Validation Of Both Precision-and Accuracy Of Automated Liquid Handler .

1y ago
6 Views
2 Downloads
1,014.07 KB
5 Pages
Last View : 1m ago
Last Download : 3m ago
Upload by : Rosemary Rios
Transcription

Application Note AN-12 Validation of Both Precision-and Accuracy of Automated Liquid Handler Pipette Tips Cleaned with ‘Cold’ Atmospheric Plasma Introduction The life science industry has been developing and utilizing automated liquid handling systems for more than two decades. Originally used in clinical chemistry environments for assays of patient samples, automated systems are now commonly used in drug discovery environments to identify and validate genomic and proteomic targets and to develop corresponding therapies. Automated liquid handling (using the newer miniaturized microtiter plates, polypropylene pipette tips and stainless pins) results in lower per-well costs and enables many laboratories to run more assays and screening campaigns. Precision and Accuracy: Automated Liquid Handling Regardless of the type of tip (polypropylene or stainless) used, reagents must be delivered with both accuracy and precision. Often thought of as being synonymous, “accuracy” and “precision” are not interchangeable terms and do, in fact, mean very different things. Pipetting accuracy describes the deviation among a group of volumes from that of a standard volume. Pipetting precision describes the closeness of a group of volumes regardless of a standard (Figure 1). automated liquid handlers and their various pipetting tools. In general, gravimetric methods require (a) weighing an empty assay plate, (b) dispensing a volume of liquid per well and (c) taking a final weight. Although relatively simple, this process allows for identifying differences among plates, rather than on a well-to-well basis. Variances associated with a faulty syringe or plunger within an automated liquid handler or false weights due to high evaporation rates can skew overall results but not be individually identified. Consequently, this method is often reserved for single-channel pipetting devices. Single dye-based methods can provide a numeric approximation on a well-to-well basis providing precision data only. In comparison, a dual-dye photometric approach allows the user to independently calculate precision and accuracy data using one dye as an internal standard and the second as a sample. Figure 1. Visual representation of precision versus accuracy. 1a) precision 1b)accuracy 1c)represents precision and accuracy. There are two commonly used methods (gravimetric and dye-based) for determining precision and accuracy for 1 In this study, the effects of ‘cold’ atmospheric plasma on pipetting precision and accuracy for polypropylene pipette tips and stainless steel slotted pins were demonstrated using the highly standardized, patented dual-dye, dualwavelength MVS ratiometric absorbance method (ARTEL, Westbrook ME). www.ionfieldsystems.com

AN-12 Plasma Cleaning of Pipette Experimental FactorsTips for TipChager Validation The TipCharger System is provided in 8, 96 and 384-well plate The TipCharger by IonField Systems uses selfdensities compliant with standard SBS footprints. TipCharger contained, low-temperature, plasmawithin to the liquid can be taught as eitheratmospheric a device or consumable handler this study weand used polypropylene clean pipette tips,software. metalIn cannula pin tools tips Sciclone ALH liquid 3000 (Caliper Life Sciences, associated specific with to the automated handlers. Hopkinton MA), the TECAN EVO-MCA (The Tecan Group, Implementation of the TipCharger system provides Zurich SWITZERLAND), and the Biomek FX (Beckman cleaning equivalent to thatCA)ofas awellfresh setpinof tips, Coulter, Fullerton as 50nL tools mounted on a 96-well array (V&P Scientific,costs San Diego CA). The exterior reducing both the direct and indirect associated of each pipetting exposed to TipCharger-generated with replacing pipette tips.tool Inisthis study polypropylene plasma, while contaminants inside polypropylene pipet tips are tips specific removed to the through Sciclone ALH 3000 (Caliper Life a series of aspirate-and-dispense steps while Sciences, Hopkinton MA), the TECAN Figure 2. ARTEL Dual-Dye Photometric MVS Detection Method. tips are in the Cleaning Station. EVO-MCA (The Figure 2. ARTEL Photometric Detection Method. The Dual-DyeDual-Dye Ratiometric Photometry methodMVS employs two colorometric dyes Tecan Group, Zurich SWITZERLAND), and the Biomek (blue dye, with distinctRatiometric absorbance maxima at 520nm (red dye, ar) and 730nm The Dual-Dye Photometry method employs twoab). ARTEL Dual-Dye Photometric MVS Verification Method FX (Beckman Coulter, Fullerton CA) were used as well as Each sample solution contains twoabsorbance dyes. The red dyemaxima increases in at concentration, colorometric dyes with distinct 520nm The Dual-Dye Ratiometric Photometry method employs two col- while the blue dye concentration remains constant. 50nL pin tools mounted on a 96-well array (V&P (red dye, ar) and 730nm (blue dye, ab). orometric dyes with distinct absorbance maxima at 520nm (red) Scientific, Sanand Diego CA). 730nm (blue). MVS solutions include a series of sample solutions as well as a diluent and a baseline. Five sample soluwell. An automated liquid handler is used to dispense sample Precision and tions containing different concentrations of the red dye are Validating used solution and diluent intoAccuracy the wells of a microtiter plate, and for testing the performance of instruments over different volume the absorbanceExposure at both wavelengths is measured for every well. Post-TipCharger TipCharger Integration ranges from 10nL to 200µL. The concentration of blue dye is By applying the Beer-Lambert law, the MVS determines both Polypropylene tips (both 96 andof384 and stainless constant in all sample solutions across the volume ranges and is the precision and accuracy the formats) volume delivered by each Contaminants on the exterior of pipette tips and pin equal to that in the diluent. Therefore, the blue dye is usedsteel as an pins pipetting channel ofto theTipCharger-based automated liquid handler (Figure 2). were exposed atmospheric tools exposed to TipCharger-generated plasma are internal standard by which to calculate solution depth in each immediately ionized; contaminants on the inside of pipette tips are removed through a series of aspirate and dispense steps while the tips are in a cleaning station. The TipCharger cleaning stations are provided in 8, 96 and 384-well plate densities and can be taught as either a device or a consumable within the liquid handler software. plasma for a total of 30 seconds across a varying number of cycles. Post exposure,Validating the samePrecision pipettorsand were then used to Accuracy Post-TipCharger Exposure perform the appropriate ARTEL MVS volume verification Polypropylene tips (both 96 and 384 formats) (Table 1). and stainless steel pins were exposed to TipCharger-based atmospheric plasma for a total Table 1. ARTEL Dual Dye Photometric MVS aVerification Method of 30 seconds across varying number of cycles. Post exposure, the same pipettors were then Plasma Exposure for 96-well Caliper/Tecan Polypropylene Pipet Tips (5 μL) used to perform the appropriate ARTEL MVS volume verification (Table 1). ARTEL Dual-Dye Photometric MVS Verification Method MVS Diluent The Dual-Dye Ratiometric Photometry method employs two colorometric dyes with distinct absorbance maxima at 520 nm (red) and 730 nm (blue). MVS solutions include a series of sample solutions as well as a diluent and a baseline. Five sample solutions containing different concentrations of the red dye were used for testing the performance of instruments over different volume ranges from 10nL to 200 µL. The concentration of blue dye was constant in all sample solutions across the volume ranges and was equal to that in the diluent. Therefore the blue dye was used as an internal standard by which to calculate solution depth in each well. An automated liquid handler was used to dispense both the sample solution and diluent into the wells of a microtiter plate. By applying the BeerLambert law, to the absorbance values measured for both dyes, the MVS determines both the precision and accuracy of the volume delivered by each pipetting channel of the automated liquid handler (Figure 2). atmospheric plasma does not negatively impact either precision or accuracy for the various pipetting tools. In total, plasma exposure proMVS Diluent 100 μL vided equivalent - if not more accurate - liquid handling for each of the systems described. MVS range A (50 ml to 200 ml) 100 μL Overall, the standard deviations and CVs for plasma treated polypropylene tips(5and Plasma Exposure for 384-well Beckman Polypropylene Pipet Tips μL) pins were similar to that of untreated control sets across a wide range of plasma exposure MVS Diluent 50 μL cycles. 2 195 μL Results & Conclusion MVS range C (2.0 ml to 9.9 ml)results demonstrate that 5 μLTipCharger The Plasma Exposure for 96-well Caliper/Tecan Polypropylene Pipet Tips (100 μL) Caliper/Tecan 100µL Polypropylene Pipet Tips MVS range B (2.5 ml to 9.9 ml) 5 μL Plasma Exposure for 384-well Beckman Polypropylene Pipet Tips (25 μL) MVS Diluent 30 μL MVS range A (10 ml to 55 ml) 25 μL Plasma Exposure for 96-well V&P Stainless Steel Slotted pin Tools (50 nL) MVS Diluent 0.0 μL MVS range D (50 nl to 199 nl) 50 μL www.ionfieldsystems.com

AN-12 Results & Conclusion Exposure to TipCharger-generated plasma generally improved pipetting accuracy and had little effect on CVs for Caliper and Tecan 96-well automated liquid handling pipette tips. The ARTEL MVS showed that wells receiving a programmed target volume of 5 µL of Sample Solution C in reality received 4.3 µL on average (CV 3.8%). After a series of plasma exposures, tips aspirated and dispensed volumes more in line with the target volume, 4.7 µL (CV 7.6%). Similar results were observed relative to a 100 µL target volume using the same type of pipette tip (Figure 3). 25! Volume (uL)! Caliper/Tecan 100µL Polypropylene Pipette Tips 30! Biomek 25 µL polypropylene Tips ! (Target Volumes 5 µL and 25 µL)! "# %&!'()! * ,-%!.! )/&01/,! 20! 15! "*234!'()! * ,-%!"! )/&01/,! 10! 5! 0! 0! 50! 100! 150! 200! TipCharger Plasma Cleaning Cycles! Figure 4. Precision & Accuracy for 384-well Polypropylene Tips. Calliper/Tecan 100 µL polypropylene Tips (Target Volumes 5 µL and 100 µL)! 120! V&P Stainless Steel 50 nL Slotted Pins Volume )uL)! 100! "# %&!'()! * ,-%!.! )/&01/,! 80! 60! "*234!'()! * ,-%!"! )/&01/,! 40! 20! 0! 0! 25! 100! 200! TipCharger Plasma Cleaning Cycles! Figure 3. Precision & Accuracy for 96-well Polypropylene Tips. Accuracy for this tip type was virtually unaffected throughout 200 cycles. The slight increase in pipetting CVs was not considered to be significant when compared to the manufacturer specifications ( 5% CV). Nanoliter volumes dispensed for stainless steel slotted pins exposed to TipCharger-generated plasma were similar to those dispensed for untreated pin tools. The ARTEL MVS demonstrated that wells receiving a programmed target volume of 50 nL of Sample Solution D in reality received 53 nL on average (CV 18%). This slight difference in volume was attributed to suboptimal liquid handling practices and acted as the overall control. After an extensive series of plasma exposures, pins dispensed volumes of 53.1 nL (CV 10%). Overall, both precision and accuracy for this tip type was unaffected throughout 30,000 cycles (Figure 5) when compared to pin manufacturer specifications (7-8% CV). Beckman 25µL Polypropylene Pipette Tips 3 100! V&P Scientific Pin Tool Precision and Accuracy (Target Volume 5 nL)! 75! Volume (nl)! "# %&!'()! * ,-%.! )/&01/,! 50! 25! 0! 0! 10 0 20 ! 00 30 ! 00 40 ! 00 50 ! 00 60 ! 00 70 ! 00 80 ! 00 90 ! 0 10 0! 00 20 0! 00 30 0! 00 0! Exposure to TipCharger-generated plasma generally improved pipetting accuracy for low volumes versus untreated pipette tips with little effect on CVs using Beckman 384-well automated liquid handling pipette tips. The ARTEL MVS showed that control wells receiving a programmed target volume of 5 µL of Sample Solution B received 4.6 µL on average (CV 13.87%). After a series of plasma exposures, tips aspirated and dispensed volumes more in line with the target volume, 4.7 µL, with improved precision (CV 3.5%). Comparable results were observed with regard to a 25 µL target volume using the same type of pipette tip. Untreated tips pipetted 24.6 µL (CV 2.3%) while plasma treated tips pipetted on average 24.5 µL (CV 2.4%) throughout 200 cycles (Figure 4). TipCharger Plasma Cleaning Cycles! Figure 5. Precision & Accuracy for Stainless Steel Pins. www.ionfieldsystems.com

AN-12 Summary Regardless of the widespread adoption of disposable pipette tips, fixed cannula and pin tools, their use creates significant budgetary and logistical challenges for screening environments. The use of fixed tips and pin tools primarily allows for cost effective pipetting and assay miniaturization, but does require lengthy solvent-based wash procedures. The data shown in this study presents an additional option to screening environments. The data validates the premise that polypropylene tips can be repetitively cleaned with 30 second exposure to TipCharger-generated plasma and be reused at least 200 times prior to disposal with no detectable adverse effects on precision or accuracy of pipetting. It is important to note that because atmospheric plasma provides cleaning by altering surface chemistries, the length of the cleaning cycle within the TipCharger cleaning station determines the lifespan of the pipette tip. Therefore, a 200 cycle mark does not necessarily represent an upper limit for reuse since most cleaning cycles are less than 30 seconds on average. Since plasma exposure does not affect surface chemistries on stainless steel, either metal cannula or pins can, most likely, be exposed indefinitely. Automated environments can save on both direct and indirect costs associated with current practice with the implementation of novel technologies such as the TipCharger by IonField Systems. Given that life science screening environments continue to make every effort to reduce time-to-market for therapeutics, the TipCharger may help play a role achieving this objective. 4 www.ionfieldsystems.com

Integrating the TipCharger into Automated Assays Provides the Following Benefits: About the TipCharger Confidence The TipCharger Cleaning System replaces existing wash stations and easily integrates with most existing and new automation platforms. The system utilizes a low temperature, atmospheric pressure plasma process that cleans metal and plastic pipette tips and pin tools. Treated surfaces are clean, dry and have uniform surface properties. Cost Benefit Speed TipCharger cleans better than any other washing technology - in most applications the TipCharger will clean to background, so there is no difference between plasma cleaning and a new tip. TipCharger can save up to 98% on the cost of the disposable tips and extends the life of fixed tips. Incorporating the TipCharger System can result in a time savings of 10-30 seconds for every microplate processed or rack of tips cleaned. Convenience Clear away the clutter and save time: Integrating the TipCharger System eliminates the need to store cases of new pipette tips and dispose of racks of hazardous used tips. The TipCharger cleaning process reduces the incidence of microbubble formation and other random surface effects that degrade liquid handling precision and accuracy, even with new disposable tips. IonField Systems’ TipCharger improves the reproducibility of process results, shortens automation cycle times, reduces the number of lost runs, and eliminates environmental waste and liquid handling disposables. The overall result is increased confidence in results and a more effective and productive laboratory operation. TipCharger Plasma Cleaning Stations Available in 8, 96 and 384 channel versions USA and ROW Europe and Middle East IonField Systems 3411 Silverside Road Hanby Building Wilmington, DE 19810 USA IonField Systems Limited 7 Paynes Park, Hitchin Hertfordshire, SG5 1EH, UK Call: 1.302.442.6901 Call: 44 (0) 1462 429 727 Fax: 1.302.792.7407 Fax: 44 (0) 1462 429 798 Email: info@ionfieldsystems.com Email: info@ionfieldsystems.com www.ionfieldsystems.com www.ionfieldsystems.com Rev 6/11 About IonField Systems IonField Systems is an advanced technology company focused on the development of state-of-the-art pipette tip cleaning for life science laboratory research applications. IonField Systems provides on-site technical support services to assist laboratories in rapidly integrating the system into day-to-day operations. 2011 IonField Systems

precision and accuracy for this tip type was unaffected throughout 30,000 cycles (Figure 5) when compared to pin manufacturer specifications (7-8% CV). AN-12 3 www.ionfieldsystems.com Figure 3. Precision & Accuracy for 96-well Polypropylene Tips. Figure 4. Precision & Accuracy for 384-well Polypropylene Tips. Figure 5.

Related Documents:

Precision Air 2355 air cart with Precision Disk 500 drill. Precision Air 2355 air cart with row crop tires attached to Nutri-Tiller 955. Precision Air 3555 air cart. Precision Air 4765 air cart. Precision Air 4585 air cart. Precision Air 4955 cart. THE LINEUP OF PRECISION AIR 5 SERIES AIR CARTS INCLUDES: Seven models with tank sizes ranging from

Cleaning validation Process validation Analytical method validation Computer system validation Similarly, the activity of qualifying systems and . Keywords: Process validation, validation protocol, pharmaceutical process control. Nitish Maini*, Saroj Jain, Satish ABSTRACTABSTRACT Sardana Hindu College of Pharmacy, J. Adv. Pharm. Edu. & Res.

Pharmaceutical Engineers (ISPE) GAMP 5. Our validation service is executed in accordance with GxP standards producing a validation library that features the following documents: Validation and Compliance Plan The Validation and Compliance Plan (VCP) defines the methodology, deliverables, and responsibilities for the validation of Qualer.

heard. These goals relate closely to the Validation principles. Validation Principles and Group Work The following eleven axioms are the Validation Principles as revised in 2007. I have tried to find various ways of incorporating the principles into teaching Group Validation and by doing so, anchoring group work to theory. 1.

Validation of standardized methods (ISO 17468) described the rules for validation or re-validation of standardized (ISO or CEN) methods. Based on principles described in ISO 16140-2. -Single lab validation . describes the validation against a reference method or without a reference method using a classical approach or a factorial design approach.

Dipl.-Ing. Becker EN ISO 13849-1 validation EN ISO 13849-2: Validation START Design consideration validation-plan validation-principles documents criteria for fault exclusions faults-lists testing is the testing complete? Validation record end 05/28/13 Seite 4 Analysis category 2,3,4 all

ØExtent of validation and key parameters should be specified and justified in validation plan: e.g. accuracy, precision, stability etc. ØSpecific validation requirements and acceptance criteria may need to be established for each analyte Food and Drug administration. Bioanalytical method validation Guidance for industry.

Sandia National Laboratory, 23 July 2009. Burkardt Accuracy, Precision and E ciency in Sparse Grids. Accuracy, Precision and E ciency in Sparse Grids 1 Accuracy, Precision, . Burkardt Accuracy, Precision and E ciency in Sparse Grids. PRODUCT RULES: Pascal's Precision Triangle Here are the monomials of total degree exactly 5. A rule has