V5 N2 Garrett Air Removal From The Sterilization Chamber

2y ago
5 Views
3 Downloads
2.61 MB
8 Pages
Last View : 1m ago
Last Download : 3m ago
Upload by : Gannon Casey
Transcription

TMVolume 5, Number 2March 2008Air Removal from the Sterilization ChamberGarrett KrushefskiScientific & TechnicalServices ManagerI recently received a question from a customer asking why we state in the EZTestinstructions for use to “Place the EZTest biological indicators in a horizontal position ”When sterilizing by steam, it is imperative to first remove ambient air not only from thechamber but also from the devices being sterilized. Failure to do so will impede steamcontact with all surfaces which can result in non-sterile items being used from aseemingly acceptable sterilization cycle.There are two general means for removing ambient air from the sterilization chamber:dynamic air removal or gravity-displacement. Dynamic-air-removal sterilization cyclescan be either a prevacuum cycle, in which case air is actively removed by a series ofpressure and vacuum phases prior to admission of steam, or by steam-flush-pressurepulsing (SFPP) in which case a series of steam flushes and pulses of pressure areemployed. Gravity-displacement is a passive means of air removal in which incomingsteam displaces or “pushes” residual air out through the drain of the chamber.ANSI/AAMI ST79:2006 notes, “The dynamic-air-removal cycle is generally preferred toa gravity-displacement cycle because of more efficient air removal, a shorter exposuretime at higher temperatures, and a vacuum drying phase, resulting in an overall reductionin cycle time.”While the above ANSI/AAMI note is accurate, do not think that you must start upgradingand replacing your gravity-displacement sterilizers with new machinery. Gravitydisplacement is still an effective means of sterilization as long as one employs a fewprecautions. Such precautions are not necessary in a dynamic-air-removal processbecause the air is actively “sucked-out” of the chamber and load.Without the aid of active air removal, one needs to “facilitate” removal of air from theload when using gravity-displacement systems. This is achieved by positioning items insuch a manner that does not create any potential air traps. To illustrate this point, startMesa Laboratories, Inc.www.MesaLabs.com

with a simple example such as a standard laboratory beaker. Of the three orientationsshown in Figure 1, which are appropriate for a gravity-displacement steam sterilizer?Figure 1ACBNOTE: In all of the photos non-packaged items are shown for visual clarity. In actual practicethese items would be in sterilization pouches, allowing for post-exposure protection fromcontamination.The answer is (B & C). The ambient air that is contained within the beaker is heavier(i.e. more dense) than the steam that will be introduced into the chamber. In the uprightorientation (A), the steam will not be able to displace or push the ambient air out of thebeaker because the air has no means of escape (i.e. the beaker does not have a “drain”).Note that the orientation shown for (B) does allow the air to drain from the beaker.The next example is an Erlenmeyer flask which is a bit trickier. Review the fourorientations shown in Figure 2 and decide which are the most appropriate for a gravitydisplacement cycle.Mesa Laboratories, Inc.www.MesaLabs.com

Figure 2DGEFIt is easy to see that (F) or (G) is the best option and (D) must be avoided. But whatabout (E)? This orientation is also problematic because there is a zone within theErlenmeyer flask where ambient air will become trapped and steam contact to thesurfaces in that zone will be impeded (see outline of dashed red zone in Figure 3). Avery effective way to demonstrate the effects of trapped air is to run a standardsterilization cycle with three flasks oriented as shown in D, E and F (see Figure 3 below).Place an EZTest biological indicator at each location marked in the photo and incubatethe units after exposure. The EZTest from locations 1 and 2 will show surviving sporeswhile the EZTest from location 3 will have no growth. This is a perfect example to showthat the parametric data for the cycle (time, temperature, and pressure readings) willindicate acceptable cycle performance but the biological indicator indicates non-sterility.Mesa Laboratories, Inc.www.MesaLabs.com

Figure 3DEF312When conducting a validation several years ago, I performed the above recommendedtest in a 65-minute exposure. The penetration thermocouples (with attached EZTest BIs)recorded accumulated Fo values in excess of 80 minutes even from locations 1 and 2.The physical data indicated a lethal delivery that was well in excess of what was neededto kill the spore challenge, yet the EZTest from locations 1 and 2 were positive whilelocation 3 was negative. How could the spores survive 80 Fo? The reason is theprevailing condition at locations 1 and 2 was NOT saturated steam; rather, it was 121 123 C dry heat. Without an escape route, the less-dense steam molecules could notdisplace the heavier ambient air and air pockets (dashed red zones in Figure 3) becometrapped in the flasks. The penetration thermocouples recorded temperatures of 121 to123 C for most of the 65-minute dwell time, but unlike the spores in the biologicalindicator, the thermocouples can not differentiate between dry heat conditions andsaturated steam. When reviewing the cycle data, time, temperature and pressure were allwithin acceptable levels, thus indicating acceptable cycle performance, yet the sporessurvived. The biological results indicated non-sterility and did not agree with thephysical data which indicated sterilization was achieved. We refer to this condition as acatastrophic sterilization failure. SPORES DON’T LIE !Air pockets within the materials being sterilized can not be detected by anyinstrumentation available today. The only system available to detect these pockets is thebiological indicator. Air pockets occur in the “worst case” to sterilize locations in theMesa Laboratories, Inc.www.MesaLabs.com

product or load. All standards direct the user to challenge the “worst case” locations withbiological indicators.As a final example, consider the lengths of tubing shown in Figure 4. When oriented asshown in (G) multiple air pockets will form at various low spots along the length of thetube. One must carefully coil the tube in a helix shape (H) to facilitate air removal andsteam penetration.Figure 4HGUsing colored water, one can illustrate the multiple low spots that exist in (G) where airpockets will form. For Figure 5, both tubes were filled with purple water and placed onthe perforated stainless tray. All of the colored water drained from (H). Compare that tothe amount and location of water that was unable to escape the (G) configuration. Thecolored water allows one to visually discern where the trapped air will form pockets andthus, the areas in the tube where steam will not penetrate.Mesa Laboratories, Inc.www.MesaLabs.com

Figure 5HGThe water visualization method is a valuable tool that can be used when trainingpersonnel to use steam sterilizers. One must position items in the chamber such that ifthey were filled with water, the water would completely drain from the item. If any waterwere retained, the positioning is not suitable (as shown for the Erlenmeyer flask in Figure6).Mesa Laboratories, Inc.www.MesaLabs.com

Figure 6The colored water is unable to drain from the flask and shows wheretrapped air will collect and impede steam contact.Now back to the original question, “Why is it necessary to place the EZTest biologicalindicators in a horizontal position?” The EZTest, by virtue of its design is an excellentdetector of trapped air. The EZTest itself is not unlike the Erlenmeyer flask in that it toocan harbor ambient air which would impede steam contact with the spores on theinoculated paper strip. Thus, just like the beaker shown in (B), this is the correctorientation for the EZTest BI.Is it necessary to use the horizontal orientation in a dynamic-air-removal sterilizationcycle? In dynamic-air-removal cycles, one is relying on the prevacuum or SFPP toremove air from the chamber, the load and the EZTest (which may or may not be in a testpack). Regardless, it is always a good idea to use the horizontal orientation for yourEZTest as it does help avoid inadvertent trapping of air, even in dynamic-air-removalsystems.Mesa Laboratories, Inc.www.MesaLabs.com

Finally, consider Figure 7; can you see the flaw in (I) and why (J) is the preferred set-up?JIFigure 7Recall the earlier statement, “the EZTest itself is not unlike the Erlenmeyer flask in that ittoo can harbor ambient air which would impede steam contact with the spores on theinoculated paper strip.” In both (I & J) above, the Erlenmeyer flask is inverted to theextent that if filled with water, all of the water would drain from the flask. HOWEVER,in (I) the EZTest is in a slight “cap-up” orientation and if the EZTest biological indicatorwere filled with water, it would be trapped inside the EZTest. In (J) a slight “cap-down”orientation prevails (by virtue of the inverted positioning of the Erlenmeyer flask) andthus we have a “drain” which allows the ambient air from inside the EZTest to escape theunit.SPORES DON’T LIE Please email us with topics you would like to see addressed in “Spore News”.Mesa Laboratories, Inc.www.MesaLabs.com

Dynamic-air-removal sterilization cycles can be either a prevacuum cycle, in which case air is actively removed by a series of pressure and vacuum phases prior to admission of steam, or by steam-flush-pressure pulsing (SFPP) in which case a series of st

Related Documents:

AMD Rome CPU based yx5x servers Hot insertion Orderly removal Surprise removal Hot insertion Orderly removal Surprise removal Orderly removal Surprise removal Note: Linux upstream kernel v

spark plug removal 14.52 [.572] spark plug boot removal 10.09 [.397] spark plug boot fuel filter fuel pump 35.31 [1.390] spark plug removal 142.89 [5.626] 92.10 [3.626] 20 mounting hole "a" cl 130.00 [5.118] air filter cover assembly removal 72.81 [2.867] spark plug removal 52.02 [2.048] spark plug boot removal

Precision Air 2355 air cart with Precision Disk 500 drill. Precision Air 2355 air cart with row crop tires attached to Nutri-Tiller 955. Precision Air 3555 air cart. Precision Air 4765 air cart. Precision Air 4585 air cart. Precision Air 4955 cart. THE LINEUP OF PRECISION AIR 5 SERIES AIR CARTS INCLUDES: Seven models with tank sizes ranging from

Getting Over Garrett Delaney. Abby McDonald. Scholastic. Sadie has been in love with Garrett Delaney for far too long. She finally realizes that no matter how ‘perfect’ she is for him, Garrett will never see her as more than a friend. Sadie picks up the pieces of her

Diocese of Evansville St Joseph School Princeton . Diocese of Fort Wayne Saint Joseph High School South Bend Diocese of Fort Wayne Saint Joseph School South Bend . Garrett-Keyser-Butler Com Schools Garrett High School Garrett Greater Clark County Schools Utica Elementary School Jeffersonville

Garrett Retirees’ Club of Arizona Newsletter December 2015 Fix-It Guys Sponsored by Honeywell, Garrett and Sperry Retirees Clubs We can help Honeywell retirees or surviving spouses with fix-it jobs around the house that you are not able to accomplish. This will hopefully keep the job from being a budget buster for you.

2:00 pm in Garrett 86 presented by R.K. Hoddinott III. The Current Status of U.S. Economy and Jobs in the Financial Sector. 2:00 pm in Garrett 11 presented by Daryl Ferrara. Achieving Success in a Professional Selling Career. 2:00 pm and 3:30 pm in Garrett 8 presented by Caleb Puig. How a S

made to Valdosta State University including to the Langdale College of Business Administration and several athletic programs. Garrett Copeland - Garrett Copeland was born and raised in Valdosta and is a graduate of Lowndes High School and Valdosta State University. Garrett is the