Mee T N R HMasu Pe In The Lab - Assets.thermofisher

3y ago
51 Views
3 Downloads
723.04 KB
13 Pages
Last View : 1m ago
Last Download : 3m ago
Upload by : Mara Blakely
Transcription

pH Measurementin the LabpH Measurement per USP 791 Preparing your Lab

Why you should read this ebookThe purity of pharmaceutical productsdepends on the quality of raw materials.The United States PharmacopeialConvention (USP) is a non-profitorganization that sets standards for theidentity, strength, quality, and purity ofmedicines, food ingredients, and dietarysupplements manufactured, distributed andconsumed worldwide. USP 791 refersto the standardized test method for pHmeasurements used in the pharmaceutical,food and beverage, patient care, and dietarysupplement industry.In December 2014, USP 791 officiallywent into effect. This version of 791 ispart of the Second Supplement toUSP 37-NF 32.pH measurements within the pharmaceuticalindustry often reference USP 791 .Thermo Scientific Orion pH meter kitsare part of a high-quality pH test methoddesigned to assist with compliance to USP 791 pH requirements.A growing number of foods and beveragesare sweetened with plant-based stevia(Rebaudioside A). In 2009, USP releasedstandards that manufacturers can use tohelp ensure the quality and purity of thestevia ingredients they put in their products.When you take a medicine or supplement,you won’t get the therapeutic benefit unlessthe pill or capsule dissolves properly in yourbody. A medication or supplement thatmeets USP standards is able to pass teststhat ensure it dissolves so your body canabsorb it.All prescription and over-thecounter drugssold in the United States, no matter wherein the world they are made, must meet theUSP quality standards that exist for thatproduct.Visit usp.org for more information2

Measurement system and instrument requirementsAs of December 1, 2014, the requirements per USP 791 for pH measurement were revised. For details and exact language, seeUSP 791 pH General Chapter, The United States Pharmacopeial Convention. usp.orgpH Measurement system requirementsComponentElectrodeMeter*Temperature device ph sensor (H ) and suitable reference Choice of combined or half-cells Voltage measurement Handles high-input impedance Embedded in electrode Alternatively, use an external temperture device*If no meter, then embedded inside the pH sensorInstrument requirementsMeasurement 2-point calibaration capablepH Resolution 0.01 pH or better Temperature compensatingTemperature Accuracy 1 C Resolution 0.1 C3

Typical acceptable parameters and requirements1Per USP 791 pH, purchased buffers traceable to NIST and having a stated pH value accurate to 0.02 pH may be used. ThermoScientific Orion pH buffers meet these criteria. For each lot, a Certificate of Analysis is issued which documents the NIST traceabilityand the pH value accuracy to 0.02 pH or better. Alternately, the analyst may prepare buffers in compliance with Table 2 in the USP 791 method.pH Measurement system calibration requirementsComponentBuffersThree buffers minimum: at least 2 buffers for calibration and 1buffer for verificationVerification The verification buffer value should be between the calibrationbuffers /- 0.05 pH of the expected value at that temperatureTypical acceptable parametersTypical Slope90 – 105%*Typical Offset /- 30 mV (0.5 pH units at 25 C)** Typical Offset and Slope values suggested by USP 791 should be used unless otherwise defined by the pH measurement systemmanufacturer.3Slope 92-102% and Offset 59 mV are defined by Thermo Scientific Orion specifications.4

Measuring pH per USP 791 Thermo Scientific Orion pH meters, electrodes and solutions are part of a high-quality pH test method designed to assistwith compliance to USP 791 pH requirements. Orion pH meters, electrodes and buffers offer features commonly desiredby highly-regulated labs.USP 791 ure Device 1ºCVerify against a NIST-traceable thermometer2 (minimum)(USA Buffer Set)3-point calibration with pH 4.01, 7.00, and 10.01 buffers, or5-point calibration with pH 1.68, 4.01, 7.00, 10.01, and 12.46 buffers92 - 102%*Slope is displayed on the meter screen when calibration is completeSlope is also displayed in the meter calibration log 59 mV*Offset can be calculated by the meter, orOffset can be measured in pH 7 buffer2, orOffset can be calculated by the analyst 0.05 pHAnalyze pH 6.86 buffer. Compare resultsto a table of buffer values at temperatureSee table on page 10 of eBookCalibration BuffersSlope ValueOffsetVerification* Slope 92-102% and Offset 59 mV are defined by Thermo Scientific Orion specifications.Remember to consult with the SOP prepared by your internal quality or regulatory group when performing your testingper USP 791 .5

Calibration buffersConsider a 3-point calibration at pH 4.01, 6.86, and 9.18 buffers,or a 4-point calibration at pH 1.68, 4.01, 6.86, and 9.18 buffers.For DIN Buffer SetsVerify in pH 7.00 buffer.Obtain the offset from the meter calculation, or from the mV value reading during thepH 7.00 buffer verification, or by analyst calculation.Choosing Calibration Points:USP 791 specifies a minimum of two calibration points. Other multipoint calibrations are acceptable, if the calibration2accuracy and pH sensor slope and offset criteria are met.For Optimal AccuracyConsider choosing buffers that are not more than 3 or 4 pH units apart.For wide-range pH testing, a 3-point calibration at pH 4.01, 7.00, and 10.01is an option. A 2-point calibration at pH 4 and pH 10 mayimpact your accuracy.For Full-Range pH TestingConsider a 5-point calibration at pH 1.68, 4.01,7.00, 10.01 and 12.46.For Short-Range pH TestingA 2-point calibration is the minimum requirement(for example, at pH 4 and 7 OR at pH 7 and 10).Slope ValueThe slope value is always displayed by the meter after the calibration and in the meter’s calibrationlog. Compare the displayed value to the defined parameters of 92-102% per Thermo Scientific Orionspecifications.36

OffsetThere are different valid methods to determine the offset of the electrode, including by calculation or by direct measurement in pH 7.00 buffer2. Choosethe option that best fits your requirements and always consult with the SOP prepared by your internal quality or regulatory group.MeterCalculationDisplayed in calibration logCalculated from calibration data3hohoAnalystCalculationdetM etMethods toDetermineOffsetd 2t ho d 1MeMDirectMeasurementMeasured in pH 7 buffer7

Method 1 – Meter calculationDisplayed in calibration logThermo Scientific Orion meters will calculate offset during the calibration. When using the Thermo Scientific Orion Versa Star Pro pHmeter, the calculated offset (sometimes noted as E0) can be found in the calibration log display and in the computer/printer capture of thepH calibration log. When using the Thermo Scientific Orion Star A200 series or Orion Star A300 series pH meters, the calculated offsetcan be found in the computer/printer capture of the pH calibration log. The meter will calculate an offset for each segment of the calibration. For a 2-point calibration, one offset (E1)will be calculated. For a 3-point calibration, two offsets (E1, E2) will be calculated, and so on. When more than one offset is calculated, each offset should meet the /- 59 mV criteria per Thermo Scientific Orion specifications.3The calculated offsets should be similar, but may not be identical values.Method 2 – Direct MeasurementMeasured in pH 7 bufferThermo Scientific Orion meters electrodes are constructed to provide a cell zero (pH 0) of pH 7. The offset potential is the potential measured at the indicated cell zero; therefore, the offset potential is the mV reading measured in pH 7 buffer. To determine the offset, the analyst can measure pH 7 and document the mV reading. This may be achieved by including pH 7 bufferin the calibration or by using pH 7 as the verification buffer. Compare the mV reading to the /- 59 mV per Thermo Scientific Orionspecifications.38

Method 3 – Analyst calculationCalculated from calibration dataAnother method is to calculate the offset from a calibration buffer data point. This can be doneby the analyst. To calculate offset, use the following equation:E0 Es - (-59.16 * (T 273.15) * (% slope) * (pHs -7))298.15 100Where:EsT% slopepHs mV of a pH buffer used during calibration (e.g. mV for the pH 4.01 buffer)temperature of the buffer tested (ºC)percent slope determined during the calibrationthe true pH of the pH buffer at the temperature during testingExample of Method 3 EquationA pH 4.01 buffer at 35ºC is used during calibration and reads a mV value of 157.4.The displayed slope of the calibration is 95.3%. The true value of pH 4.01 buffer at 35ºC is 4.02.E0 157.4 - (-59.16* ((35 273.15)/298.15) * (95.3%/100) * (4.02-7))E0 -16.2 mV offset9

VerificationpH Buffer Values for Calibration VerificationTo verify the calibration, test a pH buffer that was notused during the calibration and has a value betweenthe calibration buffers.T ( C) pH6.86pH7.00T ( C) pH6.86pH7.00T ( C) pH6.86pH7.00T ( C) 116.987.11266.866.99516.836.96766.857.01 Consider pH 6.86 or pH 7.00 buffer (whichever isnot used during the calibration) for the purpose ofverification. If desired, an alternate buffer maybe .97806.867.03The pH reading for the buffer should agree within 0.05 pH units of the buffer value expected at .07336.856.98586.846.97836.877.04 Compare the pH reading obtained to knownexpected values for the buffer at that 5106.927.06356.846.97606.846.97856.877.05 See table at the right for expected values of pH 6.86and pH 7.00 buffers at temperatures ranging from 0to 7.01996.897.131006.897.14Thermo Scientific Orion pH Values for 6.86 and 7.00 Buffers at temperatures from 0 to 100 ºC10

Preparing to Measure pH in aHighly-Regulated LabChoose EquipmentCalibrateVerify Temperature pH pHTest SamplesRecord Results 0.05 pH Slope 92-102%Choose Buffers*Periodic calibrationFootnotes:1 For details and exact language, see: USP 791 pH General Chapter, The United States Pharmacopeial Convention,www.usp.org.Per written communication with Mr. Antonio Hernandez-Cardoso, Senior Scientific Liaison, United States Pharmacopeia,November 7, 2014. AHC@usp.org.2 Per written communication with Mr. Antonio Hernandez-Cardoso, Senior Scientific Liaison, United States Pharmacopeia,April 29, 2019. AHC@usp.org.3 11

Next stepsConduct your pH testing with confidence. Thermo Scientific Orion pH meters, electrodes, and solutions are part of a high-quality pH test methoddesigned to assist with USP 791 pH requirements. To purchase a pH meter, electrode and calibration solutions, please contact your localThermo Scientific Orion sales representative and reference the part numbers listed below.DescriptionPart NumberThermo ScientificTM OrionTM Versa Star ProTM pH benchtop meter kit, includes electrode stand, 8302BNUMDROSS UltraTM TriodeTM glass-body pH/ATC electrode, 096019 stirrer probe and 810199 ROSS pH buffer & solution kitVSTAR12Orion ROSS Ultra Triode glass-body pH/ATC electrode8302BNUMDOrion ROSS Ultra Triode epoxy-body pH/ATC electrode8157BNUMDOrion ROSS Sure-Flow glass-body pH electrode8172BNWPOrion ROSS Sure-Flow epoxy-body pH electrode8165BNWPOrion ATC temperature probe with stainless steel body927007MDSolutionsOrion pH 1.68 buffer, 475 mLOrion pH 4.01 buffer, 475 mLOrion pH 6.86 buffer, 475 mLOrion pH 7.00 buffer, 475 mLOrion pH 10.01 buffer, 475 mLOrion pH 12.46 buffer, 475 nTM Navigator ProTM Computer Software for Orion Versa Star Pro meters, includes one software license on CDand user manualVSTAR-NPCDOrion Navigator Pro Computer Software for Orion Versa Star Pro meters, includes one software license on USB flashdrive and user manualVSTAR-NPUSBProductMetersElectrodesFor more information on the Orion Navigator Pro computer software, visit thermofisher.com/orionsoftware12

PharmaceuticalDietary SupplementsFood & BeverageMedicineResources for Electrochemistry and Water Analysis EquipmentThermo Scientific Orion Lab, Field and Online ProductsVisit thermofisher.com/water for information on all of the Thermo Scientific Orion laboratory, field and online equipment forpH, ISE, conductivity and dissolved oxygen analysis, plus Thermo Scientific Orion UV-Vis spectrophotometry, colorimetryand turbidity products.Find out more at thermofisher.com/waterThis product is intended for General Laboratory Use. It is the customer’s responsibility to ensure that the performance of the product is suitable for the specific use orapplication. 2017 - 2019 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified.R-USP791-E 0819 RevDBiotechnology

Thermo Scientific Orion pH meter kits are part of a high-quality pH test method designed to assist with compliance to USP 791 pH requirements. 2 Visit usp.org for more information A growing number of foods and beverages are sweetened with plant-based stevia (Rebaudioside A). In 2009, USP released standards that manufacturers can use to

Related Documents:

Sep 28, 2021 · SAP LLC “SAP Labs“ MEE Russian Federation SAP SAP CIS, LLC MEE Serbia SAP SAP West Balkans d.o.o. MEE Slovakia SAP SAP Slovensko s.r.o. MEE Slovakia SAP Ariba Ariba Slovak Republic, s.r.o. MEE Slovenia SAP SAP sistemi, aplikacije in produkti za obdelavo podatkov d.o.o. MEE Switzerland Emarsys Emarsys Schweiz GmbHFile Size: 598KB

Mee/Beehoon/Kway Teow Soup Noodles served in a healthy & light tasting vegetable broth filled with the goodness of vegetables and tofu 8.0 8.5 8.0 8.5 8.5 8.5 8.0 8.0 8.0 8.0 8.0 9.5 8.0 8.0 ND01 Gokul Special Hokkien Mee ND02 Tom Yam Seafood Mee Soup ND03 Mee Siam ND04 Nonya Laksa ND06 Laksa Goreng ND07 Char Kwa

monteur verandert mee. ‘De nieuwe installateur’ is een compleet andere vakman dan die van tien jaar geleden. In een serie van drie artikelen beschrijft Installatie en Sanitair de veranderingen. buurt was om de ketel mee de trap op te zeulen. Man dat was beulenwerk! Ga maar na. Nederland was bezaaid met portiekflats. Vier etages mee -

83 Mee, Kway Teow atau Meehoon sup (Egg noodle, flat rice noodle or rice vermicelli soup with seafood) 9.90 84 Mee goreng udang (Fried egg noodles with eggs and prawns) 9.90 85 Mee goreng Mamak (Savoury fried egg noodles with bean curd, eggs and shrimp fritters) 9.90 86Meehoon

convection natural convection, radiation heat transfer, heat exchanger, boiling and condensation, and numerical methods in heat transfer. Prerequisites: MEE 230, MEE 360. ABET Outcomes: 1. Ability to identify, formulate and solve complex engineering problems by applying principles of engineering, science, and mathematics 2.

man,” and later “I have laid my deity aside and go disguised as man” (Bacchae 4, 53-54).9 Mee, on the other hand, does not focus on the divine nature of Dionysus or the introduction of a new religion—both major themes i

integrated study of the MEE subject and about their training needs (Magdaş et al., 2017) and about the way in which the curriculum documents facilitated the study of the MEE subject. In this paper, we analyse the primary education teachers’ answers, expressed during the

Anatomy of a journal 1. Introduction This short activity will walk you through the different elements which form a Journal. Learning outcomes By the end of the activity you will be able to: Understand what an academic journal is Identify a journal article inside a journal Understand what a peer reviewed journal is 2. What is a journal? Firstly, let's look at a description of a .