Yokogawa TUS Report Software Quick Start Model Code M1286VB; Thermal Survey Reporting Software Yokogawa Corporation of America; Newnan, GA Control Instruments Business Division Revised 10/18/15 Document Number Copyright Yokogawa Corporation of America Sept 2014 -1-

Thermal Survey Reporting Software; M1286VB Overview: – This software produces AMS2750E compliant Temperature Uniformity Survey (TUS) reports using data files from Yokogawa DXAdvanced and SMARTDAC GX/GP data acquisition stations This document will guide you through installation and the steps needed to produce a report. You can select the Help tool within the software for additional information Supported file types: – DX series .DAD; Display Data File .DAE; Event Data File .DSD; Display Data File when Advanced Security is used .DSE; Event Data File when Advanced Security is used .GDS; Display Data File .GDE; Event Data File .GSD; Display Data File when Advanced Security is used .GSE; Event Data File when Advanced Security is used – GX/GP series Note: It is recommended to use Event data files for TUS reporting because they record the instantaneous channel values at a specified interval (1S, 2S, 5S, etc.) Document Number Copyright Yokogawa Corporation of America Sept 2014 -2-

Installation PC System requirements: – – – – – – – – – Windows Windows Windows Windows Windows Windows Windows Windows Windows XP 2003 Server Vista 7 (32-bit) 7 (64-bit) 2008 Server (32-bit) 2008 Server (64-bit) 8 2012 Server Run the .exe file on the CDROM. Allow the installer to use the default settings to complete the installation Document Number Copyright Yokogawa Corporation of America Sept 2014 -3-

License activation (1) Following installation, a permanent license must be keyed to the PC that it is installed on This is done by emailing an activation file to the software development group A license file is then returned to you that you must copy into the license folder Please complete the following steps Document Number Copyright Yokogawa Corporation of America Sept 2014 -4-

License activation (2) Start the TSR software Go to the Settings menu and select the Company Details tab Click License Click License Document Number Copyright Yokogawa Corporation of America Sept 2014 -5-

License activation (3) The License Information menu will open. Click Activate Complete the information in the Activation screen and click ok You will be prompted to save the Activation.dat file. Save it in this folder: – C:\Users\Public\Documents\Yokogawa\ TUS-Y Document Number Copyright Yokogawa Corporation of America Sept 2014 -6-

License activation (4) Email the Activation.dat file to: – activate@oceandatasys.com Within 24hrs, a file “License.dat” will be sent to you Copy this file into folder: – C:\Users\Public\Documents\Yokogawa\TUS-Y Start the TSR software. Go to the Settings menu and select the Company Details tab Click License and confirm that End User is specified and a license ID is present DONE Document Number Copyright Yokogawa Corporation of America Sept 2014 -7-

Configuration basics The TSR software allows the user to build highly customized TUS reports. You have to input and assign the following range of settings before you can run a report: – Information about the company and people performing the TUS – Information about the thermal equipment that was surveyed – Information about the data source, the recording equipment used, the channels, set point, stability criteria and other data settings for the TUS The above information can be saved to a Profile – Any number of Profiles can be saved and recalled for future use Document Number Copyright Yokogawa Corporation of America Sept 2014 -8-

AMS2750E Furnace Class and Temp Uniformity Range Furnace Class Temperature Uniformity Range ( F) 1 Temperature Uniformity Range ( C) 1 1 /- 5 /- 3 2 /- 10 /- 6 3 /- 15 /- 8 4 /- 20 /- 10 5 /- 25 /- 14 6 /- 50 /- 28 (1) Uniformity range requirement is established by the specification for the material being processed Document Number Copyright Yokogawa Corporation of America Sept 2014 -9-

Move the test files to the PC Copy the test data file(s) to your PC. Create a folder that you will specify as the data source in TUS – Example: C:\GX TUS source Copy one file at a time into this folder. TUS can work with multiple files in this folder, but work with one file at a time until you become familiar with the software Tip: – Use the Universal Viewer software to preview the test data. Confirm the start/stop date and time View the Trend display and use the Statistics tool to examine the data over the stability period You can learn quickly if the TUS report will produce pass/fail results if the P-P (peak to peak) data for each channel exceeds the allowed limits for the furnace category your are testing Document Number Copyright Yokogawa Corporation of America Sept 2014 - 10 -

TC Correction Factors The correction factors provided by the TC wire supplier must be input in the TSR software and visible in the generated report – A check box is provided to enable or disable the use of these factors in the report calculations Document Number Copyright Yokogawa Corporation of America Sept 2014 - 11 -

Calibration Correction; files without Advanced Security When Calibration Correction is used on DX (/CC1 option) and GX models (standard feature) the data stored in the data file is already corrected. When this feature is used, the correction factors for each TC channel must be clearly indicated in the TUS report; a requirement of AMS2750E. The TSR software can process a TUS report with recorder correction factors in the following ways: – When Using DX .DAD and .DAE files or GX .GDS and .GDE files that were recorded using calibration correction User must manually enter the recorder correction factors Do not check “Consider Recorder correction factors in calculations”. The settings will be displayed in the report but will not be applied to the calculations – When Using DX .DAD and .DAE file or GX .GDS and .GDE files that were recorded without calibration correction If you know the correction factors and wish to apply them to the raw data, the user can manually enter them Check “Consider Recorder correction factors in calculations” and the correction settings will be applied to the TUS calculations and displayed in the report Document Number Copyright Yokogawa Corporation of America Sept 2014 - 12 -

Calibration Correction; files with Advanced Security The TSR software can directly read the calibration correction settings from data files produced by DX and GX models equipped with the Advanced Security option; these settings are stored in the data files as a result of the DX/GX audit trail functions When Calibration Correction is used on DX (/CC1 option) and GX models (standard feature) the data stored in the data file is already corrected The TSR software can process a TUS report with recorder correction factors in the following ways: – When Using DX .DSD and .DSE files or GX .GSD and .GSE files that were recorded using calibration correction Click “Read from file”. TSR will automatically read the correction factors and display them in the table and TUS report The TSR software will not apply the correction factors in the TUS calculations because the software knows the data has already been corrected. They will only be displayed in the report – When Using DX .DAD and .DAE file or GX .GDS and .GDE files that were recorded without calibration correction If you know the correction factors and wish to apply them to the raw data, the user can manually enter them Check “Consider Recorder correction factors in calculations” and the correction factors will be applied to the TUS calculations and displayed in the report Document Number Copyright Yokogawa Corporation of America Sept 2014 - 13 -

Start the TSR software The above screen appears and you are ready to configure the software to run a report Document Number Copyright Yokogawa Corporation of America Sept 2014 - 14 -

Settings; Company Details Input all required Customer and Company information in the fields on this screen Document Number Copyright Yokogawa Corporation of America Sept 2014 - 15 -

Settings; Localization Select and apply the desired report attribute settings on this screen Document Number Copyright Yokogawa Corporation of America Sept 2014 - 16 -

Settings; Logo Management A company logo can be applied to the first page header. Click here to navigate to the file folder and select the .bmp or .jpg file Software version and correction factor usage information can be input in these two fields Document Number Copyright Yokogawa Corporation of America Sept 2014 - 17 -

Settings; Survey Engineers Enter the name(s) of the survey engineer and click Add to create a list Document Number Copyright Yokogawa Corporation of America Sept 2014 - 18 -

Settings; Profiles- Saving and Loading All settings can be saved to a Profile. Any number of profiles can be created, saved and re-loaded for future use. You can return to this screen and click Save to archive your latest settings at any time as you configure the TSR software Enter a profile name and description and click Save Document Number Copyright Yokogawa Corporation of America Sept 2014 - 19 - Select a saved profile from the list and click Load to retrieve a profile

Initial Data Input a range of information about the equipment surveyed, the customer, and test personnel – Equipment data settings – Customer data – Name of person generating the report – Survey engineer – Approval name A previously saved profile can also be loaded from this screen Document Number Copyright Yokogawa Corporation of America Sept 2014 - 20 -

Data Sources 2. Select the recorder model and file type 3. Click and navigate to the source folder 4. Select data by time period or batch 1. Select Survey Data Source 9. Repeat for control channels, if any 5. Click Connect 6. Click and choose the serial number of the recorder 7. Select All Channels if the TC channels are in multiple groups. You can also choose one group only Document Number Copyright Yokogawa Corporation of America Sept 2014 - 21 - 8. Available survey data range is shown here

Channel Setup Click Survey Channels Select the number of channels to be used in the survey The selected channels will appear in this table Input a TC serial number for each channel Document Number Copyright Yokogawa Corporation of America Sept 2014 - 22 -

Setpoint Setup Click and select the number of survey setpoints. One setpoint is shown in this example Enter the setpoint value and low and high tolerances Document Number Copyright Yokogawa Corporation of America Sept 2014 - 23 -

Survey Data; Instrumentation Input the recorder model and serial number, and description for the TC set used Document Number Copyright Yokogawa Corporation of America Sept 2014 - 24 -

Survey Data; TC Correction Factors When this box is checked, the correction factors in the table below are applied to the raw data and used in the report calculations. When left unchecked, they are not applied Input the TC correction factor for each TC. This is provided by the TC supplier Document Number Copyright Yokogawa Corporation of America Sept 2014 - 25 -

Survey Data; Recorder Correction Factors When this box is checked, the recorder calibration correction settings are displayed in the report. Calibration correction factors for DX .DAD and .DAE, and GX .GDS and .GES files must be entered manually in this table By clicking Read from file, calibration correction factors are read directly from DX .DSD and .DSE, and GX .GSD and .GSE files Document Number Copyright Yokogawa Corporation of America Sept 2014 - 26 -

Survey Data; TC Locations A furnace diagram can be loaded and displayed in the report. Click here to navigate to the folder containing the .bmp or .jpg file Enter drawing notes Enter the furnace working volume Enter comments for each TC Document Number Copyright Yokogawa Corporation of America Sept 2014 Enter the furnace dimensions; volume will be automatically - 27 calculated

Survey Data; Comments A range of descriptive text comments can be input in this table and applied to the report. Input the text here and click Add for each line of text Document Number Copyright Yokogawa Corporation of America Sept 2014 - 28 -

Stability Criteria Monitoring threshold is the temperature below the setpoint that is the beginning of the trend and the first marker on the trend graph Criteria type: Manual: Stability calculations start at date & time entered here Offset: Stability calculations start at low tolerance setpoint time minutes entered here Auto: Stability calculations start at low tolerance setpoint time Document Number Copyright Yokogawa Corporation of America Sept 2014 - 29 - Select the desired stable measure ment period This table will auto-fill with the Setpoint information. Choose Criteria type and stable date/time as needed

Thermal Images Up to six thermal images can be placed on the report to supplement the survey information. Supported image files are .bmp and .jpg. Click the checkbox and enter a description for the image Click here and navigate to the folder containing the image file and select the file Document Number Copyright Yokogawa Corporation of America Sept 2014 - 30 -

Trends Setup The trend display for each available TC channel can enabled or disabled via the check box selection in this table Document Number Copyright Yokogawa Corporation of America Sept 2014 - 31 -

Generate; the final step Select the desired printer and folder for the PDF output on this menu Click Generate preview; a PDF report will be created for review. If an error message is presented, make the stated setting corrections and repeat this step until a report is generated Select Printer or PDF for the report output Click Generate Report to produce and print or save the final report Document Number Copyright Yokogawa Corporation of America Sept 2014 - 32 -

Revision History Oct 18, 2015; revise text on slide 21; step 8; control channels Nov 17, 2014; first release Document Number Copyright Yokogawa Corporation of America Sept 2014 - 33 -

- This software produces AMS2750E compliant Temperature Uniformity Survey (TUS) reports using data files from Yokogawa DXAdvanced and SMARTDAC GX/GP data acquisition stations This document will guide you through installation and the steps needed to produce a report. You can select the Help tool within the software for additional information

Related Documents:

HART Communication Foundation Yokogawa Contact Information For customer Assistance Address Yokogawa Corporation of America 2 Dart Road Newnan, GA. 30265USA Telephone 1-800-888-6400 E-mail meters-instr@us.yokogawa.com . YHC5150X Series HART Communicator Overview

YOKOGAWA Yokogawa Corporation of America 2 Dart Road Newnan, GA 30265 To: Classis Controls From: Todd Garber, Yokogawa Corporation of America Date: December I, 2003 Subject: Classic Control’ s Yokogawa Territory Coverage This memo is to serve as verifica

and features of the Yokogawa Zirconia Oxygen Analyzers: Compact design for flexible low-cost installation Long-life sensor Reduced maintenance and calibration using self-diagnostics Excellent price/ performance correlation Yokogawa’s Zirconia Oxygen analyzer can also be use

Yokogawa released the first Sushi Sensor, a wireless vibration sensor, in March 2018. In July 2019, Yokogawa released wireless temperature and pressure sensors, expanding the sensor portfolio (Note 1: The wireless vibration sensor has been released in Japan and some other countries. Although the wireless pressure and temperature sensors are .

- This software produces AMS2750E compliant Temperature Uniformity Survey (TUS) reports using data files from Yokogawa DXAdvanced and SMARTDAC . - Information about the thermal equipment that was surveyed - Information about the data source, the recording equipment used, the channels, set point, stability

9 Honra a Jehová con tus bienes y con las primicias de todos tus frutos; 10 entonces tus graneros estarán colmados con abundancia y tus lagares rebosarán de mosto. 11 No menosprecies, hijo mío, el castigo de Jehová, no te canses de que él te corrija, 12 porque Jehová al que ama castiga, como el padre al hijo a quien quiere.

4.1 Quick Start Tutorial When Collect is opened for the first time, the Quick Start tutorial is displayed. Quick Start is a series of five screens that provide a summary review of the main capabilities of the app. Tap the Forward icon to move through the Quick Start screens. Tap Start to open the app after finishing with the Quick Start screens .

The Trading and Investment Strategies Interactive Qualifying Project is an in depth examination of the methods and strategies used on investable markets in order to gain long-lasting investing experience.