Two Stage Rotary Vane Vacuum Pump Operation And Maintenance Manual

1y ago
8 Views
1 Downloads
519.24 KB
15 Pages
Last View : 9d ago
Last Download : 3m ago
Upload by : Annika Witter
Transcription

Two Stage Rotary Vane Vacuum Pump Operation and Maintenance Manual Features Applications Extremely Quiet Virtually Vibration Free Two Year Warranty Anti Suck Back Design Corrosion Resistant Motors for Any Voltage Satisfaction Guaranteed Refrigeration and A/C Service Vacuum Ovens and Furnaces Electron Microscopes Freeze Dryers Cluster Tool Load Locks Sputter and Deposition Systems Part no. 795022

Contents FORWARD.1 WARNINGS.2 PUMP SPECIFICATIONS AND PUMP SPEED CURVES.3 DIMENSIONS.3 PARTS LIST.PAGES 4-5 EXPLODED VIEW DRAWING.6 PREPARING AND INSTALLING THE PUMP.8 RECEIVING.8 REPORTING SHIPPING SHORTAGE.8 REPORTING INCORRECT SHIPMENT.8 STARTING AND OPERATING THE PUMP.9 START UP.9 OPERATION.9 ANTISUCKBACK.10 GAS BALLAST.10 MAINTAINING THE PUMP.11 PREVENTATIVE MAINTENANCE.11 TROUBLESHOOTING.12-13 May 2008

Forward This manual contains installation, operation, maintenance, and troubleshooting information for the Model 140-5 RS Rotary Vane Vacuum Pump. Please read it in its entirety before operating the pump. Our Rotary Vane Vacuum Pumps are designed to ensure safety when used properly. It is the responsibility of the user to follow safety-related warnings, cautions, notes, and other requirements described in this manual. Returned equipment will not be accepted by our company without prior authorization. Prior to shipping please call for a returned goods authorization number (RGA). Our company reserves the right to cancel the warranty if the pump is disassembled without authorization, if pump fluids are used that are not compatible with the design and materials used in the manufacture of the pump, and if unauthorized spare parts are used. 1

WARNINGS WARNING Death or serious injury can result from improper use or application of this pump. If the pump will be exposed to toxic, explosive, pyrophoric, highly corrosive, or other hazardous process gases including greater than atmospheric concentrations of oxygen, contact our company for specific recommendations. WARNING Ground the motor properly during installation. Disconnect the power before beginning installation, maintenance or repair work or before interchanging the input leads when correcting the direction of rotation. Disconnecting the power also avoids an unexpected start-up for pumps with automatically resetting thermal overloads. WARNING Don’t run the pump without an exhaust line and an adequate exhaust system if hazardous gases or vapors are expelled from the pump. Don’t exceed a maximum backpressure of 7 psig. Excessive pressure in the pump could damage the seals, blow out the sight glass, or rupture the pump housing. In addition, excessive backpressure can result in hazardous process gas or contaminated oil leaking out of the pump. Don’t install an exhaust line with a smaller ID than the exhaust port or allow restrictions or deposit build up in the exhaust line. If you are purging the oil casing with inert gas, limit the inert gas flow to avoid exceeding the 7 psig limit. Accidentally connecting the pump’s exhaust port to a vacuum line containing a closed valve also causes a dangerous excessive pressure. WARNING Hazardous process gases can concentrate in the vacuum pump, its oil, and its filters. If the pump has been used on toxic, explosive, pyrophoric, corrosive, volatile, or other hazardous substances, take the proper safety precautions before opening the pump or filters. Proper precautions could include inert gas purging before and after you drain the oil to sweep hazardous gas from the pump or filters; wear gloves or protective clothing to avoid skin contact with toxic or highly corrosive substances; specially ventilated work areas; fume hoods; safety masks; breating apparatus; etc. 2

Specifications: Model 140-5 RS Pump Speed Curve Pumping Speed 10 @ 60 Hz.140 l/min, 5 CFM, 8.5 m3/hr. Ultimate Pressure 10 Gas Ballast Closed.5x10-3 Torr, .007 mbar, 5 Microns 10 Gas Ballast Open.5x10-2 Torr, .07 mbar, 50 Microns Standard Motor.115V, 1ø, 60 Hz, 1/2 hp Includes On/Off Switch and Circuit Breaker 10 Optional Motor.230V, 1ø, 60 Hz, 1/2 hp Power Connection.6 ft. (2 m) 3 conductor cord Motor speed @ 60 Hz.1700 r.p.m. Oil capacity.500 cc; 0.53 Qt. Weight Net: 40 lbs., 18 kg.Shipping: 45 lbs., 21 kg Intake & exhaust type/diameter.Tee with 2 of 3/8 and 1 of 1/4 in. fittings Exhaust.Through Handle Ambient operating temperature.7 to 40 C, 45 to 104 F PUMPING SPEED (L/min) -3 Model 140-5 RS 5 CFM -2 -4 Model 140-5 RS WITHOUT GAS BALLAST WITH GAS BALLAST 10-3 397 (15.63) 114 (4.48) 70 (2.74) 397 (8.39) 52 2.05 Carry Handle 160 (6.30) 137 (5.39) OIL 140 (5.51) 102 (4.02) 114 (4.47) 35 47 (1.38) (1.85) 70 (2.74) 183 30 (7.20) (1.18) 183 (7.20) 30 (1.18) 130 (5.12) 41 (1.61) 140 (5.51) 168 (6.61) 27 (1.06) 129 (5.06) 396 202 (7.97) (15.59) 3 65 (2.56) 10-2 10-1 1 PRESSURE IN TORR Model 140-5 RS 26 (1.02) 5 CFM 10 100 Atmos.

BOM OF SV-140 No. CAD No. CODE No. PART NAME QTY Motor Assembly (1 phase) 1 BMT1A11B Motor 1 SV14002 BWG20201 Leg Rubber 1 4 SV14003 SD05020B Leg Rubber 1 5 SV14004 SD06020A Motor Plate Fixing Bolt 4 6 SV14005 SGW20501 Motor Plate 1 7 SV14006 BWG00601 Motor Plate 2 8 SV14007 EG005A Washer 4 9 SV14008 DG005A Motor Plate Fixing Nut 4 10 SV14009 SD05010C Coupling Fixing olt 1 11 SV14010 BWG12301 Fan 1 12 SV14011 BWD12201 Coupling Joint 1 13 SV14012 SD06012C Coupling Fixing Bolt 2 14 SV14013 BWG21301 Coupling 1 15 SV14014 BWD21401 Leg Rubber (FRONT) 2 SG05010B Leg Rubber (FRONT) Fixing Bolt 2 1 AMT1A11B 2 SV14001 3 16 17 SV14015 BAC0P006 O-Ring 1 18 SV14016 BWG21601 Main Plate (BODY) 1 19 SV14017 BADAN160 O-Ring 1 20 SV14018 BWG21801 Bush 1 21 SV14019 1ST Guide Pin 2 22 SV14020 BWG22001 1ST Body Separator 1 23 SV14021 BWG22101 1ST Stage Stator 1 24 SV14022 BWG22201 Shaft 1 25 SV14023 BWG22301 1ST Rotor Assembly 1 26 SV14024 BWG22401 Bush 1 27 SV14025 BWG22501 1ST, 2ND Separator 1 28 SV14026 2ND Guide Pin 2 29 SV14027 BWG22701 Spacer 1 30 SV14028 BWG22801 2ND Rotor Assembly 1 31 SV14029 BWG22901 2ND Stage Stator 1 32 SV14030 BWG23001 End Plate 1 33 SV14031 Oil Lever Window Fixing Nut 1 34 SV14032 SD06020A Oil Box Assembly fixing Bolt 4 35 SV14033 BWG23301 Serial No. 1 36 SV14034 BWG23401 Oil Box Assembly 1 37 SV14035 BWG23501 Name Plate 1 38 SV14036 BAC0P005 O-Ring 1 4

BOM OF SV-140 No. CAD No. CODE No. PART NAME QTY 39 SV14037 BWG23701 Oil Drain Plug 1 40 SV14038 BWG23801 Oil Lever Window 1 41 SV14039 BAC0P044 O-Ring 1 Oil Lever Window Sticker 1 42 43 SV14040 BAD30026 O-Ring 1 44 SV14041 BWD06201 Oil Filling Plug 1 45 SV14042 SD06090A Stage Fixing Bolt 4 46 SV14043 EG006B Spring Washer 4 47 SV14044 BWG24401 2nd Stage Blade 2 48 SV14045 BWG24501 2nd Stage Valve 1 49 SV14046 50 SV14047 BWG03508 2nd Stage Valve Spring 1 51 SV14048 EG005A Washer 1 52 SV14049 SD05006A 2nd Stage Valve Fixing Bolt 1 53 SV14050 Tapper Pin 1 54 SV14051 BWG25101 1st Stage Blade 2 55 SV14052 BWG25201 1st Stage Valve 1 56 SV14053 BWG03508 1st Stage Valve Spring 1 57 SV14054 EG005A Plate Washer 1 58 SV14055 SD05006A 1st Stage Valve Fixing Bolt 1 59 SV14056 SD06008A Oil Baffle Fixing Bolt 1 60 SV14057 BWG25701 Oil Baffle 1 61 SV14058 BACAN119 O-ring 62 SV14059 BWG02103 Oil Seal (Rubber) 1 63 SV14060 BWE26001 Check Valve 1 64 SV14061 BWG03509 Valve Spring 1 65 SV14064 BWG26401 Intake 1 66 SV14067 BAC0P005 O-Ring 1 67 SV14068 BWG23701 Gas Ballast Knob 1 68 SV14070 BWG27001 Exhaust & Handle 1 69 SV14071 BWG00201 Key 1 70 SV14075 SD04006B Terminal Fixing Bolt 1 71 SV14076 BWG27601 Switch 1 72 SV14077 BWG27701 Rubber Cap 4 73 SV14078 BWG15001 Earth Pin 1 74 SV14081 BWF14102 Power Cord 1 75 SV14082 BWD14002 Powet Cord Protector 1 BWG14705 Cotton Box 1 76 2nd Stage Valve Space 5

6

PREPARING AND INSTALLING THE PUMP RECEIVING Each Rotary Vane Vacuum Pump is inspected and carefully packed prior to shipment. Inspect it after carefully unpacking it. In case of external damage, retain the shipping container and notify the shipping agency and our company immediately. Because the packing materials are designed specifically for this pump, they should always be used when transporting the pump. Unpack the pump and check for shipping damage as follows: 1. Inspect the outside of the shipping container for shipping damage. If you will be making a damage claim, save the shipping container and packing materials. 2. Unpack the pump. 3. Carefully inspect the pump for damage. 4. If you find any damage, proceed as follows: a. Save the shipping container, packing material, and parts for inspection. b. Notify the carrier that made the delivery within 7 days of delivery. c. File a claim with the carrier. d. Contact our company to make arrangements for replacing the damaged part(s). REPORTING SHIPPING SHORTAGE If you did not receive all the goods that you ordered, do the following: 1. Check the number of items listed on the packing slip. If the number of pieces listed is greater than the number of shipping containers received, contact the carrier concerning the missing piece. 2. Check the packing list to see if the missing item is on back order. 3. Carefully check the packing material and container to ensure that you did not overlook the missing item. 4. If you cannot find the item, please notify our company immediately. REPORTING INCORRECT SHIPMENT If the item received is not the item ordered, contact our company immediately. 8

STARTING AND OPERATING THE PUMP START UP Before starting the pump, please complete the following checklist: 1. Be sure that the pump is filled with the appropriate amount of vacuum fluid. 2. Be sure that all electrical connections have been properly wired and that there are no bare wires that could cause an electrical shock or fire. 3. Be sure that all system connections have been secured with the appropriate seal rings and clamps. 4. Charge the pump with fluid as follows: a. Reinstall the fluid-drain plug with flat gasket into the fluid-drain port. b. Remove the fluid-fill cap and fill the pump to capacity with VPO-3000 vacuum pump fluid. Using other than VPO-3000 vacuum pump fluid may result in damage to the pump or compromise the pump performance and lifetime. c. Reinstall the fluid-fill cap with flat gasket. Long Term Storage (2 weeks or longer) Before placing a pump in long term storage, follow the procedure below: 1. Drain all fluids from the pump as described in the previous section. 2. Refill the pump with clean VPO-3000 vacuum fluid as described in section for changing the pump fluid. 3. Be sure that the pump is stored in a horizontal position with the intake and exhaust ports facing up. 5. When putting a pump into storage, put a pin hole in both the intake and exhaust port caps. Avoiding Oil Leaks During Shipping and Storage Always drain your vacuum pump of all fluids before shipping. Failure to do so can result in damaged shipping containers and delays by freight carriers due to possibility of the presence of hazardous materials in the event of a spill. 9

OPERATION 1. Periodically check the vacuum fluid level in the sight glass to be sure it is between the low and high levels. If you are operating the pump with the gas ballast open, it will be necessary to check the oil level more frequently. 2. If the vacuum fluid within the pump becomes discolored or contaminated, change the fluid as soon as possible. Operating the pump with contaminated or dirty oil will greatly reduce the life expectancy of the pump and may lead to the cancellation of the warranty. ANTI SUCK BACK If the pump stops with the inlet under vacuum the anti suck back system will stop air or oil leakage inside the module or into the vacuum chamber. The vacuum integrity is guaranteed by: Quality of machining from surfaces between the functional elements (stator, plates, housing, etc.) The exhaust valves on the exhaust orifice. A spring valve automatically closes the oil injection canal in the pump. When the pump stops, the oil pump exhaust pressure is decreased and a spring activated valve closes the oil injection canal. GAS BALLAST When condensable vapors (such as water vapor) are being pumped the gas is compressed beyond its saturated pressure and can condense, impairing pump performance. The vapor pressure of water at typical pump temperatures is over 100 Torr. Even small amounts of water in the pump fluid will have a big effect on pump performance. The gas ballast control button allows a quantity of air to be injected into the second stage of the pump during “compression” to reduce the partial pressure of the pumped gas below its saturated vapor pressure and thus prevent condensation. At the end of “compression” the pressure in the discharge chamber is greater than atmospheric. The antisuckback feature described above prevents gas and oil from being discharged into the environment. The saturated vapor pressure of pump fluid and the condensed vapors such as water is higher when it is hot than when it is cold; therefore it is necessary to wait until the pump reaches its operating temperature before pumping condensable vapor. Using the gas ballast increases the ultimate pressure of the pump as well as its operating temperature. 10

MAINTAINING THE PUMP PREVENTIVE MAINTENANCE Pump Fluid Every vacuum pump is designed to work best with a specific pump fluid and the fluid is an active part of the pumping mechanism. For best performance from your pump, care must be used to select fluid with the physical and chemical properties engineered for your pump. For our pumps the ideal fluid for general purpose pumping is VPO-3000. This is a moderately priced fluid that is engineered to give best vacuum and longest life in our pumps. Other fluids may give performance that is good enough for your needs but our specifications are based on regular use of VPO-3000. VPO-3000 Typical Data ASTM Test Method Property Density, lbs. / USG 60 F Color, ASTM Flash Point F Viscosity cSt @ 40 C cSt @ 100 C Viscosity SUV @ 100 C SUV @ 210 C Viscosity Index Pour Point, F Emulsion Characteristics Distilled Water @ 130 F ML of water separated (minutes) Aniline Point, F UV Absorptivity (250 mm) D1298 D1500 D92 7.22 0.5 421 D445 46.6 6.97 D445/D2161 D2270 D97 240 50 106 0 D1401 40 (5) D611 D2008.A4 228 0.1 Checking the pump fluid level: NOTE: When the pump is not running, the pump fluid level appears lower. 11 VPO-3000

TROUBLE SHOOTING Symptom Noisy motor, will not turn. Noisy and hot pump. Vacuum drops (gets worse). Motor runs, but not pump. Check Steps to Take Power line voltage and connections Correct voltage or connections. Any foreign materials inside the pump. Remove the foreign materials. If problem is with oil, change oil. Motor (open internal circuit). Any foreign materials inside the exhaust valve. Replace open windings. Remove foreign materials. Leakage in the system. Fix the leakage. If leakage, valve is open. Close the valve. All the valves in the vacuum line. Oil level. Close if found open. Leakage on the device connected. Close the intake and recheck. Moisture content of oil. Change oil. Oil regulator. Replace. If gas ballast is open. Worn out coupling mechanism. Close the gas ballast. Replace coupling. Worn out key between motor and pump. Replace the key and the set screws. 12 Add oil.

CAUTION: Hydrocarbon pump fluid should be changed at the following times: 1. After a 100 hour break-in period of pump operation. 2. When the pump fluid becomes contaminated. 3. When condensation in the pump fluid is present. 4. Before and after the pump has been stored for a long period of time. NOTE: Always change the pump fluid while the pump is warm to prevent condensables, such as water, from remaining in the pump. Turn the pump off and change the fluid as follows: 1. Drain the fluid from the pump. Use your fingers to remove the oil fill cap and the oil drain plug from the pump; allow the fluid to drain into a suitable container. If the fluid fill cap or fluid drain cap cannot be loosened with your fingers, cover them with a cloth and use pliers. 2. After the oil flow dimininshes, switch ON the pump, allow it to run for about 10 seconds and then switch it OFF. 3. If the fluid drained from the pump is discolored, contains particulate, has a foul odor or is very dirty, flush out the pump using the procedure below until the drained fluid is clean. If your pump requires more than 2 flushes, a foreline trap or oil filtration unit should be installed on the pump. Contact our company for more information on foreline traps and oil filtration units. a. Reinstall the fluid-drain plug with flat gasket into the fluid-drain port. b. Refill the pump with VPO-3000 vacuum pump fluid until the fluid level is visible in the lower rim of the fluid sight glass. c. Reinstall the fluid-fill plug with gasket in the fluid fill port. d. Turn ON the pump and allow it to run for about 10 minutes. e. Turn the pump OFF and refer to step 1 to drain the vacuum fluid. 13

Vacuum Research Corp. 2419 Smallman Street Pittsburgh, PA 15222 USA www.vacuumresearch.com Phone: 800-426-9340 412-261-7630 FAX: 412-261-7220 e-mail: VRC@vacuumresearch.com

b. Remove the fluid-fill cap and fill the pump to capacity with VPO-3000 vacuum pump fluid. Using other than VPO-3000 vacuum pump fluid may result in damage to the pump or compromise the pump performance and lifetime. c. Reinstall the fluid-fill cap with flat gasket. Long Term Storage (2 weeks or longer) Before placing a pump in long term .

Related Documents:

5 Fisher Scientifi c Vacuum Pumps Two Stage Rotary Vane Vacuum Pumps Type Cat. No. Cat. No. Cat. No. Cat. No. Cat. No. Rotary Vane Vacuum Pump 100-120 V/200-240 V 50/60 Hz FB654591 FB654601 FB654611 FB654641 Rotary Vane Vacuum Pump 230 V 50/60 Hz FB65458² Rotary Vane Vacuum Pump 115 V 50/60 Hz FB65458A² Maintenance kit 302012-1 302075 302076 302077 302079 .

4 RV3, RV5, RV8 and RV12 Rotary Vane Pumps 1.3 Description The Edwards RV rotary vane pump is shown in Figure 1. Refer to Figure 1 for item numbers in brackets in the following descriptions. Th e RV pumps are two-stage, oil-sealed, sliding-vane vacuum pumps. The pump has NW25 inlet (4) and outlet (7) ports, a gas-ballast control (5)

The vacuum pumps in the FRVP series are oil-sealed, two-stage rotary vane vacuum pumps. The pump shaft and motor shaft are connected to one another by an elastic coupling. The bearing points of the interior pump body are force-fed lubricated slide bearings. Rotary knob for setting 1 Screw plug for oil filling 2 the gas ballast valve 3 On/off switch

The following types of vacuum pumps are available on the market for chamber machines, thermoforming packaging machines and tray sealers: Rotary vane vacuum pumps Oil-lubricated rotary vane vacuum pumps are the state-of-the-art solution for this application and the most widely used vacuum generators (market share 99%). The benefits:

Roots Vacuum Pumps leybold Oerlikon Leybold Vacuum Full Line Catalog 2015/2016 285 Central Vacuum Supply Systems Large Roots vacuum pumps, usually in connection with single-stage rotary vane vacuum pumps serve several consumers of vacuum (packaging machines, for example) at the same time. The pressure can be controlled in a

rotary vane pumps have final pressures about 10-2 Torr range while two-step medium vacuum vane pumps reach 10-3 Torr. Rate of pumping lies between 1-650 cfm, depending on whether the pump is a coarse vane or rough vane pump. While hybrid pumps were constructed to accept the poor fore line pressure diaphragm pumps produce.

ROTARY VANE COMPRESSORS/VACUUM PUMPS FOR VACUUM TANKERS Wittig RFL. The Wittig Advantage . 4-way valve, and final filter . COMPRESSOR / VACUUM PUMP RFL 60 DV RFL 80 DV RFL 100 DV OIL CONSUMPTION fl oz/h 4.2 5.1 5.1 MASS MOMENT OF INERTIA lb/ft² 4.44 5.71 7

compressor choice for continuous duty requirements such as auto body, paint, and plant air. Rotary compressors are typically available in two styles, rotary screw or rotary vane. Rotary screw compressors are often the most commonly appli