MICROMETRIX Streaming Current Instrument Training
Like-Charged Particles Repel
Coagulant A coagulant is a charged chemical species that is added to destabilize colloidal particles, so they are free to aggregate. Positively charged metal salts are used to neutralize negatively charged natural particles
Neutral Particles Flocculate
Particles Coagulant Al3 Stable Monomeric hydrolysis products e.g. Al(OH)2 Destabilized Adso Various polymeric species rpt i o n Coagulated Colloidal hydroxide precipitate Restabilized Precipitated Al(OH)3 Flocculated “Sweep Flocculation”
Common Coagulants Aluminum Sulfate (Alum) Ferric Sulfate Ferric Chloride Polymeric Inorganic Coagulants (Polyaluminum Chloride)
SCM Charge Measurement ( ) Positive ( 0 ) Neutral ( - ) Negative Ionic and Colloidal
Background Particle charge/distances influence stability Like-charged particles repel Opposite-charged particles attract Neutral particles are free to collide and aggregate Van Der Waals forces cause particles to attract
Double Layer Model The Double Layer The left view shows the change in charge density around the colloid The right shows the distribution of ( & -) ions around the charged colloid
Double Layer The double layer consists of counter-ions in the stern layer and the charged atmosphere in the diffuse layer The thickness of the layer is determined by the concentration of ions in the solution and the type of counter-ion Type refers to the valence of the positive ( ) counter-ion Example: (Al 3) ions are much more effective in charge neutralizing colloidal material than (Na )
Stabilized Colloids Electric Repulsion Net Interaction Energy Repulsive Energy Energy Barrier Distance Between Colloids Van Der Waals Attraction Attractive Energy Energy Trap
Destabilized System Electric Repulsion Net Interaction Energy Repulsive Energy Energy Barrier Distance Between Colloids Van Der Waals Attraction Attractive Energy Energy Trap
Names for Streaming Current Devices Streaming Current Meter – SCM Streaming Current Detector – SCD Streaming Current Monitor – SCM Particle Charge Detector – PCD Particle Charge Analyzer – PCA
Applications for SCMs Municipal Drinking Water Papermaking Process Sludge Dewatering Industrial Water and High Purity
Drinking Water Treatment Plants began using these devices back in the early eighties for coagulant control The American Water Works Research Foundation funded a project in 1986 to report on the satisfaction of users of Streaming Current technology
SCM vs NTU Streaming Current Response NTU SCM at Overfeed Desired Water Quality SCM at Optimum SCM Chemical Dosage mg/L Chemical Savings
SC Sensor Motor Sideview Stainless Steel Electrode Replaceable Sleeve Teflon and Delrin Material Spare Sleeve/Piston Motor Inlet Barb Fitting Outlet Barb Fitting Probe Housing Teflon Piston Teflon Sleeve Electrode
Stationary Probe Inner Probe Wall Piston Stationary Liquid
Piston Upstroke Probe Inner Probe Wall Piston Moving Liquid
Background Mixing Unit Flocculator SCM Pump Controller Settling Basin Filters Sample Point Well Mixed Quick Response Set Point Empirical Based on Process Performance
Preventing Upsets Maintain Water Quality with SCM Control Clarified/Settled Water NTU 7 6 5 4 3 Flow Increase Coagulant Feed Loss Influent NTU Increase 2 1 Hours With SCM Without SCM
Factors Affecting Operation Chemical Factors Response Time and Lag Time Location of Sample Point
Chemical Factors Ionic strength (conductivity) – typically is sufficiently constant in most water treatment applications Ohms law E I x R An increase in conductivity decreases the SCM sensitivity
Chemical Factors pH – Similar to the effect on Zeta Potential An increase in pH moves the SC reading in the negative direction for simple systems Trends may be more complex for higher pH values 8 May see a decrease in sensitivity for very high pH 10
Response Time Lag time is of considerable importance for proper use of the SCM System Lag time equals process detention/lag time plus sample line detention/lag time The SCM value decreases with time Feedback signal produced by the SCM Controller changes the output to the pump proportionally Process controller has programmable tuning constants for lag times
Built-in alarm functions that alert the operator immediately if an overfeed or underfeed of chemicals is encounters The setpoint is established empirically by the operator to determine what amount of dosage corresponds to optimum chemical feed The process of establishing the setpoint begins when a system is producing acceptable water An incremental decrease in dosage is made and the effect is observed by evaluating the finished quality
Sample Point The sample point should follow mixing Uniform dispersion Reaction of coagulant Fluctuations in the SC reading may indicate incomplete mixing Sample near the center of a pipe or channel Avoid sample locations prone to collecting silt, sand, or grit Shorten sample lines when possible to minimize clogging and reduce response time
Factors Affecting Dose Flow Rate Coagulant Strength Solids Concentration pH Particle Charge Single Offset
Flow Rate A change in plant flow must correspond to a proportional change in chemical feed Automatic flow proportioning can utilize SCM as a trim adjustment Drastic changes in flow can affect lag time
Coagulant Strength The strength of the coagulant or flocculant may vary and can be detected with the SCM Examples have been seen in slurry or polymer makedown
Solids Concentration An increase in influent solids, particle concentration, (turbidity), and organic matter will increase the coagulant demand The SCM reading will move in the negative direction under these conditions
pH Changes in the H ion concentration will affect the SCM Sampling between the coagulant and other chemicals can reduce interference of chemicals used to adjust pH The chemicals can also be dosed proportionally to minimize effects
Particle Charge Variations in colloidal and sub-colloidal (i.e. color) particle charges are measured with the SCM
Signal Offset The absolute SCM reading which corresponds to optimum coagulation efficiency is not necessarily Zero The instruments have the capability to force a “Zero” reading at optimum conditions to allow for an easy reference
Other Factors A periodic “re-optimization” is recommended to verify the proper SCM setpoint Downstream process parameters such as settled and filtered turbidity and particle counts are useful for optimizing The floc meter/PDA can be used as well
Installation The most common problem is clogging of sample line to SCM Recommend relocation to reduce buildup of silt or other debris Cyclone separators and “Y” can be used Periodic Auto or manual flushing Increase flow rate Short sample lines with visible drains
Operational Difficulties Location of sensor at sample point vs. in the Lab (remote sensor) Signal offset due to sample limitations Provides information more rapidly than a Jar Test, but doesn’t indicate process performance only deviations from optimum Chemical savings are more substantial under changing conditions Excessive lag time can lessen response to change in dosage SCM requires scheduled preventative maintenance Samples high in Iron or manganese require more frequent cleaning A loss of response or sensitivity typically indicates cleaning is required
Recommendations for Successful Operations Sample times from coagulant addition to SCM are typically 3-5 minutes SCM must be installed far enough downstream to allow proper mixing (i.e. After rapid or incline mixer, vanturi’s or several pipe bends, etc.) Optimize pH control for proper coagulation Maintain sample flow through sensor and use flow alarm for automatic dosing Keep the SCM Maintained Get the instrument “tuned up” periodically (replace piston and probe components if necessary) Periodically check optimization of the process and verify the SC setpoint
Determination of Proper Set Point First optimize plant process – flocculation, particle and NOM removal Minimize filter effluent turbidity/particle count Minimize coagulant dose and maintain water quality Maximizing plant/process efficiency “Zero” streaming current reading
Maintenance and Troubleshooting Source water quality can affect the frequency of maintenance and cleaning High turbidity, high iron and manganese levels and “foul” the electrode bore faster Oxidized iron and manganese can be cleaned with a reducing agent Worn probe components must be replaced
Streaming Current Monitor -SCM Particle Charge Detector -PCD Particle Charge Analyzer -PCA . Applications for SCMs Municipal Drinking Water . Streaming Current technology. SCM vs NTU SCM NTU SCM at Overfeed SCM at Optimum Chemical Savings Streaming Current Response Chemical Dosage mg/L Desired Water Quality.
standard, the ISO 14001:2004 manual is being integrated into the new manual so that cross referencing between the two systems becomes easier and enhances the understanding of the differences between the new and the old standards. This is to be used as a reference document only _ ESM 43012 Issue Date: 4th October 2017 Page 1 of 40 _ Environmental Manual ISO 14001:2015 _ Introduction .
6 Blueprint II: Improving mental health and well being for all New Zealanders – How things need to be 4. Executive Summary Blueprint II champions a bold new vision to improve the mental health and wellbeing of all New Zealanders. It is a ten year vision that encompasses all of government and provides guidance on what is required to meet
This publication was developed through a consultative process led by the International Telecommunication Union (ITU) and UNICEF and benefited from the expertise of a wide range of contributors from leading institutions active in the information and communications technologies (ICT) sector and on child online safety issues. UNICEF Corporate Social Responsibility Unit: Amaya Gorostiaga, Eija .
Cambridge Checkpoint MATHEMATICS 1112/01 Paper 1 November 2005 1 hour Candidates answer on the question paper Additional Materials: Protractor Ruler NO CALCULATOR ALLOWED READ THESE INSTRUCTIONS FIRST Write your Centre number, candidate number and name on all the work you hand in. Write in dark blue or black pen in the spaces provided on the Question Paper. You are not allowed to use a .
Modernising Scientific Careers The UK Way Forward Executive Summary . p 5 1 The healthcare science workforce is at the heart of safe and effective care for patients, working across some 51 disciplines. It provides expert diagnostic advice and therapeutic care for the treatment of patients and prevention of disease. The workforce develops and applies new technologies that help improve the care .
CalPERS’ SIRI III: Climate Change Request for Information . SIRI.III.RFI.CC Page 2 of 9 12/02/2020 . Through SIRI III we seek to identify and better understand sustainability factors relevant to an interthe -generational,
than English at home.4 In the decade between the 1997-98 and 2008-09 school years, the number of English language learners in public schools increased by 51 percent while the general population of students grew by just 7 percent.5 Given the increase in number of ELL students in the United States, many U.S. teach-
Real and Healthy Chinese Food Recipes All Purpose Sauce . Real and Healthy Chinese Food Recipes Almond Boneless Chicken . Almond Chicken. Almond Cookies . Almond Float . Ants Climbing a Tree or . Ingredients: shoots) ) Braised Chinese Cabbage with Shrimps : )