Astra Tech Implant System SmartFix conceptManual and product catalogAstra Tech Implant System EV

The Astra Tech Implant System EV is designed for easeof use and versatility in providing treatment solutionsfor your implant patients.The foundation of this evolutionary system remainsthe unique Astra Tech Implant System BioManagementComplex, which has been proven to predictablyprovide long-term marginal bone maintenance andesthetic results.Animation showing:SmartFix concept for Astra Tech Implant System EVStep-by-step procedure

Astra Tech Implant System CONTENTSIntroduction – SmartFix conceptIntroduction4Treatment planningPre-operative considerationsImplant assortment for SmartFix conceptAbutment assortment for SmartFix concept566Step-by-step ��–Implant placementAbutment connectionImmediate temporizationProsthetic and laboratory procedures781012OsseoSpeed Profile EV15Simplant computer-guided implant treatment16Atlantis patient-specific suprastructures17AppendixBone Reamer EVSmartFix GuideStep-by-step procedure – Verification jig181920Product catalog22Torque guideExplanation of the symbols on labels and instructions for useIndex282829This manual is designed for use by clinicians who have undergone appropriateeducation and training in surgical and prosthetic implant treatment. Staying currenton the latest trends and treatment techniques in implant dentistry through continuededucation is the responsibility of the clinician.This manual only addresses the additional information needed to work with theSmartFix concept. For all other instructions and/or a full description of implantplacement and restorative procedures for the Astra Tech Implant System EV as wellas all the instruments and components needed, please refer to the Surgical ManualAstra Tech Implant System EV, OsseoSpeed Profile EV, Screw-retained restorationmanual, Guided surgery manual and the Astra Tech Implant System EV Product Catalog.AstraTechClinical&AstraTech EVImplantSystem Product catalAstraogTech ImplantSystem EVAll products may not be regulatory cleared/released/licensed in all markets.Please contact the local Dentsply Sirona sales office for current product assortmentand availability.To improve readability, Dentsply Sirona does not use or in body copy.However, Dentsply Sirona does not waive any right to the trademark andnothing herein shall be interpreted to the contrary.ImplantSystem laboratoryScrew-retaine manuald restorationsSystemAstraTech ImplantAstraTechImplantSystem SurgicalmaAstra Tec nualh Implant System EVAstraTechImplantSystem Manualand proOsseoSpeed duct catalogProfileEVProduct illustrations are not to scale.3

IntroductionIntroduction – SmartFix conceptWith the SmartFix treatment concept patients canbenefit from an immediate implant-supportedrestoration, as a provisional prosthesis is screwed ontothe implants on the day of surgery. Final solutions forthe SmartFix treatment concept include both fixedprostheses and removable solutions e.g. Atlantispatient-specific suprastructures.By placing two posterior implants at an angle, longerimplants can be used adjacent to areas of otherwiseanatomically compromised bone volume such asclose to the mental foramen or the maxillary antrum,increasing bone-to-implant contact and reducing theneed for vertical bone augmentation. The tilted posteriorimplants reduce cantilevers and improve prostheticsupport by increasing the anterior/posterior spread.Biomechanical measurements show that tilted implants,when part of prosthetic support, do not have a negativeeffect on the load distribution. The tilting of implants hasbeen used in clinical practice for over a decade and hasshown good results.The ease of using the SmartFix concept4 Reduces treatment time for the patient Involves all team members in the treatment plan toensure a quality outcome Uses a prosthetically driven protocol to improve theesthetic result Allows the patient to leave the clinic, with implantretained teeth, on the day of surgery Provides one uniform interface to reduce restorativecomplexity Provides a simplified restorative concept to reducechair time Increases patient acceptance by minimizing costs

Treatment planningPre-operative considerationsThe SmartFix concept is an implant-prosthetic procedurefor the immediate restoration of edentulous patients withscrew-retained bridges or bar dentures on a minimumof four implants. The posterior implants placed at anangle are restored with 17 or 30 angled Multibase EVabutments to obtain a common path of insertion. Bone quality and quantity, primary implant stability,design of restoration and loading conditions shouldalways be carefully examined and assessed by theclinician when deciding the appropriate time to loadthe implants in the individual case.Both parts of the Multibase EV abutment need tobe tightened to 25 Ncm to secure a stable screwjoint and pre-load. Thus implants need to be stableenough to withstand this tightening torque if youconsider immediate temporization. If in doubt, healingabutments or even a two stage surgical approach canbe an alternative. Possible extraction sites should be debridedthoroughly. It is advisable to place implants betweenextraction sockets. It is recommended that a new denture, to be convertedinto a temporary fixed restoration on the day ofsurgery, is fabricated in advance. If possible, the posterior implants should be placedusing maximum diameter and length within thelimitations of available bone. Extensions should be limited to one tooth bilaterallyfor an immediate acrylic bridge with a maximum of12 teeth. Fabrication of an immediate acrylic bridge can bemade from a well-adapted existing denture in goodcondition. For best esthetics and function, the final bridge shouldhave 12 teeth and a supporting metal framework.For tilted posterior implants, plan for the emergence ofthe screw access hole to be located within the occlusalsurfaces of the posterior teeth.Low cone and gingival height– more free interocclusal space.Cone taper 21 – allows for flexible bridgeinsertion by non-parallel abutments.One uniform prosthetic interface forabutments, lowering the inventory needs.The narrow abutment design reducesthe need for bone reaming, therebysimplifying the abutment connection.The angled abutment is designed withtwo pieces, which provides for full wallstrength and full thread support forbridge screw.Available as straight, 17 and 30 , indexed andindex-free, in different gingival heights to handlechallenging clinical situations.Multibase EV abutments aredelivered with a pre-mounted,flexible plastic holder for easyhandling during installation.Fitting 3.6, 4.2 and 4.8 OsseoSpeed EV andOsseoSpeed Profile EV implant diameters.5

Treatment planningImplant assortment for SmartFix conceptOsseoSpeed EV implants are available in a versatilerange of shapes, diameters and lengths for all indications,including situations with limited space and/or bonequantity.OsseoSpeed EVSpecific colors have been assigned to the differentimplant-abutment connection sizes, which areconsistently used throughout the system and identifiedby symbols and markings.Note: OsseoSpeed Profile EV implants and componentsare also marked with a “P”.The following implants can be used when using theSmartFix concept with the Astra Tech Implant System EV:OsseoSpeed EVØ 3.6Ø 4.2Ø 4.8PurpleYellowBlueOsseoSpeed Profile EV OsseoSpeed EV straight implantsDiameters: 3.6 S, 4.2 S and 4.8 SLengths: 6 mm–17 mm OsseoSpeed EV conical implantsDiameters: 4.2 C and 4.8 CLengths: 8 mm–17 mmOsseoSpeed Profile EV OsseoSpeed Profile EV straight implantsDiameters: 4.2 PS and 4.8 PSLengths: 8 mm–17 mmOsseoSpeed Profile EV conical implantsDiameters: 4.2 PC and 4.8 PCLengths: 8 mm–17 mmØ 4.2Ø 4.8YellowBlueAbutment assortment for SmartFix conceptMultibase Abutment EV for OsseoSpeed EVVersion:straight, 17 and 30 Diameter:Height:1.5, 2.5, 3.5Indexing options:Indexed abutments will seat in six availablepositionsIndex-free abutments will be seated any rotational positioninMultibase Abutment EV for OsseoSpeed Profile EVVersion:straight, 17 and 30 Diameter:Height:1.5, 2.5, 3.5Indexing option:Index-free abutments will be seated any rotational positioninNote: Some restorative components are marked with a groove or laser marking todifferentiate from other similar components of the Astra Tech Implant System EV.6

Step-by-step procedure – Implant placementStep-by-step procedure – Implant placementBelow is a step-by-step procedure for implant placementin the maxilla using OsseoSpeed EV 4.2 S implants.Implant placement1a1b1c2a2b2c1SmartFix Guide (Optional) After raising a flap, prepare an osteotomyfor the SmartFix Guide Drill in the midline using 1-Twist Drill EV 1.9to a depth of 1 1 mm (1a). The guide can be pre-shaped outside theoral cavity. Place the guide in the osteotomy – thelines on the guide help aligning the axis ofthe implant (2a).34ABImplant placement Choose a starting point and an anglewhich results in an osteotomy that doesnot interfere with the maxillary sinus (1b). Prepare the implant site. Drill toappropriate depth and check for correctangulation, by using the installation guide. Follow the drilling protocol forOsseoSpeed EV 4.2 S implant. Install the implant so that the top of theimplant is at bone level on the mesialside (2b). This could result in a subcrestalposition at the distal.Note: Regarding minimum torquerequirement for immediate temporization,see section Pre-operative considerations.Note: See surgical manual for detailedsurgical drilling protocol and options.All drilling should be performed ata maximum speed of 1500 rpm withprofuse irrigation.Bone Reamer EV (Optional) A reamer can be used when boneinterferes with a correct abutment seating. Manually connect the appropriate BoneReamer Guide EV to the implant using ahex driver (1c). Use the appropriate bone reamertogether with a driver handle. Place the bone reamer over the guideand remove bone by rotating the reamerunder irrigation (2c). You can also usethe reamer in a contra angle at low speedunder irrigation (max. 100 rpm). Remove the guide using a hex driver.Note: The depthmarkings on thebone reamer aremeasured from theimplant and up tothe indication line.5 mm4 mm3 mm2 mmThe guide isprovided with adepth stop.7

Step-by-step procedure – Abutment connectionStep-by-step procedure– Abutment connectionAbutmentheadHolderBelow is a step-by-step procedure for abutmentconnection in maxilla using Multibase EV abutments.Multibase Abutment EV30 and 17 AbutmentbodyDelivered assembledin a sterile blister packAbutmentscrewAbutment connection – Multibase Abutment 30 1a2a1b3aAbutment body 8Select the appropriate abutment angleand height. Connect the abutment body to theimplant and rotate the abutment tothe desired position (1a). The flexible holder can be bent tofacilitate placement. Perform initial tightening of the abutmentscrew with a manual hex driver (2a). Unscrew the holder from theabutment body (3a).2b1c3bAbutment body/head Use the restorative driver handletogether with the hex driver and theTorque Wrench EV to tighten theabutment screw to the recommendedtorque (25 Ncm) (1b). Flip the holder over 180 degrees to theside that holds the abutment head (2b). Screw the abutment head into theabutment body with the holder (3b).2c3cAbutment head Snap off the handle (1c). Perform initial tightening with the manualMultibase Driver EV to initially tighten theabutment head (2c). Use the restorative driver handle togetherwith the Multibase Driver EV and thetorque wrench to tighten the abutmenthead to the recommended torque(25 Ncm) (3c).

Step-by-step procedure– Abutment connectionMultibase Abutment EVHolderDelivered assembled ina sterile blister packAbutmentAbutment connection – Multibase Abutment straight1a2a1b3aAbutment connection Pick up the selected abutment in thepre-mounted plastic holder (1a). Manually seat and secure the abutmentusing the holder (2a). Snap off the holder (3a).2bAbutment connection Perform initial tightening with therestorative driver handle together withthe multibase driver (1b). Use the restorative driver handle togetherwith the multibase driver and torquewrench and tighten to the recommendedtorque (25 Ncm) (2b).9

Step-by-step procedure – Immediate temporizationStep-by-step procedure– Immediate temporizationThe following procedure is a technique where a dentureis used as the base for a temporary restoration.Polymerization SleeveUse sleeves to protect thewound from resin.1a1b2b2aMultibase EV Temporary Cylinder10 After all abutments are seated, close theflap and suture. Manually seat and secure the temporarycylinders to the abutments with theMultibase EV Bridge Screws using thehex driver (1a). Apply the polymerization sleeves toprotect the wound from resin (2a).A rubber dam can also be used.Adjust the denture Perforate the denture to allow it to seatonto the mucosa without interfering withthe cylinders.Attach temporary cylinders Use autopolymerizing resin to attach thetemporary cylinders to the denture (1b). After resin has set, unscrew the bridgescrews and remove the denture (2b).

Step-by-step procedure – Immediate temporization1a2aConnect Multibase EV Heal Cap Manually seat and secure the heal capsto the abutments using light finger force(5–10 Ncm) with the hex driver.Modify dentureTemporary bridge installation Cut off the excessive parts of thetemporary cylinders (1a). Remove the heal caps fromthe abutments. Grind away the palatal plate of thedenture and reduce the buccal baseplate (2a). Connect the temporary bridge with theMultibase EV Bridge Screws and checkthe fit. Fill in any voids with resin and adjust thesoft tissue side of the denture to allowaccess for good oral hygiene. Use the restorative driver handle togetherwith the hex driver and the torque wrenchto tighten to the recommended torque(15 Ncm). Check function and contacts for balancedocclusion and articulation.11

Step-by-step procedure – Prosthetic and laboratory proceduresStep-by-step procedure– Prosthetic and laboratory proceduresBelow is a step-by-step procedure using an open tray procedure.Note: Closed tray option also available.Multibase EV Pick-upfor open tray optionMultibase EV Transferfor closed tray optionClinical procedure – open tray1a2aMultibase EV Pick-upImpression Remove the temporary bridge. Cover an open tray with wax (1a). Connect the pick-ups using theHex Driver EV. Apply an elastomeric impression materialaround the pick-ups separately (2a).Place the tray, filled with the impressionmaterial, and take the impression. Once the impression material hasset, unscrew the pins and removethe impression. Check the impression for correct andstable retention of the pick-ups. Reinstall the temporary bridge. 12Applying impression materialSecure the pick-ups using manualtightening torque (5–10 Ncm).

Step-by-step procedure – Prosthetic and laboratory proceduresBelow a laboratory procedure for fabrication ofAtlantis suprastructures is described. You can also useconventional wax-up technique using the Multibase EVBurnout Cylinder and casting process.Laboratory procedure – open tray1a2aMultibase EV Pick-up/Multibase EV Replica Connect the replicas carefully to thepick-ups and tighten. Secure the pick-ups using manualtightening torque (5–10 Ncm).Master model / Tooth set-upOrder – Atlantis suprastructures Prepare the impression for duplicationwith a removable soft tissue mask byapplying silicone around the replica sites. Enter the order via the Atlantis WebOrder.Consult the Atlantis suprastructures –User guide for the ordering process. Pour high quality stone and fabricate themaster model (1a). Prepare a tooth set-up in wax (2a).After review and final approval ofthe design in Atlantis Viewer, thesuprastructure is fabricated. Consult the separate Atlantissuprastructure Design Guide for detailedhandling procedures in the laboratory.Note: Multibase EV Replica is for single use.13

Step-by-step procedure – Prosthetic and laboratory proceduresMultibase EVBridge ScrewLaboratory procedureFinal restoration –Atlantis suprastructures The Atlantis suprastructure is shippedto the dental laboratory.Note: The Multibase EV Lab BridgeScrew is recommended to be used duringthe laboratory procedure with Atlantissuprastructures for the Astra Tech ImplantSystem EV.Multibase EVLab Bridge ScrewClinical procedureFinal restoration –Atlantis suprastructures Fabricate the final prosthesis.Installation of the final restoration Remove the temporary bridge. Connect the final restoration with theMultibase EV Bridge Screws and checkthe fit. Use the restorative driver handle togetherwith the hex driver and the torque wrenchto tighten to the recommended torque(15 Ncm). Cover the screw head before the screwchannel is filled with a suitable materiale.g. composite resin. Check function and contacts for balancedocclusion and articulation.Note: The lab bridge screw should bereplaced with a clinical bridge screw forplacement of the final restoration in theclinical situation.14

OsseoSpeed Profile EVOsseoSpeed Profile EVPlacement of the uniquely designed OsseoSpeedProfile EV at an angle can help to eliminate/reducethe need for bone removal at installation.When placing the uniquely designed OsseoSpeedProfile EV implant distally at an angle, the implant canoften be aligned flush with the marginal bone, therebyavoiding the occurrence of a partially-submergedimplant margin. As a consequence the need for bonereamers is reduced.OsseoSpeed Profile EV 4.2 Pimplant placed in a 30 angulationOsseoSpeed EV 4.2 implantplaced in a 30 angulationImplant placement OsseoSpeed Profile EV implants;–C onsider carefully where you prefer to position themost apical point of the slope of the implant. Thisis achieved by making sure that the flat side of theImplant Driver Profile EV is facing in the directionin which you would want the most apical point ofthe slope to be positioned. This will often be in amesial position so that the slope of the implant isin alignment with the ridge of the bone.Note: Only index-free Multibase EV abutments fitOsseoSpeed Profile EV; directing the abutment istherefore not related to the implant insertion. OsseoSpeed EV implants;– Consider carefully how you prefer to direct theangulated Multibase EV abutment. For an indexedabutment the preferred direction is achieved bypositioning one of the six flat sides of the implantdriver in line with the desired direction.For a non-indexed angulated Multibase EV abutmentthe positioning of the implant driver is withoutconsequence.15

Simplant computer guided implant treatmentSimplant computer guided implant treatmentSimplant software and surgical guides can be used forthe SmartFix concept to ensure accurate planning foroptimized implant position and placement.Simplant facilitates crown-down planning by visualizingthe surgical and prosthetic aspects of the case.By virtually planning your components, a custom-madeSimplant SAFE Guide can be fabricated to assist in aguided surgery procedure.For the SmartFix concept, normally plan for a distallytilted placement of the posterior implant on each side.In most cases up to 30 inclination is appropriate.The available implant diameters for Multibase EVabutments and guided surgery are 3.6 / 4.2 / 4.8,and the maximum implant length for use with aSimplant SAFE Guide is 15 mm.Multibase EV abutments (straight, 17 and 30 )can be selected in the Simplant Library.16

Atlantis patient-specific suprastructuresAtlantis patient-specific suprastructuresAtlantis suprastructures are produced using the latestdevelopments in world-class production technologiesand are supported by computer-based industrial andmedical-device expertise.Screw-retained solutionsAtlantis Bridge and Atlantis Hybrid are full

The tilted posterior implants reduce cantilevers and improve prosthetic support by increasing the anterior/posterior spread. Biomechanical measurements show that tilted implants, when part of prosthetic support, do not have a negative effect on the load d

Related Documents:

Astra Control itself is not a standard app; it is a "system app." You should not try to manage Astra Control itself. Astra Control itself isn’t shown by default for management. To see system apps, use the “Show system apps” filter. For instructions on how to manage apps using the Astra API, see the Astra Automation and API information.

SENARAI Majalah Astra Edisi 04, April 2015 Tahun XLIII ASTRA 2015 DAFTAR ISI 03 03 14 20 Melalui RUPST Astra, pemegang saham merestui pembagian dividen tunai senilai Rp 216 per saham. Sejumlah tokoh penting seperti Mari Elka Pangestu dan Muhamad Chatib Basri memperkuat jajaran Komisaris Astra yang baru. RAGAM BERITA RAGAM BERITA Jalin Silaturahmi, Astra Kunjungi Tempo 14 UTAMA RUPST Astra .

Astra Tech Implant System EV and SmartFix concept A facially driven treatment plan was carried out for a 56 year-old female patient with gummy smile and terminal dentition. After tooth extraction, in both arches, the patient was rehabilitated wi

Astra Control Center Astra Control is now available in a new deployment model. Astra Control Center is self-managed software that you install and operate in your data center so that you can manage Kubernetes application lifecycle management for on-premise Kubernetes clusters. Go to the Astra Control Center documentation to learn more.

To launch Astra Schedule, double-click on the Astra Schedule shortcut icon on your desktop, or choose it from the Programs folder in your Start menu. The Astra Schedule main window will appear. Note: You may get a warning about Events when you open Astra for the first time and Astra

a longer implant. Immediate loading of single tooth restorations is not recommended. Implant overview The OsseoSpeed TX implants have been developed and extensively documented for both one- and two-stage surgical procedures. The Conical Seal Design of the Astra Tech Implant System allows for a strong

The Astra Tech Insight is published and distributed world-wide by Astra Tech AB, P.O. Box 14, S-431 21 Mölndal, Sweden. Astra Tech is a company in the AstraZeneca Group. Editor-in-chief Mia Jensen, DDS Clinical Information Manager, Astra Tech AB Associate editor Anders Holmén, DDS Me

Surgical manual ASTRA TECH Implant System EV. The ASTRA TECH Implant System EV is designed for ease of use and versatility in providing treatment solutions for your implant patients. The foundation of this evolutionary system