Measuring Instruments Directive (MID)

8m ago
8 Views
1 Downloads
921.46 KB
18 Pages
Last View : 30d ago
Last Download : 3m ago
Upload by : Victor Nelms
Transcription

Measuring Instruments Directive (MID) MID/EN14154 Short Overview

STARTING POSITION Approval vs. Type examination In the past, country specific approvals were needed to sell measuring instruments in EU For the future, only one type examination certificate acc. the new Measuring Instrument Directive (MID) for the EU 26. Juli 2013

MEASURING INSTRUMENT DIRECTIVE Facts & Figures Scope includes. .water meters (MI-001) .gas meters (MI-002) .electricity meters (MI-003) .heat energy meters (MI-004) .etc. Needed for measuring instruments, where levying of consumption takes place within the EU All these measuring instruments are covered by the EU directive 2004 / 22 / EC Measuring instruments acc. MID calls from now. .«Product name» with CE Conformity (MID) Declaration of CE conformity

STARTING POSITION Goal The main goal for the definition and introduction of the new measuring instrument directive was the reduction of trade barriers within the European countries. Validation period Instruments approved before 30th October 2006 may continue to be sold for up to 10 years Country specific approvals Oct. 30th 2006 Overlapping period 10 years Oct. 30th 2016 Type examination acc. MID New meters have to comply as from 30th October 2006

Harmonized norms and certification EU members under the MID Austria Poland Belgium Portugal Bulgaria Slovenia Cyprus Slovakia Denmark Spain Estonia Sweden France United Kingdom Finland The Netherlands Germany Switzerland Greece Hungary Italy Ireland Latvia Lithuania Luxemborg Malta (Bilateral Agreements)

What is required of the manufacturer? All new meters require a approval of the instrument (Type examination certificate (CH-MI004-07003-00)) Audit of the production plant before instrument can be sold Must issue a conformity declaration Instrument must be marked in accordance with the directive

The MID Certification does away with the following: MID PTB Type approvals EEC approval Local European certification (country specific)

Marking of measuring instruments according to MID Number of the certified calibration centre M 08 CE - Marking 0862627 6 Metrology – Identification and manufacture year Manufacturer Serial Number

Markings MI - 001 Cold Water Meters

Markings MI-001 Consumption Meters Unit of measurement: cubic metre Numerical value of Q3 Ratio Q3 / Q1 , preceded by “R” Ratio Q2 / Q1 , where it differs from 1,6 (during transition period) Maximum admissible pressure if it differs from 10 bar (1 MPa) Direction of flow Letter V or H, if the meter can only be operated in the vertical or horizontal position Temperature class, where it differs from T30 Pressure loss class, where it differs from DP 63 Classes on sensitivity to irregularities in velocity field

Markings MI-001 Name or trademark of the manufacturer Year of manufacture (first 2 digits) and serial number Pattern approval sign according to European regulations Climatic and mechanical environment severity level EMC Class Output signals for ancillary devices (type/levels) if any

Main changes ISO to MID – MI-001 Definition of flow rate points The MID has mean comprehensive changes to the terminology and points which were commonly used to refer to its flow rate range New definitions for flow rate points: Q1 Q2 Q3 Q4 .Minimum flow rate .Transitional flow rate .Nominal flow rate .Maximum flow rate analogue to analogue to analogue to analogue to Qmin Qt Qn Qmax R .Measuring range analogue to class Also the relationship between the points has been newly defined Q2/Q1 1,6 Q4/Q3 1,25 Q3/Q1 R .acc. to a normed chart 26. Juli 2013 analogue to analogue to Qmax/Qn 2 Qn/Qmin

Main changes ISO to MID – MI-001 The MID has mean comprehensive changes to the terminology and points which were commonly used to refer to its FLOWRATE RANGE New definitions for flowrate points Q1 Q2 Q3 Q4 Minimum flowarte Transitional flowarte Nominal (Continuous) flowrate Maximum flowarte analog to analog to analog to analog to Qmin Qt Qn Qmax Also the relationship between the points has been newly defined Q2 / Q1 1.6 Q4 / Q3 1.25 analog to Qmax / Qn 2 Q3 / Q1 according to a normed R analog to Qn / Qmin z.B. 1:100 (Cl. C)

Dimensioning of Q1 - Q4 ISO Norm Metrological Class C, 1:100 Qmin Qt Qn Qmax Qmax 2*Qn Q4 Q4 1.25*Q3 New according to MID Measuring range R160 (Similar to 1:100) Q1 Q2 Q3

Maximum Permissible Error (MPE) (%) 5 Heated water 3 2 0 -2 -3 -5 Cold water Q1 Q2 Q3 Q4 Q Cold water Heated water

Overview of changes ISO to MID MI-001 New terms (according to MID) Old terms Qmin Qt Qn Qmax Measuring range Q1 Q2 Q3 Q4 R [l/h] [l/h] [m3/h] [m3/h] [1:x] [l/h] [l/h] [m3/h] [m3/h] [-] 15 24 1.5 3 100 15.6 25 2.5 3 160 25 40 2.5 5 100 25 40 4 5 160 35 58 3.5 7 100 39 58 6.3 7.8 160 60 96 6 12 100 62.5 100 10 12.5 160 100 160 10 20 100 100 160 16 20 160 Example A cold water meter Qn 2.5 Class C will be classified as follows according to MID: Q1 25 l / h; Q2 40l / h; Q3 4m3 / h; Q4 5m3 / h; Measuring range R Q3 / Q1 160 Attention On paper the meter seems to have a more dynamic measuring range! The meter stays the same!

Dimensioning a water meter. MI-001 Q3 is defined according to ISO 3 Row R5 Q1 is defined according to Q3 Q3 / R (ISO 3 Row R 10) Q2 is defined according to Q3 Q1 1,6 Q4 is defined according to Q4 Q3 1,25 Permanent Flow Q3 1n 1,6 2,5 4 6,3 10 16 25 40 63 100 160 250 400 630 1000 1600 2500 4000 6300 The numerical value for the permanent flowrate Q3 in m3/h is selected from the row R5 according to ISO 3. Measuring Range 10 12,5 16 20 25 31,5 40 50 63 80 The permanent flowrate range is defined with the Q3 / Q1 relationship. The values are selected from R10 form the ISO 3 table. 100 125 160 200 250 315 400 500 630 800

Schematic of Conformity Assessment Application Choice of Module Module B Type Examinations Module D Quality Assurances of production process QM Documentation Technical documentation Technical documentation Choice of examination Certification audit of QM system QM Certificate Evaluation report Declaration of conformity Examinations and tests EC Type examination certificate Marking CE M XX 1259 GWF ha both Module D and Module B for MID certified meters Surveillance audit Legend MID Module Notified Body Manufacturer

STARTING POSITION Goal The main goal for the definition and introduction of the new measuring instrument directive was the reduction of trade barriers within the European countries. Validation period Country specific approvals Oct. 30th 2006 Oct. 30th 2016 Type examination acc. MID Overlapping period 10 years New meters have to comply as from 30th October 2006

Related Documents:

(f.2) “Directive 050” means Directive 050: Drilling Waste Management; (g) “Directive 056” means Directive 056: Energy Development Applications and Schedules; (g.1) “Directive 058” means Directive 058: Oilfield Waste Management Requirements for the Upstream Petroleum Industry; (h) “Directive 060” means Directive 060: Upstream

MITSUBISHI METALWOOD CUSTOM SHAFTS OPTIONS mitsubishirayongolf.com Model Flex Weight Torque Tip Size Butt Size Launch Spin Tip Stiffness Fubuki J 60 X 66 3.9 0.335 0.600 Mid Mid Mid S 64 3.9 0.335 0.600 Mid Mid Mid R 61 3.9 0.335 0.600 Mid Mid Mid Fubuki J 70 X 74 3.6 0.335 0.600 Mid

or end measurements. In end measuring instruments, the measurement is taken between two end surfaces as in micrometers, slip gauges etc. The instruments used for linear measurements can be classified as: 1. Direct measuring instruments 2. Indirect measuring instruments The Direct measuring instruments are of two types: 1. Graduated 2. Non Graduated

Low Mid High Launch Spin Low Mid High Launch Spin KBS Hi-Rev 2.0 Wedge Flex R S X Tip.355" .355" .355" Weight (g) 115 125 135 Torque N/A N/A N/A Launch Mid Mid Mid Program Stock Stock Stock KBS TOUR 105 Flex R S X Tip.355" .355" .355" Weight (g) 105 110 115 Torque 2.5 2.5 2.5 Launch Mid-High Mid-High Mid-High Spin Mid-High M

In addition to the Pressure Equipment Directive (97/23/EC), there is the Simple Pressure Vessel Directive (87/404/EC, for vessels containing air and nitrogen), the Transportable Pressure Equipment Directive (99/36/EC), and the Aerosol Dispensers Directive (75/324/EEC). This publication provides an outline of the Pressure Equipment Directive only.

The EMC directive, Low Voltage directive, ATEX directive, and Measuring Instruments directive are . EN 50581: The path to RoHS-2 Compliance In order to provide companies with a consistent approach to applying the mandates of the EU directives, the EU publishes a

Directive 75/443/EEC UNECE Regulation 39.00 Speedometer and reverse gear Directive 70/311/EEC UNECE Regulation 79.01 Steering effort Directive 71/320/EEC UNECE Regulation 13.10 Braking Directive 2003/97/EC UNECE Regulation 46.02 Indirect vision Directive 76/756/EEC UNECE Regulation 48.03 Lighting installation

3006 AGMA Toilet Additive 1338 (3006) 19.0% 2914 CERAVON BLUE V10 DC (2914) 0.05% 2922 FORMALDEHYDE REODORANT ALTERNATIVE (2922) 0.6% 3 Water (3) 80.05% Constituent Chemicals 1 Water (3) 80.05% CAS number: 7732-18-5 EC number: 231-791-2 Product number: — EU index number: — Physical hazards Not Classified Health hazards Not Classified Environmental hazards Not Classified 2 Bronopol (INN .