Validation Of Test Methods For Air Leak Rate Verification Of .

1y ago
8 Views
2 Downloads
886.30 KB
9 Pages
Last View : 21d ago
Last Download : 3m ago
Upload by : Mara Blakely
Transcription

https://ntrs.nasa.gov/search.jsp?R 20170007409 2019-08-29T22:18:22 00:00Zbrought to you byCOREView metadata, citation and similar papers at core.ac.ukprovided by NASA Technical Reports ServerVALIDATION OF TEST METHODS FOR AIR LEAK RATE VERIFICATION OFSPACEFLIGHT HARDWAREHeather A. OravecThe University of AkronAkron, Ohio, USAChristopher C. DanielsThe University of AkronAkron, Ohio, USAABSTRACTAs deep space exploration continues to be the goal ofNASAโ€™s human spaceflight program, verification of theperformance of spaceflight hardware becomes increasinglycritical. Suitable test methods for verifying the leak rate ofsealing systems are identified in program qualification testingrequirements. One acceptable method for verifying the air leakrate of gas pressure seals is the tracer gas leak detector method.In this method, a tracer gas (commonly helium) leaks past thetest seal and is transported to the leak detector where the leak rateis quantified. To predict the air leak rate, a conversion factor ofhelium-to-air is applied depending on the magnitude of thehelium flow rate. The conversion factor is based on either themolecular mass ratio or the ratio of the dynamic viscosities. Thecurrent work was aimed at validating this approach forpermeation-level leak rates using a series of tests with a siliconeelastomer O-ring. An established pressure decay method withconstant differential pressure was used to evaluate both the airand helium leak rates of the O-ring under similar temperatureand pressure conditions. The results from the pressure decaytests showed, for the elastomer O-ring, that neither the molecularflow nor the viscous flow helium-to-air conversion factors wereapplicable. Leak rate tests were also performed using nitrogenand argon as the test gas. Molecular mass and viscosity basedhelium-to-test gas conversion factors were applied, but did notcorrectly predict the measured leak rates of either gas. To furtherthis study, the effect of pressure boundary conditions wasinvestigated. Often, pressure decay leak rate tests are performedat a differential pressure of 101.3 kPa with atmospheric pressureon the downstream side of the test seal. In space applications, thedifferential pressure is similar, but with vacuum as thedownstream pressure. The same O-ring was tested at four uniquedifferential pressures ranging from 34.5 to 137.9 kPa. Up to sixcombinations of upstream and downstream pressures for eachdifferential pressure were compared. For a given differentialpressure, the various combinations of upstream and downstreamdry air pressures did not significantly affect the leak rate. Asexpected, the leak rate of the O-ring increased with increasingJanice L. MatherThe University of AkronAkron, Ohio, USAdifferential pressure. The results suggested that the current leaktest pressure conditions, used to verify spacecraft sealingsystems with elastomer seals, produce accurate values eventhough the boundary conditions do not model the spaceapplication.Keywords: conversion factor, helium leak detector, leak rate,permeation, pressure decay, verificationNOMENCLATURE๐‘Ž0 zero-order regression coefficient๐‘Ž1 first-order regression coefficient๐›ฝ bias errori,k indicesm massแน mass leak rate๐‘€ molecular mass๐œ‚ dynamic viscosityN number of samplesp absolute pressure๐œ™ precision error๐‘ž volumetric leak rateR specific gas constantT temperaturet timeU uncertaintyV volumeINTRODUCTIONThe success of any spaceflight mission is dependent on theverification program set forth to ensure that systems andcomponents meet the specifications and requirements defined atthe outset of the program. As NASA has its sights on deep spaceexploration, especially manned missions, verification ofhardware performance is imperative. Acceptable verificationmethods include analyses, demonstration, inspection, test, or anycombination thereof [1]. To verify the leak rates of pressurizedunits or sealing systems, testing is the verification method most1

commonly utilized. Acceptable leak test methods are outlined inprogram requirements [2,3]. For meaningful results, the selectedtest method must accurately simulate the functional conditionsof the unit under test following the test as you fly, fly as you testapproach [1]. For sealing systems, this may include theoperational temperature range, pressure conditions, gas type, andexposure to unique space environment elements (e.g., atomicoxygen, ultraviolet radiation, and foreign object debris). In someinstances, the test method is selected based on project resources,schedule, and feasibility and therefore, may not incorporate allaspects of the operating conditions. In other instances, the testmethod itself may not be capable of simulating the operatingconditions. In either case, flawed data resulting from poorlysimulated or misapplied test methods could be misinterpreted asa valid representation of the hardware performance. In this paper,the soundness of leak test methods used to verify the air leak rateof spacecraft sealing systems is investigated.Two common methods to evaluate the leak rate of gaspressure seals are the tracer gas leak detector method and thepressure change or pressure decay technique. These methodshave long been used to quantify the leak rates of siliconeelastomer seals used in habitat type space applications, e.g.,docking system seals [4โ€“10]. The tracer gas leak detector methodis often a preferred test method due to its sensitivity to smallleaks and relatively short test time. With this method, achemically inert tracer gas, typically helium, is used to pressurizethe volume upstream of the seal under test. The downstream sideis placed under vacuum and connected to a mass spectrometerleak detector. As the helium migrates downstream of the testarticle, it enters the mass spectrometer, which ionizes the gasparticles, and then separates the ions based on their molecularweight. A volumetric flow rate is then associated with thecollection of helium ions.For small leaks down to 10-12 sccs, the helium leak detectormethod is highly accurate and the test itself can be completed inshort order, although time and manpower are expended in thecalibration procedure and required post-processing of the testdata. Another drawback of this method is that for sealing systemsthat must perform to an air leak rate requirement (e.g., dockinghatch seals) the helium leak rate must be converted to a massflow rate of airโ€”a process that is not as obvious as it may seem.The conversion from volumetric to mass flow rate is trivial, andis achieved by multiplying the volumetric flow rate by thedensity of the test gas; however, the conversion from helium toair is more involved and often misapplied.Data sheets with conversion factors that can be applied to themeasured helium leak rate often accompany the documentationincluded with a helium leak detector [11]. In practice, helium isconverted to air by applying a factor of 0.374 to the measuredvolumetric leak rate following Eqn. 1:๐‘ž๐‘Ž๐‘–๐‘Ÿ ๐‘€๐ป๐‘’๐‘€๐‘Ž๐‘–๐‘Ÿ ๐‘ž๐ป๐‘’ ,(1)where, the square root of the molecular mass ratio of helium-toair is equal to the conversion factor (0.374 in this case). Thisvalue is derived from the volumetric flow rate of a gas throughan orifice in the molecular flow regime [12,13]. Similarconversion factors have been established based on viscous flowusing the ratio of dynamic viscosities (e.g., ๐œ‚๐ป๐‘’ ๐œ‚๐‘Ž๐‘–๐‘Ÿ ). Standardconversion factors for the viscous and molecular flow regimesare listed in Table 1 for some common gases.For a leak detected in a space sealing system, thedetermination of the flow regime can be challenging. Generally,the flow regime is estimated by the ratio of the average mean freepath to the diameter of the leak pathโ€”otherwise defined as theKnudsen number. This estimation assumes that a pore leak withphysical dimensions, such as a crack, hole, or other materialdefect is present within the sealing system [14]. It does notaccount for a leak dominated by gas permeation through the sealmaterial, as is the case with silicone elastomer space seals[15,16]. Currently, there is no universal conversion factor, basedupon theory, for the helium-to-air leak rate conversion ofpermeation dominant systems. This is due, in part, to thevariations in compounding different elastomers and the installedconfiguration of the elastomer in different systems. To date, nowork has compared the permeation dominated leak rates of airand helium to determine a ratio.Table 1. Standard helium leak rate conversion factors forviscous and molecular flow regimes [11].Multiply Helium Leak Rate by:Convert to:Viscous FlowMolecular sides the tracer gas leak detector method, anothercommonly accepted leak test method is the pressure decaytechnique with mass point leak rate analysis. The pressure decaytechnique does not require any conversions since air is utilizedas the test gas. In this method, a known volume of pressurizedgas permeates through and/or leaks at the interface of the sealunder test while the pressure and temperature are recorded withtime [17]. The Ideal Gas Law is applied to calculate the mass ofthe gas (mi) in the known volume at each time-step (ti). A linearleast-squares regression, centered about the differential pressureof interest, is used to find the best-fit line to the mass-timedataset. The mass flow rate of the test article is defined by thefirst-order coefficient (๐‘Ž1 ) of the best-fit line.Though the pressure decay method is low-cost and applicableto an extensive range of leak rates, it has several limitations thatreduce feasibility. First, the size of the internal volume and themagnitude of the leak, among other factors, determine the testduration. The combination of large volume and small leak mayresult in a test that spans several days costing time andmanpower. Conversely, large leaks from a small volume mayoccur in such a short time that the limited amount of collecteddata leads to large measurement uncertainty. Second, althoughthe method accounts for the temperature of the gas in thecalculation of the leak rate, the location of the measurement iscritical and correct sensor placement can be difficult in certain2

testing environments.Additionally, leak tests are typically performed at adifferential pressure of 101.3 kPa. To achieve this, the lowpressure region is either placed under vacuum or the ambient airpressure is utilized. Ambient pressure is more commonly used toavoid the inclusion of a secondary seal in the hardware design.However, when ambient pressure is used, changes in barometricpressure may extend the test duration and/or invalidate the leakrate measurement. More critically, the ambient downstreampressure does not simulate the vacuum of space. It is well knownthat differential pressure is a driving factor for the final leak ratevalue, but there are no reports on the effect of varying theupstream and downstream pressure boundary conditions whichgenerate the differential pressure.As such, the purpose of this study was twofold, but focusedon the verification of a spacecraft sealing systemโ€™s air leak rateusing the helium leak detector method and the pressure decaymethod. The first objective was to prove or disprove that theconversion of the measured helium leak rate to an equivalent air(or other gas) leak rate could be accomplished using standardconversion factors. The second objective was to investigate theeffect of pressure boundary conditions on the air leak rate whenusing the pressure decay method of testing. Using a modifiedpressure decay method, a series of leak tests was performed on asilicone elastomer O-ring. All tests in this series were completedat or near room temperature and with a differential pressure of101.3 kPa, applying vacuum downstream of the test article. Fourdifferent gases including helium, nitrogen, dry air, and argonwere utilized in this series. From the experimental data,conversion factors were computed for helium to each test gas andcompared to the industry standards for viscous and molecularflow. A second series of leak tests was conducted at four uniquedifferential pressures ranging from 34.5 to 137.9 kPa. Dry airwas the test gas for all tests in this series. For each differentialpressure, up to six combinations of unique upstream anddownstream pressures were tested and the final leak rate valueswere compared. The experimental set up, methodology, and testresults for each test series are discussed herein.DESCRIPTION OF EXPERIMENTSIn this study, two series of leak rate experiments wereperformed using a modified pressure decay method. Four testsets were completed in the first series: one set for each test gas.Four repeats per test set were performed for a total of sixteentests. Eighteen tests were completed in the second series with norepeats. Most tests were conducted at ambient room temperature(20 C); however, some were run within an environmentalchamber controlled to 23 C. The slight difference in temperaturewas accounted for in the data analysis and did not affect the finalresults. Additional specifics of the experimental setup and testmethodology are detailed in the following sections.Test ArticleA single test article was evaluated throughout this study. Thetest article was made from high temperature siliconemultipurpose O-ring cord stock. The cord stock had a nominalcross-sectional diameter of 9.5 mm and durometer Shore Ahardness of 70. The cord stock was cut to length and the two endsbonded with Loctite Superflex Clear RTV Silicone Sealant(#59530) to form one continuous test article. The test article hada nominal 30.5-cm outer diameter and was sized to fit in thegroove of a custom aluminum test fixture. Once installed, the testarticle remained in the test fixture, physically undisturbed, forthe duration of the study.Test GasesIn the first test series, a total of four different inert gases wereused to evaluate the leak rate of the test article. These includedhelium, nitrogen, dry air, and argon. Dry air was used as the testmedium to evaluate the leak rate under various pressureboundary conditions in the second test series. The specificationsfor all test gases are listed in Table 2.Table 2. Test gas -99.999%Ar--O2 2 ppm 2 ppmMoistureTHCCO2CO 2 ppm 0.5 ppm--- 3 ppm 0.5 ppm---Dry Air---19.523.5% 3 ppm 0.5 ppm 1 ppm 1 ppmArgon--99.998% 5 ppm 5 ppm 2 ppm---Pressure Boundary ConditionsIn the first test series, all tests were run with a differentialpressure of 101.3 kPa. In general, the upstream pressure was 1.3times atmospheric pressure. Vacuum was applied downstream ofthe test article.To determine the effect of pressure boundary conditions onthe leak rate of the test article in the second series, four individualtest sets were runโ€”each set with a unique differential pressure.Within each test set, the differential pressure was held constant,but the initial upstream pressure was varied per test. Thedifferent pressure combinations for each test set are shown inTable 3. These pressure conditions were selected to be within thelimitations of the measurement transducers used in the testassembly.Table 3. Pressure boundary conditions for dry air leak ratetests on the silicone elastomer O-ring test article.Set 1 Set 2 Set 3 Set 4Differential pressure, kPa34.568.9 103.4 137.968.9---103.4 103.4--Initial upstream pressure,137.9 137.9 137.9-kPa172.4 172.4 172.4 172.4206.8 206.8 206.8 206.8241.3 241.3 241.3 241.33

Test MethodA modified pressure decay method was used to measure theleak rate of the test article in this study. This method was similarto the standard pressure decay method with mass point leak rateanalysis, but used a control system to maintain the desireddifferential pressure across the test article. Previous work hasshown that this enhanced method is accurate, reliable, can beused to measure both large and small leaks, minimizes test time,and improves the measurement uncertainty [18].In this method, the test apparatus, Fig. 1, consisted of ahermetically sealed volume of gas on the upstream side of thetest article. The pressurized volume of gas was allowed to leakdownstream of the test article into a region of lower pressure.This low-pressure region was controlled to maintain a constantdifferential pressure across the test article throughout the testduration. A differential pressure transducer was used to measurethe pressure difference between the high- and low-pressureregions. A controller monitored the differential pressure andcompared it to the chosen set point value. As the differentialpressure varied from the set point, due to permeation or interfaceleakage, the controller reacted by sending a voltage signal to apressure regulator. The pressure regulator appropriately raised orlowered the downstream pressure through connections ofvacuum and ambient pressure. In cases where the downstreampressure was above ambient pressure (refer to Table 3), theregulator was connected to a gas supply system.In this equation, the volume, V, was determined in advancethrough application of Boyleโ€™s Law.Assuming a constant leak rate, a linear least-squaresregression was computed to determine the best-fit line to thedataset. Unlike the standard pressure decay method, all datacould be included in this computation since a constantdifferential pressure was maintained. The best-fit line wasmodeled by Eqn. 3, where the first-order coefficient, ๐‘Ž1 ,represented the mass leak rate (แน) of the test article.๐‘š(๐‘ก) ๐‘Ž1 ๐‘ก ๐‘Ž0(3)The measurement uncertainty of the leak rate was calculatedusing the generalized Eqn. 4 [17,19].2๐‘๐‘ˆ๐‘šฬ‡ ๐‘– 1 ( ๐‘šฬ‡ 2 ๐‘šฬ‡ 2 ๐‘š๐‘– ๐‘ก๐‘– ๐‘๐‘– 1 (2) ๐›ฝ๐‘š ๐‘๐‘– 1 (๐‘–) ๐›ฝ๐‘ก2๐‘– ๐‘šฬ‡ 2 ๐‘šฬ‡ 2 ๐‘š๐‘– ๐‘ก๐‘–2) ๐œ™๐‘š ๐‘๐‘– 1 (๐‘–๐‘ 2 ๐‘ 1๐‘– 1 ๐‘˜ ๐‘– 1 ( ๐‘šฬ‡ ๐‘š๐‘–๐‘ 2 ๐‘ 1๐‘– 1 ๐‘˜ ๐‘– 1 ( ๐‘šฬ‡๐‘ 2 ๐‘ 1๐‘– 1 ๐‘˜ ๐‘– 1 ( ๐‘šฬ‡ ๐‘ก๐‘– ๐‘ก๐‘–)()()( ๐‘šฬ‡ ๐‘š๐‘˜ ๐‘šฬ‡ ๐‘ก๐‘˜) ๐œ™๐‘ก2๐‘–) ๐œ™๐‘š๐‘– ๐‘š๐‘˜ ๐›ฝ๐‘š๐‘– ๐›ฝ๐‘š๐‘˜(4)) ๐œ™๐‘ก๐‘–๐‘ก๐‘˜ ๐›ฝ๐‘ก๐‘– ๐›ฝ๐‘ก๐‘˜ ๐‘šฬ‡ ๐‘š๐‘˜) ๐œ™๐‘ก๐‘–๐‘š๐‘˜ ๐›ฝ๐‘ก๐‘– ๐›ฝ๐‘š๐‘˜Assuming no errors in the measurement of time, and usingcorrelation coefficients that produce maximum uncertainty [19],the previous equation can be reduced to Eqn. 52๐‘๐‘ˆ๐‘šฬ‡ ๐‘– 1 ( ๐‘šฬ‡ 2 ๐‘šฬ‡ 2 ๐‘š๐‘– ๐‘š๐‘–2) ๐›ฝ๐‘š ๐‘๐‘– 1 (๐‘–2) ๐œ™๐‘š๐‘–(5)The partial derivative of แน with respect to mi is: ๐‘šฬ‡๐‘๐‘ก๐‘– ๐‘๐‘– 1 ๐‘ก๐‘– ๐‘2 ๐‘š๐‘– ๐‘ ๐‘– 1(๐‘ก๐‘–2 ) ( ๐‘๐‘– 1(๐‘ก๐‘– ))Figure 1. Schematic of leak rate test apparatus withcontrolled downstream pressure (in cases where downstreampressure was above ambient, the regulator was connected toa gas supply system instead of vacuum).Not unlike the standard pressure decay method, the pressureand temperature of the gas in the sealed volume were recordedwith time. Following the assumptions of the Ideal Gas Law, themass of gas within the volume was calculated at each time-step(ti, mi) using Eqn. 2.๐‘š ๐‘๐‘‰/๐‘…๐‘‡(2)(6)And the bias and precision errors, respectively, are:2๐›ฝ๐‘š (2๐œ™๐‘š (2๐‘‰๐‘…๐‘‡๐‘‰๐‘…๐‘‡๐‘๐›ฝ๐‘ ) (๐‘…๐‘‡2๐‘๐œ™๐‘ ) (๐‘…๐‘‡2๐›ฝ๐‘‰ ) (2๐‘๐‘‰๐‘…2๐‘‡๐œ™๐‘‰ ) (๐‘๐‘‰๐‘…2๐‘‡2๐›ฝ๐‘… ) (2๐‘๐‘‰๐‘…๐‘‡ 2๐œ™๐‘… ) (๐‘๐‘‰๐‘…๐‘‡ 2๐›ฝ๐‘‡ )2๐œ™๐‘‡ )(7)2(8)As shown, the bias and precision errors include contributionsfrom the measurement instruments which were obtained thoughthe instrumentsโ€™ calibration records, product specifications, orcomputations.4

Using the modified pressure decay method, the test articleโ€™sleak rate and measurement uncertainty were calculated in realtime.Test ApparatusThe complete test apparatus consisted of the test fixture withhermetic plumbing, gas supply system, measurementinstrumentation, differential pressure control system, and dataacquisition (DAQ) hardware and associated software. The testfixture consisted of two clear anodized platens manufacturedfrom 6061-T6 aluminum. The test article was installed into arecessed grove in the bottom platen, constrained only along theouter diameter of the O-ring, Fig. 2. The O-ring was free to moveinward, however, once the upper platen was installed and theinterior volume was pressurized, movement in this direction wasnot anticipated. The upper platen was installed onto the lowerplaten compressing the O-ring by 17% of its nominal crosssectional diameter.Figure 2. Schematic of leak rate test fixture cross-section.The test gas was supplied to the high-pressure side of thetest apparatus to the desired initial pressure ranging from 68.9 to241.3 kPa, refer to Pressure Boundary Conditions. The lowpressure side was controlled to achieve the desired differentialpressure set-point value which ranged from 34.5 to 137.9 kPa. Asecondary O-ring of larger inside diameter was installed in thetest fixture, concentric to the test article, such that the pressuredownstream of the test article could be increased or reduced asnecessary.To determine the mass of gas at each time-step (Eqn. 2), themeasured pressure, volume, and temperature were required. Thegas pressure in the high-pressure region was measured using twopressure transducers whose values were averaged by the dataacquisition system. This average value was used in the dataprocessing. For reference, typical bias and precision errors of thepressure transducers were 15.6 Pa and 12.0 Pa, respectively.The volume of the high-pressure region changed over thecourse of the study due to slight modifications in the fixtureplumbing. For each modification, the volume was directlymeasured using a minimum of 31 applications of Boyleโ€™s Lawwhere ๐‘1 ๐‘‰1 ๐‘2 ๐‘‰2. The total volume changed from234.3 3.8 mL to 286.2 6.6 mL depending upon theconfiguration. For the corresponding tests, the appropriatevolume was used in the computations for leak rate and did notaffect the overall results of the study.The temperature of the gas in the high-pressure region wasindirectly measured using a resistance temperature detector(RTD). The RTD was placed on the upper platen and insulatedwith a foam block to minimize changes in the temperaturereadings due to laboratory conditions. The RTD had Class Aaccuracy and typical bias and precision errors of 0.196 C and0.0225 C, respectively. Recall that some tests were run in anenvironmental chamber controlled to 23 C. Other tests wereconducted in the ambient laboratory environment. For thesetests, the temperature reading did not vary by more than 2.1 Cper test, which negligibly impacted the results. Therepresentative temperature for the ambient laboratory tests was20 C.The data acquisition system consisted of signal conditionersand an associated computer software program. The DAQ wasused to collect the pressure and temperature measurements at anominal rate of 10 Hz. These values were combined with thevolume measurement in the computer software program tocalculate the mass of the test gas at each time-step. The softwarealso calculated the test articleโ€™s leak rate and associatedmeasurement uncertainty in real-time. In general, each test rancontinuously for a maximum duration of 29 hours unlessotherwise manually stopped.TEST RESULTS AND DISCUSSIONThe results of this study are discussed in the followingsections. The first section summarizes the results of the first testseries investigating the applicability of using standard heliumto-air conversion factors for verifying air leak rates. The secondsection presents the findings of the second test seriesinvestigating the effect of pressure boundary conditions used inthe pressure decay method.Validation of Conversion FactorsIn the first test series, the internal volume of gas waspressurized to approximately 1.3 times atmospheric pressure,and the downstream pressure was controlled to maintain aconstant differential pressure of 101.3 kPa. The leak rates ofhelium, nitrogen, dry air, and argon through the siliconeelastomer test article were compared. For each gas, the leak testwas repeated four times. The test results were highly repeatableproviding confidence in the test method, Fig. 3. The argon resultsdisplayed the greatest scatter with a maximum difference of6.7x10-12 kg/s between repeat tests. The average mass leak ratevalues for each gas are plotted in Fig. 4. The error bars representthe measurement uncertainty. As shown, the leak rate increasedwith the molecular mass of the test gas.The leak rate of the O-ring was also measured using ahelium leak detector. The average volumetric leak rate,calculated from four repeat tests, was 1.03x10-4 sccs ( 5.7%).This value was converted to a mass flow rate of 1.84x10-11 kg/s.The average helium leak rate measured using the modifiedpressure decay method, Fig. 4, was 1.81x10-11 kg/s, a differenceof 1.6%. This comparison provided an additional level of5

confidence in the measured results using the modified pressuredecay test method.shown in Table 4 with the standard viscous and molecular flowregime conversion factors for comparison. As can be seen, theexperimental conversion factors did not align with the standardvalues for either flow regime.Table 4. Experimental helium-to-test gas conversion factorscompared to standard conversion factors for viscous andmolecular flow regimes.Multiply Helium Leak Rate by:Convert Figure 3. Repeat leak test results for a silicone elastomerO-ring tested with four different gases using a modifiedpressure decay method. Error bars represent measurementuncertainty.Furthermore, the standard viscous and molecular flowfactors for each gas were applied to the measured volumetricflow rate of helium to calculate the projected nitrogen, air, andargon leak rates. Figure 5 displays the experimental leak ratevalues measured for each gas compared to the projected values.No comparison was needed for helium-to-helium; therefore,only one bar is shown. As expected, neither the application ofthe viscous flow factor nor the molecular flow factor to themeasured helium leak rate correctly predicted the nitrogen, dryair, nor argon leak rates of the test article. When molecular flowwas assumed, the converted leak rates consistentlyunderpredicted the leak rate of the test article. When viscousflow was assumed, the converted leak rates for nitrogen and dryair were overpredicted, but the leak rate for argon wasunderpredicted.Figure 4. Average experimental leak rate of a siliconeelastomer O-ring for four different gases, measured using amodified pressure decay method. Error bars representmeasurement uncertainty.The average mass flow rates of helium, nitrogen, dry air, andargon from the modified pressure decay tests were converted tovolumetric flow rates at normal temperature and pressure (NTP:20 C, 101.3 kPa). For example, the average mass flow rate ofhelium (1.81x10-11 kg/s) was converted to a volumetric flow rateof 1.09x10-4 cm3/s (NTP). Experimental conversion factors (testgas-to-helium volumetric flow ratios) were computed and areFigure 5. Experimental average volumetric leak rates at NTPcompared to projected leak rates of nitrogen, dry air, andargon computed by applying standard conversion factors formolecular and viscous flow to the measured helium leak rate.6

In a similar fashion, the average volumetric leak rates fornitrogen and argon were converted to air leak rates in the viscousand molecular flow regimes. The standard nitrogen-to-air andargon-to-air conversion factors applicable for the molecular flowregime were calculated using the radical in Eqn. 1 with themolecular mass of nitrogen and argon substituted for that ofhelium (e.g., ๐‘€๐ด๐‘Ÿ ๐‘€๐‘Ž๐‘–๐‘Ÿ ). The standard conversion factors forthe viscous flow regime were calculated using the ratios of thedynamic viscosities of nitrogen-to-air and argon-to-air (e.g.,๐œ‚๐ด๐‘Ÿ ๐œ‚๐‘Ž๐‘–๐‘Ÿ ). Figure 6 shows the experimentally measured dry airleak rate compared to the projected air leak rate values computedby applying the standard conversion factors to the measuredhelium, nitrogen, and argon leak rate values. No conversion wasneeded for air-to-air; therefore, only one bar is shown. Theconversion from helium-to-air in the viscous flow regime mostclosely represented the measured air leak rate, but overpredictedthe value by 11%. In the molecular flow regime the leak rate wasunderpredicted by 62%. The conversion from nitrogen-to-airunderpredicted the measured leak rate in both flow regimes.Conversely, the conversion from argon-to-air overpredicted theleak rate.convert a measured tracer gas leak rate, the underlyingassumptions must be fully understood to prevent the acceptanceof invalid data in spaceflight hardware verification.Effect of Pressure Boundary ConditionsIn the second test series, the interior volume of the test fixturewas pressurized with dry air, while the downstream pressure wascontinuously adjusted to maintain a constant differential pressureacross the test article. The selected differential pressures rangedfrom 34.5 to 137.9 kPa and up to six different upstream pressuresfor each differential pressure were tested based on the limitationsof the pressure transducers.The results for each pressure combination are plotted inFig. 7, along with the average leak rate of the test article withrespect to differential pressure. Not surprisingly, the differentialpressure significantly influenced the leak rate of the test article.The average leak rate increased linearly with differentialpressure and by a factor of 4 from 34.5 kPa to 137.9 kPa(differential pressure).Figure 6. Experimental average volumetric air leak rate atNTP compared to projected air leak rates computed bystandard conversion of measured helium, nitrogen, andargon leak rates assuming molecular and viscous flow.Figure 7. Air leak rates for a silicone elastomer O-ring atroom temperature under various pressure boundaryconditions. Initial P-high represents the initial pressure of theinternal volume. Downstream pressure was controlled tomaintain a constant differential pressure. Error barsrepresent measurement uncertainty.These findings supported the assumption that the standardconversion factors for the viscous and molecular flow regimeswere not applicable to a silicone elastomer O-ring whose leakrate was dominated by permeation. Since there is no uniformmethod of converting the helium measurements to the gas ofinterest, using a helium leak detector and applying a standardconversion factor cannot be used to accurately

simulated or misapplied test methods could be misinterpreted as a valid representation of the hardware performance. In this paper, the soundness of leak test methods used to verify the air leak rate of spacecraft sealing systems is investigated. Two common methods to evaluate the leak rate of gas

Related Documents:

Bruksanvisning fรถr bilstereo . Bruksanvisning for bilstereo . Instrukcja obsล‚ugi samochodowego odtwarzacza stereo . Operating Instructions for Car Stereo . 610-104 . SV . Bruksanvisning i original

contents of the tool box" used for validation. The methods and techniques listed in the report are grouped as - review - models - analysis - dynamic methods - methods regarding formality - development methods The validation methods have to be combined together in a validation plan. The plan shall list requirements and validation methods.

contents of the tool box" used for validation. The methods and techniques listed in the report are grouped as - review - models - analysis - dynamic methods - methods regarding formality - development methods The validation methods have to be combined together in a validation plan. The plan shall list requirements and validation methods.

10 tips och tricks fรถr att lyckas med ert sap-projekt 20 SAPSANYTT 2/2015 De flesta projektledare kรคnner sรคkert till Cobbโ€™s paradox. Martin Cobb verkade som CIO fรถr sekretariatet fรถr Treasury Board of Canada 1995 dรฅ han stรคllde frรฅgan

service i Norge och Finland drivs inom ramen fรถr ett enskilt fรถretag (NRK. 1 och Yleisradio), fin ns det i Sverige tre: Ett fรถr tv (Sveriges Television , SVT ), ett fรถr radio (Sveriges Radio , SR ) och ett fรถr utbildnings program (Sveriges Utbildningsradio, UR, vilket till fรถljd av sin begrรคnsade storlek inte รฅterfinns bland de 25 stรถrsta

Hotell Fรถr hotell anges de tre klasserna A/B, C och D. Det betyder att den "normala" standarden C รคr acceptabel men att motiven fรถr en hรถgre standard รคr starka. Ljudklass C motsvarar de tidigare normkraven fรถr hotell, ljudklass A/B motsvarar kraven fรถr moderna hotell med hรถg standard och ljudklass D kan anvรคndas vid

Lร„S NOGGRANT Fร–LJANDE VILLKOR Fร–R APPLE DEVELOPER PROGRAM LICENCE . Apple Developer Program License Agreement Syfte Du vill anvรคnda Apple-mjukvara (enligt definitionen nedan) fรถr att utveckla en eller flera Applikationer (enligt definitionen nedan) fรถr Apple-mรคrkta produkter. . Applikationer som utvecklas fรถr iOS-produkter, Apple .

Cleaning validation Process validation Analytical method validation Computer system validation Similarly, the activity of qualifying systems and . Keywords: Process validation, validation protocol, pharmaceutical process control. Nitish Maini*, Saroj Jain, Satish ABSTRACTABSTRACT Sardana Hindu College of Pharmacy, J. Adv. Pharm. Edu. & Res.