Product Data Sheet00813-0100-4808, Rev FAJune 2008Rosemount 3051NThe Rosemount 3051N Smart PressureTransmitter for Nuclear ServiceINDUSTRY LEADING PERFORMANCE Qualified per IEEE Std 344-1987 and IEEE Std323-1983 (mild environment) Superior Performance with 0.075% accuracy 100:1 Rangeability reduces inventory costs Non-interacting Zero and Span adjustmentreduces calibration time Adjustable damping Internal diagnostics Coplanar sensor/process interface formaximum mounting flexibilityIntroduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2Transmitter Functional Operation and Design. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2Nuclear Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4Performance Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5Functional Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6Physical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 8Dimensional Drawings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 11Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 14www.rosemountnuclear.com

Product Data SheetRosemount 3051N00813-0100-4808, Rev FAJune 2008Results Driven By Proven MeasurementINTRODUCTIONThe Rosemount 3051N Coplanar Smart PressureTransmitters are designed for precision differential,gauge, and absolute pressure measurementsrequiring reliable performance and safety. Thesetransmitters are seismically qualified for use in Class1E safety related applications per IEEE Std 344-1987at SSE response spectrum levels up to 16.5g’s, andper IEEE Std 323-1983 (mild environment).FIGURE 1. The Rosemount 3051ND/NG Sensor ModuleTypical Cut-away ionTRANSMITTER FUNCTIONALOPERATION AND DESIGNThe Rosemount 3051N is designed with a uniqueCoplanar sensor/process interface. Performance andreliability improvements over traditional designs areachieved by moving the transmitter sensor from theflange interface into the module neck, thereby reducingflange stress, thermal, and process interface effects.The Coplanar sensor platform also allows greaterflexibility in process interfaces and mountingconfigurations.For differential and gauge pressure measurements,the Rosemount 3051N utilizes capacitance sensortechnology similar to that of the Rosemount 1151,1152, 1153, and 1154 transmitters (see Figure 1).Rosemount capacitance technology delivers thehighest inherent performance, stability and reliabilityin the process industry as proven in millions ofinstallations worldwide.2SiliconeOilIsolatingDiaphragms

Product Data Sheet00813-0100-4808, Rev FAJune 2008Rosemount 3051NDuring operation, process pressure is transmittedthrough isolating diaphragms and silicone oil fill fluidto a center diaphragm in the alpha-cell capacitancesensor (see Figure 1). The displacement of thecenter diaphragm is proportional to the pressuredifferential across it. The position of the centerdiaphragm is detected through differentialcapacitance between it and capacitor plates locatedon each side. The differential capacitance isprocessed electronically through a microprocessor toa 2 wire 4–20mA (digital) HART (HighwayAddressable Remote Transducer) output signal.Each unit completes a compensation and verificationprocess during manufacturing where the uniquesensor characteristics are measured over pressureand temperature and retained in the device tooptimize performance over a wide operating range.FIGURE 2. The Rosemount 3051NA Sensor ModuleTypical Cut-away DiagramPiezoelectricSilicon SensorFor absolute pressure measurements, theRosemount 3051N utilizes piezoelectric siliconsensor technology designed and manufactured at theRosemount Solid State Technology Center (seeFigure 2).FIGURE 3. The Rosemount 3051N Differential Pressure Transmitter 4–20 mA Block DiagramSensor ModuleElectronics Board4–20 mA Signalto Control SystemSignal er Sensor Linearization Rerange Damping Diagnostics Engineering CommunicationCapacitiveSensorModule Memory Rerange Values ConfigurationPressureDigital-to-AnalogSignal ConversionDigitalCommunicationLocal Spanand ZeroAdjustmentPressureRosemount HARTCommunicator 275 or 3753

Product Data Sheet00813-0100-4808, Rev FAJune 2008Rosemount 3051NSPECIFICATIONSNuclear SpecificationsQualified for nuclear use per IEEE Std 344-1987 and IEEE Std 323-1983 (mild environment) as documented inRosemount Report D2001019SeismicTABLE 1. Seismic Specifications SummaryRangeCode3051NDDifferentialModelDuring SeismicAccuracy(1)Post SeismicAccuracySpecified Seismic MaximumWorking Pressure0not specifiednot specified750 psi (5,2 MPa)1(2)0.75% of URL(adjustable damping 1.6 s)2(2)0.75% of URL(adjustable damping 0.8 s)30.75% of URL2000 psi (13,8 MPa)3000 psi (20,7 MPa)(glass-filled TFE o-ring)2000 psi (13,8 MPa)(EP o-ring)40.25% of URL3051NGGauge52(2)0.75% of URL(adjustable damping 0.8 s)30.75% of URLStructuralIntegrity0.25% of e45Upper Range Limit0.25% of URL01not specifiednot specified0.25% of URL0.25% of spanAbsolute23051NA33000 psia(glass-filled TFE o-ring)42000 psi (13,8 MPa)(EP o-ring)(1) User-adjustable damping set at 0.4 s unless otherwise noted.(2) Mounting bracket (Option Code B2, BS, or PM) required for specified “During Seismic Accuracy” performance.EnvironmentalPerformance to normal operating limits as describedin the Performance Specifications and FunctionalSpecifications sections of this document.Hydrostatic TestingModelRange CodeHydrostatic Test Pressure(1)3051ND012–5750 psi2000 psi4200 psi3051NG2–5150% of maximumworking pressure(2)Quality Assurance ProgramIn accordance with 10CFR50 Appendix B,ISO 9001:2000Nuclear CleaningTo 1 ppm chloride content4(3)3051NA1–4(1) Process o-ring Code A. (Glass filled TFE)(2) Maximum Working Pressure equals upper range limit (URL).(3) Hydrostatic testing not performed on The Rosemount 3051NARange Code 0.

Product Data Sheet00813-0100-4808, Rev FAJune 2008Performance SpecificationsBased upon zero-based calibrations, referenceconditions, 4–20mA analog output, and digital trimvalues equal to the span setpointsReference Accuracy(1)Includes hysteresis, terminal-based linearity,and repeatabilityThe Rosemount 3051NDRange Code Reference Accuracy012–5 0.10% calibrated span from 1:1 to 2:1 RDF 0.05% upper range limit from 2:1 to 30:1 RDF 0.10% calibrated span from 1:1 to 15:1 RDF (0.005% URL 0.025% span)from 15:1 to 50:1 RDF 0.075% calibrated span from 1:1 to 10:1 RDF (0.005% URL 0.025% span)from 10:1 to 100:1 RDFThe Rosemount 3051NGRosemount 3051NDriftThe Rosemount 3051ND, NG, NARange Code12–52–5Reference Accuracy 0.075% calibrated span from 1:1 to 10:1 RDF (0.005% URL 0.025% span)from 10:1 to 100:1 RDFThe Rosemount 3051NARange Code01–4Reference Accuracy 0.075% calibrated span from 1:1 to 5:1 RDF (0.01% URL 0.025% span) from 5:1 to 30:1RDF 0.075% calibrated span from 1:1 to 10:1 RDF (0.0075% URL) from 10:1 to 100:1 RDF (0.2% URL 0.2% span) for 30 months 0.2% URL for 30 monthsAmbient Temperature Effect (1) (2)The Rosemount 3051ND/NGRange CodeAmbient Temperature Effect per 50 F (28 C)0 (0.25% URL 0.05% span)1 (0.1% URL 0.25% span) from 1:1 to 30:1 (0.14% URL 0.15% span) from 30:1 to 50:12–5 (0.0125% URL 0.0625% span) from 1:1 to 5:1 (0.025% URL 0.125% span) from 5:1 to 100:1The Rosemount 3051NARange Code Ambient Temperature Effect per 50 F (28 C)0Range CodeDrift1–4 (0.1% URL 0.25% span) (0.025% URL 0.125% span) from 1:1 to 30:1 (0.035% URL 0.125% span) from 30:1 to 100:1Overpressure EffectMaximum zero shift after overpressure of MaximumWorking PressureThe Rosemount 3051NDRange Code Overpressure Effect0–3 0.5% URL4–5 3.0% URLThe Rosemount 3051NGRange Code Overpressure Effect2–4 0.25% URL5 0.30% URLThe Rosemount 3051NARange Code Overpressure Effect0–4(1)(2) 0.05% URLRDF Range Down Factor URL / Calibrated SpanExposure of isolator diaphragms to process temperatures above 185 F (85 C) but below 250 F (121 C) produces a temperatureeffect of 1.0% of calibrated span in addition to the effects listed.5

Product Data Sheet00813-0100-4808, Rev FAJune 2008Rosemount 3051NStatic Pressure EffectThe Rosemount 3051NDZero Error (can be calibrated out at line pressure)Per 1000 psi (6,9 MPa) line pressureRange CodeStatic Pressure Zero Effect(1)0(2) 0.125% URL for Ps 750 psi (5,2 MPa)1 0.25% URL for Ps 2000 psi (13,8 MPa)2, 3 0.05% URL for Ps 2000 psi (13,8 MPa) [0.1 0.1 (Ps-2000) / 1000]% URL forPs 2000 psi 3626 psi4, 5 0.1% URL for Ps 2000 psi (13,8 MPa) [0.2 0.2 (Ps-2000) / 1000]% URLfor Ps 2000 psi 3626 psiFunctional SpecificationsServiceLiquid, gas, or vaporOutput4–20 mA, user-selectable for linear orsquare root output; digital signal based on HART protocolPower SupplyLoad LimitationsMaximum loop resistance is determinedby the voltage level of the external power supply,as described by:(1) Ps equals static line pressure applied(2) Specification for The Rosemount 3051N Range 0 is expressedin [% per 100 psi (689 KPa)]Max. Loop Resistance 43.5 (Power Supply Voltage – 10.5) ohms1935Span ErrorPer 1000 psi (6,9 MPa) line pressureRange Code15001000 0.15% input reading1 0.40% input reading2, 3 0.10% input reading4, 5 0.20% input reading (uncertainty after calibrationcorrection for systematic effects)(1) Specification for The Rosemount 3051N Range 0 is expressedin [% per 100 psi (689 KPa)] up to 750 psi (5 171 KPa)DesignRegion500Static Pressure Span Effect(1)0Load (ohms)The Rosemount 3051ND010.5203040Voltage (V dc)55Communication requires a minimum loop resistance of 250 ohms.Temperature LimitsAmbient0 to 185 F (-18 to 85 C)Power Supply EffectLess than 0.005% of calibrated span per volt forRDF 10Load EffectNo load effect other than change in voltage supplied tothe transmitterMounting Position EffectThe Rosemount 3051ND/NGZero shifts up to 1.25 in H20 (0,31 KPa), whichcan be calibrated out; no span effectwith meter option:0 to 175 F (-18 to 80 C)Process(1) (2) (3)0 to 250 F (-18 to 121 C) Coplanar flange0 to 300 F (-18 to 149 C) traditional flangeStorage0 to 212 F (-18 to 100 C)with meter option:0 to 185 F (-18 to 85 C)The Rosemount 3051NAZero shifts up to 2.5 in H20 (63,5 mm), which canbe calibrated out; no span effect(1)(2)(3)6Process temperatures above 185 F (85 C) require derating the ambient temperature limitsby 1.5 F per degree above 185 F.220 F (104 C) limit in vacuum service; 130 F (54 C) for pressures below 0.5 psiaEPR process O-ring is limited to 150 F (66 C) process temperature.

Product Data Sheet00813-0100-4808, Rev FAJune 2008Rosemount 3051NSpan and Zero, Zero Elevation, and SuppressionZero and span values can be set anywhere within the range limits stated in Table 2 and Table 3, providing sensorlimits are not exceeded.Span must be greater than or equal to the minimum span stated in Table 2 and Table 3.TABLE 2. The Rosemount 3051ND and 3051NG Range and Sensor LimitsMinimum SpanRange and Sensor LimitsLower (LRL)RangeThe Rosemount 3051ND, NGUpper (URL)The Rosemount 3051NDThe Rosemount 3051NG00.1 in H20 (25 Pa)3.0 in H20 (750 Pa)-3.0 in H20 (-750 Pa)NA10.5 in H20 (0,12 kPa)25 in H20 (6,22 kPa)-25 in H20 (-6,22 kPa)NA22.5 in H20 (0,62 kPa)250 in H20 (62,2 kPa)-250 in H20 (-62,2 kPa)-250 in H20 (-62,2 kPa)310 in H20 (2,48 kPa)1000 in H20 (248 kPa)-1000 in H20 (-248 kPa)0.5 psia (3,5 kPa abs)43 psi (20,7 kPa)300 psi (2 070 kPa)-300 psi (-2 070 kPa)0.5 psia (3,5 kPa abs)520 psi (138 kPa)2000 psi (13 800 kPa)-2000 psi (-13 800 kPa)0.5 psia (3,5 kPa abs)TABLE 3. The Rosemount 3051NA Range and Sensor LimitsRange and Sensor LimitsRangeMinimum SpanUpper (URL)00.167 psia (8,6 mmHga)5 psia (260 mmHga)0 psia (0 mmHga)10.3 psia (2,07 kPa abs)30 psia (206,8 kPa abs)0 psia (0 kPa abs)21.5 psia (10,34 kPa abs)150 psia (1 034,2 kPa abs)0 psia (0 kPa abs)38 psia (55,16 kPa abs)800 psia (5 515,8 kPa abs)0 psia (0 kPa abs)440 psia (275,8 kPa abs)4000 psia (27 580 kPa abs)0 psia (0 kPa abs)Humidity Limits0-100% relative humidityVolumetric DisplacementLess than 0.005 in3 (0,08 cm3)Adjustable DampingTime constant on analog output is incrementallyadjustable from the minimum values statedabove to 25.6 seconds nominalFIGURE 4. Typical Smart Transmitter Response TimeTurn-on Time2 seconds maximumResponse TimeLower (LRL)Transmitter 4–20 mA Output vs. TimePressure Released20mATdTcDead Time (Td)Maximum dead time before analog output reactsto step change in input 0.1 seconds9.89mATd Dead TimeTc Time ConstantResponse Time Td Tc63.2% of TotalStep Change4mAUpdate RateTime20 times per second minimumMinimum Time Constant (Tc)At 70 F, with minimum damping settingRange Code012-5Minimum Time Constant (Tc)Including Dead Time (Td) 1.0 seconds 0.5 seconds 0.2 seconds7

Product Data Sheet00813-0100-4808, Rev FAJune 2008Rosemount 3051NMaximum Working Pressure(1) (2)The Rosemount 3051NDStatic pressure limitThe Rosemount 3051NG and 3051NAUpper range limitStatic Pressure LimitsOperates within specifications between static linepressures stated belowThe Rosemount 3051ND onlyRange Code00.5 psia to 750 psig (3,4 kPa abs to 5,2 MPa)12–5Static Pressure Limits0.5 psia to 2000 psig (3,4 kPa abs to 13,8 MPa)(1)0.5 psia to 3626 psig (3,4 kPa abs to 25 MPa)Overpressure LimitsTransmitters withstand following overpressurewithout damage:The Rosemount 3051ND/NGDrain/Vent Valves316 SSTProcess FlangesCF-8M (cast version of 316 SST)Process O-ringsGlass-filled TFE, Ethylene propylene (optional)Fill FluidSilicone oilFlange BoltsPlated carbon steel, per ASTM A449, Type 1(austenitic 316 SST per ASTM F593 for TheRosemount 3051N Range Code 0)Electronics HousingLow-copper aluminum withpolyester-polyurethane paint, or CF-3M (castversion of 316 SST)Non-wetted O-ringsEthylene propylene elastomerRange CodeOverpressure Limits0750 psig (5,2 MPa)12000 psig (13,8)2 – 5 (1)3626 psig (25 MPa)The Rosemount 3051NARange CodeOverpressure Limits060 psia (414 kPa)1120 psia (827 kPa)2300 psia (2 070 kPa)31600 psia (11 030 kPa)4 (1)6000 psia (41 370 kPa)Burst PressureMinimum burst pressure is 10,000 psig (69 MPa)Physical SpecificationsMaterials of ConstructionSensor Module HousingCF-3M (cast version of 316L SST perASTM-A743)Mounting BracketAISI 1010 steel with polyurethane paint (OptionCode B2), 304 SST per ASTM 554 (Option CodeB4), 316L SST (Option Code BS) or SST (OptionCode PM)Mounting Bolts (bracket-to-transmitter)Carbon steel, per ASTM A449 Type 1 (OptionCode B2, BS, PM), 316 SST per ASTM F-593(Option Code B4)Process Connections1/4-18Electrical Connections1Isolating Diaphragms/2-14 NPT conduit with screw terminals316L SST(1)(2)8NPTEPR process O-ring (Code B) is limited to 2000 psi maximum working pressure.See Table 1 for specified Seismic Maximum Working Pressure.

Product Data Sheet00813-0100-4808, Rev FAJune 2008Rosemount 3051NWeightTransmitter without options: 6.0 lb (2,7 kg) (seetable below for additional weights)Option Code DescriptionJStainless Steel HousingAdd:3.1 lb (1,4 kg)H2Traditional Flange2.4 lb (1,1 kg)M5LCD Meter for AluminumHousing0.5 lb (0,2 kg)M6LCD Meter for SST Housing1.25 lb (0,6 kg)B2Carbon Steel PanelMounting Bracket forTraditional Flange2.3 lb (1,0 kg)B4SST Mounting Bracket forCoplanar Flange1.0 lb (0,5 kg)BSUniversal SST Panel Bracketfor Traditional Flange3.4 lb (1,5 kg)PM2 inch pipe mount assemblyfor Traditional Flange6.8 lb (3.0 kg)9

Product Data Sheet00813-0100-4808, Rev FAJune 2008Rosemount 3051NFIGURE 5. The Rosemount 3051N Exploded View (with Coplanar process flange)LabelSpan and Zero Adjustments (Standard)Electronics HousingTerminal BlockO-ringCoverElectronics BoardNameplateSensor ModuleHousing Rotation Set Screw (180 Maximum Housing Rotation)Coplanar FlangeProcess O-ringDrain/vent ValveFlange AlignmentScrew (notpressure retaining)Flange Bolts10

Product Data Sheet00813-0100-4808, Rev FAJune 2008Rosemount 3051NDIMENSIONAL DRAWINGSFIGURE 6. Coplanar Flange Mounting Configurations with Optional Bracket (Code B4) for Panel Mounting2.8 (72)4.3 (110)5/16 Bolts for PanelMounting (not )(54.1)x 11/4Bolts forMounting toTransmitter3.4(85)4.8(120)NoteDimensions are nominal in inches (millimeters).FIGURE 7. The Rosemount 3051N Coplanar Flange Dimensional Drawing (Differential Pressure Transmitter Shown)5 (127)MeterCover(optional)1/2-14 NPT ConduitConnection(two places)4.3 (110)Label4.1 (105)0.75 (20)Clearancefor CoverRemoval0.75 (20)Clearancefor y7.1(181)NameplateHousing RotationSet ScrewDrain/VentValve6.4 (163)NoteDimensions are nominal in inches (millimeters).1/4-181/4-18NPT on Coplanar Flangefor Pressure ConnectionNPT on Coplanar Flangefor Pressure Connection11

Product Data Sheet00813-0100-4808, Rev FAJune 2008Rosemount 3051NFIGURE 8. Traditional Flange Mounting Configurations with Optional Brackets for Panel MountingOPTION CODE B2:TRADITIONAL FLANGE PANELMOUNTING BRACKET(PAINTED CARBON STEEL)OPTION CODE BS:TRADITIONAL FLANGEUNIVERSAL PANEL MOUNTINGBRACKET (STAINLESS STEEL)9.5 (241)9.4 (239)2.63(66.8)2.75(69.9)NameplateNameplate6.3 (181)4.93(125)5/16 Bolts for PanelMounting (Not Supplied)3/8 Bolts for PanelMounting (Not Supplied)2.81 (71)2.81 (71)2.81(71)NoteDimensions are nominal in inches (millimeters).12

Product Data Sheet00813-0100-4808, Rev FAJune 2008Rosemount 3051NFIGURE 9. Traditional Flange (Option Code H2) Dimensional Drawing5(127)1/2-14 NPT Conduit Connection(Two Places)4.3(110)0.75 (20) Clearance forCover Removal0.75 (20) Clearance forCover RemovalTerminal connectionsTransmitter CircuitryMeter Cover (optional)Center of gravity(aluminum housingwithout meter)Nameplate1.36 (36)1.63 (41.4)2.13(54.1)1/4-18 NPT on Traditional Flangefor Process Connection1.60(41)Label4.1(105)Housing Rotation Set Screw7.9(202)Center of gravity(aluminum housingwithout meter)1.36(35)1/4-18NoteDimensions are nominal ininches (millimeters).NPT onTraditional Flange forProcess 13

Product Data Sheet00813-0100-4808, Rev FAJune 2008Rosemount 3051NORDERING INFORMATIONTABLE 4. The Rosemount 3051N(1) Differential, Gage, and Absolute Pressure Transmitters— Not Applicable smitter Type (Select One)NDNGNADifferential Pressure TransmitterGage Pressure TransmitterAbsolute Pressure Transmitter ——— ——— NDNGNA — — Pressure Ranges (RANGE/MIN. SPAN)The Rosemount 3051NDThe Rosemount 3051NGThe Rosemount 3051NA–3 to 3 inH2O/0.1 inH2O(–747 to 747 Pa/25 Pa)–25 to 25 inH2O/0.5 inH2O(–6,22 to 6,22 kPa/0,12 kPa)–250 to 250 inH2O/2.5 inH2O(–62,2 to 62,2 kPa/0,6 kPa)–1000 to 1000 inH2O/10inH2O(–248 to 248 kPa/2,5 kPa)–300 to 300 psi/3 psi(–2 070 to 2 070 kPa/20,7kPa)–2000 to 2000 psi/20 psi(–13 800 to13 800 kPa/138 kPa)Not Applicable–250 to 250 inH2O/2.5 inH2O(–62,2 to 62,2 kPa/0,6 kPa)–393 to 1000inH2O/10in H2O(–101 to 249 kPa/2,5 kPa)0 to 5 psia/0.167 psia(0 to 259 mmHga/8,6 mmHga)0 to 30 psia/0.3 psia(0 to 207 kPa/2,1 kPa)0 to 150 psia/1.5 psia(0 to 1 034 kPa/10,34 kPa)0 to 800 psia/8 psia(0 to 5 516 kPa/55,16 kPa)–14.2 to 300 psi/3 psi(–101 to 2 070 kPa/20,7 kPa)0 to 4000 psia/40 psia(0 to 27 580 kPa/276 kPa) –14.2 to 2000 psig/20 psi(–101 to 13 800 kPa/138 kPa)Not Applicable — NDNGNA Not ApplicableNOTE: The Rosemount 3051NG lower range limit varies with atmospheric pressure.CodeAOutput4–20 mA with Digital Signal Based on HART ProtocolMaterials of ConstructionCode20Code2CodeAB(3)14ProcessFlange TypeFlange MaterialCoplanarSSTAlternate Flange – See Option Code H2Drain/VentSSTIsolating Diaphragm316L SSTProcess O-ringGlass-filled TFEEthylene Propylene (EPR)CodeFill Fluid1Silicone oil

Product Data Sheet00813-0100-4808, Rev FAJune 2008Rosemount 3051NTABLE 4. (continued) The Rosemount 3051N(1) Differential, Gage, and Absolute Pressure Transmitters— Not Applicable ApplicableCodeAJCodeH2CodeB2B4BSPMCodeM5M6Housing MaterialConduit Entry SizePolyurethane-covered AluminumSST½–14 NPT½–14 NPT Alternate Flange Options (Requires Materials of Construction Code 0)Traditional Flange, 316 SST, SST Drain/Vent, SST Flange AdapterMounting Bracket OptionsTraditional Flange Bracket for Panel Mounting, CS BoltsCoplanar Flange Bracket for Panel Mounting, all SSTUniversal Traditional Flange Bracket for Panel Mounting (SST), CS BoltsTraditional Flange Bracket for 2-in. pipe mounting, all SSTMeters (optional)LCD Meter for Aluminum Housing (Housing Code A only)LCD Meter for SST Housing (Housing Code J only)Typical Model Number:3051ND2A22A1AB4(1) All Rosemount 3051N transmitters are provided as standard with transient protection terminal block (T1) and cleaning for 1 PPM chloride.(2) The Rosemount 3051ND0 is available only with Process Flange Code 0 (Alternate Flange H2), O-ring Code A, and stainless steel processflange bolting.(3) EPR process o-ring is limited to 2000 psi maximum working pressure and 150 F (66 C) process temperature.NOTE:Mounting Bracket option code must be specified last in the Rosemount Model number even if optionalmeter is ordered.15

Product Data Sheet00813-0100-4808, Rev FAJune 2008Rosemount 3051NIMPORTANT NOTICE -- ERRATAModel 3051N Product Data Sheet 00813-0100-4808 Rev FA (6/2008)1AffectedPages82838No.Description of ChangeFlange Bolts and Mounting Bolts - Carbon steel, per ASTM A449 Type 1 or SAEJ429 Grade 5 Electronics Housing – Low copper aluminum with polyurethane paint, or CF-8M(cast version of 316 SST).Mounting Bracket – AISI 1010 steel or JIS G3131 SPHC P/O Steel withpolyurethane paint (Option Code B2).Effect.Date12/3/0711/18/081/24/1112/5/07The Emerson logo is a trade mark and service mark of Emerson Electric Co.Rosemount, the Rosemount logotype and SMART FAMILY are registered trademarks of Rosemount Inc.PlantWeb is a registered trademark of one of the Emerson Process Management group of companies.All other marks are the property of their respective owners.Coplanar is a trademark of Rosemount Inc.Standard Terms and Conditions of Sale can be found at www.rosemount.com/terms of saleEmerson Process ManagementRosemount Nuclear Instuments, Inc. satisfies all obligations coming fromlegislation to harmonize product requirements in the European Union.PRINTINU. S. A.EDRosemount Nuclear Instruments, Inc.8200 Market BoulevardChanhassen, MN 55317 USAT (952) 949-5210F (952) 949-5201 2008 Rosemount Inc. All rights reserved.www.rosemountnuclear.com 00813-XXXX-XXXXm 00813-0100-4808, Rev FA 00813-XXXX-XXXXm www.rosemountnuclear.com

The Rosemount 3051N is designed with a unique For differential and gauge pressure measurements, the Rosemount 3051N utilizes capacitance sensor technology similar to that of the Rosemount 1151, 1152, 1153, and 1154 transmitters (see Figure 1). Rosemount capacitance technology delivers the highest inherent performance, stability and reliability

Related Documents:

May 02, 2018 · D. Program Evaluation ͟The organization has provided a description of the framework for how each program will be evaluated. The framework should include all the elements below: ͟The evaluation methods are cost-effective for the organization ͟Quantitative and qualitative data is being collected (at Basics tier, data collection must have begun)

Silat is a combative art of self-defense and survival rooted from Matay archipelago. It was traced at thé early of Langkasuka Kingdom (2nd century CE) till thé reign of Melaka (Malaysia) Sultanate era (13th century). Silat has now evolved to become part of social culture and tradition with thé appearance of a fine physical and spiritual .

On an exceptional basis, Member States may request UNESCO to provide thé candidates with access to thé platform so they can complète thé form by themselves. Thèse requests must be addressed to esd rize unesco. or by 15 A ril 2021 UNESCO will provide thé nomineewith accessto thé platform via their émail address.

̶The leading indicator of employee engagement is based on the quality of the relationship between employee and supervisor Empower your managers! ̶Help them understand the impact on the organization ̶Share important changes, plan options, tasks, and deadlines ̶Provide key messages and talking points ̶Prepare them to answer employee questions

Dr. Sunita Bharatwal** Dr. Pawan Garga*** Abstract Customer satisfaction is derived from thè functionalities and values, a product or Service can provide. The current study aims to segregate thè dimensions of ordine Service quality and gather insights on its impact on web shopping. The trends of purchases have

Rosemount 3051 www.rosemount.com 00825-0100-4797G Quick Installation Guide 00825-0100-4797, Rev EA June 2010 Rosemount 3051 Pressure Transmitter with Profibus-PA Rosemount 3051CF Series Flowmeter Transmitter with Profibus-PA Start End Step 1: Mount the Transmitter Step 2: Consider Housing Rotation Step 3: Set the Jumpers

Rosemount 1151 Commissioning the Rosemount 1151 with a HART-Based Communicator Before putting the Rosemount 1151 Smart Pressure Transmitter into operation, commission the instrument using a HART-based communicator. To commission on the bench, connect a 17 to 45 V dc power supply and a current meter, such as the Rosemount 282 Validator.

NORTH LANARKSHIRE COUNCIL AGmA REPORT 1 1 I I 1 1 IFROM: QR8FSocWWoRK PERlQD Ollff109 - 16mm I I SoClAtWoRK DATE : 16 SEPTEMBER1896 Ref. : EMch I I 1 1. introduction This report compares actual expenditure and income against estimates both for the year to date and the prc@cted &-turn. Explanations are provided for the major &-turn variance.