BENCHTOP LCR METERLCR Meter LCR-1000 seriesProgramming Manual

Contents1. Overview . 12. Description to GPIB Common Commands . 2*RST . 2*IDN? . 2*TRG . 2FETC? . 2*SAV . 2*RCL . 23. Introduction to SCPI Commands . 33.1 Command Structure. 33.2 Parameter. 43.3 Command System . 43.3.1 SPEED . 43.3.2 DISPlay . 43.3.3 FREQuency. 53.3.4 LEVel. 53.3.5 PARAmeter. 53.3.6 EQUivalent . 53.3.7 SRESistor . 53.3.8 RANGe . 63.3.9 TRIGger . 63.3.10 CORRection . 63.3.11 COMParator . 63.3.12 LIMit . 73.3.14 HANDLER . 73.3.15 CALCULATE. 83.3.16 SYSTem. 83.3.17 PRINt . 93.3.18 COMP . 94. Format Commands . 104.1 Data Output Format . 104.2 String Format. 11I

1. OverviewThis manual provides users with guidelines in writing commands for remote control of the LCR meter. It isconsidered that readers of this manual have carefully read the User Operation Manual of this product and are familiarwith operation methods this product.The LCR meter communicates with PC via RS232, USB or GPIB interface. For details of communication methods,please refer to different chapters of this manual. After the LCR meter is reliably connected to PC, user can controlthe LCR meter via programming commands. The LCR meter supports SCPI commands and part of GIPB Commoncommands1

2. Description to GPIB Common Commands*RSTFunction: Reset the instrument to factory setup.*IDN?Function: Query the instrument ID and return a string (unique identification code of the instrument).Return: Company name, Instrument model number, Version number*TRGFunction: Trigger the instrument externally.FETC?Function: Perform with *TRG command. Query measurement data manually. If the data automatic transmission functionis turned on, no need to use this command.*SAVFunction: Save files.Command: numeric value ,[string]Description: numeric value is internal file numbers 1 to 105.[string] is fileExample:name, max.10 strings.*SAV 1,“AB”Remarks: When sending this command, if the existing file is covered, there is no notification about it.*RCLFunction: Upload the existing files.Commands: numeric value Description: numeric value is internal file numbers 1 to 105.Example:*RCL 12

3. Introduction to SCPI CommandsSCPI (Standard Commands for Programmable Instrument) is standard commands for programmable instrumentsbased on IEEE 488.2 common commands. SCPI commands include two parts: IEEE 488.2 Common Commandsand Control Commands defined for SCPI instruments.Common Commands, as defined by IEEE 488.2, are supported by the instrument. Syntax and semantics of commoncommands must follow the rules of IEEE 488.2. Common commands are generally used to control reset, self-testand status operations in stead of measurement. For more details, please refer to chapter IEEE 488.2 CommonCommands.Control Commands defined for SCPI instruments are used to measure, read data, switch ON/OFF a function or amode and so on. This chapter includes the following contents:3.1 Command StructureThe SCPI is in a tree structure. The command on the top is called “root command” or called “root” in short. Toreach command of the lowest level, user must follow its route.Command terminatorIt is a symbol to end the command. For example: NL (Line break symbol, as 10 in ASCIIcode).Colon (:)It means a command level, going to the next level.Semic olon (;)It means multiple levels of commands.Question mark (?)It means query.Comma (,)Space ( )It is a break between multiple parameters.It is a break between commands and parameters. A symbolis used to representsspace In description of commands.It means contents of the original citation, which will not be processed by commandanalysis programm.Commands follow an asterisk are common commands.Quotation mark (“”)Asterisk (*) The following figure shows how to reach command of the lowest level by using colon and semicolon.3

If sending command: AA:BB:EE;FF;GGIt means the same to send below three commands::AA:BB:EE:AA:BB:FF:AA:BB:GG3.2 ParameterCharacter data and string dataCharacter data is consisted of ASCII characters. String data is consisted of ASCII charcters with quotation mark(“”).Numercial dataNumerical data includes integer number (NR1), Fixed point number (NR2), and Floating point number (NR3). Thenumerical data range is 9.9E37.Examples of NR1:123 123-123Examples of NR2:12.3 1.234-123.4Examples of NR3:12.3E 5123.4E-563.3 Command System3.3.1 SPEEDFunction:Parameter:Return:Query or set test speed.FAST MEDium SLOWFAST MEDium SLOW3.3.2 y or set display page.MEASurement (Measurement display)BNUMber(Comparator function)MSETup(Measurement setup)SYSTem(System setup) MEAS DISP BIN DISP MEAS SETUP SYSTEM SETUP ABOUT INTER FILE LIST :DISPlay:RFONtFunction:Parameter:Return:Query or set font size. Query or set display ON/OFF.LARGe (Large font size)TINY (Small font size)OFF (Measurement data display off)ON(Measurement data display onOFF LARGE TINY4

:DISPlayFunction:Parameter:Return:Query or set measurement data display mode.DIRect (Direct reading)PERcent (Percentage deviation)ABSolute (Absolute deviation)DIRECT PERCENT ABSOLUTE3.3.3 FREQuency: FREQuencyFunction:Query or set test frequency.Parameter: 50 60 100 120 1k 10k (for LCR-1010)100 120 1k 10k 20k 30 (for LCR-1030)40 50 60 75 100 120 150 200 250 300 400 500 600 750 800 1k 1.5k 2k 2.5 3k 4k 5k 6k 7.5k 10k 12k 15k 15.7k 16.2k 20k 25k 30k 40k 50k 60k 66.6k 75k 100k 120k 150k 200k (for LCR-1030)Return:50 60 100 120 1k 10k (for LCR-1010)100 120 1k 10k 20k 30 (for LCR-1030)40 50 60 75 100 120 150 200 250 300 400 500 600 750 800 1k 1.5k 2k 2.5 3k 4k 5k 6k 7.5k 10k 12k 15k 15.7k 16.2k 20k 25k 30k 40k 50k 60k 66.6k 75k 100k 120k 150k 200k (for LCR-1030)3.3.4 LEVel: LEVelFunction:Parameter:Return:Query or set test level.0.1V 0.3V 1.0V0.1V 0.3V 1.0V3.3.5 PARAmeter: PARAmeterFunction:Query or set test parameters.Parameter: cd lq rq rd rx zd zrReturn:cd lq rq rd rx zd zr3.3.6 EQUivalent: EQUivalentFunction:Query or set equivalent shift of test parameters.Parameter: SERial PARallelReturn:SERIAL PARALLEL3.3.7 SRESistor: SRESistorFunction:Parameter:Return:Query or set source resistance.30 10030 1005

3.3.8 RANGe: RANGeFunction:Parameter:Return:Query or set range.AUTO HOLD 0 1 2 3 4 5AUTO- n HOLD- n n is number 0 5.3.3.9 TRIGger: TRIGgerFunction:Parameter:Return:Query or set trigger mode. Or immediately trigger for one time.INTernal(Internal trigger)EXTernal (External trigger)IMMediate (Immediately trigger for one time)INTERNAL EXTERNAL: TRIGger:DELayFunction:Parameter:Return:Query or set trigger delay time.0 6000 (ms)0 60003.3.10 CORRection: CORRectionFunction:Parameter:Clear data.OPENOPEN ALLSHORtSHORt ALL(Single point open circuit clearing)(All points open circuit clearing)(Single point short circuit clearing)(All points short circuit clearing)3.3.11 COMParator: COMParator[:STATe]Function:Parameter:Return:Query or set comparator status.ON OFF 1 01 0: COMParator:Auxiliary BINFunction:Parameter:Return:Query or set auxiliary status.ON OFF 1 01 0: COMParator:BIN CLEarFunction:Parameter:Return:Clear all upper and lower limits.ON OFF 1 01 06

: Query or set counter status.ON OFF 1 01 0: COMParator:COUNt:DATA?Function:Return:Query counter number of each range. count numbers of fail range , counter number of range 1 , counter number of range 2 , counternumber of range 3 , counter number of auxiliary range : COMParator:COUNt:CLEarFunction:Clear counter number of ll ranges.3.3.12 LIMit: LIMit:NOMinalFunction:Parameter:Return:Query or set nominal value. value value value is NR3 format.: LIMit:BIN n HIGH , LOW Function:Parameter:Return:Note:Query or set high and low limits of each range.n (1 3, in NR1 format.)high, low (-100 100, in NR3 format.) HIGH , LOW If there is no setting for high and low limits, the return value will be “9.9999e 37”.: LIMit:SECondary HIGH , LOW Function:Parameter:Return:Note:Query or set high and low limits of secondary parameters.high, low (in NR3 format.) HIGH , LOW If there is no setting for high and low limits, the return value will be “9.9999e 37”.3.3.14 y or set comparator output mode.CLEAr HOLD PULSeCLEAR HOLD PULSE: HANDler:PULSeFunction:Parameter:Return:Note:Query or set pulse width. numeric value MIN MAX numeric value numeric value is 1 9999. Unit is default at ms.7

: HANDler: EDGEFunction:Parameter:Return:Query or set trigger edge.RISing FALLingRISING FALLING3.3.15 rn:Query or set average time numbers.1 2551 r:Return:Query or set beeper source.MASTr EARPHone ALLMASTER EARPHONE Return:Query or set beeper for comparator pass.OFF LONG SHORT TwoSHORTOFF LONG SHORT eter:Return:Query or set beeper for comparator fail.OFF LONG SHORT TwoSHORTOFF LONG SHORT TWOSHORT3.3.16 urn:Query or set touch tone.OFF 0ON 10 1:SYSTem:SAVEFunction:Parameter:Note:Save file. numeric value ,[string] numeric value is internal file number 1 105.[string] is file name, with max.10 strings. When file name is omitted, the file is named automaticallywith file e:Load from existing files. numeric value numeric value is internal file number 1 105.SYST:LOAD 18

:SYSTem:RESetFunction:Reset system.3.3.17 PRINt:PRINtFunction:Parameter:Return:Note:Query or set automatic data transmission function.OFF 0ON 10 1When automatic data transmission is activated, the instrument automatically transmitsmeasurement data in the following format. A , B , COMP A : Measurement data of main parameters. B : Measurement data of secondary parameters. COMP : Measurement data of comparator results 0 5.3.3.18 COMP: COMParator: COMP:STATeFunction:Parameter:Return:Query or set comparator on/off.ON OFF 1 01 0: COMParator: COMP: MODEFunction:Parameter:Return:Note:Query or set comparator mode.1 01 01: ABS mode.: COMParator: COMP: LIMit:NOMinalFunction:Parameter:Return:Query or set nominal value of main parameters. numeric value (NR3 format) numeric value (NR3 format): COMParator: COMP: LIMit:A HIGH , LOW Function:Parameter:Return:Note:Query or set high or low value of main parameters. HIGH , LOW numeric value (HIGH), numeric value (LOW)If there is no high or low limits, the returned value will be “9.9999e 37”.In comparator mode %, the returned value range is -100 100.: COMParator: COMP: LIMit:B HIGH , LOW Function:Parameter:Return:Note:Query or set high or low value of secondary parameters. HIGH , LOW (NR3 format) numeric value (HIGH), numeric value (LOW)If there is no high or low limits, the returned value will be “9.9999e 37”.9

4. Format Commands4.1 Data Output FormatWhen writing commands, the “Symbol” in below table must be expressed in the way of ASCII 1Initialword {No special meaningrecognition2Indicationfor 0, 1, 2, 3, 4Main parameterSecondary parametermain/secondary0L: InductanceQ: Quality factorparameters.1C: CapacitanceD: Dissipation2R: ResistanceQ: Quality factor3R: ResistanceD: Dissipation4Other3Frequency0, 1, 2, 3, 4, 50: 10kHz1: 1kHz2: 120Hz3: 100Hz4: 60Hz5: 50Hz4Level0, 1, 20: 1V1: 0.3V2: 0.1V5Display mode0, 1, 20: % deviation 1: Direct reading 2: ABS deviation6Range0, 10: HOLD1: AUTO7Speed0, 1, 20: FAST1: SLOW2: MEDIUM8Clear0, 1, 2, 30: Short circuit clear1: Open circuit clear2: All clear3: Not clear9Beeper0, 10: ON1: OFF10Operation mode0, 10: Continuous 1: Single11Equivalent form0, 10: Series1: Parallel12Serial port0, 10: OFF1: ON13Comparator mode0, 1, 20: OFF0: Comparator P1: ON1: Comparator P32: Comparator OFF14Internal resistance0, 10:30 1:100 15 20Main parameter0 9, “.”, “-”From high to low21 26Secondary parameter 0 9, “.”, “-”From high to low27Unitofmain 0, 1, 2, %LCR/Zparameter, or %0uHpF 1mHnFk 2HuFM % Percentage deviation28Comparator output0, 1, 2, 3, 4, 5Comparator P3 Comparator P10NGNGD/QNG:badsecondaryparameter1Grade 1Grade 1 of P1PASS2Grade 2Grade 1 of P2HI: ExceedingHIGH limit3Grade 3Grade 1 of P3LO: ExceedingLOW limit4AUX failD/QNG&HI5No comparatorD/QNG&LO29The current range0, 1, 2, 3, 4, 50: 100k 1: 10k 2: 1k 3: 100 4: 31.6 5: 10 30Stop bit}No special meaning10

4.2 String FormatWhen writing commands, the “Command Code” in below table must be expressed in the way of ASCII code. Forexample, command string {A0} in ASCII code is 7B41307D. Only one of the following command can in containedin one string. Each command must be started with { and ended with }.No. Command CodeNumbers ofControl FunctionCharacters1A02Measure parameters, series and parallel: L-Q2PA12Measure parameters, series and parallel: C-D3A22Measure parameters, series and parallel: R-Q4A32Measure parameters, series and parallel: Z-r5B02Measure frequency: 10kHz6B12Measure frequency: 1kHz7B22Measure frequency: 120Hz8B32Measure frequency: 100Hz9B42Measure frequency: 60Hz10B52Measure frequency: 50Hz11C02Measure level: 1V12C12Measure level: 0.3V13C22Measure level: 0.1V14D02Display mode: %15D12Display mode: Direct reading16D22Display mode: ABS deviation17E02Hold in the current range.18E12Auto ranging.19E22Hold in 0 range 100k .20E32Hold in 1 range 10k .21E42Hold in 2 range 1k .22E52Hold in 3 range 100 .23E62Hold in 4 range 31.6 .24E72Hold in 5 range 10 .25F02Measurement speed: FAST.26F12Measurement speed: SLOW.27F22Measurement speed: MEDIUM.28G02Short circuit clear.29G12Open circuit clear.30G22Fixed point frequency short circuit clear.31G32Fixed point frequency open circuit clear.32H02Beeper: MASTER.33H12Beeper: EARPHONE.34I02Trigger: (OFF) Continuous.35I12Trigger: (ON) Single.36J02Equivalent format: (SER) series.37J12Equivalent format: (PER) parallel.38K02RS232 interface: (OFF) only receive data (single direction).39K12RS232 interface: (ON) Allow to send data (Bi-direction).40L02Comparator mode: P141L12Comparator mode: P342L22Comparator OFF43L32Comparator ON44M02Source resistance: 30 45M12Source resistance: 100 11

No.Command CodeNX XB1BXB2BXB3BXB4B46Numbers ofCharacters 10XB5BXB6BXB7BExamples and remarks:(Symbols are used in the followingexamples and remarks. Please change toASCII code when writing commands.)1) N1 122 The instrument adjusts it asN1 12.0002 (7 bit) after receiving.The last bit is code for unit. Forresistance measurement, nominalvalue here is 12 M . That is:N1 starts with numerical values andends with code of unit. Its first bitmust be 0 9, while other bits contain0 9 or decimal points. After adjustedby the instrument, it is less than 7 bit.2) N2 12 The instrument adjusts it asN2 12.000 (6 bit). That is:N2 (high and low values of D/Q)starts with numerical values andends with code of unit. Its first bitmust be 0 9, while other bits contain0 9 or decimal points. After adjustedby the instrument, it is less than 6 bit.3) N3 1% The instrument adjusts it asN3 1.000% (7 bit

This manual provides users with guidelines in writing commands for remote control of the LCR meter. It is considered that readers of this manual have carefully read the User Operation Manual of this product and are familiar with operation methods this product. The LCR met

Related Documents:

ELECTRICAL MEASUREMENT LABORATORY MODULE 2020 5 MODULE 2 IMPEDANCE MEASUREMENT I. OBJECTIVE 1. To know LCR Meter and its function 2. To know the construction of LCR Meter and how LCR Meter works II. BASIC THEORY LCR meter is an electronic electrical measurement to measure resistance, inductance and capacitance value.

LCR-2A PC INSTALLATION & OPERATING GUIDE BUNN-O-MATIC CORPORATION POST OFFICE BOX 3227 SPRINGFIELD, ILLINOIS 62708-3227 PHONE: (217) 529-6601 FAX: (217) 529-6644 To ensure you have the latest revision of the Operating Manual, or to view the Illustrated Parts Catalog, Programming Manual, or Service M

Local Rules for the Superior Court for Spokane County Effective September 1, 2022 Page 2 LCR 56 Summary Judgment LCR 69 Execution Superior Courts and Clerks LCR 77 Superior Courts and Judicial Officers LCR 79 Books and Records Kept by Clerk LCR 80 Court Reporters Local General Rules (LGR)

LCR Meter User and Service Manual. i PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT i Email: info@ietlabs.com . The DE-6000 is a portable, high-performance LCR meter that is full-featured yet cost effective. It measures in true 4-wire Kelvin . Impedance can be measur

LCR-6000 Series Firmware Update Manual 6 3. Plug in the USB drive and select the firmware file to update the meter. 6. Plug the USB drive that has the new firmware file in its root directory into the front panel USB port on the LCR-6000. 7. The meter will show you all the available firmware

the Cat 325C LCR utilizes the undercarriage of the 330C L, which allows for an extremely stable operating platform. A A A 2770 mm 1900 mm 3050 mm 320C L 321C LCR 325C L 325C LCR 320C L 325C L Tail swing radius (mm) 1900 2770 3050 A Overhang (mm) with 600 mm shoes 305 1280 1455 with 800 mm shoes – 1180 1355 with 850 mm shoes 180 – –

4w00-mm postage meter rentals meter for connect green 1000/2000 & sendpro p1000/2000 76.00 59.28 22% 4w00-mh postage meter rentals meter for connect green 3000 & sendpro p3000 100.00 78.00 22% 1r0t postage meter rentals 1r0t postage meter 137.00 106.86 22% 1h00-mh postage meter rentals meter for sendpro c400 385.00 300.00 22%

M20924891 Lopez, Jose Gerardo Citation Arraignment Filing Agency #: 20-604 M20920196 Lopez, Jose Manuel, JR Jail Release Arraignment Filing Agency #: 20-26090 M20924360 Lopez, Manuel Louis Surety Bond Jury Trial M20923837 Lopez, Michael Paul Citation Arraignment M17926273 Lopez, Raymond Fugitive Arraignment M19928461 Lopez, Robert A DPD: Public .