Determination Of Calcium Hardness No. T10 In Water By Automatic Titration

1y ago
6 Views
1 Downloads
1.29 MB
6 Pages
Last View : 1m ago
Last Download : 3m ago
Upload by : Victor Nelms
Transcription

APPLICATION NOTE Key Words EDTA, complexometric, titrimetric, ASTM D1126, ASTM D511, ISO 6059, SM 2340C, SM 3500-Ca, magnesium, Orion 9720BNWP, Orion Star T930, Orion Star T940, titrator. Drinking water, process water, cooling water, boiler water, wastewater, surface water, environmental water, raw water. Introduction Calcium Hardness in water is determined using the preprogrammed method, T10 Ca Hard. To determine calcium, sodium hydroxide is added to a sample to raise the pH to 12 to 13. The sample is then titrated to the equivalence point using ethylenediaminetetraacetic acid (EDTA) titrant. A calcium ion selective electrode (ISE) is used as the titration indicator, which gives a potentiometric change when all the calcium has been complexed by the EDTA. The calcium ISE overcomes problems typical of color indicators such as endpoints that are not clear and sharp, endpoint color changes that are subjective, and the need for careful lighting to determine the endpoint during titration. Recommended Equipment Thermo Scientific Orion Star T930 Ion Titrator or T940 All-in-One Titrator or equivalent with a 20 mL burette. Thermo Scientific Orion 9720BNWP Calcium electrode or equivalent. Pipet: 10 mL graduated; 1.00 mL (optional for standardization) Graduated cylinders: 100 mL and 250 mL Beakers: 100 mL and 250 mL Orion Star T930 ISE Titrator with 20 mL burette Required Reagents and Solutions Titrant: Purchased or prepared EDTA titrant, 0.01M (10 mM). Calcium Standard: Orion 92206 0.1M Calcium Standard. For preparation of storage solution and for standardizing titrant. Water Hardness Standard (optional): Orion 923206 Water Hardness Standard, 100 mg/L (ppm). Reagent Solutions: Purchased or prepared 1N (1M) Sodium Hydroxide; Prepared storage solution. Reagent grade water (RGW). Use suitable Personal Protective Equipment (PPE) as recommended by the Safety Data Sheets (SDS) for the chemicals utilized during this procedure. No. T10 Determination of Calcium Hardness in Water by Automatic Titration

Titrator Setup Connect the electrode and the stirrer probe to the titrator. If not previously done, import the T10 Ca Hard preprogrammed method into the titrator through the Methods screen. Rinse and fill the burette with titrant. See the titrator user manual for details on setting up the titrator. Table 1 If bubbles are visible in the tubing, dispense titrant (from the Burette screen) until the bubbles have been expelled. Tap the tubing to dislodge bubbles. Consider standardizing the titrant before titrating samples. See Titrant section below. Electrode Preparation At the start of the day, remove the electrode from storage solution. Add fill solution to the bottom of the fill hole. Leave the fill hole open during testing. Rinse thoroughly with RGW before and between titrations. At the end of the day, thoroughly rinse the electrode with RGW and store in storage solution. Cover the fill hole when in storage. Storage Solution Preparation Place 5 mL of Orion 922006, 0.1M calcium standard into a 100 mL beaker and add RGW to the 50 mL mark to make a 0.01M storage solution. Store the calcium ISE in this solution. Prepare fresh on a weekly or biweekly basis. Sample Preparation Using a graduated cylinder or pipet, accurately transfer an appropriate amount of the sample into a clean beaker. See Table 2 for guidance on recommended beaker size and sample volumes. Record the volume used. If the volume of sample is less than 50 mL, add RGW to the 50 mL mark on the side of the beaker. Add 2 mL of 1N NaOH to bring the pH above 12 (e.g., near 12.5). Proceed to Sample Titration. 2 T10 Ca Hard Method: Preprogrammed Parameters Electrode Parameter Electrode Type ISE-Titration ISE Type Calcium (Ca2 ) Electrode Name edit as desired Titrant Parameter Titrant Name EDTA Titrant ID edit as desired Conc Input Mode Standardization Nominal Concentration 10.0 mM Standardize Tech Equivalence Pt. Number of Endpoints 1 Results Units mM Standardize Reaction Ratio 1 Standard Name Calcium Standard Amount Fixed Volume, 1.0 mL Standard Concentration 0.10 M Pre-dose Titrant Volume 8 mL Max total titrant volume 16 mL Stand. Process Control Routine Pre-stir Duration 10 seconds Stir Speed Fast Titration Parameter Titration Technique Equivalence Pt Number of Endpoints 1 Display Units mV Titration Type Direct Blank Required No Result Units mg/L Reaction Ratio 1 Sample Amount Variable volume Sample MW 100.1 Pre-dose Titrant Volume 2.0 mL Max total titrant volume 4.0 mL Titration Process Control User Defined E 15 mV Vmin 0.050 mL Vmax 0.050 mL dE/dt 10 mV/min Tmin 2 seconds Tmax 10 seconds Threshold 100 Prestir Duration 10 seconds Stir Speed Fast Sample ID Manual

Table 2 Recommended sample volumes Expected sample concentration (mg/L) Sample amount (mL) Beaker size (mL) Add this volume of RGW to the beaker (mL) 15 200 250 --- 50 60 100 --- 100 30 100 To the 50 mL mark 200 15 100 To the 50 mL mark 300 10 100 To the 50 mL mark 500 6.0 100 To the 50 mL mark Sample Titration 1. From the Methods screen, select option to run the saved method T10 Ca Hard. 2. Rinse the electrode, stirrer, and dispenser with RGW. Place the electrode, stirrer, and dispenser into the prepared sample in the beaker. Ensure that the dispenser tip is inserted below the surface of the sample. 3. Start the titration. Enter the sample volume and sample ID when prompted. 4. Results reported are calcium hardness in mg/L CaCO3. Orion Star T930 ISE Titrator with 20 mL burette Table 3 Sample Results Results (mg/L as CaCO3) % Recovery RSD (n 4) Analysis Time (min:sec) Hardness QC Standard 182.0 101% 0.23% 02:30 min Tap Water 66.86 NA 0.87% 02:29 min Low Hardness Standard 15.31 102% 1.2% 02:55 min Sample 3

Figure 1. Calcium Hardness in Tap Water Range This preprogrammed titration method covers a range of about 15 to 500 mg/L of Hardness, when using 0.01 M (10 mM) EDTA titrant and the prescribed volume of sample. Titrant Over time, standard titrant solutions age and can change concentration. For higher accuracy, determine the exact concentration by standardizing the titrant. It is common to standardize on a weekly basis, but other standardization frequencies may be suitable. 1. Standardizing titrant a.Pipet 1.0 mL standardizing solution, 0.1M calcium standard, into a clean 100 mL beaker. Add RGW to the 50 mL mark on the side of the beaker. Add 2 mL of 0.1N NaOH. b.At the titration pre-check screen, select the standardize option. c.Start the titration. Standardization results will be displayed at the end of the cycle. d.For the most accurate results, run 3 or more cycles. The average value and relative standard deviation (RSD) of the cycles are automatically calculated when titrations are completed. e.The new standardized titrant concentration will automatically be saved and used for subsequent titrations for that method. 4 2. Certified Standardized Titrant Solutions a.Some customers may prefer not to standardize their titrant, instead choosing to purchase and use certified standardized titration solutions. In this case, edit the Titrant section of the method. Choose Conc. Input Mode Manual and enter the certified concentration and titrant ID (i.e., lot number, if desired). Titrator and Electrode Care Refer to the titrator and electrode user manuals for details on cleaning, storage, and maintenance recommendations to keep the titrator and electrode performing well. Main points for care are summarized below.

Daily care Weekly or biweekly care As needed If bubbles are visible in the titrator tubing, dispense titrant until bubbles have been expelled. Tap tubing to dislodge bubbles that stick. Drain and replace the fill solution of the electrode. For slow or drifty electrode response, soak the electrode in for 1 hour in RGW. Drain and replace the old filling solution, and soak the electrode for a few hours in freshly prepared storage solution. See the Storage Solution Preparation section above. Add electrode fill solution up to the bottom of the fill hole and leave the fill hole open during measurement. Change the electrode storage solution. Consider standardizing the titrant on a weekly basis, or more frequently, as desired. See the electrode user manuals for maintenance details. Rinse electrode well with RGW before and between titrations. Storage: Thoroughly rinse the electrode with RGW and store in storage solution. Cover the fill hole. Orion Star T930 ISE Titrator with 20 mL burette 5

Ordering Information Product Description Cat. No. Titrator Orion Star T930 Ion titrator without electrode Orion Star T940 All-in-One Titrator without electrode START9300 START9400 Electrode Orion Calcium ISE Electrode 9720BNWP Orion Solutions Orion Optimum Results A filling solution Orion Chloride standard solution, 0.1 M Orion Water Hardness Standard (optional) 900061 922006 923206 810017 Accessories Orion Electrode storage sleeve 100 & 250 mL beakers 100 & 250 mL graduated cylinders 10 mL graduated pipet 1.00 mL Pipete (optional for standardization) Reagents 0.01M EDTA titrant 1N (1M) Sodium Hydroxide Reagent Grade Water Thermo Scientific Barnstead Smart2Pure Pro UV Water Purification System 50129890* * Please contact a sales representative for support on ordering the best water purification system for your application, or find out more at thermofisher.com/purewater 1. SM 2340 C., Hardness, EDTA Titrimetric Method. Rice, Eugene W., Baird, Rodger B., Eaton, Andrew D., Lenore S. Clesceri. 2012. Standard methods for the examination of water and wastewater. Washington, DC: American Public Health Association. www. standardmethods.org 2. ASTM D1126, Standard Test Method for Hardness in Water. ASTM International, West Conshohocken, PA, USA. www.astm.org. 3. ISO 6059-1984, Water Quality – Determination of the sum of calcium and magnesium – EDTA titrimetric method. International Organization for Standardization, Geneva, Switzerland, 2017. www.iso.org. 4. SM 3500-Ca B., Calcium, EDTA Titrimetric Method. Rice, Eugene W., Baird, Rodger B., Eaton, Andrew D., Lenore S. Clesceri. 2012. Standard methods for the examination of water and wastewater. Washington, DC: American Public Health Association. www. standardmethods.org 5. ASTM D511, Standard Test Methods for Calcium and Magnesium in Water. ASTM International, West Conshohocken, PA, USA. www.astm.org. Find out more at thermofisher.com/titrator 2019 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. AN-CAHARDH2O-E RevA 0319

titrator. Drinking water, process water, cooling water, boiler water, wastewater, surface water, environmental water, raw water. Introduction Calcium Hardness in water is determined using the preprogrammed method, T10 Ca Hard. To determine calcium, sodium hydroxide is added to a sample to raise the pH to 12 to 13. The sample is then titrated to the

Related Documents:

This standard covers hardness conversions for metals and the relationship among Brinell hardness, Vick-ers hardness, Rockwell hardness, Superficial hardness, Knoop hardness, Scleroscope hardness and Leeb hardness. ASTM E10 (Brinell) This standard covers the Brinell test method as used by stationary, typically bench-top machines. This

g acceleration constant, 9.81m/s2 H nano, micro, macrohardness, kg/mm2,GPa HB Brinell hardness number, kg/mm2,GPa HBGM geometric mean of minimum and maximum Brinell hardness, kg/mm2,GPa HBK Berkovich hardness number, kg/mm2,GPa HK Knoop hardness number, kg/mm2,GPa HM Meyer hardness number, kg/mm2,GPa HRC Rockwell C hardness

A Leeb’s Hardness Tester measures the hardness of sample material in terms of Hardness Leeb (HL), which can be converted into other Hardness units (Rockwell B and C, Vicker, Brinell and Shore D). 1.3. Notation of Leeb’s Hardness When measuring the hardness of a sample materi

1. Define Hardness. 2. Applications of Rockwell Hardness A Scale, B-Scale, C-Scale. 3. Type of Indentor used in the Three Different Scales of Rockwell Hardness Test. 4. Different Types of Hardness Testing Methods. 5. Size of the Ball to be used in Ball Indentor of Rockwell Hardness Test. 6. Di ameters of the different Balls used in Brinell Hardness Test.

PESSE EASEMENT CORROSION AND MATERIAL SELECTION GUIDE TD/2600T/CORROSION-EN REV C 5 Material AISI 316L stainless steel Hastelloy C Monel Tantalum PTFE Elastomer (Viton) Calcium carbonate B B A A A A Calcium chlorate B A A A A Calcium chloride NR A B A A A Calcium fluoride A A A NR A Calcium hydroxide A A A A A 71 (160) Calcium hypochlorite 6 % NR NR NR A A 71 (160)

41NO2 Nitrites None Nitrites as NO2-N None 0.5L (CBWM) 250 40NH3 Ammonia None Ammonia as NH3/4-N None 0.5L (CBWM) 350 Total Hardness, Carbonate & Non-carbonate None 0.5L (CBWM) 550 Hardness Total HardnessNone & Components 39THC 38THM TH,Ca & Mg None TWH & hardness minerals (Calcium, Magnesium)None 0.5L (CBWM) 450 37THW Tot Hardness None Total Water Hardness (TWH) None 0.5L (CBWM) 250

1. ROCKWELL HARDNESS TEST 1. AIM: To determine the Rockwell Hardness of a given test specimen II. APPARATUS: Rockwell Hardness testing machine, Test specimen. III. THEORY: HARDNESS- It is defined as the resistance of a metal to plastic deformation against Indentation, scratching, abrasion of cutting. The hardness of a material by this Rockwell .File Size: 860KB

Using standard hardness conversion tables, the Rockwell hardness value is determined for the load applied, the diameter of the indenter, and the indentation depth. The hardness testing of plastics is most commonly measured by the Rockwell hardness test or Shore (Durometer D) hardness tes