FreeStyle Optium Neo H - Glucose Monitoring And Data .

3y ago
247 Views
22 Downloads
1.86 MB
49 Pages
Last View : 1d ago
Last Download : 1d ago
Upload by : Melina Bettis
Transcription

FreeStyle Optium Neo HBlood Glucose and Ketone Monitoring SystemProduct TrainingHealthcare professional portrayal.

The FreeStyle NeoH can perform both blood glucose and blood Ketonetests. Ensure to familiarise yourself with both modules.Click on the module below.GlucoseKetonesPatient testingPatient testingQuality ControlQuality ControlMaintenanceMaintenance

An Introduc on toFreeStyle Op um NeoH

The FreeStyle OpAum Neo H System at a GlanceLogbookHigh contrast displayDisplays meter home screen, test resultsand other important informationUSB portInsert a data cable foruploading test results to acomputerDown buttonUp button Used to mark a control solution test Use to mark a venous blood test Adjust settings Adjust settingsPower buttonTest strip portPress to:Where the test strip is inserted Turns meter on and off Returns to the home screen Saves settings

The FreeStyle OpAum Neo H Blood Glucose Test StripThe FreeStyle OpAum Neo H meter is designed to be used only withFreeStyle OpAum Neo H Blood Glucose Test Strips. 0.6μL sample size 5-second result countdown Assay range 1.1-27.8 mmol/L Haematocrit range 15-65% Top or Ap sample applicaAon Visual confirmaAon of sufficient sample Fill trigger ensures sufficient sample Sample Types – Capillary, venous, arterial & neonatal Each pack contains 100 individually foil wrapped test strips, calibraAonbar and test strip insert

Meter calibraAon The calibraAon procedure programmes the meter with the lotnumber, expiry date and test strip technologyAlways calibrate the meter with each new pack of test strips –failure to calibrate properly will cause incorrect resultsWith the lot number facing toward you, insert the contact bars ofthe calibrator into the meter. The meter turns on automaAcallyLOT and the lot number appear in the display windowCheck that the lot number on the meter display window matcheswith the number on the test strip calibrator and the last fivedigits on both the test strip foil packet and test strip insertRetain the calibraAon bar for the duraAon of that packCalibraAon is completeOnly use the calibrator that is packaged in the pack of test stripsyou are using

Performing a Pa ent TestThe FreeStyle Optium Neo H systemenables easy and accurate bedsidetesting of blood glucose levelsBlood glucose testing}5 second test time}Small 0.6 μL sample size}Calibrator provides expiry check andcalibrates the meter}Control solution mark

PaAent TesAng1.Prepare the sample site:i.Before you obtain a blood drop, make sure the sample site isclean, dry and warmii.Hang the arm down before lancing the sample site, to help aidblood flow2.Open the individually foil wrapped test strip at the notch andtear downInsert the test strip into the meter unAl it stops. This will turnthe meter on.Check that the LOT number of test strip you are using matchesthe LOT number shown on the meter displayTheblinks, indicaAng the meter is ready for you to apply asample to the test strip3.4.5.i.Ensure you follow organisaAonal policy for sample collecAon

PaAent tesAng6.If tesAng glucose on venous blood press thethe test unAlto markappears indicaAng that the meter is nowready for you to apply venous blood to the test strip7.Apply a sample with the meter in a horizontal (flat)posiAon to avoid it entering the test strip port8.If you are applying the sample from a syringe, apply awayfrom the direcAon of the test strip port

PaAent tesAng9.Hold sample to the test strip unAl 3 short lines appear on the screen10. The results will be displayed afer a 5-second countdown

PaAent tesAngWhen the result appears on the meter screen the test is complete.The results will be stored in the meter’s memory.Press and hold the buhon to turn off the meter.

Summary for blood glucose tesAng1. Inserts test strip,blinks2. Apply blood3. Sample detected, 5 secondscountdown begins4. Result is displayed

Out of range indicatorsHigh blood glucose results A result falling outside the range set by your organisaAon mayrequire some intervenAon from the operator, such asrepeaAng the test– The meter will display a solid yellow arrowFor guidance on out of range results, please follow yourorganisaAonal policyLow blood glucose results A result falling outside the range set by your organisaAon willrequire some intervenAon from the operator, such asrepeaAng the test– The meter will display a solid red arrowFor guidance on out of range results, please follow yourorganisaAonal policy

Display MessagesResult is less than1.1 mmol/LGlucose result isabove 27.8 mmol/LResult is higher than orequal to 13.3 mmol/L

Results & MemoryThe test is complete when the result appears on the meter screen. The result is stored in the meter’s logbook The meter logbook can store up to 1,000 events – including bloodglucose, blood ketone, and control soluAon results, and other meterinformaAon The logbook can be accessed by going to the Home Screen andselecAng To recall paAent results:With the FreeStyle OpAum Neo H turned on:1. Press theto view logbook events2. Pressorto scroll through log book eventsWhile on the Homescreen, pressto view logbookevents

Logbook: View Logbook EventsPressto scroll through to view logbook eventsBlood Glucose Result Venous Blood ResultBlood Ketone ResultControl solution test results will alternate between ‘CtL’ and resultGlucose ControlSolution ResultKetone ControlSolution Result

Logbook: View Blood Glucose Averages1. While in the logbook, press anytime to view the 7-day average2. Pressorto scroll through 7-,14- and 30-day averages. Pressto return to logbook events.AverageNumber of testsIn this average

Error MessagesMessageWhat it meansE1Temperature is too hot or too cold for the meter to work correctlyE2/E9Meter errorE3i.ii.iii.Blood drop is too smallIncorrect test procedureThere may be a problem with the test stripE4i.ii.The blood glucose level may be too high to be read by the systemThere may be a problem with the test stripE5i.Blood was applied to the test strip too soonE6i.ii.Calibration errorTest strip errorE7i.Test strip may be damaged, used or the meter does not recogniseit

Performing a Quality Control Test

Performing a Quality Control Test If QC tesAng is NOT performed, misleading results may lead to inappropriatetreatment Always follow your organisaAonal policy to avoid pumng paAents at risk The FreeStyle OpAum Neo H meter may prompt when this is due Only use MediSense Glucose and Ketone Control SoluAons and keep the meterflat to prevent excess soluAon running into the meter QC SoluAon is stable for 90 days following opening or unAl the expiry dateprinted on the soluAon bohles, whichever is sooner. Discard and do NOT usecontrol soluAon past its expiry date Check results against the test strip insert inside the pack of test strips being usedas QC ranges will vary between test strip LOTs It is the responsibility of all users to ensure the meter has been tested beforepaAent use

Performing Quality Control (QC) Tests1.Check that the LOT number printed on the test strip and instrucAons for usematch2.Open the individual foil test strip at the notch and tear down3.Insert the test strip into the meter which will turn the meter on turn on4.5.When theblinks press thebuhon unAl this icon appearsThe meter is now ready for you to apply the low level control soluAon to the teststripi.Invert the bohle to mix the soluAonii.Wipe the QC nozzle with clean gauze before and afer each testiii. Replace the cap securely on the bohle immediately afer use7. Apply a drop of the control soluAon with the meter ina horizontal posiAon to avoid it entering the test strip port

Performing Quality Control (QC) Tests8.Apply a drop of the control soluAon to the test strip unAl 3 short lines appear onthe screen9.The results will be displayed afer a 5-second countdowni.i.ii.Compare the control soluAon results to therange printed on the blood glucose test strip insert;the results should fall within this rangeThe results will be stored in memory as a control soluAon resultRecord the result in the QC record book10. If results are out-of-rangei.ii.Repeat the testIf the repeated result is out-of-range follow organisaAonal policy for out-of-range results11. Remove the test strip, repeat the procedure for high level control soluAon12. Record HI and LO QC test result in the QC record book

Quality Control TesAng Summary1. Insert test strip,blinks.2. Press and hold theappears.unAl3. The meter is now ready for theoperator to apply controlsoluAon to the test strip.4. Apply control soluAon.5. Sample detected, countdownbegins.5. Result is displayed.

Maintenance

Maintenance of the meter1.Gently clean the exterior of the FreeStyle OpAum Neo H meter with a damp clothi.ii.iii.Acceptable cleaning soluAons include mild detergent and water, 70% isopropyl alcoholor a mixture of 1 part household bleach and 9 parts waterDo not immerse the meter in water or other liquids. Avoid gemng dust, blood, controlsoluAon and any other substances in the test strip port, USB port and baherycompartmentAllow the meter to dry2.Do not try to clean the test strip port.3.Do not pour liquid into the test strip port or onto the buhons.4.Do not immerse the meter in water or any other liquid.

Maintenance of the meter1.Replace coin cell baheries when ‘Low Bahery Power’ message is displayed1.The meter requires 2 replaceable CR2032 coin cell baheries.2.To change the baheries turn the meter over and slide the bahery door(located on the side of the device) open, as shown.2.3.Perform QC test as per organisaAonal policyRefer to troubleshooAng guide in FreeStyle OpAum Neo H operator’s manual forerror messages4.When not in use, always store the FreeStyle OpAum Neo H meter as perorganisaAonal policy

Blood β-Ketone Tes ng withFreeStyle Op um Neo HThe FreeStyle Optium Neo H systemenables easy and accurate bedside testingof blood ketone levelsBlood ketone testing}10 second test time}1.5 μL sample size}Purple calibrator calibrates the meter}Control solution mark

An Introduc on toFreeStyle Op um NeoH

The FreeStyle OpAum Neo H System at a GlanceUSB portInsert a data cable foruploading test results to acomputerHigh contrast displayDisplays meter home screen, test resultsand other important informationDown button Used to mark a control solution test Adjust settingsUp button Use to mark a venous blood test Adjust settingsPower buttonTest strip portPress to:Where the test strip is inserted Turns meter on and off Returns to the home screen Saves settings

FreeStyle Optium H Blood β-Ketone Test StripsThe FreeStyle OpAum H Blood β-Ketone Test Strip: 1.5 μL sample size 10-second result countdown Assay Range 0.0-8.0 mmol/L Haematocrit Range 30-60% Top or Ap sample applicaAon Visual confirmaAon of sufficient sample Fill trigger ensures sufficient sample Sample Type – capillary (finger) and venous whole blood samples Each pack contains 10 individually foil wrapped test strips,calibraAon bar and test strip insert

β-Ketone BenefitsClinical accuracy Measures β-hydroxybutyrate (the most significant ketone body) A quanAtaAve vs. qualitaAve measure Results unaffected by many common medicaAonsPa ent convenience/compliance Blood vs. urine sampling Foil wrapped test strips Test integrated within blood glucose meters

Meter calibraAon The calibraAon procedure programmes the meter with the lotnumber, expiry date and test strip technologyAlways calibrate the meter with each new pack of test strips –failure to calibrate properly will cause incorrect resultsWith the lot number facing toward you, insert the contact bars ofthe calibrator into the meter. The meter turns on automaAcallyLOT and the lot number appear in the display windowCheck that the lot number on the meter display window matcheswith the number on the test strip calibrator and the last fivedigits on both the test strip foil packet and test strip insertRetain the calibraAon bar for the duraAon of that packCalibraAon is completeOnly use the calibrator that is packaged in the pack of test stripsyou are using

Performing a Pa ent Test

PaAent TesAng1.Prepare the sample site:i.Before you obtain a blood drop, make sure the sample site isclean, dry and warmii.Hang the arm down before lancing the sample site, to help aidblood flow2.Open the individually foil wrapped test strip at the notch andtear downInsert the test strip into the meter unAl it stops. This will turnthe meter on.Check that the LOT number of test strip you are using matchesthe LOT number shown on the meter displayTheblinks, indicaAng the meter is ready for you to apply asample to the test strip3.4.5.i.Ensure you follow organisaAonal policy for sample collecAon

PaAent tesAng6.If tesAng ketones on venous blood press thethe test unAlto markappears indicaAng that the meter is nowready for you to apply venous blood to the test strip7.Apply a sample with the meter in a horizontal (flat)posiAon to avoid it entering the test strip port8.If you are applying the sample from a syringe, apply awayfrom the direcAon of the test strip port

PaAent tesAng9.Hold sample to the test strip unAl 3 short lines appear on the screen10. The results will be displayed afer a 10-second countdown

PaAent tesAngWhen the result appears on the meter screen the test is complete.The results will be stored in the meter’s memory.Press and hold the buhon to turn off the meter.

Display MessagesMessageExplanationResult above 8.0 mmol/LThis appears when a result is higher than 8.0 mmol/LThis suggests very high blood ketone levels or that there may bean error with the test stripRepeat the test with a new test stripThis requires immediate acAon as defined by your facility’spolicies and procedures

Blood Ketone Testing summary1. Insert ketone test strip,blinks and KET appears2. Apply blood3. Sample detected, 10 secondscountdown begins4. Result is displayed

Error MessagesMessageWhat it meansE1Temperature is too hot or too cold for the meter to work correctlyE2/E9Meter errorE3i.ii.iii.Blood drop is too smallIncorrect test procedureThere may be a problem with the test stripE4i.ii.The blood glucose level may be too high to be read by the systemThere may be a problem with the test stripE5i.Blood was applied to the test strip too soonE6i.ii.Calibration errorTest strip errorE7i.Test strip may be damaged, used or the meter does not recogniseit

Performing a Quality Control Test

Performing a Quality Control Test If QC tesAng is NOT performed, misleading results may lead to inappropriatetreatment Always follow your organisaAonal policy to avoid pumng paAents at risk The FreeStyle OpAum Neo H meter may prompt when this is due Only use MediSense Glucose and Ketone Control SoluAons and keep the meterflat to prevent excess soluAon running into the meter QC SoluAon is stable for 90 days following opening or unAl the expiry dateprinted on the soluAon bohles, whichever is sooner. Discard and do NOT usecontrol soluAon past its expiry date Check results against the test strip insert inside the pack of test strips being usedas QC ranges will vary between test strip LOTs It is the responsibility of all users to ensure the meter has been tested beforepaAent use

Performing Quality Control (QC) Tests1.Check that the LOT number printed on the test strip and instrucAons for use match2.Open the individual foil test strip at the notch and tear down3.Insert the test strip into the meter which will turn the meter on turn on4.When theblinks press thebuhon unAl this icon appears5.The meter is now ready for you to apply the low level control soluAon to the teststripi.Invert the bohle to mix the soluAonii.Wipe the QC nozzle with clean gauze before and afer each testiii. Replace the cap securely on the bohle immediately afer use7. Apply a drop of the control soluAon with the meter ina horizontal posiAon to avoid it entering the test strip port*KET will appear on the screen as a purple ketone test strip has been inserted

Performing Quality Control (QC) Tests8.Apply a drop of the control soluAon to the test strip unAl 3 short lines appear onthe screen9.The results will be displayed afer a 10-second countdowni.i.ii.Compare the control soluAon results to therange printed on the blood glucose test strip insert;the results should fall within this rangeThe results will be stored in memory as a control soluAon resultRecord the result in the QC record book10. If results are out-of-rangei.ii.Repeat the testIf the repeated result is out-of-range follow organisaAonal policy for out-of-range results11. Remove the test strip, repeat the procedure for high level control soluAon12. Record HI and LO QC test result in the QC record book

Quality Control TesAng Summary1. Insert test strip,blinks.2. Press and hold theappears.unAl3. The meter is now ready for theoperator to apply controlsoluAon to the test strip.4. Apply control soluAon.5. Sample detected, countdownbegins.5. Result is displayed.

Maintenance

Maintenance of the meter1.Gently clean the exterior of the FreeStyle OpAum Neo H meter with a damp clothi.ii.iii.Acceptable cleaning soluAons include mild detergent and water, 70% isopropyl alcoholor a mixture of 1 part household bleach and 9 parts waterDo not immerse the meter in water or other liquids. Avoid gemng dust, blood, controlsoluAon and any other substances in the test strip port, USB port and baherycompartmentAllow the meter to dry2.Do not try to clean the test strip port.3.Do not pour liquid into the test strip port or onto the buhons.4.Do not immerse the meter in water or any other liquid.

Maintenance of the meter1.Replace coin cell baheries when ‘Low Bahery Power’ message is displayed1.The meter requires 2 replaceable CR2032 coin cell baheries.2.To change the baheries turn the meter over and slide the bahery door(located on the side of the device) open, as shown.2.3.Perform QC test as per organisaAonal policyRefer to troubleshooAng guide in FreeStyle OpAum Neo H operator’s manual forerror messages4.When not in use, always store the FreeStyle OpAum Neo H meter as perorganisaAonal policy

CongratulaAons on compleAng this moduleFreeStyle blood glucose meters assist in the monitoring of blood glucose (and ketone) levels for people diagnosedwith diabetes mellitus. People diagnosed with diabetes with elevated blood ketone levels should seek medicaladvice. Information contained herein for distribution outside of the U.S. Only. FreeStyle and related brand marksare trademarks of Abbott Diabetes Care Inc. in various jurisdictions. All other trademarks are the property of theirrespective owners. Abbott Diabetes Care Australia, 666 Doncaster Road, Doncaster VIC 3180. For furtherinformation call Customer Service on 1800 801 478 or visit www.myfreestyle.com.au MSE160420044321

glucose, blood ketone, and control soluon results, and other meter informaon The logbook can be accessed by going to the Home Screen and selecng To recall paent results: With the FreeStyle Opum Neo H turned on: 1. Press the to view logbook events 2. Press or to scroll through log book events While on the Home

Related Documents:

Optium Neo Blood Glucose and Ketone Monitoring System. A multicentre clinical study was conducted to evaluate performance with fresh capillary whole blood. Additional laboratory studies were performed to verify performance claims under various testing conditions. Comparative Methods The YSI 2300 Stat Plus glucose analyser served as the

3 neo.lign Full teeth A3 neo.lign A I47 A3 neo.lign A D48 A3 neo.lign A M48 A3 neo.lign A D49 A3 neo.lign A B51 A3 neo.lign A S46 Upper anterior I47 D48 M48 D49 B51 48,2 47,9 48,7 49,0 52,2 Scale 1:1, dimensions in mm Minor deviations in the range of tenths of a millimeter reserved

The FreeStyle Libre 2 sensor can be scanned with either the free, FreeStyle LibreLink app4 or with a FreeStyle Libre 2 reader. However, patients choose which device they want to receive alarms: FreeStyle Libre 2 reader or FreeStyle LibreLink app. They must start their FreeStyle Libre 2 sensor with that selected device. Once the patient

STARTING THE NEO 1 OR SHUTTING IT DOWN 4 NEO 1 Quick Guide STARTING THE NEO 1 OR SHUTTING IT DOWN To start the NEO 1, press the on/off key in the top-left corner of the keyboard. The NEO 1 will start AlphaWord Plus, and you can begin typing immediately in the open file if you wish.

The Dario Blood Glucose Test Strips are for use with the Dario Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip. The Dario Glucose Control Solutions are for use with the Dario Blood Glucose Meter and the Dario Blood Glucose Test Strips to check that the meter and test

FreeStyle EN AirSep Corporation EN-1 MN131-1 rev. B - 02/13 AirSep’s FreeStyle Portable Oxygen Concentrator This Patient Manual will acquaint you with AirSep’s FreeStyle Portable Oxygen Concentrator (POC). Make sure you read and understand all the information contained in this manual

FreeStyle 5 EN AirSep Corporation EN-1 MN167-1 rev. – 02/13 AirSep’s FreeStyle 5 Portable Oxygen Concentrator This Patient Manual will acquaint you with AirSep’s FreeStyle 5 Portable Oxygen Concentrator (POC). Make sure you read and understand all of the information contained in this manual b

Peter G. Harris SHERFIELD Ian Buckbury Farm, Buckbury Lane, Newport, PO30 2NL UKIP Paul S. Martin . Anne E.V. Robertson Ivy D. Sykes Frank Vecsei ( ) Janet Champion Stephen G. Phillips Nicholas H. Finney Jean C. Burt KENDALL Gordon Sutherland 29 Beachfield Road, Bembridge, Isle of Wight, PO35 5TN Independent Patrick D. Joyce ( ) Jennifer A. Austen John L. Gansler Richard C. Beet Roger F .