Mirage FX for HerNASAL MASKUser GuideEnglish

Mirage FX for Hernasal maskFitting/配戴步骤/配戴步驟123456Mirage FXfor HeriMirage FXfor Her

sassembly123ii

Reassembly/重新组装/重新組合1234iii

Mirage FX for HerA2314756Biv

Mask �目Description/描述/描述Part number/配件编号/配件編號APACANZ1Mask frame/面罩框/面罩框6211362143 (S)6211362143 (S)2Forehead �垫/軟墊6211162136 (S)6211162136 /頭帶62129621296Elbow assembly/弯头组合/彎頭組合——7Elbow side e system/全套系统/完整系統6210762144 (S)6213462122 (S)BElbow assembly/弯头组合/彎頭組合6211462114S Small / 小号 / 小號v

Nasal MASKThank you for choosing the Mirage FX for Her. Please read the entiremanual before use.Intended useThe Mirage FX for Her channels airflow noninvasively to a patient froma continuous positive airway pressure (CPAP) or bilevel device.The Mirage FX for Her is: to be used by patients ( 30 kg) for whom positive airwaypressure has been prescribed intended for single-patient re-use in the home environment andmultipatient re-use in the hospital/institutional environment.WARNING The vent holes must be kept clear. This mask should only be used with CPAP or bilevel devicesrecommended by a physician or respiratory therapist. The mask should not be used unless the device is turned on.Once the mask is fitted, ensure the device is blowing air.Explanation: CPAP and bilevel devices are intended to be usedwith special masks (or connectors) which have vent holesto allow continuous flow of air out of the mask. When thedevice is turned on and functioning properly, new air fromthe device flushes the exhaled air out through the mask ventholes. However, when the device is not operating, insufficientfresh air will be provided through the mask, and the exhaledair may be rebreathed. Rebreathing of exhaled air for longerthan several minutes can, in some circumstances, lead tosuffocation. This applies to most models of CPAP or bileveldevices. Follow all precautions when using supplemental oxygen. Oxygen flow must be turned off when the CPAP or bileveldevice is not operating, so that unused oxygen does notaccumulate within the device enclosure and create a riskof fire.1EnglishMirage FX for Her

Oxygen supports combustion. Oxygen must not be used whilesmoking or in the presence of an open flame. Only use oxygenin well ventilated rooms. Remove all packaging before using the mask.Using your maskWhen using your mask with ResMed CPAP or bilevel devices thathave mask setting options, refer to the Technical specifications sectionin this user guide for mask setting options.For a full list of compatible devices for this mask, see the Mask/Device Compatibility List on www.resmed.com on the Products pageunder Service & Support. If you do not have internet access, pleasecontact your ResMed representative.Note:If you experience nasal dryness or irritation, using a humidifier isrecommended.FittingThe fitting illustrations section in this user guide show the sequencefor obtaining the best fit and comfort when using your mask.Fitting stepsPrior to fitting your mask, remove the elbow from the mask bypressing the side tabs. Connect the elbow to your device air tubing.1.With both lower headgear straps released, hold the maskagainst your face and pull the headgear over your head.2.Bringing the lower straps below your ears, loop the headgearinto the lower hooks on the mask frame.3.Unfasten the Velcro , pull the upper headgear straps evenlyuntil they are comfortable and reattach the Velcro.4.Repeat step 3 with the lower headgear straps.5.Attach the combined elbow assembly and air tubing to themask by pressing the side-tabs and pushing into mask frame.6.Your mask and headgear should be positioned as shown.Ensure the headgear label is facing outwards.2

Refer to the removal illustrations section in this user guide. To removeyour mask, disconnect one of the lower headgear hooks and pull themask and headgear up over your head.DisassemblyRefer to the disassembly illustrations in this user guide.Disassembly steps1.Unhook the upper headgear straps from the mask frame, thereis no need to undo the Velcro.2.Remove the combined elbow assembly and air tubing from themask by pressing the side-tabs and pulling away from the maskframe. Remove the elbow assembly from the air tubing.3.Gently separate the cushion from the mask frame.ReassemblyRefer to the reassembly illustrations in this user guide.Reassembly steps1.Ensure the cushion is oriented correctly by aligning the topcushion tab with the recess in the top of the mask.2.Gently pull the cushion through the front of the mask frameuntil the cushion locks securely into the mask frame. The tabson either side of the cushion should be sitting against the maskframe. Ensure the cushion is not twisted or distorted.3.Attach the top headgear straps to the mask frame by loopingonto each of the upper headgear hooks.4.Replace the elbow assembly into the air tubing and attachthe combined elbow assembly and air tubing to the mask bypressing the side-tabs and pushing into the mask.3EnglishRemoving the mask

Cleaning your mask in the homeYour mask and headgear should only be gently handwashed.Daily/After each use: Disassemble the mask components according to the disassemblyinstructions. Thoroughly clean the separated mask components (excludingheadgear), by gently rubbing in warm (approximately 30 C) waterusing mild soap. To optimise the mask seal, facial oils should be removed from thecushion after use. Use a soft bristle brush to clean the vent. Inspect each component and if required, repeat washing untilvisually clean. Rinse all components well with drinking quality water and allow toair dry out of direct sunlight. When all components are dry, reassemble according to thereassembly instructions.Weekly: Handwash the headgear.WARNINGDo not use aromatic-based solutions or scented oils (eg,eucalyptus or essential oils), bleach, alcohol or products thatsmell strongly (eg, citrus) to clean any of the mask components.Residual vapours from these solutions can be inhaled if notrinsed thoroughly. They may also damage the mask.CAUTIONIf any visible deterioration of a system component is apparent(cracking, crazing, tears etc), or, if after cleaning, any componentremains unclean, the component should be discarded andreplaced.4

This mask should be reprocessed when used between patients.Cleaning, disinfection and sterilisation instructions are available fromthe ResMed website, www.resmed.com/masks/sterilization. If you donot have internet access, please contact your ResMed representative.TroubleshootingProblem/possible causeSolutionMask is too noisyVent is blocked or partially blocked.If the vent requires cleaning, use asoft bristle brush.Elbow incorrectly inserted.Remove the elbow from the maskand then reassemble according tothe instructions.Mask leaks around the faceMask not positioned correctly.Gently pull the cushion away fromthe face to reinflate. Repositionyour mask according to theinstructions.Air leaking around the bridge of the nose.Mask is not tightened correctly.Tighten top headgear straps.5EnglishReprocessing the mask between patients

Air leaking around the sides of the noseMask is not tightened correctly.Tighten both upper and lowerheadgear straps.Turn off the CPAP or bilevel deviceand remove mask. Refit the maskaccording to the instructions.Ensure you position the cushioncorrectly on the face before pullingheadgear over the head.Do not slide mask down the faceduring fitting as this may lead tofolding or curling of the cushion.The mask does not fit properlyMask is incorrectly assembled.Disassemble the mask, thenreassemble according to theinstructions.Cushion may be dirty.Clean cushion according to theinstructions.WARNING The technical specifications of the mask are provided for yourclinician to check that they are compatible with the CPAP orbilevel device. If used outside specification or if used withincompatible devices; the seal and comfort of the mask maynot be effective; optimum therapy may not be achieved; andleak, or variation in the rate of leak may affect the CPAP orbilevel device function. Discontinue using this mask if you have ANY adverse reactionto the use of the mask, and consult your physician or sleeptherapist. Using a mask may cause tooth, gum or jaw soreness oraggravate an existing dental condition. If symptoms occur,consult your physician or dentist. As with all masks, some rebreathing may occur at low CPAPpressures. Refer to your CPAP or bilevel device manual for details onsettings and operational information.6

Technical specificationsThe mask contains passive venting to protect againstrebreathing. As a result of manufacturing variations,the vent flow rate may vary.Pressure–flowcurveVent flow rate (L/min)6050403020Pressure(cm 0Mask pressure (cm H2O)Dead spaceinformationPhysical dead space is the empty volume of the maskto the end of the swivel.FX: 93 mLFX Small: 71 mLTherapypressure4 to 20 cm H2OResistanceDrop in pressure measured (nominal)at 50 L/min: 0.2 cm H2Oat 100 L/min: 1.0 cm H2OEnvironmentalconditionsOperating temperature: 5 C to 40 COperating humidity: 15% to 95% non-condensing.Storage and transport: -20 C to 60 CStorage and transport humidity: up to 95% noncondensing.SoundDECLARED DUAL-NUMBER NOISE EMISSIONVALUES in accordance with ISO 4871. The A-weightedsound power level of the mask is 26 dBA, withuncertainty 3 dBA. The A-weighted sound pressurelevel of the mask at a distance of 1 m is 18 dBA, withuncertainty 3 dBA.7English At a fixed rate of supplemental oxygen flow, the inhaledoxygen concentration varies, depending on the pressuresettings, patient breathing pattern, mask, point of applicationand leak rate.

Technical specificationsGrossdimensionsGross MassMask fully assembled with elbow assembly –no headgear.FX:138 mm (H) x 86 mm (W) x 88mm (D)FX Small:124 mm (H) x 82 mm (W) x 88mm (D)Mask fully assembled, including headgear.Fx:70gFx Small: 65gMask settingoptionsFor S9: Select ‘NASAL’For other devices: Select ‘STANDARD’, otherwiseselect ‘VISTA’ as the mask option.Notes: The mask system does not contain latex, PVC or DEHP materials. This product does not contain natural rubber latex. The manufacturer reserves the right to change thesespecifications without notice.StorageEnsure that the mask is thoroughly clean and dry before storing itfor any length of time. Store the mask in a dry place out of directsunlight.DisposalThis mask does not contain any hazardous substances and may bedisposed of with your normal household refuse.8

The following symbols may appear on your product or packaging.Caution, consult accompanying documents;Code;REFCatalogue Number;Humidity limitation;Manufacturer;LOTBatchTemperature limitation;Does not contain natural rubber latex;European Authorised Representative;Indicates a Warning or Caution and alerts you to a possibleinjury or explains special measures for the safe and effectiveuse of the device;Keep away from rain;This way up;Fragile, handle with care.Consumer WarrantyResMed acknowledges all consumer rights granted under the EUDirective 1999/44/EG and the respective national laws within theEU for products sold within the European Union.9EnglishSymbols

Manufacturer:ResMed Ltd 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 AustraliaFor patent information, see www.resmed.com/ip.Mirage is a trademark of ResMed Ltd and is registered in U.S. Patent and Trademark Office.Velcro is a trademark of Velcro Industries B.V. 2017 ResMed Ltd.

213456Mirage FXfor HerMirage FXfor Her628026/4 2017-07MIRAGE FX for HERUSERAPAC-ANZ628026Global leaders in sleep and respiratory medicine   www.resmed.com

When using your mask with ResMed CPAP or bilevel devices that have mask setting options, refer to the Technical specifications section in this user guide for mask setting options. For a full list of compatible devices for this mask, see the Mask/ Device Compatibility List on www.resmed.com on the Products page under Service & Support. If you do .

Related Documents:

Homeopathic combination ingredients Nosodes . Luffeel Nasal Spray Treatment of itchy, watery eyes, sneezing, runny nose and nasal congestion Contains . nasal polyps – Allergies causing nasal congestion and a runny nose – Dry nasal membranes – Cold and flu symptoms Can be used as a nasal spray or given orally in oral drops

Nasal cavity: Initially, bilateral nasal placodes (ectoderm thickenings) appears at the rostral end of the frontonasal prominence Subsequently, growth of surrounding medial and lateral nasal processes establishes bilateral nasal pits Continued growth of nasal processes produces a primitive

Fig 1: Schematic of a saggital section of nasal cavity 1. Nasal vestibule: Most anterior part of the nasal cavity is nasal vestibule, just inside the nostrils, and presents an area about 0.6 cm2 [14]. Nasal hairs are present in this area, a

Nov 12, 2018 · 2 The Sleep in America poll was run alongside the National Sleep Foundation’s validated sleep health assessment tool, the Sleep Health Index , which has been fielded quarterly since 2016.The Index is based on measures of sleep duration, sleep quality and disordered sleep. Am

uals experience disturbed sleep at least a few nights each week12 . Research presented at SLEEP 2013, the 27th Annual Meeting of the Associated Professional Sleep Societies, LLC, addressed sleep-related topics ranging from basic sleep science, such as cell and molecular genetics, to such clinical topics as sleep disorders and sleep and aging .

Rhinological and Otological Society, Inc. Nasal Valve Suspension Revisited Michael J. Nuara, MD; Steven Ross Mobley, MD Objectives: Nasal valve suspension (NVS) is a sim-ple technique to correct nasal valve obstruction or col-lapse by providing a lateral vector of pull on the nasal sidewall. The purpose of this research was to review our

USING A RESUSCITATION MASK Note: Always follow standard precautions when providing care. Always select the properly sized mask for the victim. ASSEMBLE THE RESUSCITATION MASK Attach the one-way valve to the resuscitation mask, if necessary. POSITION THE MASK Kneel to the side of or above the victim’s head and place the rim of the mask between the .

A. Anatomi Tulang Belakang 1. Anatomi Tulang Kolumna vertebralis atau yang biasa disebut sebagai tulang belakang merupakan susunan dari tulang-tulang yang disebut dengan vertebrae. Pada awal perkembangan manusia, vertebrae berjumlah 33 namun beberapa vertebrae pada regio sacral dan coccygeal menyatu sehingga hanya terdapat 26 vertebrae pada manusia dewasa. 26 vertebrae tersebut tersebar .