Updated USP Monograph 1092

3y ago
86 Views
5 Downloads
5.04 MB
91 Pages
Last View : 5d ago
Last Download : 3m ago
Upload by : Genevieve Webb
Transcription

Updated USP Monograph 1092

Updated USP Monograph 1092 USP 711 (Dissolution) late 1960USP 724 (Drug Release) 1985USP 1088 (In Vitro and In Vivo Evaluation of Dosage Forms) 1995USP 1092 (The Dissolution Procedure Development andValidation): Total Revision August 2015 USP 1094 CAPSULES—DISSOLUTION TESTING AND RELATEDQUALITY ATTRIBUTES USP 2040 Disintegration and Dissolution of Dietary Supplements EP 2.9.3 Dissolution late 1960 EP 2.9.4 Dissolution for Transdermal Systems late 1970Harmonization in the year 2006 between USP, EP and JP

Updated USP Monograph 1092 USP 1092 1.1 Performing Filter Compatibility1.2 Determining Solubility and Stability of DrugSubstance in Various Media1.3 Choosing a Medium and Volume1.4 Choosing an ApparatusMETHOD DEVELOPMENT2.1 Deaeration2.2 Sinkers2.3 Agitation2.4 Study Design2.4.1 Time Points2.4.2 Observations2.4.3 Sampling2.4.4 Cleaning2.5 Data Handling2.6 Dissolution Procedure AssessmentANALYTICAL FINISH3.1 Sample Processing3.2 Filters3.3 Centrifugation AUTOMATION4.1 Medium Preparation4.2 Sample Introduction and Timing4.3 Sampling and Filtration4.4 Cleaning4.5 Operating Software and Computation ofResults4.6 Common Deviations from the CompendiaProcedures That May Require ValidationVALIDATION5.1 Specificity/Placebo Interference5.2 Linearity and Range5.3 Accuracy/Recovery5.4 Precision5.4.1 Repeatability of Analysis5.4.2 Intermediate Precision/Ruggedness5.4.3 Reproducibility5.5 Robustness5.6 Stability of Standard and Sample Solutions5.7 Considerations for AutomationACCEPTANCE CRITERIA6.1 Immediate-Release Dosage Forms6.2 Delayed-Release Dosage Forms6.3 Extended-Release Dosage Forms

1.1. Performing Filter Compatibility Filtration is a key sample-preparation step in achieving accurate testresults The purpose of filtration is to remove undissolved drug and excipientsfrom the withdrawn solution The filter material has to be compatible with the media and the drug.Common pore sizes range from 0.20 to 70 µm, however, filters of otherpore sizes can be used as needed Prewetting of the filter with the medium may be necessary. In addition, itis important that leachables from the filter do not interfere with theanalytical procedure. This can be evaluated by analyzing the filtereddissolution medium and comparing it with the unfiltered medium.

1.1. Performing Filter Compatibility The filter size should be based on the volume to be withdrawn and theamount of particles to be separated. Use of the correct filter dimensionswill improve throughput and recovery, and also reduce clogging. Use of alarge filter for small-volume filtration can lead to loss of sample throughhold-up volume, whereas filtration through small filter sizes needs higherpressures and longer times, and the filters can clog quickly. Flow rate through the filter and clogging may be critical for filters used inautomated systems.

1.1. Performing Filter Compatibility

1.1. Performing Filter Compatibility

1.2 Determining Solubility and Stability of DrugSubstance in Various Media When deciding the composition of the medium for dissolutiontesting, it is important to evaluate the influence of buffers, pH, and ifneeded, different surfactants on the solubility and stability of thedrug substance Typical media for dissolution may include the following (not listed inorder of preference): diluted hydrochloric acid, buffers (phosphate oracetate) in the physiologic pH range of 1.2–7.5, simulated gastric orintestinal fluid (with or without enzymes), and water Use of Surfactants for poorly soluble drugs

Table 1. Commonly Used Surfactants withCritical Micelle ConcentrationsCMC(% wt/volume)Reference0.18%–0.23%(2–4)Taurocholic acid sodium salt0.2%(3)Cholic acid sodium saltDesoxycholic acid sodiumsaltCetyltrimethyl ammoniumbromide (CTAB,Hexadecyltrimethylammonium bromide)Benzethonium chloride(Hyamine 1622)Polysorbate 20(Polyoxyethylene (20)sorbitanmonolaurate, Tween 20)Polysorbate 80(Polyoxyethylene (20)sorbitanmonooleate, Tween 80)Caprylocaproyl polyoxyl-8glycerides .0 mM)(5,6)0.18% (4 SurfactantSodium dodecyl sulfate(SDS), Sodium lauryl sulfate(SLS)AnionicCationic

1.3 Choosing a Medium and Volume When developing a dissolution procedure, one goal is to have sinkconditions, which are defined as having a volume of medium at leastthree times the volume required to form a saturated solution of drugsubstance. The use of enzymes in the dissolution medium is permitted, inaccordance with Dissolution 711, when dissolution failures occur asa result of cross-linking with gelatin capsules or gelatin-coatedproducts

1.3 Choosing a Medium and Volume

1.4 Choosing an Apparatus For solid oral dosage forms, Apparatus 1 and Apparatus 2 are usedmost frequently Some changes can be made to the compendial apparatus; forexample, a basket mesh size other than the typical 40-mesh basket(e.g., 10-, 20-, or 80-mesh) may be used when the need is clearlydocumented by supporting data. Care must be taken that basketsare uniform and meet the dimensional requirements specified in 711. Peak Vessels do not comply with USP .can be used forresearch and development.

1.4 Choosing an Apparatus A noncompendial apparatus may have some utility with properjustification, qualification, and documentation of superiority over thestandard equipment. For example, a small-volume apparatus with minipaddles and baskets may be considered for low-dosage strength products. A rotating bottle or dialysis tubes may have utility for microspheres andimplants, peak vessels, and modified flow-through cells for special dosageforms including powders and stents.

1.4 Choosing an Apparatus

2.METHOD DEVELOPMENT 2.1 Deaeration The significance of deaeration of the dissolution medium should bedetermined because air bubbles can act as a barrier to the dissolutionprocess if present on the dosage unit or basket mesh and can adverselyaffect the reliability of the test results. Furthermore, bubbles can causeparticles to cling to the apparatus and vessel walls. Bubbles on the dosageunit may increase buoyancy, leading to an increase in the dissolution rate,or may decrease the available surface area, leading to a decrease in thedissolution rate

FDA Dissolution inFor OralSuspensionII (Paddle)75Phosphate Buffer, pH 6.8 10005, 10, 15, 20, 07/21/230 and 45011II (Paddle)50Water (degassed)5, 10, 15, 20, 06/06/230 and 45013900Amphetamine ERCapsuleII (Paddle)50750 ml of dilute HCl,pH 1.1 for the first 2hours, then add 200 mlof 200 mM phosphatebuffer, and adjust topH 6 (w/ HCl or NaOH)for the remainder750 ml ofdilute HCl,200 ml ofphosphatebuffer1, 2, 3, 4,and 6hoursAspirin/DipyridamoleCapsuleI (Basket)1000.01 N HCl for firsthour, 0.1 M PhosphateBuffer, pH 5.5,thereafter0-1 hrs:900 mL.900 mLthereafterAcid: 10,20, 30, 45and 60min;Buffer:1, 2,5, and 7 tion/dsp getallData.cfm

2.METHOD DEVELOPMENT Peak Vessels are availableAlfuzosin HolderFelodipine BasketDispension Releaser for Nanoparticles containing TestmaterialNew PTWS 1220 12 2 Vessel Dissolution TesterNew Automatic Media Addition System

Method Development, Software structure

Method development, Calibration

2.2 SinkersTable 2. Wire Sinkers Used With Common Capsule Shell SizesCapsule Shell Size#0, elongated#1 and #2#3 and #4Length of Wire(cm)12108Diameter Size(cm)0.80.70.55Cork Bore Number432Sinkers are often used to adjust the buoyancy of dosage forms thatwould otherwise float during testing with Apparatus 2.For materials, use 316 stainless steel wire, typically 0.032 inch/20gauge, or other inert material and wind the wire around cylinders ofappropriate diameter (e.g., cork borers) for an appropriate number ofturns to fit the capsule shell type

2.2. Sinkers

2.3 Agitation For immediate-release capsule or tablet formulations, Apparatus 1(baskets) at 50–100 rpm or Apparatus 2 (paddles) at 50 or 75 rpm are usedcommonly. Other agitation speeds are acceptable with appropriatejustification. Rates outside 25–150 rpm for both the paddle and the basketare usually not appropriate because of mixing inconsistencies that can begenerated by stirring too slow or too fast. Agitation rates between 25 and50 rpm are generally acceptable for suspensions.

2.4 Study Design, 2.4.1 time points For immediate-release dosage forms, the duration of the dissolutionprocedure is typically 30–60 min; in most cases, a single time pointspecification is adequate for pharmacopeial purposes. Industrial and regulatory concepts of product comparability andperformance may require additional time points, which may also berequired for product registration or approvalMonograph with Method description for Biowaivers:www.fip.org/bcs monographs

2.4 Study Design, 2.4.1 time pointsSo-called infinity points can be useful during development studies. To obtainan infinity point, the paddle or basket speed is increased at the end of therun (after the last time point) for a sustained period (typically, 15–60 min),after which time an additional sample is taken. Although there is norequirement for 100% dissolution in the profile, the infinity point can becompared to content uniformity data and may provide useful informationabout formulation characteristics during initial development or aboutmethod bias.Pharma Test Features Infinity Tests

2.4 Study Design, 2.4.1 time pointsMedia Addition for „Half Change“ Dissolution TestsPharma Test automatedMedia Addition Station

2.4 Study Design, 2.4.1 time pointsMedia Addition for „Half Change“ Dissolution TestsPharma Test automatedMedia Addition Station

2.4 Study Design, 2.4.1 time points According to the Biopharmaceutics Classification System referred toin several FDA Guidances, highly soluble, highly permeable drugsformulated into very rapidly dissolving products need not besubjected to a profile comparison if they can be shown to release85% or more of the drug substance within 15 min. For these types ofproducts, a one-point test or disintegration will suffice. However,most products do not fall into this category. Dissolution profiles ofimmediate-release products typically show a gradual increasereaching 85%–100% at about 30–45 min. If the f2 similarity factor is to be used, multiple time points for thedissolution test are required, with at least two time points with meanpercent dissolved (typically for n 12) below 85% dissolved andonly one point above 85% for both products (16). Therefore, theaddition of early time points may be useful.

BIOWAIVER Requiremments/BCS SystemClass IClass IIHighly solublePoorly solubleHighly permeableClass IIIHighly permeableClass IVHighly solublePoorly solublePoorly permeablePoorly permeableMonday, 07 March, 2016www.pharma-test.com28

2.4 Study Design, 2.4.1 time lowlowBCS classificationI(e.g. Propranolol)II (e.g. Glibenclamide)III(e.g. Atenolol)IV (e.g. Azathioprine)

Comparative Dissolution Tests ieBiowaiver and Stability Tests using f2Pharma Test 12 2 Vessel Dissolution Tester PTWS 1220And PTWS D620 (Double Motor)

Comparative Dissolution Tests ieBiowaiver and Stability Tests using f2PTWS 1220 and PTWSD620 (12 2 Vessel Dissolution TesterAdvantages:- f2 Tests always require the Test of 12 Tablets- Same Testtime, Same Temperature, Same Stirrer Speed, Same Sampling Times forall 12 Tablets- Staggered Start is possible to allow timeshifted Teststart- Mechanical Calibration According to all Pharmacopoeia, Only one Protocol for IQ,OQ, PQ and other Validation Documents are needed.- Automation with only one Set of Equippment such as Pumps andSpectrophotometer. Only One Softwarepackage. And only one Set of ValidationDocuments for the additional Instruments- Wider Range for Prednisone Test according to USP (Only one Test is required for12 Vessel Bathes)- With PTWSD620 high degree of Flexibility. 6 Front Row Vessels and 6 Back RowVessels are driven by a individual Motors.- Minimum Space required- Convinient Price compared to 2 Dissolution Instruments with 6/8 Vessels

2.4.2 Observations Visual observations and recordings of productdissolution and disintegration behavior are usefulbecause dissolution and disintegration patterns canbe indicative of variables in the formulation For visual observation, proper lighting (withappropriate consideration of photo-degradation) ofthe vessel contents and clear visibility in the bathare essential. Formulation or manufacturing process It is important to record observations of all sixvessels to determine if the observation is seen in allsix vessels,

2.4.2 ObservationsDeaeratedNot DeaeratedNote the coneshapes and amountof flowing particles!Monday, 07 March, 2016www.pharma-test.com33

2.4.2 ObservationsObservations are: Uneven distribution of particles throughout the vessel Air bubbles on the inside of the vessel or on the apparatus or dosage unit Dancing or spinning of the dosage unit Adhesion of particles to the paddle or the inside of the basket Observation of the disintegration rate Whether the dosage form lands in the vessel center or off-center, and ifoff-center, whether it sticks there

2.4.3 Sampling For manual sampling, use chemically inert devices (e.g., polymeric or glasssyringes, and polymeric or stainless steel cannula), a filter, and/or a filterholder The sampling site must conform to specifications in 711 Replacement is not preferred because the dosage unit may be disturbedduring delivery of the media. However, replacement may be necessary ifmaintaining sink conditions is a challenge. With replacement, the volumeused in the calculations remains the same throughout the time points, butthere is some drug substance withdrawn with each sample that will needto be accounted for in the calculations.

2.4.3 SamplingOuter TubeITM TemperatureSensorInner Tube

2.4.4 Cleaning Importance is placed on evaluation of the cleaning process betweentests. Residues on the vessels can affect the results (e.g., adsorbedresidues may dissolve and alter subsequent media properties orinterfere with the sample analysis), and effective cleaning will returnthem to a suitable state Vessel Washer

2.4.4 Cleaning New Vessel WasherEasy-Ease of Use. It’s small and compact size inaddition with small weight makes the apparatus easyto move.Comfortable use through its touch-screen display andits silent operation.Versatile instrument that can be used with all 1 literUSP dissolution vessels, as well as for 2 liters and500ml vessels, including plastic vessels and specialdissolution vessels like rimless vessels.On-site cleaning makes unnecessary vessel removal,avoiding vessels repositioning, breakage andrecalibration.Fast cleaning operation, typically 20 seconds per vesselhighly increasing laboratory dissolution testingcapacity.Efficient and homogeneous vessel cleaning.Standard Validated Method included with calibratingtablets

3.1 Sample Processing After the samples are withdrawn from the dissolution medium, they mayrequire additional processing to make them suitable for the analyticalmethodology Filtration for UV-VIS, HPLC

3.4 Analytical Procedure The usual assay for a dissolution sample employs either aspectrophotometric procedure or a liquid chromatographicprocedure. Spectrophotometric determination may be direct or mayprovide the detection for HPLC. Spectrophotometric determination isused often because results can be obtained faster, the analysis issimpler, it is easier to automate, and fewer solvents are needed.

3.5 Spectrophotometric Analysis Direct spectrophotometric analysis may be performed on samplesthat are manually introduced to the cuvette samples may be automatically introduced into thespectrophotometer Cells with path lengths ranging from 0.02 cm to 1 cm are typicallyused

3.5 Spectrophotometric AnalysisEP 2.2.25

R3.5 SpectrophotometricAnalysisRequirements EP, USP, FDA, ASTM Spectrophotometer

3.5 Spectrophotometric AnalysisRequirements EP, USP, FDA, ASTM SpectrophotometerControl of absorbance

3.5 Spectrophotometric AnalysisWhen using 2mm, 5mm or 10 mm kuvettes?If the concentration and the Absorbance of the solution is to high for a measurement witha spectrophotometer it’s possible to dilute the sample or to take a smaller pathlength ofthe küvette.Law of Lambert Beer:A e xbxcWhereA is absorbancee is the extinction koefficient (the molar absorbtivity with units of L mol-1 cm-1)b is the path length of the sample - that is, the path length of the cuvette in which the sample is contained.c is the concentration of the compound in solution, expressed in mol L-1Most of the spectrophotometers have an Absorbance measurement limit of 1,2.If you take a look on the Equation the only possibility to change the Absorbance is to change b (the pathlength) or c(concentration by dilution) because e is a constant.

3.5 Spectrophotometric AnalysisTurbidity EffectsCorrected SpectraTC

3.5 Spectrophotometric Analysis Using a validated analytical finish, standard solutions are typicallyprepared in dissolution media and analyzed at just one concentration,either at 100% of the dosage strength or the selected Q value becauselinearity of the analytical finish has been established.

4. AUTOMATION Online analysis returns the sample aliquot to the test system, as in thecase of spectrophotometry with flow-through cuvettes. Offline analysisremoves the sample aliquot from the dissolution medium for subsequentanalysis, typically by HPLC, where the analysis consumes the sample.

4. AUTOMATION/ Offline Systems What is an Offline Automated System? An offline automated dissolution system includes a dissolution bath, a pump and a fractioncollector.Monday, 07 March, 2016www.pharma-test.com49

4. AUTOMATION/ Offline Systems Offline with DSR and CAT

4. AUTOMATION/ Online Systems What is a Online Dissolution System? A closed loop or online dissolution system includes a dissolution bath, a pump and aspectrophotometer with a multiple cell changer.Pump 1 Runtime18152637263754412345678910111213141516Demo Bath 1 RuntimeJM SA500 Spectrophotometer 4567Monday, 07 March, 2016www.pharma-test.com51

4. AUTOMATION/ Online Systems Online System

4.1 Medium Preparation Automated media preparation systems typically dispense the volume ofmedium into the vessel by monitoring either the weight or volume If deaeration of the medium is required, the level of deaeration should bespecified. The concentration of the dissolved oxygen can be used to evaluate theefficiency of deaeration procedures discussed in section 2.1 Deaeration.

4.1 Medium PreparationFactors Influencing the PQ Result - Deaeration The Influence of Dissolved Oxygen 50% increasebetween nondeaerated anddeaerated media*) This value is out of range: Note that this study was carried out when thePrednisone LOT POE203 ranges were : 37 - 70% for Apparatus 2 (paddles)Monday, 07 March, 2016www.pharma-test.com54

4.1 Medium PreparationFactors Influencing the PQ Result - Deaeration% Dissolution of Prednisone Test with Deaeration and NonDeaeration% Dissolution10370374123456VesselMonday, 07 March, 2016www.pharma-test.com55

4.1 Medium PreparationFactors Influencing the PQ Result - DeaerationDeaeratedNot DeaeratedNote the coneshapes and amountof flowing particles!Monday, 07 March, 2016www.pharma-test.com56

4.1 Medium PreparationFactors Influencing the PQ Result - DeaerationMonday, 07 March, 2016www.pharma-test.com57

4.1 Medium PreparationFactors Influencing the PQ Result - DeaerationMonday, 07 March, 2016www.pharma-test.com58

4.1 Medium PreparationFactors Influencing the PQ Result - DeaerationNithyanandan et al. (USP), Dissolution Technologies 2006, 15-18Monday, 07 March, 2016www.pharma-test.com59

4.1 M

USP 1092 (The Dissolution Procedure Development and Validation): Total Revision August 2015 USP 1094 CAPSULES—DISSOLUTION TESTING AND RELATED QUALITY ATTRIBUTES USP 2040 Disintegration and Dissolution of Dietary Supplements EP 2.9.3 Dissolution late 1960 EP 2.9.4 Dissolution for Transdermal Systems late 1970

Related Documents:

USP Reference Standards for USP or NF. section, under Quantitative determinations, the text is revised as follows: ”For the USP Reference Standards where an International Standard (IS) established by the WHO exists, the reference standards documentation will indicate when the USP RS has been established by

Y-site compatible with several IV bag fluids, including: — Water for Injection, USP — 0.9% Sodium Chloride Injection, USP — Lactated Ringer’s Injection, USP — 10% Amino Acid — 5% Dextrose Injection, USP — 5% Dextrose in 0.9% Sodium Chloride Injection, USP — 5% Dextrose in Lactated Ringer’s Injection, USP Titrate DOSE .

USP 1092 The Dissolution Procedure: Development and Validation (USP 38 NF 33, 2015) USP Pharmacopeial Forum – In Process Revision Chapter 1092 addresses the development and validation of dissolution methods, with a focus on solid oral dosage forms. Many of the concepts presented, however, may be applicable to

USP 1092 The Dissolution Procedure: Development and Validation (USP 38 NF 33, 2015) USP Pharmacopeial Forum –In Process Revision Chapter 1092 addresses the development and validation of dissolution methods, with a focus on solid oral dosage forms. M

Understanding the Revisions to USP Monograph 467 : Residual Solvents Page 4 Figure 1. USP Method 467 Procedure A. A) Class 1 for water soluble compounds. B) Class 2 mix A for water soluble compounds. C) Class 2 mix B for water soluble compounds. D) Class 3 mix C for water non-soluble compounds. Conditions same for all runs Column: Zebron ZB-624

DSC USP–NF General Chapters25 USP–NF General Chapters This section contains selected official general chapters that are reprinted from the USP.Only those USP chapters considered relevant for the analysis and/or manufacturing of dietary supplements are included in this section.

Apr 08, 2021 · About United States Pharmacopeia (USP) USP, a scientific organization that sets standards for identity, strength, quality, and purity of medicines, food ingredients, and . Follow containment requirements in USP 800 Perform an Assessment o

Academic writing is iterative and incremental. That is, it is written and rewritten numerous times in a number of stages. Pre-writing: approaches for getting the ideas down The first step in writing new material is to get your ideas down without attempting to impose any order on them. This process is often called ‘free-writing’. In “timed writing” (Goldberg 1986) or “free writing .