QuickSite & QuickFlex CRT Lead Update Presentation [PDF]

1y ago
22 Views
2 Downloads
592.13 KB
14 Pages
Last View : 19d ago
Last Download : 3m ago
Upload by : Camille Dion
Transcription

QuickSite & QuickFlex CRT Lead UpdateSilicone Insulated Left Ventricular Lead Advisory, April 3, 2012

QuickSite and QuickFlex LV Lead Update St. Jude Medical is proactively informing clinicians of the potential for visualobservations of externalized conductors in QuickSite and QuickFlex leads withsilicone insulation There have been no reports of death or serious injury associated with theexternalized conductors; likewise there have been no electrical dysfunctionsattributable to the externalized conductors. St Jude Medical and its Medical Advisory Board recommend continuing to monitorpatients at regularly scheduled intervals X-ray and fluoroscopy is not recommended, replacement is not recommended As a conservative measure, the sale of QuickSite and QuickFlex leads is beingdiscontinued immediately.pversion of the QQuickSite lead,, model 1056K,, and the QuickFlexQµ The unipolar model 1258T and Quartet model 1458Q leads are not subject to thiscommunication as their construction is different than the subject leads and theQuickFlex µ and Quartet leads utilize Optim insulation rather than siliconeinsulation2 2012.St Jude Medical. All Rights Reserved.

QuickSite and QuickFlex LV Lead Update Silicone insulated bipolar CRT Leads This voluntary advisory includes the QuickSite models 1056T and 1058T, andQuickFlex models 1156T and 1158T. Since 20042004, approximately 171171,000000 of these leads have been implantedworldwide (82,000 QuickSite and 89,000 QuickFlex). Approximately 101,000 QuickSite and QuickFlex bipolar leads are estimated tobe currently in service worldwide (59,000 in the US and 42,000 Internationally). Externalized Cable Observations SJM has confirmed a total of 39 cases of externalized conductors (11 throughfield reportspand 28 throughg laboratoryy analysis).y ) The reported externalization rate is 0.023% (2.3 per 10,000 leads) No electrical dysfunctions have been attributable to externalized conductors Based on a review of 1,219 leads returned for any reason and availableflfluoroscopici andd x-ray images,iit iis estimatedti t d ththatt 3 tot 4% off QuickSiteQ i kSit anddQuickFlex leads may demonstrate this anomaly.3 2012.St Jude Medical. All Rights Reserved.

Examples of Externalization in Silicone Insulated LV LeadsVisual ExampleFluoroscopic ExampleLead BodyTransitionETFE CoatedCableSiliconeCablesterminate atring electrodePolyurethaneSource: SJM Product ReturnsQuickFlex lead model 1158T/864 2012.St Jude Medical. All Rights Reserved.Source: SJM Technical ServicesQuickSite lead model 1056T/86

Confirmed Electrical Failure Rates of Industry CRT Leads50.25%O c c u rre n c e R a te (% ) No electrical dysfunctions have beenattributable to the externalizedconductors Conductor cables are coated withETFE insulation, which providesadequate dielectric strength for thelead to continue to function normally The ETFE coating on the cableshas remained intact in all analyzedcases All-cause electrical failure rate forQ i kSi andQuickSitedQQuickFlexi kFl lleadsd iis0.019% (1.9 per 10,000 leads). Data from returned product analysisand Product Performance Reports(PPR) show all-cause electricalfailure rates for these leadscompares favorably to the other LVleads on the marketCRT Lead Malfunction RatesAll-cause Electrical FailuresBased on Returned Product Analysis (PPR)0.20%EasyTrak (Bipolar)AcuitySteerable0.054%0.019%0.00%Year leads Introduced:St. lit0.008%0.006%Boston Scientific20072004Medtronic20042009Sources: St Jude Medical worldwide returned product analysis data in April 3, 2012 Medical DeviceAdvisory Letter (US data in Nov 2011 PPR and supplemental US and International data on file).Boston Scientific worldwide data in Q4 2011 PPR, Medtronic US data in 2011-2 PPR.Definitions: For St Jude Medical, all-cause electrical failure rate is calculated as fractures,crimp/weld/bond failures, and insulation breaches not resulting in externalized conductors (All-causeelectrical failure rate is also described as the all-cause mechanical failure rate less the externalizedconductor rate). For Boston Scientific and Medtronic, data shown represents lead conductor fracturesreportedepo ted in PPRs.s 2012.St Jude Medical. All Rights Reserved.

Perspective from Medical Advisory Board The clinical implications are much different from, and should not be viewed assimilar to, the Riata lead externalized conductor issueg therapypy ((defibrillation,, LV CRT leads are not used to deliver immediate life-savinganti-tachycardia pacing, or primary pacing for pacemaker-dependent patients). LV CRT leads are used to provide resynchronization therapy for heart failure Long term loss of biventricular pacing could lead to heart failure exacerbation No electrical dysfunction attributable to externalized conductors has beenreported to date and the rate of occurrence of an electrical anomaly for theseleads is expected to be low If an electricall t i l abnormalityblit were tot occur, it can potentiallyt ti ll beb correctedt d withithreprogramming to an alternative configuration at a follow-up clinic visit Therefore, the risk of a clinically significant event is low The MAB provided input to the Dear Doctor Letter and have concurred with itscontents.6 2012.St Jude Medical. All Rights Reserved.

Recommendations from Medical Advisory Board Continue to monitor patient’s implanted system at regularly scheduled intervals Review diagnostic information related to CRT pacing performance, inparticular LV lead impedance and capture thresholds. Programming of alerts that monitor lead impedance changes outside of thenominal range and enabling the patient notifier should be considered ifavailable. X-ray or fluoroscopic imaging is not recommended for leads with normalelectrical function. Only leads that exhibit electrical anomalies and cannot be reprogrammed todeliver effective CRT pacing should be considered for replacement.7 2012.St Jude Medical. All Rights Reserved.

Available Pacing and Impedance Diagnostics LVCap Confirm automatically measures pacing threshold every 8 or 24 hours(programmable), and displays 1-year capture threshold trend LV lead impedance diagnostic measures the programmed pacing vector daily, anddisplaysp y 1-yearyimpedanceptrend Alerts can be set and transmitted if impedance goes outside specified rangeLVCap Confirm Trend is availableon Promote Accel , Unify CRT-Dsand on Anthem RF and AnthemCRT-Ps.8Daily LV Lead Impedance Trend is available on all St Jude Medical CRT-Ds. Programmablealert range and automatic alert transmissions are available on Promote RF/ /Accel, Unify CRTDs and on Anthem RF CRT-Ps. 2012.St Jude Medical. All Rights Reserved.

QuickFlex µ and Quartet CRT Lead ReliabilityNext Generation Optim Insulated Left Ventricular Leads

Optim Insulated QuickFlex µ and Quartet Lead Reliability QuickFlex µ and Quartet are made with all Optim insulation in the entire lead body anddistal section (i.e. they are full body Optim leads). Optim insulation has been shown to be greater than50 times more abrasion resistant than silicone1Full-Body Optim for entire leadQuickFlex µ 1258T (shown) and Quartet 1458Q QuickSite and QuickFlex leads are made with siliconeinsulation in the distal portionPolyurethaneQuickSiteQi kSit 1056T,1056T 1058T1058T,and QuickFlex 1156T, 1158TSiliconeLead BodyTransition1. Helland J, et al. Heart Rhythm, 2005.10 2012.St Jude Medical. All Rights Reserved.

QuickFlex µ and Quartet Lead Clinical Performance With 65,000 Optim insulated QuickFlex µ and Quartet leads implanted starting in 2008,there are no reports or laboratory observations of externalized conductors in Optiminsulated QuickFlex µ and Quartet LV leads Extensive testingg on new and returned Optimpleads shows no signsg of abrasion in cablelumensNameModelYearIntroducedQ i kSitQuickSite1056T20040 043%0.043%0 063%0.063%0 025%0.025%QuickSite 12%0.023%0.012%QuickFlex bined QuickSite/QuickFlexRate of All-CauseAbrasionsRate of All-CauseAbrasionsRate of All-CauseRate of All-CauseMechanical Failures Electrical DysfunctionNameModelYearIntroducedRate of All-CauseRate of All-CauseMechanical Failures Electrical DysfunctionyQuickFlex .008%0.008%Sources: St Jude Medical worldwide returned product analysis data (US data in Nov 2011 PPR and supplemental US and International data on file).Definitions: All-cause mechanical failure rate includes all-cause abrasionsabrasions, fracturesfractures, and crimp/weld/bondfailures. All-cause electrical failure rate is calculated as fractures, crimp/weld/bond failures, and insulationbreaches not resulting in externalized conductors (All-cause electrical failure rate is also described as theall-cause mechanical failure rate less the externalized conductor rate).11 2012.St Jude Medical. All Rights Reserved.

Confirmed Mechanical Failure Rates of Industry CRT Leads Data from returned productanalysis and Product PerformanceReports (PPR) show the all-causemechanical failure rate (includeselectrical and insulation failures) ofQuickFlex µ leads comparesfavorably to the other LV leads onthe market SCORE Registry data onQuickFlex µ leads through 40months of follow-up showexcellent performance The PPR represents US fieldexperience, so Quartet leads dataare not yet available St JudeJ d MedicalM di l iis alsol ttrackingkiperformance of these leads in theOPTIMUM registry and postapproval studies12CRT Lead Malfunction RatesAll-cause Mechanical FailuresBased on Returned Product Analysis (PPR)*µSt JudeMedicalYear leads Introduced:2008Boston Scientific2007-082004Medtronic2004Sources: St Jude Medical worldwide returned product analysis data (US data in Nov 2011PPR and supplemental US and International data on file). Boston Scientific worldwide data inQ4 2011 PPR, Medtronic US data in 2011-2 PPR.Definitions: Category of all-cause mechanical failures includes all-cause abrasions, fractures,and crimp/weld/bond failures. 2012.St Jude Medical. All Rights Reserved.2009

Summary QuickSite and QuickFlex silicone insulated leads We are proactively informing clinicians of the existence of visual observations ofexternalized conductors despite the low potential for clinical consequences and thelow reportedpincidence All-cause electrical failure rates are low, and to date no instances of electricaldysfunction have been attributable to externalized conductors Recommendation is to monitor according to normal follow-up practice QuickFlex µ and Quartet Optim insulated leads QuickFlex µ and Quartet leads are made with all Optim insulation, which is greaterthan 50 times more abrasion resistant than silicone No reports of externalized conductors in over 65,000 Optim insulated QuickFlex µand Quartet LV leads implanted worldwide St Jude Medical has active, prospective registries and post-approval studies to trackperformance of Optim insulated CRT leads13 2012.St Jude Medical. All Rights Reserved.

Brief Summary: Prior to using these devices, please review the User’s Manual for a complete listing of indications,contraindications, warnings, precautions, potential adverse events and directions for use.Unless otherwise noted, or indicates that the name is a trademark of, or licensed to, St. Jude Medicalor one of its subsidiaries. ST. JUDE MEDICAL, the nine-squares symbol and MORE CONTROL. LESS RISK.are trademarks and service marks of St. Jude Medical, Inc. and its related companies. 2012 St.St Jude Medical,Medical IncInc. All Rights ReservedReserved.14

Sources: St Jude Medical worldwide returned product analysis data in April 3, 2012 Medical Device Advisory Letter (US data in Nov 2011 PPR and supplemental US and International data on file). Boston Scientific worldwide data in Q4 2011 PPR, Medtronic US data in 2011-2 PPR.

Related Documents:

CRT CRT 101 Introduction to Commercial Art 25.00 25.00 25.00 25.00 CRT 107 Photo Imaging Assembly 25.00 25.00 25.00 25.00 CRT 108 Line and Halftone Photography 25.00 25.00 25.00 25.00 CRT 112 Electronic Pagination 35.00 35.00 35.00 35.00 CRT 204 Photography 50.00 50.00 50.00 50.00 CRT 206 Digital Imaging 40.00 40.00 40.00 40.00

PSI AP Physics 1 Name_ Multiple Choice 1. Two&sound&sources&S 1∧&S p;Hz&and250&Hz.&Whenwe& esult&is:& (A) great&&&&&(C)&The&same&&&&&

and compression ratio. The results indicate that Huffman-CRT on text outperforms LZW-CRT for compression size by 55.64% to 57.11%, compression time for Huffman-CRT is 44.61s better than LZW-CRT with 55.39s while Huffman-CRT also outperformed LZW-CRT by 3.70x to 2.37x for compression ratio. 1. INTRODUCTION

Argilla Almond&David Arrivederci&ragazzi Malle&L. Artemis&Fowl ColferD. Ascoltail&mio&cuore Pitzorno&B. ASSASSINATION Sgardoli&G. Auschwitzero&il&numero&220545 AveyD. di&mare Salgari&E. Avventurain&Egitto Pederiali&G. Avventure&di&storie AA.&VV. Baby&sitter&blues Murail&Marie]Aude Bambini&di&farina FineAnna

The program, which was designed to push sales of Goodyear Aquatred tires, was targeted at sales associates and managers at 900 company-owned stores and service centers, which were divided into two equal groups of nearly identical performance. For every 12 tires they sold, one group received cash rewards and the other received

1 P a g e CRT 302 – ACLS RENEWAL COURSE CRT 302 ACLS Renewal Course (Baltimore) CRT 302 ACLS Renewal Course (DC) Please bring your MedStar ID if you are an employee or a photo ID if you are a non-employee, as well as your current AHA BLS card or the print out from completion of the Online BLS course (CRT 107) as

College"Physics" Student"Solutions"Manual" Chapter"6" " 50" " 728 rev s 728 rpm 1 min 60 s 2 rad 1 rev 76.2 rad s 1 rev 2 rad , π ω π " 6.2 CENTRIPETAL ACCELERATION 18." Verify&that ntrifuge&is&about 0.50&km/s,∧&Earth&in&its& orbit is&about p;linear&speed&of&a .

with API 650’s level of risk of tank failure. Likewise, the rules in the external pressure appendix will be consistent with the basic part of API 650 with regard to loading conditions and combinations. Thus, starting with a specified design external pressure, roof live or snow load, and wind pressure (or velocity), the total roof design pressure is calculated as the greater of DL (Lr or S .