IMPLEMENTATION OF LEAK TEST METHODS FOR THE

2y ago
24 Views
2 Downloads
778.87 KB
17 Pages
Last View : 20d ago
Last Download : 3m ago
Upload by : Pierre Damon
Transcription

IMPLEMENTATION OF LEAK TEST METHODS FOR THE INTERNATION SPACESTATION (ISS) ELEMENTS, SYSTEMS AND COMPONENTSSteve Underwood(1), Oleg Lvovsky(2)The Boeing Company 499 Boeing Boulevard Huntsville, AL, USA, Email: steve.d.underwood@boeing.com(2)ARES Corp., 16441 Space Center Blvd., Bldg. A, Houston, TX, USA, Email: olvosvsky@arescorporation.com(1)ABSTRACTThe International Space Station (ISS has Qualificationand Acceptance Environmental Test Requirementsdocument, SSP 41172 that includes manyenvironmental tests such as Thermal vacuum &Cycling, Depress/Repress, Sinusoidal, Random, andAcoustic Vibration, Pyro Shock, rence(EMI)/Electromagnetic Compatibility (EMCO), etc.This document also includes (13) leak test methods forPressure Integrity Verification of the ISS Elements,Systems, and Components. These leak test methods arewell known, however, the test procedure for specificleak test method shall be written and implementedpaying attention to the important proceduralsteps/details that, if omitted or deviated, could impactthe quality of the final product and affect the crewsafety. Such procedural steps/details for differentmethods include, but not limited to:- Sequenceoftesting,forexample,pressurization and submersion steps forMethod I (Immersion);- Stabilization of the mass spectrometer leakdetector outputs for Method II (vacuumChamber or Bell jar);- Proper data processing and taking aconservativeapproachwhilemakingpredictions for on-orbit leakage rate forMethod III(Pressure Change);- Proper Calibration of the mass spectrometerleak detector for all the tracer gas (mostlyHelium) Methods such as Method V (DetectorProbe), Method VI (Hood), Method VII(Tracer Probe), Method VIII(Accumulation);- Usage of visiblility aides for Method I(Immersion), Method IV (Chemical Indicator),Method XII (Foam/Liquid Application), andMethod XIII (Hydrostatic/Visual Inspection);While some methods could be used for the total leakage(either internal-to-external or external-to-internal) raterequirement verification (Vacuum Chamber, PressureDecay, Hood, Accumulation), other methods shall beused only as a pass/fail test for individual joints (e.g.,welds, fittings, and plugs) or for troubleshootingpurposes (Chemical Indicator, Detector Probe, TracerProbe, Local Vacuum Chamber, Foam/LiquidApplication, and Hydrostatic/Visual Inspection). Anyisolation of SSP 41172 requirements have led to eitherretesting of hardware or accepting a risk associatedwith the potential system or component pressureintegrity problem during flight.INTRODUCTIONBoeing in was the prime contractor for building,outfitting, and testing of the U.S. pressurized modulesfor the International Space Station: the Airlock,Laboratory, and Node 1 modules. This work wasperformed at Marshall Space Flight Center inHuntsville, Alabama. Each of these modules has a totalallowable leakage rate of about one tenth of a pound ofair per day, a requirement based on the available resupply gas.Component-level and system-levelrequirements based on hardware capability are morestringent, and are the basis for the lower-level leak teststo ensure good hardware installations.The Node 1 or “Unity” module contains six ports and isused for interconnection of modules in space. The U.S.Laboratory module will house equipment andexperiment racks, and the Airlock module will providecrew members access to space for extravehicularactivity (EVA). Figure 1 shows the Node 1, U.S. Lab,and Airlock modules. Potential leakage paths to spaceinclude the module structure, made of machined andformed aluminum panels welded together, andfeedthrough penetrations in the module structuresthrough which fluids and power pass to different areason the station. Other leakage paths are the CommonBerthing Mechanisms (CBMs) that mate the variousmodules together, the Hatches that close out themodules, and windows that allow crew observations ofexternal activities and space. A summary flow chart ofthe leak testing activities for acceptance of the flighthardware is shown in Figure 2.

Hatch (6 places)Common Berthing Mechanism (CBM)(6 places)Node 1 End ViewU.S. LaboratoryNode 1 Side ViewAirlockFigure 1: Boeing-built International Space Station Pressurized Modules

ModuleStructureFab & AssyModule ProofPressure/Leak Rate TestFeedthroughInstallation &Leak TestFeedthroughFabrication& AssemblyAModuleFunctionalTestingCBMInstallation& Leak TestStandoff/EndconePlumbing Install& Leak TestHose/TubeFabrication& AssemblyHatchInstallation& Leak TestEquipmentRackInstallationAEquip. RackFab & Assy,Plumbing Leak TestTransportModuleto KSCBHatch Fab &Assy, ProofPress/Leak RateBWindow Install& Leak Test(Lab Module)ModuleAcceptanceLeak TestFinal ModuleCloseout andGross Leak TestLaunchFigure 2: International Space Station Module Acceptance Leak Test FlowSTRUCTURAL WELD LEAK TESTINGLeak tests of the structural welds were performedimmediately following a 22.8 psig proof pressure testof the modules in 1996 and 1997. The leak testmethod was a helium detector probe test usinghelium mass spectrometer leak detectors (MSLDs).The modules were pressurized to 12-14.7 psig for an87% helium concentration. Portable counter-flowMSLDs were used in conjunction with 75 foot snifferhoses, and welds were sniffed at a speed of 1foot/minute. Probe heads which conformed to theweld surfaces were installed over the tip of thedetector probes to increase test sensitivity.The requirement for weld leakage was 1E-4 sccshelium per linear inch of weld. Calibrated heliumleaks sized slightly smaller than the requirement wereinstalled on weld test plates and were used tocalibrate the detector probes by passing the probehead over the leak and recording the MSLDresponse. This calibration was performed prior tostart of sniffing, a minimum of every hour thereafter,whenever detector probe operators were changed,and before shutting down the test. The calibrationverified that the operator’s technique and theequipment’s operation were sufficient to detect a leakequal to or less than the requirement. Over 1000 feetof structural welds were tested in this manner.FEEDTHROUGH LEAK RATE TESTINGFeedthroughs are installed in the module structure toroute power, data, and fluids to the ISS modules. TheNode 1 module contains 160 feedthroughs, theLaboratory module contains 100 feedthroughs, andthe Airlock module contains 40 feedthroughs. Thefeedthroughs were leak tested following installationto verify a proper installation. The leak raterequirement for the feedthrough leak test was 5E-6sccs air, and was verified with a helium massspectrometer leak detector (MSLD) hood test. Thetest setup is shown in Figure 3.

The feedthrough leak test procedure is as follows: Abell jar is placed over the external side of thefeedthrough and evacuated to 1E-2 torr with anMSLD. A helium containment hood is installed overthe internal side of the feedthrough. Helium isinjected into the hood and the MSLD output isrecorded until the feedthrough seals are fullypermeated and a leak rate is obtained.In addition to verifying the total leak rate of thefeedthrough seals, the ISS also has a requirement toverify that the seal installation meets its redundancyrequirement. Each ISS sealed joint is required toHeliumContainment Hoodhave a minimum two seals, and must be tested on theground prior to launch. ISS seals larger than sixinches in diameter have leak check ports whichprovide access between the two seals for leak testingeach seal individually. The feedthrough seals,however, have no leak check ports, so theredundancy is verified through analysis of thefeedthrough permeation curves. (Permeation throughtwo seals is slower and lower in magnitude thanpermeation through one seal.) Many installationproblems were detected and corrected by using thepermeation data to verify seal redundancy.Bell JarFeedthroughUnder TestMass SpectrometerLeak Detector(MSLD)Figure 3: Feedthrough Leak Rate TestingPLUMBING LEAK TESTINGTubing and hoses in the International Space Stationroute fluid and gases throughout the Station. TheU.S. Laboratory Module also has a Vacuum Systemwhich provides experiment racks with a spacevacuum for experimentation and with an exhaustvacuum for venting unneeded gases. Plumbinginternal to the modules is routed through the modulestandoffs and endcones and in the equipment racks.The plumbing was thoroughly leak tested as it wasinstalled to ensure crew safety.Leak testing of the vacuum system plumbing wasperformed on the pre-integrated standoffs prior toinstallation of the standoffs into the module, and thensystem level tests were performed followingcompletion of the flight systems. The test methodwas a helium MSLD hood test and is illustrated inFigure 4 for one manifold of the Vacuum System.The manifolds under test were evacuated with anMSLD. A helium standard leak was connected to therack location farthest from the MSLD; the leak wasvalved onto the manifold in order to determine thesystem sensitivity to a known helium leak. The jointsin the manifold were bagged using plastic and tapeand injected with helium one at a time. The seals inthe vacuum system are metal gaskets, and each joint

had a leak rate requirement of 1E-8 sccs helium.RoughingPumpHeliumStandard LeakQuick Disconnectsat Equipment Rack LocationsTest EquipmentB1B3B5B7Bellows in VacuumSystem ManifoldB2B4B6B8B11B12B13MSLDTestEquipmentFigure 4: U.S. Laboratory Standoff Vacuum System Manifold Leak TestPressurized Plumbing SystemsISS pressurized plumbing systems include theThermal Control System which provides coolingwater to the electronic hardware and Oxygen andNitrogen Systems for air re-supply. Flex hoses andsystem components were acceptance tested by thevendor prior to delivery and installation. Theinstalled plumbing was leak tested by pressurizingthe systems with helium to maximum operatingpressure and performing helium detector probe testson the joints. A major challenge with the leak testingof the plumbing was the teflon-lined flex hoses thatpermeated helium, thus increasing the heliumbackground levels during testing. High quality/purityair was used to purge the helium from the jointsunder test during sniffing. A helium leak sizedslightly lower than the per joint requirement of 1E-4sccs helium was used to verify sensitivity before,during, and after testing of the flight joints.The CBM is used to join modules together in space.Each module port has a CBM bolted to it while onthe ground, and this joint is referred to as the CBMpressurized element, or CBM-PE joint. This jointcontains two elastomeric O-rings, the inner one madeof fluorocarbon and the outer of silicone. The ISSprogram has a requirement that all sealed joints thatcan leak to space must have two seals forredundancy. This requirement applies to joints thatcan be tested on the ground before launch. If aconnection is made in space, then three seals arerequired. This is the case when two modules aremated on-orbit to comprise the CBM-CBM joint,which contains three seal beads. The CBM-CBMseals consist of four sections of silicone seals moldedinto an aluminum substrate. Figure 5 shows a crosssection of a CBM with the CBM-PE and CBM-CBMseals installed.COMMON BERTHING MECHANISM (CBM)LEAK TESTINGCBM-PE Leak Testing

Leak testing of the CBM-PE seals is performed usingthe helium mass spectrometer hood method. All ISSseals greater than six inches in diameter weredesigned to accommodate leak testing with theincorporation of leak check ports which provideaccess to the volume between the seals. Two leakcheck ports are included for every seal volume, andthe ports are located one hundred and eighty (180)degrees apart.Module StructureModule Structure“Passive” CBM“Active” CBMLeak Check PortCBM-PE SealsCBM-CBM SealFigure 5: Common Berthing Mechanism SealsThe ports are fitted with leak check port adaptersduring leak testing, and are then plugged with “blindglands” following testing of the seals. Figure 6shows the leak test setup for a CBM-PE joint. AnMSLD is connected to one leak check port, and athree-valve manifold is connected to the other leakcheck port. The manifold is used with a roughingpump to evacuate the helium standard leak used tocalibrate the test volume during the systemcalibration.For the leak test, the volume between the seal beadsis evacuated by the MSLD, then the systemcalibration is performed during which a helium leakis introduced to the test volume and detected by theMSLD. After the MSLD is stable, the leak is valvedoff and the MSLD background returns to its originalvalue. The system sensitivity is calculated and islater used to determine the seal leak rate.Helium containment hoods are installed over theinner and outer CBM-PE joints. One of these hoodsis then inflated with 100% helium until the MSLD isstable. The hood is then removed, and the MSLDbackground cleans up, and then the other hood isinflated to test the other seal.

Outer O-ring (Silicone)CBM-PE JointInner hHeliumContainmentHoodFlight ModuleHeliumContainmentHoodFigure 6: CBM-PE Seal Leak TestingBlind Gland Leak TestingFollowing the leak test of the CBM-PE seals, theblind glands are installed and leak tested. This test iscritical because if a blind gland leaks, then the sealredundancy would be lost. First one of the leak testport adapters is removed and a blind gland isinstalled. The blind gland seals off the leak checkport volume with a 1/16” diameter silicone seal thatseats in the bottom of the leak check port hole. Ahelium containment hood is installed over the blindgland, and the volume between the seal beads isevacuated by the MSLD. The helium containmenthood is injected with helium and the seal is allowedto permeate.Following testing of the first blind gland installed,the volume between the seal beads, while still undera vacuum, is back-filled with 100% helium. The leakcheck port adapter is removed and the second blindgland is installed. The goal during this operation isto leave the leak check port uncovered for as littletime as possible to keep a 100% helium concentrationin the volume. A test fixture connected to the MSLDis sealed over the blind gland and evacuated with theMSLD, and the blind gland leak rate is determined.CBM-CBM Leak TestingLeak tests of the CBM-CBM seals for Acceptancewere not generally performed on the ground, as thisjoint will usually be made on-orbit. Node 1,however, was launched with two Pressurized MatingAdapters (PMAs) mated to it to provide berthingports for the shuttle, and these CBM-CBM jointswere tested prior to launch. The test procedure forthe CBM-CBM seals is very similar to that used onthe CBM-PE seals. Leak check ports provide accessto the volumes between each of the three seal beads.MSLDs are used to evacuate the volume between theseal beads. System calibrations are performed on thevolume between the outer and middle bead and on

the volume between the middle and inner bead.Helium containment hoods are installed over theouter and inner joint of the CBM-CBM to enabletesting of the outer and inner seal beads. The onlyreal difference in the CBM-CBM leak test conduct isthat in order to test the middle seal bead, the sealvolume between the middle and inner seals beads isused as the helium containment hood. This volume isevacuated and then back-filled with 100% helium totest the middle bead.HATCH LEAK TESTINGEach ISS module includes at least one commonHatch, and the Node modules have six hatches each.When two modules are connected, the hatchesbetween them will generally be open, however, foran unused port the Hatch is exposed to the vacuum ofspace. See Figure 7 for an illustration of the Hatch;Figure 8 shows cross sections of the Hatch seals.The aluminum Hatch seals to the module via aPerimeter Seal, a two-beaded silicone rubber moldedseal with a stainless steel substrat

Figure 2: International Space Station Module Acceptance Leak Test Flow STRUCTURAL WELD LEAK TESTING Leak tests of the structural welds were performed immediately following a 22.8 psig proof pressure test of the modules in 1996 and 1997. The leak test method was a helium detector probe test using helium m

Related Documents:

the tank itself, API standards prescribe provisions for leak prevention, leak detection, and leak containment. It is useful to distinguish between leak prevention, leak detection and leak containment to better understand the changes that have occurred in tank standards over the years. In simple terms, leak prevention is any process that is designed to deter a leak from occurring in the first .

LIST OF FIGURES . Number Page . 1.1 Subsea Development USA Gulf of Mexico 5 1.2(a) Liquid Only Leak 6 1.2(b) Gas Only Leak 6 2.1 Categorization of Leak Detection Technology Used 10 in this Study 2.1.1 Leak Detection by Acoustic Emission Method 12 2.1.2 Leak Detection by Sensor Tubes 14 2.1.3 Leak Detection by Fiber Optical Sensing 17 2.1.4 Leak Detection by Soil Monitoring 19

8 Leak Detection Techniques Using Vacuum Leak Detectors 8 9 Industrial Leak Test 9. 3 Fundamentals of Leak Detection . by a special apparatus and fed to the leak detector. This procedure can be carried out using either helium, refrigerants, SF6 as the test gas. . Leak test using vacuum gauges which are sensitive to the type of gas. 6

simulated or misapplied test methods could be misinterpreted as a valid representation of the hardware performance. In this paper, the soundness of leak test methods used to verify the air leak rate of spacecraft sealing systems is investigated. Two common methods to evaluate the leak rate of gas

3.10 Evaluation of the Vacuus Leak Test 25 3.11 Evaluation of Large Experimental Leaks 30 3.12 Evaluation of the Liquid Nitrogen Leak Test 35 3.12.1 Tests Using Glass Ampuls 35 3.12.2 Tests on 8wSr Sources 36 4. REFERENCES 36 APPENDIX A, Approximate Leak Test Sensitivities 38 APPENDIX B, Examples of Typical Leak Test Procedures 39

The wide range of leak rates from Besides the determination of the total several 100 mbar x l/s to below 10-11 mbar x l/s as they occur in practi-ce necessitates the use of different leak detection principles and hence leak detectors (see figure). leak tightness, it is usually important to locate the leak, quickly and precisely,

constant leak sound value of a leak and the total current sound value. In addition, the results of each measurement can be automatically stored for a comparison of leak sound levels along a series of sequential listening points (see Section 3.8 on Memory Mode, pp 18 - 19). The HL 5000 is the first leak locator that when in leak

Division and 3-505 Parachute Infantry Regiment on 4 August 1990. My company, Charlie 3-505, had been conducting night live-fire exercises at Fort Bragg, North Carolina. Around 2230 hours on the night of 4 August, I received a Warning Order from my commander, Captain Charles Dydasco, to prepare for movement to the Battalion Area. Shortly after midnight, in a torrential downpour, we began .