IntelliVue Patient Monitor - Medaval

1y ago
5 Views
2 Downloads
8.25 MB
534 Pages
Last View : 2m ago
Last Download : 3m ago
Upload by : Francisco Tran
Transcription

In stru ctio ns fo r UseIntelliVue Patient MonitorMX400/450/500/550/600/700/800MX 40 0 / 4 5 0 / 5 0 0 / 5 5 0 Re l e a s e K w i t h Re v . K. 1 x . x xMX600/ 700/80 0 Release J with Re vison J.xx. xxPatient Monitori ng

1Table of Contents1 IntroductionSafety InformationSecurity InformationIntroducing the MonitorDevices for Acquiring MeasurementsOperating and NavigatingOperating ModesUnderstanding ScreensConnecting Additional Displays to the MonitorUsing the XDS Remote DisplayUsing the Visitor ScreenUnderstanding ProfilesUnderstanding SettingsChanging Wave SpeedsFreezing WavesUsing LabelsEntering Measurements ManuallyChanging Monitor SettingsChecking Your Monitor RevisionGetting StartedDisconnecting from PowerNetworked MonitoringUsing the Integrated PCUsing Your Monitor with a Monitor in Companion Mode2 What's New?What's New in Release K.1 (for MX400/450/500/550 only)What's New in Release K.0What's New in Release J.03 AlarmsVisual Alarm IndicatorsAudible Alarm IndicatorsAcknowledging AlarmsPausing or Switching Off AlarmsAlarm LimitsReviewing AlarmsLatching AlarmsTesting AlarmsAlarm Behavior at Power 55555556596061636467717273733

Alarm Recordings4 Patient Alarms and INOPsPatient Alarm MessagesTechnical Alarm Messages (INOPs)5 Managing Patients and EquipmentPatient ConceptsEquipment ConceptsManaging PatientsManaging EquipmentCare GroupsInformation Center Compatibility737575811051051051061191261326 ECG, Arrhythmia, ST and QT Monitoring133Skin Preparation for Electrode PlacementConnecting ECG CablesSelecting the Primary and Secondary ECG LeadsChecking Paced ModeUnderstanding the ECG DisplayMonitoring Paced PatientsChanging the Size of the ECG WaveChanging the Volume of the QRS ToneChanging the ECG Filter SettingsSelecting Positions of Va and Vb Chest Leads (for 6-lead placement)Choosing EASI or Standard Lead PlacementAbout ECG LeadsECG Lead FallbackECG Lead PlacementsEASI ECG Lead PlacementCapture 12-LeadECG and Arrhythmia Alarm OverviewUsing ECG AlarmsECG Safety InformationAbout Arrhythmia MonitoringSwitching Arrhythmia Analysis On and OffChoosing an ECG Lead for Arrhythmia MonitoringAtrial Fibrillation AlarmAberrantly-Conducted BeatsIntermittent Bundle Branch BlockUnderstanding the Arrhythmia DisplayArrhythmia RelearningArrhythmia AlarmsAbout ST MonitoringSwitching ST or STE On and OffUnderstanding the ST 4

Updating ST Baseline SnippetsRecording ST SegmentsAbout the ST Measurement PointsST AlarmsSTE AlarmsViewing ST MapsAbout QT/QTc Interval MonitoringQT AlarmsSwitching QT Monitoring On and Off7 Monitoring Pulse RateEntering the Setup Pulse MenuSystem Pulse SourceSwitching Pulse On and OffUsing Pulse Alarms8 Monitoring Respiration Rate (Resp)Lead Placement for Monitoring RespUnderstanding the Resp DisplayChanging Resp Detection ModesChanging the Size of the Respiration WaveChanging the Speed of the Respiration WaveUsing Resp AlarmsChanging the Apnea Alarm DelayResp Safety Information9 Monitoring SpO2SpO2 SensorsApplying the SensorConnecting SpO2 CablesMeasuring SpO2SpO2 Signal Quality Indicator (FAST SpO2 only)Assessing a Suspicious SpO2 ReadingChanging the Averaging TimeUnderstanding SpO2 AlarmsPleth WavePerfusion NumericPerfusion Change IndicatorSetting SpO2/Pleth as Pulse SourceSetting Up Tone ModulationSetting the QRS VolumeCalculating SpO2 Difference10 Monitoring NBPIntroducing the Oscillometric NBP MeasurementPreparing to Measure 002002002002012012012032032055

Starting and Stopping MeasurementsEnabling Automatic Mode and Setting Repetition TimeEnabling Sequence Mode and Setting Up The SequenceChoosing the NBP Alarm SourceSwitching Pulse from NBP On/OffAssisting Venous PunctureCalibrating NBP11 Monitoring TemperatureMaking a Temp MeasurementCalculating Temp Difference20720820820920921021021121121212 Monitoring Invasive Pressure213Setting up the Pressure MeasurementZeroing the Pressure TransducerAdjusting the Calibration FactorDisplaying a Mean Pressure Value OnlyChanging the Pressure Wave ScaleOptimizing the WaveformUsing the Wave CursorNon-Physiological Artifact SuppressionChoosing the Pressure Alarm SourceCalibrating Reusable Transducer CPJ840J6Calculating Cerebral Perfusion PressureCalculating Pulse Pressure VariationMeasuring IAPMeasuring Pulmonary Artery Wedge PressureEditing the WedgeIdentifying the Pressure Analog Output 222322413 Monitoring Cardiac OutputHemodynamic ParametersUsing the C.O. Procedure WindowAccessing the Setup C.O. and Setup CCO MenusEntering the HemoCalc WindowMeasuring C. O. Using the PiCCO MethodMeasuring C.O. Using the Right Heart Thermodilution MethodDocumenting C.O. MeasurementsC.O. Injectate GuidelinesC.O./CCO Curve Alert MessagesC.O./CCO Prompt MessagesC.O./CCO Warning MessagesC.O./CCO Safety Information14 Monitoring Carbon DioxideMeasurement 241242

Measuring CO2 using M3014A or X2Measuring Mainstream CO2 using M3016AMeasuring Microstream CO2 using M3015A/BSetting up all CO2 MeasurementsUnderstanding the IPI Numeric15 Monitoring Airway Flow, Volume and PressureAttaching the Flow SensorZero CalibrationAutomatic PurgingManual PurgingGas CompensationSetting up Spirometry16 Monitoring tcGasIdentifying tcGas Module ComponentsSetting the tcGas Sensor TemperatureUsing the tcGas Site TimerSetting the tcGas Barometric PressureRemembraning the tcGas TransducerCalibrating the tcGas TransducerApplying the tcGas TransducerFinishing tcGas MonitoringZeroing the tcGas Relative Heat PowerTcGas Corrections17 Monitoring Intravascular Oxygen SaturationSelecting a Measurement LabelPreparing to Monitor with the M1021A Wide ModulePreparing to Monitor with the M1011A Narrow ModuleFurther Information for Both Modules18 Monitoring EEGEEG Monitoring SetupUsing the EEG Impedance/Montage WindowAbout Compressed Spectral Arrays (CSA)Changing EEG SettingsEEG ReportsEEG Safety InformationEEG and Electrical Interference19 Monitoring BISBIS Monitoring SetupBIS Continuous Impedance CheckBIS Cyclic Impedance CheckBIS 822832842842842852862882882897

Changing the BIS Smoothing RateSwitching BIS and Individual Numerics On and OffChanging the Scale of the EEG WaveSwitching BIS Filters On or OffBIS Safety Information20 Monitoring NMTStimulation ModesPreparing to Measure NMTTaking NMT MeasurementsChanging the NMT Measurement SettingsAlarmsUnderstanding NMT Numerics21 Guardian Early Warning ScoringPerforming the Scoring ProcedureUnderstanding Guardian Early Warning ScoringViewing EWS Trend DataUsing Different Types of Scoring22 Using a Telemetry Device and a Monitor (PIIC only)How Can You Combine Devices?Use Models With Telemetry23 TrendsViewing TrendsSetting Up TrendsDocumenting TrendsTrends DatabasesScreen Trends24 CalculationsViewing CalculationsReviewing CalculationsPerforming CalculationsEntering Values for CalculationsDocumenting Calculations25 High Resolution Trend WavesChanging the Hi-Res Trend Waves DisplayedHi-Res Trend Wave ScalesHi-Res Trend Waves and OxyCRGPrinting Hi-Res Trend Wave ReportsHi-Res Trend Wave 7328329331331331331332332

26 Event Surveillance333Levels of Event SurveillanceEvent GroupsEvent EpisodesEvents Pop-Up KeysEvent TriggersThe Events DatabaseViewing EventsAnnotating EventsDocumenting Events33333433433533633934034234227 ProtocolWatch349SSC Sepsis Protocol34928 Recording357Paper-Strip RecordingElectronic Recording35736529 Printing Patient Reports369Starting Report PrintoutsStopping Reports PrintoutsSetting Up ReportsSetting Up Individual Print JobsChecking Printer SettingsPrinting a Test ReportSwitching Printers On or Off for ReportsDashed Lines on ReportsUnavailable Printer: Re-routing ReportsChecking Report Status and Printing ManuallyPrinter Status MessagesSample Report Printouts36937137137237337437437437437537537630 Using the Drug Calculator381Accessing the Drug CalculatorPerforming Drug CalculationsCharting Infusion ProgressUsing the Titration TableDocumenting Drug Calculations38238238538538631 VueLink Modules387Connecting an External DeviceChanging VueLink Waves and Numerics DisplayedViewing the VueLink Device Data WindowUsing VueLink ScreensSwitching VueLink On and OffAlarms/INOPs From External Devices3883883893893893899

Language Conflict with External Device Drivers32 IntelliBridge EC10Connecting an External DeviceChanging Waves and Numerics DisplayedViewing the IntelliBridge Device Data WindowUsing Screens with External Device DataAlarms/INOPs from External Devices33 Using TimersViewing TimersTimer Setup Pop-up KeysSetting Up TimersDisplaying a Timer On The Main ScreenDisplaying A Clock On The Main Screen34 Respiratory wing LoopsCapturing and Deleting LoopsShowing/Hiding LoopsChanging Loops Display SizeUsing the Loops CursorChanging Loops TypeSetting Up Source DeviceDocumenting Loops39940040040040040140140135 Laboratory Data403Viewing Received Data36 Using BatteriesBattery Power IndicatorsChecking Battery ChargeWhen Battery Lifetime is ExpiredReplacing a BatteryOptimizing Battery PerformanceBattery Safety Information37 Care and CleaningGeneral PointsCleaning the EquipmentDisinfecting the EquipmentSterilizing the EquipmentCleaning, Sterilizing and Disinfecting Monitoring AccessoriesCleaning the SO2 Optical ModuleCleaning the Recorder Printhead (M1116B only)Cleaning Batteries and the Battery 13413413413414

38 Maintenance and TroubleshootingInspecting the Equipment and AccessoriesInspecting the Cables and CordsMaintenance Task and Test ScheduleTroubleshootingReturning Equipment for RepairDisposing of the MonitorDisposing of Empty Calibration Gas Cylinders39 AccessoriesECG/Resp AccessoriesNBP AccessoriesInvasive Pressure AccessoriesSpO2 AccessoriesTemperature AccessoriesCardiac Output (C.O.) AccessoriesMainstream CO2 AccessoriesSidestream CO2 AccessoriesMainstream CO2 Accessories (for M3016A)Microstream CO2 AccessoriesSpirometry AccessoriestcGas AccessoriesEEG AccessoriesBIS AccessoriesSO2 Accessories for M1021ASO2 Accessories for M1011ANMT AccessoriesRecorder AccessoriesBattery Accessories40 SpecificationsIndications for UseRestricted AvailabilityUse EnvironmentManufacturer's InformationSymbolsInstallation Safety InformationMonitor Mounting PrecautionsAltitude SettingMonitor Safety SpecificationsPhysical SpecificationsEnvironmental SpecificationsEMC and Radio Regulatory ComplianceMonitor Performance SpecificationsInterface SpecificationsDisplay 344344444444444544745545645645745846246447047411

M4605A Battery SpecificationsMeasurement SpecificationsSafety and Performance Tests41 Default Settings AppendixCountry-Specific Default SettingsAlarm and Measurement Default SettingsAlarm Default SettingsECG, Arrhythmia, ST and QT Default SettingsPulse Default SettingsRespiration Default SettingsSpO2 Default SettingsNBP Default SettingsTemperature Default SettingsInvasive Pressure Default SettingsCardiac Output Default SettingsCO2 Default SettingsSpirometry Default SettingstcGas Default SettingsIntravascular Oxygen Saturation Default SettingsSvO2 Default SettingsScvO2 Default SettingsEEG Default SettingsBIS Default SettingsNMT Default SettingsVueLink Default 4514515517518518519519520520520521521522523

11IntroductionThese Instructions for Use are for clinical professionals using the IntelliVue MX400/MX450, MX500/MX550, and MX600/MX700/MX800 patient monitor.This basic operation section gives you an overview of the monitor and its functions. It tells you how toperform tasks that are common to all measurements (such as entering data, switching a measurementon and off, setting up and adjusting wave speeds, working with profiles). The alarms section gives anoverview of alarms. The remaining sections tell you how to perform individual measurements, andhow to care for and maintain the equipment.Familiarize yourself with all instructions including warnings and cautions before starting to monitorpatients. Read and keep the Instructions for Use that come with any accessories, as these containimportant information about care and cleaning that is not repeated here.This guide describes all features and options. Your monitor may not have all of them; they are not allavailable in all geographies. Your monitor is highly configurable. What you see on the screen, how themenus appear and so forth, depends on the way it has been tailored for your hospital and may not beexactly as shown here.In this guide: A warning alerts you to a potential serious outcome, adverse event or safety hazard. Failure toobserve a warning may result in death or serious injury to the user or patient. A caution alerts you to where special care is necessary for the safe and effective use of theproduct. Failure to observe a caution may result in minor or moderate personal injury or damageto the product or other property, and possibly in a remote risk of more serious injury.MX400/ Whenever a monitor's identifier appears to the left of a heading or paragraph, it means that theMX450 information applies to that monitor only. Where the information applies to all models, no distinction ismade.For installation, repair, testing and troubleshooting instructions, refer to the Service Guide for yourmonitor model.Rx only: U.S. Federal Law restricts this device to sale by or on the order of a physician.13

1 IntroductionSafety InformationThe following warnings apply to the monitors in general. Warnings that apply to specificmeasurements or procedures can be found in the corresponding chapters.Electrical Hazards and InterferenceWARNINGGrounding: To avoid the risk of electric shock, the monitor must be grounded during operation. If athree-wire receptacle is not available, consult the hospital electrician. Never use a three-wire to twowire adapter.Electrical shock hazard: Do not open the monitor or measurement device. Contact with exposedelectrical components may cause electrical shock. Refer servicing to qualified service personnel.Leakage currents: If multiple instruments are connected to a patient, the sum of the leakage currentsmay exceed the limits given in IEC/EN 60601-1, IEC 60601-1-1, UL 60601-1. Consult your servicepersonnel.Radio frequency interference: The monitor generates, uses and radiates radio-frequency energy, andif it is not installed and used in accordance with its accompanying documentation, may causeinterference to radio communications.Use EnvironmentWARNINGExplosion Hazard: Do not use in the presence of flammable anesthetics or gases, such as aflammable anesthetic mixture with air, oxygen or nitrous oxide. Use of the devices in such anenvironment may present an explosion hazard.Positioning Equipment: The monitor should not be used next to or stacked with other equipment.If you must stack the monitor, check that normal operation is possible in the necessary configurationbefore you start monitoring patients.Environmental Specifications: The performance specifications for the monitors, measurements andaccessories apply only for use within the temperature, humidity and altitude ranges specified in“Environmental Specifications” on page 458.Liquid Ingress: If you spill liquid on the equipment, battery, or accessories, or they are accidentallyimmersed in liquid, contact your service personnel or Philips service engineer. Do not operate theequipment before it has been tested and approved for further use.Prohibited Environments: The monitors are not intended for use in an MRI environment or in anoxygen-enriched environment (for example, hyperbaric chambers).14

1 IntroductionAlarmsWARNING Do not rely exclusively on the audible alarm system for patient monitoring. Adjustment of alarmvolume to a low level or off during patient monitoring may result in patient danger. Rememberthat the most reliable method of patient monitoring combines close personal surveillance withcorrect operation of monitoring equipment. Be aware that the monitors in your care area may each have different alarm settings, to suitdifferent patients. Always check that the alarm settings are appropriate for your patient before youstart monitoring.AccessoriesWARNINGPhilips' approval: Use only Philips-approved accessories. Using other accessories may compromisedevice functionality and system performance and cause a potential hazard.Reuse: Never reuse disposable transducers, sensors, accessories and so forth that are intended forsingle use, or single patient use only. Reuse may compromise device functionality and systemperformance and cause a potential hazard.Electromagnetic compatibility: Using accessories other than those specified may result in increasedelectromagnetic emission or decreased electromagnetic immunity of the monitoring equipment.Damage: Do not use a damaged sensor or one with exposed electrical contacts.Cables and tubing: Always position cables and tubing carefully to avoid entanglement or potentialstrangulation.MR Imaging: During MR imaging, remove all transducers, sensors and cables from the patient.Induced currents could cause burns.Security InformationProtecting Personal InformationProtecting personal health information is a primary component of a security strategy. Each facilityusing the monitors must provide the protective means necessary to safeguard personal informationconsistent with country laws and regulations, and consistent with the facility’s policies for managingthis information. Protection can only be realized if you implement a comprehensive, multi-layeredstrategy (including policies, processes, and technologies) to protect information and systems fromexternal and internal threats.As per its intended use, the patient monitor operates in the patient vicinity and contains personal andsensitive patient data. It also includes controls to allow you to adapt the monitor to the patient's caremodel. To ensure the patient's safety and protect their personal health information you need a securityconcept that includes:15

1 Introduction Physical security access measures - access to the monitor must be limited to authorized users.It is essential that you consider physical security measures to ensure that unauthorized userscannot gain access. Operational security measures - for example, ensuring that patients are discharged aftermonitoring in order to remove their data from the monitor. Procedural security measures - for example, assigning only staff with a specific role the right touse the monitors.In addition, any security concept must consider the requirements of local country laws and regulations.Always consider data security aspects of the network topology and configuration when connectingpatient monitors to shared networks. Your medical facility is responsible for the security of thenetwork, where sensitive patient data from the monitor may be transferred.Note: Log files generated by the monitors and measurement modules are used for systemtroubleshooting and do not contain patient data.About HIPAA RulesIf applicable, your facility’s security strategy should include the standards set forth in the HealthInsurance Portability and Accountability Act of 1996 (HIPAA), introduced by the United StatesDepartment of Health and Human Services. You should consider both the security and the privacyrules and the HITECH Act when designing policies and procedures. For more information, pleasevisit http://www.hhs.gov/ocr/privacy/.About the EU DirectivesIf applicable, your facility’s security strategy should include the practices set forth in the Directive onthe protection of individuals with regard to the processing of personal data and on the free movementof such data (Directive 95/46/EC of the European Parliament and of the Council of24 October 1995). In addition, your facility should also take into account any additional, morestringent standards put forward by any individual EU countries; that is, Germany, France, and so on.For more information, please visit http://eur-lex.europa.eu/en/dossier/dossier 27.htm.Philips Product Security Policy StatementAdditional security and privacy information can be found on the Philips product security web site ufacturer Disclosure Statement for Medical Device Security –MDS2You can view the Manufacturer Disclosure Statements for Medical Device Security (MDS2) forspecific devices at ducing the MonitorThe IntelliVue MX400/MX450, MX500/MX550, and MX600/MX700/MX800 patient monitoroffers a monitoring solution optimized for the high-end surgical, cardiac, medical and neonatal careenvironments. Combining patient surveillance and data management, it allows multi-measurementmonitoring by linking separate modules. The MX600 uses the navigation knob as primary input device16

1 Introductionand the MX400/MX450, MX500/MX550, and MX700/MX800 use the touch screen as primary inputdevice. All monitors have a remote control for convenient access to the five main keys and numericdata input.The monitor stores data in trend, event, and calculation databases. You can see tabular trends (vitalsigns) and document them on a printer. You can view measurement trend graphs, with up to threemeasurements combined in each graph, to help you identify changes in the patient's physiologicalcondition. You can view fast-changing measurement trends with beat to beat resolution and see up tofour high resolution trend segments. Event surveillance enhances documentation and review ofphysiologically significant events by automatically detecting and storing up to 50 user-defined clinicalevents over a 24 hour period.MX600/700/ With the optional Integrated PC, you have computer functionality directly in the monitor. You can use800 standard applications (e.g. Web browsers), connect to the hospital network or intranet, and run asecond independent display with content from the patient monitor.An IntelliVue X2 can be connected to your monitor, where it acts as a multi-measurement module,acquiring measurements for the host monitor. When the X2 is disconnected from the original hostmonitor, it continues to monitor the patient as a fully independent, battery powered patient monitor,eliminating the need for a separate transport monitor. On connection to a new host monitor, the X2resumes its role as multi-measurement module, ensuring fully continuous monitoring.Major Parts and KeysMX400/450/500/550The MX400/450/500/550 monitors have the same parts, controls and indicators. Here the MX400 isshown.12345678Color coded alarm lampsAlarms Off lampPower on/Standby switch withintegrated LED: Green - On/Standby, Red - ErrorAC power LEDBattery LEDMounting quick-release lever(when this is pressed themonitor is not fixed on themounting)Service number and serialnumberDevice type17

1 IntroductionMX600/70012345678Color coded alarm lampsAlarms Off lampPower on/Standby switch withintegrated LED: Green - On/Standby, Red - ErrorAC power LEDMounting quick-release lever(when this is pressed themonitor is not fixed on themounting)Part number and serial numberHardkeys (Silence, Alarms Off,Main Screen)Navigation knobMX800123456Color coded alarm lampsAlarms Off lampPower on/Standby switch withintegrated LED: Green - On/Standby, Red - ErrorAC power LEDMounting quick-release lever(when this is pressed themonitor is not fixed on themounting)Part number and serial numberDevices for Acquiring MeasurementsThe patient monitor acquires patient measurements using the devices described in this section. Youcan also extend the measurement capabilities of your monitor with such devices. Of thesemeasurement devices, only the X2 has its own power on/standby switch, and can be powered from anexternal power supply or a rechargeable battery when not directly connected to the monitor (refer tothe IntelliVue X2 Instructions for Use for details). All the rest take their power exclusively from the18

1 Introductionmonitor, and switch on automatically when you turn on the monitor. A green power-on LED indicateswhen they are drawing power from the monitor. A permanently illuminated, or flashing, red LEDindicates a problem with the unit that requires the attention of qualified service personnel.All symbols used on the front panels are explained in “Symbols” on page 445.WARNINGWhen connecting devices for acquiring measurements, always position cables and tubing carefully toavoid entanglement or potential strangulation.Flexible Module Rack (M8048A)MX600/ The 8-slot flexible module rack (FMS-8) lets you use up to eight plug-in physiological measurementMX700/ modules. For the MX800, you can connect two FMSs to use up to 10 measurement modules.MX800 The maximum number of specific module types that can be used simultaneously in an FMS-8 is: fivepressure modules, four temperature modules, four VueLink or IntelliBridge modules (anycombination).When two FMSs are used, in total a maximum of 10 pressure modules can be used.Connect the FMS to the monitor via the measurement link cable (MSL). Use the MSL connector onthe left-hand side to connect an additional MMS. Use the connector on the right to connect to themonitor.12345X1 Multi-Measurement ModuleMulti-Measurement ModulemountFlexible Module Rack FMS-8Power on LEDInterruption indicator4-Slot Flexible Module Rack (FMS-4)MX600/ The 4-Slot flexible module rack (FMS-4) lets you use up to four plug-in physiological measurementMX700/ modules.MX80019

1 IntroductionThe maximum number of specific module types that can be used simultaneously in an FMS-4 is: fourpressure modules, four temperature modules, four VueLink or IntelliBridge modules (anycombination).Connect the FMS to the monitor via the measurement link cable (MSL). Use the MSL connector onthe left-hand side (if you have the appropriate option) to connect an additional MMS. Use theconnector on the back to connect to the monitor.Measurement ModulesMX500/ You can use up to three plug-in modules in the optional module slots. Available modules are:MX550 Invasive blood pressure (M1006B) Temperature (M1029A) Oxygen saturation of arterial blood (SpO2) (M1020B) Cardiac output (M1012A), and Continuous cardiac output with M1012A Option #C10 Intravascular Oxygen Saturation - ScvO2 or SvO2 (M1011A) Spirometry (M1014A) EEG (M1027A) NMT (865383) IntelliBridge EC10 (865115) Recorder (M1116B/C)MX600/ You can use up to eight measurement modules with the Flexible Module Rack (M8048A). AvailableMX700/ modules are:MX800 Invasive blood pressure (M1006B) Temperature (M1029A) Oxygen saturation of arterial blood (SpO2) (M1020B) Cardiac output (M1012A), and Continuous cardiac output with M1012A Option #C10 Transcutaneous gas (M1018A) Mixed venous oxygen saturation - SvO2 (M1021A) Intravascular Oxygen Saturation - ScvO2 or SvO2 (M1011A) EEG (M1027A) Bispectral Index - BIS (M1034A) Spirometry (M1014A) NMT (865383) VueLink device interface (M1032A) IntelliBridge EC10 (865115) Recorder (M1116B/C)MX500/550/ You can plug in and unplug modules during monitoring. Insert the module until the lever on the600/700/800 module clicks into place. Remove a module by pressing the lever upwards and pulling the module out.A measurement automatically switches on when you plug the module in, and switches off when you20

1 Introductionunplug it. Reconnecting a module to the same monitor restores its label and measurement settings,such as alarms limits. If you connect it to a different monitor, the module remembers only its label.The connector socket on the front of each module is the same color as the corresponding connectorplug on the transducer or patient cable.Press the Setup key on the module's front to display the measurement's setup menu on the monitorscreen. When the setup menu is open, a light appears above the key. Some modules have a second key.On the pressure module, for example, it initiates a zeroing procedure.Example Module (SpO2)MX500/550/600/700/8001Module name2Setup key LED3Setup key to enter setup menu of measurement modules orexternal device data window. Some modules have a secondmodule-specific key next to this one, for example Zero.4Connector socket for patient cable/transducerX1 Multi-Measurement Module (M3001A)The X1 Multi-Measurement Module (MMS) can simultaneously monitor 3-, 5-, 6- or 10-lead ECG(including arrhythmia and ST monitoring), respiration, SpO2, NBP and either invasive pressure ortemperature.You can connect it to the monitor via a cable or mount it on the left side of the FMS.21

1 IntroductionX1 Connectors and Symbols1White ECG/Resp connector2Blue SpO2 connector3Red NBP connector4Combined pressure (red) and temperature(brown) connector - connect either invasivepressure transducer or temperature probe.You might have a version of the MMS thatdoes not have this connector.5NBP STAT key - starts NBP STAT seriesof measurementsorZero key - initiates a zero procedure for theconnected pressure transducer whenpressed and held for a second6NBP Start/Stop key - starts or stops NBPmeasurements7Silence: acknowledges all active alarms byswitching off audible alarm indicators andlampsX2 Multi-Measurement Module (M3002A)The X2 Multi-Measurement Module (MMS) can simultaneously monitor 3-, 5-, 6- or 10-lead ECG(including arrhythmia and ST monitoring), respiration, SpO2, NBP and either invasive pressure andtemp

Changing Monitor Settings 47 Checking Your Monitor Revision 48 Getting Started 48 Disconnecting from Power 49 Networked Monitoring 50 Using the Integrated PC 50 Using Your Monitor with a Monitor in Companion Mode 52 2 What's New? 55 What's New in Release K.1 (for MX400/450/500/550 only) 55 What's New in Release K.0 55 What's New in Release J.0 .

Related Documents:

the IntelliVue patient monitor to be accessed and s hown on a PC which is connected to the IntelliVue patient monitor on an IntelliVue Clinical Network (ICN). This is intended for use as an additional, independent display for viewing and operation by clinicians and nurses. For additional information please refer to the connected patient monitor .

transport. The IntelliVue MX400 establishes device interfacing and connectivity to the EMR at virtually any point of care in the hospital. Along with these exciting features, the IntelliVue MX400 offers the same high standard of patient monitoring for which our IntelliVue family is known. IntelliVue MX400 patient monitor 5

5 Connection to IntelliVue X3 IntelliVue Dock 1. MSL connector for monitor connection 2. Flexible Sync Output connector, to provide signal to other medical devices 3. AC power connector 4. LAN connector for connection to Hemo workstation The IntelliVue Dock provides an external power supply for the monitor and extension when connected to mains .

The IntelliVue PC Printing Solution is a PC softwa re application which allows printer use and printer sharing for IntelliVue patient monitors on a LAN ne twork. You can print out reports generated by the patient monitor or store them as files in a choice of formats. You can print reports on A4, A3, Letter or Ledger (11x17in) paper format.

Integrated Monitor Theory of Operation The IntelliVue MP60/MP70 Patient Monitor: displays real-time data controls the attached measurement servers alarms in the case of patient or equipment problems offers limited data storage and retrieval (trending) interfaces to the Philips Clinical Network and other equipment

The Philips IntelliVue MX800 patient monitor offers a flexible and modular monitoring solution, designed to suit a broad spectrum of needs. The monitor can . The iPC is a fan-less, medical grade PC residing within the MX800 and as such designed for continuous operation in the patient vicinity. The iPC uses MS Windows 7 (or XP) as operating .

The IntelliVue Patient Monitor can generate three types of respiratory loops and display one real-time loop and up to six stored loops simultaneously. This assists in early detection of patient airway problems (for example, atelectasis, bronchospasm) and ventilator problems (for example, leaks and kinked tubes).

A First Course in Scientific Computing Symbolic, Graphic, and Numeric Modeling Using Maple, Java, Mathematica, and Fortran90 Fortran Version RUBIN H. LANDAU Fortran Coauthors: KYLE AUGUSTSON SALLY D. HAERER PRINCETON UNIVERSITY PRESS PRINCETON AND OXFORD