Chapter 26: An Introduction Chromatographic Separations

2y ago
9 Views
2 Downloads
6.85 MB
84 Pages
Last View : 6d ago
Last Download : 3m ago
Upload by : Angela Sonnier
Transcription

Chapter 26: An Introductionto Chromatographic Separations Column Chromatography Migration Rates– Distribution Contstants– Retention Times– Selectivity Factor Zone Broadening & Column Efficiency Optimizing Performance ResolutionIntro to ChromatographyChromatography is a separation techniqueMany determinations involve separation followed by analysisChromatographyelectrophoresisHPLC & GC are our primary focusAlso discuss low pressure column chromatography & TLC (thinlayer) All chromatographic techniques have– Stationary phase – solid or viscous liquid phasetypicallyin a column– Mobile phase – moves sample in contact withstationary phase

Chromatography:sample transported by mobile phaseelectrostatic or van der Waals'some components in sample interact more strongly withstationary phase and are more strongly retainedsample separated into zones or bandsElution Chromatography:flushing of sample through column by continual mobile phase(eluent) additionmigration rate fraction time spent in mobile phasePlanar chromatography - flat stationary phase, mobile phasemoves through capillary action or gravityColumn chromatography - tube of stationary phase, mobile phasemoves by pressure or gravity

Overview ofchromatographicprocess – packedcolumn‐ Inject at to‐ Separate t1 to t3‐ Detect at t4Å Resultingchromatogram A & B retained bycolumn differentlyB has higher KB takes longer to elutefrom columnDetector sees A firstthen BPeak heights & peakareas are proportionalto conc.Band broadening

Important: chromatogram (concentration versus elution time) more strongly retained species elutes last (elution order) analyte is "diluted" during elution (dispersion) zone broadening proportional to elution timeBy changing experimental conditions, non-separated bands can beseparated(A) adjust migration rates for A and B (increase band separation)(B) adjust zone broadening (decrease band spread)

Partitioning type of equilibrium where the analyte divides itselfbetween two phasesFor liquid‐liquid extraction – two liquidsFor chromatography – mobile vs. stationary phases Analyte A in equilibrium with two phases AmobileAstationaryDefine a partition ratio K (or distribution constant)CsK ‐‐‐‐‐‐‐CMwhere Cs & CM areconcentrations of analyte instationary & mobile phases Prefer if K is constant over conc. range If not constant we can work in a narrow range whereit is constant This is linear chromatography From now on everything is linear chromatography In linear chromatography a constant flow rate ofmobile phase moves through column K is typically constant or nearly constant Elution process by which analyte is flushed throughthe column by mobile phase (which could be a liquidor a gas)

tM time for unretained molecule to reach detector or dead timetR retention time, time for retained species to reach detectorDefine ν as average linear rate of solutemigration & L as column length, thenLν ��‐‐‐ velocitytimeSimilarly if define µ as average linear rate ofmovement of molecules of mobile phaseLµ ‐‐‐‐‐tM

tM time for unretained molecule to reach detector or dead timetR retention time, time for retained species to reach detectorRelating retention time tR to K ( Cs/CM)ν µ x fraction of time analyte is in mobile phasemoles of analyte in mobile phaseν µ x ��‐‐‐‐‐‐‐‐‐‐‐number of moles of analyte1CM VMν µ x �‐‐‐ µ x �‐‐‐‐CM VM Cs Vs1 CsVs/CMVMSubstituting K Cs/CMGives1ν µ x �‐1 K Vs/VM

More useful relationships ‐ capacity factor k’(comes from K) K in concentration, k’ in molesamount of analyte in stationary phasek’ ��‐‐‐‐‐‐‐‐‐‐‐‐amount of analyte in mobile phaseKAVsnsSo for A Æ kA’ ‐‐‐‐‐‐‐ ‐‐‐‐‐‐VMnMn # ofmolesFrom previous slide1ν µ x �‐1 K Vs/VMFrom previousequation ÆCan plug in ν L/tR & µ L/tMRearrangeand get1ν µ x ‐‐‐‐‐‐‐‐‐‐‐1 kA’1ν µ x ‐‐‐‐‐‐‐‐‐‐1 kA’tR – tMkA’ ‐‐‐‐‐‐‐‐‐‐‐tMNow have kA’ in terms of something easily measured inchromatogramCompares how long it takes a species to move through systemcompared to unretained speciesRelative because ratio, Numerator Net RetentionWhen k'A is 1.0, separation is poorWhen k'A is 30, separation is slowWhen k'A is 2-10, separation is optimum

One step further Æ Selectivity factor (α) describes differentialmigrationFor twocomponentsAnd fromchromatogramKBkB’α ‐‐‐‐‐‐ ‐‐‐‐‐‐‐KAkA’(tR)B ‐ tMα ‐‐‐‐‐‐‐‐‐‐‐‐‐‐‐(tR)A – tMAllows calculation of the resolving power ofa chromatographic system (i.e. column with A & B)larger α better separation(B) Adjusting Zone Broadening: Individual molecule undergoes "random walk" Many thousands of adsorption/desorption processes Average time for each step with some ve or -ve differences Add up to give Gaussian peak (like random errors) Breadth of band increases down column because morea time Zone broadening is affected by separation efficiency - moreefficient, less broadening

Chromatographic Plate Theory vs. Rate Theory Plate theory based in liquid‐liquid extraction(successive extractions) K Corg/Cwater Chromatographic column can be thought ofin the same way (only continuous process) K Cs/CMStationary phase beadMobile phase (liquid)L Divide chromatographic column upinto steps or segments calledtheoretical plates The theoretical concept is that thesetheoretical plates are equilibriumunits for K Cs/CM The more theoretical plates a columnhas, the more efficient it isL NH If column length L & N number ofplates, then H height equivalent toortheoretical plateN L/H

Gausianpeaks –statisticaldistributionofmoleculesWb 4σGausian distribution (bell curve)W 4σ

Can deriveN 16 (tR/Wb)2N number of platesWb base widthN 16 (tR/4σ)2 (tR/σ)2N 5.54 (tR/W½)2W½ width athalf heightColumn manufacturers use Nto characterize column – N varies widelyOther Variables Affecting Peak Width (Zone Broadening):Mobile Phase Velocity:Higher mobile phase velocity, less time on column, less zone broadeningHowever, plate height H also changes with flow rate - plot of H versus u called vanDeemter plot (Fig 26-8)

Rate Theory of ChromatographyH HL HS HM HSMH height equivalent to theoretical plate (as in Plate Theory)HL contribution due to longitudinal diffusionHS stationary phase mass transfer contributionHM diffusion associated with mobile phase effectsHSM diffusion into or mass transfer across a stagnant layer ofmobile phase (neglect)H B/µ Cµ Avan Deemter Equation A, B & C are coefficients, µ velocity1) Uneven Flow or Eddy DiffusionPath 1 is shorter than path 2HM A Moleculesmovethrough different paths Larger difference inpathlengths for largerparticles At low flow rates,diffusion allows particlesto switch betweenpaths quickly andreduces variation intransit time

2) Longitudinal Diffusion HL (B/µ)tRt 00 t tRσL2 0σL2 2 DM tMVariance due to longitudinal diffusion 0 at startVariance increases with time & diffusion coefficient D Diffusion from zone (front and tail) Proportional to mobile phase diffusion coefficient Inversely proportional to flow rate - high flow, less time fordiffusion2) Mass transfer in & out of stationary phaset2t3t1Mobile PhaseStationary PhaseResulting PeaksBroadening of peaks is a function of mobile phase velocity(moving molecules faster than those in stationary phase)Not the same as longitudinal diffusionHS CµIn Plate Theory condition at t1 assumed to hold throughout

Putting it all togetherVan DeemterOverallFinding optimum

Optimizing Column Performance – seldomoperate at optimum Æ too slowNormally want to get required separation inshortest time, this may be at 2X µoptCan optimize a separation by varyingexperimental conditions, usually goals are1) reduce band broadening (zone)2) alter relative migration rates ofcomponents (allowing better separation oftwo components)

This bringsus toResolution(RS) Measureof columnsability toseparate 2analytesNote ΔZ spread ofpeaks &W or W/2 peak widthΔZ2 ΔZ2[(tR)B – (tR)A]RS �‐ ‐‐‐‐‐‐‐‐‐‐‐ �‐W A WBWA/2 WB/2 WA WBIf RS 1.0 then ΔZ WA/2 WB/2and peaks touch with about 4% overlapThis is too big an error to tolerateIf RS 1.5 then about 0.3% overlapCan lengthen column to improve resolution byincreasing N Æ this also increases time foranalysis

Avoidvaluesabove10

k’ in optimalrange of 2 to 5Commonly found problem in chromatographyGeneral Elution ProblemSolution – change conditions duringchromatographic run so that k’ changesStart with conditions for chromatogram (a), after1 & 2 eluteChange to conditions for chromatogram (c), after3 & 4 eluteChange to conditions for chromatogram (b) toget 5 & 6

k’ in optimalrange of 2 to 5Since k’ is related to partitioning of solutebetween mobile phase and stationary phase,can easily change mobile phaseIn GC do temperature programmingIn HPLC do solvent programming (a.k.a. gradientelution)

Homework 26‐126‐326‐926‐14

Chapter 27: Gas Chromatography PrinciplesInstrumentationDetectorsColumns and Stationary PhasesApplicationsBasic Principle of GC – sample vaporized by injectioninto a heated system, eluted through a column byinert gaseous mobilephase and detectedThree types (or modes)gas – solid chromatographyearly(stationary phase: solid)gas – liquid“important(stationary phase: immobilized liquid)gas – bonded phase “relatively newAn estimated 200,000 GC in use worldwide

Discusscomponentsstarting hereCarrier gas: He (common), N2, H2Pinlet 10-50 psiF 25-150 mL/min packed columnF 1-25 mL/min open tubular columnColumn: 2-50 m coiled stainless steel/glass/TeflonOven: 0-400 C average boiling point of sampleaccurate to 1 CDetectors: FID, TCD, ECD, (MS)

Carrier gases (mobile phase) – must bechemically inert He, Ar, N2, CO2 even H2 andmixtures 95/5 N2/CH4Often detector dictates choice of carrier gasIn GC sample doesn’t really interact with carriergas (unlike HPLC), temp controls partitioningOften necessary to purify cylinder gas with a trap,scrubber or cartridge of molecular sieves (orbuy high purity gas) O2 ppm HcThe move today is away from gas cylinderstoward gas generators (extract pure carriergas from air)

Next is gasflow controlIn this regionFlow control – 10 to 50 psi with regulatorRegulators vary in quality, material & control,typically use a 2 stage regulator with the bestmaterial being stainless steelUltimately flow rate is checked by a soap bubblemeter for accurate flow

InjectorNext is gasflow controlIn this region

Injector – use micro syringe 99.9 % of the timeinjecting 1 to 20 µL, rapidly shoot in plug ofsampleOld GCs had separate injection areaToday use on‐column & microflash vaporizers –all have septum of synthetic rubber which ispunctured by syringeInjector usually 50 oC hotter than boiling point ofsample – also hotter than columnCan use rotary injector valve (as for HPLC)

Rotary Injection ValveCommon for HPLC, rare in GCSplit injection: routine method0.1-1 % sample to columnremainder to wasteSplitless injection: all sample to columnbest for quantitative analysisonly for trace analysis, low [sample]On-column injection: for samples that decompose aboveboiling point - no heated injection portcolumn at low temperature to condense sample in narrowbandheating of column starts chromatography

Column housed in Column Oven to maintaintemperatureTypes – packed, open tubular, capillaryoldest �‐ newestCapillary columns will take over completelyPacked – tube (steel, glass, fused silica, Teflon)packed with materialOpen Tubular – coated on wallsCapillary – coated on walls, long & narrowLength range – 2 to 50 m (typically 30 m)

Column ConceptsIn GC since mobile phase is under pressure & weoperate at various temperaturesgiven that P V is proportional to TSometimes use retention volumes (VR, VM)VR tR F for retained species tR retentiontimeVM tM F for unretainedtM retentiontimeF flow rateProblem ‐ pressure drop across a columnPressure at head of column may be 5 atm & atend of column may be 1 atmNeed a correction factor3[(Pi/P)2 ‐1]j �‐2[(Pi/P)3 ‐ 1]Where Pi inlet pressure &P outlet pressure (atmospheric)

Detectors – dozens of detectors availableCharacteristics of an ideal detector:1) Adequate sensitivity for desired analysis(typical 10‐8 to 10‐15 g analyte/sec)2) Stable – background constant with time3) Reproducible – good precision4) Linear response over several orders ofmagnitude5) Temperature range – room temp ‐ 400 oCCharacteristics of ideal detector: (continued)6) Rapid response time7) Independent of flow rate8) Reliable9) Easy to Use – inexperienced operators10) Either selective or universal response11) NondestructiveNo detector exhibits all these characteristics

Flame IonizationDetector (FID)‐ one of most widelyused GC detectors‐ good sensitivity toalmost all organiccompoundsFID Basics‐ column effluent mixed with air and burned inH2 flame producing ions & electrons thatconduct electricity‐ a few hundred volts applied between burnertip & a collector electrode above the flameproducing currents on the order of 10‐12 amps‐ amplify & measure‐ signal approximately proportional to numberof reduced carbon atoms in flame

FID Basics (continued)‐ mass sensitive rather than concentration‐ insensitive to non combustible gases – H2O,CO2, SO2, NOxFID exhibits‐ High sensitivity (as low as 10‐13 g/s)‐ Large linear response range (107)‐ Easy to use‐ Rugged‐ DESTRUCTIVEThermal ConductivityDetector (TCD)‐ One of earliest GCdetectors‐ Not popular today‐ Low sensitivity‐ Several designs‐ Use heated wire orsemiconductor‐ Resistance of wirechanges withanalyte vs carrier

TCD uses bridge circuit withSample & Reference CellsTCD‐ New TCDs use pulsed current to increasesensitivity & reduce drift‐ Thermal conductivity of He & H2 are about 6to 10 times greater than most organiccompounds (must use these carrier gases)‐ Other carrier gases (N2, Ar, etc) have thermalconductivities too close to organics

Advantages of TCD‐ Simple Æ Reliable & Easy to use‐ Universal response (organic & inorganic)‐ Large linear dynamic range 105‐ Nondestructive, can use in tandem‐ Older instruments have built‐in TCDDisadvantages‐ Low sensitivity‐ Often can’t use with capillary columnsbecause amount of analyte is smallECD

Electron Capture Detector‐ Sample passes over β emitter (radioactive) like63Ni foil or 3H adsorbed on Pt or Ti foil2‐ β particles (i.e. electrons) hit carrier gas(usually N2) causing a burst of e‐ to be released& measured by electrode standing current orconstant signal‐ When analyte molecule that absorbs e‐ passesthrough, current is reduced signal‐ Response is non‐linear unless pulsedECD Advantages‐ Responds well to molecules withelectronegative atoms like halogens (F, Cl, Br,I), peroxides, quinones, & nitro groups‐ Insensitive to amines, alcohols, hydrocarbons‐ Chlorinated pesticides are big application‐ Highly sensitive‐ Easy to use‐ Pretty reliable, although foil can get coated‐ Selective

ECD Disadvantages‐ Narrow linear range‐ Radioactive‐ Regular wipe test‐ Bake out contaminants‐ Some limits to applicability because highlyselectiveOther Conventional DetectorsThermionic Detector (TID)‐ Selective for N & P compounds‐ 500 x more sensitive than FID for P‐ 50 x more sensitive than FID for N‐ Bad for C‐ Design similar to FID with rubidium silicatebead at 180 V vs collector Æ get hot plasma600 ‐ 800 oC‐ Produces large number of ions with N & P

Flame Photometric Detector (FPD)‐ Selective for P & S compounds‐ Again sample goes through H2/air flame‐ Observe optical emission of HPO at 510 nm &526 nm & S2 at 394 nm‐ Use optical filters to isolate signal‐ Can also measure halogens, N, some metals(e.g. Cr, Ge, Se)Photoionization Detector (PID)‐ Column effluent irradiated with intense UVlight source‐ Ionizes molecules‐ Measure ions with electrodes in detector cell

Unconventional Detectors(Hyphenated Techniques)Atomic Emission Detector (AED)‐ Fairly new‐ Very powerful‐ Sample eluent introduced to He microwaveplasma atomizing all atoms in sample‐ Uses diode array detector measuring opticalemission over wide spectral range (170 ‐ 780nm)‐ Measure many elements simultaneouslyGC‐AED

GC‐AED‐ Potentially can measure70 or more elements‐ If look at C signal fromAED get chromatogramwith hundreds of peaks‐ If look at O signal getvery simplechromatogram with onlya few peaksGC – Mass Spectrometry (GC‐MS)‐ Already covered Mass Spec‐ Interfacing GC & MS normally difficult‐ GC at pressure above atmospheric while MSunder high vacuum‐ Need special interfaces for packed columns‐ Jet separator – discussed below‐ Membrane separator – a membrane sandwichbetween spiral channels, column efluent on oneside under pressure, MS on other side undervacuum – relies on differential permeability ofcarrier gas vs analyte molecules

GC‐MS SchematicInterface less critical for capillary columns

Jet Separator‐ Purpose is to get more analyte into MS thancarrier gas‐ Usually an all glass device‐ Principle is that heavier atoms have greatermomentum and travel a fairly straight pathinto the MS, lighter carrier gas molecules aredeflected outward by vacuum & pumpedawaySeveral types of Mass Specs available‐ Rarely magnetic sector or time of flight‐ Usually quadrapole or ion trap for GC‐MS‐ Less expensive‐ Less maintenance‐ Easy to use‐ Normally use electron multiplier as detector‐ All MS systems need ion source, either electronimpact or chemical ionization

Three modes of operation for GC‐MS1) Spectral mode – look at mass spectrumevery second or so during chromatogram ‐gives most information for research ormethod development2) Total ion current – sum signal for all ions asone large signal – highest sensitivity3) Selective ion monitoring (SIM) – look atcertain mass/charge ratios for compounds ofinterest – routine analysisGC‐MS‐ sensitive‐ can be very selective in SIM mode‐ powerful for qualitatively & quantitativelyThere is also one other kind of Mass SpecIon Cyclotron MS which is a very high resolution,Fourier transform instrument not used for GC

GC‐FTIR‐ Powerful technique for identifying compounds‐ Use heated light pipe 1 to 3 mm diaGC‐FTIR‐ Powerful technique for identifying compounds‐ Use heated light pipe 1 to 3 mm dia and 10 to 40cm long‐ Heat to prevent condensation of sample‐ Cool detector for sensitivity‐ Gives structural information from spectrum‐ Not very common

GC Columns & Stationary Phases‐ Historically used packed columns‐ Stationary phase coated as a thinfilm ona high surface area solid support‐ Theoretical studies showed that unpackedcolumns with narrow diameters were better‐ Open tubular columns first developed‐ Capillary columns came later because‐ Very fragile, difficult to construct, hard to connectto GCs, small samples hard to detect, difficult tocoat column walls, etc.Packed Columns‐ Tubing of metal, glass, Teflon, etc.‐ 2 to 3 m long and 2 to 4 mm in dia‐ Packed with diatomaceous earth (SiO2), clay,carbon particles, glass microbeads, polymer‐ Diameter 150‐250 µm (60‐100 mesh) 1 m2/g‐ Thin coating of liquid stationary phase

Stationary Phase CoatinggoodbadSilicaParticleSilicaParticleDon’t want stationary phase liquid coating to bleed orpuddle in column – gives zone broadening & poorresolutionOpen Tubular Columns Æ Capillary ColumnsColumn evolutionThree typesWall Coated Open Tubular (WCOT) – open glasstube with coating on wall – duhSupport Coated Open Tubular (SCOT) – opentube with particles of support material stuck tothe wallsFused Silica Open Tubular (FSOT) – WCOT madeof fused silica

Surface chemistry – glass & silicawith ‐OH at surfaceOH OH OH –O–Si–O–Si–O–Si–O‐are SiO2SilanolgroupSilicacoreOH is a problem because it can adsorb polarsubstances with strong affinity causing peaktailing – must deactivate by reacting

React Si‐OH groups with silaneSometimes still have –OH groupsIf silica not pure may have metal impurities M‐OHtypically use high purity silica – acid washSame chemistry to making specialty bonded phaseLiquid coatings on stationaryphase should exhibit:1) Chemical inertness2) Low volatility (b.p. 100 oC max temp)3) Thermal stability4) Good solvent characteristics (i.e. k’ & αsuitable)Many different liquid coatings have been usedor attempted for GC, only about 10 havewithstood the test of time

Retention time of a solute depends on K (partitioncoefficient) which is dependent on stationaryphase – must have different K’s for differentanalytesHowever, if K’s too large Æ long retention timeif K’s too small Æ short retention timeresulting in incomplete separationIn choosing a stationary phase use generalprinciples such as “like dissolves like”, polargroups interact with polar groups, non polar withnon polar, etc.

Polar groups include –CN, –CO, –OHPolar analytes include alcohols, acids, aminesNon polar Æ hydrocarbonsWhere analyte & stationary phase match is goodÆ elution order is determined by boiling pointsBonded Stationary PhasesUse silylation chemistry to covalently attachstationary phase to solid support or columnwall

Bonded Stationary PhasesAdvantages‐ monolayer coverage can be obtained‐ reduced bleeding of stationary phase‐ longer lasting‐ better stability‐ can be solvent washedChiral Stationary Phases – separating stereo‐isomers is the ultimate in chromatography,separate molecules that are mirror imagesPredicting retention (or identifying compoundsI) Selectivity Factorsby retention)If B is a standard compound & we know α, canthen be able to identify compound A even if wechange the the chromatographic conditionsor go to another chromatograph, etc.This is limited to specific applications where adatabase is available, not universally applicable

II) Retention Index (I)Proposed by Kovats in 1958Index based on normal alkanesIf have a mixture of 2 known alkanes & 1unknown compound & the 2 knownsbracket unknown in tR can then determineI for unknown & identify itI 100 x # of carbon atomsRegardless of column packing, temp. or otherconditionsKovats Retention IndexDoesn’t work as well for other typesof compounds (Hc), but useful insome cases e.g. homologous seriesPlot log adjusted retention time (tR’ tR – tM)vs number of carbon atoms is linearUseful in particular fields – petroleum industry,cosmetics, pharmaceuticals, etc. since havetheir own unique “standards”

Note number ofcarbons thatwould becalculated forthese 3compoundsbased on IHomework 27‐4 27‐5 27‐19

Chapter 28: High-Performance LiquidChromatography (HPLC) Scope Instrumentation – eluants, injectors, columns Modes of HPLC– Partition chromatography– Adsorption chromatography– Ion chromatography– Size exclusion chromatographyHPLC Most widely used separation technique Broad applicability – organic & inorganic Can be very sensitive, accurate & precise Suitable for separation of nonvolatile species Has found numerous uses in industry, clinicalsettings, environmental areas, pharmaceuticals,etc.

Instrumentation for HPLC: For reasonable analysis times, moderate flow raterequired but small particles (1-10 μm) Solvent forced through column 1000-5000 psi - moreelaborate instrument than GC Solvents degassed - "sparging“ High purity solventsSingle mobile phase composition - isocratic elutionProgrammed mobile phase composition - gradient elution

Gradient elutiondramaticallyimproves theefficiency ofseparation

Solvents (mobile phase) – are storedin special reservoirs connected to thepumping system – must be free ofparticles that can clog components & free ofbubble forming gases that get trapped incolumn or detectorThree basic ways to degas solvents1) vacuum or suction filter (0.4 or 0.2 µm)2) ultrasonicate (with vacuum)3) He purge (sparge units often built in)Can purchase HPLC solvents & water - stillHPLC pumping systems typically employ tworeciprocating or piston pumpsCheck valves& pump sealsneed to bereplacedPulse‐freeflow is neverreallyachieved Up to 10,000 psi, small internal volumes Produces pulsation

In GC the analyte affinity for the column isinfluenced by tempIn HPLC the solvent strength affects an analytesretention on columnTherefore, analogous to temp programming inGC, do solvent programming in HPLCThis is also referred to as gradient elutionHPLC sample injectors are exclusively 6 port valvesthat are overfilled by syringe giving extreme accuracy& precision – typical volumes are 10 to 50 µL but canbe larger

Rotary Injection ValveCommon for HPLC, rare in GCInjector for HPLC 6 port rotary valve Similar to FIA, GC Introduce small sample (0.1-100 μL) without depressurization Microsyringe/septum system (only 1500 psi)

Columns- usually stainless steel- can be PEEK (poly ether ether ketone)- may cost 200- 1000 packed- Length 10-30 cm, ID 4-10 mm- Packings are 3, 5, or 10 µm particle size- Most common 25 cm, 5 µ, 4.6 mm ID- N 40,000 to 60,000- Normally packed under 6000 psi pressure atfactory as a slurryGuard columns are normally used before theanalytical column to protect & increaselifetime of column – operator usually slurry ordry packs short guard column regularly withsame or similar packing used in analyticalcolumn (old column material) – can purchaseguard systems, cartridges, etc.

Detectors for HPLC- Ideal characteristics same as GC- Exception is temp range- Low dead volume 1 to 10 µLBulk property detectors - measure property of mobile phase(refractive index, dielectric constant, density)Solute property detectors - measure property of solute not presentin mobile phase (UV absorbance, fluorescence, IR absorbance)Most common detector is UV-vis absorbanceThree types1) Filter instrument – optical filters, Hg lamp2) Variable wavelength – monochromator3) Diode array detector- provide spectraMany HPLC detectors availableFor universal & selective detection

1) Filter based UV-vis detector –Typically set at254 nm using the most prominent band inHg spectrum – can also use 313, 365, 334 nmand other lines as well2) Variable wavelength detectors – usecontinuum source like (D2 or H2) & amonochromator, select any λ, less sensitive3) PDA - D2 or H2 source, disperse & focus ondiode array, get complete spectrum every 1sec, powerful, expensive, less sensitive, lotsof data generatedFrom columnCell for UV‐visdetectorfor HPLC‐ Low vol

sources: single line (arc or hollow cathode lamp, laser) continuum (Xe, D2 lamp) detector: photodiode/photomultiplier tube photodiode arrayCombination of separation and analysis (GC-MS, HPLC-UV-Vis) very powerfulDiode ArrayDetectorFluorescence detector – normally fixed wavelengthfilter fluorometer excitation filter & emissionfilter can be changed for particular λ of interestgives selectivity based on:- ability to exhibit fluorescence- excitation wavelength- emission wavelengthVariable λ monochromator based fluorescencedetectors also availableFilter based detectors usually more sensitive

Refractive index detector (RI) - responds tonearly all solutes but has poor sensitivity –detects changes in refractive index as samplepasses through as long as solute has differentRI than solvent – analogous to TCD in GCElectrochemical Detection Amperometric – fix potential & measurecurrent (i) Conductometric – measure conductivity Coulometric – fix potential & integrate i Voltammetric – vary potential & measure i Potentiometric – measure potentialCan use 2 or 3 electrode design with Pt or carbonelectrodes (glassy C or C paste)Electrochem. detector nearly universal

Other HPLC detectors LC-MS using thermospray – new popularity(pharmaceuticals) Evaporative light scattering - polymers LC-FTIR LC-plasma emission or ICP-MSModes of SeparationPartition Chromatography – most used form ofHPLC primarily for nonionic compounds of varyingpolarity with low MW ( 3000)Most common form is bonded phase chrom. using silicabased packing materials functionalized by silylation(as for GC)Partition Chromatography: Most popular method Low molecular weight (mw 3000) analytes Polar or non-polar Bonded stationary phase column (liquid chemically bonded to support particles)3, 5 or 10 μm hydrolyzed silica particles coated with siloxanesNormal phase HPLC nonpolar solvent/polar columnReversed phase HPLC polar solvent/nonpolar column

Early work with partition chrom. was done with polarstationary phases (like bare silica) & nonpolar solutes normal phase chromatog.Later bonded phases were introduced using C18 groupsÆ very non-polar with polar solvents reversed-phasechromatographyToday almost all partition chrom. done in reversed-phasemode with many different bonded phases (althoughC18 very popular)Normal- (polar column) versus Reversed Phase(nonpolar) elution:Reversed-phase HPLC most common (high polarity solvent, high polarity solutes elutefirst)

‐Si‐CH2‐(CH2)16‐CH318 carbon chainLong chain acts as if it were an alkane coated onsilica Æ analyte molecules partition into it,hence the nameIn chromatogram, most polar compounds elutefirst because they partition into C18 least – likedissolves like – most non‐polar compoundscome out last

Besides C18 can have C8, C4, C3, C2, C1 plusfunctionalities like cyano (-C2H4CN), amino (C2H4NH2), diol (-C3H6O-CH2-CHOHCH2OH)Each has different polarityCan also do Ion Pair Chromatography or Paired-IonChromatography – type of RP-HPLC used to separateionic speciesStill partition chrom. but use a reagent like a quaternaryammonium salt (C4H9)4N to pair with analyte ions toseparate by RPColumn Optimization in HPLC:Can optimize k' and αMore difficult than GC- in GC mobile phase just transported solute- in HPLC mobile phase interacts with soluteAnalyte Polarity:hydrocarbons ethers esters ketones aldehydes amines alcoholsStationary Phase Choice:Choose column with simila

Chapter 26: An Introduction to Chromatographic Separations Column Chromatography Migration Rates – Distribution Contstants – Retention Times – Selectivity Factor Zone Broadening & Column Efficiency Optimizing Performance Resolution Intro to Chromatography

Related Documents:

Part One: Heir of Ash Chapter 1 Chapter 2 Chapter 3 Chapter 4 Chapter 5 Chapter 6 Chapter 7 Chapter 8 Chapter 9 Chapter 10 Chapter 11 Chapter 12 Chapter 13 Chapter 14 Chapter 15 Chapter 16 Chapter 17 Chapter 18 Chapter 19 Chapter 20 Chapter 21 Chapter 22 Chapter 23 Chapter 24 Chapter 25 Chapter 26 Chapter 27 Chapter 28 Chapter 29 Chapter 30 .

TO KILL A MOCKINGBIRD. Contents Dedication Epigraph Part One Chapter 1 Chapter 2 Chapter 3 Chapter 4 Chapter 5 Chapter 6 Chapter 7 Chapter 8 Chapter 9 Chapter 10 Chapter 11 Part Two Chapter 12 Chapter 13 Chapter 14 Chapter 15 Chapter 16 Chapter 17 Chapter 18. Chapter 19 Chapter 20 Chapter 21 Chapter 22 Chapter 23 Chapter 24 Chapter 25 Chapter 26

DEDICATION PART ONE Chapter 1 Chapter 2 Chapter 3 Chapter 4 Chapter 5 Chapter 6 Chapter 7 Chapter 8 Chapter 9 Chapter 10 Chapter 11 PART TWO Chapter 12 Chapter 13 Chapter 14 Chapter 15 Chapter 16 Chapter 17 Chapter 18 Chapter 19 Chapter 20 Chapter 21 Chapter 22 Chapter 23 .

About the husband’s secret. Dedication Epigraph Pandora Monday Chapter One Chapter Two Chapter Three Chapter Four Chapter Five Tuesday Chapter Six Chapter Seven. Chapter Eight Chapter Nine Chapter Ten Chapter Eleven Chapter Twelve Chapter Thirteen Chapter Fourteen Chapter Fifteen Chapter Sixteen Chapter Seventeen Chapter Eighteen

18.4 35 18.5 35 I Solutions to Applying the Concepts Questions II Answers to End-of-chapter Conceptual Questions Chapter 1 37 Chapter 2 38 Chapter 3 39 Chapter 4 40 Chapter 5 43 Chapter 6 45 Chapter 7 46 Chapter 8 47 Chapter 9 50 Chapter 10 52 Chapter 11 55 Chapter 12 56 Chapter 13 57 Chapter 14 61 Chapter 15 62 Chapter 16 63 Chapter 17 65 .

HUNTER. Special thanks to Kate Cary. Contents Cover Title Page Prologue Chapter 1 Chapter 2 Chapter 3 Chapter 4 Chapter 5 Chapter 6 Chapter 7 Chapter 8 Chapter 9 Chapter 10 Chapter 11 Chapter 12 Chapter 13 Chapter 14 Chapter 15 Chapter 16 Chapter 17 Chapter

Chapter 3 Chapter 4 Chapter 5 Chapter 6 Chapter 7 Chapter 8 Chapter 9 Chapter 10 Chapter 11 Chapter 12 Chapter 13 Chapter 14 Chapter 15 Chapter 16 Chapter 17 Chapter 18 Chapter 19 Chapter 20 . Within was a room as familiar to her as her home back in Oparium. A large desk was situated i

The Hunger Games Book 2 Suzanne Collins Table of Contents PART 1 – THE SPARK Chapter 1 Chapter 2 Chapter 3 Chapter 4 Chapter 5 Chapter 6 Chapter 7 Chapter 8. Chapter 9 PART 2 – THE QUELL Chapter 10 Chapter 11 Chapter 12 Chapter 13 Chapter 14 Chapter 15 Chapter 16 Chapter 17 Chapt