Installation Instruction for NP210 Gas Control Conversion Panel Tri-Tech Medical Inc. Converting a Medaes, Hill Rom or Ohmeda Gas Control Panel 37 Degree JIC Female Flare 3/8 Nom Risers, 45 Degree SAE Male Flare 5/16 Nom Risers or 3/8” Nom Tube Compression Figure 1 Removal of existing components Back Box in Wall Control Bracket Mounting Screws Front Panel Mounting Screws Gauge Covers Regulator Knob Switch Knob Back Box in Wall Figure 2 Installation of the Conversion Components Conversion Panel Supplied Mounting Hardware Gas Control Panel Mounting Screws Conversion Fitting Assembly Hardware . e7060rB 12/05/22 Tri-Tech Medical, Inc. 35401 Avon Commerce Parkway, Avon, Ohio 44011 Tel 80 0-253-8692 or 440-937-6244 Fax. 440-937-5060 Domestic e-mail: sales@tri-techmedical.com International e mail: internationalsales@tri-techmedical.com Web site address: www.tri-techmedical.com Page 1 of 2 No. 99-0236

Installation Instruction for NP210 Gas Control Conversion Panel Tri-Tech Medical Inc. Gas Control Medaes, Hill Rom or Ohmeda Back Box Disassembly: Close the area Gas supply. Warning: Do not close zone gas supply until all personnel have been advised of the intended service and all systems requiring medical gas are being supplied from portable or alternate supplies. Bleed the residual gas pressure from the outlet. Regulator Bracket Front Panel 1 2 Unscrew and remove the mounting screws from the front panel and the gauge covers, along with the regulator knob and switch knob so that the back box can be revealed. Remove the bracket mounting screws, dropping the bracket. Then loosen the incoming and the outgoing connections from the risers. Now remove the control bracket from the back box Tri-Tech Medical Conversion Installation Instruction: Inlet 3 4 5 Black Tube to the Inlet Riser Connection Fitting Outlet Mounting Screws Conversion Sub- Panel Select the proper brass riser adaptor assemblies that mates to the original supply riser and outlet riser. Tighten the swivel nuts securely. Yellow tube to the Outlet Riser Install the conversion sub-panel to the back box, using the #6 X 1” or #8 x 1” flat head screws lining up the appropriate holes. Screws heads should be flush with panel. If nitrogen is not being supplied downstream, remove yellow outlet tubing from the tee fitting on the front panel, and insert the supplied plastic plug. 6 While holding the Control Panel, push the Black tube firmly into the inlet supply line fitting and, if applicable the yellow tube firmly into the Outlet supply line fitting. Pull on the tubing to ensure it is locked in place. 7 Mounting Screws e7060rB 12/05/22 Secure the New Control Panel to the subpanel with the four short #6 x 5/16” pan head screws. Tighten mounting screws and test in accordance with NFPA Pamphlet 99. Conversion completed. Tri-Tech Medical, Inc. 35401 Avon Commerce Parkway, Avon, Ohio 44011 Tel 80 0-253-8692 or 440-937-6244 Fax. 440-937-5060 Domestic e-mail: sales@tri-techmedical.com International e mail: internationalsales@tri-techmedical.com Web site address: www.tri-techmedical.com Page 2 of 2 No. 99-0236

Installation Instruction for NP220 Gas Control Conversion Panel Tri-Tech Medical Inc. Converting a Allied/Chemetron Gas Control Panel 5/16 or 3/8 Tube Male Compression Fitting on Risers Figure 1 Removal of existing components Back Box in Wall Front Panel Mounting Screws Figure 2 Installation of the Conversion Components Back Box in Wall Conversion Panel Supplied Mounting Hardware Gas Control Panel Mounting Screws Conversion Fitting Assembly Hardware . e7060rB 12/05/22 Tri-Tech Medical, Inc. 35401 Avon Commerce Parkway, Avon, Ohio 44011 Tel 80 0-253-8692 or 440-937-6244 Fax. 440-937-5060 Domestic e-mail: sales@tri-techmedical.com International e mail: internationalsales@tri-techmedical.com Web site address: www.tri-techmedical.com Page 1 of 2 No. 99-0243

Installation Instruction for NP220 Gas Control Conversion Panel Tri-Tech Medical Inc. Gas Control Allied/Chemetron Back Box Disassembly: Close the area Gas supply. Warning: Do not close zone gas supply until all personnel have been advised of the intended service and all systems requiring medical gas are being supplied from portable or alternate supplies. Bleed the residual gas pressure from the outlet. Mounting Screws 1 Front Panel 2 Unscrew and remove the mounting screws from the front panel. Remove the front panel. Then loosen the incoming and the outgoing connections from the risers. Now remove the control panel from the back box. Inlet Tri-Tech Medical Conversion Installation Instruction: 3 4 5 Black tube to the Inlet Riser Outlet Mounting Screws Connection Fittings Conversion Sub- Panel Select the proper adaptor compression fitting assemblies that mates to the original supply riser and outlet riser. Tighten the compression nuts securely. Yellow tube to the Outlet Riser Install the conversion sub-panel to the back box, using the #6 x 1” or #8 x 1” flat head screws lining up the appropriate holes. Screw heads should be flush with sub-panel. If nitrogen is not being supplied downstream, remove yellow outlet tubing from the tee fitting on the front panel, and insert the supplies sub-plastic plug. 6 While holding the Control Panel, push the Black tube firmly into the inlet supply line fitting and, if applicable, the yellow tube firmly into the Outlet supply line fitting. Pull on the tubing to ensure it is locked in l 7 Mounting Screws e7060rB 12/05/22 Secure the New Control Panel to the subpanel with the four short #6 x 5/16” pan head screws. Tighten mounting screws and test in accordance with NFPA Pamphlet 99. Conversion completed. Tri-Tech Medical, Inc. 35401 Avon Commerce Parkway, Avon, Ohio 44011 Tel 80 0-253-8692 or 440-937-6244 Fax. 440-937-5060 Domestic e-mail: sales@tri-techmedical.com International e mail: internationalsales@tri-techmedical.com Web site address: www.tri-techmedical.com Page 2 of 2 No. 99-0243

Tri-Tech Medical Inc. Installation Instruction for NP230 Gas Control Conversion Panel Converting a Oxequip Gas Control Panel 1/4 OD Plastic Tube Connectors on Risers Figure 1 Removal of existing components Back Box in Wall Front Panel Mounting Screws Figure 2 Installation of the Conversion Components Back Box in Wall Conversion Panel Supplied Mounting Hardware Gas Control Panel Mounting Screws Tube Connectors Fittings . e7060rB 12/05/22 Tri-Tech Medical, Inc. 35401 Avon Commerce Parkway, Avon, Ohio 44011 Tel 80 0-253-8692 or 440-937-6244 Fax. 440-937-5060 Domestic e-mail: sales@tri-techmedical.com International e mail: internationalsales@tri-techmedical.com Web site address: www.tri-techmedical.com Page 1 of 2 No. 99-0246

Tri-Tech Medical Inc. Installation Instruction for NP230 Gas Control Conversion Panel Gas Control Oxequip Back Box Disassembly: Close the area Gas supply. Warning: Do not close zone gas supply until all personnel have been advised of the intended service and all systems requiring medical gas are being supplied from portable or alternate supplies. Bleed the residual gas pressure from the outlet. Front Panel Mounting Screws 1 2 Remove the front panel. Then loosen the incoming and the outgoing connections from the risers. Now remove the control panel from the back box. Unscrew and remove the mounting screws from the front panel. Tri-Tech Medical Conversion Installation Instruction: 3 Inlet 4 5 Black tube to the Inlet Riser Mounting Screws Connection Fittings Conversion Sub- Panel Change out the tube connection fittings, in the original supply riser and outlet riser. Tighten the connection fittings securely. 6 Mounting Screws e7060rB 12/05/22 Outlet Install the conversion sub-panel to the back box, using the #6 X 1” flat head screws lining up the appropriate holes. Screws heads should be flush with panel. If nitrogen is not being supplied downstream, remove yellow outlet tubing from the tee fitting on the front panel, and insert the supplied plastic plug. Yellow tube to the Outlet Riser 7 While holding the Control Panel, push the Black tube firmly into the inlet supply line fitting and, if applicable the yellow tube firmly into the Outlet supply line fitting. Pull on the tubing to ensure it is locked in place. Secure the New Control Panel to the subpanel with the four short #6 x 5/16” pan head screws. Tighten mounting screws and test in accordance with NFPA Pamphlet 99. Tri-Tech Medical, Inc. 35401 Avon Commerce Parkway, Avon, Ohio 44011 Tel 80 0-253-8692 or 440-937-6244 Fax. 440-937-5060 Domestic e-mail: sales@tri-techmedical.com International e mail: internationalsales@tri-techmedical.com Web site address: www.tri-techmedical.com Page 2 of 2 Conversion completed. No. 99-0246

Tri-Tech Medical Inc. Installation Instruction for NP240 Gas Control Conversion Panel Converting an Ohio Medical (Squire Cogswell) or Amico Gas Control Panel with 3/8” Female NPT Risers Figure 1 Removal of existing components Back Box in wall Front Panel Mounting Screws Figure 2 Installation of the Conversion Components Back Box in wall Conversion Panel Supplied Mounting Hardware Gas Control Panel Mounting Screws Tube Connectors Fittings e7060rB 12/05/22 Tri-Tech Medical, Inc. 35401 Avon Commerce Parkway, Avon, Ohio 44011 Tel 80 0-253-8692 or 440-937-6244 Fax. 440-937-5060 Web-site www.tri-techmedical.com Page 1 of 2 No. 99-0347

Installation Instruction for NP240 Gas Control Conversion Panel Tri-Tech Medical Inc. Gas Control Ohio Medical (Squire Cogswell) or Amico Back Box Disassembly: Close the area Gas supply. Warning: Do not close zone gas supply until all personnel have been advised of the intended service and all systems requiring medical gas are being supplied from portable or alternate supplies. Bleed the residual gas pressure from the outlet. 3/8” NPT Connection Fittings Mounting Screws 1 2 Unscrew and remove the mounting screws from the front control panel. 3 Remove the front panel. Then loosen the incoming and the outgoing 3/8” NPT connections from the risers. Clean the female 3/8 NPT threads on the risers. Inlet Tri-Tech Medical Conversion Installation Instruction: 4 5 6 Black tube to the Inlet Riser Mounting Screws Connection Fittings Yellow tube to the Outlet Riser Conversion Sub- Panel Apply Teflon tape or other suitable pipe thread sealing compound to the threads of the two brass adaptors assemblies supplied. Install the two brass adaptors that mate to the original nitrogen supply riser and nitrogen outlet riser. Tighten the pipe thread joint securely. 7 Outlet Install the sub-plate with the large cutout to the original back box using the #6 x 1” or #8 x 1” flat head screws in the appropriate holes. Screw heads should be flush with panel. If nitrogen is not being supplied downstream, remove the yellow outlet tubing from the tee fitting on the front panel, and insert the supplied plastic plug. While holding the new Nitrogen Control Panel, push the Black Tubing firmly into the Nitrogen supply line fitting and, if applicable, the yellow tubing firmly into the nitrogen outlet line fitting. Pull on the tubing to ensure it is locked in place. 8 Mounting Screws Conversion completed. Secure the New Control Panel to the subpanel with the four short #6 x 5/16” pan head screws. Tighten mounting screws and test in accordance with NFPA Pamphlet 99. e7060rB 12/05/22 Tri-Tech Medical, Inc. 35401 Avon Commerce Parkway, Avon, Ohio 44011 Tel 80 0-253-8692 or 440-937-6244 Fax. 440-937-5060 Web-site www.tri-techmedical.com Page 2 of 2 No. 99-0347

Tri-Tech Medical Inc. Installation Instruction for NP210 Gas Control Conversion Panel e7060rB 12/05/22 Tri-Tech Medical, Inc. 35401 Avon Commerce Parkway, Avon, Ohio 44011 Tel 80 -253-8692 or 440 -937-6244 Fax. 440-937-5060 Domestic e-mail: sales@tri-techmedical.com International e mail: internationalsales@tri- techmedical.com

Related Documents:

Bruksanvisning för bilstereo . Bruksanvisning for bilstereo . Instrukcja obsługi samochodowego odtwarzacza stereo . Operating Instructions for Car Stereo . 610-104 . SV . Bruksanvisning i original

Tri-Sprintec Tri-Vylibra Trinessa Generic For Ortho Tri-Cyclen Lo sold as: Norgest/Ethi Estradiol Tri-Lo-Estaryll Tri-Lo-Marzia Tri-Lo Mili Tri-Lo-Sprintec Tri-Vylibra Lo Generic Ovcon-35 sold as: Balziva Briellyn Philith Vyfemla Generic Quartette sold as: Fayosim Levonor/Ethi Estradiol Rivelsa Generic Safyral sold as:

Tri-Sprintec Tri-Vylibra Trinessa : Generic For Ortho Tri-Cyclen Lo sold as: Norgest/Ethi Estradio Tri-Lo-Estaryll Tri-Lo-Marzia Tri-Lo Mili Tri-Lo-Sprintec Tri-Vylibra Lo: Generic Ovcon-35 sold as: Balziva Briellyn Philith Vyfemla : Generic Ovral sold as: Ogestrel : Generic Quartette sold as: Fayosim

10 tips och tricks för att lyckas med ert sap-projekt 20 SAPSANYTT 2/2015 De flesta projektledare känner säkert till Cobb’s paradox. Martin Cobb verkade som CIO för sekretariatet för Treasury Board of Canada 1995 då han ställde frågan

service i Norge och Finland drivs inom ramen för ett enskilt företag (NRK. 1 och Yleisradio), fin ns det i Sverige tre: Ett för tv (Sveriges Television , SVT ), ett för radio (Sveriges Radio , SR ) och ett för utbildnings program (Sveriges Utbildningsradio, UR, vilket till följd av sin begränsade storlek inte återfinns bland de 25 största

Hotell För hotell anges de tre klasserna A/B, C och D. Det betyder att den "normala" standarden C är acceptabel men att motiven för en högre standard är starka. Ljudklass C motsvarar de tidigare normkraven för hotell, ljudklass A/B motsvarar kraven för moderna hotell med hög standard och ljudklass D kan användas vid

LÄS NOGGRANT FÖLJANDE VILLKOR FÖR APPLE DEVELOPER PROGRAM LICENCE . Apple Developer Program License Agreement Syfte Du vill använda Apple-mjukvara (enligt definitionen nedan) för att utveckla en eller flera Applikationer (enligt definitionen nedan) för Apple-märkta produkter. . Applikationer som utvecklas för iOS-produkter, Apple .

kesehatan mata menggunakan media permainan ular tangga. Tindakan preventif yang dilakukan adalah dalam bentuk pemeriksaan kesehatan mata pada masyarakat. Kegiatan diharapkan dapat mencangkup masyarakat lebih luas. Oleh karena itu, even Car Free Day dipilih sebagai lokasi kegiatan. Gambar 1. Lokasi Pendidikan dan Penyuluhan Kesehatan Mata di Car Free Day Cilembang . April 2020 [JURNAL .