Agilent 1200 Infinity II LC Specification Compendium

2y ago
40 Views
7 Downloads
646.71 KB
58 Pages
Last View : 14d ago
Last Download : 3m ago
Upload by : Lee Brooke
Transcription

Agilent 1200 Infinity LCSpecification CompendiumAgilent Technologies

Notices Agilent Technologies, Inc. 2014, 2015WarrantyNo part of this manual may be reproducedin any form or by any means (including electronic storage and retrieval or translationinto a foreign language) without prior agreement and written consent from AgilentTechnologies, Inc. as governed by UnitedStates and international copyright laws.The material contained in this document is provided “as is,” and is subject to being changed, without notice,in future editions. Further, to the maximum extent permitted by applicablelaw, Agilent disclaims all warranties,either express or implied, with regardto this manual and any informationcontained herein, including but notlimited to the implied warranties ofmerchantability and fitness for a particular purpose. Agilent shall not beliable for errors or for incidental orconsequential damages in connectionwith the furnishing, use, or performance of this document or of anyinformation contained herein. ShouldAgilent and the user have a separatewritten agreement with warrantyterms covering the material in thisdocument that conflict with theseterms, the warranty terms in the separate agreement shall control.Manual Part Number01200-90061Edition08/2015Printed in GermanyAgilent TechnologiesHewlett-Packard-Strasse 876337 Waldbronnreceive no greater than Restricted Rights asdefined in FAR 52.227-19(c)(1-2) (June1987). U.S. Government users will receiveno greater than Limited Rights as defined inFAR 52.227-14 (June 1987) or DFAR252.227-7015 (b)(2) (November 1995), asapplicable in any technical data.Safety NoticesCAUTIONA CAUTION notice denotes ahazard. It calls attention to anoperating procedure, practice, orthe like that, if not correctly performed or adhered to, couldresult in damage to the productor loss of important data. Do notproceed beyond a CAUTIONnotice until the indicated conditions are fully understood andmet.Technology LicensesThe hardware and/or software described inthis document are furnished under a licenseand may be used or copied only in accordance with the terms of such license.Restricted Rights LegendIf software is for use in the performance of aU.S. Government prime contract or subcontract, Software is delivered and licensed as“Commercial computer software” asdefined in DFAR 252.227-7014 (June 1995),or as a “commercial item” as defined in FAR2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June1987) or any equivalent agency regulationor contract clause. Use, duplication or disclosure of Software is subject to AgilentTechnologies’ standard commercial licenseterms, and non-DOD Departments andAgencies of the U.S. Government willWA R N I N GA WARNING notice denotes ahazard. It calls attention to anoperating procedure, practice,or the like that, if not correctlyperformed or adhered to, couldresult in personal injury ordeath. Do not proceed beyond aWARNING notice until the indicated conditions are fully understood and met.Specification Compendium

ContentsContents1 Pumps5Agilent 1290 Infinity II Flexible Pump (G7104A) 6Agilent 1290 Infinity II High Speed Pump (G7120A)2 Injectors913Agilent 1290 Infinity II Multisampler (G7167B) 14Agilent 1260 Infinity Multisampler (G7167A) 20Agilent 1290 Infinity II Vialsampler (G7129B) 26Agilent 1260 Infinity Autosampler (G7129A) 313 UV-Detectors37Agilent 1290 Infinity II Variable Wavelength Detector (G7114B)Agilent 1290 Infinity II DAD (G7117B) 41Agilent 1290 Infinity II DAD FS (G7117A) 444 Special Detectors3847Agilent 1290 Infinity II Evaporative Light Scattering Detector (G7102A)5 Ovens51Agilent 1290 Infinity II Multicolumn Thermostat (G7116B) 52Agilent 1200 Infinity Integrated Column Compartment (G7130A)Specification Compendium48563

Contents4Specification Compendium

Specification Compendium1PumpsAgilent 1290 Infinity II Flexible Pump (G7104A)Physical Specifications 6Performance Specifications 76Agilent 1290 Infinity II High Speed Pump (G7120A)Physical Specifications 9Performance Specifications 10Agilent Technologies95

1PumpsAgilent 1290 Infinity II Flexible Pump (G7104A)Agilent 1290 Infinity II Flexible Pump (G7104A)Physical SpecificationsTable 16Physical SpecificationsTypeSpecificationCommentsWeight16.1 kg (35.5 lbs)Dimensions (height width depth)180 x 396 x 436 mm (7.1 x 15.6 x 17.2inches)Line voltage100 – 240 V , 10 %Line frequency50 or 60 Hz, 5 %Power consumption120 VA / 110 WAmbient operatingtemperature4 – 55 C (39 – 131 F)Ambient non-operatingtemperature-40 – 70 C (-40 – 158 F)Humidity 95 % r.h. at 40 C (104 F)Operating altitudeUp to 3000 m (9842 ft)Non-operating altitudeUp to 4600 m (15092 ft)For storing the moduleSafety standards: IEC, EN,CSA, ULInstallation category II, Pollution degree 2For indoor use tion Compendium

PumpsAgilent 1290 Infinity II Flexible Pump (G7104A)1Performance SpecificationsTable 2Agilent 1290 Infinity II Flexible Pump (G7104A) Performance SpecificationsFeatureSpecificationHydraulic systemDual pistons in series pump with proprietary servo-controlled variablestroke design and smooth motion control for active damping.Pump resolution stepsize300 pL step sizeSettable flow range0.001 – 5 mL/min, in 0.001 mL/min increments (executed in 300 pL/stepincrements).Flow precision 0.07 % RSD or 0.01 min SD, whatever is greaterFlow accuracy 1 % or 10 µL/min, whatever is greaterPressure rangeup to 130 MPa (1300 bar) at 0 – 2 mL/minramping down to 80 MPa (800 bar) at 5 mL/minPressure pulsation 1 % amplitude or 0.5 MPa (5 bar), whatever is ndedpH-range1.0 – 12.5, solvents with pH 2.3 should not contain acid which attackstainless steel.Gradient formationLow pressure quaternary mixingDelay volumeAs low as 350 µLComposition rangeSettable range: 0 – 100 %Recommended range: 1 – 99 % or 5 µL/minCompositionprecision 0.15 % RSD or 0.02 min SD, whatever is greaterCompositionaccuracy 0.4 % absolute (1 – 99 % B)Number of solvent4 out of maximum 26 solventsSolvent selectionvalveInternal 4-solvent gradient formation valve included.External 2x 12 solvent valve as option, fully integrated in the pump controlinterface.Specification Compendium7

1PumpsAgilent 1290 Infinity II Flexible Pump (G7104A)Table 28Agilent 1290 Infinity II Flexible Pump (G7104A) Performance SpecificationsFeatureSpecificationDegassing unitIntegrated.Number of channels: 4, Internal volume per channel: 1.5 mLMaterials in contactwith solventTFE/PDD copolymer, FEP, PEEK, PPS, stainless steel, polyimideAutomatic PurgeValveIncluded, allows automatic inline-filter back-flush and automatic mixerchange, e.g. for optional TFA-mixerActive Seal washIncludedIntelligent SystemEmulationTechnology (ISET)IncludedCommunicationsController-area network (CAN), RS232C, APG remote: ready, start, stop andshutdown signals, LANSafety andmaintenanceExtensive diagnostics, error detection and display through includedAgilent LabAdvisor, leak detection, safe leak handling, leak output signalfor shutdown of the pumping system. Low voltage in major maintenanceareas.GLP featuresEarly maintenance feedback (EMF) for continuous tracking of instrumentusage in terms of seal wear and volume of pumped mobile phase withpre-defined and user settable limits and feedback messages. Electronicrecords of maintenance and errors.HousingAll materials are recyclable.Specification Compendium

PumpsAgilent 1290 Infinity II High Speed Pump (G7120A)1Agilent 1290 Infinity II High Speed Pump (G7120A)Physical SpecificationsTable 3Physical SpecificationsTypeSpecificationWeight21.0 kg (46.3 lbs)Dimensions (height width depth)200 x 396 x 436 mm (7.9 x 15.6 x 17.2inches)Line voltage100 – 240 V , 10 %Line frequency50 or 60 Hz, 5 %Power consumption210 VA / 180 WAmbient operatingtemperature4 – 55 C (39 – 131 F)Ambient non-operatingtemperature-40 – 70 C (-40 – 158 F)Humidity 95 % r.h. at 40 C (104 F)Operating altitudeUp to 3000 m (9842 ft)Non-operating altitudeUp to 4600 m (15092 ft)For storing the moduleSafety standards: IEC, EN,CSA, ULInstallation category II, Pollution degree 2For indoor use only.Specification sing9

1PumpsAgilent 1290 Infinity II High Speed Pump (G7120A)Performance SpecificationsTable 410Agilent 1290 Infinity II High Speed Pump (G7120A) Performance SpecificationsFeatureSpecificationHydraulic systemTwo dual pistons in series, pumps with proprietary servo-controlledvariable stroke design and smooth motion control.Pump resolution stepsize300 pL step sizeSettable flow range0.001 – 5 mL/min, in 0.001 mL/min increments (executed in 300 pL/stepincrements).Flow precision 0.07 % RSD or 0.005 min SD, whatever is greaterFlow accuracy 1 % or 10 µL/min, whatever is greaterPressure rangeup to 130 MPa (1300 bar) at 0 – 2 mL/minramping down to 80 MPa (800 bar) at 5 mL/minPressure pulsation 1 % amplitude or 0.5 MPa (5 bar), whatever is ndedpH-range1.0 – 12.5, solvents with pH 2.3 should not contain acid which attackstainless steel.Gradient formationHigh pressure binary mixingDelay volumeAs low as 45 µL (10 µL without mixer)Compositionprecision 0.15 % RSD or 0.01 min SD, whatever is greaterCompositionaccuracy 0.35 % absoluteNumber of solvent2 out of maximum 26 solventsSolvent selectionvalveInternal 4-solvent selection valve included.External 2x 12 solvent valve as option, fully integrated in the pump controlinterface.Integrated degassingunitIncludedNumber of channels: 2 Internal volume per channel: 1.5 mLMaterials in contact with solvent: TFE/PDD Copolymer, FEP, PEEK, PPS.Specification Compendium

1PumpsAgilent 1290 Infinity II High Speed Pump (G7120A)Table 4Agilent 1290 Infinity II High Speed Pump (G7120A) Performance SpecificationsFeatureSpecificationAutomatic PurgeValveIncludedActive Seal washIncludedIntelligent SystemEmulationTechnology (ISET)IncludedCommunicationsController-area network (CAN), RS232C, APG remote: ready, start, stop andshutdown signals, LANSafety andmaintenanceExtensive diagnostics, error detection and display through includedAgilent LabAdvisor, leak detection, safe leak handling, leak output signalfor shutdown of the pumping system. Low voltage in major maintenanceareas.GLP featureEarly maintenance feedback (EMF) for continuous tracking of instrumentusage in terms of seal wear and volume of pumped mobile phase withpre-defined and user settable limits and feedback messages. Electronicrecords of maintenance and errors.HousingAll materials are recyclable.Specification Compendium11

112PumpsAgilent 1290 Infinity II High Speed Pump (G7120A)Specification Compendium

Specification Compendium2InjectorsAgilent 1290 Infinity II Multisampler (G7167B) 14Physical Specifications 14Performance Specifications 15Physical Specifications of the Sample Cooler 18Agilent 1260 Infinity Multisampler (G7167A) 20Physical Specifications 20Performance Specifications (G7167A) 21Physical Specifications of the Sample Cooler 24Agilent 1290 Infinity II Vialsampler (G7129B) 26Physical Specifications 26Performance Specifications 27Physical Specifications of the Sample Cooler 29Agilent 1260 Infinity Autosampler (G7129A) 31Physical Specifications 31Performance Specifications 32Physical Specifications of the Sample Cooler 34Agilent Technologies13

2InjectorsAgilent 1290 Infinity II Multisampler (G7167B)Agilent 1290 Infinity II Multisampler (G7167B)Physical SpecificationsTable 514Physical SpecificationsTypeSpecificationCommentsWeight22 kg (48.5 lbs)w/o sample coolerDimensions (height width depth)320 x 396 x 468 mm (12.6 x 15.6 x 18.4inches)Line voltage100 – 240 V , 10 %Line frequency50 or 60 Hz, 5 %Power consumption180 VA, 180 WAmbient operatingtemperature4 - 40 C (39 - 104 F)Ambient non-operatingtemperature-40 – 70 C (-40 – 158 F)Humidity 95 % r.h. at 40 C (104 F)Operating altitudeUp to 3000 m (9842 ft)Non-operating altitudeUp to 4600 m (15092 ft)For storing the moduleSafety standards: IEC, EN,CSA, ULInstallation category II, Pollution degree 2For indoor use only.ISM ClassificationISM Group 1 Class BAccording to CISPR n Compendium

InjectorsAgilent 1290 Infinity II Multisampler (G7167B)2Performance SpecificationsTable 6Agilent 1290 Infinity II Multisampler (G7167B) Performance SpecificationsTypeSpecificationCommentInjection range forSingle-needleinstrumentsDefault: 0.1 – 20 µL in 0.1 µL increments; optional:40 µL or 100 µL (using 100 µL analytical head)Up to 1300 bar using 40 µL (default) oroptional 100 µL analytical head0.1 – 500 µL or 900 µL in 0.1 µL increments (using900 µL analytical head)Pressure range up to 400 bar due to 900 µLanalytical head0.1 – 120 µL in 0.1 µL increments with 1290Infinity large volume injection kit (hardwaremodification required) G4216-687110.1 – 500 µL or 1500 µL in 0.1 µL increments with100 µL upgrade kit (hardware modificationrequired) G7167-68711Pressure range up to 1300 bar Multi-drawmodus (Injection into needle-seat capillary)Injection range forDual-needleinstrumentsDefault: 0.1 – 20 µL in 0.1 µL increments; optional:40 µL or 100 µLUp to 1300 bar using 100 µL analytical headUp to 500 µL in 0.1 µL increments depending oninstalled loop sizeUp to 1300 bar using 100 µL analytical head Multi-loadPrecision forSingle-needleinstruments 0.15 % RSD or SD 10 nL, whatever is greaterMeasured caffeinePrecision forDual-needleinstruments 0.2 % RSD or SD 10 nL, whatever is greaterMeasured caffeinePressure rangeUp to 1300 bar (G7167B)Max pressure for basic instrumentSample viscosityrange0.2 – 5 cpSample capacity1H Drawer up to 8 drawers and 16 positionsShallow well plates (MTP)Max. 6144/1536 samples (384MTP/96)2H Drawer up to 4 drawers and 8 positionsMTP, deep well plates, vials, Eppendorf3072 samples, 432 vials (2 mL)3H Drawer up to 2 drawers and 4 positionsMTP, deep well plates, vials up to 6 mL, Eppendorf1536 samples, 60 vials (6 mL), 384 vials(1 mL), 216 vials (2 mL)Specification Compendium15

2InjectorsAgilent 1290 Infinity II Multisampler (G7167B)Table 6Agilent 1290 Infinity II Multisampler (G7167B) Performance SpecificationsTypeSpecificationCommentInjection cycle time 10 s using following standard conditions:Default draw speed: 100 µL/minUsing standard Single-needle setupDefault eject speed: 400 µL/min Injection volume:1 µLTime between 2 injections is notmechanically limited, time delay depends oncommunication speed of software, OS ornetwork connectionsCarry Over 0.003 % (30 ppm) Multisampler Standard andDual Needle 0.0009 % (9 ppm) Multisampler MultiwashUsing the following conditions: Column: Agilent Pursuit XRs 3 C18, 2.0 x50 mm Mobile phase: A: 0.1 % TFA in water B: 0.1 % TFA in Acetonitrile Isocratic : % B 40 % Flow rate: 0.5 mL/min Temperature: 25 C Wavelength: 257 nm Sample: 1200 ng/µL Chlorhexidine(dissolved with mobile phase A), 1 µLinjected and measured on G4212A DAD Wash solution: H2O with 0.1 % TFA (3 s)MultiwashOuter needle wash and seat backflush forcarryover reduction with up to 3 different solventsControl and dataevaluationAgilent Open Lab CDSA.02.01 or above (A.02.02 supports SampleEntry UI)MassHunter QQQB.07.00 SP11 or aboveMassHunter QTOFB.05.01 SP31 or aboveLab AdvisorB.02.05 or aboveICF for 3rd party SW controlA.02.01 or aboveLC and CE DriversA.02.10 or aboveAgilent Instant Pilot (G4208A)B.02.17 or above (currently notsupported/official release 2015)Local Control16Specification Compendium

InjectorsAgilent 1290 Infinity II Multisampler (G7167B)Table 6Agilent 1290 Infinity II Multisampler (G7167B) Performance oller-area network (CAN), Local AreaNetwork (LAN)ERI: ready, start, stop and shut-down signalsSafety andmaintenanceExtensive support for troubleshooting andmaintenance is provided by the Instant Pilot,Agilent Lab Advisor, and the ChromatographyData System.Safety-related features are leak detection, safeleak handling, leak output signal for shutdown ofpumping system, and low voltages in majormaintenance areas.GLP featuresEarly maintenance feedback (EMF) for continuoustracking of instrument usage with user-settablelimits and feedback messages. Electronic recordsof maintenance and errors.HousingAll materials recyclable.Metering deviceMetering device in high pressure flow path12Commentonly for the integration in an Infinity I LC setupSpecification Compendium17

2InjectorsAgilent 1290 Infinity II Multisampler (G7167B)Physical Specifications of the Sample CoolerCooling unit is designed as vapor- compression refrigeration system.Contains fluorinated greenhouse gas (refrigerant) according to the Kyotoprotocol. For specifications of refrigerant, charge capacity, carbon dioxideequivalent (CDE), and global warming potential (GWP) see instrumentlabel.Table 718Physical Specification of the Sample CoolerTypeSpecificationCommentsWeight 6 kgDimensions (height width depth)205 mm x 340 mm x 370 mmRefrigerant gasHFC-134a (0.042 kg)Line voltage24 VDC (nominal)Current10 A max.Ambient operating temperature4 – 40 C (39.2 – 104 F)Ambient non-operatingtemperature-40 – 70 C (-20 – 158 F)Operating altitudeUp to 3000 m (9842 ft)Non-operating altitudeUp to 4600 m (15091 ft)Safety standards:IEC, CSA, ULInstallation category II, Pollutiondegree 2Ozone depletionpotential (ODP) 0For indoor use only.Specification Compendium

InjectorsAgilent 1290 Infinity II Multisampler (G7167B)Table 8Performance Specifications Agilent 1290 Sample CoolerTypeSpecificationsOperating principleHigh performance, low-energy consumptionmicro-compressor based cooler withozone-friendly HFC-134a coolant (42 g),user-upgradable.Temperature rangefrom 4 C to ambientTemperature settablefrom 4 – 40 C in 1 incrementsTemperature accuracy ( 25 C, 50 % r.H.)2 C to 6 C at a setpoint of 4 CSpecification Compendium219

2InjectorsAgilent 1260 Infinity Multisampler (G7167A)Agilent 1260 Infinity Multisampler (G7167A)Physical SpecificationsTable 920Physical SpecificationsTypeSpecificationCommentsWeight22 kg (48.5 lbs)w/o sample coolerDimensions (height width depth)320 x 396 x 468 mm (12.6 x 15.6 x 18.4inches)Line voltage100 – 240 V , 10 %Line frequency50 or 60 Hz, 5 %Power consumption180 VA, 180 WAmbient operatingtemperature4 - 40 C (39 - 104 F)Ambient non-operatingtemperature-40 – 70 C (-40 – 158 F)Humidity 95 % r.h. at 40 C (104 F)Operating altitudeUp to 3000 m (9842 ft)Non-operating altitudeUp to 4600 m (15092 ft)For storing the moduleSafety standards: IEC, EN,CSA, ULInstallation category II, Pollution degree 2For indoor use only.ISM ClassificationISM Group 1 Class BAccording to CISPR n Compendium

InjectorsAgilent 1260 Infinity Multisampler (G7167A)2Performance Specifications (G7167A)Table 10Agilent 1260 Infinity Multisampler (G7167A) Performance SpecificationsTypeSpecificationCommentInjection range forSingle-needleinstrumentsDefault: 0.1 – 100 µL in 0.1 µL increments;optional: 20 µL or 40 µL (using optional 40 µLanalytical head)Up to 600 bar using 100 µL (default) oroptional 40 µL analytical head0.1 – 500 µL or 900 µL in 0.1 µL increments (using900 µL analytical head)Pressure range up to 400 bar due to 900 µLanalytical head0.1 – 120 µL in 0.1 µL increments with 1290Infinity large volume injection kit (hardwaremodification required) G4216-687110.1 – 500 µL or 1500 µL in 0.1 µL increments with100 µL upgrade kit (hardware modificationrequired) G7167-68711Pressure range up to 600 bar Multi-drawmodus (Injection into needle-seat capillary)Default: 0.1 – 100 µL in 0.1 µL increments;optional: 20 µL or 40 µL (using 100 µL analyticalhead)Up to 600 bar using 100 µL analytical headUp to 900 µL in 0.1 µL increments depending oninstalled loop sizeUp to 600 bar using 100 µL analytical headPrecision forSingle-needleinstruments 0.15 % RSD or SD 10 nL, whatever is greaterMeasured caffeinePrecision forDual-needleinstruments 0.2 % RSD or SD 10 nL, whatever is greaterMeasured caffeinePressure rangeUp to 600 bar (G7167A)Max pressure for basic instrumentSample viscosityrange0.2 – 5 cpSample capacity1H Drawer up to 8 drawers and 16 positionsShallow well plates (MTP)Max. 6144/1536 samples (384MTP/96)2H Drawer up to 4 drawers and 8 positionsMTP, deep well plates, vials, Eppendorf3072 samples, 432 vials (2 mL)3H Drawer up to 2 drawers and 4 positionsMTP, deep well plates, vials up to 6 mL, Eppendorf1536 samples, 60 vials (6 mL), 384 vials(1 mL), 216 vials (2 mL)Injection range forDual-needleinstrumentsSpecification Compendium21

2InjectorsAgilent 1260 Infinity Multisampler (G7167A)Table 10Agilent 1260 Infinity Multisampler (G7167A) Performance SpecificationsTypeSpecificationCommentInjection cycle time 10 s using following standard conditions:Default draw speed: 100 µL/minUsing standard Single-needle setupDefault eject speed: 400 µL/min Injection volume:1 µLTime between 2 injections is notmechanically limited, time delay depends oncommunication speed of software, OS ornetwork connectionsCarry Over 0.003 % (30 ppm) Multisampler Standard andDual Needle 0.0009 % (9 ppm) Multisampler MultiwashUsing the following conditions: Column: Agilent Pursuit XRs 3 C18, 2.0 x50 mm Mobile phase: A: 0.1 % TFA in water B: 0.1 % TFA in Acetonitrile Isocratic : % B 40 % Flow rate: 0.5 mL/min Temperature: 25 C Wavelength: 257 nm Sample: 1200 ng/µL Chlorhexidine(dissolved with mobile phase A), 1 µLinjected and measured on G4212A DAD Wash solution: H2O with 0.1 % TFA (3 s)MultiwashOuter needle wash and seat backflush forcarryover reduction with up to 3 different solventsControl and dataevaluationAgilent Open Lab CDSA.02.01 or above (A.02.02 supports SampleEntry UI)MassHunter QQQB.07.00 SP11 or aboveMassHunter QTOFB.05.01 SP31 or aboveLab AdvisorB.02.05 or aboveICF for 3rd party SW controlA.02.01 or aboveLC and CE DriversA.02.10 or aboveAgilent Instant Pilot (G4208A)B.02.17 or above (currently notsupported/official release 2015)Local Control22Specification Compendium

InjectorsAgilent 1260 Infinity Multisampler (G7167A)Table 10Agilent 1260 Infinity Multisampler (G7167A) Performance oller-area network (CAN), Local AreaNetwork (LAN)ERI: ready, start, stop and shut-down signalsSafety andmaintenanceExtensive support for troubleshooting andmaintenance is provided by the Instant Pilot,Agilent Lab Advisor, and the ChromatographyData System.Safety-related features are leak detection, safeleak handling, leak output signal for shutdown ofpumping system, and low voltages in majormaintenance areas.GLP featuresEarly maintenance feedback (EMF) for continuoustracking of instrument usage with user-settablelimits and feedback messages. Electronic recordsof maintenance and errors.HousingAll materials recyclable.Metering deviceMetering device in high pressure flow path12Commentonly for the integration in an Infinity I LC setupSpecification Compendium23

2InjectorsAgilent 1260 Infinity Multisampler (G7167A)Physical Specifications of the Sample CoolerCooling unit is designed as vapor- compression refrigeration system.Contains fluorinated greenhouse gas (refrigerant) according to the Kyotoprotocol. For specifications of refrigerant, charge capacity, carbon dioxideequivalent (CDE), and global warming potential (GWP) see instrumentlabel.Table 1124Physical Specification of the Sample CoolerTypeSpecificationCommentsWeight 6 kgDimensions (height width depth)205 mm x 340 mm x 370 mmRefrigerant gasHFC-134a (0.042 kg)Line voltage24 VDC (nominal)Current10 A max.Ambient operating temperature4 – 40 C (39.2 – 104 F)Ambient non-operatingtemperature-40 – 70 C (-20 – 158 F)Operating altitudeUp to 3000 m (9842 ft)Non-operating altitudeUp to 4600 m (15091 ft)Safety standards:IEC, CSA, ULInstallation category II, Pollutiondegree 2Ozone depletionpotential (ODP) 0For indoor use only.Specification Compendium

InjectorsAgilent 1260 Infinity Multisampler (G7167A)Table 12Performance Specifications Agilent 1290 Sample CoolerTypeSpecificationsOperating principleHigh performance, low-energy consumptionmicro-compressor based cooler withozone-friendly HFC-134a coolant (42 g),user-upgradable.Temperature rangefrom 4 C to ambientTemperature settablefrom 4 – 40 C in 1 incrementsTemperature accuracy ( 25 C, 50 % r.H.)2 C to 6 C at a setpoint of 4 CSpecification Compendium225

2InjectorsAgilent 1290 Infinity II Vialsampler (G7129B)Agilent 1290 Infinity II Vialsampler (G7129B)Physical SpecificationsTable 1326Physical SpecificationsTypeSpecificationCommentsWeight19 kg (41.9 lbs)w/o sample coolerDimensions (height width depth)324 x 396 x 468 mm (12.8 x 15.6 x 18.4inches)Line voltage100 – 240 V , 10 %Line frequency50 or 60 Hz, 5 %Power consumption350 VA / 350 W / 1195 BTU/hAmbient operatingtemperature4 - 40 C (39 - 104 F), without chiller up to55 C (131 F)Ambient non-operatingtemperature-40 – 70 C (-40 – 158 F)Operating altitudeUp to 3000 m (9842 ft)Non-operating altitudeUp to 4600 m (15092 ft)For storing the moduleSafety standards: IEC, CSA,ULInstallation category II, Pollution degree 2For indoor use only.ISM ClassificationISM Group 1 Class BAccording to CISPR 11Wide-rangingcapabilitySpecification Compendium

InjectorsAgilent 1290 Infinity II Vialsampler (G7129B)2Performance SpecificationsTable 14Performance Specifications (G7129B)TypeSpecificationCommentInjection range0.1 – 20 µL in 0.1 µL increments (default)0.1 – 40 µL in 0.1 µL increments if 40 µL loop isinstalled0.1 – 120 µL in 0.1 µL increments with 1290Infinity large volume injection kit (hardwaremodification required)0.1 – 100 µL in 0.1 µL (if 100 µL-loop and100 µL-head is installed)Up to 1500 µL with 1400 µL-multi-draw kit and100 µL-analytical headPrecision 0.25 % RSD of peak areas from 5 µL to 100 µLMeasured caffeinePressure rangeUp to 130 MPa (1300 bar, 18854 psi)Sample viscosityrange0.2 – 5 cpSample capacity132 x 2 mL vial (two trays default)100 x 2 mL vial (two classic trays optional)36 x 6 mL vials (two trays optional)Carry over 0.004 % (40 ppm) with needle washSpecification Compendiumup to 130 MPa (1300 bar, 18854 psi)up to 60 MPa (600 bar, 8702 psi)Using the following conditions: ZORBAX Eclipse Plus C18, RRHD, 2.1 x 50mm, 1.8 µm (959757-902) Mobile Phase: A: 0.1 % TFA in water B: 0.1 % TFA in acetonitrile Isocratic : % B 33 % Flow rate: 0.5 mL/min Column temperature: 50 C Wavelength detection: 257/4 nm, ref.wavelength 380/100 nm, 20 Hz Injection volume: 1 µL Sample: 1200 ng/µL Chlorhexidine for UV,(dissolved with mobile phase A), 1 µLinjected and measured both on AgilentG7117B DAD Wash solution: H2O with 0.1 % TFA (5 s)27

2InjectorsAgilent 1290 Infinity II Vialsampler (G7129B)Table 14Performance Specifications (G7129B)TypeSpecificationInjection cycle time18 s for draw speed 200 µL/minEjection speed: 200 µL/minInjection volume: 1 µLMinimum samplevolume1 µL from 5 µL sample in 100 µL microvial, or1 µL from 10 µL sample in 300 µL microvial.Needle height offset has to be adapted toensure that needle doesn’t touch vial bottom.Default needle height 0 equates to 2 mmabove the vial bottom.Control and dataevaluationAgilent Open Lab CDSMass hunter QQQMass hunter TOF/QTOFLab AdvisorICF for 3rd party SW controlLC and CE DriversA.02.02 or aboveB.08.01 or aboveB.07.02 or aboveB.02.07 or aboveA.02.04 or aboveA.02.12 or aboveLocal controlAgilent Instant Pilot (G4208A)B.02.17 or aboveCommunicationsController-area network (CAN),Local AreaNetwork (LAN)ERI: ready, start, stop and shut-down signalsSafety andmaintenanceExtensive support for troubleshooting andmaintenance is provided by the Instant Pilot,Agilent Lab Advisor, and the ChromatographyData System. Safety-related features are leakdetection, safe leak handling, leak output signalfor shutdown of pumping system, and lowvoltages in major maintenance areas.GLP featuresEarly maintenance feedback (EMF) forcontinuous tracking of instrument usage withuser-settable limits and feedback messages.Electronic records of maintenance and errors.HousingAll materials recyclable.Metering deviceMetering device in high pressure flow path28CommentSpecification Compendium

InjectorsAgilent 1290 Infinity II Vialsampler (G7129B)2Physical Specifications of the Sample CoolerCooling unit is designed as vapor- compression refrigeration system.Contains fluorinated greenhouse gas (refrigerant) according to the Kyotoprotocol. For specifications of refrigerant, charge capacity, carbon dioxideequivalent (CDE), and global warming potential (GWP) see instrumentlabel.Table 15Physical Specification of the Sample CoolerTypeSpecificationWeight 6 kgDimensions (height width depth)205 mm x 340 mm x 370 mmRefrigerant gasHFC-134a (0.042 kg)Line voltage24 VDC (nominal)Current10 A max.Ambient operating temperature4 – 40 C (39.2 – 104 F)Ambient non-operatingtemperature-40 – 70 C (-20 – 158 F)Operating altitudeUp to 3000 m (9842 ft)Non-operating altitudeUp to 4600 m (15091 ft)Safety standards:IEC, CSA, ULInstallation category II, Pollutiondegree 2Specification CompendiumCommentsOzone depletionpotential (ODP) 0For indoor use only.29

2InjectorsAgilent 1290 Infinity II Vialsampler (G7129B)Table 1630Performance Specifications Agilent 1290 Sample CoolerTypeSpecificationsOperating principleHigh performance, low-energy consumptionmicro-compressor based cooler withozone-friendly HFC-134a coolant (42 g),user-upgradable.Temperature rangefrom 4 C to ambientTemperature settablefrom 4 – 40 C in 1 incrementsTemperature accuracy ( 25 C, 50 % r.H.)2 C to 6 C at a setpoint of 4 CSpecification Compendium

2InjectorsAgilent 1260 Infinity Autosampler (G7129A)Agilent 1260 Infinity Autosampler (G7129A)Physical SpecificationsTable 17Physical SpecificationsTypeSpecificationCommentsWeight19 kg (41.9 lbs)w/o sample coolerDimensions (height width depth)324 x 396 x

Agilent 1290 Infinity II Vialsampler (G7129B) 26 Agilent 1260 Infinity Autosampler (G7129A) 31 3UV-Detectors 37 Agilent 1290 Infinity II Variable Wavelength Detector (G7114B) 38 Agilent 1290 Infinity II DAD (G7117B) 41

Related Documents:

Agilent 1290 Infinity Agilent G1888A X X Agilent 7673A Agilent 7683A Agilent HS7694 X X Agilent 7695A X Agilent 79855(A) X Agilent 5880 Agilent 5890 Agilent 6850 (27 Pos. Einlegeschale) . Autosampler-Kompatibilitätstabelle 2. H eadline as disp

G4226A 1290 Infinity Autosampler G4227A 1290 Infinity Flexible Cube G4240A 1260 Infinity Chip Cube MS Interface G4260A Agilent 380-ELSD G4260B 1260 Infinity Evaporative Light Scattering Detector G4261A Agilent 385-ELSD G4261B 1290 Infinity Evaporative Light Scattering Detector G4277A 1290 Infinity LC

Apr 01, 2012 · The Agilent 1290 Infinity 2D‑LC Solution used for the performance evaluation comprised the following modules: Two Agilent 1290 Infinity Pumps (G4220A) Agilent 1290 Infinity Autosampler (G4226A) with thermostat (G1330) Agilent 1290 Infinity Thermostatted Colu

Agilent 1290 Infinity X Agilent G1888A Agilent 7673A X Agilent 7683A X Agilent HS7694 Agilent 7695A Agilent 79855(A) X Agilent 5880 X Agilent 5890 X . Autosampler Compatibility Chart Crimp Neck ND8 1 1. Snap Ring ND11 Screw Neck ND13 Shell Vials Shell Vials Shell Vials Shell Vials Headspa

Agilent 1290 Infinity X X Agilent G1888A Agilent 7673A X X X Agilent 7683A X X X X Agilent HS7694 Agilent 7695A Agilent 79855(A) X X Agilent 5880 X X Agilent 5890 X X . Autosampler Compatibility Chart 2. H eadline as disp

Agilent 1290 Infinity UHPLC system consisting of an Agilent 1290 Infinity Binary Pump (G4220A), an Agilent 1290 Infinity High Performance Autosampler (G4226A), a sample cooler (G1330B), and an Agilent 1290 Infinity Thermostatted Column compartment (G1316C). The UHPLC system was coupled to an

The Agilent 1290 Infinity 2D-LC Solution comprises the following modules: Two Agilent 1290 Infinity Pumps (G4220A), Agilent 1290 Infinity Autosampler (G4226A) with cooler (G1330A) Agilent 1290 Infinity Thermostatted Column Compartment (TCC) (G1316C) with 2-position-4 port-duo valve (

Alex Rider: Never say Die by Anthony Horowitz Below are the complete reviews, written by the Lovereading4kids members. George Hutton - Dormston Secondary School Alex Rider receives a suspicious email from who could be Jack Starbright who was kidnapped on his previous mission. However, whilst trying to locate Jack, he accidentally manages to get tangled up in another MI6 Mission which could put .