ERBE - Frank's Hospital Workshop

2y ago
25 Views
3 Downloads
1.40 MB
72 Pages
Last View : 7d ago
Last Download : 3m ago
Upload by : Amalia Wilborn
Transcription

ERBEERBOTOM ICC 200Instruction manual04.05

ERBOTOM ICC 200 V 2.X10128-002, 10128-009, 10128-303, 10128-027, 10128-403ERBOTOM ICC 200 E V 2.X10128-010, 10128-015, 10128-304,10128-028ERBOTOM ICC 200 EAV 2.X10128-023, 10128-036, 10128-305, 10128-058, 10128-400Instruction manual04.05

EN ISO 13485 EN ISO 9001All rights to this instruction manual, particularly the right to reproduction, distribution and translation, are reserved.No part of this instruction manual may be reproduced in any form (including photocopying, microfilm or othermeans), or processed, reproduced or distributed by means of electronic systems without prior written permissionfrom ERBE ELEKTROMEDIZIN GmbH.The information contained in this instruction manual may be revised or extended without prior notice and representsno obligation on the part of ERBE ELEKTROMEDIZIN GmbH. ERBE ELEKTROMEDIZIN GmbH, Tübingen 2005Printed by: ERBE ELEKTROMEDIZIN, TübingenInstruction manual no. 80104-401

Chapter1TitlePage1.2INTRODUCTION . 1-1Intended purpose of the ICC 200 . 1-1Explanation of the safety instructions . 1-12INITIAL OPERATION . 2-13RISKS AND SAFETYOF HIGH-FREQUENCY SURGERY. 3-1Unintentional thermal tissue damage . 3-1- due to HF leakage currents . 3-1- due to unintentional activation of an HF generator . 3-2- due to inappropriate application . 3-3- due to inappropriate or nonapplication ofthe neutral electrode . 3-3- due to unsuitable and/or faulty accessories . 3-4- due to inattentiveness . 3-5- due to an output error . 3-5- due to the ignition of flammable liquids,gases and/or vapors . 3-5Unintentional burns due to hot electrodes . 3-6Electric shock . 3-6Stimulation of nerves and muscles . 3-6Cardiac pacemaker . 3-7Danger of explosion . 3-7Interference with other electronic equipment . 3-7Portable and mobile communication equipment HF . .1.93.23.33.43.53.63.744.14.2DESCRIPTION OF THE HIGH-FREQUENCYSURGICAL UNIT . 4-1General description . 4-1Description of the controls . 4-3123456789101112134.3Power switch . 4-3AUTO CUT function field . 4-4AUTO COAG function field . 4-5Connecting socket for neutral electrodes . 4-6Connecting socket for monopolar cutting or coagulation instruments . 4-6Connecting socket for bipolar coagulation instruments . 4-7Safety field . 4-7Connecting socket for a dual-pedal footswitch . 4-7Terminal for potential equalization . 4-7Volume of the acoustic signal . 4-7Loudspeaker for the acoustic signals . 4-7Power connection . 4-7Power fuses . 4-7Description of the safety features . 4-855.15.25.3TECHNICAL DATA, SIGNALS, DIAGRAMS . 5-1Technical data . 5-1Visual and acoustic signals . 5-4Diagrams . 5-5

6INSTALLATION . 6-17CLEANING AND DISINFECTION OF THE UNIT . 7-188.18.28.3PERFORMANCE CHECKS . 8-1Automatic performance test after switching on the unit . 8-1Automatic performance check during activation . 8-1Automatic error documentation . 8-2Error list . 8-39SAFETY CHECKS . 9-110MAINTENANCE, CARE, DISPOSAL . 10-111GUARANTEE . 11-112INFORMATION ONELECTROMAGNETIC COMPATIBILITY (EMC) . 12-1ADDRESSES

1INTRODUCTION1.1Intended purpose of the ICC 200The ICC 200 is a high-frequency surgical unit for cutting and coagulation. The ICC 200 isavailable in the variations Basic Model, Basic Model with ENDO CUT (ICC 200 E), BasicModel with ENDO CUT and Argon Plasma Coagulation (ICC 200 EA).1.2Explanation of the safety instructionsThe WARNING! safety instruction indicates a danger which can result in personal injury.The CAUTION! safety instruction indicates a danger which can result in property damage.The IMPORTANT safety instruction indicates a danger which can cause functional failure of theunit.1-1

1-2

2INITIAL OPERATIONRead carefully before initial operation of the unit.In the development and production of this high-frequency surgical unit, the relevant, generallyrecognized rules of technology, as well as the valid occupational safety and accident preventionregulations have been taken into consideration. This ensures that patients, employees and thirdparties are protected from dangers to life and health during intended application of the highfrequency surgical unit, to the extent permitted by the type of application intended.Initial operationBefore delivery, every high-frequency surgical unit is tested by the manufacturer in regard to itsfunction and safety. To ensure that the unit also functions safely after shipping and installation atthe operator’s site, the following points should be observed:The operator should only operate the high-frequency surgical unit if the manufacturer or supplier1.has subjected the unit to a performance test on site2.has instructed the parties responsible for operation of the unit in handling of the unit bymeans of the instruction manual.2-1

2-2

3RISKS AND SAFETY OF HIGH-FREQUENCY SURGERY3.1Unintentional thermal tissue damageHigh-frequency surgery is associated in principle with various risks for the patient, the personneland surroundings. In order to avoid these risks in practice, the surgeon and his/her assistantsmust recognize these risks and observe the appropriate rules for prevention of damage. In thefollowing, these risks and rules for prevention of damage are explained.3.1.1 Unintentional thermal tissue damage due to HF leakage currentsDuring high-frequency surgery, the patient unavoidably conducts high-frequency electrical currentto ground potential. If the patient makes contact with electrically conductive objects duringhigh-frequency surgery, a high-frequency electrical current can result at the contact point betweenthe patient and this object, which can in turn cause thermal necroses. Not just objects made ofmetal are electrically conductive objects, but also wet cloths.WARNING!The patient must be insulated against electrically conductive objects during high-frequencysurgery. The black elastic table covers on operating tables demonstrate a certain electricalconductivity for diverting electrical charges. Therefore they are never suitable for ensuring therequired insulation of the patient against metal parts of the operating table. For this reason, anelectrically insulating intermediate layer, for example dry cover cloths, must be laid between thepatient and this black operating table cover during the application of high-frequency surgery.Grounded operating tableElectrically insulatedsurfaceFig.:Insulated positioning of the patient on the operating tableIf it is possible for this intermediate layer to become wet during the operation, for example dueto perspiration, irrigation liquid, urine etc., wetting of these intermediate layers must be preventedby a watertight sheet of plastic. Urine should be carried away via catheter.ŸExtremities lying against the trunk or skin-to-skin contact points should be insulated fromone another by laying dry cover cloths between them.ŸDo not apply ECG electrodes closer than 15 cm next to the operating field.ŸNeedle electrodes or injection cannulae should not be used as ECG electrodes duringhigh-frequency surgery.3-1

3.1.2 Unintentional activation of an HF generatorUnintentional activation of an HF generator can lead to burns on the patient if the active electrodehereby touches the patient directly or indirectly through electrically conductive objects or wetcloths.Unintentional activation of an HF generator can, for example, be caused by:ŸUnintentionally pressing a footswitch pedalŸUnintentionally pressing a fingerswitchŸDefective fingerswitches, footswitches or cablesŸPenetration of electrically conductive liquids (blood, amniotic fluid, urine, physiologicalsaline solution, irrigation fluids etc.) into fingerswitches or footswitches.ŸErrors within the high-frequency surgical unitWARNING !To prevent burns on the patient due to unintentional activation of a high-frequency generator, thefollowing application rules should be heeded:ŸNever lay active electrodes onto or beside a patient in such a way that they can touch thepatient directly or indirectly through electrically conductive objects or wet cloths.ŸThe lines to the active electrodes should be positioned in such a way that they touchneither the patient nor other lines.ŸAlways set the acoustic signal, which indicates the active status of the high-frequencygenerator, so that it can be easily heard.ŸFor operations in which the cutting or coagulation electrode unavoidably remains in contactwith the patient even in a nonactive condition, e.g. for endoscopic operations, particularcare is required. If such an electrode is unintentionally activated due to an error, this activatedelectrode should then not be removed from the body without special supervision. Whenremoving the activated electrode from the patient’s body, burns can result on all areaswithin the body which come into contact with the activated electrode. For this reason, incase such errors occur, the power switch for the high-frequency surgical unit should beswitched off immediately before an attempt is made to remove the activated electrodefrom the body.3-2

3.1.3 Unintentional thermal tissue damage due to inappropriate applicationGenerally speaking, the bipolar coagulation technique should be applied in preference to themonopolar coagulation technique. This particularly applies to coagulations on straight organs,on which the high-frequency current flows over longer areas through diameters which areapproximately equal or become even smaller.Fig.: Thermal damage of lateral tissueThe tissue is always first heated at places on the tissue where the diameter is smallest. If the HF currentflows through the same diameter (a) over longer distances, the tissue coagulates over this entire distance.If the diameter of the tissue next to the application point of the coagulation electrode is smaller than at thepoint of application, coagulation will also occur next to the application point (b).WARNING!Always make certain that the HF current does not flow through thin tissue structures or vesselswith a small diameter.3.1.4 Unintentional thermal tissue damage due to inappropriate ornonapplication of the neutral electrodeWith inappropriate or even nonapplication of the neutral electrode, there is a large risk ofunintentional thermal tissue damage both at the application point of the neutral electrode as wellas to other areas on the patient’s body.The neutral electrode must be applied with its entire surface as closely and reliably as possible tothe operating field on the patient’s body.3-3

WARNING!The effective contact surface, i.e. the electrical conductive value between the neutral electrodeand the patient must correspond to the HF capacity used, meaning the intensity of the HF current.Here the effective contact surface means the surface of the neutral electrode which has electricallyconductive contact to the skin of the patient during high-frequency surgery.a)b)Fig.: The neutral electrode must be applied at an appropriate location on the patient’s skin using theentire contact surface available (a). If the neutral electrode has only partial contact to the patient’s skin (b),there is a risk that burning will occur at this location3.1.5 Unintentional thermal tissue damage due to unsuitable and/or faultyaccessoriesIt must be ensured that only accessories in perfect condition are used for high-frequency surgery.Only accessories that are compatible or tested by the unit manufacturer must be used. Thisapplies both to the active electrodes including cable and plugs, as well as to the neutral electrodesincluding cables and plugs.When using an instrument with electric insulation, it is necessary to be certain that these insulationsare not overloaded and destroyed by overly high electric voltages. The electric output voltagesfor the high-frequency surgical unit are indicated for the various cutting and coagulation modesrelative to the possible settings in this instruction manual. The electric strength of the instrumentinsulation can be found in the technical data for the instruments or, in case of doubt, can berequested from the manufacturer of the respective instrument.WARNING!All insulation on electrodes, electrode holders, cables, plugs etc. must be in perfect condition.3-4

3.1.6 Unintentional thermal tissue damage due to inattentivenessLike a scalpel, high-frequency surgery is always a potential source of danger if handled withoutcare.WARNING!The cutting or coagulation electrodes should always be handled with care and laid aside in theintervals between use so that neither the patient nor other persons can come in contact with theelectrodes.Laying unused electrode handles or coagulation forceps on the patient, next to the patient orwithin folds on the cover cloths is dangerous. Cases of burns on patients are known which werecaused by laying the coagulation forceps within folds on the cover cloths which penetratedthrough the cloths into the patient’s skin and resulted in burns without being noticed.3.1.7 Unintentional thermal tissue damage due to output errorThe risk of unintentional thermal tissue damage is proportionate to the intensity and time limitset on the unit for cutting or coagulation.WARNING!The intensity for cutting or coagulation should only be set and only activated for as long asnecessary for the intended purpose.An insufficient effect at a standard setting can, for example, be caused by poor attachment of theneutral electrode, poor contact in the connectors, defective cables or electrically insulating tissueremnants on the active electrode. This must be checked before setting at a higher power.3.1.8 Unintentional thermal tissue damage due to the ignition of flammableliquids, gases and/or vaporsDuring high-frequency surgery, electric sparks or arcs that can ignite flammable liquids, gasesor vapors occur at the active electrode.3-5

WARNING!Make certain during high-frequency surgical operations that anesthetics, skin cleaning agentsand disinfectants are nonflammable. If their use is unavoidable, they must have completelyevaporated and the vapor must be removed from the area of spark formation before switching onthe high-frequency surgical unit.Before application of high-frequency surgery in the gastro-intestinal tract, it must be ensuredthat no flammable (endogenous) gases are present here. There is danger of explosion if flammablegases are present. For this reason, these gases must be extracted and/or eliminated by flushingout the affected lumen with CO2before using high-frequency surgery.During transurethral resection (TUR), H2O molecules may dissociate into H2 and O2 in the arcbetween the resection loop and the irrigation liquid. These gases may collect on the roof of theurinary bladder as a highly explosive gas mixture. If resection is performed in this gas mixture,dangerous explosions may occur.3.1.9 Unintentional burns due to hot electrodesCutting and/or coagulation electrodes become hot during cutting and/or coagulation proceduresindirectly through the heated tissue and through the electric arc.WARNING!Tissue can be unintentionally burnt immediately after cutting and/or coagulation procedures ifelectrodes that are still hot touch the tissue. Attention must be especially paid to this duringendoscopic operations, such as during pelviscopic fallopian tube coagulation or during endoscopicpolypectomy.3.2Electric shockAn electric shock may occur if the high-frequency surgical unit delivers a too heavy low-frequencycurrent or if a too heavy low-frequency current flows through the patient into the high-frequencysurgical unit from another voltage source.3.3Stimulation of nerves and musclesA known risk of high-frequency surgery is the unintentional electric stimulation of the patient’snerves and muscles. This stimulation can result from low-frequency electrical currents that arecaused either by low-frequency current sources or due to electrical arcs between an active electrodeand the patient’s tissue.Electric alternating current with a frequency above 300 kHz is unable to stimulate nerves andmuscles.3-6

During cutting procedures, forced coagulation and spray coagulation, the unavoidable electricarcs between an active electrode and the tissue nevertheless have the effect that a portion of thehigh-frequency alternating current is rectified, from which more or less strongly modulated,low-frequency current components result which stimulate electrically stimulable structures suchas nerves and muscles.This can result in more or less strong spasms or muscle contractions.WARNING!When using high-frequency surgery on electrically stimulable structures, contractions of theaffected muscles must be taken into account. This can occur, for example, during endoscopicoperations in the urinary bladder in the vicinity of the obturator nerve and during operations inthe area of the facial nerve.3.4Cardiac pacemakerFor patients with implanted cardiac pacemakers or pacemaker electrodes, irreparable damage tothe pacemaker and disturbance of the pacemaker function, which can lead to ventricular fibrilation,must be reckoned with.3.5Danger of explosionHigh-frequency surgical units always generate sparks during operation on the active electrode.For this reason, it is necessary to make certain during interventions that anesthetics, degreasersand disinfectants are neither flammable nor explosive. They should at least have evaporatedcompletely before switching on the high-frequency surgical unit and be removed from the areaof spark formation.3.6Interference with other electronic equipmentHigh-frequency surgical units normally generate high-frequency electrical voltages and currentswhich can interfere with other electronic equipment.When installing or arranging sensitive electronic equipment in the operating room, this problemshould be taken into consideration. In principle, sensitive electronic equipment should be set upas far as possible from the high-frequency surgical unit and particularly from the cables providingHF current. In addition, the cables providing HF current, which act like broadcast antennas,should not be unnecessarily long and should never be positioned parallel or too close to cablesfrom sensitive electronic equipment.The unit has been fitted with a special generator in consideration of the disturbance of sensitiveelectronic equipment, which generates a relatively low interference level as compared toconventional high-frequency surgical units.3.7Portable and mobile communication equipment HFATTENTION: Portable and mobile communication equipment HF can influence the device.3-7

3-8

StandardInternational

4DESCRIPTION OF THE HIGH-FREQUENCY SURGICAL UNIT4.1General descriptionCutting with automatic control of the HF voltage (Auto Cut)The ERBOTOM ICC is equipped with automatic open and closed loop control systems whichcontrol and regulate the parameters relevant to the cutting quality so that each respectively selectedcutting quality is guaranteed to be reproducible and constant.Adjustable power limitation in the cutting modeSince the ICC units are equipped with automatic control of the HF voltage in the cutting mode,a power setting in regard to cutting quality is not required. The adjustable power limitation isprimarily intended to guarantee the safety of the patient from unintentional thermal tissue damage,and to protect fine cutting instruments, such as fine needle electrodes, from destruction due tooverly high HF currents if these come in contact in activated condition with other metallicinstruments. The latter, for example, is a risk during laparoscopic operations. This adjustablepower limitation must not be confused with the power setting for conventional high-frequencysurgical units, where the cutting quality is directly dependent upon the power setting.PPS (Power Peak System)The initial incision phase can represent a special problem during an incision, particularly if thecutting electrode is firmly pressed against the tissue to be cut before activating the HF generator,so that the cutting electrode has a relatively large-surface, and therefore low-resistance, contact.This is the case, for example, for TUR and for endoscopic polypectomy. In such cases, the HFgenerator must provide higher-than-normal power so that the initial incision can proceed withoutdelay, for otherwise a very large coagulation necrosis may result at the cutting location. The ICCis equipped with an automatic power control which recognizes low-resistance loads and controlsthe HF generator in such a way that it briefly provides enough power so that the HF voltage, i.e.the intensity of the electric arc, required for the set cutting quality is ensured even for a lowresistance load. Thanks to this device, the average power can be limited to relatively smallquantities, which corresponds to an improvement in protection from unintentional thermal tissuedamage.ENDO CUTA further special problem during endoscopic operations, for example during polypectomy andpapillotomy, consists in the fact that the electrodes used for cutting, for example polypectomyloops and papillotomes, must be guided on long wire pulls through narrow working channels onflexible endoscopes, and therefore the operator has no direct control over the cutting procedure.However, particularly for polypectomy and papillotomy, a controlled incision is a requirementin preventing complications. An incision that is too fast can lead to bleeding of the cut edges dueto lack of sufficient coagulation. An incision that is too slow can cause thermal damage, forexample to the intestinal wall.For units equipped with a special cutting control (ENDO CUT), fractionation automaticallyoccurs in such a way through the incision that alternating short, automatically arc-controlledcutting intervals with defined pause intervals result. In this way, for example, a polypectomyloop cannot cut through a polyp at just any speed. The cutting speed and the degree of coagulationfor the cut edges is more uniform. ENDO CUT is additionally supported by PPS.It is activated by footswitch.4-1

Soft coagulationSoft coagulation can be activated by key or pedal.Forced coagulationForced coagulation is advantageous if an efficient hemostasis is to be achieved with relativelysmall-surface electrodes, such as TUR resection loops.Adjustable power limitation in the various coagulation modesFor the ERBOTOM ICC units, the surgically relevant coagulation qualities, i.e. the coagulationeffects Soft Coag., Forced Coag. and Bipolar Coag., are delimited by definition from one anotherand selectable by the press of a key. Nevertheless, the intensity of the different effects can bevaried by power limitation.Operating mode for Argon Gas CoagulationFor units equipped with the Argon Coag operating mode, the ERBOTOM ICC 200 supplies apulsating HF voltage with peak values up to 4,000 VP for Argon Gas Coagulation.Bipolar coagulationIn this coagulation mode, the HF voltage is, similar to Soft coagulation, automatically andconstantly controlled, and its peak value always remains lower than 200 Vp, so that the currentdensity and thus the coagulation effect is, for the most part, independent of the effective contactsurface between the coagulation electrode and the tissue, provided the effective contact surfaceis not too large relative to the currently set power limitation.The adjustable power limitation serves the purpose of protecting fine bipolar coagulationinstruments, such as pointed bipolar coagulation forceps, from being thermally destroyed incase of a short between the two forcep tips.The footswitch or Auto Start is used for activation.In the Auto Start mode, the HF generator is automatically activated if both poles of the bipolarcoagulation instruments used contact electrically conductive tissue simultaneously. Auto Startcan occur either immediately at the moment of contact with the tissue or more or less temporallydelayed. With immediate activation, it is possible to work very quickly, especially if severalcoagulations must be performed one after another. Delayed activation offers the operator theadvantage that he/she can prepare and securely grasp the tissue to be coagulated with the bipolarcoagulation forceps before the HF generator is automatically activated. Automatic activation ofthe HF generator only occurs if both poles of the bipolar coagulation instruments contact thetissue uninterruptedly for at least as long as the respectively selected delay lasts. If the contact isinterrupted before the respectively selected period of delay is over, the respective period of delayrestarts as of the next contact.4-2

4.2Description of the controlsThis symbol, in accordance with EN 60 601-1, is intended to indicate to the userthat this unit must only be used on the patient if the user is acquainted with theoperation and features of this unit.The figures set in cursive relate to the ICC illustration for this chapter, or to the functionfields in the text.1Power switchUsing this power switch, the unit is switched on and off.Each time after being switched on, the unit automatically proceeds with various performancechecks. If an error in the unit or in the accessories is recognized here, a warning signalsounds and the determined error is indicated by a corresponding error number. (See Chapter8.1, Automatic performance checks after switching on the unit). If no error is determined,the unit is ready to operate.If the unit was switched off for less than approx. 15 seconds, all settings for the programused before switching off appear after the automatic performance check on the front panel,and the unit can be immediately reactivated. This is advantageous if, for example, thepower supply briefly fails.If the unit was switched off for longer than approx. 15 seconds, the basic setting of theprogram used before switching off appears after the automatic performance check on thefront panel, whereby all relevant visual displays continue to blink and the unit cannot beactivated until any key on the front panel is briefly pressed as confirmation that this programshould be used. Then the relevant displays are continously illuminated and the unit can beactivated using the available settings. These settings can be changed or adapted to thecurrent requirements at any time. However, other programs can be selected as well.2 - 3 Function fieldsThe AUTO CUT and AUTO COAG function fields can be adjusted separately from oneanother, although not activated simultaneously for reasons of safety.WARNING!Function fields that are not used may be switched off completely to prevent unintentionalactivation. To do this, the power limitation must be set down so far in the corresponding functionfield until a beep is heard and “ —“ appears on the digital display. The corresponding functionfield cannot be activated in this condition.4-3

2AUTO CUT function fieldAll parameters can be set in this function field that are relevantto cutting:2.1 Setting of the coagulation EFFECT when cuttingHere the required cutting quality in regard to the coagulation effect on the cutting edgescan be adjusted.Level 1 corresponds to minimum coagulation effect.Level 4 corresponds to maximum coagulation effect.2.2 Setting the power limitationThe HF power output can be limited in one watt

1.1 Intended purpose of the ICC 200 The ICC 200 is a high-frequency surgical unit for cutting and coagulation. The ICC 200 is available in the variations Basic Model, Basic Model with ENDO CUT (ICC 200 E), Basic Model with ENDO CUT and Argon Plasma Coagulation (ICC 200 EA). 1.2 Explanation of the safety instructions The WARNING!File Size: 1MBPage Count: 72

Related Documents:

Conmed System 7550 / 7500 with ABC Conmed System 5000 Conmed System 2500 / 2450 Conmed Excalibur Plus PC Conmed Sabre 2400 Conmed Sabre 180 Conmed 1000 SES Smoke Evacuator ERBE VIO ERBE ICC 350 ERBE ICC 200 Codman Malis CMC III Codman Malis CMC II BLANKET WARMERS Axia DWC 36 / 36G / 24 / 24G (new)

o ERBE ICC 200 o ERBE VIO 300S o ERBE VIO 200D AXIOS Stent Design Bi-flange or double anchor for Staged and Precise positioning Flange/anchor designed to: hold tissue layers in apposition prevent migration MR Conditional Provided sterile, for single-patient use AXIOS Stent Lumen Large stent lumen diameter and short flow .

Anne Frank (1992) Le monde de Anne Frank (1990) Anne Frank, les sept derniers mois (1989) Journal (1986) Anne Frank in the world, 1929-1945 (1985) Anne Frank (1983) Vérité historique ou vérité politique ? (1980) Documents multimédia (3) Mallette Anne Frank (2010) Le journal d'Anne Franck (2000)

XSEDE HPC Monthly Workshop Schedule January 21 HPC Monthly Workshop: OpenMP February 19-20 HPC Monthly Workshop: Big Data March 3 HPC Monthly Workshop: OpenACC April 7-8 HPC Monthly Workshop: Big Data May 5-6 HPC Monthly Workshop: MPI June 2-5 Summer Boot Camp August 4-5 HPC Monthly Workshop: Big Data September 1-2 HPC Monthly Workshop: MPI October 6-7 HPC Monthly Workshop: Big Data

23 Eastman Dental Hospital 24 Royal National Throat, Nose & Ear Hospital 25 The Nuffield Hearing and Speech Centre 26 Moorfields Eye Hospital 27 St. Bartholomew's Hospital 28 London Bridge Hospital 29 Guy's Hospital 30 Churchill Clinic 31 St. Thomas' Hospital 32 Gordon Hospital 33 The Lister Hospital 34 Royal Hospital Chelsea 35 Charter .

ISO 5145 no. 17 Gas bottle adapter 20410-003 20402-060 20410-016 20410-011 20402-061 Switch valve for connecting two gas cylinder adapters 20402-052 Gas bottle adapter CO 2. erbe-medom 05 No. 20402-410 5 30402-406 STERILE EO No. 20402-411 5

Conmed System 7550 / 7500 with ABC Conmed System 5000 Conmed System 2500 / 2450 Conmed Excalibur Plus PC Conmed Sabre 2400 Conmed Sabre 180 Conmed 1000 SES Smoke Evacuator ERBE ICC 350 ERBE ICC 200 Codman Malis CMC III Codman Malis CMC II Bovie Aaron 2250 Bovie Aaron 1250 BLANKET WARMERS

to the first generation ERBE system (i.e., APC Model 300 /ESU ICC Model) mode, which was introduced globally in 1992 (USA 1996). Characteristics High-frequency voltage increases when the output setting (i.e., wattage) is increased (Fig. 3) High-frequency output can be adjusted to