Image guided therapyHemo SystemMonitor - measure - recordPhilips Interventional Hemodynamicsystem with IntelliVue X3 Technical specifications

IntroductionKey advantages of the Hemo systemImproving productivity and patient outcome is vital for healthcare facilities to meet thegrowing demand for cath lab procedures. To further simplify cath lab workflow, Philipsintroduces the Interventional Hemodynamic system (Hemo system).Hemo system brings advanced hemodynamic measurements into the interventionallab to support clinical decision making. This system is integrated with the marketleading Philips IntelliVue X3 patient monitor mounted at the table side and a hemoworkstation in the control room to perform hemodynamic analysis. Furthermore thesystem can be operated from the table side in the exam room via the Philips AzurionTouch screen module.Integrated IntelliVue X3 patient monitorControl of Philips Hemo on Touch Screen ModuleHemodynamic analyses performed in the control room can be shown inthe exam room to help the users to stay focused on the task at hand.Integrated iFR Spot and Scout pullbackincoporated Azurion’s intuitive workflow approachConfidently used by all staff members with minimal trainingContent Key advantages of the Hemo system Technical specification of the patient monitorIntelliVue X3, extension and Dock Patient cables, sensors and accessories Workstation specification23

Technical specifications of the patient monitorIntelliVue X3, measurement extensions and DockThis section describes IntelliVue X3 patient monitor, extension and Dock as a signalacquisition device that provides input to the Hemo system during interventionalprocedures where all interaction is managed by the Hemo workstation (with XperInformation Management System software and Hemodynamic Application).The IntelliVue X3, extension, Dock with the mounting bracket allow for flexiblepositioning at the table side.IntelliVue X3, extension and Dock:Weight 2,6 kgSize 120 x 180 x 190 mmThe device front panel has patient cable/accessoryconnectors for invasive blood pressure, ECG, cardiacoutput, surface temperature, SpO2, etCO2, and noninvasive blood pressure.12IntelliVue X31 Temperature2 Pressure connector for 2 pressures33 SpO2 (option from Philips FAST , Masimo or Nellcor)IntelliVue Dock1. MSL connector for monitor connection2. Flexible Sync Output connector, to provide signal to othermedical devices3. AC power connector4. LAN connector for connection to Hemo workstation12The IntelliVue Dock provides an external power supply forthe monitor and extension when connected to mains powervia AC power connector. IntelliVue Dock has Flexible SyncOutput connector to output and synchronize signals to othermedical devices. 1/4 stereo phone jack connector: 2 channels (tip and ring) 2 programmable analogue outputs: Analog ECG output (configurable at tip or ring) 2V/mV(default, 4V/mV; 1V/mV; 0.5V/mV; 0.25 V/mV selectable)) Analog pressure output (configurable at tip or ring):1V/100 mmHG; voltage swing 4V Digital ECG pulse output (configurable at ring)34Possibility to split the signal if you want to use ECG lead I/II for different devices as the input. On X3 you can choosebetween QRS, primary lead or one of the pressure channels.Signals can be split if necessary.Care and CleaningThe X3 measurement extensions and Dock deploychemically‑resistant surface materials, designed to resistdeterioration from cleaning and disinfection agents. Evenagainst very aggressive disinfectants, the X3’s the housingmaterials have been tested, and found to resist deteriorationabout 60 times longer than the housing material used forpreceeding products. See the list of tested agents in themonitor’s Instructions for Use.Mounting bracket4 ECG45 Noninvasive blood pressure512Extension1 EtCO2 (optional Mainstream/Sidestreamor Microstream connector)2 Cardiac Output Thermodilution343 Temperature4Power Consumption: 20W when on IntelliVue DockOperating Voltage36 to 60 V dc floatingOperating Temperature Range:0 to 35 C (32 to 95 F) when charging the batteryOperating Relative Humidity Range:15% to 95% humidity, non-condensingIngress ProtectionIntelliVue X3, extension and Dock IP32 when mountedhorizontallyDefinition of IP 32: Protected against ingress of water when the water is dripping vertically and monitor is tilted up to 15 ,and ingress of solid foreign objects 2,5 mm in diameter or larger4 Pressure connector for 2 pressures5 Connection to IntelliVue X35General – IntelliVue X3, extension and DockInternal Battery (453564526811)The battery is required for the operation of the monitor. The battery lifetime is 3 years from manufacturing date or 500 charge/discharge cycles.5

Fits your clinical workflowChoose the set-up that works best with your lab6BasicPerformanceBasic monitoring and hemodynamic analysis capabilities for main streamInterventional Cardiac proceduresComprehensive monitoring and hemodynamic analysis capabilitiesfor wide range of Interventional Cardiac proceduresClinical and workflow functionality: Non-invasive blood pressure Body surface temperature 12 lead ECG 2 invasive blood pressures Calculated Cardiac Output Fick Respiration rate SpO2 Philips FAST, Nellcor (Covidien) or Masimo Capture and store hemodynamic waveforms and ECG’s Full disclosure (record and store all waveforms data forpost case review and analysis) End case report (hemodynamic measurements andcalculations) Print waveforms and hemodynamic analysis Store patient data Visualization of ST values Aortic Regurgitation (AR) IndexClinical and workflow functionality: Non-invasive blood pressure Body surface temperature 12 lead ECG 4 invasive blood pressures Thermodilusion Cardiac Output (and calculated Fick) Respiration rate SpO2 Philips FAST, Nellcor (Covidien) or Masimo Capture and store hemodynamic waveforms and ECG’s Full disclosure (record and store all waveforms data forpost case review and analysis) End case report (hemodynamic measurements andcalculations) Print waveforms and hemodynamic analysis Store patient data Visualization of ST values Aortic Regurgitation (AR) IndexClinical Options: Integrated Philips iFR/FFR Integrated FFR (compatible with Abbott)Work Flow Options: Ability to operate in patient area TrolleyIntegrated with Philips image guided therapy system Hemo control from Touch Screen Module Patient demographics Connected to FlexVision or Monitor Ceiling SuspensionClinical Options: EtCO2: Respironics LowFlo CO2 Covidien Microstream CO2 Integrated Philips iFR/FFR Integrated FFR (compatible with Abbott)Work Flow Options: Ability to operate in patient area TrolleyIntegrated with Philips image guided therapy system Hemo control from Touch Screen Module Patient demographics Connected to FlexVision or Monitor Ceiling Suspension7

Measurement Specifications - ECGRespirationIEC 60601‑2‑27:2011ECG SpecificationsRespiration Performance SpecificationsHRRespiration RateLow15–300 bpm maximum delay:10 seconds according to IEC60601‑2‑27Range Adult/pedi: Neo:AccuracyRange-20–20 mmAccuracy 0.5 mm or 15% whichever isgreaterResolution0.1 mmRangeRange Adult/pedi: Neo: At 0–120 rpm 1 rpm At 120–170 rpm 2 rpmAdjustment 20 rpm: 1 rpm steps 20 rpm: 5 rpm stepsResolution1 rpmDelayBandwidth0.3–2.5 Hz (-6 dB)Noise 25 mΩ (rms) referred to the input F or limits from 0 to 20 rpm:maximum 4 seconds For limits above 20 rpm: maximum14 secondsST NumericQT-HR NumericRange - Adult15–150 bpmRange - Pedi/neo15–180 bpmResolution1 bpmRespiration Alarm SpecificationsApnea AlarmHighRange10–40 secondsAdjustment5 second stepsRange Adult/pedi: Neo:Adjustment 20 rpm: 1 rpm steps 20 rpm: 5 rpm stepsDelayMaximum 14 secondsSinus and SV Rhythm RangesBrady Adult: Pedi: Neo:15–59 bpm15–79 bpm15–89 bpmNormal Adult: Pedi: Neo:60–100 bpm80–160 bpm90–180 bpmTachy Adult: Pedi: Neo: 100 bpm 160 bpm 180 bpm0–120 rpm0–170 rpm10–100 rpm30–150 rpm0–95 rpm0–145 rpmBandwidthDiagnostic modeAdult/neo/pedi: 0.05–150 HzMonitoring mode Adult: Neo/pedi:Notch filter50/60 Hz0.5–40 Hz0.5–55 HzDifferential Input Impedance 2 MΩ RA-LL leads (Resp) 5 MΩ at all other leads (at 10 Hz including patient cable)Common Mode Rejection Ratio Diagnostic Mode: 86 dB (with a 51 kΩ/47 nF imbalance) Filter Mode: 106 dB (with a 51 kΩ/47 nF imbalance)Electrode Offset Potential Tolerance 500 mVAuxiliary Current (Leads off Detection) Active Electrode: 100 nA Reference Electrode: 900 nAInput Signal Range 5 mV89

Noninvasive Blood Pressure (NBP)Invasive Blood Pressure TemperatureComplies with: IEC 80601‑2‑30:2010 A1:2013 EN 80601‑2‑30:2010 A1:2015Supports up to two pressure transducersvia one connector and one Y-cable.Complies with: IEC 60601‑2‑34:2011 EN 60601‑2‑34:2014NBP Performance SpecificationsNBP Alarm SpecificationsInvasive Pressure Performance SpecificationsTemperature Performance eratureRange Adult: 30–270 mmHg (4–36 kPa) Pedi: 30–180 mmHg (4–24 kPa) Neo: 30–130 mmHg (4–17 kPa)DiastolicRangeRangeRangeAdjustment Adult: 10–245 mmHg (1.5–32 kPa) Pedi: 10–150 mmHg (1.5–20 kPa) Neo: 10–100 mmHg (1.5–13 kPa)MeanRange Adult: 20–255 mmHg (2.5–34 kPa) Pedi: 20–160 mmHg (2.5–21 kPa) Neo: 20–120 mmHg (2.5–16 kPa)Max. Std.Deviation8 mmHg (1.1 kPa)Max. MeanError 5 mmHg ( 0.7 m time Adult: Neo: Stat:30 seconds25 seconds20 seconds Adult/pedi: 180 seconds Neo:90 secondsPediTypical fornormal adultcuff 10 secondsTypical forneonatal cuff 2 secondsInitial CuffInflationPressure Adult: Pedi: Neo:Maximum CuffPressure Adult/pedi: 300 mmHg Neo:150 mmHgNeo165 15 mmHg130 15 mmHg100 15 mmHg 1 0–30 mmHg (1.5–4 kPa):2 mmHg (0.5 kPa) 30 mmHg ( 4 kPa): 5 mmHg (1 kPa) Adult: Pedi: Neo:10–245 mmHg (1.5–32 kPa)10–150 mmHg (1.5–20 kPa)10–100 mmHg (1.5–13 kPa) 1 0–30 mmHg (1.5–4 kPa):2 mmHg (0.5 kPa) 30 mmHg ( 4 kPa): 5 mmHg (1 kPa) Adult: Pedi: Neo:20–255 mmHg (2.5–34 kPa)20–160 mmHg (2.5–21 kPa)20–120 mmHg (2.5–16 kPa) 1 0–30 mmHg (1.5–4 kPa):2 mmHg (0.5 kPa) 30 mmHg ( 4 kPa): 5 mmHg (1 kPa)NBPOverpressureSettings(Not UserAdjustable)AdultCuff InflationTime30–270 mmHg (4–36 kPa)30–180 mmHg (4–24 kPa)30–130 mmHg (4–17 kPa)MeanTypical at HR 60 bpmAuto/Manual Adult: Pedi: Neo:DiastolicAccuracy10Complies with: ISO 80601‑2‑56:2009 EN ISO 80601‑2‑56:2012 300 mmHg (40 kPa) 2 seconds 300 mmHg (40 kPa) 2 seconds-40–360 mmHgInput SensitivitySensitivity5 μV/V/mmHg (37.5 μV/V/kPa)Adjustment range 10%Transducers (Compliant with ANSI/AAMI BP22)Range (absolute)-1–45 C (30–113 F)Range (differential) 46 C ( 115 F)Resolution0.1 C (0.1 F)Accuracy 0.1 C ( 0.2 F) 10 secondsLoad impedance200–2000 Ω (resistive)Average TimeConstantOutput impedance 3000 Ω (resistive)Temperature Alarm SpecificationsFrequencyResponseDC to 12 Hz or 40 HzTemperature High/Low AlarmsZero AdjustmentRange 200 mmHg ( 26 kPa)Accuracy 1 mmHg ( 0.1 kPa)Drift 0.1 mmHg/ C (0.013 kPa/ C)Range-1–45 C (30–113 F)Adjustment 1–30 C (30–86 F),0.5 C (1.0 F) steps 30–45 C (86–113 F),0.1 C (0.2 F) stepsGain AccuracyAccuracy 1%Drift 0.05%/ CNon-linearity andHysteresisError of 0.4% FS (@CAL 200 mmHg)Overall Accuracy(IncludingTransducer) 4% of reading or 4 mmHg( 0.5 kPa), whichever is greaterInvasive Pressure Alarm SpecificationsPressureRange-40–360 mmHg (-5.0–48 kPa)Adjustment -40–50 mmHg (-5–4 kPa):2 mmHg (0.5 kPa) 50 mmHg ( 4 kPa):5 mmHg (1 kPa)DelayMaximum 12 seconds 150 mmHg (20 kPa) 2 secondsTransducer and pressure cable should be ordered viaestablished supplier chain in the hospital.We suggest to order:Edwards TrueWare PX series Single adult transducer and thepressure cable from the front end to the transducer.11

Cardiac OutputThermodilutionSpecificationsCardiac OutputBlood Temperature RangeComplies with: ISO 80601‑2‑55:2011 EN ISO 80601‑2‑55:2011Mainstream CO2 Performance Specifications17–43 C (62.6–109.5 F)Blood Temperature Accuracy(Excluding Probe)0.1 C (0.2 F)Injectate Temperature Range-1–30 C (30.2–86.0 F)Injectate TemperatureAccuracy (Excluding Probe)0.1 C (0.2 F)CO2Range0–150 mmHg (0–20 kPa)AccuracyAfter two minutes warm-up: For values between 0 and 40mmHg (0 and 5,3 kPa): 2.0 mmHg( 0.29 kPa). For values from 41–70 mmHg(5.4–9.3 kPa): 5% of reading. For values from 71–100 mmHg(9.4–13.3 kPa) 8% of reading. For values from 101–150 mmHg(13.4–20 kPa): 10 % of reading thespecifications are valid for standardgas mixtures, balance air, fullyhydrated at 35 C, Pabs 760 mmHg(101.3 kPa), flow rate 2 l/minCardiac Output (Right Heart)C.O. Range0.1–20 l/minInstrument Specification (Measured Electronically)C.O. Instrument Accuracy 3% or 0.1 l/minC.O. Repeatability 2% or 0.1 l/minResolutionStability:Short-term DriftLong-term Drift Numeric: 1.0 mmHg (0.1 kPa) Wave: 0.1 mmHg (0.01 kPa) 0.8 mmHg (0.11 kPa) over four hoursAccuracy specification is maintainedover a 120‑hour periodWarm-up TimeTwo minutes with CO2 transducerattached for full accuracyspecificationResponse Time 60 ms (with adult or infant reusableor disposable adapter)Sidestream CO2 Performance SpecificationsCO212Microstream CO2(867041)End tidal CO2 (867040)Range0–150 mmHg (0–20 kPa)AccuracyAfter two minutes warm-up: For values between 0 and 40mmHg (0 and 5,3 kPa): 2.0 mmHg( 0.29 kPa). For values from 41–70 mmHg(5.4–9.3 kPa): 5% of reading. For values from 71–100 mmHg(9.4–13.3 kPa) 8% of reading. For values from 101–150 mmHg(13.4–20 kPa): 10% of reading.At respiration rates above 80 rpm,all ranges are 12% of reading. Thespecifications are valid for gasmixtures of CO2, balance N2, drygas at 760 mmHg (101.3 kPa) withinspecified operating temperaturerange.ResolutionStability:Short-term DriftLong-term Drift Numeric: 1.0 mmHg (0.1 kPa) Wave: 0.1 mmHg (0.01 kPa) 0.8 mmHg (0.11 kPa) over four hoursAccuracy specification is maintainedover a 120‑hour periodWarm-up TimeTwo minutes with CO2 sensorattached for full accuracyspecificationSample Flow Rate50 10 ml/minuteTotal SystemResponse Time3 secondsCO2 Alarm SpecificationsEtCO2 HighRange20–95 mmHg (2–13 kPa)Adjustment1 mmHg (0.1 kPa) stepsDelay 14 secondsEtCO2 LowRange10–90 mmHg (1–12 kPa)Adjustment1 mmHg (0.1 kPa) stepsDelay 14 secondsPerformance SpecificationsMeasurementRange0–150 mmHg (0–20.0 kPa) or 20%CO2, whichever is lowerAccuracy(After 5 minuteswarmup)These specifi cations are valid for: 21% O2 and balance N2 Up to 35 C ambient temperature Up to 60 rpm for adults and100 rpm for neonates Values between 0 and 40 mmHg(0 and 5,3 kPa): 2.2 mmHg 0.30 kPa) Values above 40 mmHg (5.3 kPa): (5% 0.08% per mmHg above40 mmHg) of readingResolution Numeric: 1 mmHg (0.1 kPa) Wave: 0.1 mmHg (0.02 kPa)Warm-up TimeUp to 5 minutes, with an accuracyof 4 mmHg or 12% of reading,whichever is greaterSample Flow Rate50 15 ml/min - 7.5 ml/minRise TimeStep Response10–90% 190 ms for neonatal patients(measured with M1923A FilterLine HSet Infant/Neonatal) 240 ms for adult patients (measuredwith M1921A FilterLine H Set Adult/Pediatric)Gas Sampling Delay TimeSampling delay time from an input step change at theairway adapter until the measured signal changes by 10%of the input step.2 m Sample LinesMaximum 3 seconds4 m Sample LinesMaximum 6 secondsTotal SystemResponse TimeSum of Gas Sampling Delay Time andRise TimeEndtidal CO2 (et CO2) Alarm LimitsRange EtCO2 low: 10–90 mmHg (1–12 kPa) EtCO2 high: 20–95 mmHg (2–13 kPa)Adjustments1 mmHg (0.1 kPa) stepsCO2 Alarm Delay 14 seconds (excluding Total SystemResponse Time)13

Philips FAST SpO2(867030 SP1)Nellcor OxiMax SpO2(867030 SP6)Masimo rainbow SETSpO2 (867030 SP5)Complies with: ISO 80601‑2‑61:2011 EN ISO 80601‑2‑61:2011Complies with: ISO 80601‑2‑61:2011 EN ISO 80601‑2‑61:2011Complies with: ISO 80601‑2‑61:2011 EN ISO 80601‑2‑61:2011Philips FAST SpO2 Performance SpecificationsPulse Oximetry Performance SpecificationsMeasurementAccuracyRange and ResolutionSpO2SpO2, no range 60–80 3%, Adult/pedi/infant 70–100 2%, Adult/pedi/infant, 3% NeoResolution1%AccuracyFor information about accuracy seePhilips 867030 Technical Data SheetRange0–100%2% ion0.1, 0.01 for small valuesLow perfusionaccuracyaPulse14SpO2PulseRange30–300 bpmPulseAccuracy 2% or 1 bpm, whichever is greaterRange25–300 bpmResolution1 bpmResolution1 bpmAccuracy 3 bpm (20–250 bpm)aMeasurement Range and ResolutionRange25-240 bpmResolution1 bpmFor more information on Philips IntelliVue X3(867030), measurement extensions (867039,867040, 867041), and Dock (867043), refer to theseparate technical data sheets. pecification applies to the performance of the device.SReading accuracy in the presence of low perfusion(detected IR pulse modulation amplitude 0.03–1.5%) wasvalidated using signals supplied by a patient simulator.SpO2 and pulse rate values were varied across themonitoring range over a range of weak signal conditionsand compared to the known true saturation and pulse rateof the input signals.15

Patient cables, sensors and Accessoriesfor Philips Hemo systemMainstream CO2 sensorM2501AAirway Adapter Adult/Pediatric Reusable Use with ET tube 4mmM2513AAirway Adapter Infant Reusable Use with ET tube 4mm Deadspace 1ccM2516ASingle patient use adult airway adapterM2533ASingle patient use infant airway adapterM2536ACategoryNamePhilips ID old P/NGas cylinder regulatorM2505AECGCBL 5 5: 10 lead ECG trunk cable, AAMI/IEC, 2.7mM1949AGAS Verification gasM2506ACBL 5 lead ECG trunk, AAMI/IEC, 2.7mM1668ACBL 5 leadset, grabber, chest, AAMI/ICUM1976ACBL 5 leadset, grabber, limb, AAMI/ICUSpO2NIBP (NBP)COTemp16RespironicsetCO2 Capnostat5 MainstreamintubatedRespironics Lo-FloetCO2 Sidestreamnon-intubatedSidestream CO2 sensorM2741ACO2 nasal cannula - adultM2744AM1968ACO2 nasal cannula - pediatricM2745ACBL 5 leadset, grabber, chest, IEC/ICUM1978ACO2 nasal cannula - infantM2746ACBL 5 leadset, grabber, limb IEC/ICUM1971ACO2 /O2 nasal cannula - adultM2750ADisposable radiolucent leads IEC989803156271CO2 /O2 nasal cannula - pediatricM2751ADisposable radiolucent leads AAMI989803156261CO2 oral-nasal cannula - adultM2756ACBL SpO2 9-pin D-sub adapter 1.1 m (8-pin)M1943ACO2 oral-nasal cannula - pediatricM2757AReusable clip adult SpO2 sensorM1196ACO2 /O2 oral-nasal cannula - adultM2760ADisposable adult/pedi SpO2 sensorM1131ACO2/O2 nasal cannula - pediatricM2761AInfant disposable SpO2 sensorM1132AAirway Adapter Set - ET 4.0 mmM2768ANeo / infant / adult disposable SpO2 sensorM1133AAirway Adapter Set H - ET 4.0 mmM2772AWristbandM1627AAirway Adapter Set H - ET 4.0 mmM2773AReusable NIBP Comfort Cuff assortmentM1579AAirway Adapter Set ET 4.0 mm989803144531NIBP HoseM1599BEtCO2 /O2 Nasal Cannula - Infant/ Neonate989803144471Reusable NIBP Comfort Cuff Adult Long Kit - 3 sizesM1579XLStraight Sample Line, non-humidifiedM2776AReusable NIBP Comfort Cuff, infantM1571AStraight Sample Line H, humidifiedM2777AReusable NIBP Comfort Cuff, pediatricM1572AStraight airway adapter. Single patient useM1612AReusable NIBP Comfort Cuff, small adultM1573AReusable Nafion sample tube13901AReusable NIBP Comfort Cuff, small adult XLM1573XLElbow airway adapter. Single patient use13902AReusable NIBP Comfort Cuff, adultM1574ABacteria filter (0.45 micron). Single patient use13904AReusable NIBP Comfort Cuff, adult XLM1574XL13905AReusable NIBP Comfort Cuff, large adultM1575AHybrid Nafion polyethylene sample tube. Single patient use Nafion length:6 ft (1.8m) Polyethylene length: 9 ft (2.7m)Reusable NIBP Comfort Cuff, large adult XLM1575XLFilterLine Set, Adult/Pedi, IntubatedM1920AFilterLine H Set, Adult/Pedi, IntubatedM1921AFilterLine H Set, Infant/Neonatal, IntubatedM1923AFilterLine Set Long, Adult/Pediatric989803160241FilterLine H Set Long, Adult/Pediatric989803160251FilterLine H Set Long, Infant/Neonatal989803160261VitaLine H Set, Adult/Pediatric989803159571VitaLine H Set, Infant/Neonatal989803159581Smart CapnoLine O2 PediatricM2520ASmart CapnoLine O2 Adult/IntermediateM2522ASmart CapnoLine PediatricM2524AReusable NIBP Comfort Cuff, thighM1576AReusable NIBP Comfort Cuff assortment, smaller sizes(infant, pediatric, small adult, adult)M1577AReusable NIBP Comfort Cuff assortment, larger sizes(small adult, adult, large adult, thigh)M1578AIce bath temperature probe23002ACardiac output cable, 4.8 mM1643ACO-Set injectate temp probe, 0.5 m23001ASkin surface temperature probe21078AGobi reusable skin probe989803203581CO2 Microstreamintubated andnon-intubated17

Workstation specificationCO2 Microstreamintubated andnon-intubatedOtherSmart CapnoLine Adult/IntermediateM2526ASmart CapnoLine O2 Pediatric Long989803160271Smart CapnoLine O2 Plus Adult Long989803160281Smart CapnoLine Plus Adult Long989803160301CapnoLine H O2 AdultM4680ACapnoLine H O2 PediatricM4681ANIV AdultM4686ANIV PediatricM4687ACapnoLine H AdultM4689ACapnoLine H Infant/NeonatalM4691ASmart CapnoLine H O2 Adult989803177951Smart CapnoLine H O2 Adult Long989803177961Smart CapnoLine H O2 Pediatric989803177971Smart CapnoLine H O2 Pediatric Long989803177981CapnoLine H O2 Infant/Neonatal989803178001CapnoLine H Infant/Neonatal Long989803178011Nasal FilterLine Infant/Neonatal989803178021Smart CapnoLine Guard989803178031Smart CapnoLine Guard O2989803178041Smart CapnoLine Guard O2 Long989803178051Hook and Loop Strap989803178071Nasal FilterLine O2 Adult989803179101Nasal FilterLine O2 Adult Long989803179111Nasal FilterLine O2 Pediatric989803179121Calibration RegulatorM2267ATRADE COMPLIANT: FILTERLINE, ADULT/PED989803182911TRADE COMPLIANT:FILTERLINE H, ADULT/PED989803182921TRADE COMPLIANT:FILTERLINE H, INFANT/NEO989803182931TRADE COMPLIANT:VITALINEH SET, ADULT/PED989803182941TRADE COMPLIANT:SMART CAPNOLINE O2 PED989803182951TRADE COMPLIANT:SMART CAPNOLINE O2 ADULT989803182961TRADE COMPLIANT:SMARTCAPNOLINEH O2 ADULT989803182971TRADE E COMPLIANT:SMART CAPNOLINE GUARD O2989803182991TRADE COMPLIANT:SMARTCAPNOLINEGUARDO2 4M989803183001TRADE COMPLIANT:CAPNOLINE H O2 ADULT989803183071Patient cable organizerM2281AControl room workstatioin:8 GB RAMIntel Core i5 8500 6C CPUHDD 500GBHDD 1TB (Standalone only)NVIDIA Quadro P400 2GBExam room workstation:4 GB RAMIntel Core i5-6300UHDD 500GBDisplays:Control room display: 1920x1200 24” or 1920x1080 24”Exam Room displays 1280x1024 19” or 1920x1080 27”Size of displayed ECG waves:To ensure that the size of ECG waves on the attacheddisplays are within 10% of the size indicated by the PhilipsHemodynamic Application, the pixel density of the displayshould be 89 /- 10% dpi (i.e. the dot size should be0.285 /- 10% mm).For more information about s complete set of the supplies and accessories,refer to the separate “Philips IntelliVue Accessories” Technical Data Sheet1819

2020 Koninklijke Philips N.V. All rights reserved. Specifications are subject to changewithout notice. Trademarks are the property of Koninklijke Philips N.V.or their respective owners.4522 991 57361 * Mar 2020How to reach usPlease visit www.philips.comhealthcare@philips.com

5 Connection to IntelliVue X3 IntelliVue Dock 1. MSL connector for monitor connection 2. Flexible Sync Output connector, to provide signal to other medical devices 3. AC power connector 4. LAN connector for connection to Hemo workstation The IntelliVue Dock provides an external power supply for the monitor and extension when connected to mains .

Related Documents:

The Philips wordmark – optional The . optional. Philips wordmark is only for Philips employees. External agencies using @philips.com email accounts must not add the Philips wordmark to their email signature. Never . replace the Philips wordmark with another logo or Philips

with Philips UV-C lamp systems. As a result, illnesses that are easily transmitted via the air are minimized and the overall air quality is improved. Philips TUV PL-S page 10-11 Philips TUV PL-L page 24-25 Philips TUV TL Mini page 12-13 Philips TUV T8 page 26-27 Philips TUV T5 page 20-21 Philips drivers page 28-29 8 9

INTERVENTIONAL RADIOLOGY: MAPS 99–102 253 Context Interventional radiology (IR) refers to a range of techniques that use radiological image guidance to target therapy, and interventional radiologists are trained in both radiology and interventional therap

is an personal video about me servicing a Philips N4520 tape recorder.I use the dutch Service Manual and I try to explain what I am doingas much as Philips N4520, Service Manual, Repair Schematics Philips N4520. Descrição: (Description). Service Repair. Datasheet. Dicas Reparação Tips. Service Manual - Philips N4520 - Tape recorder Philips N4520 - Tape recorder .

For more information contact us at lighting.india@philips.com or call at 1800 102 2929 www.lighting.philips.co.in North Philips India Ltd, 9th Floor, DLF - 9B, DLF Cyber City DLF Phase 3, Gurgaon 122002 Tel :- 91 124 460 6000 lighting.india@philips.com Philips India Ltd, 8th floor, DLF - 9B, DLF Cyber

Fig. 2 Interventional oncology procedure volumes by year. Fig. 3 Spinal intervention procedures by year. Seminars in Interventional Radiology Vol. 36 No. 1/2019 Evolution in Private Practice Interventional Radiology Morrison 19 This document was downloaded for personal us

Philips Senior Living Solutions Royal Philips Electronics is a 34 billion global leader in healthcare, consumer lifestyle and lighting products, technologies and solutions. Philips delivers products to market through the brand promise of "Sense and Simplicity". Philips Senior Living Solutions is part of the Philips Healthcare group.

Philips Lifeline. Philips Lifeline Service Welcome to Philips Lifeline Thank you for choosing the Philips Lifeline Medical Alert Service. These Instructions for Use will provide you with information about your equipment and the Lifeline Medical Alert Service. Please read the manual carefully, and if you have questions, call Lifeline at any time.