction sheet that adisconnection incorporated in the fixed wiring isto be providedN/A- an appliance inletN/ASinge-pole switches and single-pole protectivedevices for the disconnection of heatingelements in single-phase permanentlyconnected class I appliances, connected in thephase conductorN/ASinge-pole switches and single-pole protectivedevices for the disconnection of heatingelements in single-phase, permanentlyconnected class 01 and class I appliances,connected to the phase conductor (IEC/EN60335-1/A2)N/AAppliance provided with pins: no unduestrain on socket-outletsN/AApplied torque not exceeding 0,25 NmN/APull force of 50 N to each pin after the appliancehas being placed in the heating cabinet; whencooled to room temperature the pins are notdisplaced by more than 1 mmN/AEach pin subjected to a torque of 0,4 Nm; thepins are not rotating unless rotating does notimpair compliance with the standardN/A22.4Appliance for heating liquids and appliancecausing undue vibration not provided with pinsfor insertion into socket-outletsN/A22.5No risk of electric shock when touching thepins of the plug, the appliance beingdisconnected from the supply at the instant ofvoltage peak.(IEC/EN 60335-1/A1)P22.3Tests as described22.12Handles, knobs etc. fixed in a reliable mannerN/ANo Handles,KnobsN/AFixing in wrong position of handles, knobs etc.indicating position of switches or similarcomponents not possibleN/AAxial force 15 N applied to parts, the shapebeing so that an axial pull is unlikely to beappliedN/AAxial force 30 N applied to parts, the shapebeing so that an axial pull is likely to be appliedN/A

22.13Unlikely that handles, when gripped as innormal use, make the operators hand touchparts having a temperature rise exceeding thevalue specified for handles which are held forshort periods onlyN/A22.14No ragged or sharp edges creating a hazard forthe user in normal use, or during usermaintenancePNo exposed pointed ends of self tapping screwsetc., liable to be touched by the user in normaluse or during user maintenanceP22.15Storage hooks and the like for flexible cordssmooth and well roundedN/A22.16Automatic cord reels cause no undueabrasion or damage to the sheath of theflexible cord, no breakage of conductorsstrands, no undue wear of contactsN/ACord reel tested with 6000 operations, asspecifiedN/AElectric strength test of 16.3, voltage of1000 V appliedN/ASpacers not removable from the outside by handor by means of a screwdriver or a spannerN/ANot applicable to rollers or feet preventoverheating of walls or floor if the appliancemeets the requirement of clause 19 withoutthese parts in place. (IEC/EN 60335-2-30/A2)N/A22.18Current-carrying parts and other metal partsresistant to corrosion under normal conditions ofuseN/A22.19Driving belts not used as electrical insulationN/A22.20Direct contact between live parts and thermalinsulation effectively prevented, unless materialused is non-corrosive, non-hygroscopic andnon- combustibleN/ACompliance is checked by inspection and, ifnecessary, by appropriate testN/A22.1722.21Wood, cotton, silk, ordinary paper and fibrous orhygroscopic material not used as insulation,unless impregnatedThis requirement does not apply to magnesiumoxide and mineral ceramic fibres used for theelectrical insulation of heating elements (IEC/EN60335-1/A2)None of mentioned materialsusedPN/A22.22Appliances not containing asbestosP22.23Oils containing polychlorinated biphenyl (PCB)not usedN/A22.24Bare heating elements adequately supportedto prevent contact with accessible metal partsin case of rupture or saggingN/A

22.25Sagging heating conductors cannotcome into contact with accessible metalpartsN/A22.26The insulation between parts operating atsafety extra-low voltage and other live partscomplies with the requirements for double orreinforced insulationN/A22.27Parts connected by protective impedanceseparated by double or reinforced insulationN/A22.28Metal parts of Class II appliances conductivelyconnected to gas pipes or in contact with water:separated from live parts by double or reinforcedinsulationN/A22.29Class II appliances permanently connected tofixed wiring so constructed that the requireddegree of access to live parts is maintainedafter installationN/A22.30Parts serving as supplementary orreinforced insulation fixed so that theycannot be removed without beingseriously damaged, or22.31Pso constructed that they cannot be replaced inan incorrect position, and so that if they areomitted, the appliance is rendered inoperableor manifestly incompleteN/AClearances and creepage distances oversupplementary and reinforced insulation notreduced below values specified in clause 29 asa result of wearN/AClearances and creepage distances betweenlive parts and accessible parts not reducedbelow values for supplementary insulation, ifwires, screws etc. become looseN/A22.36Handles continuously held in the hand innormal use are so constructed that whengripped as in normal use, the operatorshand is not likely to touch metal parts,unless they are separated from live partsby double or reinforced insulationN/A22.37Capacitors in Class II appliances notconnected to accessible metal parts,unless complying with 22.42N/AMetal casings of capacitors in Class IIappliances separated from accessiblemetal parts by supplementary insulation,unless complying with 22.42N/A22.38Capacitors not connected between thecontacts of a thermal cut-outN/A22.39Lamp holders used only for the connectionof lampsN/AFor replaceable heat lamps in ceilingmounted heat lamp appliances, ceramicinsulating parts of lampholders are used.(IEC/EN 60335-2-30/A2)N/A

22.40Motor-operated appliances and combinedappliances intended to be moved while inoperation, or having accessible movingparts, fitted with a switch to control themotor. The actuating member of theswitch being easily visible and accessibleN/AUnless the appliance can operatecontinuously, automatically or remotelywithout giving rise to a hazard, appliancesfor remote operation being fitted with aswitch. The actuating member of theswitch being easily visible and accessibleN/A(IEC/EN 60335-1/A2)22.41No components, other than lamps,containing mercuryN/A22.42Protective impedance consisting of atleast two separate componentsN/AValues specified in 8.1.4 not exceeded ifany one of the components are shortcircuited or open-circuitedN/A22.43Appliances adjustable for differentvoltages, accidental changing of thesetting of the voltage unlikely to occurN/A22.44Appliances are not allowed to have anenclosure that is shaped and decorated sothat the appliance is likely to be treated asa toy by childrenPAppliances shall not have an enclosurethat is shaped or decorated like a toy(IEC/EN 60335-1/A2)P22.45When air is used as reinforced insulation,clearances not reduced below the valuesspecified in 29.1.4 due to deformation as aresult of an external force applied to theenclosureN/A22.46Software used in protective electroniccircuits is software class B or C (IEC/EN60335-1/A1)N/A22.47Appliances connected to the water mainswithstand the water pressure expected innormal use (IEC/EN 60335-1/A1)N/ANo leakage from any part, including anyinlet water hoseN/A22.48Appliances connected to the water mainsconstructed to prevent backsiphonage ofnon- potable water (EN 60335-1/A1)N/A22.49For remote operation, the duration ofoperation shall be set before the appliancecan be started, unlessN/A

the appliance switches off automatically orcan operate continuously without hazardN/A(IEC/EN 60335-1/A2)22.50Controls incorporated in the appliancetake priority over controls actuated byremote operation ((IEC/EN 60335-1/A2))22.51A control on the appliance being manuallyadjusted to the setting for remoteoperation before the appliance can beoperated in this mode(IEC/EN 60335-1/A2)N/AThere is a visual indication showing thatthe appliance is adjusted for remoteoperation (IEC/EN 60335-1/A2))N/AManual setting and visual indication notnecessary on appliances that can operateas follows, without giving rise to a hazard(IEC/EN 60335-1/A2) :N/A- operate continuously,N/A- operate automatically, orN/A- be operated remotelyN/A22.52Socket-outlets on appliances accessible tothe user in accordance with the socketoutlet system used in the country in whichthe appliance is sold(IEC/EN 60335-1/A2)22.101Heater other than heaters for mounting athigh level shall be guarded in order toprevent contact with heating elements(IEC/EN 60335-2-30)2323.1N/AN/AWireways smooth and free from sharpedgesPWires protected against contact with burrs,cooling fins etc.PWiring effectively prevented from coming intocontact with moving parts23.3N/AINTERNAL WIRINGWire holes in metal well rounded orprovided with bushings23.2No appropriate controlsPPBeads etc. on live wires cannot change theirposition, and are not resting on sharp edges orcornersN/ABeads inside flexible metal conduits containedwithin an insulating sleeveN/AElectrical connections and internalconductors movable relatively to eachother not exposed to undue stressFlexible metallic tubes not causingdamage to insulation of conductorsPN/A

Open-coil springs not usedN/AAdequate insulating lining provided inside acoiled spring, the turns of which touch oneanotherN/ANo damage after 10 000 flexings for conductorsflexed during normal use or 100 flexings forconductors flexed during user maintenanceN/AElectric strength test, 1000 V between liveparts and accessible metal partsN/A23.4Bare internal wiring sufficiently rigid and fixedN/A23.5The insulation of internal wiring withstanding theelectrical stress likely to occur in normal usePNo breakdown when a voltage of 2000 V isapplied for 15 min between the conductor andmetal foil wrapped around the insulationN/A23.6Sleeving used as supplementary insulation oninternal wiring retained in position by positivemeansP23.7The colour combination green/yellow used onlyfor earthing conductorsP23.8Aluminium wires not used for internal wiringN/A23.9No lead-tin soldering of stranded conductorswhere they are subject to contact pressure,unlessN/Aclamping means so constructed that there is norisk of bad contact due to cold flow of the solderN/AThe insulation and sheath of internal wiring,incorporated in external hoses for theconnection of an appliance to the water mains,at least equivalent to that of light polyvinylchloride sheathed flexible cord (60227 IEC 52)(IEC/EN 60335-1/A1)N/A23.102525.1SUPPLY CONNECTION AND EXTERNAL FLEXIBLE CORDSAppliance not intended for permanentconnection to fixed wiring, means forconnection to the supply:- supply cord fitted with a plug25.2P- an appliance inlet having at least the samedegree of protection against moisture asrequired for the applianceN/A- pins for insertion into socket-outletsN/AAppliance not provided with more than onemeans of connection to the supply mainsP

25.325.425.525.6Stationary appliance for multiple supply maybe provided with more than one means ofconnection, provided electric strength test of1250 V for 1 min between each means ofconnection causes no breakdownN/AConnection of supply conductors for applianceintended to be permanently connected to fixedwiring possible after the appliance has beenfixed to its supportN/AAppliance provided with a set of terminals for theconnection of cables or fixed wiring, crosssectional areas specified in 26.6N/AAppliance provided with a set of terminalsallowing the connection of a flexible cordN/AAppliance provided with a set of supplyleads accommodated in a suitablecompartmentN/AAppliance provided with a set of terminals andcable entries, conduit entries, knock-outs orglands, allowing connection of appropriate typeof cable or conduitN/ACable and conduit entries, rated current ofappliance not exceeding 16 A, dimensionsaccording totable 10N/AIntroduction of conduit or cable does notreduce clearances or creepage distancesbelow values specified in 29N/AMethod for assemble supply cord with theappliance:- type X attachmentN/A- type Y attachmentP- type Z attachmentN/AType X attachment, other than those with aspecially prepared cord, not used for flat twintinsel cordsN/APlugs fitted with only one flexible cordPSupply cords of single-phase portableappliances having a rated current not exceeding16 A, fitted with a plug complying with thefollowing standard sheets of IEC 60083:75 (EN60335-1)- for Class I appliances: standard sheet C2b,C3b or C4.:- for Class II appliances: standard sheet C5 orC6 . :25.7Supply cords being one of the following types(IEC/EN 60335-1/A2):N/AC5 varante IIPP

- rubber sheathed (at least 60245 IEC 53)N/A- polychloroprene sheathed (at least 60245 IEC57)N/A- cross-linked polyvinyl chloride sheathed (atleast 60245 IEC 87)N/APolyvinyl chloride sheathed:Not used if they are likely to touch metal partshaving a temperature rise exceeding 75K duringthe test of Clause 11.N/A- light polyvinyl chloride sheathed cord (atleast 60227 IEC 52), appliances notexceeding 3 kgH03VVH2-Fonly for 4595 and 6090 SHBN/A- ordinary polyvinyl chloride sheathed cord (atleast 60227 IEC 53), other appliancesAll other H05VV-FN/AHeat resistant polyvinyl chloride sheathed:Not used for type X attachment other thanspecially prepared cords.N/A- Heat-resistant light polyvinyl chloridesheathed cord (at least 60227 IEC 56),appliances not exceeding3 kgN/ACord anchorages only accessible with theaid of a tool, orN/Aso constructed that the cord can only be fittedwith the aid of a toolN/AType X attachment, glands not used ascord anchorage in portable appliancesN/ATying the cord into a knot or tying the cord withstring not usedN/A25.20Conductors of the supply cord for type Y and Zattachment adequately additionally insulatedN/A27PROVISION FOR EARTHING27.1Accessible metal parts of Class 0I and Iappliances, permanently and reliably connectedto an earthing terminal or contact of theappliance inletPEarthing terminals not connected to neutralterminalPClass 0, II and III appliance have no provision forearthingP25.1825.1927.2Safety extra-low voltage circuits not earthed,unless protective extra-low voltage circuitsN/AClamping means adequately secured againstaccidental looseningN/A

27.327.427.527.6Terminals used for the connection of externalequipotential bonding conductors allowconnection of conductors of 2,5 to 6 mm², andN/Ado not provide earthing continuity bet

f) failure of an integrated circuit (IEC/EN 60335-1/A1) N/A g) failure of an electronic power switching device (IEC/EN 60335-1/A2) N/A 19.11.3 If the appliance incorporates a protective electronic circuit which operates to ensure complian

Related Documents:

Effective date: Apr 30, 2020 . IEC/EN 60335-1, IEC/EN 60335-2-2 CNS 3765, CNS 3765-2 100128 電熨斗 IEC/EN 60335-1, IEC/EN 60335-2-3 CNS 3765, CNS 3765-3 100129 旋轉式脫水機 IEC/EN 60335-1, IEC/EN 60335-2-4 CNS 3765-4, CNS 2926 sub-clause 3.3, 3.4 100130 洗碗機 IEC/EN 60335-1

Mar 26, 2019 · (IEC 60335-2-40:2018) Not a refrigerating system NA 5.101 Motor-compressor subjected to relevant test of clause 19 of IEC 60335-2-34, unless (IEC 60335-2-40:2018) Not a refrigerating system NA motor-compressor comply with that standard (IEC 60335-2-40:2018) Not a refrigerating

IEC 61215 IEC 61730 PV Modules Manufacturer IEC 62941 IEC 62093 IEC 62109 Solar TrackerIEC 62817 PV Modules PV inverters IEC 62548 or IEC/TS 62738 Applicable Standard IEC 62446-1 IEC 61724-1 IEC 61724-2 IEC 62548 or IEC/TS 62738 IEC 62548 or IEC/TS 62738 IEC 62548 or IEC/TS 62738 IEC 62548 or IEC/

(IEC 60335-2-30) N/A 5.3 Appliance used for tests of Cl. 19 also used for the test of Cl. 22.24 (IEC 60335-2-30) N/A Test of Cl. 22.24 carried out after test of Cl. 29 (IEC 60335-2-30) N/A 5.6 Thermostats short-circuited if sensible to room air temper

-A concerning IEC/EN 60335-2-29 (4th Edition) and IEC/EN 60335-1 (4th Edition) -B concerning IEC/EN 60335-1 (5th Edition) the two reports should be used together Based on the above information, the sample has been tested and found compliant with the requirement of the safety standards l

IEC/EN 62031 IEC/EN 62471 IEC 60598-1 ENEC IEC/EN 61347-2-13 (Relevant clauses) Relevant clauses of IEC/EN 60335-1 Annex BB of IEC/EN 60335-2-89 Annex CC of IEC/EN 60335-2-24 NSF UL 2108 CSA C22.2 No.9.0-96 Environmental RoHS/REACH Application IP rating No IP-rating, fit for use in

DIN EN 60335-1;VDE 0700-1:2020-08 AS/NZS 60335.1:2011 A1:2012 A2:2014 A3: 2015 A4:2017 A5:2019 IEC 60335-2-2:2009 A1:2012 A2:2016 EN 60335-2-2:2010 A11 A1:2013 DIN EN 60335-2-2:2014 A2:2015 VDE 0700-2: 2014 A1:2015 Household and similar electrical appliances – Safety – Part 2-2:

heater is first switched on; (IEC 60335-2-73) P – a statement that the immersion heater is to be installed only in a system where the heating element is always below water level, such as in a cistern-fed tank, if applicable (IEC 60335-2-73/A1) P 7