Agilent 7890B Gas Chromatograph

2y ago
27 Views
2 Downloads
3.60 MB
200 Pages
Last View : 24d ago
Last Download : 3m ago
Upload by : Sasha Niles
Transcription

Agilent 7890B Gas ChromatographOperation ManualAgilent Technologies

Notices Agilent Technologies, Inc. 2015WarrantyNo part of this manual may be reproducedin any form or by any means (includingelectronic storage and retrieval or translation into a foreign language) without prioragreement and written consent fromAgilent Technologies, Inc. as governed byUnited States and international copyrightlaws.The material contained in this document is provided “as is,” and is subject to being changed, without notice,in future editions. Further, to themaximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied,with regard to this manual and anyinformation contained herein, including but not limited to the impliedwarranties of merchantability and fitness for a particular purpose. Agilentshall not be liable for errors or forincidental or consequential damagesin connection with the furnishing,use, or performance of this documentor of any information containedherein. Should Agilent and the userhave a separate written agreementwith warranty terms covering thematerial in this document that conflict with these terms, the warrantyterms in the separate agreementshall control.Manual Part NumberG3430-90054EditionThird edition, October 2015 Second edition, December 2013 First edition, January 2013Printed in USA or ChinaAgilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808-1610 USAAgilent Technologies, Inc. 412 Ying Lun Road Waigoaqiao Freed Trade Zone Shanghai 200131 P.R.ChinaSafety NoticesCAUTIONA CAUTION notice denotes a hazard.It calls attention to an operatingprocedure, practice, or the like that, ifnot correctly performed or adhered to,could result in damage to the productor loss of important data. Do notproceed beyond a CAUTION noticeuntil the indicated conditions are fullyunderstood and met.WA RNINGA WARNING notice denotes a hazard.It calls attention to an operatingprocedure, practice, or the like that, ifnot correctly performed or adheredto, could result in personal injury ordeath. Do not proceed beyond aWARNING notice until the indicatedconditions are fully understood andmet.

Contents1IntroductionChromatography Using a GC10The Operating Panel11The display11Status lights 12Beeping instrument12Blinking setpoint 13The keypad142Operating BasicsOverview16Instrument Control17To Start Up the GC18To Shut Down the GC for Less Than a Week19To Shut Down the GC for More Than a Week20Correcting Problems321Keypad OperationThe Run Keys24The GC Component KeysThe Status KeyThe Info Key252627The General Data Entry KeysThe Supporting Keys2829Method Storage and Automation Keys30Keypad Functionality When the GC Is Controlled by an Agilent DataSystem 31The Service Mode Key32About GC Status33Status board33Alert tones 33Error conditions34Blinking setpoint 34About Logs35Maintenance logOperation Manual353

4Methods and SequencesWhat Is a Method?38What Is Saved in a Method?38What Happens When You Load a Method?39Creating Methods 40To load a method41To store a method41What Is a Sequence?Creating Sequences4343Automating Data Analysis, Method Development, and SequenceDevelopment47Recoverable Errors548Running a Method or a Sequence from the KeypadRunning Methods from the Keypad50To manually inject a sample with a syringe and start a runTo run a method to process a single ALS sample 50To abort a method5050Running Sequences from the Keypad51To start running a sequence51To pause a running sequence51To resume a paused sequence52To stop a running sequence52To resume a stopped sequence52Aborting a sequence 52To resume an aborted sequence526Chromatographic CheckoutAbout Chromatographic Checkout54To Prepare for Chromatographic CheckoutTo Check FID Performance57To Check TCD Performance62To Check NPD Performance67To Check uECD Performance5572To Check FPD Performance (Sample 5188-5953)Preparation 77Phosphorus performance 78Sulfur performance 82477Operation Manual

To Check FPD Performance (Sample 5188-5245, Japan)Preparation 84Phosphorus performance 85Sulfur performance 89To Check FPD Performance (Sample 5188-5953)Preparation 91Phosphorus performance 92Sulfur performance 9691To Check FPD Performance (Sample 5188-5245, Japan)Preparation 98Phosphorus performance 99Sulfur performance 103798Resource ConservationResource Conservation106Sleep Methods 106Wake and Condition Methods108To Set the GC to Conserve ResourcesTo Edit an Instrument Schedule110113To Create or Edit a Sleep, Wake, or Condition MethodTo Put the GC to Sleep NowTo Wake the GC Now884114115116Early Maintenance FeedbackEarly Maintenance Feedback (EMF)Counter types118Thresholds118Default Thresholds120Available Counters121118To Enable or Change a Limit for an EMF CounterTo Disable an EMF CounterTo Reset an EMF Counter124125126EMF Counters for Autosamplers127Counters for 7693A and 7650 ALS with EMF-enabled firmwareCounters for ALS with earlier firmware 127EMF Counters for MS InstrumentsOperation Manual1271285

9Intelligent Instrument FeaturesSystem-Level Communications130GC/MS Systems131Venting the MS131MS Shutdown events 131GC Pressure Shutdown events 132To Set Up a Vent Method 132To Manually Prepare the GC for Venting the MSTo Manually Exit the MS Vent State 133To Use the GC When the MS is Shut Down134To Enable or Disable MS or HS CommunicationsSystem EMF Counters133135136Instrument Schedule for an Intelligent Instrument System10137ConfigurationAbout Configuration 140Assigning GC resources to a device140Setting configuration properties141General Topics142To Unlock the GC ConfigurationIgnore Ready 142Information displays143Unconfigured:143142Oven144To configure the oven for cryogenic coolingFront Inlet/Back Inlet147To configure the Gas type147To configure the PTV or COC coolantTo configure the MMI coolant 149147Column #151To view a summary of column connectionsComposite Columns160To configure composite columnsLTM Columns162LTM Series II column modulesCryo Trap154161162163About Heaters6145165Operation Manual

Front Detector/Back Detector/Aux Detector/Aux Detector 2To configure the makeup/reference gas167Lit offset167To configure the FPD heaters167To ignore the FID or FPD ignitor 168Analog out 1/Analog out 2Fast peaks169169Valve Box170To assign a GC power source to a valve box heaterThermal Aux174Pressure aux 1,2,3/Pressure aux 4,5,6/Pressure aux 7,8,9Time170171PCM A/PCM B/PCM CStatus167176177178Valve #179Front injector/Back injectorSample tray (7683 ALS)Instrument180182183Using the Optional Barcode Reader184Barcode reader power184Installing the barcode reader185To scan configuration data using the G3494B RS-232 barcodereader185To scan configuration data using the G3494A USB barcode readerTo delete scanned configuration data186To uninstall the RS-232 barcode reader18711OptionsAbout Options190Calibration190To zero a specific flow or pressure sensorColumn calibration 192Hydrogen sensor calibration196Communication198Configuring the IP address for the GCKeyboard and DisplayOperation Manual1861921981997

8Operation Manual

Agilent 7890BGas ChromatographOperation Manual1IntroductionChromatography Using a GCThe Operating Panel 1110This document provides an overview of the individualcomponents that make up the Agilent 7890B GasChromatograph (GC).Agilent Technologies9

1IntroductionChromatography Using a GCChromatography is the separation of a mixture of compoundsinto individual components.There are three major steps involved with separating andidentifying components of a mixture using a GC. They are:1 Injecting a sample into the GC. (This takes place at the inlet.)2 Separating the sample into individual components. (Thistakes place inside the column in the oven.)3 Detecting what compounds were in the sample. (This is donein the detector.)During this process, status messages from the GC are displayed,and user changes to parameter settings can be made throughthe operating panel or data system.InletOvenDetectorOperating panelOperating panelRefer to the Advanced Operating Manual and the GettingStarted manual for more details.10Operation Manual

Introduction1The Operating PanelThe operating panel consists of the display, status lights, andkeypad. See “Keypad Operation” and the Advanced OperationManual, along with the complete suite of documentationincluded on the Agilent GC and GC/MS User Manuals & ToolsDVDs that are included with your GC shipment for moredetailed information.DisplayShows status, setpoints, currentactivity, and messages.Status lightsLEDs indicate general status, runstate, program state, externalcontrol, and maintenance due.KeyboardUse to enter settings and program theGC.The displayThe display shows details of what is currently happening in theGC and allows you to make changes to parameters as necessary.TemperatureInitial timeRate 1OVEN150.0ActualOperation Manual150.01.000 20.000Setpoint11

1IntroductionThe cursor, , shows the current active line. Use the scroll keysto select a different line in the display and to viewadditional lines in the display.A blinking asterisk (*) prompts you to press [Enter] to store avalue or [Clear] to abort the entry. You cannot perform any othertask until this is done.Status lightsThe status lights provide a basic look at what is currentlyhappening inside the Agilent 7890B GC.A lit LED on the status board indicates: The current progress of a run (Pre Run, Post Run, and Run). Items that may require attention (Rate, Not Ready, ServiceDue, and Run Log). The GC is controlled by an Agilent data system (Remote). The GC is programmed for events to occur at specifiedtimes (Clock Table). The GC is in gas saver mode (Gas Saver).Beeping instrumentA single beep means that a problem exists, but the problem willnot prevent the GC from executing the run. The GC will emitone beep and display a message. The GC can start the run andthe warning disappears when a run starts.A series of warning beeps sound if the GC encounters a moreserious problem. The GC starts with one beep. The longer theproblem persists, the more the GC beeps. For example, a seriesof beeps sound if the front inlet gas flow cannot reach setpoint.The message Front inlet flow shutdown is briefly displayed. Theflow shuts down after 2 minutes. Press [Off/No] to stop thebeeping.12Operation Manual

Introduction1A continuous beep sounds if a hydrogen flow is shut down or athermal shutdown occurs. Press [Clear] to stop the beep.Fault messages indicate hardware problems that require userintervention. Depending on the type of error, the GC will beeponce or not at all.Blinking setpointIf a gas flow, multiposition valve, or the oven is shut down bythe system, Off or On/Off will blink on the appropriate line of thecomponents parameter listing.Operation Manual13

1IntroductionThe keypadAll of the parameters required to operate the Agilent 7890B GCcan be entered through the GC’s keypad. Normally, however,most of these parameters are controlled using an attached datasystem, such as Agilent’s OpenLAB CDS or MassHuntersoftware.When an Agilent data system is controlling your 7890B GC, it ispossible for the data system to disable editing of the GC’scurrent method from the keypad.The Run keysGC Component keysService Mode keyStatus keySupporting keysGeneral data entrykeysMethod Storage andAutomation keys14Operation Manual

Agilent 7890B Gas ChromatographOperation Manual2Operating BasicsOverview 16Instrument Control 17To Start Up the GC 18To Shut Down the GC for Less Than a Week 19To Shut Down the GC for More Than a Week 20Correcting Problems 21This section describes a few basic tasks that an operatorperforms when using the Agilent 7890B GC.Agilent Technologies15

2Operating BasicsOverviewOperating the GC involves the following tasks: Setting up the GC hardware for an analytical method. Starting up the GC. See “To Start Up the GC”. Preparing the automatic liquid sampler. Install themethod-defined syringe; configure solvent and waste bottleusage and syringe size; and prepare and load solvent, waste,and sample vials. For the 7693A ALS, see its Installation, Operation, andMaintenance manual. For the 7683 ALS, see manual Operating the 7683B ALSon a 7890 Series GC. Loading the analytical method or sequence into the GCcontrol system. See the Agilent data system documentation. For standalone GC operation see “To load a method” and“To load a stored sequence”. Running the method or sequence. See the Agilent data system documentation. For standalone GC operation, see “To manually inject asample with a syringe and start a run”, “To run a methodto process a single ALS sample”, and “To start running asequence”. Monitoring sample runs from the GC control panel or theAgilent data system program. See “About GC Status” or theAgilent data system documentation. Shutting down the GC. See “To Shut Down the GC for LessThan a Week” or “To Shut Down the GC for More Than aWeek”.16Operation Manual

2Operating BasicsInstrument ControlThe Agilent 7890B GC is typically controlled by an attacheddata system such as Agilent OpenLAB CDS. Alternately, the GCcan be controlled entirely from its keypad, with output databeing sent to an attached integrator for report generation.Agilent data system users – Please refer to the online helpincluded in the Agilent data system for details on how to load,run, or create methods and sequences using the data system.Standalone GC users – If you are running your GC without anattached data system, see the following for details on loadingmethods and sequences from the keypad: “To load a method” “To load a stored sequence”For details on running methods and sequences from the keypadsee: “To manually inject a sample with a syringe and start a run” “To run a method to process a single ALS sample” “To start running a sequence”For details on how to create methods and sequences using theGC keypad, see “Methods and Sequences”.Operation Manual17

2Operating BasicsTo Start Up the GCSuccessful operation begins with a properly installed andmaintained GC. The utility requirements for gases, powersupply, venting of hazardous chemicals, and requiredoperational clearances around the GC are detailed in theAgilent GC, GC/MS, and ALS Site Preparation Guide.1 Check gas source pressures. For required pressures, see theAgilent GC, GC/MS, and ALS Site Preparation Guide.2 Turn on the carrier and detector gases at their sources andopen the local shutoff valves.3 Turn on the cryo coolant at its source, if used.4 Turn on the GC power. Wait for Power on successful to display.5 Install the column.6 Check that the column fittings are leak free. See theTroubleshooting manual.7 Load the analytical method. See “To load a method”.8 Wait for the detector(s) to stabilize before acquiring data.The time required for the detector to reach a stable conditiondepends on whether the detector was turned off or itstemperature was reduced while the detector remainedpowered.Table 118Detector stabilization timesDetector typeStabilization time startingStabilization time startingfrom a reduced temperature from detector off (hours)(hours)FID24TCD24uECD418 to 24FPD212NPD418 to 24Operation Manual

2Operating BasicsTo Shut Down the GC for Less Than a Week1 Wait for the current run to finish.2 If the active method has been modified, save the changes.WA RNINGNever leave flammable gas flows on if the GC will beunmonitored. If a leak develops, the gas could create a fire orexplosion hazard.3 Turn off all gases, except the carrier gas, at their sources.(Leave the carrier gas on to protect the column fromatmospheric contamination.)4 If you are using cryogenic cooling, turn off the cryo coolantat the gas source.5 Reduce detector, inlet, and column temperatures to between150 and 200 C. If desired, the detector can be turned off.See the following table to determine if it is advantageous toshut down the detector for a short time period. The timerequired to return the detector to a stable condition is afactor. See Table 1.Operation Manual19

2Operating BasicsTo Shut Down the GC for More Than a WeekSee Maintaining Your GC manual for procedures for installingcolumns, consumables, and so on.1 Load a GC maintenance method and wait for the GC tobecome ready. For more information about creatingmaintenance methods, see the Maintaining Your GC manual.(If a maintenance method is not available, set all heatedzones to 40 C.)2 Turn off the main power switch.3 Shut off all gas valves at the gas source.4 If you are using cryogenic cooling, shut off the cryo coolantvalve at the source.WA RNINGBe careful! The oven, inlet, and/or detector may be hot enough tocause burns. If they are hot, wear heat-resistant gloves to protectyour hands.5 When the GC is cool, remove the column from the oven andcap both ends to keep out contaminants.6 Cap the inlet and detector column fittings and all GCexternal fittings.20Operation Manual

Operating Basics2Correcting ProblemsIf the GC stops operation because of a fault, check the displayfor any messages. Press [Status] and scroll to view anyadditional messages.1 Use the keyboard or data system to stop the alert tone. Press[Off/No] on the keyboard or turn off the offending componentin the data system.2 Resolve the problem, for example, by changing gas cylindersor fixing the leak. See the Troubleshooting Guide for details.3 Once the problem is fixed, you may need to either powercycle the instrument, or use the software keyboard or datasystem to turn the problem component off, then on again.For shutdown errors, you will need to do both.Operation Manual21

222Operating BasicsOperation Manual

Agilent 7890B Gas ChromatographOperation Manual3Keypad OperationThe Run Keys 24The GC Component Keys 25The Status Key 26The Info Key 27The General Data Entry Keys 28The Supporting Keys 29Method Storage and Automation Keys 30Keypad Functionality When the GC Is Controlled by an Agilent DataSystem 31The Service Mode Key 32About GC Status 33About Logs 35This section describes the basic operation of the Agilent 7890BGC keypad. For additional information on keypad functionality,see the Advanced Operation Manual.Agilent Technologies23

3Keypad OperationThe Run KeysThese keys are used to start, stop, and prepare the GC to run asample.24[Prep Run]Activates processes required to bring the GC tothe starting condition dictated by the method(such as turning off the inlet purge flow for asplitless injection or restoring normal flow fromgas saver mode). See the Advanced OperationManual for details.[Start]Starts a run after manually injecting a sample.(When you are using an automatic liquid sampleror gas sampling valve, the run is automaticallyactivated at the appropriate time.)[Stop]Immediately terminates the run. If the GC is inthe middle of a run, the data from that run maybe lost. Also see “To resume an aborted sequence”on page 52.Operation Manual

3Keypad OperationThe GC Component KeysThese keys are used to set the temperature, pressure, flow,velocity, and other method operating parameters.To display the current settings, press any one of these keys.More than three lines of information may be available. Use thescroll keys to view additional lines, if necessary.To change settings, scroll to the line of interest, enter thechange, and press [Enter].For context-sensitive help, press [Info]. For example, if youpress [Info] on a setpoint entry, the help provided would besimilar to: Enter a value between 0 and 350.[Oven][Front Inlet] [Back Inlet][Col 1] [Col 2] [Aux Col #][Front Det] [Back Det] [MS/Aux Det][Analog Out 1] [Analog Out 2][Front Injector] [Back Injector][Valve #][Aux Temp #][Aux EPC #][Column Comp]Operation ManualSets oven temperatures, both isothermal andtemperature programmed.Controls inlet operating parameters.Controls column pressure, flow, or velocity.Can set pressure or flow ramps.Controls detector operating parameters. Ifconfigured with an MS, control GC-MScommunications and special functions.Assigns a signal to the analog output. Theanalog output is located on the back of theGC.Edits injector control parameters such asinjection volumes and sample and solventwashes. If configured with an HS, controlGC-HS communications and specialfunctions.Allows control of a sampling valve and/orswitching valves 1 to 8 (on or off). Setsmultiposition valve position.Controls extra temperature zones such as aheated valve box, a mass selective detector(or other) transfer line, or an “unknown”device. Can be used for temperatureprogramming.Provides auxiliary pneumatics to an inlet,detector, capillary flow technology (CFT)device, or other device. Can be used forpressure programming.Creates a column compensation profile.25

3Keypad OperationThe Status Key[Status]Displays “ready,” “not ready,” and “fault”information.When the Not Ready status light is blinking, afault has occurred. Press [Status] to see whichparameters are not ready and what fault hasoccurred.The order in which items appear in the scrollingdisplay window for [Status] can be modified. Youmay, for example, want to display the things youmost frequently check in the top three lines sothat you do not need to scroll to see them. Tochange the order of the Status display:1 Press [Config] [Status].2 Scroll to the setpoint you want to appear firstand press [Enter]. This setpoint will nowappear at the top of the list.3 Scroll to the setpoint you want to appearsecond and press [Enter]. This setpoint willnow be the second item on the list.4 Continue as above until the list is in the orderyou require.26Operation Manual

Keypad Operation3The Info Key[Info]Operation ManualProvides help for the currently shownparameter. For example, if Oven Temp is theactive line in the display (has a next to it),[Info] will display the valid range of oventemperatures. In other cases, [Info] will displaydefinitions or actions that need to be performed.27

3Keypad OperationThe General Data Entry Keys[Mode/Type] Accesses a list of possible parameters associatedwith a component’s nonnumeric settings. Forexample, if the GC is configured with asplit/splitless inlet and the [Mode/Type] key ispressed, the options listed will be split, splitless,pulsed split, or pulsed splitless.[Clear]Removes a misentered setpoint before pressing[Enter]. It can also be used to return to the top lineof a multiline display, return to a previous display,cancel a function during a sequence or method, orcancel loading or storing sequences and methods.[Enter]Accepts changes you enter or selects an alternatemode.Scrolls up and down through the display one lineat a time. The in the display indicates the activeline.Numeric Keys Are used to enter settings for the methodparameters. (Press [Enter] when you are finished toaccept the changes.)[On/Yes][Off/No]Are used when you are setting up parameters,such as the warning beep, method modificationbeep, and key click or for turning on or off a devicelike a detector.[Front] [Back] Are mostly used during configuration operations.For example, when configuring a column, usethese keys to identify the inlet and detector towhich the column is attached.[Delete]28Removes methods, sequences, run table entries,and clock table entries. [Delete] also aborts theadjust offset process for nitrogen-phosphorusdetectors (NPD) without interrupting otherdetector parameters. See the Advanced OperationManual for more details.Operation Manual

Keypad Operation3The Supporting KeysOperation Manual[Time]Displays the current date and time on the first line.The two middle lines show the time between runs,the elapsed time and time remaining during a run,and the last run time and post-time during apost-run.The last line always displays a stopwatch. While onthe stopwatch line, press [Clear] to set the clock tozero and [Enter] to start or stop the stopwatch.[Post Run]Is used to program the GC to do something after arun, such as bakeout or backflush a column. See theAdvanced Operation Manual for details.[Logs]Access three logs: the Run Log, the MaintenanceLog, and the System Event Log. The information inthese logs can be used to support Good LaboratoryPractices (GLP) standards.[Options]Accesses the instrument parameters setup optionsfor calibration, communications, and the keyboardand display. Scroll to the desired line and press[Enter] to access the associated entries. See“Options” on page 189.[Config]Is used to set up components that are notautomatically detectable by the GC but are essentialto running a method, such as column dimensions,carrier and detector gas types, makeup gasconfigurations, sample tray settings, and columnplumbing to inlets and detectors. These settings arepart of, and are stored with, the method.To view the current configuration for a component(such as the inlet or detector), press [Config], thenthe component key of interest. For example,[Config][Front Det] opens front detectorconfiguration parameters.29

3Keypad OperationMethod Storage and Automation KeysThese keys are for loading and storing methods and sequenceslocally on your GC. They cannot be used to access methods andsequences stored by your Agilent data system.[Load] [Method] [Store] [Seq][Run Table]Are used together to load and store methods andsequences on your GC.For example, to load a method, press [Load][Method] and select one from the list of methodsstored in the GC. See “To load a method” onpage 41.Is used to program special events you requireduring a run. A special event could be switching avalve, for example. See the Advanced OperationManual for details.[Clock Table] Is used to program events to occur at a time of day,as opposed to during a run, and to access theInstrument Schedule. The clock table events could,for example, be used to start a shutdown run at5:00 p.m. every day. See the Advanced OperationManual and “Resource Conservation” on page 106.[Seq Control] Starts, stops, pauses, or resumes a sequence, orviews the status of a sequence. See “RunningSequences from the Keypad” on page 51.[Sample Tray] Displays whether the tray and/or bar codereader/mixer is enabled.[Prog] Allows you to program a series of keystrokes[User Key 1] commonly used for specific operations. Press User[User Key 2] Key 1 or User Key 2 to record up to 31 keystrokes asa macro. See the Advanced Operation Manual.30Operation Manual

3Keypad OperationKeypad Functionality When the GC Is Controlled by an Agilent Data SystemWhen an Agilent data system controls the GC, the data systemdefines the setpoints and runs the samples. If configured to lockthe keypad, the data system can prevent the changing ofsetpoints. The Remote LED is lit when a data system iscontrolling the GC. Lit LEDs on the status board show thecurrent progress of a run.When an Agilent data system controls the GC, the keypad can beused: To view run status by selecting [Status] To view the method settings by selecting the GC componentkey To display the last and next run times, the run timeremaining, and the post-run time remaining by repeatedlyselecting [Time] To abort a run by selecting [Stop] To find which computer is controlling the GC by pressing[Options] Communication, then scrolling. The name of thecomputer controlling the GC is listed after the Enable DHCPsetting, along with the number of hosts connected to the GC.Pressing [Stop] during a GC run immediately ends the run. Thedata system may retain the data already collected, but nofurther data is collected for that sample. Agilent data systemsmay allow the next run to begin, depending on the data systemand its settings for handling errors.Operation Manual31

3Keypad OperationThe Service Mode Key[Service Mode] Is used to set up Early Maintenance Feedbackand to access inlet leak checks for selected inlettypes. See “Early Maintenance Feedback(EMF)” on page 118 and the Troubleshootingmanual. This key also accesses settingsintended for service personnel. Because theseadvanced settings can cause problems ifmisused, avoid the service settings unlessspecifically directed to use them.32Operation Manual

3Keypad OperationAbout GC StatusWhen the GC is ready to begin a run, the display screen showsSTATUS Ready for Injection. Alternately, when a component of theGC is not ready to begin a run, the Not Ready LED is lit on thestatus board. Press [Status] to see a message explaining why theGC is not ready.Status boardA lit LED on the status board indicates: The current progress of a run (Pre Run, Post Run, and Run). Items that may require attention (Rate, Not Ready, ServiceDue, and Run Log). The GC is controlled by an Agilent data system (Remote). The GC is programmed for events to occur at specifiedtimes (Clock Table). The GC is in gas saver mode (Gas Saver).Alert tonesA series of warning beeps sounds before a shutdown occurs.The GC starts with one beep. The longer the problem persists,the more the GC beeps. After a short time the component withthe problem shuts down, the GC emits one beep, and a briefmessage is displayed. For example, a series of beeps sounds ifthe front inlet gas flow cannot reach setpoint. The message Frontinlet flow shutdown is briefly displayed. The flow shuts downafter 2 minutes. Press [Off/No] to stop the beep.A continuous tone sounds if a hydrogen flow is shut down or athermal shutdown occurs.Operation Manual33

3Keypad OperationWA RNINGBefore resuming GC operations, investigate and resolve the causeof the hydrogen shutdown. See Hydrogen Shutdown in theTroubleshooting manual for details.One beep sounds when a problem exists, but the problem willnot prevent the GC from executing the run. The GC will emitone beep and display a message. The GC can start the run andthe warning will disappear when a run starts.Fault messages indicate hardware problems that require userintervention. Depending on the type of error, the GC emits nobeep or a single beep.Error conditionsIf a problem occurs, a status message appears. If the messageindicates broken hardware, more information may be available.Press the applicable compone

Automating Data Analysis, Method Development, and Sequence Development 47 Recoverable Errors 48 5 Running a Method or a Sequence from the Keypad Running Methods from the Keypad 50 To manually inject a sample with a syringe and start a run 50 To run a method to process a s

Related Documents:

Gas Chromatograph Mass Spectrometry (GC/MS) - Agilent 7890B GC & 240 Ion Trap MS The Agilent 7890B is a state-of-art gas chromatograph (GC) that provides superior performance for all GC applications with the use of advanced electronic pneumatic control (EPC) modules and high

Agilent 1290 Infinity Agilent G1888A X X Agilent 7673A Agilent 7683A Agilent HS7694 X X Agilent 7695A X Agilent 79855(A) X Agilent 5880 Agilent 5890 Agilent 6850 (27 Pos. Einlegeschale) . Autosampler-Kompatibilitätstabelle 2. H eadline as disp

Agilent’s flagship 7890B GC system has everything you need to boost productivity, protect our . and Parts Finder tools Reduce downtime and operating costs with simplified maintenance and status monitoring. Find consumables and replacemen

Agilent 1290 Infinity X Agilent G1888A Agilent 7673A X Agilent 7683A X Agilent HS7694 Agilent 7695A Agilent 79855(A) X Agilent 5880 X Agilent 5890 X . Autosampler Compatibility Chart Crimp Neck ND8 1 1. Snap Ring ND11 Screw Neck ND13 Shell Vials Shell Vials Shell Vials Shell Vials Headspa

Agilent 1290 Infinity X X Agilent G1888A Agilent 7673A X X X Agilent 7683A X X X X Agilent HS7694 Agilent 7695A Agilent 79855(A) X X Agilent 5880 X X Agilent 5890 X X . Autosampler Compatibility Chart 2. H eadline as disp

centaur pharmaceuticals we impart health to life sr no name of instruments make model 1. gas chromatograph agilent 6850 2. gas chromatograph agilent 6890 3. gc x 2 no. agilent 7890 4. hplc-auto sampler agilent 1100 5. hplc-auto sampler x 7 nos. agilent 1200 6. polarimeter jasco p-1020 7. ft

Tip 1: How to use Agilent 82357B USB/GPIB converter in NI’s MAX or LabVIEW? Figure 2. Typical setup for Agilent IO Libraries Suite. 1 Agilent I/O Libraries Each Agilent IO product is bundled with the Agilent I/O libraries. There are four I/O libraries included in Agilent IO libraries Suite: Agilent

Digital Multi Meter No Substitute Agilent 3458A, Agilent 34401A, Agilent 34410A and Agilent 34411A 1 Dynamic Accuracy Test Kit No Substitute Agilent Z5623A Opt H01 1 Power Sensor No Substitute Agilent E9304A with Opt H18 Agilent 8482A 1 1 Calibration Kit No Substitute Agilent 85032F 1 Fixed