DE-5000 Portable, Full-Featured LCR Meter User And Service .

2y ago
136 Views
2 Downloads
1.84 MB
52 Pages
Last View : 16d ago
Last Download : 2m ago
Upload by : Karl Gosselin
Transcription

PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT DE-5000Portable, Full-FeaturedLCR MeterUser and Service ManualIET LABS, INC.534 Main Street, Westbury, NY 11590www.ietlabs.comTEL: (516) 334-5959 (800) 899-8438 FAX: (516) 334-5988

PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT IET LABS, INC.534 Main Street, Westbury, NY 11590www.ietlabs.comTEL: (516) 334-5959 (800) 899-8438 FAX: (516) 334-5988

PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT DE-5000Portable, Full-FeaturedLCR MeterUser and Service ManualCopyright 2011 IET Labs, Inc.Visit www.ietlabs.com for manual revision updatesDE-5000 im/October 2011IET LABS, INC.534 Main Street, Westbury, NY 11590www.ietlabs.comTEL: (516) 334-5959 (800) 899-8438 FAX: (516) 334-5988

PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT IET LABS, INC.534 Main Street, Westbury, NY 11590www.ietlabs.comTEL: (516) 334-5959 (800) 899-8438 FAX: (516) 334-5988

WARRANTYWe warrant that this product is free from defects in material and workmanship and, when properly used,will perform in accordance with applicable IET specifications. If within one year after original shipment,it is found not to meet this standard, it will be repaired or, at the option of IET, replaced at no charge whenreturned to IET. Changes in this product not approved by IET or application of voltages or currents greaterthan those allowed by the specifications shall void this warranty. IET shall not be liable for any indirect,special, or consequential damages, even if notice has been given to the possibility of such damages.THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED,INCLUDING BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF MERCHANTABILITY ORFITNESS FOR ANY PARTICULAR PURPOSE.

WARNINGOBSERVE ALL SAFETY RULESWHEN WORKING WITH HIGH VOLTAGES OR LINE VOLTAGES.Dangerous voltages may be present inside this instrument. Do not open the caseRefer servicing to qualified personnelHIGH VOLTAGES MAY BE PRESENT AT THE TERMINALS OF THIS INSTRUMENTWHENEVER HAZARDOUS VOLTAGES ( 45 V) ARE USED, TAKE ALL MEASURES TOAVOID ACCIDENTAL CONTACT WITH ANY LIVE COMPONENTS.USE MAXIMUM INSULATION AND MINIMIZE THE USE OF BARECONDUCTORS WHEN USING THIS INSTRUMENT.Use extreme caution when working with bare conductors or bus bars.WHEN WORKING WITH HIGH VOLTAGES, POST WARNING SIGNS ANDKEEP UNREQUIRED PERSONNEL SAFELY AWAY.CAUTIONDO NOT APPLY ANY VOLTAGES OR CURRENTS TO THE TERMINALS OF THISINSTRUMENT IN EXCESS OF THE MAXIMUM LIMITS INDICATED ONTHE FRONT PANEL OR THE OPERATING GUIDE LABEL.

Table of ContentsChapter 1: Introduction . 11.1. Overview .11.2. Introduction to measuring principles .21.2.1. What is impedance . 21.2.2. Measuring impedance. 41.3. Equivalent circuit .61.4. Instrument layout.71.5. LCD display layout .10Chapter 2: Operation. 122.1. How to obtain optimum precision .122.2. Default settings .122.3. Zeroing the meter .142.4. Attaching DUT’s to the meter .172.5. Primary measurements and functions .192.5.1. Measuring inductance, capacitance, and resistance . 192.5.2. Measuring dissipation, quality, ESR, and phase angle. 222.5.3. Sorting components . 242.5.4. Making relative measurements . 262.6. Additional settings.302.6.1. Selecting test frequency . 302.6.2. Making measurements in series and parallel . 32- iii -

2.7. Additional features .332.7.1. Connecting to a PC . 332.7.2. Using the backlight. 352.7.3. Holding a reading on the display . 352.8. Replacing Batteries .36Chapter 3: Specifications. 373.1. General specifications .373.2. Accuracy specifications .393.3. Ordering information .42- iv -

Chapter 1: Introduction1.1. OverviewThe DE-500 is a portable, high-performance LCR meter that isfull-featured yet cost effective. It measures in true 4-wire Kelvinmode and rivals the capabilities and options of many of its benchcounterparts.It measures:Ls/Lp-- Series and parallel inductanceCs/Cp-- Series and parallel capacitanceRs/Rp-- Series and parallel resistance (ac)Rdc/Rp -- Series and parallel resistance (dc)ESR/Rp -- Series and parallel equivalent resistanceD-- Dissipation factorQ-- Quality factorΘ-- Phase angleThis LCR meter can transfer data to a PC via a standard, fullyisolated, optical IR-USB interface.It also features a Sortingmode, allowing users to quickly sort components.DE-5000 has automatic LCR selection. This allows the user tomeasure the L/C/R components in Auto LCR mode withouthaving to select the type of measurement.-1-

To accommodate various test requirements, the DE-5000 offersselectable test frequencies: 100 Hz / 120 Hz / 1 kHz / 10 kHz /100 kHz.The unit is powered by a standard 9V battery.For additionalconvenience, it may also use an optional ac adapter(DE-5000-AC).To transfer data to a PC, the unit comes with a built-in IRinterface.IET offers an optional Data Transfer Kit(DE-5000-DTK). This kit includes: IR-USB interface adapter,a USB cable, and a CD with software for the PC.1.2. Introduction to measuring principles1.2.1. What is impedanceImpedance (Z) consists of resistance (real part) and reactance(imaginary part). Series impedance (Zs) can be defined as acombination of series resistance (Rs) and series reactance (Xs). Itcan be represented mathematically as magnitude Z (Rs2 Xs2)at a phase angle Θ.-2-

Zs Rs jXs or Zs ΘRs Zs cosΘXs Zs sinΘXs/Rs tanΘΘ tan-1(Xs/Rs)There are two types of reactance. One is inductive reactance –XL, and the other is capacitive reactance – XC.If Θ 0, the reactance is inductive. If Θ 0, the reactance iscapacitive.The inductive and capacitive reactances (XL and XC) can bedefined as:XL 2πƒLXC 1 (2πƒC)Where:L InductanceC Capacitanceƒ signal frequency)-3-

1.2.2. Measuring impedanceImpedance can be measured in series or in parallel. In parallelmode, impedance can be represented as reciprocal of admittance(Y). The admittance can be defined as Y G jB, where:G ConductanceB SusceptanceSeries impedanceParallel admittanceY 1/Z 1/Rp 1/jXp G jBRs Series resistanceXs Series reactanceCs Series capacitanceLs Series InductanceRp Parallel resistanceXp Parallel reactanceCp Parallel capacitanceLp Parallel inductanceTo understand the ratio of resistance and reactance, it isimportant to consider two factors: quality factor (Q) anddissipation factor (D).Usually Q is used when measuringinductance and D is used when measuring capacitance. D isdefined as the reciprocal of Q.Q 1/D tanΘQ Xs/Rs 2πf Ls/Rs ½πf CsRsQ B/G Rp/ Xp Rp/2πf Lp 2πf CpRp-4-

Both Rs and Rp are part of the equivalent circuit of capacitorsand inductors. When measuring capacitance and inductance, itis best to use the settings as shown in the table ghParallelCapacitanceInductance-5-

1.3. Equivalent circuit-6-

1.4. Instrument layoutFront panelSideGUARD provides a shield to reduce noise for device under test(DUT), test leads, and other equipment.-7-

21. TL-21Alligator –lead test-lead adapter(4-wire joined at alligator clips)22. TL-22SMD tweezers, 4-wire(optional)Rear23. TL-23Guard LineIR- USB interface(optional)-8-

.21.22.23.LCD displayPOWERTurns the instrument on/off.LCR auto mode, Inductance,Capacitance, Resistance and DCLCR AUTOresistance measurement selectionFREQTesting frequency selectionBacklight displaySORTINGSorting mode controlPC UART output controlCALOpen/Short calibration modeD/Q/ESR/ΘD/Q/ ESR/Θ parameters selectionSetup menu controlSETUP(in sorting mode )SER/PALSeries and Parallel selectionSetup menu controlENTER(in sorting mode )REL%Relative modeHOLDData holdInput sockets and Terminals (4-terminal)AC adapter plugBattery coverTilt-StandIR SlotIR-USB adapter (optional)TL-21 Alligator-clip test-lead adapterTL-22 SMD tweezers (optional)TL-23 Guard Line-9-

1.5. LCD display 12.13.AutoLCR Ls/LpCs/CpRs/RpDCRSorting function is enabledTolerance indicator in sorting mode : 0.25%, 0.5%, 1%, 2%, 5%, 10%, 20%, & 80%-20%Testing frequency indicator:1kHz,10kHz,100kHz,100Hz & 120HzPC connection is activeBattery capacity indicatorRange selection is enabled on setup menu ofsorting modeAuto range for L, C or R measurementsChecking for L/C/R mode automaticallyRelative function is enabledInductance in series or parallel mode is activeCapacitance in series or parallel mode is activeac resistance in series or parallel mode is activedc resistance mode is selected- 10 -

Ω 23.24.25.%26.27.28.29.MkΩDissipation factor, Quality factor or Phase angle isactive for L/C measurement modeac Resistance in parallel mode is activeOpen/Short calibration modeData HoldAuto power off modeSeries equivalent resistance modeSecondary DisplayPhase angleUnit for Resistance (Ω, kΩ and MΩ )– on secondary displayUnit for Capacitance ( pF, nF, µF and mF )– on secondary displayUnit for Inductance ( µH, mH and H)– on secondary displayThe percentage display in relative mode– on secondary displayPrimary DisplayUnit for Resistance ( Ω, kΩ and MΩ )– on primary displayUnit for Inductance ( µH, mH and H)– on primary displayUnit for Capacitance ( pF, nF, µF and mF )– on primary display30.Bar-graph displaySpecial Indication CharactersIndicates short calibrationIndicates open calibration- 11 -

Chapter 2: Operation2.1. How to obtain optimum precisionTo access optimum precision for all L, C, and R measurements,especially at the highest and the lowest ranges, zero the instrumentbefore use (pages 16-18).To secure the specified accuracy, connect the device under test(DUT) to the measuring socket, or use either TL-21 (standardaccessory) or TL-22 (optional accessory).If you are using test leads other than the ones specified above, use4-wire leads and avoid using long lead wires to reduce measurementerrors.2.2. Default settingsWhen the power is turned on, the monitor displays all symbols for 2seconds as shown below.- 12 -

When the meter is powered by the battery, it is in anautomatic-power-off mode. APO is shown on the display.In thismode, if the unit is inactive for 5 minutes, it shuts itself off.First,the buzzer beeps three times to remind the user, then OFF isdisplayed on the monitor as shown below while the unit powersdown.Note that when the unit is powered via an ac adapter, theautomatic-power-off mode is inactive.The default settings for the meter set LCR in auto mode and testfrequency at 1 kHz.The battery condition is continuously displayed.the battery capacity is full.means thatmeans that battery power is lowand the battery needs to be replaced.The LCR meter uses beeps to indicate whether a particular key has afunction in a given mode. If a functional key is pressed, there is asingle beep.If a non-functional key is pressed, there is a dual beep.- 13 -

2.3. Zeroing the meterZeroing the instrument gets better accuracy for impedancemeasurements. The purpose of this procedure is to reduce theparasitic effect of the test fixture.ZM is the total impedance measured on the device under test (DUT)by a test fixture which has some parasitic impedance.ZM (Rs jωLs) ((Go jωCo)-1 ZDUT)ZDUT is the target impedance user wants to measure. It is necessaryto use the zeroing process to cancel the effect of Rs jωLs andGo jωCo.Ex. Operation for open andshort calibration with TL-21Open Cal.Short Cal.- 14 -Ex. Operation for open and shortcalibration with .TL-22Open Cal.Short Cal.

To zero the meter, proceed as follows:1. Make sure the leads are completely disconnected.2. Press the CAL key for 2 seconds.The monitor should display OPEn as shown below.3. Pres the CAL key again.The unit should begin a countdown as it performs OPENcalibration.After the countdown is complete, the monitor should say PASSas shown below.If it says fail, the procedure has to be restarted.- 15 -

4. Connect the test leads for form a short circuit.5. Press the CAL key again.The monitor should display Srt as shown below.6. Press the CAL key one more time.The unit should begin a countdown as it performs SHORTcalibration.After the countdown is complete, the monitor should say PASSas shown below.If it says fail, the procedure has to be restarted.7. Press the CAL key one more time to exit OPEN/SHORTcalibration mode.- 16 -

2.4. Attaching DUT’s to the meterDevices under test (DUT’s) may be connected to themeter as follows: Insert DIP component leads to the sockets directly. Attach Alligator-clip test-lead adapter (TL-21)4-Wire AlligatorTest Lead Case(TL-21)Guard line (TL-23) providesa shield for DUT, preventinginterference when measuring- 17 -

Attach SMD tweezers (TL-22, optional).Guard line (TL-23) providesa shield for DUT, preventinginterference whenmeasuring high-impedancecomponents4-wire SMDTweezers(TL-22)- 18 -

2.5. Primary measurements and functions2.5.1. Measuring inductance, capacitance, andresistanceThe DE-5000 starts out in Auto LCR mode which ly – either inductance (L), capacitance (C),or resistance (R).Dc resistance (DCR) can only beselected manually. The value of impedance is shown ndaryshowsthedisplaysecondaryparameter – either quality factor (Q), dissipation factor(D), or phase angle (Θ). The secondary parameter isbased on the L/C/R measurement.When the meter automatically selects impedanceautomatically, it uses the following procedure:If Q 0.2, the meter measures resistance.The parameter on sub-display is Θ.If Q 0.2, the meter measures inductance.The parameter on sub-display is Q.If Q -0.2, the meter measures capacitance.The parameter on sub-display is D.If C 5pF. The parameter on sub-display is Rp.- 19 -

Impedance can also be selected manually by pressing theLCR AUTO key.This is what the process would looklike:The meter starts out in AutoLCR mode.Press LCR AUTO key.The meter enters Auto L mode.The secondary parameter is Q.Press LCR AUTO key.The meter enters Auto CLmode.The secondary parameter is D.Press LCR AUTO key.- 20 -

The meter enters Auto R mode.The secondary parameter isblank. (The meter only displaysΘ automatically in Auto LCRmode.)Press LCR AUTO key.The meter enters DCR mode.The secondary parameter isblank.Press LCR AUTO key.The meter returns to AutoLCR mode.- 21 -

2.5.2. Measuring dissipation, quality, ESR, andphase angleThe DE-5000 can measure dissipation factor (D), qualityfactor (Q), equivalent series resistance (ESR), and phaseangle (Θ). These readings are shown in the secondarydisplay. To select the appropriate measurement, cyclethrough the options using the D/Q/ESR/Θ key.Note: that this function is available in Auto L and AutoC modes only.In other modes the D/Q/ESR/Θ key isdisabled.Example: In Auto C modeThe secondary displayshows D.Press D/Q/ESR/Θkey.The secondary displayshows Q.Press D/Q/ESR/Θkey.- 22 -

The secondary displayshows ESR.Press D/Q/ESR/Θkey.The secondary displayshows Θ.Press D/Q/ESR/Θkey.The secondary displayreturns to D.- 23 -

2.5.3. Sorting componentsThe DE-5000 can sort components into PASS/FAILcategories.Components may be sorted based onresistance, capacitance, or inductance.Note: this function is not available in Auto LCRmode.Before using this function, set the meterto either Auto L mode, Auto C mode, or Auto Rmode.To enter the Sorting mode, proceed as follows:1. Zero the meter.Refer to pages 14-16 for details.2. Press LCR AUTO to cycle through measurementoptions until the desired impedance type isselected.Note: If you select Auto LCR, the sorting functionwill not work.3. Connect one of the items that is going to bemeasured to the test points.4. Press SORTING, then press SETUP.Note: If you press the SORTING key while themeter is reading outside limits (OL) or while thereading is less than 200 counts, the Sorting functionis disabled.- 24 -

5. To adjust the nominal value, proceed as follows:a. Use / keys to adjust the position ofthe decimal point as necessary.PressENTER when finished.b. Use / / / keys to adjust thedigits as necessary.Press ENTERwhen finished.6. To adjust tolerance, use / keys to cyclethrough tolerance options until the desiredtolerance is reached.Press ENTER whenfinished.Available tolerances are: 0.25% 0.5%, 1%, 2%, 5%, 10%, 20%, and -20% 80%.The components may now be sorted.The primarydisplay will show either PASS or FAIL as eachcomponent is measured.The secondary display willshow the value of the measured component, as shown inthe sample below.- 25 -

To make adjustments to the nominal value or tolerance,press SETUP and repeat steps 5 and 6 above.To exit the Sorting mode, press SORTING.2.5.4. Making relative measurementsThe DE-5000 meter can make relative measurements.This option is controlled by the REL% key.When theRelative mode is active, the display shows the symbol.Note: this option is not available in the AutoLCR mode. Nor can it be activated when theparameter being tested is outside of meter limits.The meter uses the following formula to calculaterelative measurements:REL% (DCUR – DREF) / DREF * 100%REL% Percent differenceDCUR Device currently under testDREF Device used as a standard- 26 -

To access the Relative mode proceed as follows:1. Zero the meter. See pages 14-16 for details.2. Select the parameter to be tested.L, C, R, or DCR3. Attach the chosen standard to the test points.4. Press the REL% key.The meter should now display the symbol.5. Remove the standard and attach a DUT to the testpoints.The primary display should show the value of theDUT and the secondary display should show the %difference from the standard.6. Repeat Step 5 for each DUT.Note: The range for the % difference is -99.9% to99.9%.If the DUT falls outside that range, thesecondary display will show OL.7. To exit, press and hold the REL% key for2 seconds.- 27 -

Example: In Capacitance modeDisplay shows themeasurement reading.Ex. 669.3 nFPress REL% key appears on the screenThe reading on the display isstored as reference value.0.0% is shown on thesecondary display since themeasured value and thereference are the same at thispoint.Remove the standard andattach a different deviceThe new reading is shown onthe primary display.The % difference is shownon the secondary display.Remove the current DUTdevice under test and insertanother.- 28 -

The new reading is shown onthe primary display.The % difference is shownon the secondary display.Press REL% key again flashes on the s

The DE-500 is a portable, high-performance LCR meter that is full-featured yet cost effective. It measures in true 4-wire Kelvin mode and rivals the capabilities and options of many of its bench counterparts. It measures: Ls/Lp-- Series and parallel inductance Cs/Cp-- Series and parallel capacitance Rs/Rp-- Series and parallel resistance (ac)File Size: 1MB

Related Documents:

element is the Studio 5000 Logix Designer application. The Logix Designer application is the rebranding of RSLogix 5000 software and will continue to be the product to program Logix 5000 controllers for discrete, process, batch, motion, safety, and drive-based solutions. The Studio 5000 environment is the foundation for the future ofFile Size: 1MBPage Count: 71Explore furtherEmulate a Studio 5000 Controller and Connect to a .www.dmcinfo.comStudio 5000 Logix Emulate (Virtual PLC) - PLC Programmin www.plctr.comDownload Center Rockwell Automation United Stateswww.rockwellautomation.comStudio 5000 Logix Emulate FactoryTalk United Stateswww.rockwellautomation.comStudio 5000 Design Software FactoryTalk United Stateswww.rockwellautomation.comRecommended to you based on what's popular Feedback

engine rpm chart johnson rpm . 100 hp 1989 thru 4750-5250 2.07 115 hp 1989-2005 4750-5250 2.07 125 hp 1994-2005 4750-5250 2.07 135 hp 1990-1996 5000-5600 2.00 150 hp efi 2004 5000-6000 1.87 150 hp efi 2005 5000-5600 1.87 175 hp 1990-1999 5000-5600 1.87 175 hp efi 2000-2002 5000-5600 1.87 200 hp 1991-2005 5000-5600 1.87 200 hp efi 1994-2002 5000-5800 1.87 225 hp 1996-1998 5000-5500 1.64 .

Ricoh Aficio MP 4000/5000 series Refer to Table 1 about product names for "Ricoh imagio MP 4000/5000 series" and "Ricoh Aficio MP 4000/5000 series". TOE Version : "Ricoh imagio MP 4000/5000 series" and "Ricoh Aficio MP 4000/5000 series" are identified by following software and hardware. Software System/Copy 1.09 Network Support 7.23

Nov 29, 2010 · Roundup Original Max 5000 5000 Moderate Moderate Roundup Pro Concentrate 5000 5000 Moderate Slight Roundup ProMax 5000 5000 Moderate Slight Roundup UltraDry 3700 5000 Moderate Slight Roundup

TSI 147kW GTI DSG Engine Cylinders 4 3 3 3 3 3 4 Capacity (cm3) 1 598 999 999 999 999 999 1 998 Power output (kW @ r/min) 66 @ 5000 70 @ 5000 70 @ 5000 70 @ 5000 85 @ 5000 85 @ 5000 147 @ 4400 Torque output (Nm @ r/min) N/A 175 @ 2000 175 @ 2000 175 @ 2000 200 @ 200

MasterSeal HLM 5000 is a one-component, moisture-curing, bitumen-modified polyurethane elastomeric waterproofing membrane for exterior below-grade or between-slab applications. It is available in four grades: MasterSeal HLM 5000 SL (self-leveling/squeegee) MasterSeal HLM 5000 S (spray) MasterSeal HLM 5000 T (trowel) MasterSeal HLM 5000 R (roller)

MasterSeal HLM 5000 SL, HLM 5000 S y HLM 5000 R: cubetas de 18.95 L (5 gal) y tambores de 208 L (55 gal) disponible por pedido MasterSeal HLM 5000 T: cubetas de 18.95 L (5 gal) MasterSeal 995 (Sonoshield Reinforcing Fabric): rollos de 91 por 0.9 m (300 ft

M14 40.2 103.7 148.4 173.4 M16 62 161 230 269 M18 86 222 318 372 M20 122 314 449 525 M22 166 428 613 716 M24 211 543 777 908 M27 309 796 1139 1331 . 6 BILL OF MATERIALS PER KIT 5000-040A – 6R30 JD StalkMaster 606C/706C – 2) 5000-160 Roller Bearing Bolt Bag, 4) 5000-167 Torsion Pivot Assembly, 4) 5000- 168 Torsion Mount Bolt Bag, 2) 5000-272 .