TECHNICAL SPECIFICATIONSBARTON Floco/Flotrac Meters, Floco Samplers and PulseTransmittersFloco Series F Meters Models F-500/F-2500Cameron’s BARTON Floco Series F meter is a positiveSeries F meters are available in various models with pressure ratingsdisplacement meter with a unique rotor design that providesup to 2000 psi (138 bar), an operating temperature range of -20 sustained accuracy even under adverse conditions. It accuratelyF to 400 F (-29 C to 205 C), and a flow rate capacity of 6 to 90measures viscous, waxy, corrosive and abrasive liquids to withinUSGPM based on 20 API gravity oil. 1% maximum uncertainty.Standard connections include 1” female and 2” or 3” male NPTThe Series F meter measures liquid by separating it into equalthreads. Raised face and ring joint flanges can also be supplied.portions and counting them. Liquid enters the meter throughStandard units for registering totals are US gallons, 42-gallon barrels,the inlet port, where the bridge deflects the liquid downwardliters or cubic meters.to strike the rotor blades and turn the rotor. The liquid thenpasses through the outlet port, which is aligned with theinlet port. The unique rotor design allows solid particles andsediment to pass through the meter without causing damageSeries F meters are available with Buna-N , Viton or Teflon components for compliance with process demands, and a long lifebetween maintenance cycles.or malfunction. Bridge seals prevent the liquid from passingto the outlet port without being measured.As the process fluid viscosity increases the measurementperformance of the Floco also increases. The enhancementis most notable in the capability to measure low flow rates.Because of this many process fluids can be measured at lowerrates than published.Key Features Unpowered mechanical operation Suited to viscous fluids laden with solids 25,000 cP upper viscosity limit Optional integral sampler Optional electronic output

Applicationassembly and magnetic coupling seal). Almost all spare partsare interchangeable among Series F models.A wide variety of liquids can be metered with the Series F meter,including heavy oil (25,000 centistokes max),asphalt emulsion, brine, bunker C oil,crude oil, kerosene, liquid fertilizers,paraffin, refined oils and water.Body and SideplatesThe meter body is A216 WCB cast steel with amaximum hardness of HRC22.The Series F meter does not requirethe installation of straight-run pipeupstream or downstream of themeter. The meter can be installed in anyposition and the register can be rotated in twoplanes for optimal visibility.The sideplates are A105 forged steel. The numberand type of sideplate bolts determines the safeworking pressure of the meter, as shown in the table below.All meters are compliant with ASME B31.1 & B31.3requirements. When equipped with NACE bolts, Series Fmeters are compliant with NACE MR0175/ISO 15156: 2009revision. All meters supplied with flanged end connectionsutilize slip-on flanges as standard. Optional weldneck flangesfor B31.1 compliance are available. All flanged connections arewelded per ASME Section IX procedures.ConstructionThe Series F meter consists of four basic parts:body, rotor, sideplates and register (with gear caseSafe Working PressureModelNumber of Boltsper rd (A574 or SAE Grade 8)Safe Working Pressure(psi)(Mpa)7505.2F-50043/8NACE (A320 L7M & A193 B7M)4252.9F-250083/8Standard (A574 or SAE Grade 8)150010.3F-250083/8NACE (A320 L7M & A193 B7M)8505.9F-250087/16Standard (A574)200013.8F-250087/16NACE (A564)150010.3Liner and WearplatesBody parts subject to mechanical wear or fluid abrasion aredesigned for economical field replacement. The body liner andside wearplates are constructed of polished 316 stainless steel toassure a low-friction seal with the rotor blades. Wearplates arereversible for extended life.BridgeThe bridge is available in either Delrin or 316 stainlesssteel. Delrin is a plastic material that can withstand chemicalattack and temperatures to 180 F (82 C). Alternatively,316 stainless steel is extremely resistant to abrasive andhigh-temperature fluids.Bridge seals are constructed of Viton.BearingsBearing selection should be based on the following guidelines: A luminum Bronze – General bearing, durable in mostapplications including crude oil2 M eehanite – Recommended for use with abrasive process fluid,which is often apparent by indications of severe wear on the rotorshaft C arbon Graphite – Recommended for use where yellow metals arenot acceptable or where the process fluid has very low lubricity G lass-Filled Teflon – Recommended for use where other materialsfail due to chemical attackRotorThe rotor, which is the measuring element of the flow meter, isconstructed of 316 stainless steel and has chrome-plated shaft endsfor bearing surfaces. The standard rotor hub is made of non-clad 316stainless steel for superior corrosion resistance. An optional rotor hubis clad with a Viton or Buna-N elastomer for improved flow at verylow flow rates. The spring-loaded blades are made of a stainless steelsubstrate with a Buna-N, Viton, or Teflon covering.

RegistersThe register and gear case assembly, featuring a 316 stainless steelproven magnetic coupling design, is sealed from contact with themetered liquid for accurate registration and long life.All registers display flow totals in seven digits with the resolutionindicated in the table below. A standard sweep hand provides 10times greater resolution than that shown. Reset registers have twodisplays: one reset and one small-digit non-reset. The displaysfeature 1/4” digits for good visibility.Available RegistersSeries F Meter with Flanged Ends (in.)Face-to-Face, Raised Face or Ring Joint MetersMeter Size11150 or 300 ANSI600 ANSI900 ANSI2111213-1/23121313-1/2Flanged ends are not commonly available on 1” meters.SpecificationsCommon Trim ConfigurationsUnitsResolutionInternalsStandardHigh Temp.NACEUS Gallons (1” and 2”)1 galUS Gallons (3” only)10 galMaximumTemperature180 F (80 C)300 F (150 C)410 F (210 C)optional180 F (80 arrels (42 US gal)0.1 bblLiters10 litersCubic Meters0.01 cubic metersMaintenanceBridgeDelrinStainless SteelDelrinFloco meters can be serviced without being removed from theline, and without special tools or training. Removal of the metersideplate provides easy access to all internal parts. Using theassembly drawing and parts list shipped with each meter, a usercan readily identify all parts and assembly tandardL7MFlow RatesDimensionsSeries F Meter with Threaded or Victaulic Ends (in.)*MeterSizeFlowCapacityIdeal FlowRatesPressure Dropat Max. Female6 min to 60max10 min to 35 max121 and 1/268-3/46 min to 60max10 min to 35 max32Male3Male9 min to 90max15 min to 60 max5* Victaulic ends are available only on 2” meters.Based on pale hydraulic oil 0.89 S.G. at 60 F (16 C), 110 SSU at 100 F (38 C)with 1.0% accuracy. Minimum capacity improves with higher viscosities.2These flow rates are determined to provide the best accuracy and durability.1 BFADGEH3

Floco Automatic SamplersThe Floco Automatic Sampler is an accessory to the Flocometer that provides proportional-to-flow sampling of fluids asthey are metered. The sampler is used to determine the fluidquality which aids in determining:obe preferred for their transparency, ease of cleaning andimmunity to static electricity charge. Plastic receptaclesmay be preferred for their shatter resistance.o Easy The ratio of oil and water being producedadaption to large-capacity remotemounted receivers, including those usedwith sour fluids. Cameron’s CLIF MOCK receivers are available in capacitiesup to five gallons and include floatoperated full bottle shut-offs. They areavailable in plastic or stainless steel.The stainless steel containers include amechanical level gauge. The ratio of dissolved gasses in heavy oil The physical and chemical properties of the flowing fluidAs defined by API and ISO, the proportional-to-flow techniqueis superior to manual-grab sampling or automatic samplingbased on time or event in producing representative results.Proportional-to-flow sampling is achieved by automaticallyextracting a series of small, consistently sized samples from theflowing stream. The sample size is easily adjusted and thesample interval can be adjusted by changing a gear within thesampler.The Floco sampler is extraordinarily economical to install. Itmounts on the side of a Floco meter and is mechanically drivenby a gear train connected to the flow-driven rotor within themeter. The sample is typically stored in a container mountedon the sampler. The container is vented locally, thereforeentrained or dissolved gas and light hydrocarbons with vaporpressures higher than atmospheric pressure will break out ofsolution and be lost to atmosphere. To avoid exposingpersonnel to toxic vapors, never use this container forsour fluids.The Floco Sampler is available in two models: T he FRA Sampler can be attached to an F-500 Floco meterand is limited to a 500 psi safe working pressure. A wide-mouth glass or plastic receiver. Glass receivers mayCLIF MOCK large-capacity receivero Adaptableto vertical installation whenflow is oriented upwardso Anintegral manually operated ball valve shut-off tostop operation or to allow removal of the sampler formaintenance without interrupting the meter’s operation.oA magnetic drive coupling for reduced maintenanceo 1000psi standard operating pressure, enabling thesampler’s use with F-500 meters up to 750 psi and withspecially adapted F-2500 meters.Samplers ordered with Floco meters are preassembled with themeter before shipment. Samplers may also be ordered individuallyfor assembly with an existing meter in the field. T he GLA Sampler is based on the original FRA design, butoffers a number of advanced features not available in theFRA including:oAn improved weight-activated full bottleshut-off mechanismModel FRA samplerModel GLA sampler4

Floco Automatic Samplers (continued)SpecificationsGLA SamplerFRA SamplerBodyASTM A48 Cast Grey Iron No. 30(not NACE-listed)ASTM A48 Cast Grey Iron No. 30(not NACE-listed)InternalsHardened 440 stainless steeland Meehanite – standardHigh performance treatment for abrasive process services – optionalHardened 440 stainless steeland Meehanite – standardHigh performance treatment for abrasive process services – optionalElastomerVitonBuna-N/ VitonDrive CouplingMagneticMechanicalIntegral Sample ReceiverWide mouth glass – standardWide mouth plastic – optionalRemote receiver – optionalNarrow mouth polyethylene plasticReceiver Volume with Scaleand Transport Lid1800 ml1600 mlMinimum Process Pressure10 PSIG10 PSIGMaximum Process Pressure1000 PSIG500 PSIGMaximum Process Temperature200 F (93 C)400 F (205 C)1Maximum Viscosity (flowing)5000 cP5000 cPDimensions (additional to horizontal pipecenterline when mounted on a Floco meter Includes clearance to remove bottle)13.75” high X 5.5” wide35 cm high X 14 cm wide16.5” high X 7.5” wide42 cm high X 19 cm wideVertical Pipe - OptionDownward flowUpward flowSample VolumeAdjustable 0.6 to 5.7 cc (ml)Adjustable 0.6 to 5.7 cc (ml)34.6 samples per cubic meter69.2 samples per cubic meter138.4 samples per cubic meter34.6 samples per cubic meter69.2 samples per cubic meter138.4 samples per cubic meterSample Rate2Gear ratio 80:1 standardGear ratio 40:1Gear ratio 20:112Temperature rating does not preclude the effects of process water flashing to steam and the suitability of the receiver for resistance to thermal shock or melting.Reduce sample rate by half when used with a 3” Floco meterFloco Model 1334 Pulse TransmitterThe Floco Pulse Transmitter (FPT) adds a frequency output toa Floco Series F meter. It is ideally suited for retrofitting in-situmeters to provide the necessary electronic output associatedwith an automation or SCADA project. The transmitter consistsof a gear-like target that is driven by the Floco meter andmounted within the magnetic field of a standard turbine flowmeter pick-up coil. As the teeth of the gear pass under thepick-up coil, an electronic pulse is generated in the sameway the movement of the rotor blades of a turbine metergenerates a pulse. The FPT can generate up to 126 pulses perUS gallon for any 1” or 2” Floco meter.The FPT can be installed easily and quickly without removingthe meter from service. The transmitter mounts directly to themeter body and can be ordered for use with or without amechanical register.Optional NUFLO electronicsare available to meet desiredoutput signal requirements.As an option, Cameron offersthe Model 1334 with aneconomical poweredcoil for applicationsrequiring a highamplitude 2V to 24Vsquare wave signal.Electrical certificationsfor this model aredifferent than those forthe standard model.Model 1334 pulse transmitterwith optional junction box5

Floco Model 1334 Pulse Transmitter (continued)Approval Classification OptionsWhile the standard FPT is suitablefor 250 F (121 C), optionaltemperature ratings up to 450 F (232 C) are available. ATEX, Explosion-proof, Exd IIc, T4, -40 F to 212 F(-40 C to 100 C) C SA, Explosion-Proof, Class I, Groups B, C, D; Class II,Groups E, F, G; Class III; Enclosure 4 (US and Canadianelectrical code)Explosion-proof models feature a 1/2”FNPT electrical connection, while theintrinsically safe model has anoptional 1” MNPT hub. C SA, Intrinsically Safe, Class I, Groups A, B, C, D; Class II,Groups E, F, G; Class III; Enclosure 4 (with approved barrier) CSA, Intrinsically Safe, Class I, Groups A, B, C, D; Class II,Groups E, F, G; Class III; Enclosure 4 (without barrierswhen connected to a NUFLO Scanner FlowComputer, or a NUFLO MC-II Flow Analyzer)Blind 2” F-500 meter withMC-III flow analyzer CSA, General PurposeOutput SpecificationsOutput FormRequired AccessoryLow-amplitude pulseModel 1334 standardAmplified pulseModel 1334 option (powered coil)LCD total and rate, 4-20mA, amplified pulse, ModbusModel 1334 standard plus MC-III Flow AnalyzerLCD total and rate, amplified pulse, datalogging, Modbus,4-20mA or FOUNDATION FieldbusModel 1334 standard plus Scanner 2000 microEFMFloco Model 308 Pulse TransmitterLike the Model 1334 transmitter, the Model 308 transmits adiscrete electrical pulse to represent that a specific amount ofliquid has transferred through the Floco meter. The rate of pulsetransmission can be interpreted to determine the flow rate whilecounting the pulses representing the flow total. With the Model308, each pulse is exactly divisible by a factor of ten to alloweither very simple or sophisticated remote devices to scale thepulses into preferred units of measure.Optional pulse rates from 1 to 1000 pulses per barrel or 1 to 100pulses per gallon may be used to operate electric counters, batchingcounters or combined with preset electrical counters to controlpumps, motors, valves, or solenoid-actuated equipment.A glass-encapsulated dry reed switch is actuated by the magneticfield of a gear-driven magnet. Although the Model 308 transmitteris not explosion-proof or agency-certified for hazardous areas,Pulses Per Unit of VolumeRegister Type610:1 Gear Ratio(Part No. 9A-0308-0004A)100:1 Gear Ratio(Part No. ubic Meters100010,000

Floco Model 308 Pulse Transmitter (continued)the reed switch is hermetically sealed for use in hazardouslocations as allowed by localelectrical code.The wafer-type cast aluminum caseof the Model 308 mounts on a Flocometer just beneath the register. Thepulse transmitter is easily retrofit toexisting meters in the field withoutrecalibrating the meter. The Model 308Model 308 pulsetransmitterPulse Transmitter incorporates an integralweatherproof junction box with a threeposition terminal strip that is easily replaced in the field.General SpecificationsDimensionsLength: 4.5”Width: 3.75”Height: 2.0”Weight1.5 lbTemperature Rating-25 F (-32 C) to 160 F (71 C)Conduit Connection1/2” NPTContact Rating12 volt-ampere AC10 watts DC resistive1/2 amp or 250 volts max.Contact Resistance10 to 60 milliohms plus40 milliohms lead resistanceFlotrac MetersFeaturesLike positive displacement meters,Cameron’s Flotrac meter is designedfor mechanically measuring volumesof low-viscosity liquids flowing athigh pressure. The Flotrac meteremploys the unique constrainedvortex principle to provide anaccurate measurement over a 10:1range.All construction materials used in the Flotrac meter are resistant tocorrosion and abrasion, and only one moving part in the meter—the rotor — comes in contact with the process fluid. Therotor assembly, comprised of bearings and an integralmagnet, can be quickly removed or replaced withoutremoving the meter from the line. The register assembly ismagnetically coupled to the rotor assembly, eliminatingfriction and leakage from packing glands.Process liquid enters the meteringchamber through the 1” (25mm) inletconnection. As the liquid makes a 360-degreeloop, it is separated equally into two streams. The specialconfiguration of the metering chamber forces these streams into aseries of vortices, causing the rotor assembly to rotate in directproportion to the flow rate. The two liquid streams are thencombined at the meter outlet port.Flotrac models 306 and 380 share several interchangeable parts,enabling the user to easily convert a standard flow Model 306meter to a low flow Model 380 meter and vice versa.Similar to a PD meter, the Flotrac meter does not require theinstallation of straight-run pipe upstream or downstream ofthe meter.Model 306Model 380Accuracyand Capacity 1.0% of reading,9 to 90 USGPM 2.0% of reading,1.5 to 15 USGPM(water) 2.0% of reading,4 to 15 USGPM(10 centistokes)Pressure Drop50 PSID (345 Kpad)in water atmaximum flow50 PSID (345 Kpad)in water atmaximum flowRegister Resolution1/100 US Barrels1/1000 US BarrelsRegister Resolution1 Liters1/10 LitersRegister Resolution1/1000 CubicMeters1/10,000 Cubic MetersSpecificationsOperating Pressure 2500 psig (172 bar) SWP withstandard bolting1500 psig (103 bar) SWP withL7M NACE boltingOperating Temperature32 F (0 C) to 200 F (95 C)Viscosity10 centistokes, max.Pipe Connections1” (25 mm) female, inlineFlanged connections optionalRegister8 digitsWeight22 lb (10 kg)DimensionsLength: 8.75” (220 mm)Width: 5.75” (145 mm)Height: 6.0” (150 mm)7

Flotrac Meters (continued)ConstructionThe Flotrac meter body is constructed of SA-216 WCB epoxycoated cast steel with an SA-105 forged steel cover. The impellerand housing inserts are made of glass-filled Ryton , which ispractically impervious to dissolved salts and alkalis, water, mostacids, and low-viscosity hydrocarbon liquids. Other wetted partsinclude 300 series stainless steel and a Buna-N body seal O-ring.The assembly is suitable for sour non-NACE service when thenon-wetted body bolts are supplied as optional grade L7M andthe body O-ring is supplied in Viton. The rotor bearings andjournals are made of special materials that provide exceptionalservice life with water or non-lubricating liquids. The registerassembly features a sealed weatherproof aluminum case. Themeter is compliant with ASME B31.1 and B31.3 requirements.NORTH AND SOUTHEUROPE, AFRICA,ASIA PACIFICMIDDLE EASTAMERICACASPIAN AND RUSSIASuite 16.02 Menara AmFirstLevel 9, Al Jazira Club Tower A14450 JFK Blvd.3 Steyning WayNo. 1 Jalan 19/3P.O. Box 47280, Muroor RoadHouston, TX 77032Southern Cross Trading Estate46300 Petaling JayaAbu DhabiUSABognor RegisSelangor Darul EhsanUnited Arab EmiratesTel 1 281 582 9500West Sussex PO22 9TTMalaysiaTel 971 2 596 8400ms-us@c-a-m.comEngland, UKTel 603 7954 0145ms-uk@c-a-m.comTel 44 1243 826741ms-kl@c-a-m.comms-uk@c-a-m.comA L G E R I A C A N A DA C H I N A I N D I A M A L AYS I A R U SS I A UA E U K U S Awww.c-a-m.com/measurement 2013 Cameron NUFLO and MC-III are trademarks of Cameron. BARTON, Floco, Flotrac, and Scanner are registered trademarks of Cameron. Modbus is a registered trademark of the Modbus Organization, Inc.Foundation is a trademark of the Fieldbus Foundation. Viton, Teflon, Buna N are registered trademarks of E.I. Dupont Pennwalt Corporation. Meehanite is a registered trademark of Meehanite WorldwideCorporation. SWP 2M 04/13 AD00621M

BARTON Floco/Flotrac. Meters, Floco Samplers and Pulse Transmitters. Floco Series F Meters Models F-500/F-2500. Cameron’s BARTON Floco Series F meter is a positive displacement meter with a unique rotor design that provides sustained accuracy even under adverse conditions. It accu

Related Documents:

BARTON Floco Meters, Floco Samplers and Pulse Transmitters Floco Series F Meters Models F-500/F-2500 Cameron’s BARTON Floco Series F meter is a positive displacement meter with a unique rotor design that provides sustained accuracy even under adverse conditions. It accurately measure

Mary Barton by Elizabeth Gaskell Mary Barton by Elizabeth Gaskell This etext was prepared by Les Bowler, St. Ives, Dorset. Additional proof reading by Joseph E. Loewenstein, M.D. MARY BARTON by Elizabeth Gaskell CONTENTS I. A mysterious disappearance. II. A Manchester tea-party. III. John Barton's great trouble. IV. Old Alice's history. V.

Min Longitude Error: -67.0877 meters Min Altitude Error: -108.8807 meters Mean Latitude Error: -0.0172 meters Mean Longitude Error: 0.0028 meters Mean Altitude Error: 0.0066 meters StdDevLatitude Error: 12.8611 meters StdDevLongitude Error: 10.2665 meters StdDevAltitude Error: 13.6646 meters Max Latitude Error: 11.7612 metersAuthor: Rafael Apaza, Michael Marsden

size meters. Better than 0.25% of reading over linear flow range of meters 1’’ size and above. Repeatability — 0.02% of reading. Compact and efficient — Compared to other metering techniques, Barton turbine meters are able to

ration of water meters in Smart-Metering measurement systems. 3 The EDC module was developped for: Single-jet dry dial meters ETKD/ETWD Multi-jet dry dial meters MTKD/MTWD Positive displacement meters RTKD Bulk meters WPD / WPHD / WSD

flow rate changes Two-bladed rotor offers large unobstructed flow area Electrical output signal adapts readily to a variety of readout devices Barton High Accuracy Gas and Liquid Turbines For more than 30 years BARTON turbine meters have pr

Clara Barton National Historic Site National Park Service During Miss Barton's long lifetime of service, she was honored by powerful leaders of world nations, lone survivors of American disasters, the rich and the poor. The world's admiration of Clara Barton is strikingly evident in the collection of medals and jewels that are preserved as part .

Lisa Little, Joan Wagner, and Anne Sutherland Boal 216 13. Emergency Preparedness and Response Yvonne Harris 232 14. Nursing Leadership through Informatics Facilitating and Empowering Health Using Digital Technology Shauna Davies 249 15. Regulation, the Law, Labour Relations, and Negotiations Beverly Balaski 261 16. Emerging Nursing Leadership Issues Brendalynn Ens, Susan Bazylewski, and Judy .