PC-1000/Laser 1000 Service ManualPanoramic Corporation Dental Panoramic/Cephalometric X-ray MachineSM6000 Rev C Copyright 2008 Panoramic Corporation

Table of ContentsTable of Contents.2-4Introduction .5-6Purpose . 5Statement of Compatibility . 5Limited Warranty . 6Voltage Regulator Warning . 6X-Ray Shielding Requirements . 6Pre-Installation Check . 7Electrical RequirementsControl Panel OrientationCephalometric Arm OrientationRemote SwitchDarkroom RequirementsPC-1000 InstallationPC-1000 Components . 8PC-1000 Installation .9-17Tools Required. 9Verify Power . 9Remove Packaging and Shipping Restraints . 9Control Panel Relocation . 9Install Side Covers. 10Install Film Drum Assembly . 11Install Tubehead Assembly . 12Install Temple Supports . 12Level Machine . 12Perform mA Calibration . 13Perform Pulse Count Calibration .14-15Perform X-ray Beam Alignment . 16Install Top and Rear Covers . 17Complete Installation . 17PC-1000/Laser 1000 InstallationPC-1000/Laser 1000 Components . 18PC-1000/Laser 1000 Installation.19-39Tools Required. 19Verify Power . 19Remove Packaging and Shipping Restraints . 19Control Panel Relocation . 19Install Side Covers. 20Install Film Drum Assembly . 21Install Tubehead Assembly . 22Install Temple Supports . 22Level Machine . 22Install Cephalometric Arm. 23Install Cephalostat Head Positioner . 23Perform mA Calibration .24-252 Copyright 2008 Panoramic Corporation

Table of ContentsPC-1000/Laser 1000 Installation (continued)Perform Pulse Count Calibration .26-27Perform Panoramic X-ray Beam Alignment . 28Remove Soft Tissue Shields. 29Perform Cephalometric X-ray Beam Alignment .29-33Perform Ear Rod Alignment .34-36Install Soft Tissue Shield . 37Install Suggested Settings Decal . 37Install Top and Rear Covers . 37Complete Installation . 38Notes . 39Service ProceduresRemote Switch Installation . 40Screw Motor Height Limit . 41Screw Motor Assembly Exchange .42-43Panoramic Radiography . Appendix ADarkroom Procedures .Appendix BManual ProcessingAutomatic ProcessingDarkroom Light Leak TestLoading The Panoramic CassetteLoading The Cephalometric CassetteMaintenance Schedule . Appendix CPC-1000/Laser 1000 Labeling . Appendix DPC-1000/Laser 1000 Specifications .Appendix EPC-1000/Laser 1000 Space Requirements .Appendix FPC-1000/Laser 1000 Parts List . Appendix GFinal AssemblyBase AssemblyMain Chassis AssemblyChinrest Arm AssemblyHead Support AssemblyFilm Drum AssemblyUpper Belt Drive AssemblyBelt Drive AssemblyRoller Orbit Drive AssemblyRotating Arm AssemblyScrew Motor AssemblyTubehead AssemblyElectronics AssemblyControl Panel Assembly Copyright 2008 Panoramic Corporation3

Table of ContentsPC-1000/Laser 1000 Parts List (continued) . Appendix GLaser Collimator AssemblyElectromagnet AssemblyPC-1000/Laser 1000 Schematics . Appendix HSystem WiringControl PanelTimer Circuit BoardX-ray ControlLaser Power/Control4 Copyright 2008 Panoramic Corporation

IntroductionPurposePanoramic Corporation provides this printed manual as a guide for the operation of the PC-1000 dentalpanoramic X-ray machine and the PC-1000/Laser 1000 dental panoramic/cephalometric dental X-raymachine.The PC-1000 will enable the user to take panoramic X-ray images. The PC-1000/Laser 1000 will enablethe user to take panoramic X-ray images, as well as cephalometric X-ray images. A laser alignmentdevice is incorporated into the PC-1000/Laser 1000.The information contained in this manual is not all inclusive and Panoramic Corporation should becontacted for assistance and clarification when necessary.It is imperative that this equipment be installed, serviced, and used by personnel familiar with theprecautions required to prevent excessive exposure to both primary and secondary radiation. Thisequipment features protective designs for limiting both the primary and secondary radiation producedby the X-ray beam. However, design features cannot prevent carelessness, negligence, or lack ofknowledge.Only personnel authorized by Panoramic Corporation are qualified to install and service this equipment.Any attempt to install or service this equipment by anyone not so authorized will void the warranty.Statement of CompatibilityJanuary 1, 1988Please address any comments/questions concerning this statement of compatibility to:Panoramic Corporation 4321 Goshen Road Fort Wayne, IN 46818 USA Attn: Director of EngineeringThe only components compatible with the PC-1000 are those supplied with the machine.Regardless of possible statements made by other manufacturers, no one is authorized or certified tomake additions or deletions to this machine. Only the combination of components delivered with themachine is certified compatible by the manufacturer. As compatible accessories become available, Panoramic Corporation will certify them as compatible and make them available to the user.Statement of Compatibility AddendumOctober 1, 1988The Laser 1000 Cephalometric Attachment is certified by Panoramic Corporation to be compatible withthe PC-1000 dental X-ray machine, provided installation is performed by an authorized representativeutilizing specific installation instructions furnished by Panoramic Corporation.Statement of Compatibility AddendumOctober 1, 1995Laser 1000 Cephalometric Attachments manufactured after October 1, 1995 are compatible only with Copyright 2008 Panoramic Corporation5

IntroductionVoltage Regulator WarningDo not plug this machine into ANY voltage regulating device. Contact Panoramic Corporation with anyquestions regarding this.X-ray Shielding RequirementsThe requirements for panoramic and cephalometric shielding for building, operator, and patient, dependon state and local regulations. Contact your state Department of Health for compliance information.Compliance could involve a blueprint review, facility check, wall construction, film badge implementation,remote switch installation, and/or a lead apron. It is beyond the scope of this manual to advise on theseregulations.6 Copyright 2008 Panoramic Corporation

Pre-Installation CheckElectrical RequirementsOptimally, the PC-1000 and the PC-1000/Laser 1000 should have a dedicated 105-125 VAC, 20 A circuitwith line regulation of 5% or better. If a dedicated circuit is not available, a regular 20 A circuit will workas long as it is not taxed by other loads beyond 5 A. A standard 115 VAC, three-wire, grounded, electricaloutlet should be installed by an electrician behind the machine.Control Panel OrientationThe control panel on the PC-1000 and the PC-1000/Laser 1000 is mounted directly on the side of themachine and requires no separate installation, wiring, or mounting. Although the control panel is mountedon the patient's left side from the factory, it can be easily relocated to the opposite side during installation.Verify with the doctor on which side the control panel should be located.Cephalometric Arm OrientationThe cephalometric arm for the PC-1000/Laser 1000 is shipped in a separate box and can be mounted oneither side of the machine. Verify with the doctor on which side the cephalometric arm should be located.Remote SwitchSome states and local governments require that the exposure switch to be remotely installed. A remoteswitch kit is available from Panoramic Corporation. Refer to Appendix G for detailed procedures.Darkroom RequirementsProper darkroom facilities must be present before installing the PC-1000 or PC-1000/Laser 1000. Thedarkroom will be used to process test films at the time of installation to verify proper calibration. Refer toAppendix A for darkroom requirements and processing procedures. Copyright 2008 Panoramic Corporation7

PC-1000 ComponentsPC-1000Top CoverRotating ArmTemple SupportsCassette SleeveFilm DrumForehead SupportMirrorTubeheadkVp SettingKnobChinrestControl PanelHandlesBase8 Copyright 2008 Panoramic Corporation

PC-1000 InstallationIf there is any question concerninginstallation or calibration, pleasecontact Panoramic Corporationimmediately.Tools RequiredMultimeter, regular and phillips screwdrivers, allenwrenches, small fluorescent screen, knife, level,and a pulse counter or stopwatch.Verify PowerVerify that the outlet is a 110 VAC grounded outlet.Remove Packaging and Shipping Restraints1. Carefully remove all shipping packaging fromthe PC-1000, including the plastic wrap,cardboard, and wooden cover. Unpack theside and top covers shipped in the separatebox.2. Remove the 3/4" hex head shipping bolt fromthe top rear of the overhead chassis. Theshipping bolt is located between the aluminumchassis and the rear of the machine. Theremay be 1 or 2 shipping bolts. Leave thesebolts with the office for future relocation.Control Panel RelocationVerify with the doctor, on which side the controlpanel needs to be located. To relocate it theopposite side:1. Remove the 4 7/16" mounting bolts fromthe control panel.2. Move the control panel to the opposite sideof the machine, while carefully routing thewiring harness.3. Reinstall the 4 7/16" bolts in the controlpanel. Copyright 2008 Panoramic Corporation9

PC-1000 InstallationInstall Side Covers1. Install 3 metal standoffs in the pre-threadedholes on the side of the column opposite thecontrol panel.2. Place the side cover without the power cordover the standoffs.3. Insert 3 allen head bolts into the holes in theside cover and tighten into the metal standoffs.4. Install 3 metal standoffs in the pre-threadedholes on the control panel side of the column.5. Locate the wire with the 15-pin molexconnector J7 in the overhead and connect itto the 15-pin P7 connector on the top of thecontrol panel.6. Connect the 6-pin molex connector P15 on themain power cord in the side cover to the 6-pinconnector J15 on the top of the control panel.7. Route the power cord through the notches onthe side of the control panel while placing theside cover over the standoffs.8. Insert 3 allen head bolts into the holes in theside cover and tighten into the metal standoffs.10 Copyright 2008 Panoramic Corporation

PC-1000 InstallationInstall Film Drum Assembly1. Slide the black plastic cover off of the end ofthe rotating arm.2. Remove the cardboard packing material fromthe rotating arm.3. Remove the set screw, crown nut, and washerfrom the film drum spindle.4. Insert the film drum spindle down through therotating arm.5. Install the film drum with the knob closest to thefloor. Slide the film drum onto the spindle upas far as possible and tighten the film drumknob to temporarily hold the film drum in place.6. Slide the flat washer up on the film drumspindle.7. Install the crown nut on the film drum spindlewith the slots in the crown nut facing up.8. Thread the crown nut until the lower edge ofthe threaded hole in the film drum spindle isaligned to the lower edge of a slot in the crownnut.9. Insert the set screw in the threaded hole in thefilm drum spindle and tighten the screw.10. While supporting the film drum, loosen the filmdrum knob to allow the film drum to rest on thewasher and crown nut.11. Slide the black plastic cover back on the endof the rotating arm.Note: The film drum will rotate with increasedfriction after the film drum knob is tightened.Manual rotation, while the knob is tight, shouldbe avoided to prevent abnormal wear. Copyright 2008 Panoramic Corporation11

PC-1000 InstallationInstall Tubehead Assembly1. Connect the 6-pin molex connector from theend of the rotating arm to the 6-pin connectoron the tubehead.2. Carefully raise the tubehead into position facingthe film drum, ensuring that the wires are notstressed or pinched.3. Tighten the three set screws, then back eachout slightly to allow alignment of the tubehead.Install Temple Supports1. Install the 2 temple supports using the existingscrews and washers in the head supportassembly.2. Ensure that the temple supports are angledaway from the machine and that the horizontalred lines on the temple supports are facingout.Level MachineWith a level on the underside of the mainoverhead chassis, level the machine in alldirections using the 4 four adjustable feet underthe base.12 Copyright 2008 Panoramic Corporation

PC-1000 InstallationPerform mA Calibration1. Disconnect the yellow wire, 1-pin molexconnector in the overhead chassis.2. Connect a DC milliammeter in series withthe molex connectors, positive lead to themale connector, negative lead to the femaleconnector.3. Power the machine on.4. Using the function switch on the control panel,select PANORAMIC L or PANORAMIC R.5. Using the kVp setting knob on the controlpanel, set the kVp meter to 80 kVp on thePANORAMIC scale.6. Toggle the RUN/STOP switch on the printedcircuit board in the overhead chassis to STOP.7. Locate the mA adjustment knob P1 in theoverhead chassis.CAUTION: X-rays will be emitted when theexposure switch is depressed.8. Depress the exposure switch.9. The milliammeter should show 6.0 mA .05mA. If it does not:A. Adjust P1 clockwise (CW) to increase themA until the milliammeter shows 6.0 mA .05 mA.ORB. Adjust P1 counterclockwise (CCW) todecrease the mA until the milliammetershows 6.0 mA .05 mA.Note: Contact Panoramic Corporation if the mAcannot be set to 6.0 mA .05 mA.10. Disconnect the milliammeter.11. Reconnect the yellow wire, 1-pin molexconnectors.12. Toggle the RUN/STOP switch on the printedcircuit board in the overhead chassis to RUN. Copyright 2008 Panoramic Corporation13

PC-1000 InstallationPerform Pulse Count CalibrationIf a pulse counter IS available:1. Disconnect the blue wire, 1-pin molexconnector in the overhead chassis.2. Connect the positive lead of a pulse counterto the male connector, negative lead to thealuminum chassis or neutral wire from theplug.3. Power the machine on.4. Using the function switch on the control panel,select PANORAMIC L.5. Using the kVp setting knob on the controlpanel, set the kVp meter to 80 kVp on thePANORAMIC scale.6. Ensure that the RUN/STOP switch on theprinted circuit board in the overhead chassis isset to RUN.7. Locate potentiometer R57 on the printed circuitboard in the overhead chassis.CAUTION: X-rays will be emitted when theexposure switch is depressed.8. Depress the exposure switch for the entire 12second exposure.9. The pulse counter should show 720 pulses 10 pulses. If it does not:A. Adjust R57 counterclockwise (CCW) toincrease the pulses and repeat the testuntil the pulse counter shows 720 pulses 10 pulses.ORB. Adjust R57 clockwise (CW) to decrease thepulses and repeat the test until the pulsecounter shows 720 pulses 10pulses.10. Disconnect the pulse counter.11. Reconnect the blue wire, 1-pin molex14 Copyright 2008 Panoramic Corporation

PC-1000 InstallationIf a pulse counter IS NOT available:1. Power the machine on.2. Using the function switch on the control panel,select PANORAMIC L.3. Using the kVp setting knob on the controlpanel, set the kVp meter to 0 kVp on thePANORAMIC scale.4. Ensure that the RUN/STOP switch on theprinted circuit board in the overhead chassis isset to RUN.5. Use a stopwatch to time the duration of anexposure by timing the red exposure indicatoron the control panel.6. Locate potentiometer R57 on the printed circuitboard in the overhead chassis.CAUTION: X-rays will be emitted when theexposure switch is depressed.7. Depress the exposure switch for the entire 12second exposure.8. The stopwatch should show that the exposureis 12 seconds .5 seconds. If it does not:A. Adjust R57 counterclockwise (CCW) toincrease the time and repeat the test untilthe stopwatch shows that the exposure is12 seconds .5 seconds.ORB. Adjust R57 clockwise (CW) to decreasethe time and repeat the test until thestopwatch shows that the exposure is12 seconds .5 seconds. Copyright 2008 Panoramic Corporation15

PC-1000 InstallationPerform X-ray Beam Alignment1. Power the machine on.2. Using the function switch on the control panel,select PANORAMIC L.3. Using the kVp setting knob on the controlpanel, set the kVp meter to 90 kVp on thePANORAMIC scale.4. Toggle the RUN/STOP switch on the printedcircuit board in the overhead chassis to STOP.5. Temporarily affix a fluorescent screen behindthe film drum mask (slotted plate in front of thefilm drum).CAUTION: X-rays will be emitted when theexposure switch is depressed.6. Darken the room and depress the exposureswitch.7. The fluorescent screen should glow, denotingthe presence of X-rays. Adjust the tubeheadso that the X-ray beam is centered horizontallyin the film mask slot, and that the top edge ofthe X-ray beam is aligned to the top edge ofthe film drum mask slot.8. Tighten the 3 set screws locking the tubeheadassembly horizontally and the large allen bolton the right side of the tubehead locking itvertically. Verify that the X-ray beam is stillcentered.9. Remove the fluorescent screen.10. Toggle the RUN/STOP switch on the printedcircuit board in the overhead chassis to RUN.11. Using the kVp setting knob on the controlpanel, set the kVp meter to 70 kVp on thePANORAMIC scale.12. Place a loaded film cassette sleeve on the filmdrum and align the L1 pointer.13. Depress the exposure switch for the entire 12second exposure.14. Process the film to verify proper alignment.The film should be black with clear edges.16 Copyright 2008 Panoramic Corporation

PC-1000 InstallationInstall Top and Rear Covers1. Ensure that the RUN/STOP switch on theprinted circuit board in the overhead is in theRUN position.2. Slide the top cover on the machine from thefront. Ensure that the wiring on the controlpanel side is properly routed.3. Install the 6 screws on the underside of the topcover.4. Place the cover on the rear of the machine andinstall the 3 screws.Complete Installation1. Verify all steps have been completed.2. Refer to the PC-1000/Laser 1000 User Manualand train the staff thoroughly.3. Complete the Panoramic CorporationInstallation Report and the FDA 2579paperwork and distribute accordingly. Copyright 2008 Panoramic Corporation17

PC-1000/Laser 1000 ComponentsHorizontal Locking KnobVertical Locking KnobMid-Sagittal to Film ScaleLaser SwitchCephalostat Locking KnobCollimator LeverSuggested Settings DecalPC-1000Cephalostat ArmCephalometricExposure Time ChartPanoramic Corporation Seconds90 kVP 10 mARare Earth Screens And FilmLASER 1000Patient Size:Small ChildLarge ChildAverage AdultLarge 0-2.51.5-3.52.5-4.0Ear PostsNasion SupportEar RingsEar RodsFilm Cassette18 Copyright 2008 Panoramic Corporation

PC-1000/Laser 1000 InstallationIf there is any question concerninginstallation or calibration, pleasecontact Panoramic Corporationimmediately.Tools RequiredMultimeter, regular and phillips screwdrivers, allenwrenches, small fluorescent screen, knife, level,and a pulse counter.Verify PowerVerify that the outlet is a 110 VAC grounded outlet.Remove Packaging and Shipping Restraints1. Carefully remove all shipping packaging fromthe PC-1000/Laser 1000, including the plasticwrap, cardboard, and wooden cover. Unpackthe side and top covers, and the cephalometricarm shipped in the separate box.2. Remove the 3/4" hex head shipping bolt fromthe top rear of the overhead chassis. Theshipping bolt is located between the aluminumchassis and the rear of the machine. Theremay be 1 or 2 shipping bolts. Leave thesebolts with the office for future relocation.Control Panel RelocationVerify with the doctor, on which side the controlpanel needs to be located. To relocate it theopposite side:1. Remove the 4 7/16" mounting bolts fromthe control panel.2. Move the control panel to the opposite sideof the machine, while carefully routing thewiring harness.3. Reinstall the 4 7/16" bolts in the controlpanel. Copyright 2008 Panoramic Corporation19

PC-1000/Laser 1000 InstallationInstall Side Covers1. Install 3 metal standoffs in the pre-threadedholes on the side of the column opposite thecontrol panel.2. Place the side cover without the power cordover the standoffs.3. Insert 3 allen head bolts into the holes in theside cover and tighten into the metal standoffs.4. Install 3 metal standoffs in the pre-threadedholes on the control panel side of the column.5. Locate the wire with the 15-pin molexconnector J7 in the overhead and connect itto the 15-pin P7 connector on the top of thecontrol panel.6. Connect the 6-pin molex connector P15 on themain power cord in the side cover to the 6-pinconnector J15 on the top of the control panel.7. Route the power cord through the notches onthe side of the control panel while placing theside cover over the standoffs.8. Insert 3 allen head bolts into the holes in theside cover and tighten into the metal standoffs.20 Copyright 2008 Panoramic Corporation

PC-1000/Laser 1000 InstallationInstall Film Drum Assembly1. Slide the black plastic cover off of the end ofthe rotating arm.2. Remove the cardboard packing material fromthe rotating arm.3. Remove the set screw, crown nut, and washerfrom the film drum spindle.4. Insert the film drum spindle down through therotating arm.5. Install the film drum with the knob closest to thefloor. Slide the film drum onto the spindle upas far as possible and tighten the film drumknob to temporarily hold the film drum in place.6. Slide the flat washer up on the film drumspindle.7. Install the crown nut on the film drum spindlewith the slots in the crown nut facing up.8. Thread the crown nut until the lower edge ofthe threaded hole in the film drum spindle isaligned to the lower edge of a slot in the crownnut.9. Insert the set screw in the threaded hole in thefilm drum spindle and tighten the screw.10. While supporting the film drum, loosen the filmdrum knob to allow the film drum to rest on thewasher and crown nut.11. Slide the black plastic cover back on the endof the rotating arm.Note: The film drum will rotate with increasedfriction after the film drum knob is tightened.Manual rotation, while the knob is tight, shouldbe avoided to prevent abnormal wear. Copyright 2008 Panoramic Corporation21

PC-1000/Laser 1000 InstallationInstall Tubehead Assembly1. Connect the 6-pin molex connector from theend of the rotating arm to the 6-pin connectoron the tubehead.2. Carefully raise the tubehead into position facingthe film drum, ensuring that the wires are notstressed or pinched.3. Install and tighten the horizontal locking knob,the large knob with the smooth unthreadedend, into the rear of the rotating arm.4. Remove the two small set screws on each sideof the horizontal locking knob.5. Install the two slotted screws. Snug, thenbreak the snug fit to allow alignment of thetubehead.6. Install the two short set screws (from thecephalometric packaging, not those previouslyremoved).7. Remove the large allen bolt from the rightend of the tubehead assembly and install thevertical locking knob in it's place.8. Loosen both knobs and verify that the tubeheadcan move freely.Install Temple Supports1. Install the 2 temple supports using the existingscrews and washers in the head supportassembly.2. Ensure that the temple supports are angledaway from the machine and that the horizontalred lines on the temple supports are facingout.Level MachineWith a level on the underside of the mainoverhead chassis, level the machine in alldirections using the 4 four adjustable feet under22 Copyright 2008 Panoramic Corporation

PC-1000/Laser 1000 InstallationInstall Cephalometric Arm1. Verify with the doctor on which side thecephalometric arm should be mounted. Thearm is typically installed to extend on the sameside as the control panel.2. Mount the arm, with the spacers facing themachine, on the rear of the machine using the2 3/8" allen bolts with lock washers.3. Level the cephalometric arm.Install Cephalostat Head Positioner1. Install the chrome mounting tubes on thecephalometric arm with the 2 allen bolts andlockwashers provided.2. Slide the head positioner onto the 2 chromemounting tubes.3. Tighten the 2 allen bolts on the top rear of thehead positioner. Copyright 2008 Panoramic Corporation23

PC-1000/Laser 1000 InstallationPerform mA Calibration1. Disconnect the yellow wire, 1-pin molexconnector in the overhead chassis.2. Connect a DC milliammeter in series withthe molex connectors, positive lead to themale connector, negative lead to the femaleconnector.3. Power the machine on.4. Using the function switch on the control panel,select CEPHALOMETRIC 2.0.5. Using the kVp setting knob on the controlpanel, set the kVp meter to 80 kVp on theCEPHALOMETRIC scale.6. Using the lever on the front of the collimator,select LATERAL.7. Toggle the RUN/STOP switch on the printedcircuit board in the overhead chassis to STOP.8. Locate and turn the mA adjustment knob P1 inthe overhead chassis

panoramic X-ray machine and the PC-1000/Laser 1000 dental panoramic/cephalometric dental X-ray machine. The PC-1000 will enable the user to take panoramic X-ray images. The PC-1000/Laser 1000 will enable the user to take panoramic X-ray images, as well as cephalometric X-ray images. A laser alignment device is incorporated into the PC-1000 .

Related Documents:

PANASONIC LASER MARKING SYSTEMS. 03 LP-100 CO 2 Laser Marker LP-200 CO Laser Marker LP-F FAYb Laser Marker LP-D Diode Laser Marker LP-300 CO Laser Marker LP-V FAYb Laser Marker 1996 1999 2001 2003 2004 LP-400 Laser Marker LP-G FAYb Laser Marker LP-Z FAYb Laser Marker

Epilog Laser The Leading Worldwide Provider of Laser Marking Technology Since 1988, Epilog Laser has been the leading provider of laser technology. From industrial fiber laser marking solutions to consumer CO2 laser models, Epilog Laser is known for the highest-quality laser etching and marking.

panoramic X-ray machine and the PC-1000/Laser 1000 dental panoramic/cephalometric dental X-ray machine. The PC-1000 will enable the user to take panoramic X-ray images. The PC-1000/Laser 1000 will enable the user to take panoramic X-ray images, as well as cephalometric X-ray images. A laser alignment device is incorporated into the PC-1000 .

Laser treatment parameters for the Laser RAP sites included an average laser spot size of 4.1mm (range: 4-8mm). The average laser fluence used was 5.22J/cm2 (range 1.5-8.3J/cm2). At the Laser‐Only treated sites, a single laser pass was administered using a laser spot size of 4mm at an average laser fluence of 3.9J/cm2(range: 3.-4.6J/cm2).

Laser workers are responsible for complying with all aspects of this laser safety program: 1. Attend laser safety training ; 2. Develop laser safety protocols for the laser systems for which they are responsible ; 3. Comply with all laser safety controls recommended by the LSO. IV. LASER OPERATIONS

Laser Safety Training Laser Safety Offi cer Advanced Laser Safety Offi cer Principles of Lasers and Laser Safety Safety of Lasers Outdoors Laser Safety Masterclass Laser Safety for Art and Entertainment Measurements for Laser Safety Non-Ionizing Radiation ILSC 2015 Conference Agenda* Sunday, March 22 9:00am ASC Z136 Annual Meeting

3.LASER UNIT REF No. CODE Q'TY DESCRIPTION REMARKS 1 LY4628001 1 LASER UNIT (SP) (SJ type) *1 1 LU9361001 1 LASER UNIT (SP) (SM type) *1 ADD *1 When replacing the Laser unit, open the front cover of the machine, check the first two characters of the Laser serial label attached on the Laser plate and order the same type of the Laser unit that .

ASTM F 891 Cellular Core PVC DWV Pipe ASTM D 2665 PVC DWV Pipe & Fittings NSF Standard 14 Dimensional Standard Schedule 40 Iron Pipe Size (IPS) Cell Class 12454 PVC Solid Wall Pipe & Fittings 11432 PVC DWV Cellular Core Pipe Maximum Working Temperature 140 F Maximum Working Pressure 0 (zero) PSI PVC DWV is NOT a pressure-rated piping system .