Cytokine, Chemokine, And Growth Factors Instruction Manual

1y ago
14 Views
2 Downloads
9.04 MB
44 Pages
Last View : 9d ago
Last Download : 3m ago
Upload by : Mara Blakely
Transcription

.NEW magnetic protocol and software requirements!NBio-Plex Pro AssaysCytokine, Chemokine, and Growth FactorsInstruction ManualFor technical support, call your local Bio-Rad office, or in the U.S., call 1-800-4BIORAD (1-800-424-6723).For research use only. Not for diagnostic procedures.

Table of ContentsSection 1Introduction1Section 2Principle3Section 3Product Description5Section 4Recommended Materials8Section 5Bead Regions9Section 6Sample Preparation10Section 7Standard Preparation12Section 8Assay Instructions8.1 Assay Plate and Wash Format8.2 Plan Experiment8.3 Calibrate Vacuum Apparatus8.4 Prepare Coupled Magnetic Beads8.5 Assay Procedure161616171819Section 9Data Acquisition23Section 10Troubleshooting Guide29Section 11Safety Considerations35Section 12Plate Layout Template36Section 13Legal Notice38

Section 1IntroductionCytokines, chemokines, and growth factors are cell signaling proteins,mediating a wide range of physiological responses, including immunity,inflammation, and hematopoiesis. They are also associated with aspectrum of diseases ranging from tumor growth to infections toParkinson’s disease. These molecules are typically measured either bybioassay or immunoassay. Both techniques are time consuming and canfacilitate the analysis of only a single target at a time. The Bio-Plex system, which incorporates novel technology using color-coded beads,permits the simultaneous detection of up to 100 cytokines in a single wellof a 96-well microplate.Bio-Plex Pro cytokine, chemokine, and growth factor assays aremagnetic bead-based multiplex assays designed to measure multiplecytokines, chemokines, and growth factors in diverse matrices like serum,plasma, and tissue culture supernatants. The multiplexing feature makesit possible to quantitate the level of multiple proteins in a single well, in just3 hrs, using as little as 12.5 µl of serum or plasma, or 50 µl of tissueculture supernatants.As one of the most recent additions to the Bio-Plex system, these assaysincorporate magnetic beads into their design. The magnetic beads allowfor the option of using magnetic separation during wash steps instead ofvacuum filtration. Magnetic separation allows for greater automationwithout significant alterations to the standard Bio-Plex assay protocol.Bio-Plex Manager software is recommended for Bio-Plex Pro cytokine,chemokine, and growth factor assays. For instructions using other xMAPsystem software packages, contact Bio-Rad technical support or yourBio-Rad field application specialist.For a current listing of Bio-Plex cytokine, chemokines, and growth factorassays, visit us on the Web at www.bio-rad.com/bio-plex/.1

2

Section 2PrincipleTechnologyThe Bio-Plex system is built around three core technologies. The first is a noveltechnology that uses up to 100 unique fluorescently dyed beads (xMAPtechnology) that permit the simultaneous detection of up to 100 different typesof molecules in a single well of a 96-well microplate. The second is a dedicatedflow cytometer with two lasers and associated optics to measure the differentmolecules bound to the surface of the beads. The third is a high-speed digitalsignal processor that efficiently manages the fluorescent data.Assay FormatThe principle behind the 96-well plate-formatted, bead-based assays is similarto a capture sandwich immunoassay. An antibody directed against the desiredcytokine, chemokine, or growth factor target is covalently coupled to internallydyed beads. The coupled beads are allowed to react with a sample containingthe target biomolecules. After a series of washes to remove unbound protein, abiotinylated detection antibody specific to an epitope different from that of thecapture antibody is added to the reaction. This results in the formation of asandwich of antibodies around the cytokine, chemokine, or growth factortarget. A streptavidin-phycoerythrin (streptavidin-PE) reporter complex is thenadded to bind to the biotinylated detection antibodies on the bead surface.Data Acquisition and AnalysisData from the reaction are acquired using the Bio-Plex system (or Luminexsystem), a dual-laser, flow-based microplate reader system. The contents ofthe well are drawn up into the reader. The lasers and associated opticsdetect the internal fluorescence of the individual dyed beads as well as thefluorescent reporter signal on the bead surface. This identifies each assayand reports the level of target protein in the sample. Intensity of fluorescencedetected on the beads indicates the relative quantity of target molecules inthe tested samples. A high-speed digital processor efficiently manages thedata output, which is further analyzed and presented as fluorescenceintensity (FI) and target concentration on Bio-Plex Manager software.3

4

Section 3Product DescriptionBio-Plex Pro cytokine, chemokine, and growth factor assays are offeredin a convenient kit format that includes assay, reagent, and diluentcomponents all in a single box. If magnetic separation is being used, flatbottom plates are sold separately. Instructions for this kit may be used forthe detection of the following:Premixed Multiplex AssaysOff-the-Shelf Singleplex AssaysThe following are catalog numbers required for ordering singleplex assays:5

Bio-Plex Pro Human Cytokine AssaysBio-Plex Pro Mouse Cytokine AssaysGroup IIL-1IL-1raCatalog #171-B5001M171-B5002MGroup IIIL-1IL-2RCatalog #171-B6001M171-B6002MGroup IIL-1IL-1Catalog #171-G5001M171-G5002MGroup IIIL-15IL-18*Catalog #171-G6001MIL-2IL-4IL-5IL-6IL-7IL-8IL-9IL-10IL-12 (p70)IL-13IL-15IL-17Basic FGFEotaxinG-CSFGM-CSFIFNIP-10MCP-1 B5024M171-B5025M171-B5026M171-B5027MIL-3IL-12 (p40)IL-16IL-18*CTACKGROHGFICAM-1IFN- IL-9IL-10IL-12 (p40)IL-12 (p70)IL-13IL-17EotaxinG-CSFGM-CSFIFNKCMCP-1 G5020M171-G5021M171-G5022M171-G5023MBasic 20M171-B6021M171-B6022MN/A*IL-18 is not available as singleplex assays.Bio-Plex Express AssaysVisit Bio-Rad’s Bio-Plex Express assay service at www.bio-rad.com/bioplex/x-plex to select from available Bio-Plex cytokine, chemokine, andgrowth factor targets and mix your own multiplex assay.Bio-Plex x-Plex AssaysVisit Bio-Rad’s x-Plex blending service at www.bio-rad.com/bio-plex/xplex to configure QC tested custom-mixed multiplex assays.6

Required MaterialsBio-Plex Pro cytokine, chemokine, and growth factor assays contain thefollowing components all in one kit:Components1 x 96-WellFormat10 x 96-WellFormatCoupled magnetic beads (10x)600 l6,000 lDetection antibodies (10x)320 l3,200 lStandard (additional vials sold separately)1 vial10 vialsUniversal standard diluent10 ml100 ml8 ml80 mlAssay buffer50 ml500 mlWash bufferUniversal sample diluent130 ml1,300 mlDetection antibody diluent5 ml50 mlStreptavidin-PE (100x)1 vial1 vial1 plate10 platesFilter plate (96-well)Sealing tape1 pack of 45 packs of 4Instruction Manual11Tube holder11Sold separately are the Bio-Plex Pro flat bottom plates, required for magneticseparation on the Bio-Plex Pro wash stations. Order Bio-Rad catalog #171025001.Storage and StabilityKit components should be stored at 4 C and never frozen. Coupledmagnetic beads and streptavidin-PE should be stored in the dark. Allcomponents are guaranteed for up to 6 months from the date of purchasewhen stored as specified in this manual.7

Section 4Recommended Materials8

Section 5Bead RegionsEach bead has been assigned a number (bead region) from 1 to 100according to the ratio of the two dyes contained. For correct dataacquisition using the Bio-Plex (or Luminex) system and associatedsoftware, use the chart below to identify the bead region for the targets inyour Bio-Plex assay.NOTE: DO NOT use preset panels found in Bio-Plex Manager version 5.0or lower. For details, see Section 9 (Data Acquisition). Bead regions forBio-Plex Pro cytokine, chemokine and growth factor assays are notidentical to existing preset panels.NEWNEWNEWNEWNEWNEWNEWNEWNEWNEWNEW9

Section 6Sample PreparationThis section provides instructions for preparing samples derived fromserum, plasma, and tissue culture supernatant. For sample preparationsnot mentioned here, consult the publications listed in Bio-Rad bulletin5297, available for download at discover.bio-rad.com. Additionalinformation can also be found in Section 10 (Troubleshooting Guide).Serum and Plasma SamplesNOTE: For plasma samples, both EDTA plasma and citrate plasma areacceptable. Heparin plasma may absorb certain cytokines and istherefore not preferred. Avoid using hemolyzed samples.1.To prepare serum, allow blood to clot at room temperature for 30 minto 1 hr. For plasma, centrifuge immediately after collection in tubescontaining anticoagulant. Centrifuge serum or plasma samples at13,200 rpm for 10 min at 4 C to clear the samples of precipitate.Alternatively, carefully filter the samples with a 0.8/0.22 µm dual filterto prevent instrument clogging. Assay immediately or aliquot andstore samples in single use aliquots at –70 C. Avoid repeatedfreeze/thaw cycles.2.Dilute 1 volume of sample with 3 volumes of sample diluent. Keepthe samples on ice until ready for use.NOTE: Physiological levels of VCAM-1 and ICAM-1 are typicallyfound at much higher concentrations. Sample dilutions of 1:100 arefrequently required to achieve concentrations in the measurablerange of the standard curve. Dilute serum 1:50 or 1:100 as follows:1) dilute serum 1:4 in sample diluent, and 2) dilute further 1:25 usingstandard diluent.10

Tissue Culture Supernatant Samples1. Collect and process the tissue culture supernatant samples andassay immediately or store samples in single use aliquots at –70 C.Centrifuge samples at 13,200 rpm for 10 min at 4 C to clear thesamples of precipitate. Avoid repeated freeze/thaw.2.If required, dilute the culture supernatant with culture medium. Forcell lines cultured in serum-free culture medium, collect samples,centrifuge as above, and add a carrier protein (such as BSA) at a finalconcentration of at least 0.5%. Keep the samples on ice until readyfor use.Tissue LysatesRefer to bulletin 5297 for a list of published articles on cytokine analysis intissue samples.Lavage, Sputum, and Other Biological Fluid SamplesKeep all samples on ice until ready for use. If dilution is required, use abuffer that is similar to the sample. Reconstitute and dilute the cytokinestandard using a buffer that is as similar to the sample as possible. Addcarrier protein (such as BSA) at a concentration of at least 0.5%.11

Section 7Standard PreparationOne vial of lot-specific Iyophilized standards is required for each Bio-PlexPro cytokine, chemokine, and growth factor assay. By choosing amongavailable assays, mixing standards may be necessary. This procedure willprepare enough standard to run each diluted standard point in duplicate.Prior to performing this procedure, verify the contents of the cytokinestandard by referring to the lot-specific peel-off sicker provided with thestandard. This also lists the concentrations of the reconstitutedstandards, which are required for Data Acquisition (Section 9).Reconstitute Standards1. Gently tap the glass vial containing the lyophilized standard on a solidsurface to ensure the pellet is at the bottom.2.Reconstitute lyophilized standard:Premixed Standards1. Reconstitute one vial of lyophilized standard with 500 µl of thesame matrix as samples. Do not use assay buffer to dilutestandards.2.Gently vortex reconstituted standard 1–3 sec and incubate onice for 30 min. Be consistent with the incubation time foroptimal assay performance and reproducibility.Preparation of Master Stock From Two Premixed Standards1. Refer to diagram on the following page to reconstitute each vial oflyophilized standard with 200 µl of the same matrix as samples.Do not use assay buffer to dilute standards.2.Gently vortex reconstituted standard 1–3 sec and incubate onice for 30 min. Be consistent with the incubation time foroptimal assay performance.12

3.Add 96 µl from each of the reconstituted standards into a single1.5 ml tube containing 183 µl of the standard diluent. This willbecome the master stock and will be at a final volume of375 µl. This will serve as Standard I in the standardPhotomultiplier Tube (PMT) setting standard curve.4.Gently vortex for 1–3 seconds.Prepare Standard Dilution SeriesThe cytokine, chemokine, and growth factor concentrations specified forthe eight-point standard dilution set have been selected for optimizedcurve fitting using the five-parameter logistic (5PL) or four-parameterlogistic (4PL) regression in Bio-Plex Manager software. Resultsgenerated using dilution points other than those listed in this manual havenot been optimized.1.Label a set of 1.5 ml Eppendorf tubes as shown in the diagrams onthe following page.13

2.Use the provided serum standard diluent, and pipet the appropriatevolume of standard diluent into the tubes (refer to the diagrams on thefollowing page). Alternatively use culture medium for culture samples.NOTE: If the concentrations are expected to be in the range10–1,000 pg/ml, such as in serum, then use the high PMT settingand prepare serial dilutions accordingly (see following page). It isrecommended to run a low PMT setting standard curve first.3.Transfer the respective volume of reconstituted standard:Premixed StandardsAdd 128 µl of reconstituted standard to the first 1.5 ml tube with 72 µlof standard diluent or appropriate matrix. This is identified as S1 inthe diagram on the following page.Master Stock (Two Cytokine Standards)Add 375 µl of reconstituted standard to the first 1.5 ml tube with 0 µlof standard diluent or appropriate matrix. This is identified as S1 inthe diagram on the following page.4.Continue making serial dilutions of the standard as shown. Aftermaking each dilution, vortex gently for 1–3 sec and change the pipettip after every transfer.NOTE: Running an additional two 0 pg/ml blanks is stronglyrecommended. The blank wells are useful for troubleshooting anddetermining limit of detection (LOD). Use 50 µl of the appropriate standarddiluent as the blank sample. The 0 pg/ml points should be formatted asblanks, not as points in the curve, when using Bio-Plex Manager software.Formatting the wells as blanks automatically subtracts the backgroundmean fluorescence intensity (MFI) values from the reading and may resultin negative MFI values in some wells. If negative MFI values areundesirable, format the 0 pg/ml wells as background controls.5.Keep the standards on ice until ready for use. Reconstitutedstandards should be used immediately and not frozen for future use.14

Low PMT – Standard Dilution Series for Premixed StandardsLow PMT – Standard Dilution Series for Master Stock(Two Cytokine Standards)High PMT – Standard Dilution Series for Premixed Standards15

Section 8Assay InstructionsThe following instructions apply to Bio-Plex Pro cytokine, chemokine, andgrowth factor assays in a premixed multiplex, singleplex, Bio-Plex Express,and Bio-Plex x-plex assay formats. The assay protocol to run assays thatare manually mixed is the same as that for assays that are premixed, or in asingleplex format. The only difference is the additional step of manuallymixing the coupled beads and detection antibodies prior to use.8.1 Assay Plate and Wash FormatAssays can be prepared in 96-well filter plates for vacuum based washing, orBio-Plex Pro flat bottom plates for magnetic based washing. The wash stepscan be performed with the filter plates using a manual vacuum manifold, or thevacuum feature of the Pro II Wash Station. Magnetic washing can beperformed using the magnetic settings of the Pro, or Pro II Wash Station.8.2 Plan ExperimentFor calculating coupled beads and detection dilutions, see tables onpages 18 and 21.1.Bring all buffers to room temperature. Avoid bubbles when pipetting.NOTE: To ensure adequate volume to run 10 plates using 10x96reagent kits, remove only the required amount from the bottles foreach assay run. For example, transfer a one-time volume of assaybuffer into a 50 ml reservoir, sufficient to perform all steps of the assayprocedure (i.e. prewetting the filter plate, diluting 10x coupled beads,diluting 100x streptavidin-PE, and resuspending the beads).2.Assign wells of a 96-well plate to be used for each standard, control,and sample (see the example on the following page).3.Determine the total number of wells that will be used in the assay anddesign the plate layout. Notice that the template in Section 12 canbe used as a reference.16

Example Plate8.3 Calibrate Vacuum Manifold (if appropriate)Prior to performing any Bio-Plex assay, the vacuum apparatus must becalibrated to ensure optimal bead yield. For more detailed instructionsrefer to the Bio-Plex suspension array system hardware instructionmanual.NOTE: Calibration for Bio-Plex Pro washers is not required.1.Place a standard flat-bottom microplate (not a filter plate) on thevacuum apparatus.2.Turn on the vacuum source to maximum level and press down on theplate until a vacuum is established (typically 20–30 Hg).3.Adjust the vacuum pressure using the control valves on the unit. Thepressure should be set to 1–3” Hg.4.Once the vacuum is set correctly, remove the flat-bottom plate. Checkthe vacuum periodically as house vacuum systems can fluctuate. Ensurethat all wells are exposed to vacuum, as residual fluid can lead toimprecise results. As a general guideline, 100 µl of liquid should takeapproximately 3–4 sec to completely clear the well.17

8.4 Prepare Coupled Magnetic BeadsProtect the beads from light. Keep all tubes on ice until ready for use.1.Quick-spin the coupled beads (10x). Carefully open the cap andgently resuspend the coupled beads with a pipet before removingdesired volume.2.Prepare coupled beads (1x) in assay buffer. Each well requires 5 µl ofcoupled beads (10x) adjusted to a final volume of 50 µl using assaybuffer. For mixing different bead sets, use equal volumes of eachcoupled bead. The final volume should always be the sameregardless of how many beads are combined. Refer to the examplebead calculations below, which include 20% excess in coupled beadsto compensate for transfer loss.Example Coupled Magnetic Bead Calculations forPremixed or Singleplex Assays (1x)Example Coupled Magnetic Bead Calculations forMixing Coupled Beads (1x from10x stocks)# of Wells10x Beads (µµl)G-CSF, for example10x Beads (µµl)8-plex, for exampleAssay Buffer (µµl)Total Volume 33,3542,87518

8.5 Assay ProcedureAssay Key – The terms below are repeated throughout the assayprocedure. Refer to these detailed instructions when wash and incubateon shaker are shown in bold.StepDetailed DirectionsVacuum ManifoldPlace the filter plate on a calibrated vacuum apparatus and removethe buffer by vacuum filtration. Add 100 µl of wash buffer to eachwell and remove the liquid as before. Repeat as specified.Thoroughly blot the bottom of the filter plate with a clean paper towelbetween each vacuum step to prevent cross contamination and plateleakage. Place the filter plate on the plastic plate holder.Bio-Plex Pro Wash StationFor preloaded magnetic programs with the Bio-Plex Pro flat bottomplates, use MAG X2 for two cycles of initial bead wash and MAG X3for three cycles of washing following three incubation steps.WashAlways rinse the instrument after use. Prime the instrument withoutliquid to empty all liquid from the system before removing the manifold.Bio-Plex Pro II Wash StationFor preloaded magnetic programs with the Bio-Plex Pro flat bottomplates use MAG X2 for two cycles of initial bead wash and MAG X3 forthree cycles of washing folllowing three incubation steps.For preloaded vacuum programs using filter plates, use VAC X2 fortwo cycles of initial bead wash and VAC X3 for three cycles of washfollowing three incubation steps.Always rinse the instrument after use. Prime the instrument withoutliquid to empty.all liquid from the system before removing the manifold.Incubateon shakerGently cover the filter plate with a new sheet of sealing tape andpress down on edges to prevent liquid escaping from the bottom ofthe wells. Place the filter plate on a microplate shaker andthen cover with aluminum foil. Shake the filter plate at roomtemperature at 1,100 rpm for 30 sec, and then at 300 rpm for thespecified incubation time.19

1.Equilibrate the diluted standards, samples, and controls at roomtemperature for 20 min prior to use.2.Prewet (skip this step if using a flat bottom plate).a. Vacuum ManifoldPrewet the wells of a 96-well filter plate using 100 µl of assaybuffer and remove the liquid by vacuum filtration. Dry the bottomof the filter plate thoroughly by blotting on a clean paper towel.If fewer than 96 wells are required mark the plate to identify theunused wells for later use and cover the unused wells withsealing tape prior to vacuum filtration.b.Vacuum With Pro II Wash StationPrewet the wells of a 96-well filter plate using 125 µl of assaybuffer and remove the liquid following the PREVAC program onthe Pro II Wash Station.3.Vortex the coupled beads (1x) for 30 sec at medium speed. Add 50 µlto each well.4.Wash twice with the vacuum manifold or Bio-Plex Pro wash stations(VAC 2X or MAG 2X). See page 19 for detailed instructions.5.Gently vortex the diluted standards, controls, and samples for 1–3 sec.Add 50 µl of standard, control, or sample to each well, changing thepipet tip after every volume transfer. Incubate on shaker at roomtemperature for 30 min (see page 19 for detailed instructions).6.While the samples are incubating, vortex the detection antibody (10x)for 15–20 sec at medium speed. Always perform a 30 sec spin tocollect the entire volume at the bottom of the vial.7.Prepare detection antibodies (1x) using detection antibody diluent.Each well requires 2.5 µl of detection antibodies (10x) adjusted to afinal volume of 25 µl with detection antibody diluent. Refer to theexample detection antibody calculations on the following page, whichinclude 25% excess in detection antibodies to compensate fortransfer loss.20

Example Detection Antibody Calculations forPremixed or Singleplex AssaysExample Detection Antibody Calculations forMixing Detection Antibodies# of Wells9688807264564810x Detection10x DetectionAntibody (µµl)G-CSF, for exampleAntibody (µµl)8-plex, for exampleDetectionAntibodyDiluent ,4002,2002,0001,8001,6001,4001,200Total Volume (µµl)3,0002,7502,5002,2502,0001,7501,5008.After incubating the samples, slowly remove and discard the sealingtape.9.Wash three times with the vacuum manifold or Bio-Plex Pro washstations (VAC 3X or MAG 3X). See page 19 for detailed instructions.10. Vortex the detection antibodies gently for 1–3 sec and add 25 µl toeach well. Incubate on shaker at room temperature for 30 min (seepage 19 for detailed instructions).11. While the detection antibodies are incubating, vortex the streptavidin-PE(100x) for 15–20 sec at medium speed. Always perform a 30 sec spin tocollect the entire volume at the bottom of the vial.21

12. Prepare streptavidin-PE (1x) using assay buffer. Each well requires 0.5 µlof streptavidin-PE (100x) adjusted to a final volume of 50 µl with assaybuffer. Refer to the example streptavidin-PE calculation below, whichincludes 25% excess in streptavidin-PE to compensate for transfer loss.ExampleStreptavidin-PE CalculationsExample Streptavidin-PE Calculations13. After the detection antibody incubation, slowly remove and discardthe sealing tape.14. Wash three times with the vacuum manifold or Bio-Plex Pro washstations (VAC 3X or MAG 3X). See page 19 for detailed instructions.15. Vortex the streptavidin-PE (1x) vigorously for 3–5 sec and add 50 µlto each well. Incubate on shaker at room temperature for 10 min(see page 19 for detailed instructions).NOTE: It is important to be consistent with this incubation step toreduce inter-assay variation.16. After the streptavidin-PE incubation, slowly remove and discardthe sealing tape.17. Wash three times with the vacuum manifold or Bio-Plex Pro washstations (VAC 3X or MAG 3X). See page 19 for detailed instructions.18. Add 125 µl of assay buffer to each well. Cover the plate with a newsheet of sealing tape. Shake the plate at room temperature at 1,100rpm for 30 sec and remove the sealing tape before reading the plate.22

Section 9Data AcquisitionBio-Plex Manager software is recommended for Bio-Plex Pro cytokine, chemokine, and growth factor assays. To minimize protocolsetup, lot-specific protocols for Bio-Plex Manager version 4.0 and higherare available for download at www.biorad.com/bio-plex/standards.For instructions using other xMAP system software packages, contact BioRad technical support or your regional Bio-Rad field applications specialist.Prepare System1. Empty the waste bottle and fill the sheath fluid bottle before starting, ifhigh throughput fluidics (HTF) are not present. This will prevent fluidicsystem backup and potential data loss.2.Turn on the reader, XY platform, and HTF (if included). Allow thesystem to warm up for 30 min (if not already done).3.Select Start upand follow the instructions. If the system is idlefor 4 hours without acquiring data, the lasers will automatically turnoff. To reset the 4-hour countdown, select Warm upand wait forthe optics to reach operational temperature.Calibrate With Low RP1 Target ValueDaily calibration is recommended before acquiring data. Calibrate usingBio-Plex calibration beads and target values. Bio-Plex cytokine,chemokine, and growth factor assays are run using the low RP1 targetvalue on the Cal 2 bottle.Bio-Plex Manager Software version 5.0 and HigherThis software version requires calibration only with the low RP1target value.1.Select Calibrateand confirm that the default values for CAL1and CAL2 are the same as the values on the Bio-Plex calibrationbead labels. Use the Bio-Plex low RP1 target value.23

2.Select OK and follow the instructions for CAL1 and CAL2calibration.Bio-Plex Manager Software version 4.0, 4.1, and 4.1.1These software versions require calibration at either low or high RP1target values. Recalibration is necessary between assays requiringdifferent RP1 target values. Daily calibration is recommended beforeacquiring data.1.Select Calibrateand confirm that the default values for CAL1and CAL2 are the same as the values on the Bio-Plex calibrationbead labels. Use the appropriate Bio-Plex low RP1 target value.2.Select OK and follow the instructions for CAL1 and CAL2calibration.Prepare Protocol1. Open the lot-specific assay protocols (available for download atwww.biorad.com/bio-plex/standards) or create a new protocol byselecting File, then New from the main menu. Locate the steps at theleft of the protocol menu.2.Select Step 1 (Describe Protocol) and enter information about theassay.3.Select Step 2 (Select Analytes) and create a new panel. Do not usepreset panels found in the Bio-Plex Manager software, version 5.0 orlower. If using a downloaded assay protocol from the Bio-Plexwebsite, analytes will already be entered. Confirm the selectedanalytes and proceed to #4.a)Click the Add Panel buttonb)Enter a name for the new panel in the top field. If using Bio-PlexManager version 5.0 (or higher), select the MagPlex assay typefrom the pull-down menu. Then click Add to add analytes.24in the Select Analytes toolbar.

c)Enter the bead region number of the first analyte in the Regionfield and the analyte name in the Name field.NOTE: The bead region number must be correct for properdetection of analytes. Refer to Section 5 (Bead Regions) toconfirm the bead region for each analyte in the assay is correct.4.d)Click Add Continue to add the analyte to the panel and continueadding more analytes. When you have entered your last analyte,click the Add button to add it to the list.e)When you are finished creating the panel, click OK to save yourchanges and return to the Select Analytes window.f)After selecting the panel, move analytes to the Selected view bychoosing from analytes in the Available view using the Addbutton or Add All button. Note that the analytes will already beentered in the Selected view when using a downloaded protocol.Select Step 3 (Format Plate) and click on the Plate Formatting tab.Select the number of replicates and the auto fill direction. Then clicktheicon and drag the cursor over all the wells that containstandards. Then click on theicon and drag the cursor over thewells that contain blanks. Repeat withandto identify all thewells that contain controls and samples.NOTE: If the protocol was downloaded from at www.biorad.com/bioplex/standards, the standards will already be added to the plateformat. Make any necessary changes to the plate layout to matchyour plate requirements.25

Plate Formatting Example5.Select Step 4 (Enter Standards Info) to enter standards information.Skip to the next step if using a downloaded protocol from the website; this information will already be entered.NOTE: If running the assay at high PMT setting, divide thestandard values provided on the lot-specific peel-off sticker by ten.a)

Bio-Plex Manager software is recommended for Bio-Plex Pro cytokine, chemokine, and growth factor assays. For instructions using other xMAP system software packages, contact Bio-Rad technical support or your Bio-Rad field application specialist. For a current listing of Bio-Plex cytokine, chemokines, and growth factor

Related Documents:

Entamoebalysyl-tRNA Synthetase Contains a Cytokine-Like Domain with Chemokine Activity towards Human Endothelial Cells Manuel Castro de Moura1., Francesc Miro1., Jung Min Han2, Sunghoon Kim2, Antonio Celada1,3, Lluı s Ribas de Pouplana1,4* 1Institute for Research in Biomedicine, Barcelona, Spain, 2Center for Medicinal Protein Network and Systems Biology,

with a normal immune system attempts at cytokine fine tuning, but in this setting, it is commensurate to sustained abnormal biomechanical stressing. Unlike SpA, where restoration of aberrant and excessive cytokine “fine tuning” is efficacious, antag-onism of these pathways in biomechani

PDGF-driven proliferation, migration, and IL8 chemokine secretion in human corneal fibroblasts involve JAK2-STAT3 signaling pathway Neeraj Vij,1 Ajay Sharma, 1,3 Mahesh Thakkar,3,4 Sunilima Sinha, Rajiv R. Mohan1,2,3 1Mason Eye Institute, School of Medicine, University of Missouri-Columbia, Columbia, MO; 2College of Veterinary Medicine, University of Missouri-Columbia, Columbia, MO; 3Harry S .

Journal of NeuroVirology, 9: 300–314, 2003 c 2003 Taylor & Francis Inc. ISSN: 1355-0284 print / 1538-2443 online DOI: 10.1080/13550280390201010 The chemokine receptor CXCR4 regulates cell-cycle

Journal of NeuroVirology, 10(suppl. 1): 1–5, 2004 2004 Journal of NeuroVirologyc ISSN: 1355-0284 print / 1538-2443 online DOI: 10.1080/13550280490268287 Regulation of cell cycle proteins by chemokine

er (NK)-like T cells that can exert potent MHC-unrestricted antitumor activity. A number of pre-clinical and clinical studies have demonstrated that CIK cells can serve as a safe and potent immunotherapy of malignant tumors. N-ace-tylcysteine (NAC) has been demonstrated to enhance the T-cell functions by increasing their proliferation and cytokine

β-sheet breaker peptide iAβ5, has been shown to not only reduce plaque load, but also lessen inflammation-induced cerebral damage [29]. Aβ has also been shown to induce cytokine production in various ways. The nuclear factor kappa B (NFκB) pathway leads to cytokine production and this

ASAM FOLAT (FOLIC ACID) Berbentuk kristal kuning oranye, tidak berasa, tidak berbau Tahan cahaya matahari bila dlm larutan asam Fungsi utama : pematangan sel darah merah Kekurangan : anemia, lelah Sumber : sayuran hijau, hati, gandum, kacang hijau, daging, ikan Kebutuhan : Lk 170 mg, Pr 150 mg . Tugas 1. Sebutkan pengelompokkan vitamin 2. Jelaskan penyebab dan terjadinya anemia pada wanita ! 3 .