Technical Quality Control Protocol

2y ago
13 Views
3 Downloads
4.30 MB
142 Pages
Last View : 6d ago
Last Download : 3m ago
Upload by : Milena Petrie
Transcription

Technical Quality Control ProtocolHandbookScintillation Camera andSPECT SystemsMedical Radiation Surveillance DivisionMINISTRY OF HEALTH MALAYSIA

TECHNICAL QUALITY CONTROL PROTOCOL HANDBOOKFORSCINTILLATION CAMERA AND SINGLE PHOTON EMISSIONCOMPUTED TOMOGRAPHY (SPECT) SYSTEMSPREPARED BYMEDICAL RADIATION SURVEILLANCE DIVISIONAUGUST 2015Scintillation Camera & SPECT Version August 2015: Technical QC Protocol Handbook, MRSD Page 1 of 41

CONTENTSPAGESAIntroduction4BList of Test Parameter6CList of Test Tools8DTesting Procedures10Test Parameter 1PHYSICAL INSPECTION11Test Parameter 2BACKGROUND COUNT RATE13Test Parameter 3CENTRING PULSE HEIGHT ANALYSER (PHA)WINDOW SETTINGS14Test Parameter 4ENERGY RESOLUTION15Test Parameter 5INTRINSIC FLOOD FIELD UNIFORMITY FOR 99MTC16Test Parameter 6INTRINSIC SPATIAL RESOLUTION18Test Parameter 7SYSTEM FLOOD FIELD UNIFORMITY (ON COMMONCLINICALLY USED COLLIMATOR)20Test Parameter 8SYSTEM SPATIAL RESOLUTION21Test Parameter 9SYSTEM PLANAR SENSITIVITY23Test Parameter 10COUNT RATE PERFORMANCE10.1 Intrinsic Count Rate Performance2410.2 Maximum Count Rate26Test Parameter 11MULTIPLE WINDOW SPATIAL REGISTRATION(WITH GALLIUM 67 APPLICATION)27Test Parameter 12DETECTOR HEAD SHIELDING LEAKAGE28Test Parameter 13IMAGE DISPLAY13.1 Visual Display2913.2 Hardcopy Display30ADDITIONAL TEST PARAMETER FOR SPECTTest Parameter 14ABSOLUTE PIXEL SIZE31Test Parameter 15TOMOGRAPHIC UNIFORMITY (IF APPLICABLE)32Test Parameter 16TOMOGRAPHIC RESOLUTION (TR) IN AIR (SPECTRECONSTRUCTED SPATIAL RESOLUTIONWITHOUT SCATTER)34Scintillation Camera & SPECT Version August 2015: Technical QC Protocol Handbook, MRSD Page 2 of 41

PAGESTest Parameter 17CENTER OF ROTATION (COR) / MULTI HEADREGISTRATION (MHR) TEST / IMAGEREGISTRATION CORRECTION (IRC) ANALYSIS35EPerformance and Safety Standards36FReferences40Scintillation Camera & SPECT Version August 2015: Technical QC Protocol Handbook, MRSD Page 3 of 41

A. IntroductionScintillation Camera & SPECT Version August 2015: Technical QC Protocol Handbook, MRSD Page 4 of 41

INTRODUCTIONThe Ministry of Health (MOH) is continuously taking steps to improve the quality of nuclearmedicine services. This is to ensure that necessary clinical information is obtained from theoptimum use of ionizing radiation to the patients. Over the years, the MOH has taken bothadministrative and legislative measures to enforce the various requirements under theAtomic Energy Licensing Act 1984 (Act 304).In order to further upgrade and enhance the quality, safety, and efficacy of nuclear medicineservices, the MOH has formulated and initiated the implementation of Quality AssuranceProgramme (QAP) for both government and private nuclear medicine centres since 2013.The MOH is empowered under Section 17 of Act 304 to implement such programme. Thespecific requirement of the programme will be imposed under Regulation 53 of Basic SafetyRadiation Protection Regulations (BSRP) 2010.Quality Control (QC) is one of the elements of QAP that has to be carried out at intervalperiod as specified by MOH. QC tests shall be performed at the time of the acceptance andcommissioning of the nuclear medicine imaging and non-imaging equipments prior to theuse; after replacement of major components that could cause a change in the performanceof the machine including image quality and safety of patient; or as routine tests. Theperformance and safety standards of the nuclear medicine imaging and non-imagingequipments shall be in accordance with the regulatory requirements and relevant code ofpractice.This Handbook is designed to be guidance to conduct QC checking on Scintillation Cameraand SPECT systems in the aspect of performance and safety standard requirement inaccordance with MOH. Wherever applicable, manufacturer’s protocol can be followed toperform the test procedures. A complete annual QC report by an approved medical physicistor qualified expert shall be submitted to MOH annually.Scintillation Camera & SPECT Version August 2015: Technical QC Protocol Handbook, MRSD Page 5 of 41

B. List of Test ParameterScintillation Camera & SPECT Version August 2015: Technical QC Protocol Handbook, MRSD Page 6 of 41

List of Test ParameterNo.Parameter1.Physical Inspection2.Background Count Rate3.Centring Pulse Height Analyser (PHA) Window Settings4.Energy Resolution5.Intrinsic Flood Field Uniformity For 99mTc6.Intrinsic Spatial Resolution7.System Flood Field Uniformity (On Common Clinically Used Collimator)8.System Spatial Resolution9.System Planar Sensitivity10.Count Rate Performance10.1Intrinsic Count Rate Performance10.2Maximum Count Rate11.Multiple Window Spatial Registration (With Gallium 67 Application)12.Detector Head Shielding Leakage13.Image Display13.1Visual Display13.2Hardcopy Display14.Absolute Pixel Size15.Tomographic Uniformity (If Applicable)16.Tomographic Resolution (TR) in Air (SPECT Reconstructed SpatialResolution Without Scatter)17.Center Of Rotation (COR) / Multi Head Registration (MHR) Test / ImageRegistration Correction (IRC) AnalysisScintillation Camera & SPECT Version August 2015: Technical QC Protocol Handbook, MRSD Page 7 of 41

C. List of Test ToolsScintillation Camera & SPECT Version August 2015: Technical QC Protocol Handbook, MRSD Page 8 of 41

List of Test ToolsTest toolsNo.Test Tools1.Point sources2.Source mounting for point source3.Lead mask4.Quadrant bar phantom with bar widths of approximately 2, 3, 3.5 and 4 mm5.Slit mask for X-axis and Y-axis6.Flood phantom7.Planar sensitivity phantom8.Movable stand with mounting for point source9.Measuring tape10.Quadrant bar or orthogonal hole test pattern (OHTP) phantom11.An accurate ruler12.Tomographic uniformity phantom13.COR / MHR / IRC source holder14.57Co flood sourceOthers ToolsNo.Test Tools1.SMPTE Test Pattern Installed2.Printer To Print ImagesScintillation Camera & SPECT Version August 2015: Technical QC Protocol Handbook, MRSD Page 9 of 41

D. Testing ProceduresScintillation Camera & SPECT Version August 2015: Technical QC Protocol Handbook, MRSD Page 10 of 41

TEST PARAMETER 1: PHYSICAL INSPECTIONTest ParameterPhysical InspectionObjectiveTo inspect a scintillation camera, control console, computer, and datastorage, display devices for shipping damage, production and designflaws.Test ToolNot aqpplicable.Test Procedure1.Detector housing and support assemblyInspect the aluminium casing surrounding the NaI(Tl) crystal for signs ofindentation or puncture and the gantry for loose parts or mechanicaldifficulties. Move the gantry, bed and detector head through all possiblemotions and to the fullest extent of travel, noting any grinding noises,loose parts, inability to move, or improperly functioning controls.2.Control panels & Hand controlInspect the switches and other controls for loose or broken parts. Checkfor switches that do not throw securely. Check computer keyboards andaccessories for proper operation. Inspect the hand control for propermechanical operation and confirm that the cable has acceptable strainrelief at maximum extension.3.Image display devicesInspect the display screen for scratches, fingerprints, dust or otherdebris. Inspect the image monitors for any interference patterns, rolling,lines or other signs of improper operation or electrical interference.4.Image recording devices/data storageInspect the performance of the recording devices/data storage.5.Emergency devicesTest each emergency button to ensure that all gantry motion ceaseswhen each button is activated.6.CollimatorsInspect the collimators for damage. Load each onto the camera headand ensure that the collimator mounting mechanism is aligned and isworking properly. While mounted, activate each motion sensor device toensure that all gantry motion ceases.7.Electrical connections, fuses and cablesInspect for any loose or broken cable connectors and pinched ordamaged cables. Locate all fuses and circuit breakers to enable promptchecking during equipment failure. Ensure that cables are housed,wherever possible, in conduits and are not loose on the floor. Also,ensure that they were placed to allow maximum patient access.Scintillation Camera & SPECT Version August 2015: Technical QC Protocol Handbook, MRSD Page 11 of 41

Result8.Operation and service manualsCheck that all appropriate documentation (softcopy and hardcopy) isavailable, including performance specifications.9.Room condition - Temperature & Humidity (daily)Make sure that the room temperature and humidity are according tomanufacturer’s specification.No. Visual inspection of component1. Detector housing and supportassembly2. Control panels & Hand control3. Image display devices4. Image recording devices/datastorage5. Emergency devices6. Collimators7. Electrical connections, fuses andcables8. Operation and service manuals9. Room condition - Temperature &Humidity (daily)Pass/Fail CommentsToleranceFunctional and according to manufacturer’s specifications.Test FrequencyAcceptanceDaily (except for emergency button)AnnuallyScintillation Camera & SPECT Version August 2015: Technical QC Protocol Handbook, MRSD Page 12 of 41

TEST PARAMETER 2: BACKGROUND COUNT RATETest ParameterBackground Count RateObjectiveTo check the background count rate of a scintillation camera under theconditions of routine clinical imaging with a particular radionuclide.Test ToolNot applicable.Test Procedure1.Mount the collimator to be used for daily examination (all collimatorsshould be used for acceptance test) and turn the detector head to ascanning position downward.2.Position the detector head over the center of the table bed.3.Adjust all the routine settings on the gamma camera console, for aroutine acquisition by the radionuclide concerned.4.Perform a count measurement for a 100 sec recording time without anyradiation source at the vicinity of the camera.5.Record the counts and calculate the background count rate.ResultRecord the results.Tolerance 20% of the reference value.Test FrequencyAcceptanceDailyScintillation Camera & SPECT Version August 2015: Technical QC Protocol Handbook, MRSD Page 13 of 41

TEST PARAMETER 3: CENTRING PULSE HEIGHT ANALYSER (PHA) WINDOW SETTINGSTest ParameterCentring Pulse Height Analyser (PHA) Window SettingsObjectiveTo test that all preset PHA windows for clinical imaging are properlycentred for every radionuclide to be used with the scintillation camera.Test ToolTest Procedure1.Point sources consisting of the radionuclides concerned, with sourcestrength of about 10 MBq (250 µCi) or source strength specified bymanufacturer, in suitable containers.2.Source mounting for point source.1.Remove the collimator from the detector head (if applicable). Align thehead and the source mounting.2.Mount the source in the source mounting.3.In the acquisition mode, select the default energy setting for theradionuclide concerned, which sets the energy and window width to beused for clinical imaging.4.Observe the display to ensure that the respective photopeaks arecentred in the window settings. If they are not centred, manually adjust( 5%) each photopeak so that it is centred. Record the peak value thatproperly centres the photopeak.5.Remove the source.6.Repeat steps (2) – (6) for other radionuclides in turn (during acceptancetest only).Mount the collimator to be used for daily examination (allcollimators should be used for acceptance test) and turn the detectorhead to a scanning position downward.ResultRecord the results.TolerancePhotopeaks must be properly centred.Test FrequencyAcceptanceDailyScintillation Camera & SPECT Version August 2015: Technical QC Protocol Handbook, MRSD Page 14 of 41

TEST PARAMETER 4: ENERGY RESOLUTIONTest ParameterEnergy ResolutionObjectiveTo test the intrinsic response of a scintillation camera to a spatiallyuniform flux of incident photons over the field of view using a symmetric(centred) energy window over the photopeak.Test ToolTest Procedure1.Point source consisting of 10 – 20 MBq (0.3 – 0.5 mCi) of 99mTc and57Co disc (source strength specified by manufacturer) in a suitablecontainer mounted in the source holder.2.Source mounting for point source.3.Lead mask.1.Remove the collimator. Align the head and the source mounting.2.Place the lead mask to establish Useful Field of View (UFOV) andposition the stationary support upon the camera.3.Place the 99mTc point source on the support at a distance of 5 UFOVfrom the detector face. The source should be positioned to give the leastpossible scatter component.4.A second radionuclide, 57Co, shall be employed as a reference in orderto determine the keV per channel calibration factor.5.Remove the source, the stationary support and the lead mask.ResultRecord the results.ToleranceManufacturer’s specifications.Test FrequencyAcceptanceUnder condition(1): Whenever detector system performance is suspected to havechanged significantly.Scintillation Camera & SPECT Version August 2015: Technical QC Protocol Handbook, MRSD Page 15 of 41

TEST PARAMETER 5: INTRINSIC FLOOD FIELD UNIFORMITY FOR 99MTCTest ParameterIntrinsic Flood Field Uniformity For 99mTcObjectiveTo test the intrinsic response of a scintillation camera to a spatiallyuniform flux of incident photons over the field of view using a symmetric(centred) energy window over the photopeak.Test ToolTest Procedure1.Point source consisting of 10 – 20 MBq (0.3 – 0.5 mCi) or sourcestrength specified by manufacturer of 99mTc in solution in a suitablecontainer mounted in the source holder. The count rate should not begreater than 50 kcounts/s with the manufacturer’s default PHA window.2.Source mounting for point source.1.Remove the collimator from the detector head. Align the head and thesource mounting.2.Mount the source on the support.3.Center a 15 – 20% energy window on the 99mTc photopeak.4.Acquire an image of minimum 15 x 106 counts in selectable pixel matrixand record the time.5.Remove the source and the support.ResultRecord the results.a)Qualitative method (comparable with reference image)Inspect the image for any brightness variations caused by lack ofuniformity. For proper examination the contrast must be more than 10%.Variations smaller than 5% can be detected.b)Quantitative method for UFOV and CFOV (Each Comprises Of IntegralUniformity and Differential Uniformity)Calculate the integral and differential uniformity for the UFOV and theCFOV, using the available software or if this is not the case, thefollowing method can be used, in order to calculate uniformity:1) Normalize the image with a nine-point filter.2) Determine the minimum (min) and the maximum (max) countrate in the pixels within the UFOV and the CFOV.3) The integral uniformity (IU), can be calculated using theequation: max - min IU x100 % max min Scintillation Camera & SPECT Version August 2015: Technical QC Protocol Handbook, MRSD Page 16 of 41

4) Determine the maximum count difference in any 5 contiguouspixels for each row or column of pixels within the UFOV and theCFOV. Find the highest (high) and lowest (low) value of thesedifferences.5) The differential uniformity (DU), is given by: high - lowDU high lowToleranceTest Frequencya)b) x100 % Qualitative method : No cold or hot spot.Quantitative method : Manufacturer’s specifications.AcceptanceWeeklyNote: This test should be done daily if Test Parameter 7 is done weekly.Scintillation Camera & SPECT Version August 2015: Technical QC Protocol Handbook, MRSD Page 17 of 41

TEST PARAMETER 6: INTRINSIC SPATIAL RESOLUTIONTest ParameterIntrinsic Spatial ResolutionObjectiveTo test the intrinsic spatial resolution of scintillation camera in terms ofthe FWHM of its line spread function.Test ToolTest Procedure1.Point source consisting of 20 – 40 MBq (0.5 – 1 mCi) or source strengthspecified by manufacturer of 99mTc in solution in a suitable container.2.Quadrant bar phantom with bar widths of approximately 2, 3, 3.5 and 4mm or slit mask for X-axis and Y-axis.a)Qualitative (Visual Image) method using quadrant bar phantom1.Remove the collimator. Align the head and the source mounting.2.Place the point source on the support.3.Center a 15 – 20% energy window on the 99mTc photopeak.4.Position the quadrant bar phantom upon the camera with the lead barsaligned with the X and Y-axis.5.Acquire an image of 6 x 106 counts depending on the size of the FOV.The matrix size of the image should be 512 x 512 (pixels).6.Rotate the quadrant bar phantom at 90o, 180o, 270o and 360o andrepeat acquiring.7.Remove the source, the stationary support, the quadrant bar phantom.Replace the collimator.b)Quantitative method using slit mask1.Remove the collimator. Align the head and the source mounting.2.Cover the detector with the slit mask and center its central slit on thedetector, perpendicular to the axis of measurement.3.Careful alignment is necessary. Place the stationary support upon thecamera.4.Place the point source on the support.5.Center the 20% energy window on the 99mTc photopeak.6.Acquire an image by a count rate not greater than 10000 counts/secs.7.Acquisition should be performed in X as well as Y-axis.Scintillation Camera & SPECT Version August 2015: Technical QC Protocol Handbook, MRSD Page 18 of 41

8.ResultToleranceTest FrequencyRemove the source, the support and the slit phantom. Replace thecollimator.Record the results.a)b)Qualitative method : 20% of manufacturer’s specifications.Quantitative method : 10% of reference value.Acceptance (Qualitative and Quantitative methods)Semi-annually (Qualitative method)Scintillation Camera & SPECT Version August 2015: Technical QC Protocol Handbook, MRSD Page 19 of 41

TEST PARAMETER 7: SYSTEM FLOOD FIELD UNIFORMITY (ON COMMON CLINICALLYUSED COLLIMATOR)Test ParameterSystem Flood Field Uniformity (On Common Clinically UsedCollimator)ObjectiveTo test the system flood field response of a scintillation camera for allgeneral purpose collimators used.Test ToolFlood phantom containing 200 – 400 MBq (5 – 10 mCi) ofsolution or 57Co flood source of similar activity.Test ProcedureTest FrequencyTc in1.Mount the collimator on the detector head for daily examination (allcollimators should be used for acceptance test except pinholecollimator).2.Place the 57Co or the 99mTc flood source upon the collimator.3.Center a 20% window on the radionuclide photopeak.4.Acquire an image at preset 2 x 106 counts. If uniformity correction circuitis fitted at the system, it must be enabled.5.Remove the flood source.6.Repeat the above steps for all common clinically used collimators.ResultTolerance99mRecord the results.a)b)Qualitative method : No cold or hot spot.Quantitative method : Manufacturer’s specifications.AcceptanceDailyNote: This test should be done weekly if Test Parameter 5 is done daily.Scintillation Camera & SPECT Version August 2015: Technical QC Protocol Handbook, MRSD Page 20 of 41

TEST PARAMETER 8: SYSTEM SPATIAL RESOLUTIONTest ParameterSystem Spatial ResolutionObjectiveTo test the system spatial resolution of a scintillation camera in terms ofthe FWHM of its line spread function. This test should be performed foreach parallel hole, low energy collimator.Test ToolTest Procedure99m1.Flood phantom containing about 400 - 800 MBq (10 - 20 mCi) ofor 57Co flood source of similar source strength.2.Quadrant bar phantom with bar widths and bar spacings ofapproximately 2, 3, 3.5 and 4 mm.a)Qualitative (Visual Image) method1.Mount the collimator to be tested on the camera.2.Place the 57Co or 99mTc flood source upon the quadrant bar phantom.3.Center a 15 – 20% energy window on the 99mTc photopeak.4.Position the quadrant bar phantom upon the camera with the lead barsaligned with the X and Y-axis.5.Acquire an image of minimum 6 x 106 counts depending on the size ofthe FOV. The matrix size of the image should be 512 x 512 (pixels).6.Rotate the quadrant bar phantom at 90o, 180o, 270o and 360o andrepeat acquiring.7.Remove the source, the stationary support and the quadrant barphantom.8.Repeat this procedure for all low energy collimators except the pinholeone.9.Remove the source and the quadrant bar phantom.b)Quantitative method1.Mount the collimator to be tested on the camera and turn to facevertically upward.2.Position the line source phantom on the face of the collimator and alignit with the X-axis and then with the Y-axis of the camera.3.Center the 20% energy window on the selected isotope photopeak.Scintillation Camera & SPECT Version August 2015: Technical QC Protocol Handbook, MRSD Page 21 of 41Tc

4.Acquire an image in a count rate less than 10000 counts/seconds.5.Repeat this procedure for all low energy collimators except the pinhole.6.Remove the line source phantom.ResultToleranceTest FrequencyRecord the results.a)b)Qualitative method : 20% of manufacturer’s specifications.Quantitative method : 10% of reference value.AcceptanceSemi-annuallyScintillation Camera & SPECT Version August 2015: Technical QC Protocol Handbook, MRSD Page 22 of 41

TEST PARAMETER 9: SYSTEM PLANAR SENSITIVITYTest ParameterSystem Planar SensitivityObjectiveTo test the count rate response of a scintillation camera to aradionuclide source of known radioactivity.Test ToolPlanar sensitivity phantom containing an accurately known amount ofradioactivity, about 40 – 160 MBq (1 – 4 mCi) of 99mTc or otherradionuclide such as 131I, in solution.Test Procedure1.Mount a low energy, parallel hole collimator on the detector head. Turnthe head to face vertically upward.2.Cover the face of the collimator with a plastic sheet. Place the phantomcontaining the 99mTc 10 cm from the surface of the covered collimatorface.3.Center the manufacturer’s default PHA window on the photopeak, or thewindow width used by the manufacturer in determining the specifiedperformance values.4.Collect an image over a total time of 100 s. Record the total counts inthe image frame and the exact time of day.5.Remove the phantom and count the background for the same timeperiod. Record the total counts in the image frame.6.Repeat steps (1) – (5) for all other low energy multihole collimators.7.Repeat steps (1) – (5) with the phantom containing 67Ga or 111In formedium energy multihole collimators (if applicable) and 131I for highenergy multihole collimators (if applicable).ResultRecord the results.ToleranceManufacturer’s specifications (Acceptance). 10% of acceptance value (Semi-annually).Test FrequencyAcceptanceSemi-annuallyScintillation Camera & SPECT Version August 2015: Technical QC Protocol Handbook, MRSD Page 23 of 41

TEST PARAMETER 10: COUNT RATE PERFORMANCETest Parameter10.1Intrinsic Count Rate PerformanceObjectiveTo test the intrinsic count rate performance of a scintillation camera interms of the count rate corresponding to a 20% count loss (two sourcemethod).Test ToolTwo point sources each consisting of about 2 MBq (50 Ci) ofsolution in suitable containers.Test ProcedureResult99mTc in1.Remove the collimator. Align the head and the source mounting.2.Place the stationary support upon the gamma camera.3.Center the 20% energy window on the 99mTc photopeak.4.Place the first source at more than 1m from the surface of the crystaland near its center axis, away of objects to minimize scatter radiation.5.Record the counts for 100 seconds - sufficient time to accumulate 106counts. Register the count rate R1.6.Place the second source beside the first and record the R12 count rate.7.Remove the first source and register the second source count rate, R2.8.Remove the second source and record the background count rate.9.Repeat the count rate measurements, R2, R12, R1 placing the sourcesin reverse order.10.Remove the last source and the lead mask. Replace the collimator.Record the results.Express all data as net count rates (counts/s) corrected forbackground.Calculate for each set of data the pulse pair resolving time, in secondsfrom: 2 R12 R1 R2 2 R R2 x ln 1 R12 Where R1 and R2 are the net count rates of the first and secondsources and R12 is the net count rate of the two sources together, all incounts per second. Average the two values to obtain .Calculate the Input Count Rate (ICR) for a 20% loss, R-20%, from:R-20% 10 0.2231x ln 8 1Scintillation Camera & SPECT Version August 2015: Technical QC Protocol Handbook, MRSD Page 24 of 41

Calculate the Observed Count Rate (OCR) for a 20% count loss, C-20%,from:C-20% 0.8xR 20%ToleranceThe R-20% is within 10% of the manufacturer’s specifications.Test FrequencyAcceptanceSemi-annuallyScintillation Camera & SPECT Version August 2015: Technical QC Protocol Handbook, MRSD Page 25 of 41

Test Parameter10.2ObjectiveTest ToolTest ProcedureMaximum Count RateTo test the maximum count rate of s scintillation camera.99m1.Point source consisting of about 4 MBq (100 – 500 µCi) ofsolution in a suitable container.2.Movable stand with mounting for point source.3.Measuring tape.1.Remove the collimator from the detector head. Turn the head to facehorizontally.2.Centre the manufacturer’s default PHA window on the photopeak.3.Mount the source on the movable stand. Position the latter so that thesource is on the central axis of the detector. To minimize radiationscatter, the source should not be close to other objects. Remove thecollimator. Align the head and the source mounting.Tc inDetector HeadSyringe containingradioactive materialwrapped in plasticbag or gloveStand with wheelFigure 1: Positioning of point source in relation to detector.4.Register the count rate as the source is moved progressively closer tothe detector face. The count rate will increase to a maximum and thendecrease. Record the maximum count rate.5.Remove the source and stand. Replace the collimator.ResultRecord the results.ToleranceThe maximum count rate is within 10% of the manufacturer’sspecifications.Test FrequencyAcceptanceSemi-annuallyScintillation Camera & SPECT Version August 2015: Technical QC Protocol Handbook, MRSD Page 26 of 41

TEST PARAMETER 11: MULTIPLE WINDOW SPATIAL REGISTRATION (WITH GALLIUM 67APPLICATION)Test ParameterMultiple WindowApplication)ObjectiveTo test that the images acquired at different photon energiessuperimpose when more than one PHA is used simultaneously in anadditive or subtractive mode.Test ToolTest ProcedureSpatialRegistration(WithGallium671.Point source consisting of about 40 MBq (1 mCi) of 67Ga in solution ina small vial, in a lead shield 6 mm thick and having a circular aperture3 mm in diameter.2.Quadrant bar or orthogonal hole test pattern (OHTP) phantom.1.Turn the head to face vertically upward.2.The activity of the source must ensure a count rate lower than 10,000c/s through each symmetric 20% photopeak energy window.3.Images, with 20,000 counts each, shall be acquired with thesource located at 5 specific points: 67GaOne point at the center of UFOV.Four off-central points, at a distance from the center equal to0.75 of the UFOV radius and on X , X-, Y , Y- axisrespectively.4.Acquire images in 256 x 256 matrix size so that pixel size is below 2mm.5.Separate images should be acquired through separate energywindows of the 67Ga energies and at each location of the source.ResultRecord the results.Tolerance 10% of displacement.Test FrequencyAcceptanceAnnually (Applicable only to the system using radionuclideimaging purposes)67Scintillation Camera & SPECT Version August 2015: Technical QC Protocol Handbook, MRSD Page 27 of 41Ga for

TEST PARAMETER 12: DETECTOR HEAD SHIELDING LEAKAGETest ParameterDetector Head Shielding LeakageObjectiveTo test that the detector head of scintillation camera responds only toradiation incident upon the crystal after transmission through thecollimator.Test ToolPoint sources consisting of about 4 MBq (100 µCi) of 99mTc and aradionuclide with a photon energy corresponding to the specifieddesign energy of the camera in a suitable container.Test Procedure1.Mount a collimator appropriate to the gamma radiation energy of thesource on the detector head.2.Centre the manufacturer’s default PHA window for the radionuclideconcerned on the photopeak.3.Position the source consecutively at thirteen sites around the detectorhead shielding and record the number of counts at each site for apreset time of 100 s. In addition, investigate sites of joints in theshielding, exit points of cables and other reduced shielding areas.4.Position the source in the centre of the field of view at a distance of 10cm from the face of the collimator. Record the number of counts for apreset time of 100 s.5.Remove the source and measure the background count, B, for thesame time Figure 2: Thirteen sites at which to position the source in order to test forshielding leakage.ResultRecord the results.ToleranceNegligible (No leakage).Test FrequencyAcceptanceScintillation Camera & SPECT Version August 2015: Technical QC Protocol Handbook, MRSD Page 28 of 41

TEST PARAMETER 13: IMAGE DISPLAYTest Parameter13.1ObjectiveTest ToolTest ProcedureVisual DisplayTo determine that all of the information in the video signal isdisplayed on the video display.1.The Society of Motion Pictures and Television Engineers(SMPTE) test pattern installed in the CT scanner.2.Display Monitor (CRT or LCD).1.Display the SMPTE test pattern on the (CRT or LCD).Figure 3: Image of SMPTEResult2.Adjust the window width to just encompass the range of numberscomprising the SMPTE test pattern.3.Adjust the window level to either the lower or middle value of thewindow so that the entire test pattern is visible.4.If required, clean the front surface of the monitor and reduceroom illumination to the normal viewing level for better evaluationcondition.1.The 5% patch should just be visible inside of the 0% patch.2.The line pair pattern at the center and all four sides of the imageshould be clearly resolvable.3.The 95% patch should be visible inside the 100% patch.4.The alphanumeric should be sharp and clear.ToleranceManufacturer’s specifications.Test FrequencyAcceptanceWeeklyScintillation Camera & SPECT Vers

4. Quadrant bar phantom with bar widths of approximately 2, 3, 3.5 and 4 mm 5. Slit mask for X-axis and Y-axis 6. Flood phantom 7. Planar sensitivity phantom 8. Movable stand with mounting for point source 9. Measuring tape 10. Quadrant bar or orthogonal hole test pattern (OHTP) phantom 11. An accurate ruler 12. Tomographic uniformity phantom 13.

Related Documents:

EGP Exterior Gateway Protocol OSPF Open Shortest Path First Protocol IE-IRGP Enhanced Interior Gateway Routing Protocol VRRP Virtual Router Redundancy Protocol PIM-DM Protocol Independent Multicast-Dense Mode PIM-SM Protocol Independent Multicast-Sparse Mode IGRP Interior Gateway Routing Protocol RIPng for IPv6 IPv6 Routing Information Protocol PGM

SNMP V1/V2/V3 Simple Network Management Protocol SNTP Simple Network Time Protocol RFC RFC 768 UDP (User Datagran Protocol) RFC 783 TFTP (Trivial File Transfer Protocol) RFC 791 IP (Internet Protocol) RFC 792 ICMP (Internet Control Message Protocol) RFC 793 TCP (Transmission Control Protocol) R

PERFORMANCE QUALIFICATION PROTOCOL FOR CARTON PACKING MACHINE PROTOCOL No.: Prepares the performance qualification protocol. Ensures that the protocol is in compliance with current policies and procedures on system Qualification. Distributes the finalized protocol for review and approval signatures.

Rebright & Texas Ruby Jessner Solutions Ultrasonic Rejuvenation Facial Protocol Dry Skin and Aging Protocol Oily Skin and Acne Protocol Rosacea Protocol Microdermabrasion Peel Protocol Trio Rejuvenation Facial Peel Protocol Body Peel Protocol . ADVANCED REJUVENATING CONCEPTS 102 .

BACnet Application Protocol BACnet Network Protocol BACnet Virtual Link Protocol User Datagram Protocol (UDP) Internet Protocol (IP) Logical Link Control ISO8802-2 / IEEE802.2 Ethernet ISO8802-3 (IEEE802.3) Application Layer Network Layer Data Link Layer Physical Layer 10666Z02de Figure 2: BACnet over Ethernet/IP protocol layers

4.1 Program Manager (PO) 3 4.2 Project Manager (PM) 4 4.3 Corporate Health and Safety Manager (CHSO) 4 4.4 Quality Assurance/Quality Control (QA/QC) Manager 4 4.5 Site Quality Control Supervisor (QCS) 5 4.6 Technical Quality Control Assistant (QCA) 6 4.7 Quality Control Laboratory (QCL) 6 4.8 Quality Control Field Testing (QCFT) 6

Protocol Creation Guidance OnCore Clinical Trials Management System . ONCORE USERS MANUAL 2 ONCORE USERS MANUAL Protocol Creation and Setup PC Console: CREATE A NEW PROTOCOL Purpose: This section covers setting up a new protocol. In This Section: Setting up basic protocol information

(1) This water management protocol may be cited as the Cape York Water Management Protocol. (2) Reference in this document to ‘this protocol’ means the Cape York Water Management Protocol. 2 Commencement of the water management protocol (1) The following provisions commence on