Copper Mountain TechnologiesCalibration LaboratoryQuality ManualDocument Name:Revision:Date:QMS.01:20181November 2018

Copper Mountain TechnologiesDOCUMENT CHANGE RECORDRevision DateBrief Description of Changes18-November-2018The manual cancels and replaces the QMS.01:2015, which hasbeen technically revised.Document Name: QMS.01:2018Page 2 of 58Revision: 1

Copper Mountain TechnologiesDocument Name: QMS.01:2018Page 3 of 58Revision: 1

Copper Mountain TechnologiesCONTENTSDOCUMENT CHANGE RECORD . 2CONTENTS . 4AUTHORIZATION APPROVALS . 6DISTRIBUTION . 61 SCOPE . 72 NORMATIVE REFERENCES . 73 TERMS AND DEFINITIONS . 74 MANAGEMENT REQUIREMENTS . 74.1 Impartiality .74.2 Confidentiality .85 STRUCTURAL REQUIREMENTS . 96 RESOURCE REQUIREMENTS . 106.1 General . 106.2 Personnel . 106.3 Facilities and environmental conditions . 116.4 Equipment. 116.5 Metrological traceability . 136.6 Externally provided products and services . 147 PROCESS REQUIREMENTS . 157.1 Review of requests, tenders and contracts . 157.2 Selection, verification and validation of methods . 167.3 Sampling. 177.4 Handling of test or calibration items . 177.5 Technical records . 187.6 Evaluation of measurement uncertainty . 187.7 Ensuring the validity of results . 197.8 Reporting of results . 20Document Name: QMS.01:2018Page 4 of 58Revision: 1

Copper Mountain Technologies7.9 Complaints . 257.10 Nonconforming work . 267.11 Control of data and information management . 278 MANAGEMENT SYSTEM REQUIREMENTS . 288.1 Options . 288.2 Management system documentation (Option A) . 288.3 Control of management system documents (Option A) . 298.4 Control of records (Option A) . 298.5 Actions to address risks and opportunities (Option A) . 308.6 Improvement (Option A). 308.7 Corrective actions (Option A) . 308.8 Internal audits (Option A) . 318.9 Management reviews (Option A) . 319 USE OF ACCREDITATION BODY SYMBOL . 33ANNEX A ORGANIZATION CHART . 34ANNEX B PROCESSES INTERACTION CHART . 35ANNEX C QUALITY POLICY STATEMENT. 53ANNEX D STRUCTURE OF DOCUMENTATION . 56ANNEX E PERSONNEL COMPETENCE. 57Document Name: QMS.01:2018Page 5 of 58Revision: 1

Copper Mountain TechnologiesAUTHORIZATION APPROVALSElectronic authorization approval is the preferred method for approving quality systemdocuments. Signed hardcopies are only available upon request.Approver:Laboratory HeadApprover:Consulting EngineerDocument Reviewer:Quality ManagerDocument Owner:Quality ManagerDISTRIBUTIONPrinted copies of this document are uncontrolled and users must verify the revision iscurrent before use. All previous revisions must be discarded. Current documents and revisionindex are available on Copper Mountain Technologies network. All documents will have electronic approval. An uncontrolled copy is available to customers via the Copper MountainTechnologies website.Document Name: QMS.01:2018Page 6 of 58Revision: 1

Copper Mountain Technologies1 SCOPEThis Calibration Laboratory Quality Manual defines or identifies the policies, proceduresand requirements for the competence, impartiality and consistent operation that choose tocomply with the requirements of ISO/IEC 17025 as a calibration laboratory. This documentis applicable to all laboratory activities, regardless of accreditation exists. Any local documents, procedures and policies associated with ISO/IEC 17025 compliance for calibrationlaboratories must comply with this document.2 NORMATIVE REFERENCESThe following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latestedition of the referenced document (including any amendments) applies.ISO/IEC 17025 General requirements for the competence of testing and calibration laboratoriesISO/IEC Guide 99 International vocabulary of metrology – Basic and general conceptsand associated terms (VIM, also known as JCGM 200), issued by BIPM, IEC, IFCC, ISO, IUPAC,IUPAP and OIMLISO/IEC Guide 98-4 Uncertainty of measurement – Part 4: Role of measurement uncertainty in conformity assessment (also known as JCGM 106), issued by BIPM, IEC, IFCC, I L A C ,ISO, IUPAC, IUPAP and OIMLEURAMET Calibration Guide No. 12 Guidelines on the Evaluation of Vector Network Analyzers (VNA)3 TERMS AND DEFINITIONSFor the purposes of this document, the relevant terms and definitions given in VIM apply.4 MANAGEMENT REQUIREMENTS4.1 Impartiality4.1.1 Laboratory activities are undertaken impartially and structured and managed so asto safeguard impartiality.Document Name: QMS.01:2018Page 7 of 58Revision: 1

Copper Mountain Technologies4.1.2 The laboratory management is committed to impartiality.4.1.3 The laboratory is responsible for the impartiality of its laboratory activities anddoes not allow commercial, financial or other pressures to compromise impartiality.4.1.4 The laboratory identifies risks to its impartiality on an on-going basis. This includesthose risks that arise from its activities, or from its relationships, or from the relationshipsof its personnel. However, such relationships do not necessarily present a laboratory with arisk to impartiality.4.1.5 If a risk to impartiality is identified, the laboratory is able to demonstrate how iteliminates or minimizes such risk.4.2 Confidentiality4.2.1 The laboratory is responsible, through legally enforceable commitments, for themanagement of all information obtained or created during the performance of laboratoryactivities. The laboratory informs the customer in advance, of the information it intends toplace in the public domain. Except for information that the customer makes publicly available, or when agreed between the laboratory and the customer (e.g. for the purpose of responding to complaints), all other information is considered proprietary information and isregarded as confidential.4.2.2 When the laboratory is required by law or authorized by contractual arrangements to release confidential information, the customer or individual concerned is, unlessprohibited by law, notified of the information provided.4.2.3 Information about the customer obtained from sources other than the customer (e.g.complainant, regulators) is confidential between the customer and the laboratory. The provider (source) of this information is confidential to the laboratory and is not be sharedwith the customer, unless agreed by the source.4.2.4 Personnel, including any committee members, contractors, personnel of externalbodies, or individuals acting on the laboratory's behalf, keep confidential all information obtained or created during the performance of laboratory activities, except as required by law.Document Name: QMS.01:2018Page 8 of 58Revision: 1

Copper Mountain Technologies5 STRUCTURAL REQUIREMENTS5.1 The laboratory is a defined part of Copper Mountain Technologies that is legally responsible for its laboratory activities.5.2 The laboratory has identified management that has overall responsibility for the laboratory (annex A).5.3 The laboratory has defined and documented the range of laboratory activities forwhich it conforms with this document. Copper Mountain Technologies laboratory performscalibration activities. The laboratory only claims conformity with this document for thisrange of laboratory activities, which excludes externally provided laboratory activities on anongoing basis. Laboratory whose work includes both accredited and non-accredited calibrations complies to this document, but non-accredited calibrations are not required to fullycomply with sections 6.5, 7.6 and reporting requirements of 7.8.5.4 Laboratory activities are carried out in such a way as to meet the requirementsof this document, the laboratory’s customers, regulatory authorities and organizationsproviding recognition. This includes laboratory activities performed in its permanent facilities.5.5 The laboratory has (annex B):a) defined the organization and management structure of the laboratory, its place inparent organization (see also annex A), and the relationships between management, technicaloperations and support services. The structure is described to a level of detail sufficient toidentify key personnel / activities and in order to reveal possible conflicts of interest involvingboth calibration services and manufacturing activities involving calibration;b) specified the responsibility, authority and interrelationship of all personnel who manage, perform or verify work affecting the results of laboratory activities;c) documented its procedures to the extent necessary to ensure the consistent application of its laboratory activities and the validity of the results.5.6 The laboratory has personnel named as a quality manager who, irrespective of otherresponsibilities, have the authority and resources needed to carry out their duties, including:a) implementation, maintenance and improvement of the management system;b) identification of deviations from the management system or from the proceduresfor performing laboratory activities;c) initiation of actions to prevent or minimize such deviations;Document Name: QMS.01:2018Page 9 of 58Revision: 1

Copper Mountain Technologiesd) reporting to laboratory management on the performance of the management systemand any need for improvement;e) ensuring the effectiveness of laboratory activities.NOTE Individuals may be assigned more than one function and it may be impractical toappoint deputes for every function.5.7 Laboratory management ensures that:a) communication takes place regarding the effectiveness of the management systemand the importance of meeting customers’ and other requirements;b) the integrity of the management system is maintained when changes to the management system are planned and implemented.6 RESOURCE REQUIREMENTS6.1 GeneralThe laboratory has available the personnel, facilities, equipment, systems and supportservices necessary to manage and perform its laboratory activities.6.2 Personnel6.2.1 All personnel of the laboratory, either internal or external, that could influencethe laboratory activities act impartially, are competent and work in accordance with the laboratory's management system.6.2.2 The laboratory documents the competence requirements for each function influencing the results of laboratory activities, including requirements for education, qualification, training, technical knowledge, skills and experience.6.2.3 The laboratory ensures that the personnel have the competence to performlaboratory activities for which they are responsible and to evaluate the significance of deviations.6.2.4 The management of the laboratory communicates to personnel their duties, responsibilities and authorities.6.2.5 The laboratory has procedure (s) and retains records for:a) determining the competence requirements;b) selection of personnel;c) training of personnel;Document Name: QMS.01:2018Page 10 of 58Revision: 1

Copper Mountain Technologiesd) supervision of personnel;e) authorization of personnel;f) monitoring competence of personnel.6.2.6 The laboratory authorizes personnel to perform specific laboratory activities, including but not limited to, the following:a) development, modification, verification and validation of methods;b) analysis of results, including statements of conformity or opinions and interpretations;c) report, review and authorization of results.6.3 Facilities and environmental conditions6.3.1 The facilities and environmental conditions are suitable for the laboratory activities and not adversely affect the validity of results.6.3.2 The requirements for facilities and environmental conditions necessary for theperformance of the laboratory activities are documented.6.3.3 The laboratory monitors, controls and records environmental conditions in accordance with relevant specifications, methods or procedures or where they influence the validity of the results.6.3.4 Measures to control facilities are implemented, monitored and periodically reviewed and include, but not be limited to:a) access to and use of areas affecting laboratory activities;b) prevention of contamination, interference or adverse influences on laboratory activities;c) effective separation between areas with incompatible laboratory activities.6.3.5 When the laboratory performs laboratory activities at sites or facilities outsideits permanent control, it ensures that the requirements related to facilities and environmental conditions of this document are met.6.4 Equipment6.4.1 The laboratory has access to equipment (including, but not limited to, measuringinstruments, software, measurement standards, reference materials, reference data, reagents,consumables or auxiliary apparatus) that is required for the correct performance of laboratoryDocument Name: QMS.01:2018Page 11 of 58Revision: 1

Copper Mountain Technologiesactivities and that can influence the results.6.4.2 When the laboratory uses equipment outside its permanent control, it ensuresthat the requirements for equipment of this document are met.6.4.3 The laboratory has a procedure for handling, transport, storage, use and plannedmaintenance of equipment in order to ensure proper functioning and to prevent contamination or deterioration.6.4.4 The laboratory verifies that equipment conforms to specified requirements before being placed or returned into service.6.4.5 The equipment used for measurement is capable of achieving the measurementaccuracy and/or measurement uncertainty required to provide a valid result.6.4.6 Measuring equipment is calibrated when: the measurement accuracy or measurement uncertainty affects the validity of the reported results, and/or calibration of the equipment is required to establish the metrological traceability ofthe reported results.6.4.7 The laboratory establishes a calibration program, which is reviewed and adjustedas necessary in order to maintain confidence in the status of calibration.6.4.8 All equipment requiring calibration or which has a defined period of validity islabelled, coded or otherwise identified to allow the user of the equipment to readily identifythe status of calibration or period of validity.a) Laboratory has a procedure stating its policy for establishing and changing calibrationintervals for equipment that it controls.6.4.9 Equipment that has been subjected to overloading or mishandling, gives questionable results, or has been shown to be defective or outside specified requirements, is taken outof service. It is isolated to prevent its use or clearly labelled or marked as being out of serviceuntil it has been verified to perform correctly. The laboratory examines the effect of the defect or deviation from specified requirements and initiates the management of nonconforming work procedure (see 7.10).6.4.10 When intermediate checks are necessary to maintain confidence in the performance of the equipment, these checks are carried out according to a procedure.Document Name: QMS.01:2018Page 12 of 58Revision: 1

Copper Mountain Technologies6.4.11 When calibration and reference material data include reference values or correction factors, the laboratory ensures the reference values and correction factors are updatedand implemented, as appropriate, to meet specified requirements.6.4.1

This Calibration Laboratory Quality Manual defines or identifies the policies, procedures and requirements for the competence, impartiality and consistent operation that choose to comply with the requirements of ISO/IEC 17025 as a calibration laboratory.

Related Documents:

1 Manual calibration introduction 3 1.1 Purpose 3 1.2 Manual calibration software 3 2 Manual calibration overview 3 3 Manual calibration process 4 3.1 Setup MCR program 4 3.2 Calibration pattern positioning 9 3.3 Capturing pattern image 14 3.4 Using the Synergy Calibration tool 18 3.5 Fine tuning side images 24

Calibration (from VIM3) Continued NOTE 1 A calibration may be expressed by a statement, calibration function, calibration diagram, calibration curve, or calibration table. In some cases, it may consist of an additive or multiplicative correction of the indication with associated measurement uncertainty. NOTE 2 Calibration should not be .

Host action Target needed * : Timings are given for information only, they can vary depending on the Host capabilities Calibration Data result Calibration data Calibration data Calibration data Calibration data Device initialization SPADs calibration Temperature calibration Offset calibrat

calibration. The vertical comparator was built during the year 2003. The calibration facility is designed to calibrate up to 3-m-long invar rods, both for system calibration of digital levels and for tradi-tional rod calibration. SLAC System Calibration Facility The procedure of system calibration of digital levels is described

Calibration C-0942, Mechanical(Dimensional, Pressure, Mass & Volume) Calibration C-0943 & Thermal Calibration C-0944. Our company is also ISO 17025:2005 Certified for Mechanical Calibration, Elctro Technical Calibration & Thermal Calibration. The firm is proud to have earned distinction in its various services and we are performing as a

1.7.1.1 e) v) The laboratory may replace a calibration standard provided that a. the laboratory analyzes the replacement standard within twenty-four (24) hours of the original calibration standard analysis for that particular calibration level; b. the laboratory replaces all analytes of the replacement calibration standard if a standard within the

4.2.3 Fresh air calibration in a contaminated atmosphere 21 4.3 Gas Calibration 22 4.3.1 Gas calibration failure: All sensors except oxygen 22 4.3.2 Gas calibration failure: Oxygen sensors 22 4.4 Special Calibration Instruction for NDIR CO2 sensor 23 4.4.1 CO2 Sensor True Zero 23 4.5 Special Calibration Instructions for NDIR-CH4 Sensor 23 5.

Quantify – Multivariate Calibration Overview Calibration Model Development Steps Load calibration spectra Merge all spectra into one data view Create a project (and a folder) Add labels with concentrations (Label Editor) Start the calibration wizard Create and optimize a calibration