Portable And Handheld Systems For Energy-dispersive X-ray Fluorescence .

1y ago
8 Views
3 Downloads
743.22 KB
18 Pages
Last View : Today
Last Download : 3m ago
Upload by : Cade Thielen
Transcription

Alfredo CastellanoUniversità di Lecce, Lecce, ItalyStefano RidolfiArsmensurae, Rome, Italy122234PR5671GPASTRFIINTRODUCTION7E1 Introduction2 X-ray Sources2.1 X-ray Tubes2.2 Radioactive Sources3 X-ray Detectors4 Focusing Optics with CapillaryCollimators5 Portable Energy-Dispersive X-rayFluorescence Systems5.1 Optimization of the X-ray Beam5.2 X-ray Spectrum; Spurious X-rayPeaks and Background5.3 Quantitative Evaluation of anElement from the Photoelectric Peak6 Examples of Applications of Portableor Handheld Energy-Dispersive X-rayFluorescence Equipment6.1 Analysis of Works of Art6.2 Analysis of Alloys and Metals6.3 Analysis of Lead in Paint and Toys6.4 Analysis of Environmental Samples6.5 Analysis of Soil6.6 Mining and Ore Analyzer6.7 The Mars MissionAbbreviations and AcronymsRelated ArticlesReferencesFurther ReadingFSGiovanni E. GiganteUniversità di Roma ‘‘La Sapienza’’, Rome, ItalyORoberto CesareoUniversità di Sassari, Sassari, ItalyEnergy-dispersive X-ray fluorescence (EDXRF) portablespectrometers are becoming very popular in many fieldsfor the on-site analysis of elements. This is mainly becauseEDXRF is a nondestructive, multielemental technique thatis extremely well suited for the analysis of any material.An EDXRF spectrometer mainly consists of an X- orγ -ray excitation source, an X-ray detector with electronics,and a pulse-height analyzer. Recent technological developments have resulted in small, low-power, dedicated X-raytubes, thermoelectrically cooled semiconductor detectors,and small pulse-height analyzers. Therefore, completelyportable EDXRF spectrometers are available that can beassembled on-site, having the size of a book and a weightranging from as light as 500 g (using a radioactive source)to a few kilograms (using an X-ray tube). These spectrometers can be employed for on-site analysis in various fields,such as works of art, alloys, soil, environmental samples,forensic medicine, paper, waste materials, mineral ores andtheir products, or anywhere a portable apparatus would berequired.This article reviews the present status of the developmentand application of EDXRF portable systems. The variouscomponents of a portable system are described: theradiation source, i.e. small, low-power, dedicated X-raytubes or, alternatively, radioactive sources that emit X-raysor low-energy γ -rays; and X-ray detectors, i.e. proportionalgas counters and semiconductor detectors, with specialemphasis on the more recent thermoelectrically cooled Xray detectors: Si-PIN (silicon positive-intrinsic-negative),Si-drift, CdTe, CdZnTe, HgI2 , and others.Commercial systems are considered, and finally the mostcommon and significant applications are described, withparticular emphasis to the field of works of art.OPortable and HandheldSystems for Energy-dispersiveX-ray Fluorescence Analysis9913131515151516161617Encyclopedia of Analytical Chemistry R.A. Meyers (Ed.)Copyright 2009 John Wiley & Sons LtdEnergy-dispersive X-ray fluorescence (EDXRF) is anondestructive technique that is well suited for theanalysis of any material. Its ease of use and adaptabilityhas made EDXRF the most widely used method ofnondestructive elemental analysis. Whether the sampleis a solid, powder, liquid, thin film or coating, elementswith concentrations ranging approximately from ppm topercentage levels can be determined simultaneously withlittle or no sample preparation. EDXRF analysis simplyconsists of irradiating a sample with X- or γ -rays and ofdetecting the secondary X-ray spectrum emitted by thesample itself.(1) Each element emits a typical set of Xray lines and is characterized by their energy. The X-rayspectrum is therefore composed (at least theoretically) ofas many lines as the elements composing the sample.(1)The intensity of each line is proportional in some way tothe concentration of the elements.

2X-RAY SPECTROMETRYPhoton output from an X-ray tube of the type describedin Table 1 depends on many parameters, such as voltage,current, and anode material. Typical outputs are 3–5orders of magnitude larger than those of the radioactivesources.2.2 Radioactive Sourcesα-, β-, X- and γ -ray sources may be employed for EDXRFanalysis.(7) Generally, they are characterized by their verysmall size and low photon output compared with X-raytubes and may be usefully employed for portable EDXRFsystems.EPRFor portable EDXRF equipment, X-ray tubes orradioactive sources may be employed.An X-ray tube typically emits bremsstrahlung radiationwith energy from zero (theoretically) to the value (inkilo electron volts) of the maximum tube voltage, pluscharacteristic X-lines due to the anode material (seeTable 1).In this section, the characteristics of X-ray tubes andradioisotopes emitting X- or γ -rays used for EDXRFanalysis are described along with the detectors that canbe used for portable systems. a length of about 10–15 cm; a diameter of about 3–5 cm; a Be-window 100–500 µm thick, and 0.5–1 cm diameter; a weight of about 200–500 g.FSX-RAY SOURCESIt should be noted that a 35–40 kV X-ray tube is adequatefor the excitation of almost all elements of the periodictable, being able to excite K-lines of elements up to aboutZ 56 and L-lines of heavy elements.In the last few years, small-sized and low-powered Xray tubes of various anode materials have been producedexpressly for EDXRF analysis. Manufacturers includeAMPTEK,(2) Oxford Analytical Systems,(3) HamamatsuPhotonics,(4) Varian,(5) and Moxtek.(6) The low cost ofthese tubes allows the use of various dedicated X-raytubes for the different problems.Figure 1 shows the most common and useful small-sizeX-ray tubes for portable EDXRF equipment, and Table 1lists the characteristics of these X-ray tubes (voltage,current, and anode material). Typical characteristics ofthese X-ray tubes areO22.1 X-ray TubesOThe sample also scatters (both coherently and incoherently) the incident radiation, generating lines duemainly to Rayleigh (coherent) and Compton (inelastic)scatterings.(1)An apparatus for EDXRF analysis consists of anX- or γ -ray excitation source, an X-ray detectorwith electronics, and a multichannel analyzer. Portable‘‘laboratory’’ systems can be assembled, and can beemployed for on-site analysis in various fields such asanalysis of works of art, alloys, environmental samples,forensic medicine samples, paper, waste materials, ferrousmetal mineral ores and their products, etc.Recent technological developments have resulted inboth small-sized, low-power X-ray sources and small,thermoelectrically cooled semiconductor detectors.This article reviews the present state of developmentand application of EDXRF analysis, using miniatureX-ray tubes or radioactive X- or γ -ray sources forexcitation, and semiconductor detectors for detection.Special emphasis is given to thermoelectrically cooledX-ray s(µm)Weight(g) Size (length diameter in cm)W (transmission)9.5200100545 1Ti, Cr, Fe, Co, Ni,Cu, Mo, Pd, Rh,Ag, Ta, W, Pt,Au (diffusion)Mo, Pd, Ag, Rh,Ta, W(transmission)Ti, Mo, Pd, Rh(transmission)Ag (transmission)10–35200–1500125–25020011 3.50–400–1002504505 2.55010008050011 65–3010012530016 4Anode materialHamamatsuN7599-01Oxford-TF pottedRSTPAProducer and codeMoxtek 40-kVMagnumFIVarian VF-50AMPTEKOxford Eclipse(II or III)GTable 1 Characteristics of commercial X-ray tubes for portable energy-dispersive X-ray fluorescence (EDXRF) analysis

PORTABLE AND HANDHELD SYSTEMS FOR ENERGY-DISPERSIVE X-RAY FLUORESCENCE ANALYSIS3 cm10SteelTungsten(d)Source(a)VF-50 Shielded7 cm2Source1013 cmSteel(b)(e)FSavailable from, for example, the Radiochemical Centre,Amersham, UK.(8)β-ray sources can also be employed both for directEDXRF excitation of a sample and for producingbremsstrahlung radiation in a target to successively excitethe sample. Typical sources of the first type are 22 Na,85Kr, and 63 Ni, with half-lives of 2.6, 10.7, and 100 years,respectively, and emissions of 0.55, 0.67 and 0.066 MeVβ particles. Typical sources of the second type are 147 Pm(with a zirconium target) and 3 H (with a titanium target),with half-lives of 2.6 and 12.4 years, respectively, andemitting photons up to 225 keV and 19 keV, respectively.Few radioactive sources emitting X- or γ -rays are usedfor EDXRF analysis; the most useful are listed in Table 2.The sources detailed here are available, for example, fromthe Radiochemical Centre, Amersham, UK.Radioactive sources for EDXRF analysis are muchsmaller than X-ray tubes. However, the energy ofthese sources cannot be changed. Moreover, theiroutput is often not adequate for an efficient excitation.Additionally, X-ray tube output can be monochromatizedat various energies using proper filters. Finally, the costsof an EDXRF tube are currently comparable to those ofa radioactive source.5 cm54 cmX-RAY DETECTORSO318 cmFigure 1 Small-size X-ray tubes for portable EDXRFPRequipment. (a) Moxtek 40 kV, 1 mA; (b) Varian 50 kV, 1 mA;and (c) Eclipse III, 30 kV 100 µA by AMPTEK-Oxford;(d) and (e) For comparison, radioactive sources are also shown(dimensions in millimeters).In the past, portable EDXRF analyzers traditionallyinvolved the use of scintillation or proportional gas detectors, while laboratory systems typically used nitrogencooled semiconductor detectors, such as Si(Li) and HPGe(hyperpure germanium).(9) Sometimes proportional gascounters are employed for EDXRF analysis filled withneon, argon, krypton, or xenon. Though having a poorenergy resolution, they have a large area.(10)In the last decade, there has been a true revolutionin the field of X-ray detectors. In fact, thermoelectricallycooled Si-PIN (silicon positive–intrinsic–negative),(2) Sidrift,(11) CdZnTe,(2) CdTe,(2) and HgI2 (12) detectors haveO(c)PAGEα-ray sources are especially suited for the analysisof low atomic number elements. Those most used are244Cm, with a half-life of 17.8 years and emission of 5.76and 5.81 MeV α particles, and 210 Po, with a half-life of138 days and emission of 5.3 MeV α particles. Both areSTTable 2 Radioactive sources that can be used for energy-dispersive X-rayfluorescence (EDXRF) analysis. (Available from the Radiochemical Centre,Amersham, f-life (years)X- or γ -ray energy(keV)Typical output(photons/‘s sr)2.7881.34330.660.745.9–6.5 (Mn X-rays)14.6–22 (U L X-rays)22–25 (Ag X-rays)59.6 (γ -rays)41.5 (Eu K X-rays)122, 136 (γ -rays)7 1068 1066 1074 1084 106

4X-RAY SPECTROMETRY1000FWHM (eV)Proportional gascounters E between adjacent Ka linesSi-PIN100Si-driftSi(Li)10110Energy (keV)100Figure 2 Energy resolution (in keV), as the full width at half-maximum (fwhm) of the X-ray peak, versus energy for differentdetectors. From top to bottom: proportional gas counter, Si-PIN and Si-drift thermoelectrically cooled semiconductor detector, andnitrogen-cooled Si(Li). The Kα energy difference of contiguous elements is also shown.4FOCUSING OPTICS WITH CAPILLARYCOLLIMATORSOOFSCapillary collimators may be employed for focusing anX-ray beam into a micrometric area, allowing EDXRFanalysis of micrometric areas (µ-EDXRF analysis).(13,14)This method is largely employed with synchrotronradiation, where high-intensity X-ray beams are available.A combination of a polycapillary conic collimator(poly-CCC) with an X-ray detector (e.g. Si-PIN or Sidrift) allows local detection of X-ray fluorescence (XRF)signals from a small surface area of the sample. In thiscase, a portable EDXRF equipment may be assembled,and µ-EDXRF analysis can be carried out by using apolycapillary conic collimator located as a cap at thedetector entrance (Figure 3).FIRSTPAGEPRsubstituted, for portable apparatus, proportional gascounters, or nitrogen-cooled Si(Li) and HPGe detectors.The thermoelectrically cooled Si-PIN or Si-driftdetectors have a thickness typically of about 300–500 µm,which makes this detector useful up to X-ray energies of30 keV, and an energy resolution of about 140–180 eVat 5.9 keV, which is only a little worse than that of anitrogen-cooled Si(Li) detector.Si-drift detectors have a better energy resolution thanSi-PIN, and a capacity of processing 104 –105 photons/swithout loss of energy resolution. This detector is,therefore the best current option for EDXRF analysisin the range 1–25 keV approximately; however, Si-drift ismuch more expensive than Si-PIN.The CdTe and CZT (cadmium–zinc–telluride) detector have a thickness of about 2 mm, and thereforean efficiency of 100% up to 150 keV, with an energyresolution of about 250 eV at 5.9 keV and about 1 keVat 60 keV.(2) The HgI2 detector has a thickness of a fewmillimeters, sufficient for an efficiency of about 100%in the whole range of X-rays. It typically has an energyresolution of about 200 eV at 5.9 keV.(12)Figure 2 shows the energy resolution versus energyof various X-ray detectors, proportional gas counters,Si-PIN, Si-drift, CdTe, and HPGe in the X-ray energyrange.The only present limitation of the newthermoelectrically cooled Si-PIN, Si-drift, CdTe, andHgI2 detectors in their useful energy range is the reducedarea, which limits the efficiency of the system. If highefficiency is required, or only one element is to beanalyzed, detectors with larger area can be used, suchas proportional gas counters, Si(Li), or HPGe.DetectorPoly-CCCPrimaryexcitationby X-ray orelectron beamsSampleX-Y-scanFigure 3 –Polycapillary conic collimator (Poly-CCC) coupledto an X-ray detector. The detector only collects photons from asmall area, while a much larger area is irradiated by an X-raybeam. (With permission from Institut for Scientific Instruments.)

5PORTABLE AND HANDHELD SYSTEMS FOR ENERGY-DISPERSIVE X-RAY FLUORESCENCE ANALYSISPORTABLE ENERGY-DISPERSIVEX-RAY FLUORESCENCE SYSTEMS55109Cd (22.1 keV)244Cm (14.6 and 18.8 keV)125I (27.4 and 31 keV)153Gd (41.5 and 47 keV)241Am (59.6 keV)EGPARSTThermo Niton Analyzers LLC(15)X-MET by Oxford Instruments(3)Innov-X Systems(16)Warrington Inc.(17)RMD Instruments(18)Applitek(19)FI A typical handheld instrument is shown in Figure 5.The following are the characteristics of the most commoncommercial handheld instruments:57L-linesCo (122 and 136 keV)K-lines(a)Tungsten 50 kVTungsten o different types of X-ray sources (in some cases γ ray sources) can be employed for portable (or handheld)EDXRF Systems, generally coupled to a Si-PIN detector(but also Si-drift or gas proportional counter and foranalysis of high-energy X-rays, also CdTe) and areavailable on the market: radioisotopic sources or smallsize X-ray tubes.Radioactive sources, despite the low photon emissionand emission energy flexibility, and the legislation limitingtheir use, are still employed, because of a very small sizeand, therefore, high portability, and because of theirintrinsic stability.For EDXRF analysis of single elements (for example,lead in paints or toys), where energy resolution is nota problem, the best and more convenient solution canbe that of a radioactive source with a gas proportionalcounter or a scintillation detector.The coupling of a high-intensity flexible source such asthe X-ray tube with a high-resolution thermoelectricallycooled detector (for example, Si-drift or Si-PIN) is themost common solution and is in our opinion the bestpossible current system for in situ XRF analysis, especiallywhen many elements are involved. The combination ofthe appropriate X-ray tube (in terms of anode, highvoltage, current, and collimation) coupled to a properX-ray detector (thickness of Si and of the Be-window)and sample geometry gives the best analytical results.Figure 4 shows the range of elements that may beusefully analyzed, versus the type of radioactive sourceor voltage and anode of X-ray tubes.Several portable systems based on the use of anX-ray tube or a radioactive source, coupled to athermoelectrically cooled semiconductor detector (morerarely, proportional gas counter or scintillator) arecurrently commercial. They are generally handheld XRFinstruments, and are employed for various applications,such as analysis of lead in paint and toys, of alloys,of environmental samples, of geological samples, ofheavy metals in soil, of works of art, and so on, andare manufactured and supplied by several companies,namely,Fe (5.9 keV)O55(b)1525L-linesK-lines35455565758595Atomic number, ZFigure 4 The range of elements that can be analyzed using(a) radioactive sources and (b) X-ray tubes with differentanodes, showing excitation of K-and L-lines.1. Thermo Niton Analyzers LLC (by Thermo FisherScientific) traditionally produces handheld instruments for various applications, i.e. the models Xli andXLt, the first one using radioactive sources (20 mCi ofFe-55, 10 mCi of Cd-109 and 14 mCi of Am-241), thesecond, a small-size X-Ray tube (40–50 kV, hundredsof microamperes). More recently, NITON produceda specific model for analysis of works of art. Allmodels use an Si-PIN detector.2. Oxford Instruments, which traditionally producesX-ray tubes, offers the handheld XRF analyzer XMET3000TXS (or TXR), specifically dedicated forthe analysis of heavy metals in soil on pollutedlands, and for analysis of ores at mining sites. This

6X-RAY SPECTROMETRYMain featuresErgonomically designed, ideal form factorSilicon ‘‘p–i–n’’ detector, 200 eV40-kV, 50-uA transmission anodelow power X-ray sourceLightweight3 Ibs (1.36 kg)Long, 12 hours battery useLarge touch screen LC display2.25″ 3″ (57 mm 76 mm)4096 channel MCAOptimized for speed ASIC-based DSPIntegrated touch screenIntegrated bar code scannerIntegrated wireless communication linkFastest FPFP not sensitive to shape and surface irregularitiesNo transport restrictionsFigure 5 Typical modern Thermo Scientific Niton handheld EDXRF-analyzer. (With permission of Thermo FisherScientific.)PA6.ST5.GE4.5.1 Optimization of the X-ray BeamOOFSConsidering a typical X-ray tube working at 30–40 kV, itemits bremsstrahlung radiation of the order of a few units 107 photons/µA s, which are generally adequate in thecase of thin-sample analysis (for example, aerosols), butin many cases largely in excess with respect to the XRFrequirements (MDL, energy resolution, and measuringtime), equipment parameters (mainly dead time of theelectronic chain). However, this excess of photons allowsa considerable flexibility in the treatment of the photonbeam.To reduce the photon intensity at the X-ray tube outputand detector entrance, and, in some cases, also to properlyform the X-ray beam, various ways can be followed:PR3.equipment is based on an X-ray tube and the newpentaPIN detector. Recently, a special configurationof the X-MET3000TXR was released, which providesa simple method for detection and analysis of leadcontent.Innov-X Systems produces various models ofhandheld analyzers (classic, inspector, import guard,lead guard, ultrarugged, vacuum), based on the useof a miniature, rugged 10–35-kV X-ray tube and a SiPIN detector. Various applications are considered:alloy and precious metals analysis, lead analysisin paints, analysis of toxic metals in food, andso on.Warrington Inc. produces a handheld XRF analyzerspecifically dedicated to lead analysis. It is mainlycomposed of a Co-57 radioactive source and a CsIdetector, and is able to detect Pb with a minimumdetection limit (MDL) of about 0.3 mg cm 2 .RMD Instruments produces the handheldLeadTracer-RoHS system, specifically dedicated toscreen lead in toys. It is based on the use of a Co-57radioactive source and a CdTe detector.Applitek produces the eXaMiner, specificallydesigned for mining and ore analysis. It is basedon the use of an X-ray tube and a Si-PINdetector.FIRIt should finally be observed that it is not difficult toassemble a portable EDXRF equipment by separatelybuying a proper X-ray source, detector, and pulse-heightanalyzer. For this reason, there are many self-madeportable EDXRF systems; several of these are describedin the following.1. collimate the X-ray tube and/or the detector;2. collimate and filter the X-ray tube and/or collimatethe detector;3. monochromatize the X-ray beam from the X-raytube;4. use a capillary collimator at the detector entrance.The first mode is simple, of common use, and,sometimes, necessary, and it only requires propercollimators of adequate material. However, it does notallow full exploiting of all potentialities of the X-raytube.The second mode, often coupled to the first, has theaim to cut out part of the incident X-ray spectrum. Forexample, the disturbing Ar-contribution from the air canbe practically eliminated by strongly cutting the lowenergy tail of the X-ray spectrum with an Al-filter. Thisconfiguration can also be particularly useful for enhancingthe higher energy part of the spectrum, i.e. to better

PORTABLE AND HANDHELD SYSTEMS FOR ENERGY-DISPERSIVE X-RAY FLUORESCENCE ANALYSISFS Escape peaks Monoenergetic X-rays generated inthe object to be analyzed may be subject tophotoelectric effect interacting with atoms of thedetector (for example, Si). Then an acceleratedelectron is originated, and X-rays are emitted fromthe excited atom; while the electron is certainlyprocessed, X-rays may escape, especially when thephotoelectric effect happens at the border of thedetector. In this case, a second peak (escape peak, withenergy Ee ) is produced. This peak will have an energyEe E0 EK,L,. , where E0 is the energy of incidentmonoenergetic photons, and EK,L,. , represents the K,L, . . . energy of the detector material. For example, inthe most frequent case of Si-detectors, peaks at energy(EK 1.8) keV will be present. The escape-peak effectis more probable for small size detectors. Sum peaks Electric pulses related to two incidentphotons may be processed contemporaneously,according to the shaping time of the amplifier andprocessing time of the multichannel analyzer. Thiseffect, more probable in the case of high shaping timesof the amplifier and high counting rates, produces apeak that corresponds to the sum of the two incidentphotons. Pile-up effects This effect, visible in the X-rayspectrum as an enlargement of the photoelectric peaksat the left-bottom of the peaks, is increasing with thecounting rate, and is related to the processing timeof the electronic chain, including the pulse-heightanalyzer.STPAGEPRWhen an object is irradiated by an X-ray beam, andsecondary radiation emitted by the object is collected bythe detector and processed by a pulse-height analyzer,an X-ray spectrum is obtained, generally composed ofmany peaks, and of a background. Several peaks aredue to photoelectric effect by the object atoms, andgive information about the composition of the object,but other peaks may also be present, due to ‘‘spurious’’effects.Various peaks are present in the final X-ray spectrum,as well as background photons. Possible X-ray peaks maybe originated by the anode material of the X-ray tube.For example, by using a W-anode working at 30–50 kV,W/L-lines are emitted by the X-ray tube, and may bescattered by the object material. Or on using a Aganode X-ray tube, the Ag K-lines will be present in thefinal spectrum, due to scattering by the object. Further,other peaks may be present, due to the various processesof interaction (mainly photoelectric effect and coherentand Compton scattering) of primary photons with air,source and/or detector collimators, object-matrix, andchemical elements in the analyzed object. Accordingto the incident radiation, the following situations mayoccur:2. Monoenergetic radiationIn this case, the monoenergetic peaks are elastically andCompton scattered, and this appears in two peaks, onecorresponding to elastic scattering, and a second one,of lower energy and larger, due to Compton scattering.Then, also in this case Ar-K-lines may be present, and theX-ray peaks of elements of the collimator.Finally, in both cases 1 and 2, peaks will be present, dueto the elements of the object to be analyzed. These peaksmay also generate new peaks or peak alterations, i.e.O5.2 X-ray Spectrum; Spurious X-ray Peaks andBackgroundof elements of the collimator may be present; when thecollimator is made on brass, then Cu and Zn peaks maybe expected.Oexcite elements at the border of the X-ray spectrum (forexample, tin).The third mode, monochromatization (or better,partial monochromatization) of the X-ray beam, requiresa filter of a specific element, in transmission or indiffusion, taking into account that a higher level ofmonochromatization requires a higher loss of photonintensity. This procedure can be particularly usefulto excite, in the best manner, a specific element, byirradiating it with a monochromatic beam close to itsspecific fluorescent discontinuity.(20)The fourth mode was explained in Section 4; apolycapillary conic collimator is put as a cap atthe detector window,(21) allowing the detection ofsecondary X-rays from a very small area of the analyzedobject.7FIR1. Bremsstrahlung radiationIn this case, a bremsstrahlung contribution appears in thefinal spectrum, due to elastic and Compton scattering inthe matrix, reproducing, in some manner, the incidentradiation spectrum. Further, argon K X-ray peaks arepresent (at 2.96 and 3.2 keV), due to the presence of Ar inair; this peak is strongly reduced when the primary beamis filtered to enhance its high energy contribution. WhenX-ray source and/or detector are collimated, X-ray peaksFigure 6 shows a typical X-ray spectrum, obtained witha Si-PIN detector, in which several of the described‘‘spurious’’ effects are visible.5.3 Quantitative Evaluation of an Element from thePhotoelectric PeakWhen a sample containing an element a with aconcentration ca is irradiated by a beam of X-rays having

8X-RAY SPECTROMETRY35000.0Au-La30000.0Au-LhAu-La (escape)25000.0Ar-K Ag-LAu-Lb20000.0Au-MAu-La Au-LbAu-Ll15000.0Au-Lb (sum)Au-La (sum)10000.0Au-Lg5000.00.0 024681012141618202224262830Figure 6 X-ray spectrum of a gold sheet irradiated by an X-ray tube with Ag-anode working at 28 kV (dead time 20%), showingX-ray peaks at the following energies: 2.15 keV (M-lines of Au); 2.95 keV (K-line of Ar); 8 keV (escape of 9.7 keV Lα-line of Au);8.5 keV (Ll-line of Au); 9.7 keV (Lα-line of Au); 10.3 keV (Lη-line of Au); 11.5 keV (Lβ-line of Au); 13.4 and 14.3 keV (Lγ -lineof Au); 19.5 keV (sum of two 9.7 keV photons); 21.2 keV (sum of a 9.7 keV and a 11.5 keV Au-photons); 23 keV (sum of two 11.5photons) (the various parts of the X-ray spectrum are on different y scales).FSNa N0 kωa Ja σa ca MWhen a generic element a with concentration ca , inan infinitely thick and homogeneous sample is irradiatedwith N0 incident photons, the secondary fluorescent X-rayintensity Na is given byOan energy E0 and intensity of N0 photons/s, the numberNa of fluorescent X-rays emitted by the element a, isapproximately given by(1) :(1)EGPASTTwo extreme conditions related to the sample thicknessare considered in the following.5.3.1 Thick SamplesRMost objects generally appear to EDXRF analysis as‘‘infinitely thick samples’’, in the sense that the thicknessof the objects is much greater than the ‘‘radiationpenetration’’. This is the general case of solids and liquidssuch as solutions, minerals, soil samples, artifacts likestatues, columns, alloys, and so on.FIQ1(2)where µph.a (E0 ) represents the photoelectric attenuationcoefficient of element a at incident energy E0 ; µt (Eo ) andµt (Ea ) represent the total attenuation coefficient of thesample at incident and fluorescent energies (E0 and Ea ),respectively.(1) The term in square brackets representsthe ‘‘matrix effect’’. The relationship between ca and Nais, therefore, not generally linear.Besides fluorescent X-rays, given by Equation (1), theX-ray spectrum emitted by the irradiated ‘‘infinitelythick’’ sample is also composed by scattered photons(Compton and Rayleigh effects). Scattered radiation isgenerally a disturbing effect that should be reducedas much as possible; however, in some cases, scatteredpeaks can be employed for normalization purposes.Equations (1) and (2) are strictly valid for thicksamples and for monoenergetic incident radiation. Asobserved above, Equation (2) yields an approximatelinear relationship between Na and ca or a nonlinearone according to the term in square brackets.Standard samples are required for an experimentaltest of Equation (1) and to quantitatively establish thecorrelation between Na and ca .PR k is an overall geometrical factor; ωa is the fluorescent yield of the element a in the shellof interest (i.e. percent probability of a fluorescenceeffect compared with an Auger effect); σa (cross section in cm2 ) is related to the probabilityfor fluorescent effect of the element a; Ja is the branching ratio, i.e. the intensity of the X-lineof interest over the total X-ray intensity; M is a matrix term (i.e. depending on the sample),related to the attenuation of incident and secondaryfluorescent radiation and the sample composition.ONa No kωa Ja ca [µph.a (E0 )/µt (E0 ) µt (Ea )]where

9PORTABLE AND HANDHELD SYSTEMS FOR ENERGY-DISPERSIVE X-RAY FLUORESCENCE ANALYSIS5.3.2 Thin SamplesNa N0 kωa Ja σa ca(3)i.e. counts of element a are linearly proportional to itsconcentration.Intensity Nsc of scattered photons in the case ofthin samples (mainly due to Compton scattering) isapproximately given by Equation (4):Nsc N0 kµsc (E0 )m(4)Measuring headwhere µsc (E0 ) is the attenuation coefficient of the sampleat incident energy and m (in grams per centimetersquared) is the mass per unit area of the sample.FSEXAMPLES OF APPLICATIONS OFPORTABLE OR HANDHELDENERGY-DISPERSIVE X-RAYFLUORESCENCE EQUIPMENTSi-pinHamamatsu X-ray tubeX-ray tube high voltageOFigure 7 Handheld EDXRF equipment employed for analysisof sulfur and chlorine in a fresco by Pomarancio in the church ofS. Stefano Rotondo in Rome. The measuring head is composedof a Hamamatsu with Ca anode, working at 8 kV, 0.3 mA, and aSi-PIN detector.G6.1 Analysis of Works of ArtEPRPortable EDXRF equipment are especially useful whenthe object

Collimators 4 5 Portable Energy-Dispersive X-ray Fluorescence Systems 5 5.1 Optimization of the X-ray Beam 6 5.2 X-ray Spectrum; Spurious X-ray Peaks and Background 7 5.3 Quantitative Evaluation of an Element from the Photoelectric Peak 7 6 Examples of Applications of Portable or Handheld Energy-Dispersive X-ray Fluorescence Equipment 9

Related Documents:

Bruksanvisning för bilstereo . Bruksanvisning for bilstereo . Instrukcja obsługi samochodowego odtwarzacza stereo . Operating Instructions for Car Stereo . 610-104 . SV . Bruksanvisning i original

10 tips och tricks för att lyckas med ert sap-projekt 20 SAPSANYTT 2/2015 De flesta projektledare känner säkert till Cobb’s paradox. Martin Cobb verkade som CIO för sekretariatet för Treasury Board of Canada 1995 då han ställde frågan

service i Norge och Finland drivs inom ramen för ett enskilt företag (NRK. 1 och Yleisradio), fin ns det i Sverige tre: Ett för tv (Sveriges Television , SVT ), ett för radio (Sveriges Radio , SR ) och ett för utbildnings program (Sveriges Utbildningsradio, UR, vilket till följd av sin begränsade storlek inte återfinns bland de 25 största

Hotell För hotell anges de tre klasserna A/B, C och D. Det betyder att den "normala" standarden C är acceptabel men att motiven för en högre standard är starka. Ljudklass C motsvarar de tidigare normkraven för hotell, ljudklass A/B motsvarar kraven för moderna hotell med hög standard och ljudklass D kan användas vid

LÄS NOGGRANT FÖLJANDE VILLKOR FÖR APPLE DEVELOPER PROGRAM LICENCE . Apple Developer Program License Agreement Syfte Du vill använda Apple-mjukvara (enligt definitionen nedan) för att utveckla en eller flera Applikationer (enligt definitionen nedan) för Apple-märkta produkter. . Applikationer som utvecklas för iOS-produkter, Apple .

This presentation and SAP's strategy and possible future developments are subject to change and may be changed by SAP at any time for any reason without notice. This document is 7 provided without a warranty of any kind, either express or implied, including but not limited to, the implied warranties of merchantability, fitness for a .

och krav. Maskinerna skriver ut upp till fyra tum breda etiketter med direkt termoteknik och termotransferteknik och är lämpliga för en lång rad användningsområden på vertikala marknader. TD-seriens professionella etikettskrivare för . skrivbordet. Brothers nya avancerade 4-tums etikettskrivare för skrivbordet är effektiva och enkla att

Workshop 3: Basic Counselling Skills for Drug Addiction Treatment Workshop 4: Special Considerations when Involving Families in Drug Addiction Treatment. 5 Icebreaker: If I were the President If you were the President (King, Prime Minister, etc.) of your country, what 3 things would you change related to drug policies, treatment, and / or prevention? 15 Min. 6 Workshop 1: Biology of Drug .