AquaRite 100

2y ago
32 Views
2 Downloads
2.77 MB
41 Pages
Last View : 21d ago
Last Download : 3m ago
Upload by : Rosa Marty
Transcription

AquaRite 100 Troubleshooting GuideTSG-AQR100aCopyright 2019 Hayward Industries Inc.

Safety PrecautionsWarningHigh Voltage Electrocution HazardHazardous voltage can shock, burn, cause serious injuryand or death. To reduce the risk of electrocution and orelectric shock hazards: Only qualified technicians should remove the panel Replace damaged wiring immediately Insure panel is properly grounded and bonded2!

Table of ContentsHow AquaRite 100 works, main display layout, waterchemistry RequirementsHow To:1. Input Voltage Configuration2. Adjust Daily Chlorine Output3. Clean Electrolytic CellTroubleshooting:1. Inadequate Water Flow LED2. Problem Detected with MIN LED or LED #23. Problem Detected with LED # 3 or LED #44. Problem Detected with LED # 5,6,7, or 95. Problem Detected with LED # 8 or MAX LEDAdditional Information:Salt Addition TableSalt Dilution FormulaReplacement Parts ListReading Serial Numbers3Pg. 4-7Pg. 8-1291011-12Pg. 13-3614-1617-2324-2728-3132-36Pg. 37-4138394041

AquaRite: How It Works The AquaRite 100 Salt Chlorine Generator isdesigned to convert 99% pure salt into chlorinegas (HOCl). The conversion occurs in the Electrolytic Cell,also known as the Cell (used throughout thisguide). Power is sent from the AquaRite 100 ControlCenter to the Cell when chlorine production isscheduled (based on time percentage). When power is applied to the Cell, conductiveplates produce a field; which in-turn establishesan electrochemical reaction between Chlorideions & the water to create Hypochlorous Acid.4

AquaRite 100: Display LayoutFEKHDJICGBA5AInadequate Water Flow LEDBCell Life Low LEDCProblem Detected LEDDStanding By LEDEGenerating Chlorine LEDFSuper Chlorinating LEDGMinimum Chlorine Output LED (10%)HDAILY CHLORINE OUTPUT LED’sIChlorine Output Adjustment ButtonsJSuper Chlorinate ButtonKMaximum Chlorine Output LED (100%)

AquaRite 100: Display FunctionABFCEKHDEJICDFGHGBAIPress ( ) and ( - ) to increase or decreasethe chlorine generationJBoosts DAILY CHLORINE OUTPUT to 100%for 24 hoursThe Maximum DAILY CHLORINE OUTPUTamount of chlorine being generatedK6Illuminates when there is no flow or lowflow through the CellIlluminates when Cell has reached the endof its expected lifeIlluminates if there is a problem with theunitIlluminates when unit is waiting for thefilter pump to turn on or is in-betweengenerating cyclesIlluminates when unit is actively generatingchlorineIlluminates when unit is SUPERCHLORINATINGThe Minimum DAILY CHLORINE OUTPUTamount of chlorine being generatedIndicates DAILY CHLORINE OUTPUTsettings

AquaRite 100: Chemistry RequirementsSalt LevelsFree Chlorine 1.0 – 3.0 ppmPH 7.2 – 7.8 ppmAlkalinity 80 – 120 ppmCalcium Hardness 200 – 400 ppmCyanuric Acid (Stabilizer)Metals Outdoor Pools: 30 – 50 ppm Indoor Pools: 0 ppm 0 ppmSaturation Index -.2 to .2 (0 is best)Phosphates Less than 200 ppbNitrates7 Pools 15k gal and below: 1500 – 2700 ppm Pools up to 30k gal: 2800 – 4500 ppm (Ideal level 3200 ppm) As close to 0 ppb as possible

AquaRite 100 How To:

How To: Input Voltage ConfigurationThe AquaRite 100 can be powered by 120 VAC or 240 VAC. Use a multimeter to verify voltage in order to determine wire configuration.9120 VACWire Colors240 VACWire ColorsLine 1Black & VioletLine 1BlackLine 2N/ALine 2RedNeutralRed & WhiteNeutralWhite & VioletGroundGreenGroundGreen* Power must be turned off at the circuit breaker before performing any wiring. Follow ALL Local andNEC electrical codes. To provide safe operation, the AQR100 should be installed by a qualified serviceprofessional and must be properly grounded AND bonded.

How To: Adjust Chlorine OutputThe “Daily Chlorine Output” buttons set the level of production during the generating cycle. Generating cyclesoccur in 60 minute increments. Each led illuminated (from Min to MAX) represents 10%, or six(6) minutesof chlorine production time.Example: The first 4 LED’s being on implies the Cell is producing chlorine 40% of the pump run time; so theCell should produce chlorine for the first 24 minutes of each hour before repeating.Press to SuperChlorinatePress to decreaseoutput setting(Temporarily increaseschlorine output to 100%for 24 hours. Pump mustbe on during this time)SPress to increaseoutput settingNOTE: If the chlorine level does not increase within 24 hours after increasing output, test water using anindependent test kit to determine the current salt, stabilizer, phosphate, and nitrate levels.NOTE: Output is scaled back to 20% or lower (depending on where the dial is set) at 60 F and chlorineproduction stops altogether at 50 F.10

How To: Clean the CellCell cleaning frequency is dependent on several factors; pH & calcium levels have thegreatest effect on how often cells requires cleaning. In pH environments between (7.2- 7.8) cells typically require cleaning 3-4 times a year (with moderate calcium levels).Turn Pump Off & Remove CellStep 1Wear Protective EquipmentStep 2RemoveCalcium andother debrisfrom the CellInspect cell for calcium deposits. Even if a Cell appearsclean, it may still require cleaning if salt accuracy is offand/or chlorine production has diminished.If the Cell requires cleaning, always wear allrecommended protective equipment such as gloves,apron, and safety goggles.NOTE: MIX SOLUTION AND CLEAN CELL ONLY IN A WELL VENTILATED AREA.MURIATIC AND OTHER ACIDS CAN CAUSE SEVERE INJURY, BURNS AND RESPIRATORY PROBLEMS IFNOT HANDLED PROPERLY. REFER TO THE MANUFACTURER’S DIRECTIONS FOR SAFE HANDLING.11

How To: Clean the Cell (Cont.)The Cell draws amperage when power is applied during chlorination. The amperagedraw will be impaired when calcium and other debris exist within the Cell’selectrolytic grid; this in turn effects the salt reading and chlorination efficiency.Mix: 4 Parts H2O / 1 Part Muriatic AcidStep 3Carefully Pour Solution into CellStep 41 Part:Muriatic Acid4 Parts:WaterUsing an approved container, mix a solution comprisedof 4 parts water to 1 part Muriatic Acid. Always AddAcid to Water.Cleaning Tip:The electronic capof the Cell shouldNOT be submergedin waterPlace Cell into plastic bucket, pour solution into bucket.Solution should NOT cover electronic cap. Once clean,rinse cell, pour solution back into approved container.NOTE: ALWAYS ADD ACID TO WATER! NEVER ADD WATER TO ACID.The cell cleaning solution may be reused a few times.ALWAYS: STORE MIXED SOLUTIONS IN A SAFE AREA, OUT OF HARMS WAY.When the solution is depleted, follow the manufacturer’s instructions for proper disposal.12

AquaRite 100 Troubleshooting

Inadequate Water Flow LEDWith “Standing BY” LEDReason14Possible ActionWith “Problem Detected” LEDReasonPossible ActionPower to the pump isturned “off”Turn power to the pump“on”No water flowdetectedVerify pump is “on”Remove power for two minutesAQR 100 is betweengenerating cyclesWait until next generatingcycle begins or cycle powerto the AquaRite 100Low waterflow detectedCheck for obstructions or other flowrelated issues in plumbing. Verify filter isclean.

Problem Detected: Inadequate FlowThese lights illuminated together indicate Low or No water flow through Cell VesselVerify pump is turned on and runningStep 1AInspect the CellStep 1BVerify the timer andbreaker are in the “ON”positionVerify pump is on and water is flowing. If pump isrunning, remove power to the AQR100 for two minsthen re-apply. If error still does not clear continue tostep 1B.RemoveCalcium andother debrisfrom the CellThe Cell must be clean in order to report an accurate saltreading. If the Cell is dirty, clean the Cell(steps on pg.11-12). If Cell is clean, like the image above,proceed to step 1C.The cleanliness of the Cell plays a major role in the AquaRite 100’s ability to report salt levelsaccurately.15

Problem Detected: Inadequate FlowThese lights illuminated together indicate Low or No water flow through Cell VesselInadequateFlowIs pump turnedon?YESNOYESTurn off pumpand clearObstructionsNOReplace Cellp/n CCELLRemove powerfor 2 mins thenre-applyNOYES*For verification of diagnosis contact Hayward Technical Support908-355-799516Did error clear?NONOTurn pump onDid error clear?Are the Cell andvessel clear ofobstructions?Did error clear?YESProblemsolvedYES

Problem Detected LEDWith LED #2With “MIN” LEDReasonPossible ActionReasonCheck the salt level in the pool is no greaterthan 4500ppmVerify salt level is between 1500 – 4500 ppmAdjust salt level if necessaryLowSaltInspect Cell. If dirty (calcified) then follow theCell cleaning procedureInspect Cell Vessell, verify Cell Plates arecovered by moving waterRemove power to the AquaRite 100 for twominutes, then re-apply to reset system17Possible ActionHighSaltVerify water temperature is no greater than104 FBring water temperature between65 F -104 F if necessaryRemove power to the AquaRite 100 for twominutes, then re-apply to reset system

Problem Detected and MIN LEDWhen these lights illuminate together, the Cell reports salt levels below 1500.Inspect the CellTest the Salt LevelStep 2AStep 2BDesired SaltConcentration:1500-4500 PPMTest the salt concentration of the water using anindependent test, with a calibrated salt meter. If the saltlevel is below 1500, raise the salt level (refer to saltaddition table pg. 38). If fine, go to 2B.RemoveCalcium andother debrisfrom the CellThe Cell must be clean in order to report an accurate saltreading. If the Cell is dirty, clean the Cell(steps on pg.11-12). If Cell is clean, like the image above,proceed to step 2C.The cleanliness of the Cell plays a major role in the AquaRite 100’s ability to report salt levelsaccurately.18

Problem Detected and MIN LEDWhen these lights illuminate together, the Cell reports salt levels below 1500.Verify moving water fully covers Cell platesStep 2CRemove power from the AQR100Step 2DTurn power off for 2mins, then back onInspect Cell Vessel, verify Cell Plates are fully covered bymoving water. If NOT fully covered by moving water,clear obstructions. If error persists, go to step 2D19Remove power to the AQR100 for two minutes.Re-apply power to reset the system. If error does notclear replace the cell (p/n) CCELL).

Problem Detected: MIN LEDWhen these lights illuminate together, the Cell reports salt levels below 1500.MIN LEDFlashing(Low Salt)Is salt levelbetween 15004500ppm?YESNODoes movingwater fully coverCell Plates?YESNONOClean filter orclear plumbingobstructionClean Cell usingprocedures onpg. 11-12Bring salt levelwithin rangeDid error clear?Is Cell clean andfree of debris orcalcium?NODid error clear?NODid error clear?YESYESNORemove powerfor 2 mins thenre-applyProblem solvedReplace Cellp/n CCELL20YESNODid error clear?*For verification of diagnosis contact Hayward Technical Support908-355-7995YES

Problem Detected: LED #2When these lights illuminate together, the Cell reports salt levels above 4500.Test the Salt LevelStep 3AStep 3BDesired SaltConcentration:1500-4500 PPMTest the salt concentration of the water using anindependent test, with a calibrated salt meter. If saltlevel is above 4500, dilute salt level (referring to saltdilution calculation on pg. 39). If fine, go to Step 3B.21Verify water temperature is within rangeDesiredWaterTemperatureis between65 F - 104 FIf water temperature is NOT within the proper range,wait until the water temperature is between 65 F 104 F . If the temperature IS within the proper rangecontinue to step 3C.

Problem Detected: LED #2When these lights illuminate together, the Cell reports salt levels above 4500.Remove power from the AQR100Step 3CTurn to off position for 2mins, then turn back onRemove power to the AQR100 for two minutes. Re-applypower to reset the system. If error does not clear replace theCell (p/n CCELL).22

Problem Detected: LED #2When these lights illuminate together, the Cell reports salt levels above 4500.LED #2Flashing(High Salt)Is salt levelbelow4500ppm?YESIs watertemperaturebelow 104 F?Remove powerfor 2 mins thenre-applyYESNONODilute salt levelto lower than4500Did error clear?YESWait until watertemperature fallsbelow 104 FNODid error clear?YESNONOYESProblem solved*For verification of diagnosis contact Hayward Technical Support908-355-799523Did error clear?Replace Cellp/n CCELL

Problem Detected LEDWith LED #3ReasonLowInputPower24Possible ActionRemove power to AquaRite 100 for twominutes, then re-apply to reset systemVerify input power wiring is configured for thecorrect voltage (pg. of this manual)Verify proper voltage to the AquaRite 100With LED #4Possible ActionReasonTemperatureSensorErrorRemove power to AquaRite 100 for twominutes, then re-apply to reset systemInspect Cell. If dirty (calcified) then followthe Cell cleaning procedureReplace Cell

Problem Detected: LED #3When these lights illuminate together the AQR100 is registering Low Input Power.Remove power from AQR100Step 4AVerify AQR100 WiringStep 4BTurn power off for 2minutes, then back onRemove power from the AQR100 for two minutes.Re-apply power to reset the system. If error does notclear continue to step 4B25Wiring should be configured correctly for the appliedvoltage. Verify proper voltage to AQR100. If errorpersists, replace control (p/n GLXAQR100CTL).

Problem Detected: LED #4These lights illuminated together indicate a Temperature Sensor Error.Inspect CellRemove power from AQR100Step 5AStep 5BRemoveCalcium andother debrisfrom the CellTurn power off for 2minutes, then back onRemove power from the AQR100 for two minutes.Re-apply power to reset the system. If error does notclear continue to step 5B26Cell should be clean & free of debris. If Cell is dirty,clean the Cell.If Cell is clean, replace Cell (p/n CCELL).

Problem Detected: LED #3 or #4LED #3 Flashing(Low inputpower)Remove powerfor 2 mins thenre-applyDid error clear?YESYESNORemove powerfor 2 mins thenre-applyLED #4 Flashing(TemperatureSensor error)Problem solvedVerify inputpower isconfiguredcorrectlyDid error clear?NOYESDid error clear?NOIs Cell clean andfree of debris orcalcium?Replace ControlCenterp/nGLXAQR100CTLProblem solvedYESNOClean Cell usingprocedures onpg. 11-12Did error clear?NOYES*For verification of diagnosis contact Hayward Technical Support908-355-799527Replace Cellp/n CCELL

Problem Detected LEDWith LED #5With LED #6ReasonPossible ActionRemove power to theAquaRite 100 for twominutes, then re-apply toreset systemWith LED #7With LED #9InternalErrorVerify salt level is nogreater than 4500 ppmVerify water temperature isno greater than 104 FInspect Cell Vessel andverify Cell Plates are fullycovered by moving waterReplace Control Center(p/n GLXAQR100CTL)28

Problem Detected: LED #5,6,7, or 9Problem Detected with any one of these LED’s illuminated indicates an Internal Error.Test the Salt LevelRemove power from the AQR100Step 6BStep 6ADesired SaltConcentration:1500-4500 PPMTurn power off for 2mins, then back onRemove power to the AQR100 for two minutes.Re-apply power to reset the system. If error does notclear continue to step 6B.Test salt level of water, using an independent test, with acalibrated salt meter. If salt level is above 4500,dilute salt. If fine, go to Step 6C.* Salt Dilution Formula is found on pg. 39 of this manual29

Problem Detected: LED #5,6,7, or 9Problem Detected with any one of these LED’s illuminated indicates an Internal Error.Verify water temperature iswithin rangeStep 6CVerify moving water fully coverscell platesStep 6DDesiredWaterTemperatureis between65 F - 104 FIf water temperature is NOT within the proper range,wait until it is within range.If the temperature IS within range continue to step 6D.30Inspect Cell vessel and verify Cell Plates are fully coveredby moving water. If issue does not clear, replace theControl Center (p/n GLXAQR100CTL).

Problem Detected: LED #5,6,7, or 9Problem Detected with any one of these LED’s illuminated indicates an Internal Error.LED #5,6,7, or 9Flashing(Internal Error)Remove powerfor 2 mins thenre-applyDid error clear?NOIs salt levelbelow4500ppm?NOYESDilute salt levelto lower than4500Replace ControlCenterp/nGLXAQR100CTLProblem solvedYESNONONODid error clear?YESYESYESClean filter orclear plumbingobstruction31NODoes movingwater fully coverCell Plates?Did error clear?YESNODid error clear?*For verification of diagnosis contact Hayward Technical Support908-355-7995Is watertemperaturebelow 104 F?NOWait until watertemperature fallsbelow 104 FYES

Problem Detected LEDWith LED #8ReasonPossible ActionNoCurrent toCell32With “MAX” LEDReasonRemove power to the AquaRite 100 for twominutes and then re-apply to reset systemTurn on Super Chlorinate briefly then go back tooriginal settingInspect Cell Vessel and verify Cell Plates are fullycovered by moving waterPossible ActionRemove power to the AquaRite 100 for two (2)minutes and then re-apply to reset systemCell ErrorMake sure Cell Cable is secure and free ofdamageIf error persists, replace the Cell(p/n CCELL)*Make sure cell cable is secure and free of damage. If damage is found or cord does not fitsecurely onto cell contact Hayward Technical Support 908-355-7995

Problem Detected: LED #8These lights illuminated together indicate there is no current to the Cell.Super ChlorinateRemove power from the AQR100Step 7BStep 7ATurn power off for 2mins, then back onRemove power to the AQR100 for two minutes.Re-apply power to reset the system. If error does notclear continue to step 7B.33Press Super Chlorinate button to force generating cycleto begin. After a few moments press button again toturn it off. If error does not clear continue to step 7C.

Problem Detected: LED #8These lights illuminated together indicate there is no current to the Cell.Verify moving water fully covers cell platesInspect Cell cordStep 7DStep 7CInspect Cell Cordfor damage orcorrosionInspect Cell Vessel and verify Cell Plates are fullycovered by moving water. Clear Cell of any obstructions.If problem persists, continue to step 7D.Verify Cell Cordis securelyfastened to CellCell Cord must be connected securely and be free ofdamage. If error does not clear replace the ControlCenter (p/n GLXAQR100CTL).*Make sure cell cable is secure and free of damage. If damage is found or cord does not fitsecurely onto cell contact Hayward Technical Support 908-355-799534

Problem Detected: MAX LEDThese lights illuminated together indicate there is no current to the Cell.Inspect Cell cordRemove power from the AQR100Step 8AStep 8BInspect Cell Cordfor damage orcorrosionVerify Cell Cordis securelyfastened to CellTurn power off for 2mins, then back onRemove power to the AQR100 for two minutes.Re-apply power to reset the system. If error does notclear continue to step 8B.Cell Cord must be connected securely and be free ofdamage. If error does not clear replace the ControlCenter (p/n GLXAQR100CTL).*Make sure cell cable is secure and free of damage. If damage is found or cord does not fitsecurely onto cell contact Hayward Technical Support 908-355-799535

Problem Detected: LED #8 or MAXBriefly turn onSuper Chlorinatethen turn backoffLED #8 Flashing(Low or no flowcondition)Did error clear?YESNORemove powerfor 2 mins thenre-applyDid error clear?YESNODoes movingwater fully coverCell Plates?NOProblem solved(WaterTemperatureerror)YESProblem solvedDid error clear?Remove powerfor 2 mins thenre-applyDid error clear?NOYESReplace ControlCenterp/nGLXAQR100CTLNOYESDid error clear?YESYESIs Cell Cablesecure and freeof damage?YESNOIs Cell Cablesecure and freeof damage?NONOReplace Cellp/n CCELLNO36NOClean filter orclear obstructionin plumbingYESMAX FlashingDid error clear?YES*For verification of diagnosis contact Hayward Technical Support908-355-7995NO

AquaRite 100 Additional Information

Salt Addition Chart: lbs. required for 3200ppmCurrentSalt 260028003000320034003600 Pool Size - DiluteNote: Prior to adding salt, always test water with independent tests to determine current salt and stabilizer levels.How to add saltBrushing the salt around will speed up the dissolving process. DO NOT allow the salt to sit in a pile atthe bottom of the pool. Salt water is heavier than fresh water, so the salt water will tend to accumulateat the deepest part of the pool. Run the filter system with the suction coming from the main drain for24 hours to evenly distribute the salt throughout the poolNote: Refer to the Plasters recommendations for cure time before adding salt.38

Salt Dilution FormulaTo calculate how much water will need to be removed from a pool, with too muchsalt, follow the formula provided below (Proportional Method):Part I(Ave. Pool Depth” X 3200)Actual Salt level in PoolPart 1: Take the average depth of the pool in inches and multiply that by3200 (or desired salt level within 1500-4500ppm). Then divide thatnumber by the actual salt level in the pool (based on independent test).Part 2: Subtract the Average Pool Depth by the answer from part 1. Thiswill give you the total number of inches to drain and replenish with freshwater to achieve a salt level of to match desired salt level from step 1.Part IIAve. Pool Depth - Answer from Part 1 Amount of Water to DrainExample: a pool has an ave. depth of 54” and the salt level is 4500ppmExample: Part IExample: Part II54 X 3200 172800450038.454 – 38.4 15.6”Note: It is recommended to reduce the water level no more than six inches at a time beforereplenishing with fresh water. Failure to due so may result in damage to the pool structure orsurface.39

Replacement Part NumbersReplacement Part Number40DescriptionCCELLReplacement Cell OnlyGLXAQR100CCABLECell CableGLXAQR100PCORDPower CordGLXAQR100CTLAQR 100 Control Center OnlyGLXAQR100DOORDoor Flap for DisplayGLXAQR100PCBMain PCBGLXAQR100VESNUTVessel/ Cell NutGLXAQR100XFMRTransformerSP3200UNKITVessel Unions Kit

Reading Serial NumbersBeginning 08-05-20113A11284- 1234563A11284 Standard Warranty Term3A11284 Product Family3A11284 Year of Manufacture3A11284 Day of Manufacture123456 Manufacturing IDBeginning 11-01-20013A0112- 1234563A0112 Standard Warranty Term3A0112 Product Family3A0112 Year of Manufacture3A0112 Month of Manufacture123456 Manufacturing ID41

Troubleshooting: Pg. 13-36 . 1. Inadequate Water Flow LED 2. Problem Detected with MIN LED or LED #2 3. Problem Detected with LED # 3 or LED #4 4. Problem Detected with LED # 5,6,7, or 9 5. Problem Detected with LED # 8 or MAX LED 14-16 . 17-23 . 24-27 ; 28-31 : 32-36 ; Additional Inf

Related Documents:

The AquaRite is a system that controls the equipment and it is used in conjunction with a salt chlorinator for treating swimming pools. The AquaRite allows you to control the filtration system (pump) as well as the peripheral equipment (heat pump, lighting, etc.). It can also be used to treat your pool effectively by salt water electrolysis .

100%: 100%. 100%: 100%. 100%: 100%. 100%: 100%. 100%: 100%. 100%: 100%. 100%: 100%. 100%: 2. Plain Cement Concrete: 100%. 100%: 100%. 100%: 100%. 100%: 100%. 100% .

UOB Plaza 1 Victoria Theatre and Victoria Concert Hall Jewel @ Buangkok . Floral Spring @ Yishun Golden Carnation Hedges Park One Balmoral 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 101 101 101 101 101 101 101 101 101. BCA GREEN MARK AWARD FOR BUILDINGS Punggol Parcvista . Mr Russell Cole aruP singaPorE PtE ltd Mr Tay Leng .

30 100 20 100 20 100 6 60 20 100 10 100 B1 W2 85 560 4RD 933 332-031 Dust and waterproof** 30 100 20 100 20 100 6 60 20 10 10 100 B1 W1 85 - 4RD 933 332-237* 30 100 20 100 20 100 6 60 20 100 10 100 B1 W3 85 - 4RD 933 332-277* with parallel diode Nominal switching current (A)! No. of switchings (thous.)

7 Shade 50% of the whole figure. 8 Shade 75% of the whole figure. Fill in each blank. 9 43 out of 100 % 10 out of 100 1% 11 5 out of 100 % 12 out of 100 10% 13 90 out of 100 % 14 out of 100 87% 15 21 out of 100 % 16 out of 100 2% 17 8 out of 100 % 18 out of 100 3% 19 4 out of 100 % 20 out of 100 9% 21 35 out of 100

S-600PR-WH 100-130V 0-100% Forward Phase MRF2-6ND-120-AL 100-130V 0-100% Forward Phase MRF2-6CL-GR 100-130V 1-100% Forward Phase DZ6HD 100-130V 3-100% Forward Phase PD-6WCL 100-130V 1-100% Forward Phase SELV-300P 100-130V 1-100

S-600PR-WH 100-130V 0-100% Forward Phase MRF2-6ND-120-AL 100-130V 0-100% Forward Phase MRF2-6CL-GR 100-130V 1-100% Forward Phase DZ6HD 100-130V 3-100% Forward Phase PD-6WCL 100-130V 1-100% Forward Phase SELV-300P 100-130V 1-100

tacacs-server host 123.100.100.186 port 49 key lm51! tacacs-server host 123.100.100.187 port 49 key lm51! aaa group server tacacs tacgrp server 123.100.100.186 server 123.100.100.187! aaa group server tacacs eem server 123.100.100.186 server 123.100.100.187! aaa authorization exec tacauthen group tacgrp local