Quality Assurance/Quality Control Manual - California

2y ago
20 Views
2 Downloads
446.67 KB
22 Pages
Last View : 1m ago
Last Download : 3m ago
Upload by : Karl Gosselin
Transcription

County of OrangeWater Quality DepartmentQuality Assurance/Quality Control ManualFebruary 2004County of OrangeHealth Care AgencyPublic Health LaboratoryWater Quality Department700 Shellmaker RoadNewport Beach, CA 92660(949) 219-0423FAX (949) 219-0426

County of OrangeQuality Assurance/Quality Control Project PlanRev. 1, July 14, 2004County of OrangeQuality Assurance/Quality Control Project PlanTable of Contents1.0INTRODUCTION1.1Bacteriological Monitoring QA/QC Goals1.2Organizational Scheme2.0SAMPLE COLLECTION, HOLDING TIME AND TEMPERATURE3.0SAMPLE MANAGEMENT AND DOCUMENTATION3.1Chain-of-Custody Procedures3.2Sample Log-In3.3Documentation of Field Data4.0DATA QA/QC Completeness4.4Accuracy4.5Precision4.6Sensitivity (Method Detection Limits)5.0DATA HANDLING AND MANAGEMENT5.1Bacteriological Examination of Waters Worksheet5.2Data Management5.3Data Handling6.0ANALYTICAL METHODS6.1SOP Overview7.0QUALITY ASSURANCE/QUALITY CONTROL PROCEDURES7.1QA/QC Manual Overview7.2Analytical QA/QC Procedures7.3Analyst QA/QC Procedures7.4QA/QC of Laboratory Equipment7.5QA/QC of Media, Buffer, and Reagents8.0SAFETYFIGURES AND TABLESTable 1:Table 2:Quality Control Tests, Frequencies, and Acceptable Limits.Water Sample Parameters, Methods, Sampling Requirements, and Units.Appendix A:Appendix B:Chain of CustodyMembrane Filtration Media QC ProcedureCounty of OrangeHealth Care AgencyPublic Health LaboratoryWater Quality Department700 Shellmaker RoadNewport Beach, CA 92660(949) 219-0423FAX (949) 219-0426Page 1 of 22

County of OrangeQuality Assurance/Quality Control Project PlanRev. 1, July 14, 2004INTRODUCTION1.1Bacteriological Monitoring Quality Assurance/Quality Control GoalsThe Orange County Health Care Agency (OCHCA) has been monitoring water quality atnumerous locations for over ten years. The Water Quality Department Laboratory(WQDL) Quality Assurance and Quality Control (QA/QC) Plan was developed incompliance with the QA/QC requirements set by Orange County Public HealthLaboratory, Standard Methods (1020.A-C, 1030.A-D, 1040.A-C, 1050.B, 1060.A-C,1080.A-C, 1090.A-H, 1090.J, 1100.A-C, 9030.B, (20th ed., American Public HealthAssociation)) and Environmental Laboratory Accreditation Program (ELAP)requirements. The laboratory QA/QC plan consists of strict adherence to the (1) QualityAssurance and Quality Control manual and (2) Standard Operating Procedures (SOP)manual; training manuals; maintenance of QC records; ongoing review of QCprocedures; and implementation of QA/QC improvements to provide quality results.California Department of Health Services (CDHS) also verifies the QA/QC plan bymeans of a laboratory inspection and annual requirement for acceptable analyticalperformance on performance evaluation (PE) studies. There is a 32-page ELAP on-siteinspection list that is used by DHS to confirm laboratory compliance with ELAP requiredQA/QC. Additional information regarding ELAP requirements can be found athttp://www.dhs.cahwnet.gov/.Due to the large size of the electronic versions of the laboratory SOP and QA/QCmanuals that also contain Excel spreadsheets, tables, photos and figures, the body of thisQAPP contains limited sections of the manuals that are specific to meeting the objectivesof this project. The laboratory SOP, QA/QC manuals and QC records or notebooks areavailable to project managers for review. References to sections of Standard Methods(SM) for the Examination of Water and Wastewater used by the WQD laboratory havebeen included. For example, analytical procedures that are detailed in the SOP, such asthe membrane filtration technique to enumerate bacterial densities in water are notincluded in this document but referenced with the SM number. In addition, selections ofQA/QC procedures from the WQDL QA/QC manual are located in Appendix A. ThisQAPP includes a general overview of the WQDL QA/QC practices.1.2Organizational SchemeTitle/ResponsibilityNamePhone numberE-mailLaboratory DirectorDouglas Moore,(714) 834-8385Dmoore@ochca.comPh.D.Laboratory SupervisorDonna Ferguson(949) 219-0424DFerguson@ochca.comLaboratory StaffMartin Getrich(949) 219-0423Mgetrich labhca@sbcglobal.netLaboratory StaffMariam(949) 219-0428Mzhowandai labhca@sbcglobal.netZhowandaiCounty of OrangeHealth Care AgencyPublic Health LaboratoryWater Quality Department700 Shellmaker RoadNewport Beach, CA 92660(949) 219-0423FAX (949) 219-0426Page 2 of 22

County of OrangeQuality Assurance/Quality Control Project PlanRev. 1, July 14, 20042.0SAMPLE COLLECTION, HOLDING TIME, AND TRANSPORT2.1SAMPLE COLLECTION2.1.1When collecting the sample, leave enough air space in the bottle to allow forproper mixing before examination. Collect samples that are representative ofthe water being tested. Flush or disinfect sample ports, and use aseptictechniques to avoid sample contamination. Keep the sample bottle closed untilit is to be filled. Remove cap carefully to avoid contaminating the inner surfaceof the cap and neck of the bottle. Fill container without rinsing. Replace capimmediately. The volume of sample should be sufficient to carry out all testsrequired (not less than 100 ml). Provide complete and accurate sampleidentification information as specified on the sample collection sheet (chain ofcustody form).2.1.2Sample Containers2.1.2.1 For bacteriological samples, use sterilized bottles of glass or plastic ofappropriate size and shape. Bottles must be capable of holding asufficient volume of sample for all required tests, while allowing for airspace. Commercially available, wide-mouthed, autoclavable or presterilized Polypropylene bottles of suitable size are satisfactory. Presterilized plastic bags, with or without a de-chlorinating agent, mayalso be used. Water samples that may have residual chlorine or anotherhalogen require sample bottles containing a reducing agent, such assodium thiosulfate. Sodium thiosulfate neutralizes any residual halogenand prevents the continuation of bactericidal action during sampletransit.2.1.2.2 OCHD water collection bottles are quality controlled by batch (SeeMedia QC Notebook). Sample bottles are stored in the waterlaboratory for use by Environmental Health personnel and othergovernmental agencies upon request.2.1.3Sample Types2.1.3.1 Potable Water For drinking water analysis, collect samples consisting of finishedwater.Open tap fully and flush water for 2 - 3 minutes, or until a timesufficient to permit clearing the service line.Reduce water flow to permit filling bottle without splashing.Do not sample from leaking taps that allow water to flow over theoutside of the tap.When sampling from a mixing faucet remove faucet attachments,run hot water for 2 minutes, then cold water for 2 to 3 minutes, andcollect sample as indicated above.County of OrangeHealth Care AgencyPublic Health LaboratoryWater Quality Department700 Shellmaker RoadNewport Beach, CA 92660(949) 219-0423FAX (949) 219-0426Page 3 of 22

County of OrangeQuality Assurance/Quality Control Project PlanRev. 1, July 14, 2004 If the sample is taken from a well fitted with a hand pump, pumpwater for 5 minutes before collecting sample.If the well is equipped with a mechanical pump, collect samplefrom a tap on the discharge.If there is no pumping machinery, collect a sample directly fromthe well.2.1.3.2 Raw WaterFor collecting samples directly from a river, stream, lake, reservoir,spring, or shallow well, obtain samples representative of the water thatis the source of supply to consumers.2.1.3.3 Surface WatersSelect sampling locations to include a baseline location upstream fromthe study area. Where a tributary stream is involved, select thesampling point near the confluence with the mainstream points. Tomonitor stream and lake water quality, establish sampling locations atcritical sites. Sampling frequency may be seasonal.2.1.3.4 Bathing BeachesSampling locations for recreational areas should reflect water qualitywithin the entire recreational zone. Collect samples in the swimmingarea from a uniform depth of approximately 1 m. To obtain baselinedata on marine and estuarine bathing water quality include sampling atlow, high, and ebb tides. Relate sampling frequency directly to thepeak bathing period.2.1.3.5 Swimming PoolsA swimming pool is a body of water of limited size contained in aholding structure. The water is generally chlorinated potable water butmay also be derived from thermal springs or saltwater. Collect samplesin the area and time of maximum bather density. Collect sampleswhere water is 1 m. (See Std. Methods17th Edition Section 9213 B. forfurther information.)2.1.3.6 Sediment and SludgeSediments provide a stable index of the general quality of the overlyingwater. Sampling frequency may be related to seasonal changes inwater temperatures and storm water runoff. Sludge monitoring mayindicate the effectiveness of wastewater treatment processes. Bottomsediment sampling requires special apparatus. (See Std. Methods 9060A. 3. g.)2.2HOLDING TIME AND TEMPERATURE2.2.1GeneralCounty of OrangeHealth Care AgencyPublic Health LaboratoryWater Quality Department700 Shellmaker RoadNewport Beach, CA 92660(949) 219-0423FAX (949) 219-0426Page 4 of 22

County of OrangeQuality Assurance/Quality Control Project PlanRev. 1, July 14, 2004Start microbial analysis of water samples as soon as possible after collection toavoid unpredictable changes in the microbial population. Keep samples coldduring transport to the laboratory (4-10ºC), if they cannot be processed within 1hour after collection. Blue ice packs are preferred over ice. If using ice, avoiddirect contact of samples with ice using plastic packing material.2.2.2Drinking Water for Compliance PurposesHold samples at 4-10ºC during transit to the laboratory. Analyze samples onday of receipt whenever possible and refrigerate overnight if arrival is too latefor processing on same day. Do not exceed 30-hour holding time fromcollection to analysis for coliform bacteria. Do not exceed 8 hour holding timefor heterotrophic plate counts.2.2.3Non-potable Water for Compliance PurposesHold source water, stream pollution, recreational water, and wastewater samplesbelow 4-10ºC during a maximum transport time of 6 hours. Refrigerate thesesamples upon receipt in the laboratory and process within 2 hours. Whentransport of samples is longer than 6 hours consider using delayed incubationprocedures.2.2.4Water for Non-compliance PurposesHold samples between 4-10ºC during transport and until time of analysis. Donot exceed 24-hour holding time.3.0SAMPLE MANAGEMENT AND DOCUMENTATION3.1Chain-of-Custody Procedures3.1.1Samples that are transferred from one agency to another agency for analysisrequire the use of Chain-of-Custody (COC) procedures that include thefollowing requirements for the laboratory to accept custody of samples:3.1.1.1 Sample LabelSamples must be properly labeled using waterproof ink to record thesample number/description, date and time collected.3.1.1.2 Chain-of-Custody FormsOrange County WQD Laboratory provides COC forms to fieldsampling personnel for detailed record keeping. There are two separatedocuments that make up the COC form. The field data sheet, referredto as the “Bacteriological Examination of Waters (BEW)” worksheet isthe first COC form used and contains information such as the projectname, sample identification, water type (marine, freshwater or other),weather, date and time of collection, sample location, field samplername, field bottle number and tests requested. The BEW worksheetmust accompany the sample during sample collection and transport tothe laboratory. The second form is the “Water Lab Sign-In Sheet(WLSIS)” which is filled out at the lab upon sample delivery.3.1.1.3 Transfer of CustodyCounty of OrangeHealth Care AgencyPublic Health LaboratoryWater Quality Department700 Shellmaker RoadNewport Beach, CA 92660(949) 219-0423FAX (949) 219-0426Page 5 of 22

County of OrangeQuality Assurance/Quality Control Project PlanRev. 1, July 14, 2004Immediately following receipt of water samples to the laboratory, alaboratory assistant or microbiologist will conduct inventory anddocument information regarding sample transport and laboratoryprocessing according to the “Logging in Membrane Filtration Samples”SOP. The inventory consists of checking the samples for properlabeling, cross-referencing sample labels with the BEW worksheet andreading the cooler thermometer to ensure proper transport conditions.Laboratory personnel receiving samples will record the date and timeof sample receipt, number of samples, type of samples and coolertemperature and other pertinent comments on the “Water Lab Sign-InSheet”. The sample deliverer and the analyst receiving the samplesmust initial the WLSIS. Samples received leaking, broken, containinginsufficient volumes, exceeding holding times or stored in coolers withtemperatures above 10ºC will not be accepted. The sample anomalieswill be pointed out to the sample deliverer and also noted on theWLSIS and BEW worksheet. The time of sample receipt will berecorded on the WQD “Bacteriological Examination of Waters(BEW)” worksheet that serves as the field data sheet and accompaniesthe samples. The BEW worksheet will be then be checked forcompleteness of field data information. Corrections to worksheets willbe made by crossing out the incorrect information, recording thechange and recording the date and initials of the analyst making thechange. When the COC forms are completed, copies may be providedto the appropriate party(s).3.23.3Sample Log-In3.2.1Following transfer of custody, the samples are logged in with a laboratorynumber, which is used to track the sample throughout the analytical process.Laboratory numbers are assigned in sequential numeric order of receipt. Watersamples for bacteriological testing are either tested immediately or refrigeratedat 5 ºC to maintain sample integrity. Samples for compliance testing are testedwithin 2 hours of receipt in the laboratory and less than 6 hours from samplecollection or up to 24 hours for noncompliance purposes.3.2.2Upon completion of bacterial analysis, the samples are stored for a minimum of24 hours at 5ºC until they are either archived in storage areas or properlydisposed.Documentation of Field DataField data will be recorded in field notebooks and field data sheets. Field notebooksshould be used to document field observations that are supplemental to field datarecorded on the field data sheet.4.0DATA QA/QC OBJECTIVESTo produce acceptable testing results, the general data quality objectives for this QA/QC Manualare to ensure that the data is representative, comparable, complete, accurate, and precise.Acceptable results are those values that fall within the acceptable range specified. Correctiveactions for unacceptable results for specific testing methods are detailed in the SOP and QA/QCmanuals. All corrective actions taken are documented in the QA/QC manual. The laboratory willnotify the Project Manager of any samples that are impacted by unacceptable QA/QC results.County of OrangeHealth Care AgencyPublic Health LaboratoryWater Quality Department700 Shellmaker RoadNewport Beach, CA 92660(949) 219-0423FAX (949) 219-0426Page 6 of 22

County of OrangeQuality Assurance/Quality Control Project PlanRev. 1, July 14, 20044.1RepresentativenessRepresentativeness is the degree to which the data represent the actual condition of asampling site. The following factors determine the representativeness of the data:sampling location, sampling frequency, sample type, sample collection methods, samplepreservation, sample holding times and analytical methods used. These factors arecritical components of a sampling plan designed to maximize representativeness of thedata to the extent practicable.4.2ComparabilityComparability of data is the degree to which the data produced by one laboratory or studycan be compared to another. The WQD laboratory uses EPA approved analyticalmethods where possible or methods that have been determined to produce measurementdata of known and quality sufficient to meet the objectives of this project. The data willbe reported in commonly used units.4.3CompletenessThe completeness of data is the percentage of planned data that will be used to meetstatistical criteria needed to reach study conclusions. Acquiring 100% of the dataplanned is difficult due to unexpected circumstances, adverse weather conditions,equipment problems, laboratory error, loss of samples or samples that are invalid becausethey do not meet all of the laboratory sample acceptance criteria. The goal of this projectis to obtain 80% of data completeness. Percent completeness is the number of datavalues generated/number of samples collected multiplied by 100.4.4AccuracyAccuracy is the degree to which the measurement is to its true value. Accuracy of theWQD laboratory methods is determined by means of testing the following: (1)performance evaluation (PE) samples consisting of known quantities of bacteria, (2)performance of culture media, (3) laboratory and field blanks and (4) split samples. Inaddition, equipment calibration checks are routinely done to ensure accuracy ofmeasurements.4.4.1Performance evaluation studiesTo assess laboratory accuracy and comparability of bacterial density estimation,the laboratory participates in performance evaluation (PE) studies once a year.Certified PE materials or challenge samples are purchased from vendorsapproved by the CDHS. These include samples spiked with known amounts ofbacteria provided from the vendor. The vendor evaluates the PE sample testingresults using target and range values generated from data produced by severallaboratories using the same analytical methods. The results must fall within 3standard deviations of mean bacterial counts obtained from participatinglaboratories. The results are sent to the laboratory and to CDHS. If thelaboratory receives a “not acceptable” rating for a method, they mustimmediately review their work, implement the necessary corrective action andsend a summary of the correction action to the CDHS. ELAP certifiedlaboratories must receive an acceptable rating for ELAP approved fields oftesting on an annual basis to meet the certification requirements.County of OrangeHealth Care AgencyPublic Health LaboratoryWater Quality Department700 Shellmaker RoadNewport Beach, CA 92660(949) 219-0423FAX (949) 219-0426Page 7 of 22

County of OrangeQuality Assurance/Quality Control Project PlanRev. 1, July 14, 20044.4.2Culture media QCThe performance of culture media used to enumerate bacterial densities is alsotested to ensure accuracy of bacterial enumeration using membrane filtration.Refer to the “Membrane Filtration Media Quality Control” procedure inAppendix A. Briefly, a known amount of indicator bacteria is spiked intophosphate buffered water and filtered as routine samples. The membrane isplaced on culture media, incubated overnight and the colonies on membranefilters are enumerated. The media is also tested using negative control bacteriathat should not grow on selective media. The number and appearance ofbacterial colonies is recorded. Each new lot of media is checked in parallel withthe old lot.4.4.3Field blanksField blanks are used to check for background contamination as well as handlingand storage problems that may affect the results. A field blank for bacterialanalysis should be tested for each sampling event or at least once a month foreach storm drain sampling site. The blank may be prepared using reagent water(i.e., double distilled water or sterile de-ionized water) that has been bottled in alaboratory or is commercially available. The field sampler or personnel from anagency external to the lab should prepare the blank in the field by pouringreagent water into the sample bottle using sterile techniques i.e., wearing sterilegloves and avoiding aerosol production/exposure. The field blank should beincluded with the routine samples delivered to the WQD laboratory. The blankshould not yield a value higher than that allowed by the acceptance limits (TableX). In the case of membrane filtration of reagent water, the counts should bebelow detection limits i.e., no colony forming units should be detected.4.4.4Laboratory blanksThe membrane filtration equipment, membrane filters and dilution buffers usedto process the samples are tested for possible bacterial contamination that canoccur from carryover contamination due to insufficient sterilization of thefiltration apparatus between samples. The WQDL uses sterile, disposablepipettes, funnels, and forceps to minimize potential bacterial contamination andto increase sample throughput. To test the sterility of supplies and equipment,sterile dilution buffer is processed similarly as a water sample at the beginning,middle and end of the membrane filtration test run (uninterrupted series ofanalyses).If any contamination is found, the supervisor is notifiedimmediately. An investigation of the source is initiated and the analytical datafrom samples tested with these materials is rejected (SM 9020B.8.a.5). Refer to“Membrane Filter Manifold Sterility Check” and “Sterility Testing” sections inAppendix A.4.4.5Calibration checksLaboratory instruments are inspected and calibrated by laboratory personnel orequipment maintenance contractors using standards as per manufacturersinstructions or the SOP. Electronic instruments for water analyses performed inthe field will be calibrated once a day, prior to use.4.5PrecisionCounty of OrangeHealth Care AgencyPublic Health LaboratoryWater Quality Department700 Shellmaker RoadNewport Beach, CA 92660(949) 219-0423FAX (949) 219-0426Page 8 of 22

County of OrangeQuality Assurance/Quality Control Project PlanRev. 1, July 14, 2004Precision is the degree of agreement of repeated measurements of the sample, usuallyreported in standard deviation or relative percent difference (RPD). The precision ofbacterial enumeration for the membrane filtration method is determined using the“Precision of Quantitative Methods” and “Duplicate Analyses” procedures.4.5.1Precision of Quantitative MethodsInitial demonstration of capability is conducted by each analyst beforeperforming any analysis of samples and annually thereafter. The precision inperforming the membrane filtration method is determined as outlined in“Method of Precision” in Appendix A. Briefly, analysts perform duplicateanalyses on the first 15 positive samples tested for indicator bacteria testing.Since the laboratory currently analyzes over 200 samples per week, the precisioncriterion for each analyst can easily be determined. The precision criterion iscalculated as 3.27 times the mean relative range value, “R” of duplicate resultsand is determined for each analyst of the WQD laboratory. The precisioncriterion for each analyst is determined annually. Precision between analysts incounting colonies on culture media plates is also assessed monthly following the“Analyst Comparison of Plate Counts” in Appendix A.4.5.2Duplicate AnalysesDuplicate samples are analyzed to assess the reproducibility of the sampling andanalysis methods.4.5.2.1 Laboratory DuplicatesTo asses precision of the membrane filtration method, duplicateanalysis is performed on 10% of all samples or at least one sample pertest run. Refer to “Method of Precision” in Appendix A. The duplicateanalyses is also used to determine the precision criterion for eachanalyst performing water testing and the results are compared for allanalysts using an Excel spreadsheet.4.5.2.2 Field DuplicatesField duplicates will be tested to assess the repeatability of sampling.A field duplicate is a second sample that is collected at the same timeor immediately following collection of the first sample. At least onefield duplicate per sampling event or a minimum of one per monthfrom each storm drain sampling site will be tested for indicatorbacteria.4.6Sensitivity (Method Detection Limits)Method Detection Limit is the lowest possible concentration that the equipment oranalysis can detect. In the case of bacterial enumeration by membrane filtration, the“minimum detection limit” of bacterial colony forming units (CFU) using membranefiltration is based on the volume of water tested. Fecal indicator bacterial standards arebased on bacterial counts per 100 ml of water. If 100 ml of sample is filtered and nobacteria are detected, the count is reported as “less than” ( ) 1 CFU/100 ml becausebacteria may have been present but not detected. Thus, “1 CFU/ 100 ml” is also theminimum detection limit. If only 10 ml of the 100 ml sample is filtered and no bacteriaare detected, the count is reported as “less than 10 CFU/100 ml”. The minimumCounty of OrangeHealth Care AgencyPublic Health LaboratoryWater Quality Department700 Shellmaker RoadNewport Beach, CA 92660(949) 219-0423FAX (949) 219-0426Page 9 of 22

County of OrangeQuality Assurance/Quality Control Project PlanRev. 1, July 14, 2004detection limit in this case is also 10 CFU/100 ml. For membrane filtration of stormdrainwater samples, a minimum of 3 dilutions or volumes of water will be filtered to optimizedetection of bacteria. Testing volumes should result in bacterial counts in the “countablerange” as per Standard Methods to ensure accuracy of results. If the total number ofbacterial counts exceeds 200 per membrane, or if the colonies are not distinct enough foraccurate counting, the results will be reported as greater than or equal to ( ) themaximum number countable.5.0DATA MANAGEMENT AND HANDLING5.1Bacteriological Examination of Waters Worksheet (BEW)5.1.15.2 Date and time that the samples are processed Date and time analysis was completed Testing results Corrective actions5.1.2The laboratory results are checked by a second analyst (Microbiologist) foraccuracy of the calculations (colony forming units (CFU)/100 ml) andcompleteness of the worksheet. Both the analyst that records the results andmicrobiologist who confirms calculations sign off on the results.5.1.3Corrective actions regarding sample collection, preservation and transport aredocumented on the worksheet in the “Field or Lab Remarks” section. Thisdocumentation includes the date, analyst, sample affected, problem andresolution.Data Management5.2.15.3Laboratory results include the following information documented on the BEWworksheet as described in the “Reporting Water Lab Results” section of theSOP:The laboratory results will be entered into an electronic spreadsheet (MS Excel)that is available to the project manager after the laboratory has reviewed thedatabase for correctness of data entry. The BEW worksheets and back-up disksare kept at the laboratory for five years and also are made available to theProject Manager upon request.Data Handling5.3.1Distribution of analytical resultsBacterial densities in environmental waters are highly variable, ranging frommany values below detection to a few high ones. In such cases the data wouldbe positively skewed and not normally distributed about the mean. Sincestatistical analysis assumes a normal distribution of data, the numbers must beconverted to their logarithms to approximate a symmetrical distribution. Thebest estimate of central tendency of log-normal data is the geometric mean orantilog of the arithmetic mean of the logarithms.County of Orange700 Shellmaker RoadHealth Care AgencyNewport Beach, CA 92660Public Health Laboratory(949) 219-0423Water Quality DepartmentFAX (949) 219-0426Page 10 of 22

County of OrangeQuality Assurance/Quality Control Project PlanRev. 1, July 14, 20045.3.2“Less than” ( ) valuesWater samples with no indicator bacterial detected are reported with “less than”( ) values using the calculations in Standard methods (9222B.6). If no coloniesare detected, the calculation is done as if there was 1 colony detected andreported with the “less than” qualifier “ ”. With the exception to 1/100 mlvalues reported for testing 100 ml volumes of, test volumes should be adjusted,when possible to avoid the use of qualifiers.If there are large numbers of values with the “less than” or “greater than”qualifiers, the qualifier can be omitted so that the values can be included in thedata analysis. However, values with “less than” qualifiers may also beconverted to 1, 0 or ½ (values halfway between zero and the “less than” value)depending on the value, the detection limit and distribution of values in the dataset.5.3.3“Greater than or equal to” ( ) values, TNTC and Confluent CountsWater samples with high levels of indicator bacteria, particularly total coliforms,may result in counts reported as “greater than or equal to” values, “too numerousto count” (TNTC) or “confluent” (CONF).Such samples usually containbacteria other than the specific indicator bacteria being tested, such as noncoliforms that can also grow on the membranes. High levels of non-coliforms oratypical bacteria may interfere with the detection of typical coliform bacteriasuch that the actual number of total coliforms in the sample may be greater orequal to the number detected. In this case, the number of typical coliformcolonies is reported but with the “ “qualifier. If no total coliforms are detectedand there are greater than 200 colonies, the results are reported as “TNTC”. Ifthe total number of bacteria colonies (coliforms plus non-coliforms) exceeds 200per membrane, the number of total coliforms detected is reported with thenotation “(TNTC)”. For example, if 10 total coliforms are detected, but thereare greater than 200 total coliforms and non-coliforms on the membrane, theresults are reported as “10 (TNTC)”. If colonies cover the entire membrane andare not discrete enough for accurate counting, the results are reported as“confluent growth with coliforms present” or “confluent growth withoutcoliforms detected”. In each of these situations, additional samples should becollected. However, this may not be possible for samples collected duringspecial events or circu

Quality Assurance/Quality Control Project Plan Rev. 1, July 14, 2004 County of Orange 700 Shellmaker Road Health Care Agency Newport Beach, CA 92660 County of Orange Quality Assurance/Quality Control Project Plan Table of Contents 1.0 INTRODUCTION 1.1 Bacteriological Monitoring QA/QC Goals 1.2 Organizational Scheme

Related Documents:

critical issues the University has established a Quality Assurance Directorate, which is mandated to develop a Quality Assurance Framework and a Quality Assurance Policy. The Quality Assurance Framework would clearly spell out the Principles, Guidelines and Procedures for implementing institutional quality assurance processes.

Quality Assurance and Quality Control (QA/QC) policy. Quality assurance/quality control measures for water treatment utilities refer to a set of activities that are to be undertaken to ensure compliance and above all, ensure that the water is safe for public consumption in a sustainable manner. In general, quality assurance (QA) refers to the

This quality assurance manual specifies the methods to prepare and submit Quality Assurance Process Design Diagram for products and parts to be supplied to NSK by suppliers. 2. Purpose Each supplier should prepare quality assurance process design diagram clearly showing the quality assurance methods used in each products and parts production .

Quality Assurance Representative. The Site Manager will appoint a member of the Project Office to control all Quality Assurance issues including - Assisting the Site Manager to co-ordinate and develop the Quality Assurance Plan. Advise Engineers, General Foremen, Foremen and Chargehands in all matters of Quality Assurance.

Quality Assurance and Improvement Framework Guidance 2 Contents Section 1: Quality Assurance and Improvement Framework 1.1 Overview 1.1.1 Quality Assurance (QA) 1.1.2 Quality Improvement (QI) 1.1.3 Access 1.2 Funding Section 2: Quality Assurance 2.1 General information on indicators 2.1.1 Disease registers 2.1.2 Verification

Software Quality Assurance Plan (SQAP) for the SRR-CWDA-2010-00080 H-Area Tank Farm (HTF) Performance Revision 0 Assessment (PA) Probabilistic Model August 2010 Page 5 of 15 1.0 SCOPE This Software Quality Assurance Plan (SQAP) was developed in accordance with the 1Q Quality Assurance Manual, Quality Assurance Procedure (QAP) 20-1, Rev. 11.

Sep 02, 2020 · Zetec Quality Assurance Manual Revision M Page 1 of 30 QUALITY ASSURANCE MANUAL ZQA-001 Revision M Successful implementation of the requirements defined in this Quality Assurance Manual requires the concerted effort of all Zetec employees since everyone contributes to the quality of our products and services.

The NELAC Manual was used to insure the Quality Assurance Manual more fully addressed the issues that regulators and clients would be looking for in our Quality Assurance Manual and to anticipate possibly getting accredited under NELAC in the near future. Revision 9.1 (January 28, 2004) The Quality Assurance Manual underwent the annual review.