Agilent InfiniiVision2000 X-SeriesOscilloscopesUser's Guide

Notices Agilent Technologies, Inc. 2005-2011WarrantyNo part of this manual may be reproduced inany form or by any means (including electronic storage and retrieval or translationinto a foreign language) without prior agreement and written consent from AgilentTechnologies, Inc. as governed by UnitedStates and international copyright laws.The material contained in this document is provided “as is,” and is subject to being changed, without notice,in future editions. Further, to the maximum extent permitted by applicablelaw, Agilent disclaims all warranties,either express or implied, with regardto this manual and any informationcontained herein, including but notlimited to the implied warranties ofmerchantability and fitness for a particular purpose. Agilent shall not beliable for errors or for incidental orconsequential damages in connection with the furnishing, use, or performance of this document or of anyinformation contained herein. ShouldAgilent and the user have a separatewritten agreement with warrantyterms covering the material in thisdocument that conflict with theseterms, the warranty terms in the separate agreement shall control.Manual Part Number75015-97011EditionSecond edition, February 2011Printed in MalaysiaAgilent Technologies, Inc.1900 Garden of the Gods RoadColorado Springs, CO 80907 USAPrint History75015-97000, January 201175015-97011, February 2011Technology LicensesTrademarksThe hardware and/or software described inthis document are furnished under a licenseand may be used or copied only in accordance with the terms of such license.Java is a U.S. trademark of Sun Microsystems, Inc.Sun, Sun Microsystems, and the Sun Logoare trademarks or registered trademarks ofSun Microsystems, Inc. in the U.S. and othercountries.2Restricted Rights LegendIf software is for use in the performance of aU.S. Government prime contract or subcontract, Software is delivered and licensed as“Commercial computer software” asdefined in DFAR 252.227-7014 (June 1995),or as a “commercial item” as defined in FAR2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June1987) or any equivalent agency regulation orcontract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial licenseterms, and non-DOD Departments andAgencies of the U.S. Government willreceive no greater than Restricted Rights asdefined in FAR 52.227-19(c)(1-2) (June1987). U.S. Government users will receiveno greater than Limited Rights as defined inFAR 52.227-14 (June 1987) or DFAR252.227-7015 (b)(2) (November 1995), asapplicable in any technical data.Safety NoticesCAUTIONA CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the likethat, if not correctly performed oradhered to, could result in damageto the product or loss of importantdata. Do not proceed beyond aCAUTION notice until the indicatedconditions are fully understood andmet.WA R N I N GA WARNING notice denotes ahazard. It calls attention to anoperating procedure, practice, orthe like that, if not correctly performed or adhered to, could resultin personal injury or death. Do notproceed beyond a WARNINGnotice until the indicated conditions are fully understood andmet.Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide

InfiniiVision 2000 X-Series Oscilloscopes—At a GlanceTable 12000 X-Series Model Numbers, BandwidthsBandwidth70 MHz100 MHz200 MHz2-Channel 8 LogicChannels MSOMSO-X 2002AMSO-X 2012AMSO-X 2022A4-Channel 8 LogicChannels MSOMSO-X 2004AMSO-X 2014AMSO-X 2024A2-Channel DSODSO-X 2002ADSO-X 2012ADSO-X 2022A4-Channel DSODSO-X 2004ADSO-X 2014ADSO-X 2024AThe Agilent InfiniiVision 2000 X- Series oscilloscopes deliver these features: 70 MHz, 100 MHz, and 200 MHz bandwidth models. 2- and 4- channel digital storage oscilloscope (DSO) models. 2 8- channel and 4 8- channel mixed- signal oscilloscope (MSO) models.Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide3

An MSO lets you debug your mixed- signal designs using analog signalsand tightly correlated digital signals simultaneously. The 8 digitalchannels have a 1 GSa/s sample rate, with a 50 MHz toggle rate. 8.5 inch WVGA display. Interleaved 2 GSa/s or non- interleaved 1 GSa/s sample rate. Interleaved 100 Kpts or non- interleaved 50 Kpts MegaZoom IV memoryfor the fastest waveform update rates, uncompromised. All knobs are pushable for making quick selections. Trigger types: edge, pulse width, pattern, and video. Math wavefroms: add, subtract, multiply, and FFT. Reference waveforms (2) for comparing with other channel or mathwaveforms. Many built- in measurements. Built- in license- enabled waveform generator with: sine, square, ramp,pulse, DC, noise. USB ports make printing, saving and sharing data easy. Optional LAN/VGA module for connecting to a network and displayingthe screen on a different monitor. Optional GPIB module. A Quick Help system is built into the oscilloscope. Press and hold anykey to display Quick Help. Complete instructions for using the quickhelp system are given in “Access the Built- In Quick Help" on page 40.For more information about InfiniiVision oscilloscopes, see:"www.agilent.com/find/scope"4Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide

In This GuideThis guide shows how to use the InfiniiVision 2000 X- Series oscilloscopes.When unpacking and using theoscilloscope for the first time, see: Chapter 1, “Getting Started,” starting on page 19When displaying waveforms andacquired data, see: When setting up triggers or changinghow data is acquired, see: Chapter 9, “Triggers,” starting on page 111 Chapter 10, “Trigger Mode/Coupling,” starting onpage 131 Chapter 11, “Acquisition Control,” starting on page 139Making measurements and analyzingdata: Chapter 12, “Cursors,” starting on page 155 Chapter 13, “Measurements,” starting on page 163 Chapter 14, “Mask Testing,” starting on page 183When using the built-in waveformgenerator, see: Chapter 15, “Waveform Generator,” starting on page195When saving, recalling, or printing,see: Chapter 16, “Save/Recall (Setups, Screens, Data),”starting on page 201 Chapter 17, “Print (Screens),” starting on page 215When using the oscilloscope's utilityfunctions or web interface, see: Chapter 18, “Utility Settings,” starting on page 219 Chapter 19, “Web Interface,” starting on page 237For reference information, see: Chapter 20, “Reference,” starting on page 251Chapter 2, “Horizontal Controls,” starting on page 43Chapter 3, “Vertical Controls,” starting on page 55Chapter 4, “Math Waveforms,” starting on page 63Chapter 5, “Reference Waveforms,” starting on page75 Chapter 6, “Digital Channels,” starting on page 79 Chapter 7, “Display Settings,” starting on page 99 Chapter 8, “Labels,” starting on page 105Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide5

TIPAbbreviated instructions for pressing a series of keys and softkeysInstructions for pressing a series of keys are written in an abbreviated manner. Instructionsfor pressing [Key1], then pressing Softkey2, then pressing Softkey3 are abbreviated asfollows:Press [Key1] Softkey2 Softkey3.The keys may be a front panel [Key] or a Softkey. Softkeys are the six keys located directlybelow the oscilloscope display.6Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide

ContentsInfiniiVision 2000 X-Series Oscilloscopes—At a GlanceIn This Guide135Getting StartedInspect the Package Contents19Install the Optional LAN/VGA or GPIB ModuleTilt the Oscilloscope for Easy ViewingPower-On the Oscilloscope2223Connect Probes to the Oscilloscope24Maximum input voltage at analog inputsDo not float the oscilloscope chassisInput a Waveform242425Recall the Default Oscilloscope SetupUse Auto Scale222526Compensate Passive Probes27Learn the Front Panel Controls and Connectors 28Front Panel Overlays for Different Languages 35Learn the Rear Panel Connectors37Learn the Oscilloscope Display38Access the Built-In Quick Help40Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide7

2Horizontal ControlsTo adjust the horizontal (time/div) scale44To adjust the horizontal delay (position)45Panning and Zooming Single or Stopped Acquisitions46To change the horizontal time mode (Normal, XY, or Roll)XY Time Mode 48To display the zoomed time base4750To change the horizontal scale knob's coarse/fine adjustmentsetting 52To position the time reference (left, center, right)Navigating the Time BaseTo navigate time 53To navigate segments3525353Vertical ControlsTo turn waveforms on or off (channel or math)To adjust the vertical scaleTo adjust the vertical positionTo specify channel couplingTo specify bandwidth limiting5657575758To change the vertical scale knob's coarse/fine adjustmentsetting 58To invert a waveform59Setting Analog Channel Probe Options 59To specify the channel units 60To specify the probe attenuation 60To specify the probe skew 618Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide

4Math WaveformsTo display math waveforms63To perform a transform function on an arithmetic operationTo adjust the math waveform scale and offsetMultiply646565Add or Subtract66FFT Measurement 67FFT Measurement Hints 70FFT Units 72FFT DC Value 72FFT Aliasing 72FFT Spectral Leakage 74Units for Math Waveforms574Reference WaveformsTo save a waveform to a reference waveform locationTo display a reference waveform76To scale and position reference waveformsTo adjust reference waveform skew757777To display reference waveform information78To save/recall reference waveform files to/from a USB storagedevice 786Digital ChannelsTo connect the digital probes to the device under testProbe cable for digital channels80Acquiring waveforms using the digital channelsAgilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide79839

To display digital channels using AutoScaleInterpreting the digital waveform display8384To change the displayed size of the digital channelsTo switch a single channel on or off86To switch all digital channels on or off86To switch groups of channels on or off86To change the logic threshold for digital channelsTo reposition a digital channel858687To display digital channels as a bus88Digital channel signal fidelity: Probe impedance andgrounding 91Input Impedance 92Probe Grounding 94Best Probing Practices 96To replace digital probe leads7Display SettingsTo adjust waveform intensityTo set or clear persistenceTo clear the displayTo freeze the display99101102To adjust the grid brightness896102103LabelsTo turn the label display on or off105To assign a predefined label to a channelTo define a new label10106107Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide

To load a list of labels from a text file you createTo reset the label library to the factory default9108109TriggersAdjusting the Trigger LevelForcing a TriggerEdge Trigger112113113Pattern Trigger 116Hex Bus Pattern TriggerPulse Width Trigger118119Video Trigger 121To trigger on a specific line of video 124To trigger on all sync pulses 126To trigger on a specific field of the video signal 126To trigger on all fields of the video signal 127To trigger on odd or even fields 12810Trigger Mode/CouplingTo select the Auto or Normal trigger modeTo select the trigger coupling134To enable or disable trigger noise rejectionTo enable or disable trigger HF RejectTo set the trigger holdoffExternal Trigger Input132135135136136Maximum voltage at oscilloscope external trigger inputAgilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide13711

11Acquisition ControlRunning, Stopping, and Making Single Acquisitions (RunControl) 139Overview of Sampling 141Sampling Theory 141Aliasing 141Oscilloscope Bandwidth and Sample RateOscilloscope Rise Time 143Oscilloscope Bandwidth Required 144Memory Depth and Sample Rate 145142Selecting the Acquisition Mode 145Normal Acquisition Mode 146Peak Detect Acquisition Mode 146Averaging Acquisition Mode 149High Resolution Acquisition Mode 151Acquiring to Segmented Memory 151Navigating Segments 153Infinite Persistence with Segmented MemorySegmented Memory Re-Arm Time 154Saving Data from Segmented Memory 15412CursorsTo make cursor measurementsCursor Examples13156158MeasurementsTo make automatic measurementsMeasurements SummarySnapshot All 167Voltage Measurements12153164165167Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide

Peak-Peak 168Maximum 168Minimum 168Amplitude 168Top 169Base 170Overshoot 170Preshoot 171Average 172DC RMS 172AC RMS 173Time Measurements 175Period 175Frequency 176 Width 177– Width 177Burst Width 177Duty Cycle 177Rise Time 178Fall Time 178Delay 178Phase 179Measurement Thresholds180Measurement Window with Zoom Display14182Mask TestingTo create a mask from a "golden" waveform (Automask)Mask Test Setup OptionsMask Statistics183185188To manually modify a mask fileAgilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide18913

Building a Mask File 192How is mask testing done?15194Waveform GeneratorTo select generated waveform types and settingsTo output the waveform generator sync pulseTo specify the waveform generator output loadTo use waveform generator logic presetsTo restore waveform generator defaults16195198199199200Save/Recall (Setups, Screens, Data)Saving Setups, Screen Images, or Data 201To save setup files 203To save BMP or PNG image files 203To save CSV, ASCII XY, or BIN data files 204To save ALB data files 205Length Control 207To save reference waveform files to a USB storage deviceTo save masks 209To navigate storage locations 210To enter file names 210209Recalling Setups, Masks, or Reference Waveforms 211To recall setup files 211To recall mask files 212To recall reference waveform files from a USB storagedevice 212Recalling Default SetupsPerforming a Secure Erase14212213Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide

17Print (Screens)To print the oscilloscope's display215To set up network printer connectionsTo specify the print optionsTo specify the palette option18216217218Utility SettingsI/O Interface Settings219Setting up the Oscilloscope's LAN Connection 220To establish a LAN connection 221Stand-alone (Point-to-Point) Connection to a PC 222File Explorer223Setting Oscilloscope Preferences 225To choose "expand about" center or groundTo disable/enable transparent backgroundsTo load the default label library 226To set up the screen saver 226To set AutoScale preferences 227Setting the Oscilloscope's Clock225226228Setting the Rear Panel TRIG OUT Source228Performing Service Tasks 229To perform user calibration 230To perform hardware self test 232To perform front panel self test 232To display oscilloscope information 232To display the user calibration status 233To clean the oscilloscope 233To check warranty and extended services statusTo contact Agilent 233Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide23315

To return the instrument233Configuring the [Quick Action] Key19234Web InterfaceAccessing the Web Interface238Browser Web Control 239Remote Front Panel 240Remote Programming via the Web Interface 241Remote Programming with Agilent IO Libraries 242Save/Recall 243Saving Files via the Web Interface 243Recalling Files via the Web Interface 244Get Image245Identification Function20246Instrument Utilities246Setting a Password247ReferenceSpecifications and Characteristics251Measurement Category 251Oscilloscope Measurement Category 252Measurement Category Definitions 252Transient Withstand Capability 253Maximum input voltage at analog inputsMaximum input voltage at digital channelsEnvironmental ConditionsProbes and Accessories16253253253254Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide

Passive Probes 254Differential Probes 255Current Probes 255Accessories Available 256Loading Licenses and Displaying License InformationLicensed Options Available 257Other Options Available 258Upgrading to an MSO 258Software and Firmware Updates258Binary Data (.bin) Format 258Binary Data in MATLAB 259Binary Header Format 260Example Program for Reading Binary DataExamples of Binary Files 262CSV and ASCII XY files 265CSV and ASCII XY file structure 266Minimum and Maximum Values in CSV FilesAcknowledgements257262266267IndexAgilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide17

18Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide

Agilent InfiniiVision 2000 X-Series OscilloscopesUser's Guide1Getting StartedInspect the Package Contents 19Tilt the Oscilloscope for Easy Viewing 22Power-On the Oscilloscope 23Connect Probes to the Oscilloscope 24Input a Waveform 25Recall the Default Oscilloscope Setup 25Use Auto Scale 26Compensate Passive Probes 27Learn the Front Panel Controls and Connectors 28Learn the Rear Panel Connectors 37Learn the Oscilloscope Display 38Access the Built-In Quick Help 40This chapter describes the steps you take when using the oscilloscope forthe first time.Inspect the Package Contents Inspect the shipping container for damage.If your shipping container appears to be damaged, keep the shippingcontainer or cushioning material until you have inspected the contentsof the shipment for completeness and have checked the oscilloscopemechanically and electrically. Verify that you received the following items and any optionalaccessories you may have ordered: 19

1Getting Started InfiniiVision 2000 X- Series oscilloscope. Power cord (country of origin determines specific type). Oscilloscope probes: Two probes for 2- channel models. Four probes for 4- channel models. Documentation CD- ROM.20Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide

Getting Started1InfiniiVision 2000 X-Series oscilloscopeN2862B probes(Qty 2 or 4)Documentation CDDigital Probe Kit*(MSO models only)Power cord(Based on countryof origin)*N6459-60001 Digital Probe Kit contains:N6459-61601 8-channel cable (qyt 1)01650-82103 2-inch probe ground leads (qyt 3)5090-4832Grabber (qty 10)Digital probe replacement parts are listed in the"Digital Channels" chapter.See Also “Accessories Available" on page 256Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide21

1Getting StartedInstall the Optional LAN/VGA or GPIB ModuleIf you need to install a DSOXLAN LAN/VGA module or a DSOXGPIB GPIBmodule, perform this installation before you power on the oscilloscope.1 If you need to remove a module before installing a different module,pinch the module's spring tabs, and gently remove the module from theslot.2 To install a module, slide the module into the slot on the back until itis fully seated.The module's spring tabs will latch into the slot, keeping the module inplace.LAN/VGA ModuleModule SlotGPIB ModuleL6GC C / B6 CI6 C GDJC9 ID 6KD 9:A:8IG 8 H D8@s &%%"&'%K! *% % )%% os &%%"')%K! *% % o&%% LViih B6MNOTEThe LAN/VGA or GPIB module must be installed before powering on the oscilloscope.Tilt the Oscilloscope for Easy ViewingThere are tabs under the oscilloscope's front feet that can be flipped outto tilt the oscilloscope.22Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide

1Getting StartedFlip-Out TabsPower-On the OscilloscopePowerRequirementsLine voltage, frequency, and power: Line 100- 120 Vac, 50/60/400 Hz 100- 240 Vac, 50/60 Hz 100 W maxVentilationRequirementsThe air intake and exhaust areas must be free from obstructions.Unrestricted air flow is required for proper cooling. Always ensure thatthe air intake and exhaust areas are free from obstructions.The fan draws air in from the left side and bottom of the oscilloscope andpushes it out behind the oscilloscope.When using the oscilloscope in a bench- top setting, provide at least 2"clearance at the sides and 4" (100 mm) clearance above and behind theoscilloscope for proper cooling.To power-on theoscilloscope1 Connect the power cord to the rear of the oscilloscope, then to asuitable AC voltage source. Route the power cord so the oscilloscope'sfeet and legs do not pinch the cord.Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide23

1Getting Started2 The oscilloscope automatically adjusts for input line voltages in therange 100 to 240 VAC. The line cord provided is matched to the countryof origin.WA R N I N GAlways use a grounded power cord. Do not defeat the power cord ground.3 Press the power switch.The power switch is located on the lower left corner of the front panel.The oscilloscope will perform a self- test and will be operational in a fewseconds.Connect Probes to the Oscilloscope1 Connect the oscilloscope probe to an oscilloscope channel BNCconnector.2 Connect the probe's retractable hook tip to the point of interest on thecircuit or device under test. Be sure to connect the probe ground leadto a ground point on the circuit.CAUTIONMaximum input voltage at analog inputsCAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpkWith 10073C 10:1 probe: CAT I 500 Vpk, CAT II 400 VpkWith N2862A or N2863A 10:1 probe: 300 VrmsCAUTIONDo not float the oscilloscope chassisDefeating the ground connection and "floating" the oscilloscope chassis will probablyresult in inaccurate measurements and may also cause equipment damage. The probeground lead is connected to the oscilloscope chassis and the ground wire in the powercord. If you need to measure between two live points, use a differential probe withsufficient dynamic range.24Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide

Getting StartedWA R N I N G1Do not negate the protective action of the ground connection to the oscilloscope. Theoscilloscope must remain grounded through its power cord. Defeating the groundcreates an electric shock hazard.Input a WaveformThe first signal to input to the oscilloscope is the Demo 2, Probe Compsignal. This signal is used for compensating probes.1 Connect an oscilloscope probe from channel 1 to the Demo 2 (ProbeComp) terminal on the front panel.2 Connect the probe's ground lead to the ground terminal (next to theDemo 2 terminal).Recall the Default Oscilloscope SetupTo recall the default oscilloscope setup:1 Press [Default Setup].The default setup restores the oscilloscope's default settings. This placesthe oscilloscope in a known operating condition. The major default settingsare:Table 2Default Configuration SettingsHorizontalNormal mode, 100 µs/div scale, 0 s delay, center time reference.Vertical (Analog)Channel 1 on, 5 V/div scale, DC coupling, 0 V position.TriggerEdge trigger, Auto trigger mode, 0 V level, channel 1 source, DC coupling,rising edge slope, 40 ns holdoff time.DisplayPersistence off, 20% grid intensity.OtherAcquire mode normal, [Run/Stop] to Run, cursors and measurements off.LabelsAll custom labels that you have created in the Label Library are preserved (noterased), but all channel labels will be set to their original names.Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide25

1Getting StartedIn the Save/Recall Menu, there are also options for restoring the completefactory settings (see “Recalling Default Setups" on page 212) or performinga secure erase (see “Performing a Secure Erase" on page 213).Use Auto ScaleUse [Auto Scale] to automatically configure the oscilloscope to best displaythe input signals.1 Press [Auto Scale].You should see a waveform on the oscilloscope's display similar to this:2 If you want to return to the oscilloscope settings that existed before,press Undo AutoScale.3 If you want to enable "fast debug" autoscaling, change the channelsautoscaled, or preserve the acquisition mode during autoscale, pressFast Debug, Channels, or Acq Mode.These are the same softkeys that appear in the AutoScale PreferencesMenu. See “To set AutoScale preferences" on page 227.26Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide

1Getting StartedIf you see the waveform, but the square wave is not shaped correctly asshown above, perform the procedure “Compensate Passive Probes" onpage 27.If you do not see the waveform, make sure the probe is connectedsecurely to the front panel channel input BNC and to the left side,Demo 2, Probe Comp terminal.How AutoScaleWorksAuto Scale analyzes any waveforms present at each channel and at theexternal trigger input. This includes the digital channels, if connected.Auto Scale finds, turns on, and scales any channel with a repetitivewaveform that has a frequency of at least 25 Hz, a duty cycle greater than0.5%, and an amplitude of at least 10 mV peak- to- peak. Any channels thatdo not meet these requirements are turned off.The trigger source is selected by looking for the first valid waveformstarting with external trigger, then continuing with the lowest numberanalog channel up to the highest number analog channel, and finally (ifdigital probes are connected) the highest number digital channel.During Auto Scale, the delay is set to 0.0 seconds, the horizontal time/div(sweep speed) setting is a function of the input signal (about 2 periods ofthe triggered signal on the screen), and the triggering mode is set to Edge.Compensate Passive ProbesEach oscilloscope passive probe must be compensated to match the inputcharacteristics of the oscilloscope channel to which it is connected. Apoorly compensated probe can introduce significant measurement errors.1 Input the Probe Comp signal (see “Input a Waveform" on page 25).2 Press [Default Setup] to recall the default oscilloscope setup (see “Recallthe Default Oscilloscope Setup" on page 25).3 Press [Auto Scale] to automatically configure the oscilloscope for theProbe Comp signal (see “Use Auto Scale" on page 26).4 Press the channel key to which the probe is connected ([1], [2], etc.).5 In the Channel Menu, press Probe.6 In the Channel Probe Menu, press Probe Check; then, follow theinstructions on- screen.Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide27

1Getting StartedIf necessary, use a nonmetallic tool (supplied with the probe) to adjustthe trimmer capacitor on the probe for the flattest pulse possible.On the N2862/63/90 probes, the trimmer capacitor is the yellowadjustment on the probe tip. On other probes, the trimmer capacitor islocated on the probe BNC connector.Perfectly compensatedOver compensatedUnder compensated7 Connect probes to all other oscilloscope channels (channel 2 of a2- channel oscilloscope, or channels 2, 3, and 4 of a 4- channeloscilloscope).8 Repeat the procedure for each channel.Learn the Front Panel Controls and ConnectorsOn the front panel, key refers to any key (button) you can press.Softkey specifically refers to the six keys that are directly below thedisplay. The legend for these keys is directly above them, on the display.Their functions change as you navigate through the oscilloscope's menus.For the following figure, refer to the numbered descriptions in the tablethat follows.28Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide

1Getting Started5. Tools keys6. Trigger controls7. Horizontal controls8. Run Control keys9. [Default Setup] key10. [Auto Scale] key11. Additionalwaveformcontrols4. Entry knob3. [Intensity] key12. Measure controls13. Waveform keys2. Softkeys14. File keys1. Power switch15. [Help] key21. Waveformgeneratoroutput20. Digitalchannelinputs19. USBHostport18. Demo 2, Ground, 17. Analog 16 Vertical controlsand Demo 1channelterminalsinputs1.Power switchPress once to switch power on; press again to switch power off. See“Power-On the Oscilloscope" on page 23.2.SoftkeysThe functions of these keys change based upon the menus shown on thedisplay directly above the keys.TheBackBack/Up key moves up in the softkey menu hierarchy. At thetop of the hierarchy, the Back Back/Up key turns the menus off, andoscilloscope information is shown instead.3.[Intensity] keyPress the key to illuminate it. When illuminated, turn the Entry knob toadjust waveform intensity.You can vary the intensity control to bring out signal detail, much like ananalog oscilloscope.Digital channel waveform intensity is not adjustable.More details about using the Intensity control to view signal detail are on“To adjust waveform intensity" on page 99.Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide29

1Getting Started4.Entry knobThe Entry knob is used to select items from menus and to change values.The function of the Entry knob changes based upon the current menuand softkey selections.Note that the curved arrow symbolabove the entry knobilluminates whenever the entry knob can be used to select a value. Also,note that when the Entry knobsymbol appears on a softkey, youcan use the Entry knob, to select values.Often, rotating the Entry knob is enough to make a selection. Sometimes,you can push the Entry knob to enable or disable a selection. Pushing theEntry knob also makes popup menus disappear.305.Tools keysThe Tools keys consist of: [Utility] key — Press this key to access the Utility Menu, which letsyou configure the oscilloscope's I/O settings, use the file explorer, setpreferences, access the service menu, and choose other options. SeeChapter 18, “Utility Settings,” starting on page 219. [Quick Action] key — Press this key to perform the selected quickaction: measure all snapshot, print, save, recall, freeze display. andmore. See “Configuring the [Quick Action] Key" on page 234. [Analyze] key — Press this key to access analysis features like masktesting (see Chapter 14, “Mask Testing,” starting on page 183) or toset trigger levels or measurement thresholds. [Wave Gen] key — Press this key to access waveform generatorfunctions. See Chapter 15, “Waveform Generator,” starting on page195.6.Trigger controlsThese controls determine how the oscilloscope triggers to capture data.See Chapter 9, “Triggers,” starting on page 111 and Chapter 10, “TriggerMode/Coupling,” starting on page 131.Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide

Getting Started7.Horizontalcontrols1The Horizontal controls consist of: Horizontal

MSO-X 2004A MSO-X 2014A MSO-X 2024A 2-Channel DSO DSO-X 2002A DSO-X 2012A DSO-X 2022A 4-Channel DSO DSO-X 2004A DSO-X 2014A DSO-X 2024A. 4 Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide the screen on a different monitor. An MSO lets

Related Documents:

Agilent 1290 Infinity Agilent G1888A X X Agilent 7673A Agilent 7683A Agilent HS7694 X X Agilent 7695A X Agilent 79855(A) X Agilent 5880 Agilent 5890 Agilent 6850 (27 Pos. Einlegeschale) . Autosampler-Kompatibilitätstabelle 2. H eadline as disp

Agilent 1290 Infinity X Agilent G1888A Agilent 7673A X Agilent 7683A X Agilent HS7694 Agilent 7695A Agilent 79855(A) X Agilent 5880 X Agilent 5890 X . Autosampler Compatibility Chart Crimp Neck ND8 1 1. Snap Ring ND11 Screw Neck ND13 Shell Vials Shell Vials Shell Vials Shell Vials Headspa

Agilent 1290 Infinity X X Agilent G1888A Agilent 7673A X X X Agilent 7683A X X X X Agilent HS7694 Agilent 7695A Agilent 79855(A) X X Agilent 5880 X X Agilent 5890 X X . Autosampler Compatibility Chart 2. H eadline as disp

Tip 1: How to use Agilent 82357B USB/GPIB converter in NI’s MAX or LabVIEW? Figure 2. Typical setup for Agilent IO Libraries Suite. 1 Agilent I/O Libraries Each Agilent IO product is bundled with the Agilent I/O libraries. There are four I/O libraries included in Agilent IO libraries Suite: Agilent

Digital Multi Meter No Substitute Agilent 3458A, Agilent 34401A, Agilent 34410A and Agilent 34411A 1 Dynamic Accuracy Test Kit No Substitute Agilent Z5623A Opt H01 1 Power Sensor No Substitute Agilent E9304A with Opt H18 Agilent 8482A 1 1 Calibration Kit No Substitute Agilent 85032F 1 Fixed

MSO-X 2004A MSO-X 2014A MSO-X 2024A 2-Channel DSO DSO-X 2002A DSO-X 2012A DSO-X 2022A 4-Channel DSO DSO-X 2004A DSO-X 2014A DSO-X 2024A. 4 Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide the screen on a different monitor. An MSO lets

DSO 4-Kanal DSO-X 2004A DSO-X 2014A DSO-X 2024A. 4 Agilent InfiniiVision 2000 X-Series Oszilloskope Benutzerhandbuch † Mixed-Signal Oszilloskop-Modelle (MSO) mit 2 8 Kanälen und 4 8 Kanälen. Ein MSO

300-a02 abp enterprise sdn bhd. 7th floor menara lien hee no, 8 jalan tangung, 47700 petaling jaya. selangor p. j john c.o.d. 03-7804448 03-7804444 300-c01 control manufacturing 400-2 (tingkat satu) batu 1/2, jalan pahang, 51000 kuala lumpur kl lal net 60 days 03-6632599 03-6632588