System Accuracy Evaluation Of FORA MD Blood Glucose Monitoring System .

1y ago
6 Views
2 Downloads
900.34 KB
10 Pages
Last View : 1m ago
Last Download : 3m ago
Upload by : Aydin Oneil
Transcription

System accuracy evaluation of FORA MD Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

Abstract Objective: The goal of the study was to assess the accuracy and reliability of Fora MD blood glucose monitoring system (BGMS) in the clinical studies according to the international standard ISO 15197: 2013(E) criteria and more stringent guidelines. Methods: The study sample was collected from 108 subjects comprised 53 patients with types 1 diabetes and 36 patients with types 2 diabetes, and 19 people with type unknown. The capillary blood specimens was collected via fingertip stick from participants at AMCR Institute (700 West El Norte Parkway, Suite 201, Escondido, CA 92026) and tested on three different lots of test strips. Results: l The Fora MD BGMS met the requirements of ISO 15197:2013 from all three lots of test strips tested. l Within 5 mg/dl & 5% Within 10 mg/dl & 10 % Within 15 mg/dl & 15 % (Within 0.28 mmol/l & 5 %) (Within 0.56 mmol/l & 10 %) (Within 0.83 mmol/l & 15 %) 144 / 200 (72.0 %) 192 / 200 (96.0 %) 197 / 200 (98.5 %) 142 / 200 (71.0 %) 195 / 200 (97.5 %) 197 / 200 (98.5 %) 139 / 200 (69.5 %) 192 / 200 (96.0 %) 198 / 200 (99.0 %) This BGMS is also compliant with the requirements of DIN EN ISO 15197:2003 with all 3 reagent system lot(s). Within 5 mg/dl & 5% Within 10 mg/dl & 10 % Within 10 mg/dl & 15 % Within 15 mg/dl & 20 % (Within 0.28 mmol/l & 5 %) (Within 0.56 mmol/l & 10 %) (Within 0.56 mmol/l & 15 %) (Within 0.83 mmol/l & 20 %) 142 / 200 (71.0 %) 192 / 200 (96.0 %) 194 / 200 (97.0 %) 200 / 200 (100 %) 136 / 200 (68.0 %) 193 / 200 (96.5 %) 196 / 200 (98.0 %) 200 / 200 (100 %) 137 / 200 (68.5 %) 192 / 200 (96.0 %) 196 / 200 (98.0 %) 199 / 200 (99.5 %) Conclusion: The study suggest that the Fora MD blood glucose monitoring system not only met the current accuracy standard (ISO15197:2003), but also met with the minimum acceptable accuracy standard published in the newer 2013 guideline (ISO15197:2013) when compared the test results with a laboratory reference. 2/10

1. Introduction This study was to evaluate the system accuracy following ISO 15197:20032 protocol and if possible, the newly published ISO 19197:20133 guideline. EN ISO 15197:2003 is an international standard published by International Standard Origination (ISO) specifying requirements for in vitro glucose monitoring systems on self-testing in managing diabetes mellitus. The minimum acceptable accuracy for results produced by a glucose monitoring system shall be as follows: Ninety-five percent (95 %) of the individual glucose results shall fall within 15 mg/dL ( 0.83 mmol/L) of the results of the reference measurement at glucose concentrations 75 mg/dL ( 4.2 mmol/L) and within 20 % at glucose concentrations 75 mg/dL ( 4.2 mmol/L). In 2013, ISO has published a newer standard ISO 15197:2013 (E) which tightens the accuracy acceptance criteria. The minimum acceptable accuracy for results produced by a glucose monitoring system has been changed to the followings: Ninety-five percent (95 %) of the individual glucose results shall fall within 15 mg/dL ( 0.83 mmol/L) of the results of the reference measurement at glucose concentrations 100 mg/dL ( 5.55 mmol/L) and within 15 % at glucose concentrations 100 mg/dL ( 5.55 mmol/L). Comparison between 2 ISO standards: EN ISO 15197: 2003 Maximum deviation from glucose reference value At glucose levels At glucose levels 75 mg/dl 75 mg/dl ( 4.2 mmol/l) ( 4.2 mmol/l) 15mg/dl (0.83 mmol/l) ISO 15197: 2013 (E) At glucose levels 100 mg/dl ( 5.5 mmol/l) 15mg/dl (0.83 mmol/l) 20% At glucose levels 100 mg/dl ( 4.2 mmol/l) 15% Parkes Error Grid Not Required 99% results within Zones A&B 2. Material and Methods 2.1 Reference standard l The reference used in this study was a Yellow Springs 2300 (YSI 2300) manufactured by YSI incorporated (Yellow Springs, OH) which detected glucose concentration based on a glucose oxidase method l The readings of the YSI analyzer was validated by running the analyzer with the National Institute of Standards and Technology (NIST) secondary reference material SRM 965a which consists of four levels of standard glucose concentrations. Glucose reference was determined as the average of at least duplicated measurements from the YSI. l 3/10

2.2 Preparation of Samples l Capillary blood samples were collected based on the ISO 15197:2013 guideline to have certain percentage of subject samples in the following concentration categories: Number of test subjects based on 100 Concentration category Percentage of samples [%] Glucose concentration [mg/dl (mmol/l)] 1 2 3 4 5 6 7 5 15 20 30 15 10 5 50 ( 2.77) 50 - 80 ( 2.77 - 4.44) 80 - 120 ( 4.44 - 6.66) 120 - 200 ( 6.66 - 11.10) 200 - 300 ( 11.10 - 16.65) 300 - 400 ( 16.65 - 22.20) 400 ( 22.20) 5 15 20 30 15 10 5 2.3 Test of Blood Glucose l Sample capillary blood was obtained by a finger stick using a 25G safety lancet. The blood sample was immediately collected onto a Lithium-Heparin Microvette ( 200 µl) l The capillary blood was then placed on a parafilm and tested on Fora MD blood glucose monitoring systems (6 devices per station with test strip lot 1, 2, and 3). The remaining blood in the lithium-heparin microvette was then immediately placed on a centrifuge to separate plasma. The glucose concentration of the plasma was then determined by the YSI 2300 using one electrode. The measurement was repeated once and the true glucose concentration of the subject glucose was determined by the average of the YSI values. 2.6 Statistical analysis In this study, the following statistical analyses were performed: l Bias analysis according to Bland and Altman5. l Regression analysis according to Passing and Bablok6. 3. Results 3.1 Test results 3.1.1 Accuracy results following ISO 15197:2013 (E) l Summary: Within Consensus Error Grid zones A and B 600 / 600 (100 %) l Conclusion: The Fora MD BGMS met the requirements of new ISO standard (ISO 15197:2013). 4/10

l Detail results: n System accuracy results for glucose concentrations 100 mg/dl (5.55 mmol/l) n n Within 5 mg/dl Within 10 mg/dl Within 15 mg/dl (Within 0.28 mmol/l) (Within 0.56 mmol/l) (Within 0.83 mmol/l) 40 / 64 (62.5 %) 61 / 64 (95.3 %) 64 / 64 (100 %) 44 / 64 (68.8 %) 64 / 64 (100 %) 64 / 64 (100 %) 46 / 64 (71.9 %) 62 / 64 (96.9 %) 64 / 64 (100 %) System accuracy results for glucose concentrations 100 mg/dl (5.55 mmol/l) Within 5 % Within 10 % Within 15 % 104 / 136 (76.5 %) 131 / 136 (96.3 %) 133 / 136 (97.8 %) 98 / 136 (72.1 %) 131 / 136 (96.3 %) 133 / 136 (97.8 %) 93 / 136 (68.4 %) 130 / 136 (95.6 %) 134 / 136 (98.5 %) System accuracy results for glucose concentrations between 37 mg/dl (2.06 mmol/l) and 474.5 mg/dl (26.4 mmol/l) Within 5 mg/dl & 5% Within 10 mg/dl & 10 % Within 15 mg/dl & 15 % (Within 0.28 mmol/l & 5 %) (Within 0.56 mmol/l & 10 %) (Within 0.83 mmol/l & 15 %) 144 / 200 (72.0 %) 192 / 200 (96.0 %) 197 / 200 (98.5 %) 142 / 200 (71.0 %) 195 / 200 (97.5 %) 197 / 200 (98.5 %) 139 / 200 (69.5 %) 192 / 200 (96.0 %) 198 / 200 (99.0 %) 3.1.2 Accuracy results following ISO 15197:2003 l Summary: Within Consensus Error Grid zones A and B 600 / 600 (100 %) l Conclusion: Fora MD BGMS also is in compliance with the requirements of DIN EN ISO 15197:2003 for all 3 reagent system lot(s). 5/10

l Detail results: n System accuracy results for glucose concentrations 75 mg/dl (4.2 mmol/l) n n Within 5 mg/dl Within 10 mg/dl Within 15 mg/dl (Within 0.28 mmol/l) (Within 0.56 mmol/l) (Within 0.83 mmol/l) 22 / 38 (57.9 %) 35 / 38 (92.1 %) 38 / 38 (100 %) 22 / 38 (57.9 %) 37 / 38 (97.4 %) 38 / 38 (100 %) 28 / 38 (73.7 %) 36 / 38 (94.7 %) 38 / 38 (100 %) System accuracy results for glucose concentrations 75 mg/dl (4.2 mmol/l) Within 5 % Within 10 % Within 15 % Within 20 % 120 / 162 (74.1 %) 157 / 162 (96.9 %) 159 / 162 (98.2 %) 162 / 162 (100 %) 114 / 162 (70.4 %) 156 / 162 (96.3 %) 159 / 162 (98.2 %) 161 / 162 (99.4 %) 109 / 162 (67.3 %) 156 / 162 (96.3 %) 160 / 162 (98.8 %) 162 / 162 (100 %) System accuracy results for glucose concentrations between 37 mg/dl (2.06 mmol/l) and 474.5 mg/dl (26.4 mmol/l) Within 5 mg/dl & 5% Within 10 mg/dl & 10 % Within 10 mg/dl & 15 % Within 15 mg/dl & 20 % (Within 0.28 mmol/l & 5 %) (Within 0.56 mmol/l & 10 %) (Within 0.56 mmol/l & 15 %) (Within 0.83 mmol/l & 20 %) 142 / 200 (71.0 %) 192 / 200 (96.0 %) 194 / 200 (97.0 %) 200 / 200 (100 %) 136 / 200 (68.0 %) 193 / 200 (96.5 %) 196 / 200 (98.0 %) 200 / 200 (100 %) 137 / 200 (68.5 %) 192 / 200 (96.0 %) 196 / 200 (98.0 %) 199 / 200 (99.5 %) 6/10

3.1.3 Consensus Error Grid for FORA MD with three reagent system lots Region A B C D E Number 599 1 0 0 0 Percent 99.83 % 0.17 % 0.00 % 0.00 % 0.0 7/10

l Summary of regression analysis (Passing and Bablok) n 200 Reagent system lot 1: TD14I615-CDC Reagent system lot 2: TD14I815-CDC Reagent system lot 3: TD14I915-CDC Regression Y -2.12 1.01X Y -3.63 1.01X Y -2.01 0.99X 8/10

3.1.4 Absolute differences between FORA MD and YSI 2300 STAT PLUS 9/10

4. Conclusion and discussion: This study showed that the Fora MD blood glucose monitoring system not only met the current accuracy standard (ISO15197:2003), but also exceeded the minimum acceptable accuracy standard published in the newer 2013 guideline (ISO15197:2013) when compared the test results with a laboratory reference. In addition, 99% of the data points fall within Zone A & B of Consensus Error Grid, which means that when user obtained a value using Fora MD monitor, it should not result in a clinical decision that could be deemed inappropriate. There are several glucose monitoring systems out in the market. Since FDA is in preparation of tightening the accuracy standard to 15% by adopting the new ISO15197:2013 standard. Based on the results, it will be reasonable assumption that Fora MD met the new ISO standard, and therefore would be a reasonable selection for users if they like to use meter with a higher accuracy. Reference a. 2013 International Diabetes Federation, the global burden. den b. In vitro diagnostic test systems - Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. EN ISO 15197:2003. c. In vitro diagnostic test systems - Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. ISO 15197:2013. d. Parkes JL, Slatin SL, Pardo S, Ginsberg BH: A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose. Diabetes Care 2000; 23:1143-1148. e. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 1:307-10. f. Passing H, Bablok. A new biometrical procedure for testing the equality of measurements from two different analytical methods. Application of linear regression procedures for method comparison studies in clinical chemistry, Part I. J Clin Chem Clin Biochem 1983; 21:709-20 10/10

In 2013, ISO has published a newer standard ISO 15197:2013 (E) which tightens the accuracy acceptance criteria. The minimum acceptable accuracy for results produced by a glucose monitoring system has been changed to the followings: Ninety-five percent (95 %) of the individual glucose results shall fall within 15

Related Documents:

Universidade Federal de Juiz de Fora, como requisito parcial para obtenção do título de Engenheiro de Produção. Aprovada em 12 de novembro de 2010. BANCA EXAMINADORA _ DSc, Fernando Marques de Almeida Nogueira Universidade Federal de Juiz de Fora _ DSc., Marcos Martins Borges Universidade Federal de Juiz de Fora

3.1.2 Accuracy results following ISO 15197:2013 (E) Summary: Conclusion: The FORA V30 BGMS also met the requirements of new ISO standard (ISO 15197:2013) Detail results: System accuracy results for glucose concentrations 100 mg/dl (5.55 mmol/l) Within 5 mg/dl Within 15 mg/dl (Within 0.28 mmol/l) Within 10 mg/dl

The FORA V10 monitor should only be used with FORA V10 Test Strips. Using other test strips with this meter can produce inaccurate results. Absorbent Hole Apply a drop of blood here. The blood will be automatically absorbed. Confirmation Window This is where you confirm if enough blood has

46701 CATERPILLAR (Cod. Original 1318822), Escavadeiras 320B 1CS-on/ 320BL 6CR1 / 320D Fora de Estrada 46742 CATERPILLAR (Cód. Original: 6I2509) Fora de Estrada 46745 CASE (Original 2005895C1) / JOHN DEERE Diversos (O.E: AH148880) Fora de Estrada 46774 Allis Chalmers, CAT

In 2013, ISO has published a newer standard ISO 15197:2013 (E) which tightens the accuracy acceptance criteria. The minimum acceptable accuracy for results produced by a glucose monitoring system has been changed to the followings: Ninety-five percent (95 %) of the individual glucose results shall fall within 15

GlucoMen LX PLUS: Accuracy Evaluations to ISO 15197:2013 with Specification and Technical Data July 2013 - 2 - Contents: Page 3. System accuracy evaluation of GlucoMen LX PLUS according to ISO 15197:2013 Introduction to ISO 15197:2013, Objectives of accuracy study, Method 5. Results and Conclusion - Accuracy, Bias Plot 6.

tion rate, evaluation use accuracy, evaluation use frequency, and evaluation contribution. Among them, the analysis of evaluation and classification indicators mainly adopts the induction method. Based on the converted English learning interest points, the evaluation used by the subjects is deduced for classification, and the evaluation list .

Both of the human genes involved have been cloned and gene therapy is of potential use in the treatment of both diseases. Cystic fibrosis is due to a mutation in a gene that codes for a chloride channel protein in the cell membranes of epithelial cells. This protein regulates the secretion of chloride ions from the epithelial cells. If the