TITLE 13. PUBLIC SAFETY CHAPTER 10. DEPARTMENT OF

2y ago
29 Views
2 Downloads
460.36 KB
31 Pages
Last View : 8d ago
Last Download : 3m ago
Upload by : Adalynn Cowell
Transcription

Arizona Administrative CodeTitle 13, Ch. 10Department of Public Safety - Alcohol TestingTITLE 13. PUBLIC SAFETYCHAPTER 10. DEPARTMENT OF PUBLIC SAFETYALCOHOL TESTING(Authority: A.R.S. §§ 28-1322 through 28-1326 and 41-1713)Editor’s Note: This Chapter, consisting of Article 1, Sections R13-10-101 through R13-10-109, Exhibits A through D, Exhibits E-1,through E-6, F-1 through F-5, G-1 through G-6, and H-1, through H-4, made by final rulemaking at 12 A.A.R. 1916, effective May 18,2006 (Supp. 06-2).Model 8000 (Option P)Exhibit G-6. Standard Operational and Quality Assurance Procedure Intoxilyzer Model 8000Exhibit H-1. Standard Operational Procedure Alco Sensor RBTAZExhibit H-2. Standard Calibration Check Procedure Alco SensorRBT AZExhibit H-3. Standard Quality Assurance Procedure Alco SensorRBT AZExhibit H-4. Standard Calibration Procedure Alco Sensor RBTAZARTICLE 1. DETERMINATION OF ALCOHOLCONCENTRATIONArticle 1, consisting of Sections R13-10-101 through R13-10109, Exhibits A through D, and Exhibits E-1 through E-6, F-1through F-5, G-1 through G-6, and H-1 through H-4, made by finalrulemaking at 12 A.A.R. 1916, effective 9:00 a.m., May 18, 2006(Supp. .R13-10-109.Exhibit A.Exhibit B.Exhibit C.Exhibit D.Exhibit E-1.Exhibit E-2.Exhibit E-3.Exhibit E-4.Exhibit E-5.Exhibit E-6.Exhibit F-1.Exhibit F-2.Exhibit F-3.Exhibit F-4.Exhibit F-5.Exhibit G-1.Exhibit G-2.Exhibit G-3.Exhibit G-4.Exhibit G-5.June 30, 2006DefinitionsAnalyst Methods; Approval of Additional MethodsBreath-testing DevicesTesting ProceduresPermits and CertificatesQualificationsApplication ProcessesExamination and Quality Assurance Requirementsfor AnalystsRevocation or Suspension of Permits; AppealsApplication for Blood Alcohol Analyst PermitApplication for Breath Alcohol Operator PermitApplication for Breath Alcohol Quality AssuranceSpecialist PermitApplication for Breath Testing InstructorStandard Operational Procedure Intoxilyzer Model5000Standard Calibration Check Procedure IntoxilyzerModel 5000*Standard Calibration Check Procedure IntoxilyzerModel 5000 (Option P)Standard Quality Assurance Procedure IntoxilyzerModel 5000*Standard Quality Assurance Procedure IntoxilyzerModel 5000 (Option P)Standard Operational Procedure Intoxilyzer Model5000 - Without Vapor Recirculation and WithoutKeyboardStandard Operational Procedure Intoxilyzer Model5000ENStandard Calibration Check Procedure IntoxilyzerModel 5000ENStandard Calibration Check Procedure IntoxilyzerModel 5000EN (Option P)Standard Quality Assurance Procedure IntoxilyzerModel 5000ENStandard Quality Assurance Procedure IntoxilyzerModel 5000EN (Option P)Standard Operational Procedure Intoxilyzer Model8000Standard Calibration Check Procedure IntoxilyzerModel 8000Standard Calibration Check Procedure IntoxilyzerModel 8000 (Option P)Standard Quality Assurance Procedure IntoxilyzerModel 8000Standard Quality Assurance Procedure IntoxilyzerARTICLE 1. DETERMINATION OF ALCOHOLCONCENTRATIONR13-10-101. DefinitionsIn this Article, unless the context otherwise requires:1. “Alcohol concentration” or “AC” means grams of alcoholper 100 milliliters of blood or grams of alcohol per 210liters of breath.2. “Analyst” means an individual who has been issued ananalyst permit by the Department to use approved methods to make alcohol concentration determinations fromblood or other bodily substances.3. “Analyst permit” means a document issued by theDepartment indicating the permit holder has been foundqualified to utilize an approved method in the determination of alcohol concentrations.4. “Analytical procedure” means a series of operations utilized by an analyst when employing an approved methodin the determination of alcohol concentration.5. “Calibration Check” means an operation utilizing a standard alcohol concentration solution to determine whethera device is accurately measuring alcohol concentrationsthat is performed as a Standard Calibration Check Procedure by a Quality Assurance Specialist at least every 31days or performed as Concurrent Calibration Check Procedures by an Operator within a successfully completedtest sequence bracketing a duplicate breath test.6. “Concurrent Calibration Check Procedure” means anoperation performed by an Operator, utilizing a standardalcohol concentration solution, within a successfullycompleted test sequence to determine whether a device isaccurately measuring alcohol concentration during aduplicate breath test.7. “Concurrent Quality Assurance Procedure” means operations performed by an Operator, including a ConcurrentCalibration Check Procedure and diagnostic checks,within a successfully completed test sequence to determine whether a device is accurately and properly measuring alcohol concentration during a duplicate breath test.8. “Deprivation period” means at least a 15-minute periodimmediately prior to a duplicate breath test during whichperiod the subject has not ingested any alcoholic beverages or other fluids, eaten, vomited, smoked or placedany foreign object in the mouth.Page 1Supp. 06-2

Title 13, Ch. 10Arizona Administrative CodeDepartment of Public Safety - Alcohol termination” means an analysis of a specimen ofblood, breath, or other bodily substance and expressingthe results of the analysis in terms of alcohol concentration.“Device” means a breath testing instrument.“Duplicate breath test” means two consecutive breathtests that immediately follow a deprivation period, agreewithin 0.020 AC of each other, and are conducted at leastfive and no more than 10 minutes apart.“Instructor” means a person approved by the Departmentto provide breath test training to prospective Operatorsand Quality Assurance Specialists on a specific approveddevice.“Method” means an analytical technique utilized by ananalyst or a device to make an alcohol concentrationdetermination (e.g. gas chromatography, infrared spectrophotometry, or specific fuel cell detection.)“Operator” means a person who has been issued an Operator permit from the Department to operate a specificapproved device for the purpose of determining an alcohol concentration from a specimen of breath and to perform the Concurrent Quality Assurance Procedures,Concurrent Calibration Check Procedures, and diagnosticchecks to determine whether a device is operating accurately and properly.“Operator Permit” means a document issued by theDepartment indicating that the permit holder has beenfound qualified to operate and perform the associatedQuality Assurance Procedures on a specific approveddevice.“Periodic Maintenance” means a Quality Assurance Procedure consisting of either of the following, which determines whether a device is operating accurately andproperly:a. Standard Calibration Check Procedure and StandardQuality Assurance Procedure, orb. Concurrent Calibration Check Procedures and Concurrent Quality Assurance Procedures performedwithin a successfully completed test sequence bracketing a duplicate breath test.“Preliminary breath test” means a pre-arrest breath test.“Preliminary breath tester” or “PBT” means anyapproved device used prior to an arrest for the purpose ofobtaining a determination of alcohol concentration from aspecimen of breath and includes any device included onthe National Highway Traffic Safety Administration’sConforming Products List of Evidential Breath Measurement Devices as incorporated by reference in R13-10103(F).“Procedure” means a series of operations used by anOperator or a Quality Assurance Specialist when employing an approved device in the determination of alcoholconcentration or performing associated quality assurancetesting.“Quality Assurance Procedure” means Periodic Maintenance consisting of either of the following, which determines whether a device is operating accurately andproperly:a. Standard Calibration Check Procedure and StandardQuality Assurance Procedure, orb. Concurrent Calibration Check Procedures and Concurrent Quality Assurance Procedures performedwithin a successfully completed test sequence bracketing a duplicate breath test.“Quality Assurance Specialist” means a person who hasbeen issued a Quality Assurance Specialist permit fromSupp. 06-222.23.24.25.the Department to perform the Standard CalibrationCheck Procedure and the Standard Quality AssuranceProcedure to determine the accurate and proper operationof a specific approved device.“Quality Assurance Specialist permit” means a documentissued by the Department indicating that the permitholder has been found qualified to perform the StandardCalibration Check Procedure and the Standard QualityAssurance Procedure on a specific approved device.“Standard Calibration Check Procedure” means operations performed by a Quality Assurance Specialist, atleast every 31 days, to determine whether a device isaccurately measuring alcohol concentration.“Standard Operational Procedure” means operations performed by an Operator for the purpose of determining analcohol concentration from a specimen of breath.“Standard Quality Assurance Procedure” means operations performed by a Quality Assurance Specialist, atleast every 90 days.Historical NoteNew Section made by final rulemaking at 12 A.A.R.1916, effective 9:00 a.m., May 18, 2006 (Supp. 06-2).R13-10-102. Analyst Methods; Approval of Additional MethodsA. An analyst shall use one of the following methods to analyzeblood or other bodily substances to determine a person’s alcohol concentration:1. Gas chromatography, or2. Another method that has been approved by the Directorunder the procedure in subsections (B) and (C).B. An applicant for an analyst permit may submit, with the permitapplication, a request that the Director approve a method otherthan a method approved under subsection (A)(1) or (2).C. For a method to be approved by the Director, the method’saccuracy and reproducibility shall comply with the followingstandards:1. The test results of samples with a standard alcohol concentration shall agree with the established value withinthe limits of 0.01 grams per 100 milliliters of blood or 10 percent, whichever is greater.2. The accuracy and precision shall be determined on thebasis of ten measurements at four alcohol concentrationsbetween 0.020 and 0.350 grams per 100 milliliters ofblood, to include at least one value 0.100 and one value 0.250.Historical NoteNew Section made by final rulemaking at 12 A.A.R.1916, effective 9:00 a.m., May 18, 2006 (Supp. 06-2).R13-10-103. Breath-testing DevicesA. The Director may approve devices used to determine alcoholconcentration from breath after the Department successfullytests a typical model of the device for compliance with thestandards in subsection (B).B. A device shall meet the following standards of performance:1. Breath specimens tested shall be alveolar in composition.2. The device shall be capable of analysis of a solution ofknown alcohol concentration with an accuracy limit of asystematic error of no more than 0.005 grams per 210liters of breath or 5 percent, whichever is greater, and aprecision limit of an average standard deviation of nomore than 0.0042 grams per 210 liters of breath. Theaccuracy and precision of the device being evaluatedshall be determined on the basis of 10 consecutive mea-Page 2June 30, 2006

Arizona Administrative CodeTitle 13, Ch. 10Department of Public Safety - Alcohol TestingC.D.E.surements at 4 alcohol vapor concentrations that arebetween 0.020 and 0.350 grams per 210 liters of breath,to include at least one value 0.100 and one value 0.250.3. The device shall be capable of testing a breath samplethat results in alcohol concentrations of less than 0.01grams per 210 liters of breath when alcohol-free subjectsare tested.The Department, upon specific findings that a device, method,or breath test procedure is inaccurate, unreliable, or is an unacceptable test for determining alcohol concentration or that itsuse has been discontinued in the state, shall disapprove in writing further use of the device, method, or procedure.The methods approved by the Director for use by a device todetermine alcohol concentration are infrared spectrophotometry and specific fuel cell detection.The following devices are approved by the Director:Device/ModelManufacturerIntoxilyzer Model 5000CMI, Inc.with or without Vapor Recirculation and with or without KeyboardIntoxilyzer Model 5000ENCMI, Inc.Intoxilyzer Model 8000CMI, Inc.RBT AZ(Alco Sensor AZ/RBT AZ)Intoximeter, Inc.F.Products included on the National Highway Traffic SafetyAdministration’s Conforming Products List of EvidentialBreath Measurement Devices set forth in 69 FR 42237-42239(July 14, 2004) are approved by the Director as preliminarybreath testers to determine alcohol concentration. This document is incorporated by reference and does not include anylater amendments or editions. A copy of this document isavailable from the Department and may be obtained from theNational Highway Traffic Safety Administration’s web site(www.nhtsa.gov) or by contacting the U.S. Government Printing Office, 732 North Capitol Street NW, Washington, DC20401.G. Devices listed in subsection (E) may be used to administer preliminary breath tests.H. Except when a device is used as a PBT or for other non-evidential testing purposes, an Operator permit and StandardOperational Procedure are required for the operation ofdevices listed in subsection (E).I. In addition to the devices approved in subsection (E), theDirector may approve, in writing, a device and related Standard Operational and Quality Assurance Procedures after thedevice has been successfully tested for compliance with thestandards in subsection (B) for use prior to and pending thedevice being added to subsection (E). The approval shallexpire three years after its effective date unless subsection (E)is amended to include the approved device.J. In addition to devices approved as preliminary breath testers insubsection (F), the Director may approve in writing as a PBT anew device placed on subsequent National Highway TrafficSafety Administration’s Conforming Products Lists of Eviden-June 30, 2006tial Breath Measurement Devices for use pending the newConforming Products List being added to subsection (F).Historical NoteNew Section made by final rulemaking at 12 A.A.R.1916, effective 9:00 a.m., May 18, 2006 (Supp. 06-2).R13-10-104. Testing ProceduresA. Law enforcement agencies or individuals acting independentlyof law enforcement agencies who conduct alcohol concentration determinations by means of devices shall utilize a qualityassurance program that is conducted by Quality AssuranceSpecialists or Operators and generate records of periodicmaintenance. This quality assurance program shall include:1. Criteria for ensuring the accurate and proper operation ofdevices by the regular performance of Calibration Checksand Quality Assurance Procedures as referenced in subsections (A)(2) and (A)(3);2. Calibration Checks of devices that are performed within31 days of each other as Standard Calibration Check Procedures or during a test sequence bracketing a duplicatebreath test as Concurrent Calibration Check Proceduresand recorded according to the requirements of the appropriate Quality Assurance Procedures set forth in ExhibitsE-2, E-3, F-2, F-3, G-2, G-3, G-6 and H-2 or as approvedby the Director according to R13-10-103(I). Thesechecks shall indicate that the device is capable of determining the value of a standard alcohol concentrationsolution with an accuracy limit of 0.01 grams per 210liters of breath or 10 percent, whichever is greater;3. Quality Assurance Procedure checks of devices that areperformed within 90 days of each other as Standard Quality Assurance Procedures or during a test sequence bracketing a duplicate breath test as Concurrent QualityAssurance Procedures, and recorded according to therequirements of the appropriate Quality Assurance Procedures set forth in Exhibits E-4, E-5, F-4, F-5, G-4, G-5,G-6, H-3 and H-4 or as approved by the Director according to R13-10-103(I). These checks shall indicate that thedevice is capable of proper operation and is functioningas required by the Quality Assurance Procedures for thedevice;4. Standard alcohol concentration solutions, either liquid orgas, that are National Institute of Standards and Technology (NIST) traceable; and5. Records of Calibration Checks, Quality Assurance Procedures and maintenance or repairs for each device in use.B. An Operator shall utilize the Standard Operational Procedureapproved by the Department for the device being operated inperforming tests for the determination of alcohol concentration, as contained in Exhibits E-1, E-6, F-1, G-1, G-6 and H-1or as approved by the Director according to R13-10-103(I).C. Duplicate breath tests shall be administered at intervals of notless than five minutes nor more than 10 minutes. The results ofboth tests shall be within 0.020 alcohol concentration of eachother. If the second test is not within 0.020 alcohol concentration of the first test, additional tests shall be administered untilthe results of two consecutive tests are within 0.020 alcoholconcentration.Historical NoteNew Section made by final rulemaking at 12 A.A.R.1916, effective 9:00 a.m., May 18, 2006 (Supp. 06-2).R13-10-105. Permits and CertificatesA. The Department shall issue Analyst permits to qualified applicants, in accordance with R13-10-106(A), who have satisfac-Page 3Supp. 06-2

Title 13, Ch. 10Arizona Administrative CodeDepartment of Public Safety - Alcohol TestingB.C.D.E.torily demonstrated through proficiency testing as specified inR13-10-108(A) their proficiency in conducting an alcoholconcentration determination by one or more of the methodslisted in R13-10-102. The Analyst permit shall:1. State the method of alcohol concentration determinationthe permit holder is approved to utilize and the type ofspecimen the permit holder is approved to analyze (bloodor other bodily substances); and2. Be valid for one year.An Analyst shall employ, in testing for alcohol concentrationin matters arising under A.R.S. Title 28, Chapter 4, Article 3,the same analytical procedures as those employed by the analyst for proficiency testing.The Department shall issue two categories of device permits.1. Operator permits shall be issued to applicants who qualify under R13-10-106(B) or (E). This permit authorizesoperation and performance of associated Quality Assurance Procedures, including Concurrent Calibration CheckProcedures and Concurrent Quality Assurance Procedures, performed within a successfully completed testsequence bracketing a duplicate breath test on the devicespecified on the permit. Operator permits issued after theinitial effective date of this Section shall be valid for fiveyears from the date of issue. Permits issued to Operatorsbefore the initial effective date of this Section shallremain in effect and be valid for five years after the initialeffective date of this Section.2. Quality Assurance Specialist permits shall be issued toapplicants who hold a valid Operator permit and whoqualify as a Quality Assurance Specialist under R13-10106(C) or (E). This Quality Assurance Specialist permitauthorizes the holder to perform Quality Assurance Procedures, including Standard Calibration Check Procedures and Standard Quality Assurance Procedures, on thedevice specified on the permit. Quality Assurance Specialist permits issued after the initial effective date of thisSection shall be valid for five years from the date ofissue. Permits issued to Quality Assurance Specialistsbefore the initial effective date of this Section shallremain in effect and be valid for five years after the initialeffective date of this Section.3. Operator and Quality Assurance Specialist permits maybe renewed by application as required by R13-10-107 andsuccessful completion of a recertification courseapproved by the Department.4. The Department shall issue duplicate (replacement) permits upon request and upon verification of the qualifications set forth in R13-10-106.Law enforcement agencies shall supply the Department, uponrequest, with a list of current Operator and Quality AssuranceSpecialist permit holders and shall update the list as requiredby the Department, but no more frequently than annually.The Department shall issue Instructor certificates to qualifiedapplicants who hold valid Operator and Quality AssuranceSpecialist permits and who qualify as an Instructor under R1310-106(D) or (E). The Instructor certificate authorizes theholder to provide breath test training to prospective Operatorsand Quality Assurance Specialists on a specific approveddevice. Instructor certificates issued after the initial effectivedate of this Section shall be valid for five years from the dateof issue. Instructor certificates issued before the initial effective date of this Section shall remain in effect and be valid forfive years from the initial effective date of this Section.Instructor certificates may be renewed by application asrequired by R13-10-107 and successful completion of a recertification examination approved by the Department.Supp. 06-2Historical NoteNew Section made by final rulemaking at 12 A.A.R.1916, effective 9:00 a.m., May 18, 2006 (Supp. 06-2).R13-10-106. QualificationsA. To qualify for an Analyst permit, a person shall hold a degreefrom a college or university accredited by a regional accrediting body recognized by the United States Department of Education and have earned 15 or more semester credits, or theequivalent, of chemistry, including three or more credits oforganic chemistry.B. To qualify for an Operator permit, a person shall:1. Be employed by a law enforcement agency or laboratorythat has access to a device for the person’s use as set forthin R13-10-103; and2. Complete a course in the determination of alcohol concentration approved by the Department with a score of 80percent or better. The Department shall approve coursestaught by an Instructor if they contain the following:a. Instruction on the effects of alcohol on the humanbody;b. Instruction on and demonstration of the operationalprinciples of the selected device, which shall includea functional description and detailed operationaldescription of the method;c. Instruction on the legal aspects of breath tests ingeneral and on the particular method to beemployed;d. Concurrent Calibration Check Procedures (whenapplicable to the device) approved by the Department;e. Concurrent Quality Assurance Procedures (whenapplicable to the device) approved by the Department;f. Applicant participation with the appropriate deviceutilizing reference standards, testing of subjects, orother methods that will indicate the actual responseof the device; andg. Written and practical examination of the applicantfor the purpose of determining the person’s understanding of the course material and proficiency inoperating the device.C. To qualify for a Quality Assurance Specialist permit, a personshall possess a valid Operator permit to operate the approveddevice and complete a course of training approved by theDepartment with a score of 80 percent or better. The Department shall approve courses taught by an Instructor if they contain the following:1. Review of the theory of breath testing and the operationof the particular testing device;2. Standard Calibration Check Procedures approved by theDepartment;3. Standard Quality Assurance Procedures approved by theDepartment;4. Applicant participation with the appropriate device utilizing reference standards, testing of subjects, or other methods that will indicate the actual response of the device;and5. Written and practical examination of the applicant for thepurpose of determining the person’s understanding of thecourse material and proficiency in operating the device.D. To qualify as an Instructor, a person shall hold valid Operatorand Quality Assurance Specialist permits on the device forwhich instruction is given. In addition, except as provided insubsection (E), all applicants shall complete a comprehensiveinstructor examination approved and administered by theDepartment with a score of 90 percent or better. The Depart-Page 4June 30, 2006

Arizona Administrative CodeTitle 13, Ch. 10Department of Public Safety - Alcohol TestingE.ment shall approve instructor examinations that include thefollowing:1. The theory of breath testing and the operation of the specific device, and2. Procedures for testing instrument accuracy and properoperation in accordance with Calibration Checks andQuality Assurance Procedures approved by the Department.If a device is newly approved and no Operator and QualityAssurance Specialist permits have been issued for the device,a person may qualify to be an Operator, Quality AssuranceSpecialist, and Instructor for the specific device by completinga Department-administered, manufacturer-endorsed, instructortraining course and a comprehensive examination with a scoreof 90 percent or better. The Instructor training course shallinclude the following:1. Review of the theory of breath testing,2. Instruction on the operation of the device, and3. Procedures for testing instrument accuracy and properoperation in accordance with Calibration Checks andQuality Assurance Procedures approved by the Department.B.Historical NoteNew Section made by final rulemaking at 12 A.A.R.1916, effective 9:00 a.m., May 18, 2006 (Supp. 06-2).R13-10-107. Application ProcessesA. An applicant for an initial Analyst permit or the renewal of anexisting Analyst permit shall complete the form shown asExhibit A and submit it to the Department. An application forrenewal of an Analyst permit shall be submitted no later than30 days prior to the date the current permit expires. If theapplicant makes a written or verbal request and shows goodcause, the Department shall extend this deadline.B. An applicant for an initial Operator permit or the renewal of anexisting Operator permit shall complete the form shown asExhibit B and submitted to the Department. An application forrenewal of an Operator permit shall be submitted no later than30 days prior to the date the current permit expires. If theapplicant makes a written or verbal request and shows goodcause, the Department shall extend this deadline.C. An applicant for an initial Quality Assurance Specialist permitor the renewal of an existing Quality Assurance Specialist permit shall complete the form shown as Exhibit C and submittedto the Department. An application for renewal of a QualityAssurance Specialist permit shall be submitted no later than 30days prior to the date the current permit expires. If the applicant makes a written or verbal request and shows good cause,the Department shall extend this deadline.D. An applicant for an initial Instructor approval or the renewal ofan existing Instructor approval shall complete the form shownas Exhibit D and submitted to the Department. An applicationfor renewal of an Instructor shall be submitted no later than 30days prior to the date the current certificate expires. If theapplicant makes a written or verbal request and shows goodcause, the Department shall extend this deadline.Historical NoteNew Section made by final rulemaking at 12 A.A.R.1916, effective 9:00 a.m., May 18, 2006 (Supp. 06-2).R13-10-108. Examination and Quality Assurance Requirements for AnalystsA. The Department shall require an Analyst permit applicant tosuccessfully demonstrate the applicant’s proficiency in makingalcohol concentration determinations from test specimens inJune 30, 2006C.accordance with subsection (B). The applicant shall be examined only on the methods that relate to the type of determination for which the applicant desires a permit.An applicant shall, before receiving an initial Analyst permitor renewal of an existing Analyst permit, participate in andsuccessfully complete proficiency testing administered by theDepartment. An applicant shall successfully analyze samplesby testing at least three suitable reference standards or controlsamples with a known alcohol concentration in the range of0.00 to 0.40 grams per 100 milliliters of blood and having theresults agree with the established value within the limits of 0.01 grams per 100 milliliters of blood or 10 percent, whichever is greater. Proficiency testing shall be administered by theDepartment as follows:1. An applicant shall correctly analyze all proficiency samples in the set provided by the Department.2. When returning the results of analyses to the Department,the applicant shall attach an affidavit attesting that theapplicant analyzed the proficiency samples without helpor input from any other person.3. An applicant failing to correctly analyze all proficiencysamples in the set will be provided an opportunity to successfully analyze a second set of samples.4. The Department shall deny the application of an applicantwho declines or fails to correctly analyze the second setof proficiency samples and shall not issue a permit.5. An applicant who fails to successfully analyze the secondset of proficiency samples and whose application isdenied may reapply for an analyst’s permit beginning 90days from the date of denial.An analyst who conducts alcohol concentration determinationsshall implement and maintain a quality assurance program.This program shall be designed to ensure the validity of testresults by providing for:1. Chain of custody,2. Quality control,3. Analytical procedures,4. Documentation of test results, and5. Participation in proficiency testing.Historical NoteNew Section made by final rulemaking at 12 A.A.R.1916, effective 9:00 a.m., May 18, 2006 (Supp. 06-2).R13-10-109. Revocation or Suspension of Permits; AppealsA. The Department may suspend or revoke a permit for any of thefollowing reasons:1. A false statement on the permit holder’s application,2. The neglect or refusal to examine and report the results ofsample specimens given the Analyst

“Preliminary breath test” means a pre-arrest breath test. 18. “Preliminary breath tester” or “PBT” means any approved device used prior to an arrest for the purpose of . Conforming Products List of Evidential Breath Measure-ment Devices as incorporated by reference in R13-10-103(F). 19. “Procedure” means a series of operations .

Related Documents:

Part One: Heir of Ash Chapter 1 Chapter 2 Chapter 3 Chapter 4 Chapter 5 Chapter 6 Chapter 7 Chapter 8 Chapter 9 Chapter 10 Chapter 11 Chapter 12 Chapter 13 Chapter 14 Chapter 15 Chapter 16 Chapter 17 Chapter 18 Chapter 19 Chapter 20 Chapter 21 Chapter 22 Chapter 23 Chapter 24 Chapter 25 Chapter 26 Chapter 27 Chapter 28 Chapter 29 Chapter 30 .

TO KILL A MOCKINGBIRD. Contents Dedication Epigraph Part One Chapter 1 Chapter 2 Chapter 3 Chapter 4 Chapter 5 Chapter 6 Chapter 7 Chapter 8 Chapter 9 Chapter 10 Chapter 11 Part Two Chapter 12 Chapter 13 Chapter 14 Chapter 15 Chapter 16 Chapter 17 Chapter 18. Chapter 19 Chapter 20 Chapter 21 Chapter 22 Chapter 23 Chapter 24 Chapter 25 Chapter 26

DEDICATION PART ONE Chapter 1 Chapter 2 Chapter 3 Chapter 4 Chapter 5 Chapter 6 Chapter 7 Chapter 8 Chapter 9 Chapter 10 Chapter 11 PART TWO Chapter 12 Chapter 13 Chapter 14 Chapter 15 Chapter 16 Chapter 17 Chapter 18 Chapter 19 Chapter 20 Chapter 21 Chapter 22 Chapter 23 .

HUNTER. Special thanks to Kate Cary. Contents Cover Title Page Prologue Chapter 1 Chapter 2 Chapter 3 Chapter 4 Chapter 5 Chapter 6 Chapter 7 Chapter 8 Chapter 9 Chapter 10 Chapter 11 Chapter 12 Chapter 13 Chapter 14 Chapter 15 Chapter 16 Chapter 17 Chapter

About the husband’s secret. Dedication Epigraph Pandora Monday Chapter One Chapter Two Chapter Three Chapter Four Chapter Five Tuesday Chapter Six Chapter Seven. Chapter Eight Chapter Nine Chapter Ten Chapter Eleven Chapter Twelve Chapter Thirteen Chapter Fourteen Chapter Fifteen Chapter Sixteen Chapter Seventeen Chapter Eighteen

18.4 35 18.5 35 I Solutions to Applying the Concepts Questions II Answers to End-of-chapter Conceptual Questions Chapter 1 37 Chapter 2 38 Chapter 3 39 Chapter 4 40 Chapter 5 43 Chapter 6 45 Chapter 7 46 Chapter 8 47 Chapter 9 50 Chapter 10 52 Chapter 11 55 Chapter 12 56 Chapter 13 57 Chapter 14 61 Chapter 15 62 Chapter 16 63 Chapter 17 65 .

Chapter 3 Chapter 4 Chapter 5 Chapter 6 Chapter 7 Chapter 8 Chapter 9 Chapter 10 Chapter 11 Chapter 12 Chapter 13 Chapter 14 Chapter 15 Chapter 16 Chapter 17 Chapter 18 Chapter 19 Chapter 20 . Within was a room as familiar to her as her home back in Oparium. A large desk was situated i

The Hunger Games Book 2 Suzanne Collins Table of Contents PART 1 – THE SPARK Chapter 1 Chapter 2 Chapter 3 Chapter 4 Chapter 5 Chapter 6 Chapter 7 Chapter 8. Chapter 9 PART 2 – THE QUELL Chapter 10 Chapter 11 Chapter 12 Chapter 13 Chapter 14 Chapter 15 Chapter 16 Chapter 17 Chapt